![Page 1: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/1.jpg)
An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance
David FollowillIROC Houston QA Center
![Page 2: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/2.jpg)
Outline of Presentation
• Definition of the problem• Results of audits• Possible solutions• Resources to help
![Page 3: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/3.jpg)
Learning Objectives
• Recognize that small field dosimetry is not easy and that there are tools to help you be as accurate as possible.
![Page 4: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/4.jpg)
Sometimes we all like to believe that all across the kingdom it’s bright and merry!
![Page 5: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/5.jpg)
Ladies and LordsThere is a scourge across the Kingdom
Small field dosimetry
![Page 6: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/6.jpg)
What is the truth?
From Das et al 2000
![Page 7: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/7.jpg)
Recommendation:
The unshielded stereotactic diode and micro-chambers are the detectors of choice formeasurements in radiosurgery beams. However, werecommend comparing the measurements from atleast two different detectors listed in Table 2a, andafter applying corrections taking their average.
TG-155Small Field Dosimetry
![Page 8: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/8.jpg)
TomoTherapy results (since Aug 2010)
0
20
40
60
80
100
120
0.9 0.91 0.92 0.93 0.94 0.95 0.96 0.97 0.98 0.99 1 1.01 1.02 1.03 1.04 1.05 1.06 1.07 1.08 1.09 1.1
# re
sults
TLD/Institution
Average: 0.970 +/- 0.024552 irradiations
![Page 9: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/9.jpg)
But the manufacturer told me it would work, the manufacturer set the value or I sent my own measured data to the TPS company and they modeled my beams for me.
1. Accuray calibrates all of the tomotherapy units using the Exradin A1SL chamber and the user monitors constancy.
• Correction factor of 0.980 for A1SL based on Gago-Arias et al, Med Phys, 2012
![Page 10: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/10.jpg)
Manufacturer modeling of small field dataYou make the measurementsYou send the measurementsThey create the beam model
BUTThey assume you used the correct dosimeter
They assume you made measurements correctlyThey do not know to apply additional corr. factors
RESULTIncorrect Small Beam Model
![Page 11: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/11.jpg)
Francescon et al2011 data
Small Field Dosimetry Corrections
Situation is even worse if you consider using field sizes less then 0.5 x 0.5 cm2
![Page 12: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/12.jpg)
Field Size (cm x cm)
Detectors 0.5x0.5 0.75x0.75 1.0x1.0 1.25x1.25 1.5x1.5 3.0x3.0
PTW 60012-60017 0.968±0.003 0.984±0.002 0.995±0.001 1.001±0.001 1.006±0.001 1.013±0.001
Sun Nuclear Dedge 0.932±0.003 0.951±0.002 0.967±0.001 0.978±0.003 0.986±0.002 1.001±0.002
IBA SFD 0.972±0.003 0.996±0.002 1.007±0.001 1.011±0.001 1.015±0.001 1.017±0.001
Exradin D1V 0.980±0.003 0.994±0.002 0.999±0.001 1.000±0.001 1.005±0.001 1.012±0.001
Exradin A16 1.112±0.018 1.044±0.007 1.020±0.001 1.007±0.001 1.002±0.001 0.999±0.001
PTW PinPoint 31014 1.128±0.018 1.053±0.007 1.024±0.002 1.010±0.001 1.005±0.001 1.000± 0.001
PTW MicroLion 1.023±0.009 0.997±0.002 0.993±0.001 0.992±0.001 0.994±0.001 0.998±0.001
Exradin W1 PSD 0.998±0.003 0.995±0.002 0.996±0.002 0.996±0.002 0.994±0.002 0.994±0.002
calculated as the mean over all the values obtained by changing the linacmodel, (PrimusTM Siemens 6 MV and Synergy® Elekta 6 MV), the radial FWHM andenergy of the electron source as presented by Francescon et al.97 These values canalso be used with caution for Varian 6 MV Linacs.
clin 10x10clin 10x10
f , fQ ,Qk
Unpublished data from P. Francescon
![Page 13: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/13.jpg)
Gamma Knife results (since Sep 2007)
0
10
20
30
40
50
60
0.90 0.91 0.92 0.93 0.94 0.95 0.96 0.97 0.98 0.99 1.00 1.01 1.02 1.03 1.04 1.05 1.06 1.07 1.08 1.09 1.10
# re
sults
TLD/Inst.
Average: 0.997 +/- 0.021266 results
![Page 14: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/14.jpg)
Depending on the dosimeter used, the magnitude of the additional correction, down to a 0.5 x 0.5
cm2 field size can be as much as:
6%91%1%2%1% 1. 5%
2. 2%3. No correction needed4. 13%5. 8%
![Page 15: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/15.jpg)
The answer is:
13%
The magnitude of the additional correction, down to a 0.5 x 0.5 cm2 field size can be as much as:
Francescon et al. Medical Physics, Vol. 38, No. 12, December 2011
![Page 16: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/16.jpg)
Recommendation:
It is strongly recommended to independently verifydosimetric measurements in small fields, eitherthrough measurements carried out by a differentperson and/or though an independent external audit,such as that carried out by the RPC .
TG-155Small Field Dosimetry
![Page 17: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/17.jpg)
RPC Measurements during onsite visitsMeasurement configuration:
depth = 10 cm, 100 cm SSD, Exradin A16Secondary jaws kept at a 10 x 10
MLC @ 10 x 10MLC @ 6 x 6MLC @ 4 x 4MLC @ 3 x 3MLC @ 2 x 2
![Page 18: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/18.jpg)
The Problem is that our Dragon is very small!
![Page 19: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/19.jpg)
Tables of standard small field factors
Followill et al 2012 and also will be in TG-155
We even make mistakes
![Page 20: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/20.jpg)
Corrected Varian DataVarian 6 MV Varian 10 MV Varian 15 MV Varian 18 MV
Field Size
(cm x cm) RPC Institution RPC Institution RPC Institution RPC Institution
10 x 10 1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000
6 x 6 0.938 0.945 0.959 0.964 0.963 0.964 0.969 0.971(0.005) (0.008) (0.002) (0.003) (0.005) (0.007) (0.003) (0.006)
[0.8%] [0.5%] [0.4%] [0.5%](n=127) (n=22) (n=32) (n=41)
4 x 4 0.886 0.900 0.916 0.927 0.927 0.928 0.928 0.933(0.006) (0.013) (0.006) (0.010) (0.006) (0.009) (0.002) (0.009)
[1.8%] [1.2%] [0.7%] [0.9%](n=122) (n=25) (n=32) (n=37)
3 x 3 0.851 0.870 0.880 0.894 0.892 0.894 0.884 0.891(0.007) (0.012) (0.006) (0.008) (0.006) (0.012) (0.005) (0.019)
[2.4%] [1.6%] [0.9%] [1.4%](n=123) (n=25) (n=31) (n=41)
2 x 2 0.804 0.825 0.823 0.838 0.824 0.823 0.806 0.804(0.008) (0.020) (0.005) (0.015) (0.011) (0.040) (0.007) (0.020)
[2.9%] [2.0%] [2.2%] [1.6%](n=136) (n=23) (n=33) (n=40)
Varian 6 MV Varian 10 MV Varian 15 MV Varian 18 MV Field Size (cm x cm) RPC Institution RPC Institution RPC Institution RPC Institution
10 x 10 1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000
6 x 6 0.938
(0.937)* 0.945 0.959 0.964 0.963
(0.965) 0.964 0.969 0.971 (0.005) (0.008) (0.002) (0.003) (0.005) (0.007) (0.003) (0.006) [0.8%] [0.5%] [0.4%] [0.5%]
(n=127) (n=22) (n=32) (n=41)
4 x 4 0.886
(0.885) 0.900 0.916 0.927 0.927
(0.924) 0.928 0.928 0.933 (0.006) (0.013) (0.006) (0.010) (0.006) (0.009) (0.002) (0.009) [1.8%] [1.2%] [0.7%] [0.9%]
(n=122) (n=25) (n=32) (n=37)
3 x 3 0.851
(0.849) 0.870 0.880 0.894 0.892
(0.884) 0.894 0.884 0.891 (0.007) (0.012) (0.006) (0.008) (0.006) (0.012) (0.005) (0.019) [2.4%] [1.6%] [0.9%] [1.4%]
(n=123) (n=25) (n=31) (n=41)
2 x 2 0.804
(0.802) 0.825 0.823 0.838 0.824
(0.814) 0.823 0.806 0.804 (0.008) (0.020) (0.005) (0.015) (0.011) (0.040) (0.007) (0.020) [2.9%] [2.0%] [2.2%] [1.6%]
(n=136) (n=23) (n=33) (n=40)
![Page 21: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/21.jpg)
Discrepancies Regarding:Number of InstitutionsReceiving rec. (n = 206)
Review QA Program 152 (74%)Photon Field Size Dependence 138 (67%)
Wedge Factor (WF) 66 (32%)Off-axis Factors (OAF)/Beam symmetry 60 (29%)
Electron Calibration 35 (17%)Photon Depth Dose 33 (16%)Electron Depth Dose 25 (12%)Photon Calibration 16 (8%)
Discrepancies Discovered (Jan. ’05 – April ’13)
On-Site Dosimetry Review Audit
This is a beam measurement issue and TPS beam modeling challenge.
![Page 22: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/22.jpg)
There is guidance or possibly a fair maiden after the dragon is slain.
The bigger challenge is with Varian machines due to the tertiary collimation (MLC)• Varian/Pinnacle combination:
Followill et al, JACMP, vol. 13, No. 5, 2012
• Varian/Eclipse combination:Kron et al, Med Phys, vol 39 (2), 2012
![Page 23: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/23.jpg)
Huge Resource of Published data:
239 references
5 Tables
15 Figures
13 Conclusions and recommendations
TG-155Small Field Dosimetry
![Page 24: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/24.jpg)
New IROC Houston audit
OSLD
![Page 25: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/25.jpg)
Commissioning and Testing
Three dosimeters were used to estimate the dose for each field size: Exradin A16Exradin D1V diodeExradin D1V diodePTW CC04Exradin W1
![Page 26: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/26.jpg)
First Test Irradiation away from RPC
![Page 27: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/27.jpg)
All of the following are current methods to provide a verification of your small field
dosimetry data except:
2%4%86%8%1% 1. Use of published data
2. Your 2nd set of measurements3. Rely on manufacturer’s data4. Remote audit5. Indep. meas. from another physicist
![Page 28: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/28.jpg)
The answer is:
Reliance on the manufacturers data
All of the following are current methods to provide a verification of your small field dosimetry data except:
Ref: Das et al, AAPM TG 155 report, submitted to Med Phys
![Page 29: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/29.jpg)
• The challenge continues for the <2 x 2 cm2 field sizes
• The correct dosimeter is essential.
• There are independent audits and standard data to help.
• We will defeat thisdragon!
Summary|Conclusion
![Page 30: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/30.jpg)
Thank YouQuestions?
![Page 31: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/31.jpg)
ReferencesFrancescon et al. Medical Physics, Vol. 38, No. 12, December 2011
Das et al, AAPM TG 155 report, submitted to Med Phys
Followill et al. J Appl Clin Med Phys 13(5):282-9, 2012.
Followill DS: Erratum, J Appl Clin Med Phys 15 (2): 4757 3/2014.
![Page 32: An Independent Audit and Analysis of Small Field Dosimetry ...€¦ · such as that carried out by the RPC . TG-155 Small Field Dosimetry. RPC Measurements during onsite visits Measurement](https://reader034.vdocuments.us/reader034/viewer/2022051919/600bbfa3cf40ba13933a809e/html5/thumbnails/32.jpg)
AcknowledgementsWork supported by PHS grant CA10953 awarded by NCI, DHHS