asr-3000 service manual
DESCRIPTION
hvacTRANSCRIPT
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
Contents
Safety information 1.Overview------------------------------------------------------------ ------------------------3 2.Electrical-------------------------------------------------------------------------------------3 3. Mechanical----------------------------------------------------------------------------------3 4. Safety Warnings and Cautions-----------------------------------------------------------3 5.Radiation and Magnetically Intensity Warnings----------------------------------------4 6.Other Warnings -----------------------------------------------------------------------------4 7. Moving and Using Equipment Warnings-----------------------------------------------5 8. Cleaning Cautions--------------------------------------------------------------------------5 9. Radiation Safety----------------------------------------------------------------------------5 10. Safety Symbols---------------------------------------------------------------------------5
Chapter1:
Theory and Structure of the System 1.1 General -------------------------------------------------------------------------------------8 1.2 ASR-3000X-Ray Mammography System basic structure---------------------------8
1.2.1 ASR-3000 X-Ray Mammography System basic Constitute -------------------8 1.2.2 Major Shelf ---------------------------------------------------------------------------8 1.2.3 Console -------------------------------------------------------------------------------8
1.3.Working System Principle -----------------------------------------------------------------9 1.4 The control theory of the system graph---------------------------------------------------9
Chapter 2:
Maintenance Schedules 2.1 In-Service Observation----------------------------------------------------------------10 2.2 Preventive Maintenance Schedules and Checklists--------------------------------10 2.3 System Component Preventive Maintenance---------------------------------------12 2.4 Major Shelf Preventive Maintenance------------------------------------------------12
Chapter 3:
Diagnostics and Troubleshooting 3.1 The Troubleshooting graph ------------------------------------------------------------13 3.2 Troubleshooting and Diagnostics------------------------------------------------------14 3.3 Sample s-----------------------------------------------------------------------------------15
REV1.0 1
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
Chapter 4:
Service Tools 1. Common Service Tools -------------------------------------------------------------------17
Appendix I:
The module functions and system structure 1. ASR-3000 mammography structure and composition---------------------------18 2. Functions of each module------------------------------------------------------------18 3. Signal information flow and parameters of each module------------------------19 3.1 Control console display panel-----------------------------------------------------19 3.2 Machine head display panel-------------------------------------------------------20 3.3 Power supply control board-------------------------------------------------------21 3.4 Anode driver board-----------------------------------------------------------------22 3.5 Control box intermediate board---------------------------------------------------22 3.6 Control box lower board-----------------------------------------------------------23 3.7 High-voltage driver board---------------------------------------------------------24 3.8 Main unit CPU board--------------------------------------------------------------25 3.9 Main unit output power supply test parameter---------------------------------26
REV1.0 2
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
Safety information Personnel operating and maintaining the equipment should receive system technical training and be thoroughly familiar with all aspects of its operation and maintenance. To ensure their safety, all users should read this chapter carefully before using the system. Inaddition, to promote safety awareness, Service Engineers should periodically review the basic precautions outlined in this chapter.
1Overview Do not remove the covers of the exposure equipment. Do not defeat or bypass built-in equipment safety features. Do not attempt to repair the equipment if it fails. Immediately call a person
qualified and authorized to repair the equipment.
Do not attempt to move equipment, connect or disconnect any communication cables.
Observe all warnings, cautions, stated or implied, instructions, in the procedures.
Keep all safety labels on the equipment permanently.
2Electrical Only a qualified/trained Service Engineer or electrician can replace electrical
components.
3Mechanical Keep fingers, hands, and tools away from moving parts. Unless specifically instructed, do not operate the equipment with covers or
panels removed.
Route cables properly to eliminate tripping hazards.
4Safety Warnings and Cautions The sale and use of the equipment has been approved by the government
REV1.0 3
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
Only qualified personnel may operate the system equipment. Operation of the equipment by persons who have not been trained or who are unfamiliar
with the systems functions and controls may cause serious injury to the
patient, serious injury to the operator, or damage to the equipment.
The system includes no user serviceable parts. For service assistance, contact Company.
5Radiation and Magnetic field Warnings and Cautions The system produces ionizing radiation. Operators observe all state and local
requirements and regulations.
6Other Warnings and Cautions The ASR-3000 Console component of the system may not be located in high
humidity environment.
The system and associated cables must not be operated in the presence of moisture.
To avoid excessive product leakage currents and to maintain product compliance to medical protective earth and grounding requirements, the
systems power cord must be connected directly to a hardwired ac mains
receptacle.
Ensure that earth grounding connections between the components and the building are maintained at all times. The components include Stand; X-ray
Generator; and Console.
The Console cabinet includes a dedicated, bolted, protective ground connection to the buildings electrical system. Do not remove the grounding
connection for any reason. If it becomes damaged, contact a qualified
Service Engineer.
The UPS must be replaced by an authorized company representative . The UPS contains lead and poses a hazard to the environment and human health
REV1.0 4
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
if not disposed of properly.
The ASR-3000 system is not suitable for operation in the presence of corrosive gaes.
7Warnings Moving and Using Equipment Use at least two qualified people when moving equipment in order to
prevent injury or strain.
The equipment is fragile and must be handled with care.
8Cautions Cleaning Do not spray cleaning solution directly onto the equipment. Instead, moisten
a cloth with the solution and wipe the equipment.
Do not immerse the equipment, including any components or accessories, in liquid.
Do not autoclave the equipment, including any components or accessories.
9Radiation Safety Exposure to ionizing radiation such as X-rays can be hazardous. Some effects of X-radiation are cumulative. Everyone associated with X-ray work must be familiar with the recommendations of the Center for Devices and Radiological Health , the National Institute for Standards and Technology, the National Council on Radiation Protection, and the International Committee on Radiation Protection. Do not attempt to operate X-ray systems for the health, it's hazardous to health. the walls and windows of the radiologist's workstation in most X-ray facilities must be specially disposed to protect against ionizing radiation, thick lead setting inner walls is widely used to separate from ionizing radiation Use the proper dose of X-ray for the patient to result in acceptable images
REV1.0 5
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
10Safety Symbols The following safety-related symbols are found on the equipment. To avoid injury, learn to recognize them.
Radiation
Power Off (used with the Console UPS)
Power On (used with the Console UPS)
Explosive Gas (flammable)
AttentionRead the accompanying CAUTION or WARNING
Non- Anesthetic Proof
Universal Interface Connection Identifiers
Sound
Earphones
Protective Earth Ground
Hazardous Voltage
Power On Indicator
REV1.0 6
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
On Line Indicator (used with X-ray Generator switch at Console)
Off Line Indicator (used with X-ray Generator switch at Console)
Prep/Exposure Switch
X-ray Exposure Switch
Do Not Immerse In Liquid
This product contains no field-serviceable parts.
Special Cleaning Instructions
Lifting Warning
Danger-Pinch Point
REV1.0 7
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
Chapter 1
Theory and Structure of the System
1.1 General ASR-3000 X-Ray mammography system is that the X-ray tube generates
X-ray by heating filament and inputting a high voltage, and then Radiography. To avoid the hazard of X-ray radiation and high voltage, ensure that line disconnect switches are opened before repairing the equipment .if must repair in the activated current, ensure to follow the safety prescribes of the correlative manuals.
1.2 The Structure of ASR-3000 X-RAY Mammography System
1.2.1ASR-3000 X-RAY Mammography System general components Wall Stand
Console
It's a high-frequency(40KHZ) tiny focus molybdenum target eddy-anticathode special X-Ray Mammography under the control of the computer
1.2.2 Main Frame contains
X-RAY tube
Rotation anode
X-Ray generator
Control system
Radiography easel
Mammography oppression
Filter and cassette
1.2.3 Console contains
REV1.0 8
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
Console panels
Switching power
1.3 Theory of the System
The working principle of the ASR-3000 X-Ray mammography is that the X-ray tube generates X-ray by heating filament and inputting a high voltage, and then Radiography.
1.4 The control theory of the system graph
REV1.0 9
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
Chapter 2:
Maintenance Schedules Take the proper preventative maintenance to ensure the system in the best working state and to reduce the expenses and time of maintenance by all means. The information of preventative maintenance is provided in this chapter. Service procedures must be performed by trained personnel only . Preventive Maintenance Schedules and Checklists Preventive maintenance procedures for the System components
2.1 In-Service Observation All provided products and associated components are designed for replacement or service in a hospital environment.
. Observe the operation of the locking mechanisms as the Wall Stand is
prepared for use. If the mechanism fails to release, the unit should immediately be removed from service and the failure repaired.
The Wall Stand and all accessories should be inspected for worn, loose, bent,
or broken parts.
2.2 Preventive Maintenance Schedules and Checklists
In accordance with using , please Preventive Maintenance
It includes SemiAnnual Preventative Maintenance/ Annual Preventive Maintenance Record the correlative results of maintenance in the following table:
REV1.0 10
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
Preventive Maintenance Checklist
User Name Linkman Telephone
Equipment name Inspection parts state
X-Ray generator
tube
X-Ray
collimator equipment Up/down system
Radiography easel filtercassette
writhe function
Up/down function
The Wall Stand
Transmission parts
Circuit board Console
display of panel
Sensor Position sensor
The whole
performance
applying state of the
performance
Service Engineersign Hospital Superinrtendentsign
Inspection Date:
2.3 System Component Preventive Maintenance
For preventive maintenance for the Wall Stand, refer to the topic Wall Stand Preventive Maintenance.
For preventive maintenance for the X-ray tube, refer to the manufacturers
REV1.0 11
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
service manual. For preventive maintenance for the Console, refer to the manufacturers
service manual.
2.4 Wall Stand Preventive Maintenance Ensure that the table is completely disconnected and that the emergency
switches have been activated, before starting any cleaning operations. Ensure that no liquid gets into the unit. Do not use water. Water can
short-circuit the electrical installation and cause corrosion to mechanical parts.
Do not use acid or abrasive products. Use a wax appropriate for the surface. When disinfecting the examination room, ensure the unit is covered with
plastic sheets. When using disinfectants, carefully follow the instructions for disinfecting
under explosion hazard conditions.
Chapter 3:
Diagnostics and Troubleshooting
This chapter describes the diagnostic and troubleshooting procedures for the
REV1.0 12
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
ASR-3000 System.
This chapter provides: General information on error logs Recommendations for responding to error messages Troubleshooting guidelines Troubleshooting sample
3.1 The Troubleshooting graph
3.2 Troubleshooting and Diagnostics
No.
Failure Phenomenon
Causes Solutions
1
Power Supply Indicator does
a. No power in the extern power mains. b. Power switch is not turned on.
a. Check external power mains.b. Turn on the power supply.
REV1.0 13
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
not light up after power on.
c. Check if the power cable is damaged.
2 ALARM a. Computer error. b. Ground wire is not well
connected.
a. Turn on the machine again after power off.
b. Check the ground.
3 After press the hand-switch, Ready LED does not light up.
a. The interval between the previous exposure and the next one is less than 30 seconds.
b. Rotating anode working circuit error. Anode cannot rotate or start up.
c. Filament heating circuit error.
a. Wait 30 seconds, then press the hand switch.
b. Check the rotating anode circuit board and replace the board if necessary.
c. Check the filament circuit.
4 The displayed
breast thickness value is not consistent with the actuality.
a. Computer error. b. Component error. c. The steel wire in the main unit
arm becomes loose.
a. Turn off the machine and turn it on again.
b. Check CPU board, control power supply board, replace boards if necessary.
c. Adjust the steel wire in the main unit arm.
d. Adjust the weight under the cassette of the machine head.
5 No X-ray output no exposure, no mA reading, film cannot be exposed
a. Computer error. b. High voltage circuit error.
a. Turn off and restart the exposure system.
b. Check high-voltage circuit.
c. Check high-voltage driver board, replace circuit board.
6 No exposure under
Auto mode . Computer error. . -ray sensor on the X-ray
detection board error.
.Turn off the machine and restart the exposure system.
b. Check X-ray detection board, measure thevoltages of Pin 1 and Pin 4 under thecondition there is no light illumination. The value should be 40. Under sunlight or under the illumination of 60W light bulb to
REV1.0 14
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
illuminate the X-ray tester , the signal output voltage should be 0-7. If no change, then the X-ray detector has problems. Please replace the PCB.
3.3 Examples
A hospital has a set of ASR-3000 installed. During their operation,
when press the auto exposure, the machine will not perform auto
exposure and alarm occurs. We guess when X-ray is radiated to the X-ray
dose detection board, no feedback signal is transmitted to the control
exposure console. Then, X-ray dose detection board may have problems.
Remove the X-ray dose detection board, power it on. Under the condition
that there is no illumination, test J1 voltage, P2 is +12V, P2 is 12V.
These are normal. While when signal output P1 and P4 voltages are
measured, the value should be 0V. Use sunlight or 60W light bulb to
illuminate the X-ray dose detector, the signal output voltage is also 0V.
After 10 minutes, the signal output voltage is also 0V. But when light
bulb is used to illuminate the X-ray detector, under normal conditions,
X-ray dose detection board J1s P2 and P3 voltages are respectively +12V
and 12V. While signal output P1 and P4 voltages should be 0-7V. With
the change in the illuminating time, P1 and P4 voltages also change.
Therefore, based on the above test results, we determine the X-ray
REV1.0 15
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
detector has problems. We replace the X-ray detector chips, power it up,
check the detector following the previous steps. All the test data are
normal. Then secure the X-ray dose detection board, perform auto
exposure test, press exposure button, if the auto exposure is normal, then
the problem is resolved.
Chapter 4 :
Service Tools
The following tools and instruments are necessary in the system
maintenance.
Description QTY
Insulation silicon rubber resin 1pcs
Rubber hammer 1pcs
REV1.0 16
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
Hot air blower 1pcs
Level 15in 1pcs
L Wrench (set) 1pcs
Pincers 1pcs
Cross screw driver 1pcs
Shape Screwdriver 1pcs
Electrostatic wrist 1pcs
Torch 1pcs
Hex Key set, metric 1pcs
Roll scale 3m 1pcs
Aperture wrench(set) 1pcs
Cable binder 10pcs
Cable clip 1pcs
Multi-meter(FLUKE 7X) 1pcs
Electric iron25with soldering tin 1pcs
Appendix 1 Function Modules and Configuration 1 ASR-3000 system configuration
ASR-3000 mamography consists of: control console display board, machine head display (left, right) board, HV driver board, anode startup board, main unit CPU board, pre-positioned power supply board, output high voltage feedback board. See figure below.
Electric adjustment
Output HV
t Boa
rd o
f mac
hine
REV1.0 17
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
Anode startup
Console display Pre-PS Control Bd.Pre-PS Bd.
Tube
CPU board Le
ft B
d of
mac
hine
H.V driver
24 12 12 5
switch PS
2Functions Console display Mainly includes illuminating diode, alphanumeric tube, function keys and buzzer.
Machine head display (left, right) Through connecting to CPU board, control the motion of the machine head and display some working modes.
X-ray dose detection board When auto exposure is used, automatically detect the dose of the exposure.
High-voltage driver board Supply high-voltage power for X-ray to generate X-ray.
Anode startup board Supply startup power for the tube to rotate the anode.
Main unit CPU board It is also named computer main board and mother board. It completes the data processing and flow control of the complete system and also is the core of the system.
Pre-positioned power supply control board Supply power to the control operation system.
REV1.0 18
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
3. The connectors, data flow direction and related test data
3.1 Control console display board
7-S07746
7-S07747
7-S07757
7-S06911
JP1
P1P2
J2
J8J7
J6J5J4
3.2 Machine head display board (left and right)
To CPU Bd.
J6
To CPU Bd.
J5 JP6
Control signal
J1
PS
REV1.0 19
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
Machine head display left and right board 5 and 6 are connected to JP3
on the CPU board, JP6 is connected to JP5 on CPU board.
3.3 Power Supply Control Board (vertical)
REV1.0 20
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
D7R D6C7ba
YJ1-1(A:7-S07684)
D6-W
JP1 JP2
D7-
b
aba
ba
ba
ba
D2D1
T2
C1-L
R36
+ AC
T2i2
B T2i1
R35C1+
R33
JP3
R34
JP0-10(7-S07685)
C2
C1
FFi1 Fi2
Fo1C2-
Fo2
Li1 Li2 C8C6D5
Wa
176
C2+ 45
102 3
Wb11
Q1D5+
98
D8+ C8- C8+D9K D8 C30
1 2 3 4 5 6 1 2 3 4 5 6 7 8 9 10 11 12
21 3
IN(
:7-S07684)
YJ1-2(A:7-S07684)IN(
:7-S07684)
JP1-5(B
JP0-11(7-S07685
JP0-12(7-S07685
YJ2-6(A:7-S07695)
YJ2-5(A:7-S07696)
DL-3(D:7-S07741DL-4(D:7-S07741
JP0-9(7-S07685)
IN(
:7-S07684)
OUT1(
:7-S07756)
OUT2(
:7-S07756)
Related test data When 220AV is input, check fuse , test diode B signal, measuring electrolytic capacitor C1 voltage to be 320V, electrolytic capacitor C8 voltage to be 110-145V (voltage value is different when different scale is selected). If abnormal phenomenon occurs, K or Q1 may fail and should be replaced.
3.4 Anode Driver Board
REV1.0 21
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
7-S07744
7-S07747 7-S07746
7-S07754
7-S07755
7-S07755
7-S07755
JP1-3(B:7-S07741)
K18K10 K15
JP2
K6K2J1
K4K9
R1
K14
K16D3
JP1
K19
JP4
K8 K1
K13
K7K3
J2
D1
K17K20K5 K12
K11
JP1-4(B:7-S07741)
PS2-N(D:7-S07745)
JP3T1 F1
PS2-L(D:7-S07745)
DL-2D1 250V/.2A
50/60Hz
D2
KT1
KJ1R1
C1C2
C4
R3
KJ2
C3
3.5 Control Box PCB (Middle)
JP4-6(D:7-S07755)
aa1a2
1 a
JP0-27-S07732
b3JP2-15(D7-S07755)
JP2-11(D7-S07755)
JP2-19(D7-S07755)
R12
JP4-5(D:7-S07755)JP1-3(D:7-S07755)
11
a1
JP1-
3D
7-S0
7755
JP4-3(
D7-S0
7755
JP4-
1D
7-S0
7755
JP2-
12D
7-S077
55
JP4-2(D7-S07755)
R323
JP0-37-S07735
JP4-4(D:7-S07755)
R2 R4 R5
REV1.0 22
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
3.6 Control Box PBC (Lower)
JP1-4(B)
JP3-5(D)JP2-7(B)
JP3-4(D)
JP1-3(B)
JP2-6(B)
JP1-2(B)
JP1-5(B)JP1-6(B)SW-4 SW-c
220/150
R2/1.5K
Out InTo4 1
3
1
2YJ1
2
5 4
1
2
3
2 1
JP2-8(
B)1 R1/30
JP2-3(
B)JP2-6(
B)
JP2-5(
B)
2T1
KT4
T2 3
21 3 4
R3/300
58 7 6 b
YJ2
21 3 4 a
JP2-8(
B)JP2-4(
B)
JP2-12
(B)
JP2-9(
B)
6 5L1 2
1JP2-10(B)
JP1-6(B)
JP2-11(B)
Related data of Control Console: DC power supply, 2GZM-60s12 voltage 1211GZM-
40512to be 5, 121GZM-10s12 to be 121. If the test results are not the same with the above reference data, PCB may fail and should be replaced.
3.7 High Voltage Driver Board
REV1.0 23
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
E1
J301J30210-S00003
E1C2
C1
J401
J402J403 GN2VC2
JP2
P
VC3GN3
E2
JP3
C2E1 E2
J20210-S00005
VC3GN3
GN2VC2
C1 G1
V+J10 J11
E1G1
K103
K102
V-
J22E2G2
CU104
Vout
1
CM
Vout2
G2E2
V1+
V1-
T124V 24V
18V 18V
220V 220V
J401 J404
J402 J403
K1032602mm
2002mm
2002mm
JP0-9 JP0-10
Related test data:
Connect to power mains and turn on 220VAC switch, J2-1.2.3.4Ground and SG3525s Pin 15 is 15V0.5V, SG3525s pin 16 is 5.1V0.5V. On OFGB board, adjust 310, LM324s pin 6 to 4047815pin 2 voltage to be 6.00, adjust 314, 324s 10 to 4047815pin 2 voltage to be 1.00.
If the test results are not the same with the above reference data, then, the PCB fails and should be replaced.
3.8 Main Unit CPU Board
REV1.0 24
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
Related test data
JP7
JP4
PS
JP10
sensorJP11
To HV J103
JP1
Sync signal
P2
To high volt J400
JP5
JP8 JP3 J3 JP9
5V Limits and motor driver
Display and key
Sensors connector JP6
When powered up at 220VAC, the main unit CPU board C18, C15, C7, C4 voltage at the two terminals are respectively -12V, 24V, 12V, and 5VC44 is 3.65VU17pin12 is 1.96Vwhile U17s pin 13 and pin 15 voltage is lower than 5V. If there is any discrepancy between the test results and the above reference data, the board may fail and should be replaced.
3.9. Main unit output power supply test data:
REV1.0 25
-
ASR-3000 X-RAY MAMMOGRAPHY SERVICE MANUAL Total 28 Pages
100V 220V F Ground
F large focus
Y R
12 11 10 9 8 7 6 5 4 3 2 1
100V Earth Ia F small focus
G 220V
Remarks a.100V-100V are to supply power to the tube and high-voltage startup board. When starting the exposure signal, the voltages of P1 and P2 are respectively 100V-100V. b. p3, p4 pins are power supply for AC 220 c. p6, p8, p9 are power supply for tube large and small focus d. p7 Ia is tube currrent e. p10, p11, p12 pins are power supply for tube anode and motor driver.
Caution: Refer ASR-3000 Operators Manual and Technical Manual when servicing and testing PCBs and pay
attention to the capacitor discharge process to prevent electric
shock.
REV1.0 26
ContentsSafety information