anvisa - home | abracro...the validation report of the transferred method as proof that it was...

22
Ministry of Health – MS Brazilian Health Surveillance Agency - ANVISA This text does not replace the text(s) published in the Brazilian Official Gazette RESOLUTION OF COLLEGIATE BOARD RDC # 166, OF JULY 24, 2017 (Published in Brazilian Official Gazette #141 on July 25, 2017) It provides for the validation of analytical methods and other provisions. The Collegiate Board of the Brazilian Health Surveillance Agency, in the use of the attribution conferred by art. 15, III and IV allied to art. 7, III and IV of Law No. 9,782 of January 26, 1999, and to art. 53, V, Paragraph 1 and Paragraph3 of the Internal Regulations approved in accordance with Annex I of the Resolution of the Collegiate Board of Directors - RDC No. 61 of February 3 rd , 2016, resolves to adopt the following Resolution of the Collegiate Board of Directors, as resolved at a meeting held On July 11 th , 2017, and I, the Chief Executive Officer, hereby determine its publication. CHAPTER I INITIAL PROVISIONS Section I Objective Art. 1 This Resolution establishes criteria for the validation of analytical methods. Sole paragraph. Failure to comply with any of the criteria set forth in this Resolution must be technically justified and subject to analysis by ANVISA. Section II Scope Art. 2 This Resolution applies to analytical methods used in pharmaceutical ingredients, drugs and biological products at all stages of production. Paragraph 1 - Validation parameters and their respective acceptance criteria shall be defined according to the characteristics of the analyte and the nature of the method. Paragraph 2 - Analytical methods applied to the products under investigation used in clinical trials shall have their suitability demonstrated in accordance with this Resolution, as applicable for each phase of clinical development. I - The use of an alternative approach must be technically justified based on recognized scientific references. Paragraph 3 - The use of alternative approaches for the validation of analytical methods applied to biological products, such as biological and immunological tests, shall be permitted.

Upload: others

Post on 18-Feb-2021

0 views

Category:

Documents


0 download

TRANSCRIPT

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    RESOLUTION OF COLLEGIATE BOARD – RDC # 166, OF JULY 24, 2017

    (Published in Brazilian Official Gazette #141 on July 25, 2017)

    It provides for the validation of analytical methods

    and other provisions.

    The Collegiate Board of the Brazilian Health Surveillance Agency, in the use of the

    attribution conferred by art. 15, III and IV allied to art. 7, III and IV of Law No. 9,782 of

    January 26, 1999, and to art. 53, V, Paragraph 1 and Paragraph3 of the Internal Regulations

    approved in accordance with Annex I of the Resolution of the Collegiate Board of Directors -

    RDC No. 61 of February 3rd, 2016, resolves to adopt the following Resolution of the Collegiate

    Board of Directors, as resolved at a meeting held On July 11th, 2017, and I, the Chief Executive

    Officer, hereby determine its publication.

    CHAPTER I

    INITIAL PROVISIONS

    Section I

    Objective

    Art. 1 This Resolution establishes criteria for the validation of analytical methods.

    Sole paragraph. Failure to comply with any of the criteria set forth in this Resolution must be

    technically justified and subject to analysis by ANVISA.

    Section II

    Scope

    Art. 2 This Resolution applies to analytical methods used in pharmaceutical ingredients,

    drugs and biological products at all stages of production.

    Paragraph 1 - Validation parameters and their respective acceptance criteria shall be

    defined according to the characteristics of the analyte and the nature of the method.

    Paragraph 2 - Analytical methods applied to the products under investigation used in

    clinical trials shall have their suitability demonstrated in accordance with this Resolution, as

    applicable for each phase of clinical development.

    I - The use of an alternative approach must be technically justified based on recognized

    scientific references.

    Paragraph 3 - The use of alternative approaches for the validation of analytical methods

    applied to biological products, such as biological and immunological tests, shall be permitted.

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    Paragraph 4 - Microbiological methods for which a technical justification for the chosen

    approach based on the Brazilian Pharmacopoeia or other official compendia recognized by

    ANVISA are excluded from this Resolution.

    Section III

    Definitions

    Art. 3 For the purposes of this Resolution, the following definitions are adopted:

    I.- Sample: representative quantity of pharmaceutical ingredient, intermediate product or

    finished product, duly identified, within the established period of validity;

    II.- Analyte: substance or set of substances of interest to be identified or quantified;

    III.- Characterization of chemical substance: this is the set of tests which unequivocally

    guarantees the authenticity and quality of the substance as regards its identity, purity, content

    and strength and shall include data obtained from techniques applicable to the characterization

    of each substance, for example, thermogravimetry, melting point, differential scanning

    calorimetry, infrared spectroscopy, mass spectrometry, nuclear magnetic resonance, elemental

    (carbon / hydrogen / nitrogen) analysis, X-ray diffraction, optical rotation, chromatographic

    assays, among others;

    IV.- Analytical run: set of measurements performed on a group of samples within a

    predetermined time point under the same repeatability conditions, such as method, analyst,

    instrumentation, location and conditions of use;

    V.- Matrix effect: effect of matrix components on analytical response;

    VI.- Test: a technical operation consisting of the determination of one or more

    characteristics of a given ingredient or product according to a specified method;

    VII.- Limit test: tests to verify that the amount of analyte is above or below a pre-

    established level, without quantifying it accurately;

    VIII.- Response factor: ratio between analytical signal and analyte concentration;

    IX.- Relative response factor: ratio between two response factors, which is used as

    correction in the calculation of the concentration of a substance when it is measured by means

    of the analytical response of another;

    X.- Quality management: this is what determines the implementation of the "Quality

    Policy", that is, the overall intentions and guidelines regarding quality, formally expressed and

    authorized by the superior management of the company;

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    XI.- Impurities: any component present in the pharmaceutical ingredient or the finished

    product other than the active pharmaceutical ingredient or the excipient;

    XII.- Pharmaceutical ingredient: any substance being part of the formulation in a

    pharmaceutical form;

    XIII.- Active pharmaceutical ingredient (API): a pharmaceutical ingredient that, when

    administered to a patient, acts as an active component, and can exert pharmacological activity or

    direct effect in the diagnosis, cure, treatment or prevention of a disease or even affect the

    structure and functioning of the human organism;

    XIV.- Complex matrix: one containing an undefined number of unmonitored substances,

    which cannot be obtained without the presence of the analyte;

    XV.- Matrix: composition that mimics the sample without the presence of the analyte;

    XVI.- Investigational product: experimental drug, placebo, active comparator or any other

    product to be used in the clinical trial;

    XVII.- Chromatographic purity: absence of interference in the chromatographic signal of

    the analyte;

    XVIII.- Peak purity: spectral homogeneity of a chromatographic peak, indicative of its

    chromatographic purity, and the criteria for concluding whether there is spectral homogeneity

    and the parameters adopted for the purity calculation are defined as previously established for

    the software used or by means of scientifically based evaluation technique;

    XIX.- Validation report: document in which the validation procedures, records, results and

    evaluation are consolidated and summarized;

    XX.- Revalidation of analytical method: partial or complete re-validation of an analytical

    method to ensure that it continues to comply with established requirements;

    XXI.- Chemical Reference Substance (CRS): substance or mixture of high purity chemical

    or biological substances, which has been carefully characterized to ensure its identity, quality,

    content and potency, including a characterized chemical reference substance and a

    pharmacopoeial chemical reference substance;

    XXII.- Characterized Chemical Reference Substance (CCRS): Substance or mixture of

    chemical or biological substances in which identity, quality, purity, content and strength have

    been ensured by a characterization process;

    XXIII.- Pharmacopoeial Chemical Reference Substance (PCRS): Substance or mixture of

    chemical or biological substances established and distributed by official compendia recognized

    by ANVISA;

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    XXIV.- Working Chemical Reference Substance (WCRS): Substance or mixture of

    chemical or biological substances used in the laboratory routine, standardized from a

    pharmacopoeial reference chemical or, in the absence thereof, from a characterized chemical

    reference substance, traceable to the CRS used for its standardization;

    XXV.- Method transfer: a documented process that qualifies a laboratory (receiving unit)

    for the use of an analytical method from another laboratory (transfer unit), ensuring that the

    receiving unit is knowledgeable and capable of performing the analytical method according to

    intended purpose;

    XXVI.- Analytical validation: is the systematic evaluation of a method by means of

    experimental tests in order to confirm and provide objective evidence that the specific

    requirements for its intended use are met;

    XXVII.- Partial validation: demonstration, through some validation parameters, that the

    previously validated analytical method has the necessary characteristics to obtain results with

    the required quality, under the conditions in which it is practiced; and

    XXVIII.- System suitability: Procedure to be carried out prior to an analytical run to

    demonstrate that the system is fit for the intended use, and the parameters of this procedure

    must be defined during the development and validation of the method.

    CHAPTER II

    GENERAL PROVISIONS

    Art. 4 Validation shall demonstrate that the analytical method produces reliable results and

    is appropriate to the intended purpose, in a documented manner and through objective criteria.

    Art. 5 The use of analytical method not described in an official compendium recognized by

    ANVISA requires the conduction of an analytical validation, according to the parameters

    established in this resolution, taking into account the technical-operational conditions.

    Art. 6 The typical parameters to be considered for validation depend on the test to be

    carried out and are set out in Table 1 of Annex I.

    Art. 7 Compendial analytical methods must have their demonstrated suitability for the

    intended use, under the operational conditions of the laboratory, by means of the presentation of

    a partial validation study.

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    Sole paragraph. The provisions of the caput exclude basic general compendial methods

    such as pH measurement, drying loss, sulfated ash, moisture, disintegration, among others, and

    the analytical methods described in individual compendial monographs of non-active

    pharmaceutical ingredients.

    Art. 8 Partial validationvalidations shall evaluate, at least, the parameters of precision,

    accuracy and selectivity.

    Paragraph 1 - In case of analytical methods for quantification of impurities, partial

    validation shall include the quantification limit.

    Paragraph 2 - In case of a limit test, instead of the parameters of the caput, the selectivity

    and detection limit parameters must be evaluated.

    Art. 9 In case of transfer of method between laboratories, this will be considered as

    validated, provided that a partial validation study is carried out in the premises of the receiving

    laboratory.

    Paragraph 1 - Transfer of method between laboratories with the same quality management

    system can be carried out by means of a partial validation study, according to art. 8, or by

    reproducibility assessment.

    Paragraph 2 - Another approach may be accepted, based on justification and presentation

    of protocol and transfer report, based on risk analysis and considering previous experience,

    knowledge of the receiving unit, complexity of product and method and specifications, as well

    as other applicable relevant aspects.

    Paragraph 3 - If the transfer also uses comparative tests, the similarity in the results should

    be proven by means of statistical tool.

    Paragraph 4 - The method transfer documentation shall be presented containing the copy of

    the validation report of the transferred method as proof that it was originally validated in

    accordance with specific standards and regulations approved/endorsed by ANVISA.

    Paragraph 5 - In the case of transfer of methods already approved by ANVISA a copy of

    the approved validation report or an indication of the file number of the petition in which the

    final version of that report has been filed.

    Art. 10. A revalidation of analytical method may consider the following circumstances:

    I.- changes in the synthesis or obtainment of API;

    II.- changes in product composition;

    III.- changes in analytical method; and

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    IV.- other changes that could significantly impact the validated method.

    Sole paragraph. The validation parameters to be evaluated depend on the nature of the

    changes made.

    Art. 11 System verification shall be carried out at each analytical run.

    Art. 12 The validation and partial validation documents submitted shall describe the

    procedures, analytical parameters, acceptance criteria and results, with sufficient detail to enable

    their reproduction and, where applicable, their statistical evaluation.

    Art. 13 The validation report to be filed, according to registration and post-registration

    resolutions, shall contain the data and calculations obtained during the conduction of the

    analytical validation, as well as the statistical approach used for data evaluation.

    Paragraph 1 - The raw data related to the selectivity parameter should be part of the report

    mentioned in the caput.

    Paragraph 2 - The raw data related to other parameters must be available at the company

    for evaluation at ANVISA’s request.

    CHAPTER III

    CHEMICAL REFERENCE SUBSTANCES

    Art. 14 In the validation of analytical methods Pharmacopoeial Chemical Reference

    Substance (PCRS) should be preferably used, officially designated by the Brazilian

    Pharmacopoeia, or by other compendia officially recognized by ANVISA.

    Paragraph 1. The use of Characterized Chemical Reference Substance (CCRS) will be

    admitted by submitting a conclusive characterization report for the batch under study, including

    the technical reasons for choosing the tests used and the relevant raw data.

    Paragraph 2 - ANVISA and members of the Brazilian Health Surveillance System may

    request samples from the CRS for the purpose of evaluating the characterization process in the

    hypotheses of the previous paragraph and, when it is necessary to carry out a fiscal analysis, a

    sample of CRS should be provided for the purposes of realization of the necessary tests.

    Art. 15 The characterization report, depending on the analyte, shall contain the data

    obtained from techniques applicable to the characterization of each chemical such as

    thermogravimetry, melting point, differential scanning calorimetry, infrared spectroscopy, mass

    spectrometry, nuclear magnetic resonance, elemental analysis (carbon/hydrogen/nitrogen), X-

    ray diffraction, optical rotation, chromatographic methods, among others.

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    Paragraph 1 - In addition to the characterization data, the following information should be

    included in the report:

    I.- number and shelf-life of the batch of the substance used in the characterization;

    II.- Brazilian non-proprietary name or international non-proprietary name;

    III.- CAS number;

    IV.- chemical name;

    V.- synonymy;

    VI.- molecular and structural formula;

    VII.- molecular weight;

    VIII.- physical form;

    IX.- physicochemical properties;

    X.- impurity profile;

    XI.- handling and conservation care, and

    XII.- Analytical report proving the identity, content and validity of CRS.

    Paragraph 2 - For biological products, the characterization of the reference

    material/standard shall be carried out using appropriate state of the art methods.

    Art. 16 For medical gases, analytical instrument verification and analytical determinations

    shall be conducted using traceable reference materials distributed by metrology institutes or by

    bodies recognized as producers of certified reference materials.

    Sole paragraph. In the absence of reference materials, internal standards produced

    according to guides and bibliographic records may be used.

    Article 17 For biological products, the terms material/standard substitutes the term

    chemical substance in the definitions of CRS, PCRS, CCRS and WCRS.

    Art. 18 The use of WCRS for the purpose of analytical method validation is not allowed.

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    CHAPTER IV

    ANALYTICAL VALIDATION PARAMETERS

    Section I

    Selectivity

    Art. 19 The selectivity of the analytical method should be demonstrated by its ability to

    identify or quantify the analyte of interest unambiguously in the presence of components that

    may be present in the sample such as impurities, diluents and matrix components.

    Sole paragraph. In the case of chromatographic methods, the chromatographic purity of the

    analyte signal must be demonstrated, except for biological products.

    Art. 20 In identification methods, the ability to obtain a positive result for the sample

    containing the analyte and a negative result for other substances in the sample.

    Paragraph 1 - CRS shall be used in comparison with the response obtained for the analyte

    under Chapter III.

    Paragraph 2 - In order to demonstrate the selectivity of the identification methods, the tests

    shall be applied to substances structurally similar to the analyte, with the acceptance criterion

    being negative.

    Paragraph 3 - For active pharmaceutical ingredients derived from plants and medicinal

    products containing them, the ability of the method of distinguishing material of interest from

    other similar plant species, particularly those which may be present as adulterants or

    substituents.

    Paragraph 4 - To achieve the required level of selectivity, a combination of two or more

    analytical methods of identification may be required.

    Art. 21 For quantitative methods and limiting tests, the selectivity shall be demonstrated by

    verifying that the analytical response is exclusively due to the analyte, without interference from

    the diluent, matrix, impurities or degradation products.

    Paragraph 1 - To demonstrate the absence of interference degradation products, it is

    necessary to expose the sample to degradation conditions in a wide pH range, oxidation, heat

    and light.

    Paragraph 2 - The following cases shall be exempted from the demonstration described in

    Paragraph 1:

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    I.- products for which it has already been shown to comply with the resolution that

    establishes parameters for the notification, identification and qualification of degradation

    products in drugs.

    II.- performance methods;

    III.- non-chromatographic methods.

    Paragraph 3 - The use of a method with technical limitation for selectivity, under the caput,

    is only accepted by technical justification and joint application of another complementary

    method.

    Art. 22 For medicinal gases, selectivity should be demonstrated by comparing the result of

    reading the sample with the reading response of CRS under Chapter III.

    Sole paragraph. The maximum amount of possible interference should be justified.

    Section II

    Linearity

    Art. 23 The linearity of a method must be demonstrated by its ability to obtain analytical

    responses directly proportional to the concentration of an analyte in a sample.

    Art. 24 A linear relationship must be evaluated throughout the established range for the

    method.

    Art. 25 For the establishment of linearity, at least 5 (five) different concentrations of CRS

    must be used for solutions prepared in at least triplicate.

    Sole paragraph. The solutions used for linearity evaluation must be prepared

    independently, and diluted solutions of the same CRS stock solution may be used.

    Art. 26 All calculations for the assessment of linearity shall be performed from actual

    concentration data and individual analytical responses.

    Art. 27 For the assessment of linearity, the following data shall be presented:

    I.- graphical representation of the responses as a function of analyte concentration;

    II.- graph of dispersion of the residues, accompanied by its statistical evaluation;

    III.- equation of the regression line of y in x, estimated by least squares method;

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    IV.- evaluation of the linear association between the variables by means of the coefficients

    of correlation (r) and determination (r²);

    V.- significance evaluation of the angular coefficient.

    Paragraph 1 - Homoscedasticity of the data shall be investigated for use of the appropriate

    model.

    Paragraph 2 - Statistical tests should be used a significance level of 5% (five percent).

    Paragraph 3 - The correlation coefficient should be above 0.990.

    Paragraph 4 - The angular coefficient shall be significantly different from zero.

    Section III

    Matrix Effect

    Art. 28 The provisions of this section apply to complex matrices.

    Art. 29 The matrix effect must be determined by comparing the angular coefficients of the

    calibration curves constructed with the CRS of the analyte in solvent and with the sample

    fortified with the CRS of the analyte.

    Sole paragraph. The curves shall be established in the same way as in the linearity for the

    same concentration levels, using at least 5 (five) different concentrations in at least triplicate.

    Art. 30 The parallelism of the lines is indicative of interference absence of the matrix

    constituents and its demonstration must be carried out by means of adequate statistical

    evaluation.

    Sole paragraph. The level of significance of 5% (five percent) should be adopted in the

    hypothesis test.

    Section IV

    Working range

    Art. 31 The working range must be established from the linearity studies, together with the

    results of precision and accuracy, being dependent on the intended application.

    Art. 32 The following working ranges should be considered:

    I.- for content: from 80% (eighty percent) to 120% (one hundred and twenty percent);

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    II.- for uniformity of content: from 70% (seventy percent) to 130% (center and thirty

    percent);

    III.- for dissolution test: from -20% (minus twenty percent) of the lowest expected

    concentration to + 20% (plus twenty percent) of the highest concentration expected from the

    dissolution profile; and

    IV.- for determination of impurities: from the limit of quantification up to 120% (one

    hundred and twenty percent) of the concentration at the limit of the specification of each

    individual impurity;

    V.- for simultaneous determination of content and impurities by the area normalization

    procedure: from the limit of quantification (LQ) up to 120% (one hundred and twenty percent)

    of the expected concentration of the active substance.

    Paragraph 1 - Larger working ranges than those defined in the caput may be used if

    technically justified.

    Paragraph 2 - For medicinal gases, alternate working ranges shall be accepted provided that

    the approach for the choice of range is justified.

    Section V

    Precision

    Art. 33 The precision shall evaluate the proximity between the results obtained by means of

    tests with samples prepared as described in the analytical method to be validated.

    Art. 34 Precision shall be expressed by means of repeatability, intermediate precision or

    reproducibility.

    Art. 35 Precision must be demonstrated by scattering the results, calculating the relative

    standard deviation (RSD) of the series of measurements according to the formula "RSD = (SD /

    DAC) X100", where SD is the standard deviation and DAC, determined average concentration.

    Art. 36 Samples for precision evaluation shall be prepared independently from the

    beginning of the procedure described in method.

    Sole paragraph. In the case of solid and semi-solid samples, diluted solutions of the same

    stock solution are not acceptable.

    Art. 37 Where the precision assessment involves contamination of the matrix with a very

    low-quantity substance which makes direct weighing impossible, a concentrated solution of the

    substance may be used following the procedure described in the analytical method for extraction

    and dilution of the sample.

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    Paragraph 1 - In the case of known impurities absent or present in concentrations below the

    specification limit in the sample, the sample shall be fortified with known concentrations of the

    impurities standard.

    Paragraph 2 - In the case of unknown impurities, the sample shall be evaluated using the

    response of the active added to the matrix in the concentration corresponding to the limit of the

    specification established for the impurity, provided that the same response factor for impurity

    and for the active.

    Art. 38 The determination of repeatability shall comply with the following criteria:

    I - evaluate the samples under the same operating conditions, same analyst and same

    instrumentation, in a single analytical run.

    II - using at least 9 (nine) determinations, taking into account the linear interval of the

    analytical method, i.e. 3 (three) concentrations: low, medium and high, with 3 replicates at each

    level or 6 (six) replicates at 100% (one hundred percent) of the test concentration individually

    prepared.

    Art. 39 The acceptance criteria shall be defined and justified in accordance with the

    following aspects:

    I - purpose of the method;

    II - intrinsic variability of the method;

    III – work concentration; and

    IV – sample analyte concentration in the sample.

    Art. 40 The determination of intermediate precision shall meet the following criteria:

    I.- express the proximity between the results obtained from the analysis of the same

    sample, in the same laboratory, on at least two different days, performed by different operators;

    and

    II. - contemplating the same concentrations and the same number of determinations

    described in the repeatability evaluation.

    Art. 41 Reproducibility should be obtained by means of the proximity of the results

    obtained in different laboratories.

    Paragraph 1 - Reproducibility is applicable in collaborative studies or in the standardization

    of analytical methods for inclusion in official compendia, by appropriate statistical tests.

    Paragraph 2 - The acceptance criteria for the relative standard deviation must be justified as

    recommended in art. 39.

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    Section VI

    Accuracy

    Art. 42 The accuracy of an analytical method must be obtained by the degree of agreement

    between the individual results of the method under study in relation to a value accepted as true.

    Art. 43 The accuracy must be verified from at least 9 (nine) determinations, considering the

    linear interval of the analytical method, i.e. 3 (three) concentrations: low, medium and high,

    with 3 (three) replicates at each level.

    Art. 44 Samples for evaluation of accuracy should be prepared independently and diluted

    solutions of the same CRS stock solution may be used.

    Art. 45 For the determination of accuracy, the most appropriate approach must be used,

    according to the analytical method under study:

    I.- for API:

    a) apply the proposed method using a substance with known purity (CRS);

    b) compare the results with those resulting from a second validated method, whose

    accuracy has been established; or

    c) in the case of complex matrix analyte, perform analysis by the CRS addition method in

    which known quantities of CRS are added to the sample.

    II.- for finished product:

    a) apply the proposed method in the analysis of a sample, in which known quantity of CRS

    was added to the matrix;

    b) in the unavailability of samples of all components of the drug, analysis may be carried

    out by the CRS addition method in which known amounts of CRS are added to the solution of

    the finished product; or

    c) compare the results obtained with those resulting from a second validated method.

    III.- for impurities:

    a) apply the standard addition method in which known amounts of impurities or

    degradation products are added to the sample;

    b) in the unavailability of samples of certain impurities or degradation products, a

    comparison of the results obtained with a second validated method and the use of the response

    factor relative to the API;

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    c) for unknown impurities, accuracy shall be assessed by comparing the API’s CRS or

    known impurity response, according to the proposed method, in a concentration range that

    contemplates the working range of the method, provided that the same response factor is

    considered.

    Sole paragraph. In all cases, the method of calculating analyte concentrations shall be the

    same as that described in the concerned analytical method.

    Art. 46 The accuracy shall be expressed by the percentage recovery ratio of the known

    analyte added to the sample or by the ratio between the experimentally determined mean

    concentration and the corresponding theoretical concentration given by formula 1 of Annex II.

    Sole paragraph. When accuracy is determined from a previously validated method, the

    concentration of analyte determined by it should be considered, replacing the term "theoretical

    concentration".

    Art. 47 The standard relative deviation (SRD) should be calculated for each concentration.

    Art. 48 The acceptance criteria for percentages of recovery and relative standard deviation

    obtained must be justified according to the criteria recommended in art. 39.

    Section VII

    Detection Limit

    Art. 49 Detection limit shall be demonstrated by obtaining the smallest amount of analyte

    presented in a sample that can be detected but not necessarily quantified under the established

    experimental conditions.

    Art. 50 The determination of the detection limit can be carried out by means of a visual

    method, the signal-to-noise ratio, based on the determination of the blank or parameters of the

    calibration curve, considering the particularities of the analytical method used.

    Art. 51 For visual methods, the detection limit is determined by the lower concentration for

    which the visual effect expected.

    Art. 52 For instrumental methods, the limit of detection can be determined by the signal-to-

    noise ratio.

    Paragraph 1 - The method used to determine the signal-to-noise ratio shall be described and

    justified.

    Paragraph 2 - The signal-to-noise ratio must be greater than or equal to 2:1.

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    Art. 53 For the determination based on analytical curve parameters, the detection limit may

    be calculated by formula 2 of Annex II.

    Art. 54 In cases where an estimated value for the detection limit is obtained by calculation

    or extrapolation, this estimate must be confirmed according to art. 52.

    Section VIII

    Quantification Limit

    Art. 55 The quantification limit is the smallest amount of analyte in a sample that can be

    determined with acceptable precision and accuracy under established experimental conditions.

    Art. 56 The quantification limit shall be consistent with the impurity specification limit.

    Sole paragraph. For products suitable for the resolution setting out parameters for the

    notification, identification and qualification of degradation products in medicinal products, the

    quantification limit must be less than or equal to the notification limit.

    Art. 57 For the determination of this parameter the same procedure described in art. 53, and

    the signal-to-noise ratio must be at least 10:1.

    Art. 58 For the determination based on analytical curve parameters, the quantification limit

    may be calculated by formula 3 of Annex II.

    Art. 59 In cases where an estimated value for the quantification limit is obtained by

    calculation or extrapolation, this estimate must be confirmed according to art. 57.

    Art. 60 Precision and accuracy at the concentrations corresponding to the quantification

    limit must be tested.

    Section IX

    Robustness

    Art. 61 Robustness is a parameter typically carried out in the development of the analytical

    method that indicates its ability to withstand small and deliberate variations of analytical

    conditions.

    Sole paragraph. If the method is susceptible to variations in analytical conditions, these

    should be controlled by precautions described in the method.

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    Art. 62 In the case of quantitative methods, the impact of proposed variations on the results

    obtained shall be evaluated using the same criteria used for accuracy.

    Art. 63 In the case of qualitative methods, it should be checked whether the proposed

    variations interfere in the analytical response.

    Art. 64 Compliance to the system verification characteristics shall be demonstrated. Art. 65

    The assessment of the parameters described in Table 1 of Annex III shall be included in the

    validation report.

    Paragraph 1 - Parameters that are considered relevant to the result, according to the

    characteristics of the method, should be additionally evaluated.

    Paragraph 2 - The absence of the evaluation of any of the variations shall be justified.

    CHAPTER V

    TRANSITIONAL PROVISIONS

    Art. 66 Analytical method validations will be accepted in compliance with Resolution RE

    No. 899/2003, since it has been finalized prior to the validity of this resolution and the petitions

    containing them have been filed within 550 (five hundred and fifty) calendar days after the

    validity of this resolution.

    Paragraph 1 - In case of the need to execute and re-present one or more validation

    parameters, as long as a new validation is not required, the company may follow Resolution RE

    No. 899, dated May 29th, 2003.

    Paragraph 2 - In case of need to execute and submit a new validation, the company must

    follow this resolution.

    Paragraph 3 - After the deadline established in the caput for investigational products whose

    validation of the analytical method used in clinical development has been initiated before the

    validity of this regulation, the analytical validations carried out according to Resolution RE No.

    899 dated May 29th, 2003 will be accepted at the time of registration.

    CHAPTER VI

    FINAL PROVISIONS

    Art. 67 Additional documentation and tests may be requested at any time by ANVISA.

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    Art. 68 All relevant data obtained during the conduction of the analytical validation, as

    well as the formulas used for calculation, must be filed together with the petition of interest for

    the evaluation of ANVISA.

    Art. 69 Failure to comply with the provisions contained in this resolution constitutes a

    sanitary infraction, pursuant to Law No. 6.437, dated August 20th, 1977, without prejudice to the

    applicable civil and criminal liability.

    Art. 70 Resolution RE No. 899, dated May 29th, 2003, Subsection XXXI of art. 1, sole

    paragraph of art. 11 and Annex I of Resolution RDC No. 31 dated August 11th, 2010 are

    revoked.

    Art. 71 This Resolution shall enter into force within a period of 180 (one hundred and

    eighty) calendar days from the date of its publication.

    JARBAS BARBOSA DA SILVA JR.

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    ANNEX I

    Table 1. Parameters to be considered in analytical validation.

    Evaluated Parameter Identification

    Impurity

    Test Assay

    - dissolution (quantification)

    - content uniformity

    - strength Quantitative Limit Test

    Accuracy No Yes No Yes

    Repeatability precision No Yes No Yes

    Intermediate Precision No Yes (1) No Yes (1)

    Selectivity (2) Yes Yes Yes Yes

    Detection Limit No No (3) Yes No

    Quantification Limit No Yes No No (3)

    Linearity No Yes No Yes

    Interval No Yes No Yes

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    (1) In cases where reproducibility has been conducted, it is not necessary to conduct intermediate

    precision.

    (2) In the case of identification tests, it may be necessary to combine two or more analytical

    procedures to achieve the required level of discrimination.

    (3) May be necessary in some cases.

    ANNEX II

    Formula 1. Accuracy calculation.

    Experimental mean concentration

    Recovery = __________________________________________________ x 100

    Theoretical concentration

    or

    CA (added sample) - CA (sample)

    Recovery = _________________________________________________ x 100

    CTA

    Where: CA is the experimental concentration of the analyte and CTA is the theoretical

    concentration of the added analyte

    LD = 3.3. σ

    IC

    Where: IC is the slope of the calibration curve, σ is the standard deviation and can be

    obtained in 3 ways:

    I - from the standard deviation of the Y-axis intercept of at least 3 calibration curves

    constructed containing analyte concentrations close to the assumed detection limit;

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    II - from the residual standard deviation of the regression line;

    III - from the estimation of noise coming from the analysis of an appropriate number of

    blank samples.

    Formula 3. Calculation of quantification limit.

    LQ = 10. σ

    IC

    Where: IC is the slope of the calibration curve,curve; σ is the standard deviation and can be

    obtained in 3 ways:

    I - from the standard deviation of the Y-axis intercept of at least 3 calibration curves

    constructed containing analyte concentrations close to the assumed detection limit;

    II - from the residual standard deviation of the regression line;

    III - from the estimation of noise coming from the analysis of an appropriate number of

    blank samples.

    ANNEX III

    Table 1. Conditions for the evaluation of the method robustness

    Sample preparation

    Stability of analytical solutions

    Extraction time

    Filter compatibility

    Spectrophotometry

    pH variation of the solution

    Different solvent batches or manufacturers

    Liquid Chromatography

    pH variation of the mobile phase

    Different column batches or manufacturers

    Temperature

    Mobile phase flow

  • Ministry of Health – MS

    Brazilian Health Surveillance Agency - ANVISA

    This text does not replace the text(s) published in the Brazilian Official Gazette

    Gas Chromatography

    Different column batches or manufacturers

    Temperature

    Carrier gas velocity

    Other analytical techniques The variations to be tested should be evaluated critically and

    their results should be presented