appendix 1 - electrical current and the human body10.1007... · web viewappendix 1 - electrical...
TRANSCRIPT
Appendix 1 - Electrical current and the human body
There is a rather large inter-individual difference in the susceptibility for effects of electrical
current, as is illustrated in Table 1 by the difference in current inducing ventricular fibrillation in
5% and 50% of all persons. A similar variability is seen in the “let-go” current, which in women
varies between about 8 mA (5%) and 13 mA (95%), in men between 12 mA (5%) and 20 mA
(95%), with median (or 50%) values of about 11 mA and 16 mA, respectively [10] (the “let go”
current is the maximum current at which it is still possible to voluntary remove one’s hand from
a live part one is holding in his hand. Values higher than this are called “freezing” currents).
Currents through the chest above the “let-go” threshold may cause asphyxia, unconsciousness
and death if they are not interrupted in time. Induction of ventricular fibrillation forms the
greatest risk for a fatal outcome of electrical shock, however.
Small, seemingly innocent shocks can induce unintended, life threatening movements of the
sufferer. Larger currents may jolt the victim away from the apparatus due to strong, involuntary
muscle contractions. Severe muscular stimulation can cause ruptures of tendons and muscles and
even the fracture of bones.
Some factors affecting the magnitude and physiological consequences of an electrical current
through the human body are:
- The resistance met by the current. The resistance between the two hands is between about 1
and 10 kΩ, depending on contact area and skin wetness. Hand-foot values are 10% to 20%
lower than corresponding hand-hand resistances.
- The path of the current. The risk is especially high when part of the current passes through
the heart. Currents in the heart as low as 50 μA may cause ventricular fibrillation (the safe
1
limit is as low as 10 μA). The heart is most susceptible during cardiac repolarization (the T-
wave phase in the electro-cardiogram). Generally, fibrillation can only be ended by another
brief, strong electric shock that arrests the complete heart, allowing a synchronized restart.
- Shape of voltage. The thresholds for different levels of risk of fibrillation are somewhat
higher for full- and half-wave rectified currents (rms) from transformers than for the
corresponding unrectified currents. Table 2 illustrates this for situations in which contact lasts
several seconds. Thresholds for pulsed voltages (their amplitude, not the rms value) as
generated by an inductor are also shown in Table 2.
- The frequency of the source of electricity. The sensitivity is highest around 50 – 60 Hz, but
the IEC assumes a constant risk within 15 – 100 Hz. The threshold for fibrillation increases
with frequency, e.g., at 300 Hz it is about five times higher than at 60 (or 50) Hz, and at 1000
Hz about 14 times higher [11]. For DC-voltages the risks are also somewhat less than for 15
– 100 Hz AC-current, e.g. the threshold currents for ventricular fibrillation are about 3 times
higher than indicated in Table 1 when contact duration is between 1 to 10 seconds. For 10 ms
and 100 ms contacts, the thresholds are similar to those in Table 1.
As X-ray systems require high-voltage in addition to some primary source of “low voltage”
electricity, greater risks come into play. High-voltages (defined by the National Electrical Code
as > 600 V [10]) puncture the skin, lowering the resistance between two contacts on the body.
For the hand - foot resistance holds according to [10]: 5% of all persons have a resistance below
460 Ω, 50% of less than 620 Ω and 95% of less than 840 Ω. These numbers apply to AC- and
DC-current and are virtually independent of the size of the contact area. Depending on the
electrical power of the supply, high-voltages can induce very large currents (e.g., already 10 A to
2
20 A at 10 kV in case of a sufficiently powerful high-voltage source) causing severe damage to
tissues (e.g. to nerves, muscles and blood vessels) and subcutaneous burns.
A discharge may pass through the air, turning an insulating track of air into a conducting path
of rather low resistance. Fig. S1 shows the distance between sharp points high-voltage may
bridge in air through a discharge [13], which is taken as the minimum distance that should be
observed to prevent injury from a spark.
Fig. S1 Estimate of maximum distance a spark can bridge in air. It gives an indication of the
3
0 20 40 60 80 100 120 140 160020406080100120140160180200220240260280
Peak voltage [kV]Max
imum
spa
rk le
ngth
[mm
]
minimum distance that must be maintained from live parts of an X-ray system [13]. At 100 kV,
this is, for example, already 15 cm.
Appendix 2 - High-voltage power supplies used in X-ray systems
Static generators
Static generators based on the influence principle were already well developed in 1895
[16]. As they provided sufficiently high DC voltages for the production of X-rays, they found
immediate application in radiology. The early static models had the advantage of being run by
hand, and the continuous DC output eliminated the problem of “reverse current” that occurred
with induction coils after closure of the interrupter. Reverse current limited the life of the tube
and could cause X-rays to be generated at an unwanted place.
Static systems had disadvantages as well as they only produced small currents and
functioned poorly if the air was humid. Improved models, mainly variations of the Töpler, Holtz
and Wimshurst systems [17], soon became available [9,15], but the old problems were never
completely solved. When Leyden jars were used to achieve the desired high-voltage, these would
discharge through the tube once the discharge potential had been reached, resulting in a pulsed
X-ray beam much like induction coils produced.
Type (manufacturer) Diameter Plate[cm]
Numberof platesa
Speed of rotation[s-1]
Current[mA]
Wimshurst [17] 55 2 “max” 0.085Wimshurst [17] 55 8 “max” 0.21Wommelsdorf (Wehrsen) [18]
55 2 “max” 0.3 – 0.35
Toepler-Holtz (Wagner) [19] 71 4 30 1.3Toepler-Holtzb 71 100 30 10 - 20
Table S1 Current to be drawn from static machines
a a single element of any static machine consists in principle of 2 platesb very large system that was probably too expensive, complicated and noisy for common use from R.V. Wagner, Chicago, USA [20,21]
4
The manufacturers of static systems generally provided limited information on their
systems. Typically, they provided some construction details specific to their apparatus and the
maximum spark length. From this spark length the high-voltage, of the order of 50 – 150 kV,
could be deduced (Fig. S1). The current a system could deliver was usually unspecified. With a
few exceptions, we only found measured values in scientific publications, e.g., in the chapter
“Elektrisiermaschinen und Apparate” by Schmidt in Graetz’ books on physics [17] and in the
work by Wommelsdorf [22]. Some of this data is shown in Table S1. The maximum current of
the commonly used systems likely did not exceed 1 mA. When these static generators were
effectively short-circuited, e.g., by a person of low impedance touching both output terminals,
the current was not much higher as the system behaved as a current-source. Personal risks were,
therefore, small.
Because of the low maximum output current and the problematic nature of static
machines, most radiological professionals considered the more expensive induction coil superior.
Both slowly lost their place once the high-voltage transformer became available in 1907.
Induction coils
Like static generators, induction coils predated the discovery of the x-ray. [23]. Early
induction coils were mostly low power devices. For example, the Gerhardt, Bonn, system we
tested before was rated at 100 Watt, corresponding to an upper limit of the average secondary
current of 1 mA at 100 kV (in practice rather a few tenths of a mA) [42]. The new application of
inductors in generating X-rays prompted the development of more powerful systems.
Table S2 gives an overview of data for a few commercial products. Manufacturers of
inductors provided few specifications in the early years, which explains the many empty cells in
the Table.
5
Table S2 Power specifications of some induction coils (maximum high-voltage ≥ 100 kV)a
Application / manufacturer Year Primary power
(max) [kW]
Secondary power[kW]
Secondary currentb
(max.) [mA]
Diagnosis (and other purposes) Gerhardt, Bonn [42] <1896 0.1 1 Siemens & Halskec, Berlin 1901 2 Max Kohl, Chemnitz [24] 1902 3.3 Siemens & Halskec, Berlin 1908 15 45d
American X-ray, New York [45] 1915 6.6 2 20Therapy RGSe, Erlangen 1913 0.75 RGSe, Erlangen 1919 0.36 RGSe, Erlangen 1922 0.57
a All specifications from manufacturers, except for the Max Kohl coilb Time averaged currentc From Siemens MedArchiv (personal communication)d Estimated assuming an efficiency of 30% (see text)e RGS = Reiniger, Gebbert & Schall (Siemens MedArchiv, personal communication)
When the efficiency of an induction coil is known, the secondary current can be
estimated from the primary power. Salomonson performed such efficiency measurements for a
12-inch (30cm) spark induction coil from Max Kohl, Chemnitz [24]. The efficiency was between
35 – 54% when a rectifier was inserted to block the negative lobe of the output pulse and the
losses caused by the interrupter (likely a mercury interrupter) were taken into account.
According to Armagnat, the efficiency for a Wehnelt interrupter might still be lower, 20% or
even less [23]. These figures are similar to the efficiency calculated from the primary and
secondary ratings of the American X-ray coil as listed by the manufacturer [45], which was 30%
(Table S2).
6
Assuming an efficiency of 30%, the system with the highest primary power of 15 kW
(Siemens & Halske) would produce 4.5 kW at its output. At a high-voltage of 100 kV, the
corresponding time averaged output current would be 45 mA maximum. This was an impressive
output as Morton noted in 1915, “with a modern tube and powerful apparatus, a current of ten
milliamperes is quite ordinary.” [25].
In therapy the power used was generally lower than in diagnostic applications as
prolonged exposures were not a problem and low power had the advantage of sparing the tube.
However, the high-voltage for deep therapy was higher (180 – 250 kV) than for imaging or
superficial therapy (up to 100 – 150 kV). The replacement of coil systems was slower in deep
therapy than in diagnostic radiology because coil systems were easier to make for higher
voltages than transformers. Even around 1920 several authors still doubted that transformers
would overtake induction coils [24,26-28].
The frequency of the pulsed high-voltage supply is relevant for its physiological effect.
Typical frequencies are shown in Table S3. Most frequencies fall within the most dangerous
range of 15 – 100 Hz, except coils with Wehnelt interrupters which mostly worked at higher
frequencies. After 1903 the Wehnelt system and the mercury jet interrupter were the favorite
interrupters for many users, though they were far from perfect [28-31].
Table S3 Frequency of HV-pulses from induction coils with various interrupters [29,30]Type of interrupter Frequency [Hz] RemarksClassical hammer-on-spring, Pt-contacts, Wagner, Neefs 15 - 20 Low power, low voltageMechanical interrupter, with Pt-contacts, Deprez 15 - 40 Low power, low voltageMotor mercury interrupter with dipping contact 20 - 30 Larger coilsMotor mercury interrupter with double dipping contacts 40 - 50 Larger coilsMercury-jet interrupter ≈ 100 Large coils, up to 240 V inputElectrolytic interrupter, Wehnelt 300 – 400a High power
a Typical frequency range, but frequencies could be as high as 5 kHz
7
What happened if a person came into contact with the secondary circuit of an induction
coil X-ray system? It was common to leave the secondary, including the X-ray tube, electrically
floating, i.e. no point was connected to earth. A person touching a live part of such a system
received a shock, but one with little risk (see the discussion below in Appendix 7, “Types of
electrical accidents”, points 1 and 2). When contact was made with both poles of the secondary,
or equivalently with the anode and cathode of a powered tube, more severe shocks were possible.
The time-averaged currents given in Table S2 are unsuitable for estimating risk according
to IEC [11]. Peak currents are required, and these can be assessed from information on inductors
from Rühmer [32] and basic theory of inductors [23,33]. The maximum current in the secondary
after break, Ismax, is equal to I0/M, with I0 the primary current just before break, and M the
transformer ratio of the inductor [23,33]. Some information on a series of typical induction coils
from the Allgemeine Elektrizitäts Gesellschaft (AEG, Berlin) is given in Table S4 [32]. To have
such coils produce their rated spark length, using a mercury interrupter at a frequency of 18 Hz
with equal closure and opening times, the average primary current had to be about 3 A.
Assuming the primary current increased approximately linearly between closure and break, the
maximum current at break (I0) must have been about 12 A. Table S4 then shows the peak-peak
currents that might flow in the secondary in case of a short-circuit.
An approximation of the upper limit of the short-circuit current from these coils might be
calculated in the following way. Rühmer specified the maximum power the inductors could
transfer continuously when they were used as a normal transformer [32]. Although inductors
were not used to generate X-rays in this manner, we assumed the same maximum power in the
calculation of the short-circuit current for conventional interrupted DC. We further presumed
220 V DC-mains was available from which the power could be taken (Table S4).
8
Table S4 Typical properties of inductors [32] and the currents in a short-circuited secondary
Spark length 33 cm 43 cm 54 cm 65 cmWeight 42 kg 58 kg 70 kg 103 kgInductor used as transformer for normal AC as from mains Transformer ratio M, assuming 80% efficiency 240 300 350 420 Maximum power continuously allowed 600 W 900 W 1300 W 1900 WSettings for rated spark length (at 18Hz)a
Voltage (DC) 20 V 24 V 28 V 32 V Primary current I0 before break 12 A 12 A 12 A 12 A Estimated max. secondary short-circuit currentb Ismax = I0/M 50 mA 40 mA 34 mA 29 mAMax. power. 220 V-DC; frequency and current set for max powera,c
Frequency of interrupter 44 Hz 33 Hz 25 Hz 20 Hz Primary current I0 before break 11 A 16 A 24 A 35 A Estimated max. secondary short-circuit currentb Ismax = I0/M 45 mA 55 mA 68 mA 82 mA
a Mercury interrupterb Currents are peak-peak valuesc For 110 V-DC, frequencies would be halved, currents doubled, assuming M remained the same
When a Wehnelt interrupter was used instead of a mercury interrupter, the frequency was
likely in the range of 300 – 400 Hz, and the threshold for the induction of fibrillation increased
by a factor of about 5.
In practice the power supply was adjusted in such a way that a specific gas X-ray tube
received the optimized combination of voltage and current for a specific task. This adjustment
was usually done in the primary circuit by choosing one of several separate primary coils
available in inductors for use with mercury and Wehnelt interrupters, through the adjustment of
rheostats, and sometimes through the introduction of a coil for additional inductance. Thus, in
many cases Ismax would have been less than the estimates presented at the bottom of Table S4.
Comparison with the thresholds for fibrillation in Table 2 shows that the risks generally must
have been small. However, current settings presenting greater risk were possible.
High-frequency coils
9
X-ray systems using high-frequency inductors (often called Tesla coils) as high-voltage
power supply were rather popular in the US, but their specifications were hard to find.
A high-frequency coil generally used the output of an induction coil as its input. The
output consisted of high-voltage bursts of high-frequency oscillations. Specially constructed gas
tubes that suppressed reverse current were required so that only one polarity of the high-voltage
would generate X-rays. The high-voltage and current were of a similar magnitude as those of
commonly used induction coils (the quality of the X-rays was likely similar). The efficiency of a
Tesla coil could be high. A typical frequency might have been about 500 kHz, a high frequency
with low likelihood for causing muscular contraction and ventricular fibrillation. In case a person
short-circuited the output of a high-frequency system the most important effect of the electrical
current was heating of the tissue, implying that burns were possible. Note that the so-called skin-
effect is still negligible for tissue at the frequency mentioned.
Transformers
What is commonly called a transformer differs from an induction coil in that a
transformer’s iron core is normally closed, e.g. often has a rectangular or circular shape. The
core of an induction coil was generally linear and open. The transformer was normally fed by a
sinusoidally shaped AC-voltage, for instance from the mains power supply, and was well suited
for high power applications. The negative lobe of the sinus of the high-voltage had to be
eliminated for the gas tube, so some form of rectification had to be applied. For this purpose
Snook introduced a mechanical rotating rectifier in 1907, along with his transformer [9,15].
Later, suitable rectifying tubes (valves) were developed, and with the advent of the Coolidge
vacuum hot cathode X-ray tube, the X-ray tube itself could be used as rectifier for low power
10
applications. To give an idea of the behavior of a transformer when it is short-circuited, we did
some tests with a simple 220 V (50 Hz), 24 V and 20 VA transformer. The maximum allowable
current is (20/24) = 0.83 A, but when it was short-circuited the output current increased to 5.1 A,
i.e. it became more than 6 times higher. To illustrate the difference in risk between using a
resistor network and an autotransformer for setting the voltage, the primary voltage was
(arbitrarily) set to 150 V for a secondary load current of 0.6 A using each method. When the
transformer was subsequently short-circuited, the current increased to 1.46 A for the resistor
network and 3.56 A for the autotransformer, showing a higher current (corresponding to a higher
risk) of the autotransformer. Although the tested transformer was not of the high-voltage type, it
might serve to illustrate the typical behavior of transformers.
In 1919 a complete X-ray unit in an oil-filled tank, the CDX (Coolidge Dental X-ray
unit), was developed at General Electric based on a patent obtained by Waite [34]. It only
required a low voltage mains connection, and as no voltage carrying parts could be touched, it
was electrically shockproof. Its current, however, was low, 5 or 10 mA at 60 kV. High power
systems were necessarily more voluminous. A few manufacturers built large structures with all
components within one shielded container, sometimes with the power supply outside the
examination room [9]. More flexible high-power systems only became shockproof after the
shielded flexible (rubber) high-voltage cable was developed at the end of the 1920s [51]. Tubes
and power supplies, often submerged in oil, were shielded in separate housings and connected to
each other with the new cable, yielding completely electrically shockproof systems. Philips
introduced such a system, the Metalix Junior, in 1928 [35]. Siemens-Reiniger-Veifa introduced
their TUTO fully shockproof system in 1931 [Siemens MedArchiv; personal communication].
11
Appendix 3 – Table S5 Specifications of the allowed load of some X-ray tubes
Application / manufacturer Year Current & Time the current was allowedGas tubes Fluoroscopy Radiography Therapy Crookes nr. 9 [42] Early 1896 0.1 – 0.2 mA & minutesa
Green & Bauer [40] About 1910 ≈10 mAb & 30 s RGS*, Erlangenc 1915 2 mA & ? 40 mA – 60 mA & ? Philips, type R Id 1923 5 mA & some minutes 40 mA & 5 s Philips, type R II Momentd 1923 70 mA & 0.5 s; 30 mA &10 s
Philips, type T IId 1923 2 mA & 60 min (130 kV) Philips, type T IIId 1923 2.2 mA & 60 min (155 kV)
Hot cathode (Coolidge) tubes GE [43] 1913 26 mA & cont. (70 kV) GE, Universal broad focus [41] 1924 5 mA & cont. (140 kV) 80 mA & 5 s (100 kV) RGS*, Erlangenc 1922 10 mA 20 mA Philips, Metalix Type Dd 1927 6 mA & cont. (120 kV) 100 mA & 0.5 s (70 kV) RGS*, Erlangenc 1925 10 mA & cont. (65 kV) 150 mA & 0.5 s (50 kV) Philips, Rotalixd 1929 2 mA & cont. (100 kV) 230 mA & 0.04 s (100 kV)
6 mA & 10 s (100 kV) SRV, Goliathc 1929 555 mA & 0.1 s (50 kV)
86 mA & 5 s (100 kV)
Philips, Metalix Type Fd 1925 4 mA & contin. (220 kV) SRV, T III/4c 1926 7 mA & contin. (165 kV)
4 mA & contin. (220 kV)a Estimate [42]b Assuming coil efficiency of 50%c from Siemens MedArchiv (personal communication)d Philips (personal communication)
12
Appendix 4 - Safety recommendations and regulations
The American Röntgen Ray Society established a committee to investigate electrical
hazards and provide recommendations to improve safety in 1920 [55]. The results were
published in 1923 [56]. Some of the recommendations were: high ceilings, dry rooms, insulating
floors, no footswitches, quick acting automatic circuit breakers acting on a 20% overload (in
addition to fuses), easily accessible operating switches, overhead high tension conductors at
sufficient height made of tubing not less than half an inch in diameter and firmly mounted,
strong high tension wire on automatically winding reels to connect the X-ray tube to the
overhead tubing and to keep the wires straight, insulating patient table and the application of
sufficient insulation material and distance between persons (both patients and operators) and live
parts.
In the US new rules were provided in the 1931 “National Electrical Code” (NEC)
regulations [7], with revisions in 1933 and 1937. Even in 1936 US legislation did not yet require
that newly installed systems be of the shockproof type [57]. Today’s requirements are laid down
in the 2011 edition of the National Electrical Code NFPA 70, with articles 517 (medical) and 660
(non-medical) dealing with X-ray installations.
In Germany DIN RÖNT I was issued in 1930, providing rules to increase electrical safety
(DIN stands for Deutsches Institut für Normung). It distinguished several classes (A – D) of
systems depending on the degree to which the various parts were shielded with an earthed
conductor [8]. The rules are too detailed to summarize here. The document included a class of
apparatus, A, with all live parts fully protected by an earthed conductor, but such electrically
shockproof systems were still rather exceptional at that time. Electrically shock- proof systems
have been legally required in Germany since 1954.
13
Appendix 5 – Fatal electrical accidents with X-ray systems
Case Approximate Name Age Profession Country (state)
No. date
1 18-Sep-1906 Murphy MF 58 Patient, financier USA, Pennsylvania
2 1-Nov-1917 Clapp AM 41 MD USA, Massachusetts
3 6-May-1919 Fishpaugh RS 21 Private (soldier) USA, Maryland
4 28-Nov-1919 Jaugeas F 39 MD France
5 16-Jul-1920 Cope CL 29 MD, Dentist USA, Indiana
6 11-Nov-1920 Ilg C 35 Dental patient USA, New Jersey
7 16-Aug-1921 Bagley W ?? Demonstrator, nurse USA, Illinois
8 20-Oct-1921 Bryant WE 21 Assistant USA, Massachusetts
9 10-Mar-1922 Law IT ?? MD USA, Florida
10 21-Oct-1924 ?? ?? MD Finland
11 21-Oct-1924 ?? ?? Helper, female Finland
12 1925 Remy E ?? MD USA, Ohio
13 15-Oct-1925 Farago S 30 MD Hungary
14 23-May-1927 Matevich L 38 Father of patient USA, Illinois
15 9-Jun-1927 Cardona A 32 X-ray technician France, Algeria
16 5-Jan-1927 Klubien E 48 MD Denmark
17 6-Feb-1928 Bolton AT 49 MD USA, Ohio
18 7-Jun-1928 Steuer H 35 MD USA, Ohio
19 18-Dec-1928 Laverack D 8 Patient, girl UK
20 6-Feb-1929 Ledon P ?? X-ray technician France
21 8-Mar-1929 Robb HJ 40 MD USA, Montana
22 5-Sep-1930 Yancey BA 25 MD USA, Missouri
23 20-Sep-1930 Chadwick B 22 Nurse USA, Iowa
24 26-Oct-1930 Nyitrai B 36.5 MD Hungary
14
25 2-Mar-1931 Lemire WA 53 MD USA, Michigan
26 1931 ?? ?? Patient Italy
27 29-Oct-1931 K., Lissy ?? Patient Germany, Ostpreussen
28 28-Mar-1932 ?? 46/48 Patient Italy
29 2-Jul-1932 Magnusson HV 65 MD USA, California
30 16-Nov-1932 Eberle JD 52 MD Swiss
31 19-Jan-1933 Davis DD 32 MD USA, Iowa
32 16-Dec-1933 Barnes D 26 Nurse UK
33 5-Sep-1934 Berger S 55 Patient USA, New York
34 20-Mar-1937 Coates WM 29 Physicist USA, New York
35 27-Jun-1937 Thane BJ 51 MD USA, Montana
36 24-Sep-1937 Tracy HM 24 Electrical engineer USA, Illinois
37 2-Oct-1939 Kostas G 10 Patient, boy USA, Idaho
38 18-Jul-1940 Pompa M 29 Patient USA, Connecticut
39 24-Oct-1940 Brixey HH 45 MD, Dentist USA, California
40 17-Dec-1940 Cummings RE 56 MD USA, Pennsylvania
41 24-Sep-1941 Walker H 36 MD, Dentist USA, Illinois
42 11-Mar-1948 Kellet EG 23 X-ray repair man USA, Maryland
43 7-Nov-1949 Mandula I ?? Doctoral candidate Hungary
44 27-Jun-1951 Roderick J 24 X-ray technician USA, Pennsylvania
45 25-Jul-1951 Rius JM ?? MD Spain
46 3-Nov-1955 Walton PK 28 Mother of patient Australia
47 5-May-1956 Esclangon 40 MD?, Professor France
48 21-Jul-1959 Frehner A 6 Child in shoe shop Swiss
49 14-Jun-1962 Jones RJ 80 Patient USA, California
50 1919- 1933 ?? ?? Excited kin of patient Germany (?)
51 1919-1933 ?? ?? MD Germany (?)
15
Appendix 6 - Serious but non-fatal electrical accidents with X-ray systems
Case Approximate Name Age Profession Country (state)
No. date
1 22-Feb-1908 Maloney JJ ?? MD USA,
2 20-Dec-1910 Gaze WH ?? MD Australia
3 5-May-1913 Dr. R. ?? MD (lecturer) Austria
4 1-Oct-1921 ?? ?? Patient Germany ?
5 22-Jan-1922 Dr. B. ?? MD Austria
6 21-Oct-1924 ?? ?? Patient Finland
7 19-Nov-1924 ?? ?? Nurse Germany ?
8 13-Aug-1925 ?? ?? Patient Germany ?
9 15-Dec-1925 Dr. H. ?? MD Austria ?
10 18-Dec-1925 ?? ?? Nurse Austria ?
11 29-Jun-1926 Pike CM ?? Father of boy patient USA, Pennsylvania
12 28-Feb-1928 Matthews A ?? Nurse, student - USA, Iowa
13 8-Mar-1929 Waite ?? Nurse USA, Montana
14 27-Mar-1930 Sidenberg SS ?? MD USA, Ohio
15 27-Mar-1930 Hahn WF ?? Custodian children's ward USA, Ohio
16 27-Mar-1930 ?? ?? Patient, baby/child USA, Ohio
17 1930 ?? ?? MD, intern USA, Missouri
18 7-Aug-1931 Robinson WW ?? MD USA, Washington
19 29-Oct-1931 ?? ?? MD Ostpreussen
20 28-Mar-1932 Gortan M 59 MD Italy
21 28-Mar-1932 D'Este ?? MD Italy
22 28-Mar-1932 ?? ?? MD Italy
23 25-Sep-1933 Geering ?? Fire brigade officer USA, Pennsylvania
24 21-Feb-1934 Selby H ?? MD UK
25 12-Dec-1934 Bethune P ?? Hospital porter Australia
26 11-May-1936 Ratner I 33 Nurse USA, New York
27 25-Oct-1936 Coats EB ?? MD USA, Montana
16
28 17-Dec-1936 Russell 23 MD Australia
29 13-Jun-1939 Speirs RB ?? MD Australia
30 24-Jun-1939 Epstein ?? MD Australia
31 1940 Bonnie (pseudonym) ?? Radiographer UK
32 14-Dec-1944 Barron WH ?? MD USA, Texas
33 1919-1922 Hemler WF ?? MD USA
34 1919-1922 ?? ?? Orderly USA
35 1919-1933 ?? ?? ?? Germany ?
36 1921 ?? ?? Nurse Germany ?
37 1922-1927 ?? ?? Patient Germany ?
38 1922-1927 ?? ?? MD Germany ?
39 1928 Pansdorf H ?? MD Germany
40 1919-1922 ?? ?? Patient USA
41 1919-1922 Dr. F.V.M. ?? MD USA
42 1919-1933 ?? ?? MD Germany ?
43 before 1929 ?? ?? Assistant Germany ?
44 1919-1933 ?? ?? Patient Germany ?
45 1919-1933 ?? ?? MD Germany ?
46 1919-1933 ?? ?? MD Germany ?
47 1919-1933 ?? ?? Engineer Germany ?
48 1919-1933 ?? ?? Nurse Germany ?
49 before 1922 Hemler WF ?? MD USA
50 before 1922 Dr. S.C. ?? MD USA
51 before 1922 ?? ?? Anesthesist USA
52 before 1934 ?? 35 Patient USA
53 1928 Clarke BLW ?? MD Australia
54 before 1918 Thimgan D ?? Father of patient little girl USA
55 2-Mar-1920 Nixon SF ?? Patient USA, Pennsylvania
17
56 before 1926 Pugh-Grant EM ?? Nurse, test-patient USA, Texas
57 before 1928 Curley E ?? Patient USA, Massachusetts
58 before 1929 Ragin ?? Patient USA. California
59 1-Dec-1933 Kelly ?? Mother, helped USA, Pennsylvania
60 before 1944 Rabasco ?? Father of a child patient USA, New York
61 before 1944 McLaughlin ?? Patient USA, Arkansas
62 2-Mar-1955 Sinatra ?? Patient USA, New Jersey
18
Appendix 7 - Types of electrical accidents
We will restrict our evaluation to accidents with transformers as we assume that they
were responsible for nearly all accidents (merely two cases clearly involved coils). Only the most
common accidents will be addressed; more complicated sequences of events and additional
information are found in the work of Grossmann [38,39] and in Grashey’s Chapter 4 of [37].
1. By its construction, the secondary coil of a transformer was normally floating, i.e. not
connected to earth at any point. In this configuration both poles had a similar variation in
potential with respect to earth. For instance, if the secondary voltage was 70 kV peak-to-
peak, each pole’s potential would approximately vary between -35 kV and +35 kV with
respect to ground, and both poles would momentarily have opposite potential. If a person
who was well isolated from earth touched one pole of the secondary, a small current may
have flowed through corona discharges. In addition, a very small current would have been
present as the person and the other pole were both capacitively coupled to ground, forming a
closed circuit of high impedance. These currents would have posed negligible risk. Note that
static machines and induction coils normally had a floating output when used for generating
X-rays as well.
2. If the person was grounded when he touched a floating secondary, e.g., because he was in
contact with a grounded part of the X-ray stand or was standing on a good conducting floor,
the point of contact of the secondary would effectively obtain ground potential. The charge of
the parasitic capacitive system that the secondary circuit forms with earth would change,
leading to a brief current through the person that was felt as a shock. Though it may have
19
been painful, it was likely harmless. In addition, a small continuous current ran as mentioned
above.
3. There was a risk at higher potentials when a grounded person touched a floating secondary. If
one pole of a formerly floating secondary was grounded, the other pole’s potential varied
between plus and minus the full potential, e.g., between – 70 kV and +70 kV for a 70 kV
peak-to-peak output. A transformer, and other parts of the X-ray system, designed to
withstand half of the full potential as was sufficient under normal conditions, could not
necessarily do so with the full potential. If it led to a shortcut, e.g., in the transformer, or to a
discharge to a grounded part, a closed current loop was formed from the contacted pole,
through the person and earth, to the other pole whose isolation failed. The risk of serious
injury or death was high in such an occurrence.
4. To ascertain that the full potential was indeed equally distributed over both poles of the
secondary, some manufacturers grounded the middle of the secondary. These systems were
the most dangerous even if one was exposed to only half of the full potential. If a grounded
person touched one of the poles, a closed circuit was immediately in existence: contacted
pole of secondary, person, earth, middle of secondary.
5. The last type of severe hazard to consider occurred when someone came into contact with
both poles, for instance during repositioning a tube by taking the anode in one hand and the
cathode in the other while the high-voltage was on. Alternatively, another person could also
inadvertently switch the power on, or the victim himself could mistakenly step on the foot
switch intended for fluoroscopy. It was not necessary to make physical contact with a live
part; coming close to it might have had the same effect, although an air gap somewhat limited
the magnitude of the current. It also appeared that in several cases the path of the current was
20
considerably more complicated than discussed above. It was not uncommon for metal parts
or several persons to come in series in the path of the current. In an accident of type 5, for
instance, two persons were standing directly next to (or against) each other, and one took the
cathode of the X-ray tube in his hand and the other tried to take the anode while the high-
voltage was on. The current passed through both persons, and both were killed [60]. The
footswitch was a dangerous device because a misstep could turn on the high voltage. It
provided also an effective connection to earth, because either the housing was grounded, or
the electrical wires inside were well within sparking distance when a “hot” foot was on it
(Fig. 1). These wires generally had some connection with mains voltage, and mains voltage
has a low impedance connection to earth.
21