promax preinstall instructions v16 - planmeca images (2d) • dimaxis software ... wind osxp /erv 2...
Post on 30-Apr-2018
230 Views
Preview:
TRANSCRIPT
Planmeca ProMax pre install guide
Page 1 of 7 August, 2008
ProMax PRE INSTALLATION INSTRUCTIONS AND DIMENSIONS
INPUT / OUTPUT OUTLETS All outlets must be within 3ft from the back of the column and 12”-48” from the �oor.
1. Power: 110V, 20A dedicated circuit. 2. Remote exposure switch (See page 6 for wiring diagram):
2”x4” electrical outlet boxes at ProMax and remote exposure switch location. ½ ” conduit between the 2"x4" electrical boxes required.
All parts supplied for exposure switch connection. 3. Pan/ceph �lm units w/ Admark: 1½ ” conduit to the darkroom, 30ft. cable supplied. 4. ProMax Ethernet interface: Standard Ethernet RJ45 port to the network HUB/SWITCH.
Wall preparation: The pan or pan/ceph creates pull-out force of 1216 pounds. Wall bracket has 4 pre drilled holes, 16” and 12” on center. Either ones can be used. If metal studs are in wall, use a mounting board to expand the load over several studs to meet the pull -out requirement. If needed, sandwich the wall on both sides.
YOU MUST USE EITHER 1. Top wall bracket and �oor screws OR 2. Two wall brackets.
Only one wall bracket is supplied with each system. If planning for 2 wall brackets, you must add part# 10001073 to your order.
48”
– 50
”
8-16
”
16”
Version 17
Planmeca ProMax pre install guide Version 17 Page 2 of 7
August, 2008
ProMax Panoramic
60 ½
” – 9
5 ¾
”
38 ½
” – 7
4”
chin
rest
48” –
50”
8”-1
6”
22” (9”)
27 ½ ”
64” (
49 ½
”)
39” (33 ½ ”) 37
” re
st p
ositi
on
(ent
ry1)
61” (42 ½ ”)
27 ½ ” 42” (39”) 64” (48”)
Digital Film
Recommended operating dimensions (actual product dimensions in parenthesis)
Top
limits
can
be
set a
t the
inst
all s
ite.
22” (9”)
Planmeca ProMax pre install guide Version 17 Page 3 of 7
August, 2008
Promax Film Ceph
60 ½
” – 9
5 ¾
”
51” –
86
½ ”
46” (44 ½ ”)
27 ½”
28”
42” (39”)
38” –
73
½ ”
ear s
uppo
rts
32 ½
“ –
67 ¾
”
38 ½
” – 7
4”
chin
rest
37”
rest
pos
ition
(e
ntry
1)
48” –
50”
8-
16”
88” (83 ½ ”)
Top
limits
can
be
set a
t the
inst
all s
ite
Recommended operating dimensio ns (actual product dimensions in parenthesis)
64” (
49 ½
”)
Planmeca ProMax pre install guide Version 17 Page 4 of 7
August, 2008
ProMax Digital ceph
38” –
73
½”
ear s
uppo
rts
33”
– 68
½”
51 ¼
” –
86 ½
”
60 ½
” –
95 ¾
”
38 ½
” – 7
4”
chin
rest
46” (44 ½”)
30”
27 ½”
37”
rest
pos
ition
(e
ntry
1)
39” (33 ½”)
85” (78”)
Recommended operating dimensions (actual product dimensions in parenthesis)
Top
lim
its c
an b
e se
t at t
he in
stal
l site
64” (
49 ½
”)
Planmeca ProMax pre install guide Version 17 Page 5 of 7
August, 2008
Other install information: Each �lm based machine comes with one Autoprint Pan Cassette ( pan unit ) OR Standard Pan Cassette plus Ceph Cassette for Admark use ( pan/ceph unit ) Extra cassettes Standard panoramic, order number w/ Lanex screens: Autoprint panoramic, order number 10001288
,cirtemolahpeC Film recommended: pan: Agfa Dentus Ortholux 6”x12”
”01x”8 xulohtrO sutneD afgA :hpec Bite sticks: Bite sticks are autoclavable
08240001 rebmun redro ,kap-01,skcits etib artxE 1 rebmun redro :sdor 2 /w kap-01 ,setalp etib omoT 0005520
2nd wall bracket: order number 10001073 (optional) Handheld exposure order number 10004040 switch w/ wall mount (optional) Darkroom: Light tight, GBX2 �lter on safe light, 15W frosted maximum.
.mlif nepo morf ”84 naht resolc oN
Computer specs: 512 MB of RAM for digital Win 2000, XP
PCAnywhere connection in Imaging Station and Image server computers Hard drive: 2x 40GB for average size practice. Please consult with Planmeca certi�ed computer provider for your computer needs. Requirements vary greatly based on usage.
For more technical information, please contact Planmeca Customer Care (630) 529-230 0 or visit www.PlanmecaUSA.com
50024021
order number 50024015
Page 6 of 7 Planmeca ProMax pre install guide
August, 2008 Version 17
Ethernet ReadyThe new, built-in Ethernetconnection for Planmeca
Digital X-rays o�ers directnetwork connectivity.
You can literally plug theEthernet connection
directly into your network.
Note: An UninterruptedPower Supply (UPS) isrecommended forthe X-Ray.
Image AcquisitionWorkstation (2D)• Dimaxis Software acquires the
Panoramic, Tomographicand Cephalometric images (2D).
• Dimaxis Imaging Software is usedto enhance, �lter, measure,diagnose, archive and exportthe images.
Note: If using Vista, only VistaBusiness or Vista Ultimate(32 bit)
Server• Stores and archives all data
to MS SQL or SOLID.The data server should havelarge hard drive disk space,hardware redundancy and alogical backup system.
Note:
Additional WorkstationFor Panoramic,Tomographic andCephalometric images (2D)• Dimaxis Software
PlanmecaDigital X-Ray(ProMax Shown)
Local Area Network ConnectionRequired for sharing printer or addingadditional workstations.
Dimaxis Imaging Softwarefor 2D X-Ray Imaging
OptionalCrossover
OptionalPrinter
CPU Speci�cationsDigital ImageWorkstationFor Panoramic, Bitewing, Tomographicand Cephalometric 2D images• Pentium 4 or Higher• 60 GB or better HD• 2 GB or more RAM• WINDOWS XP / Server 2003 OS /
Vista (32 bit OS Only)• Ethernet 10/100 Mbit• 24-32 bit graphic card• monitor 1024 x 768 res. min.• High Speed Internet, DSL or better
CPU Speci�cationsAdditional WorkstationsFor Panoramic,Tomographic and Cephalometricimages• Refer to Digital Image
Acquisition Workstation (2D)
Database ServerSpeci�cations• Pentium 4 or Higher• 2 Hard Drives, 160 GB each min.• 1 GB or more RAM• WINDOWS XP / Server 2003 OS /
Vista (32 bit OS Only)
• Ethernet 10/100 Mbit
P L A N M E C AX -R a y Im a g in g S y s t e m s
Digital X-ray Imaging Hardware Diagramsand System Requirements
NOTE:Dimaxis is NOT compatiblewith 64Bit OS Systems
OS is requiredand requires Dimaxis 4.3.Xor newer.
forPlanmeca Digital Panoramicand Dixi X-Ray Systems
If using Vista, only VistaBusiness or Vista Ultimate(32 bit) OS is requiredand requires Dimaxis 4.3.Xor newer.
top related