Pharmaceutical Guidelines
Document No :
PG/FAT/001
Factory Acceptance Test
Protocol Page No : 1 of 6
Tray Dryer (96 Trays)
1. Purpose :
1. To ensure that the Tray Dryer (96 Trays) is manufactured as per designed
specification at manufacturer's site.
2. To check the basic performance of the machine at manufacturer's site.
2. Scope and References:
1. This protocol is applicable to the machine manufactured by M/S XYZ Pvt. Ltd.
3. Procedure:
1. Take out photocopy of the approved protocol for the FACTORY ACCEPTANCE
TEST to be carried at the manufacturer's site.
2. Carry out the below tests mentioned in Table Number 1 at manufacturer's site
& mention the observation in ACTUAL column of the protocol photocopy.
3. In case of any additional information or mode of checking mention in the
REMARKS column.
4. Check the basic operation of the machine.
5. In case any deviation observes during FAT, document it in Observed deviation
section.
6. Ensure that corrective action taken is meeting the user requirement
specification.
7. If deviation is acceptable, document in Observed Deviation section and provide
proper justification.
8. If deviation does not acceptable, inform to manufacturer for necessary action.
9. If deviation is acceptable, provide final conclusion.
10. Ensure that following listed documents are provided.
1. Operational and Maintenance manual
2. Electrical / mechanical schematics diagram
3. Process and utility diagram
4. Critical Spare part list
4. Documentation
All documentation work shall be completed concurrently during execution of the
FAT. However the protocol does not define the sequence of the test/
documentation to be carried out.
1. Use indelible black / blue ink for the recording.
2. Fill out complete information in the format provided.
3. Do not leave any blank space.
4. Correct the wrong entry by drawing single line through incorrect data,
recording the correct data and then signing and dating the change.
5. Enter NA in space that is not applicable. 6. During execution of the protocol any deviation must be implemented as per
defined procedure.
7. During protocol execution, report shall be prepared with following details.
1. Discussion of all study in observation column.
-
Pharmaceutical Guidelines
Document No :
PG/FAT/001
Factory Acceptance Test
Protocol Page No : 2 of 6
Tray Dryer (96 Trays)
2. FAT is successfully completed or not
3. Conclusion
4. Recommendation, if any
5. Acceptance criteria:
1. The Equipments shall fulfill the selection criteria & its purpose of application.
1. The Equipments shall be as per purchase order.
2. Accessories received shall be as per purchase order.
3. The Equipments shall meet Pre-selected design parameters.
6. Verification Sheet:
Sr.
No.
Item/
Parameter
Expectation Actual Remarks
1 Make M/s Precikot Pharma Pvt. Ltd.
2 Model Tray Dryer (96 Trays)
3 Capacity : Approx. 300.00 kg with 0.5 Bulk Density
4 Design: Should have basic four parts,
Process Chamber
Trolley
Trays
Air Filtration system
5 Process
Chamber
comprising
Air Deflectors
Excess pressure flap
Door lock
Trolleys & Trays
MOC SS 316
6 Trolley: MOC : SS 316
Qty. 2 Nos.
With PU Wheel
7 Trays : MOC : SS316
Qty. 96 Nos.
8 Air Filtration
system :
Feed Blower
Pre Filter
Cooling Coil
Steam Coil
HEPA Filter
DOP Test report
MOC: SS 304
9 Feed Blower Make : Paan Fabriwork Engineering
-
Pharmaceutical Guidelines
Document No :
PG/FAT/001
Factory Acceptance Test
Protocol Page No : 3 of 6
Tray Dryer (96 Trays)
Capacity: 1500 CFM
Qty. 1 No
Motor : Reputed
HP : 3
Qty. 1 No.
10 Pre Filter
(Combination
type)
Filtration : 5 & 10
Make : Fine Airsys
Qty. 1 Nos.
11 Cooling Coil Make : Precikot
Qty. 1 No
MOC : SS316
12
Steam Coil
Make : Precikot
Qty. 1 No
MOC : SS316
13 HEPA Filter Filtration : 0.3
Make : Fine Airsys
Qty. 1 Nos.
Effeciency : 99.97 %
14 Exhaust
Blower
Make : Paan Fabriwork Engineering
Capacity: 1000 CFM
Qty. 1 No
Motor : Reputed
HP : 2
Qty. 1 No.
15 Control Panel
:
Qty. 1 No.
Operation : Through Auto/ Manual
Mode
Emergency Stop
Manual ON/ OFF switch
16 PLC Make : Allen Bradly
Model : Micro Logix
Qty. 1 No
17 HMI Make : Allen Bradly
Model : PV 550 Touch display
Qty. 1 No
18 Temperature
Sensor
Type : RTD
Make : Trister
Model : PT - 100
Qty. 5 Nos.
-
Pharmaceutical Guidelines
Document No :
PG/FAT/001
Factory Acceptance Test
Protocol Page No : 4 of 6
Tray Dryer (96 Trays)
19 Air Velocity
Sensor
Make : Microset
Model : HD403TS1
Qty. 1 No.
20 % RH Sensor Make : Microset
Model : RHT-DM
Qty. 1 No.
21 Operation
Selector
switch
PLC should Switch ON/ OFF by turning
the switch to ON / OFF position
22 Timers &
temperature
When selected should able to run the m/c
for set period of time & temperature
within 10C of set value,
23 Emergency
STOP Push
button
When pressed should stop the machine.
Start the machine & check.
24 Safety
Features
All moving
parts
All electrical
connections
Should be suitably covered under
replaceable guards or box
Should be appropriately numbered and
identif ied and covered with suitable
guard.
25 Documents
Machine
Manual
Should be available
MOC
Certificates
Should be available
Calibrati
on
Certificates
Should be available
Test and
Guarantee
Certificates
Should be available for Blower, motors &
Filters
7. Observed Deviation
Details of Deviation/ s :
-
Pharmaceutical Guidelines
Document No :
PG/FAT/001
Factory Acceptance Test
Protocol Page No : 5 of 6
Tray Dryer (96 Trays)
Investigation:
Corrective action:
Acceptable : Yes / No
Deviation Approved by / Date:
8. Conclusion:
FAT CARRIED OUT BY: IN PRESENCE OF MANUFACTURER:
Sign : Date:
Name:___________________________
Name: Sign : Date:
-
Pharmaceutical Guidelines
Document No :
PG/FAT/001
Factory Acceptance Test
Protocol Page No : 6 of 6
Tray Dryer (96 Trays)
Sign : Date:
Name:
___________________________
Site address :-
9. Approval
Name & Designation Department Signature Date
Prepared By