ars 1 - fraunhofer€¦ · ars 1.0 – working principle through an inlet neutral air is drawn into...

2
For evaluations of olfactory parameters it is still necessary to use human pannelists. Naturally the olfactory perception between individuals varies. Therefore several approaches have been tried out to linearize the pannelists responses. An approved and reliable approach for this is the use of a reference substance. An unknown olfactory impression is compared to a reference substance with regard to its e.g. odor intensity. Besides butanol, in some regulations acetone is used as reference substance. For example in DIN ISO 26000-28 the perceived intensity PI is referred to defined concentrations of acetone. For evaluations using a reference substance it is necessary to have a device, which provides the reference substance in a defined and reproducible concentration to the panelists. The ARS 1.0 was developed to accomplish these targets. The device generates at its outlet an acetone atmosphere in a concentration range suitable for olfactory evaluations. The acetone concentration can be defined by the user. It is provided to the panelists as long as he needs it, but not longer than necessary to avoid contamina- tion of the surrounding area. So the ARS 1.0 can be used for measure- ments according to DIN ISO 16000-28 and ISO 16000-30 in the laboratory but also out in the field. ARS 1.0 – Working principle Through an inlet neutral air is drawn into the device. To remove possible contami- nations from the neutral air, an air purifier is placed in the inlet. This neutral air is the constantly flowing main gas stream and is regulated by a mass flow controller. In a reservoir a small amount of liquid ace- tone is provided. The reservoir is connected with an airtight syringe. If the syringepiston is pulled, air is passing through the acetone and the saturated vapour phase is drawn Fraunhofer Institute for Building Physics IBP Department Building Chemistry, Building Biology, Hygiene Fraunhoferstraße 10 83626 Valley Working Group Chemistry Dr. Christian Scherer Phone +49 8024 643-246 [email protected] Stefan Mair Phone +49 8024 643-294 [email protected] COMPETENCES IN DETAIL ARS 1.0 FRAUNHOFER INStItUtE FOR BUIldING PHySICS IBP

Upload: others

Post on 18-Aug-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: ARS 1 - Fraunhofer€¦ · ARS 1.0 – Working principle Through an inlet neutral air is drawn into the device. To remove possible contami-nations from the neutral air, an air purifier

For evaluations of olfactory parameters it

is still necessary to use human pannelists.

Naturally the olfactory perception between

individuals varies. Therefore several

approaches have been tried out to linearize

the pannelists responses.

An approved and reliable approach for this

is the use of a reference substance. An

unknown olfactory impression is compared

to a reference substance with regard to

its e.g. odor intensity. Besides butanol,

in some regulations acetone is used as

reference substance. For example in DIN

ISO 26000-28 the perceived intensity PI

is referred to defined concentrations of

acetone.

For evaluations using a reference substance

it is necessary to have a device, which

provides the reference substance in a

defined and reproducible concentration to

the panelists.

The ARS 1.0 was developed to accomplish

these targets. The device generates at

its outlet an acetone atmosphere in a

concentration range suitable for olfactory

evaluations. The acetone concentration can

be defined by the user. It is provided to the

panelists as long as he needs it, but not

longer than necessary to avoid contamina-

tion of the surrounding area.

So the ARS 1.0 can be used for measure-

ments according to DIN ISO 16000-28 and

ISO 16000-30 in the laboratory but also

out in the field.

ARS 1.0 – Working principle

Through an inlet neutral air is drawn into

the device. To remove possible contami-

nations from the neutral air, an air purifier

is placed in the inlet. This neutral air is the

constantly flowing main gas stream and is

regulated by a mass flow controller.

In a reservoir a small amount of liquid ace-

tone is provided. The reservoir is connected

with an airtight syringe. If the syringepiston

is pulled, air is passing through the acetone

and the saturated vapour phase is drawn

Fraunhofer Institute

for Building Physics IBP

Department Building Chemistry,

Building Biology, Hygiene

Fraunhoferstraße 10

83626 Valley

Working Group Chemistry

Dr. Christian Scherer

Phone +49 8024 643-246

[email protected]

Stefan Mair

Phone +49 8024 643-294

[email protected]

COMPETENCES IN DETAILARS 1.0

F R A U N H O F E R I N S t I t U t E F O R B U I l d I N G P H y S I C S I B P

Page 2: ARS 1 - Fraunhofer€¦ · ARS 1.0 – Working principle Through an inlet neutral air is drawn into the device. To remove possible contami-nations from the neutral air, an air purifier

into the syringe. The syringepiston is always

pulled up completely. If this state is reached

the device can be seen as loaded.

On demand the acetone vapour in the

syringe is ejected into the main gas stream.

For high concentrations, the syringe is

emptied faster, for low concentrations

slower.

The concentration of acetone in the satu-

rated vapour depends on the temperature.

For this reason a temperature sensor is

implemented in the device and a regression

term with regard to temperature regulates

the ejection speed of the syringe.

By switching valves in an appropriate way

it is guaranteed, that the acetone vapour

is first led to the syringe and then to the

main gas stream.

ARS 1.0 – Product properties

– Free selectable acetone concentrations

– Release of acetone on demand

No contamination of the test area