analysis, purification and validation of target compounds
TRANSCRIPT
Page 1
Analysis, purification and validation of target
compounds
Covering the complete compound generation workflow with Agilent 1200 Series LC/MS solutions
Udo Huber Product Specialist
October 31, 2007
Page 2
Overview
• The compound generation workflow
• Analytical and preparative tasks in the compound generation workflow
• Agilent solutions for:
• Walk-up systems
• Fast result generation
• High sample throughput
• Sample purification
• Compound confirmation
Page 3
From building blocks to hits – the compound generation workflow
Buildingblocks HitTarget
compounds
Compoundlibrary
Synthesis Screening
Storage
Page 4
Analytical and preparative tasks
Buildingblocks HitTarget
compounds
Compoundlibrary
Synthesis
Storage
QC PurificationControl ofreaction outcome
Final QCCharacterization
Identity & purity
Prep LC/MS
Routine randominspection
Purification
Screening
Page 5
Identity & purity
Standard can be any compound with nitrogen (adjacent
nitrogens)
Without standard compound 3D calibration is
required (inverse gradient)
Requires standard of compoundAbsolute
Only compounds with nitrogen are
detected
Non-linear response, response
depends on composition of mobile phase
Area % report, response is compound
dependent, can be zero
Relative
CLNDELSDUV (DAD)Purity
Standard can be any compound with nitrogen (adjacent
nitrogens)
Without standard compound 3D calibration is
required (inverse gradient)
Requires standard of compoundAbsolute
Only compounds with nitrogen are
detected
Non-linear response, response
depends on composition of mobile phase
Area % report, response is compound
dependent, can be zero
Relative
CLNDELSDUV (DAD)Purity
Identity: SQ mass + knowledge of compound structureAccurate mass (TOF), NMR, IR, etc. for final QC of hits
Page 6
Compound generation – facts
Medium or large numbers of compoundscertain level of automation is required
Instrument users are often synthetic chemistsnot chromatographers, system must be simple to use
Structure and mass of building blocks/compounds known
Non-regulated environment
Page 7
Analysis of compounds – the basic solution
The workhorse: 1200 Series LC system with 6100 SQAgilent ChemStation
• Binary or quaternary pump with degasser
• Standard or well-plate autosampler
• Column compartment• DAD (MWD)• 6110, 6120 or 6130 SQ
Page 8
The Agilent 6100 Series single quadruple
Research GradeHigh Throughput6140
m/z 10 – 1350 mass range1 pg sensitivity spec (50 pg scan)Native 2500 u/s scan speed10000 u/sec scan speed1.4 MHz quadrupolePOS/NEG and divert valveMatched to 1200 RRLC performance
Entry Level6110
m/z 10-1500 mass range10 pg sensitivity2500 u/s scan speed1 MHz quadrupoleOptional Automated Divert Valve
Mid Range6120
m/z 10-1500 mass range10 pg sensitivity2500 u/s scan speed1 MHz quadrupoleAutomated Divert Valve (MMS)POS/NEG switching
Research Grade6130
m/z 2 – 3000 mass range1 pg sensitivity spec (50 pg scan)Native 2500 u/s scan speedFast scanning mode 5250 u/sec1.0 MHz quadrupolePOS/NEG and divert valve
Upgrade kitG2735F
Upgrade kitG2735E
Upgrade kitG2735G
Page 9
Analysis of compounds – Easy Access
• Easy Access Software allows chemists untrained in detailed operation of the instrument data system and possibly unskilled as chromatographers to run their own samples safely.
• Same software interface for LC, LC/MSD and LC TOF.
• It is not a software for “inexperienced users”! It is a software for a system-user/system-administrator environment!
“The system users are chemists who need results, not chromatographers who want to play with the instrument or tweak the software!”
Page 11
External tray for standard autosampler SL
Vials discarded here on completion (with optional delay)
Chemists put up to 17 vials in external tray. Sampler arm moves them inside.
Page 14
Easy Access in the compound generation workflow
Buildingblocks HitTarget
compounds
Compoundlibrary
Synthesis Screening
Storage
Routine randominspection
QC PurificationControl ofreaction outcome
Final QCCharacterization
Purification
Page 15
Analysis of compounds – fast chromatography
Medium or large numbers of compoundsMore compounds require faster analysis:The 1200 Series RRLC system with 6140 SQ MS
• Binary pump SL with degasser
• High-performance autosampler SL
• Column compartment SL• Sub-2-µm particle column• DAD SL (MWD SL)• 6140 SQ
Page 16
Analysis of compounds – fast chromatography
The 1200 Series RRLC system with 6210/6220 TOF MSAdditional accurate mass information gives higher confidence of target compound
• Binary pump SL with degasser
• High-performance autosampler SL
• Column compartment SL• Sub-2-µm particle column• DAD SL (MWD SL)• 6210/6220 TOF
Page 17
Analysis of compounds – fast chromatography
In a walk-up environment the number of samples is usually low.With fast chromatography chemists do not have to wait a long time to get their results and can continue work earlier:The 1200 Series RRLC system with 6140 SQ MS
• Binary pump SL with degasser
• Standard autosampler SL with external tray
• Column compartment SL• Sub-2-µm particle column• DAD SL (MWD SL)• 6140 SQ• Easy Access software
Page 18
Analysis of compounds – matching capacity to throughput
Medium or large numbers of compoundsMore compounds require higher sample capacity:The 1200 Series (RRLC) system with CTC PAL
• Binary pump SL with degasser
• HTS or HTC PALwith 600 bar injection valve
• Column compartment SL• Sub-2-µm particle column• DAD SL(MWD SL)• 6140 SQNow compatible with Easy-Access
Page 19
Analysis of compounds – matching capacity to throughputMedium or large numbers of compoundsMore compounds require higher sample capacity:The 1200 Series (RRLC) system with sample capacity extension
• Binary pump SL with degasser
• High-performance autosampler SLwith sample capacity extension
• Column compartment SL• Sub-2-µm particle column• DAD SL (MWD SL)• 6140 SQ
Page 20
High throughput and RRLC in the compound generation workflow
Buildingblocks HitTarget
compounds
Compoundlibrary
Synthesis Screening
Storage
Routine randominspection
QC PurificationControl ofreaction outcome
Final QCCharacterization
Purification
Page 20
High throughput and RRLC in the compound generation workflow
Buildingblocks HitTarget
compounds
Compoundlibrary
Synthesis Screening
Storage
Routine randominspection
QC PurificationControl ofreaction outcome
Final QCCharacterization
Purification
Page 21
The Agilent 1200 Series purification system
• 2 preparative pumps• Dual-loop autosampler• ZORBAX preparative
column• MWD (DAD)• 1-3 fraction collector PS• 6110, 6120 SQ
with active splitter
Preparative HPLC is the method of choice for compound purification in compound generationThe 1200 Series purification system with 6100 SQ MS
Page 22
Mass-based fraction collection
Totals formulas of target compound known
Mass of target compound known
Mass-based fraction collection
Advantages:• Small number of fractions per
sample, often only one fraction per sample
• Only target compound collected (or compounds with the same mass).
• No need to analyze fractions to determine which one contains the compound of interest.
Disadvantages:• Target mass must be known• Purity of fractions slightly lower
than for UV based collection.• System is more expensive• Seems to be more complicated
(splitter, etc.)
Page 23
Sample purification
Buildingblocks HitTarget
compounds
Compoundlibrary
Synthesis Screening
Storage
QC PurificationControl ofreaction outcome
Final QCCharacterization
Routine randominspection
Purification
Page 24
Final characterization of compounds
High-resolution LC gives confidence to separate most compounds.High mass accuracy of TOF MS gives high confidence of compound identity.The 1200 Series RRLC system with 6210/6220 TOF MS
• Binary pump SL with degasser
• High-performance autosampler SL
• Column compartment SL• Sub-2-µm particle column• DAD SL (MWD SL)• 6210/6220 TOF
Page 25
How much accuracy is needed?
Example:Reserpine (C33H40N2O9) has a protonated ion at 609.28066Single quad reports mass to +/- 0.1 = 165 ppm
Number of possible formulas using only C, H, O & N:165 ppm 20910 ppm 135 ppm 73 ppm 42 ppm 2
+ information about totals formula
Page 26
Typical performance of TOF instrumentMass error histogram of the analysis of 140 real screening compounds under ultra-fast LC conditions (90 s cycle time 1000 samples/day):
0
5
10
15
20
25
30-5
.0
-4.5
-4.0
-3.5
-3.0
-2.5
-2.0
-1.5
-1.0
-0.5 0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
91 % within ± 2.0 ppm
ppm
n pe
r 0.2
5 pp
m
71 % within ± 1.0 ppm
Page 27
Summary
• The compound generation workflow
• Analytical and preparative tasks in the compound generation workflow
• Agilent solutions:
• Easy Access software for walk-up systems
• RRLC for fast results or higher sample throughput
• Solutions for high sample capacity – HTC and HTS autosampler and sample capacity extension
• Preparative HPLC – Agilent purification system
• Reliable compound confirmation – Agilent TOF MS