    Factory Acceptance Test

    Protocol Page No : 1 of 6

    Tray Dryer (96 Trays)

    1. Purpose :

    1. To ensure that the Tray Dryer (96 Trays) is manufactured as per designed

    specification at manufacturer's site.

    2. To check the basic performance of the machine at manufacturer's site.

    2. Scope and References:

    1. This protocol is applicable to the machine manufactured by M/S XYZ Pvt. Ltd.

    3. Procedure:

    1. Take out photocopy of the approved protocol for the FACTORY ACCEPTANCE

    TEST to be carried at the manufacturer's site.

    2. Carry out the below tests mentioned in Table Number 1 at manufacturer's site

    & mention the observation in ACTUAL column of the protocol photocopy.

    3. In case of any additional information or mode of checking mention in the

    REMARKS column.

    4. Check the basic operation of the machine.

    5. In case any deviation observes during FAT, document it in Observed deviation

    section.

    6. Ensure that corrective action taken is meeting the user requirement

    specification.

    7. If deviation is acceptable, document in Observed Deviation section and provide

    proper justification.

    8. If deviation does not acceptable, inform to manufacturer for necessary action.

    9. If deviation is acceptable, provide final conclusion.

    10. Ensure that following listed documents are provided.

    1. Operational and Maintenance manual

    2. Electrical / mechanical schematics diagram

    3. Process and utility diagram

    4. Critical Spare part list

    4. Documentation

    All documentation work shall be completed concurrently during execution of the

    FAT. However the protocol does not define the sequence of the test/

    documentation to be carried out.

    1. Use indelible black / blue ink for the recording.

    2. Fill out complete information in the format provided.

    3. Do not leave any blank space.

    4. Correct the wrong entry by drawing single line through incorrect data,

    recording the correct data and then signing and dating the change.

    5. Enter NA in space that is not applicable. 6. During execution of the protocol any deviation must be implemented as per

    defined procedure.

    7. During protocol execution, report shall be prepared with following details.

    1. Discussion of all study in observation column.

  • Pharmaceutical Guidelines

    Document No :

    PG/FAT/001

    Factory Acceptance Test

    Protocol Page No : 2 of 6

    Tray Dryer (96 Trays)

    2. FAT is successfully completed or not

    3. Conclusion

    4. Recommendation, if any

    5. Acceptance criteria:

    1. The Equipments shall fulfill the selection criteria & its purpose of application.

    1. The Equipments shall be as per purchase order.

    2. Accessories received shall be as per purchase order.

    3. The Equipments shall meet Pre-selected design parameters.

    6. Verification Sheet:

    Sr.

    No.

    Item/

    Parameter

    Expectation Actual Remarks

    1 Make M/s Precikot Pharma Pvt. Ltd.

    2 Model Tray Dryer (96 Trays)

    3 Capacity : Approx. 300.00 kg with 0.5 Bulk Density

    4 Design: Should have basic four parts,

    Process Chamber

    Trolley

    Trays

    Air Filtration system

    5 Process

    Chamber

    comprising

    Air Deflectors

    Excess pressure flap

    Door lock

    Trolleys & Trays

    MOC SS 316

    6 Trolley: MOC : SS 316

    Qty. 2 Nos.

    With PU Wheel

    7 Trays : MOC : SS316

    Qty. 96 Nos.

    8 Air Filtration

    system :

    Feed Blower

    Pre Filter

    Cooling Coil

    Steam Coil

    HEPA Filter

    DOP Test report

    MOC: SS 304

    9 Feed Blower Make : Paan Fabriwork Engineering

  • Pharmaceutical Guidelines

    Document No :

    PG/FAT/001

    Factory Acceptance Test

    Protocol Page No : 3 of 6

    Tray Dryer (96 Trays)

    Capacity: 1500 CFM

    Qty. 1 No

    Motor : Reputed

    HP : 3

    Qty. 1 No.

    10 Pre Filter

    (Combination

    type)

    Filtration : 5 & 10

    Make : Fine Airsys

    Qty. 1 Nos.

    11 Cooling Coil Make : Precikot

    Qty. 1 No

    MOC : SS316

    12

    Steam Coil

    Make : Precikot

    Qty. 1 No

    MOC : SS316

    13 HEPA Filter Filtration : 0.3

    Make : Fine Airsys

    Qty. 1 Nos.

    Effeciency : 99.97 %

    14 Exhaust

    Blower

    Make : Paan Fabriwork Engineering

    Capacity: 1000 CFM

    Qty. 1 No

    Motor : Reputed

    HP : 2

    Qty. 1 No.

    15 Control Panel

    :

    Qty. 1 No.

    Operation : Through Auto/ Manual

    Mode

    Emergency Stop

    Manual ON/ OFF switch

    16 PLC Make : Allen Bradly

    Model : Micro Logix

    Qty. 1 No

    17 HMI Make : Allen Bradly

    Model : PV 550 Touch display

    Qty. 1 No

    18 Temperature

    Sensor

    Type : RTD

    Make : Trister

    Model : PT - 100

    Qty. 5 Nos.

  • Pharmaceutical Guidelines

    Document No :

    PG/FAT/001

    Factory Acceptance Test

    Protocol Page No : 4 of 6

    Tray Dryer (96 Trays)

    19 Air Velocity

    Sensor

    Make : Microset

    Model : HD403TS1

    Qty. 1 No.

    20 % RH Sensor Make : Microset

    Model : RHT-DM

    Qty. 1 No.

    21 Operation

    Selector

    switch

    PLC should Switch ON/ OFF by turning

    the switch to ON / OFF position

    22 Timers &

    temperature

    When selected should able to run the m/c

    for set period of time & temperature

    within 10C of set value,

    23 Emergency

    STOP Push

    button

    When pressed should stop the machine.

    Start the machine & check.

    24 Safety

    Features

    All moving

    parts

    All electrical

    connections

    Should be suitably covered under

    replaceable guards or box

    Should be appropriately numbered and

    identif ied and covered with suitable

    guard.

    25 Documents

    Machine

    Manual

    Should be available

    MOC

    Certificates

    Should be available

    Calibrati

    on

    Certificates

    Should be available

    Test and

    Guarantee

    Certificates

    Should be available for Blower, motors &

    Filters

    7. Observed Deviation

    Details of Deviation/ s :

  • Pharmaceutical Guidelines

    Document No :

    PG/FAT/001

    Factory Acceptance Test

    Protocol Page No : 5 of 6

    Tray Dryer (96 Trays)

    Investigation:

    Corrective action:

    Acceptable : Yes / No

    Deviation Approved by / Date:

    8. Conclusion:

    FAT CARRIED OUT BY: IN PRESENCE OF MANUFACTURER:

    Sign : Date:

    Name:___________________________

    Name: Sign : Date:

  • Pharmaceutical Guidelines

    Document No :

    PG/FAT/001

    Factory Acceptance Test

    Protocol Page No : 6 of 6

    Tray Dryer (96 Trays)

    Sign : Date:

    Name:

    ___________________________

    Site address :-

    9. Approval

    Name & Designation Department Signature Date

    Prepared By