OPERATION MANUALMODE 4/MODE STEST SYSTEM
1-2-1 Page 1
Aug 1/98
SECTION 2 - OPERATION
1. INSTALLATION
1.1 GENERAL
The Mode 4/Mode S Test System is a bench test set requir ing cont inuous ac power to operate. Mode 4/Mode S Test System instal lat ion includes connect ing the SI-1404 Test Auxi l iary with the ATC-1400A Transponder/DME Test Set, powering up the system and accomplishing an instal lat ion cal ibrat ion.
1.2 SAFETY PRECAUTIONS
The fol lowing safety precautions must be observed dur ing instal lat ion and operat ion. IFR Systems, Inc. assumes no l iabi l i ty for fai lure to comply with any safety precaution out l ined in this manual.
1.2.1 Complying with Instructions
Instal lat ion/operat ing personnel should not attempt to instal l or operate the Mode 4/ Mode S Test System without reading and complying with instruct ions contained in this manual. Al l procedures contained in this manual must be performed in exact sequence and manner descr ibed.
1.2.2 Grounding Equipment and Power Cord
WARNING: DO NOT USE A THREE-PRONG TO TWO-PRONG ADAPTER PLUG. DOING SO CREATES A SHOCK HAZARD BETWEEN THE CHASSIS AND ELECTRICAL GROUND.
The power cord, equipped with standard three-prong plug, must be connected to a proper ly grounded three-prong receptacle. I t is the customer 's responsibi l i ty to:
● Have a qual i f ied electr ic ian check receptacle(s) for proper grounding.
● Replace any standard two-prong receptacle(s) with proper ly grounded three-prong receptacle(s) .
1.2.3 Operating Safety
Due to potent ial for electr ical shock within test equipment, Test Set covers must not be removed by operators. The Instal lat ion Cal ibrat ion, Procedure, component replacement and internal adjustments must only be performed by qual i f ied service personnel. Refer to the Instal lat ion Cal ibrat ion Procedure or Mode 4/Mode S Test System Maintenance Manual (1002-2401-400) for maintenance procedures.
1.2.4 CAUTION and WARNING Labels
Exercise extreme care when performing operat ions preceded by a CAUTION or W ARNING label. CAUTION labels appear where possibi l i ty of damage to equipment exists. W ARNING labels denote condit ions where bodi ly injury or death may result .
OPERATION MANUALMODE 4/MODE STEST SYSTEM
1-2-1 Page 2
Aug 1/98
1.3 POWER REQUIREMENTS
The Mode 4/Mode S Test System operates over a voltage range of 100 to 120 VAC at 60 Hz or 220 to 240 VAC at 50 Hz. No internal wir ing changes are required before applying ac power to the Test Auxi l iary. The VOLTAGE SELECT Switch sett ing on SI-1404 rear panel must match the input l ine voltage. Instantaneous surge current at power up is ≤50 A. The specif ied fuse rat ings are l is ted in 1-2-1, Table 1.
CAUTION: FOR CONTINUOUS PROTECTION AGAINST FIRE, REPLACE ONLY W ITH FUSES OF THE SPECIFIED VOLTAGE AND CURRENT RATINGS.
INPUT
VOLTAGE
F1 AND F3
AC IN FUSES
F2 AND F4 LINE SUPPLY
FUSES 100 to
120 VAC 3.0 A, 250 V
Fast Blo (Type F) ( IFR PN:
5106-0300-600) (Bussman AGC3)
1.0 A, 250 V Fast Blo (Type F)
(IFR PN: 5106-4501-000)
(Bussman AGC1)220 to
240 VAC 3.0 A, 250 V
Fast Blo (Type F) ( IFR PN:
5106-0300-600) (Bussman AGC3)
0.5 A, 250 V Fast Blo (Type F)
(IFR PN: 5106-0000-016)
(Bussman AGC1/2)
Specif ied Fuse Ratings Table 1
1.4 INSTALLATION PROCEDURE
Refer to 1-2-1, Figure 1.
STEP PROCEDURE 1. Place the ATC-1400A on the bench.
2. Place the SI-1404 on top of the ATC-1400A.
3. Connect two 25-Pin D style cables:
● One to the SI-1404 IFR BUS Connector (J11) and ATC-1400A IFR BUS Connector (J5).
● One to the SI-1404 AUX BUS Connector (J10) and ATC-1400A AUXILIARY Connector (J6).
4. Connect three SMB-to-SMB coaxial cables:
● One to the SI-1404 RF LVL Connector (J13) and ATC-1400A RF LEVEL INPUT Connector (J4).
● One to the SI-1404 PPMG Connector (J14) and ATC-1400A EXTERNAL MEASUREMENT GATE Connector (J3).
● One to the SI-1404 DPSK OUT Connector (J15) and ATC-1400A DABS INPUT Connector (J1).
5. Connect ac power cable to the SI-1404 1400A POW ER Connector and ATC-1400A AC INPUT Connector.
OPERATION MANUALMODE 4/MODE STEST SYSTEM
1-2-1 Page 3
Aug 1/98
TYPE F
TYPE F
VOLTAGE SELECT
FE
F3
E S UF
3A @
110/230
FUS E
S U
VAC
F1
F4
115
USF E
.230
E S UF
FUS E
.
163 WATTS MAX
F2
E S UF
FUS E1400A POWER
1.0A @ 110VAC
0.5A @ 230VAC
02407007
87654321ONOFF
GPIB ADDRESSIEEE 488 - 1978
SH1
L4
PP0
AH1
LE0
DC1
T6
SR1
DT1
TE0
RL1
C0
DECODER
WIDE EQUALIZER
SELF-
INTERR
INST DIM
HI
NARROW
R/NAV
OFF
GEN
OFF
XMTR
LOW
J9 J10 J11 J22
CLOCK
FREQ
SERIAL
J1
J2
J3
J4J23
TEST
ON
OFF
1
0
1
0
1
0
1
0
1
0
A1A5
SYNC IN
EXT
PULSE/PROCESSOR
MECH ASSY,
EXT
SYNC OUTTRIG OUT
SCOPE
J7
MOD IN
EXT
J3
DPSK ADJRS-232
J12
PPMGR28
RF LVL
J13J14J15
DPSK OUT
J17
J23
VIDEO OUT
GEN OUT
PREPULSE
OUT
J8 J9
OUT
ATCRBS
J4 J5
J22
ANT B DISPARTY
OUT
J6
ANT B EXT
PRF OUT TRIG IN
REPLY GRP
REPLY
(TTL) OUT
ENABLE
TRIG IN
J24
MODE GRP
TRIG IN
J18
REPLY
(3-27V) OUT
J25 J26
VIDEO IN
DECODED
J19 J20
IFR BUS J11
AUX BUS J10
GPIB J21
24
1
FOR CONTINUED PROTECTION
AGAINST FIRE, REPLACE ONLY
WITH FUSE OF THE SPECIFIED
VOLTAGE AND CURRENT RATINGS.
CAUTION
F2F1
AC LINE FUSES
SSE
FU
SUF
E
E
F
UF
ESU
J6
AUXILIARY
IFR BUS
J5
J8
INTERROGATOR
J7
INDICATOR
120 WATTS MAX
100-120VAC 60Hz
220-240VAC 50Hz
3.0A @ 110/230 VAC
TYPE F
SI-1404 to ATC-1400A Interconnect ions
Figure 1
The Mode 4/Mode S Test System can be instal led in ei ther bench-top or rack-mount fashion. Al l IFR Systems, Inc., test sets are normally shipped f rom the factory with plast ic feet instal led for bench-top instal lat ion. Instal lat ion k its are l is ted in 1-2-1, Table 2.
KIT INSTRUMENT IFR SYSTEMS PART NUMBER
Rack-Mount ATC-1400A SI-1404
7001-7636-800 7001-6740-800
Bench-Top/ Stack-Mount
ATC-1400A SI-1404
7005-6743-000 One ki t required
per Mode 4/ Mode S Test
System
Instal lat ion Kits Table 2
CAUTION: AVOID RESTRICTION OF AIR FLOW TO INTAKE VENT. W HEN OPERATING IN THE NORMAL HORIZONTAL POSITION, MAINTAIN AT LEAST TW O INCHES (FIVE CENTIMETERS) OF CLEARANCE BETW EEN THE FAN SIDE OF THE EQUIPMENT AND OBJECTS OR W ALLS. IF OPERATING IN A RACK, MAXIMUM AMBIENT TEMPERATURE MUST BE AT OR BELOW 40° C.
OPERATION MANUALMODE 4/MODE STEST SYSTEM
1-2-1 Page 4
Feb 1/99
1.5 POWER-UP PROCEDURE
The Mode 4/Mode S Test System is conf igured for one-touch power control . W hen the ATC-1400A POW ER Switch is ON ( I) , the SI-1404 POW ER Switch controls power to the SI-1404 Test Auxi l iary and ATC-1400A Transponder/DME Test Set.
STEP PROCEDURE 1. Connect ac power cable to SI-1404
AC IN Connector and standard 3-pin grounded power receptacle.
2. Press ATC-1400A POW ER Switch ON ( I) .
3. Press SI-1404 POW ER Switch ON ( I) to energize Test Auxi l iary and ATC-1400A.
4. Ver i fy ATC-1400A Displays i l luminate.
5. Ver i fy SI-1404 MENU Display shows an act ive screen.
1.6 INSTALLATION CALIBRATION
Before operat ion, the Mode 4/Mode S Test System requires cal ibrat ion according to the Instal lat ion Cal ibrat ion Procedure. Any t ime the SI-1404 or ATC-1400A is replaced, another instal lat ion cal ibrat ion should be performed (only by qual i f ied service personnel) to maintain measurement accuracy.
1.7 EXTERNAL CLEANING
The fol lowing procedure contains rout ine instruct ions for c leaning the outs ide of the Test Set.
CAUTION: DISCONNECT POW ER FROM TEST SET TO AVOID POSSIBLE DAMAGE TO ELECTRONIC CIRCUITS.
STEP PROCEDURE 1. Clean f ront panel, switches and
display face with sof t l int- f ree c loth. I f dir t is di f f icul t to remove, dampen cloth with water and a mild l iquid detergent.
2. Remove grease, fungus and ground- in dir t f rom surfaces with sof t l int- f ree c loth dampened (not soaked) with isopropyl alcohol.
3. Remove dust and dir t f rom connectors with sof t-br is t led brush.
4. Cover connectors, not in use, with suitable dust cover to prevent tarnishing of connector contacts.
5. Clean cables with sof t l int- f ree c loth.
6. Paint exposed metal surface to avoid corrosion.