Calibration technology
Multipurpose integrated calibration systemModel CBS-SC9000
Calibration bench, model CBS-SC9000
Applications
■ Multipurpose calibration station for exhaustive testing of process instruments and sensors
Main features
■ 19” modular construction for table operation ■ Aluminium structure with painted steel plates on sides,
top and rear; front panels, are anodised aluminium with engraved mimics.
■ Digital measurement simulation/generation of:pressure, vacuum, voltage, current, frequency, resistance/RTD’s, thermocouples, switch status test
Description
General informationThe calibration system CBS-SC9000 is a multipurpose calibration station that features a single 19” table top rack incorporating all it is needed to ensure the exhaustive calibration of a wide range of process instruments and sensors. System CBS-SC9000 is very easy to use and shortens the way and the time required to conform user’s calibration procedures to ISO 9000. Its exclusive redundancy of operations with PASCAL calibrators, the Scandura’s multi-function calibrator, remains unique, allowing a real integration between lab and fi eld calibration under identical quality procedures and calibration formats. As stand-alone unit System CBS-SC9000 opens the doors of your maintenance & calibration laboratories to total quality management.
The user interface is represented by a wide LCD display with touch screen through which it is possible to setup the instrument in total. The procedure to confi gure is very simple. The display screen varies in a dynamic way according to the choices made and drives the operator to perform his job, decreasing his time for learning and the possibility to make mistakes. The touch screen can also be used with hand gloves and in case of dirt you can easily clean it with a cloth or a sponge and can be locked to avoid unintentional pressing of keys during the process of measurement.
The backlight illumination with LED guarantees a perfect view in case of low environmental visibility
System CBS-SC9000 allows the user to calibrate the whole industrial instrumentation like for example:
■ Electronic and pneumatic pressure/vacuum transmitter ■ Electronic and pneumatic diff erential pressure transmitter ■ Electronic and pneumatic temperature transmitter
(thermocouple and RTD) ■ Signal converter V, mV, mA, Hz, Pulse, Ω ■ Signal converter I/P and P/I ■ Signal isolator mA, mV ■ Mathematical device (adder, subtractor, multiplier, divider) ■ Manometer, pressure switch ■ Temperature monitor switch ■ Thermocouple and resistance thermometer ■ Compensation device (up to 4 in-out signals) ■ Electronic and pneumatic receiver ■ Electronic and pneumatic controller ■ Electronic and pneumatic recorder ■ Miscellaneous instrument
WIKA data sheet CT 92.35
Page 1 of 8WIKA data sheet CT 92.35 ∙ 03/2015
Specifications
1) PASCAL RACK
Display Touchscreen + 5 keysDimensions 320 x 240 Dots
Dot size: 0.34 x 0.34 mm (0.013 x 0.013 inch)Backlight LED
Electrical input and output
Type banana-plug inputs for electrical parameters, resistance thermometers and thermocouples
HART® communication
HART® module based on HART® universal and common practice commands
Resistance HART® resistance 250 Ω (activable)
Loop current max. DC 24
mA voltage supply DC 24 V
Permissible media clean, dry, non-corrosive gases
Permissible media clean, dry, non-corrosive gases
Temperature coefficient 0.001 % of reading/°C, outside of 19 ... 23 °CUnits bar, mbar, psi, psf, Pa, hPa, kPa, MPa, torr, atm, kg/cm², kg/m², mmHg (0 °C), cmHg (0 °C), mHg (0
°C), inHg (0 °C), mmH2O (4 °C), cmH2O (4 °C), mH2O (4 °C), inH2O (4 °C), ftH2O (4 °C)
Permissible ambient conditions
Operating temperature -10 ... +50 °C (14 ... 122 °F)
Storage temperature -30 ... +80 °C (-22 ... +176 °F)
Relative humidity Operating humidity: 10 ... 90 % r. h. (no condensation)
Storage humidity: 0 ... 90 % r. h. (no condensation)
Page 2 of 8 WIKA data sheet CT 92.35 ∙ 03/2015
3
1
2
Page 3 of 8WIKA data sheet CT 92.35 ∙ 03/2015
Pressure moduleInternal sensors (other pressure ranges available on request)
■ One year specifications ■ Temperature effect: 0.002 % of reading * |t – tc| for t : 0 °C ≤ t ≤ 18 °C and 28 °C ≤ t ≤ 50 °C and tc = 20 °C ■ Pneumatic connection: depending on the Pascal model
Measuring range Precision (% FS) Accuracy (% FS) ResolutionRelative pressure-60 ... +60 mbar 0.08 0.1 0.01 mbar-500 ... +500 mbar 0.015 0.025 0.01 mbar-900 ... +1,500 mbar 0.015 0.025 0.01 mbar0 ... 7 bar 0.015 0.025 0.1 mbar0 ... 21 bar 0.015 0.025 0.1 mbar0 ... 50 bar 0.015 0.025 1 mbar0 ... 100 bar 0.015 0.025 1 mbarAbsolute pressure600 ... 1,300 mbar abs. 0.015 0.025 0.01 mbar0 ... 1,500 mbar abs. 0.015 0.025 0.01 mbar0 ... 2,500 mbar abs. 0.015 0.025 0.01 mbar0 ... 2,500 mbar abs. 0.010 0.015 0.01 mbar0 ... 5 bar abs. 0.015 0.025 0.1 mbar0 ... 7 bar abs. 0.015 0.025 0.1 mbar0 ... 21 bar abs. 0.015 0.025 0.1 mbar0 ... 81 bar abs. 0.015 0.025 1 mbar0 ... 100 bar abs. 0.015 0.025 1 mbar
Internal sensors (other pressure ranges available on request) ■ One year specifications ■ Temperature effect: 0.002 % of reading * |t – tc| for t : 0 °C ≤ t ≤ 18 °C and 28 °C ≤ t ≤ 50 °C and tc = 20 °C ■ Pneumatic connection: depending on the Pascal model
Measuring range Precision (% FS) Accuracy (% FS) ResolutionRelative pressure-60 ... +60 mbar 0.1 0.15 0.01 mbar-500 ... +500 mbar 0.015 0.025 0.01 mbar-900 ... +1,500 mbar 0.015 0.025 0.01 mbar0 ... 7 bar 0.015 0.025 0.1 mbar0 ... 21 bar 0.015 0.025 0.1 mbar0 ... 50 bar 0.015 0.025 1 mbar0 ... 100 bar 0.015 0.025 1 mbar0 ... 200 bar 0.015 0.025 10 mbar0 ... 400 bar 0.015 0.025 100 mbar0 ... 700 bar 0.025 0.05 100 mbar0 ... 1,000 bar 0.025 0.05 100 mbarAbsolute pressure0 ... 1,500 mbar abs. 0.015 0.025 0.01 mbar0 ... 2,500 mbar abs. 0.015 0.025 0.01 mbar0 ... 5 bar abs. 0.015 0.025 0.1 mbar0 ... 7 bar abs. 0.015 0.025 0.1 mbar0 ... 21 bar abs. 0.015 0.025 0.1 mbar0 ... 81 bar abs. 0.015 0.025 1 mbar0 ... 100 bar abs. 0.015 0.025 1 mbar
Page 4 of 8 WIKA data sheet CT 92.35 ∙ 03/2015
Electrical / temperature input signal
Electrical signal Measuring range Full scale Precision% of rdg ±% FS
Accuracy% of rdg ±% FS Max. resolution
Voltage DC 1) 2) ±100 mV 3) 100 mV 0.008 % ±0.002 % FS 0.01 % ±0.003 % FS 0.0001 mV±2 V 3) 2 V 0.008 % ±0.002 % FS 0.01 % ±0.003 % FS 0.000001 V
±80 V 4) 80 V 0.008 % ±0.002 % FS 0.01 % ±0.003 % FS 0.00001 VCurrent DC 1) 5) ±100 mA 100 mA 0.008 % ±0.003 % FS 0.01 % ±0.003 % FS 0.0001 mAResistance 1) 6) 0 ... 400 Ω 400 Ω 0.008 % ±0.002 % FS 0.01 % ±0.003 % FS 0.001 Ω
0 ... 10,000 Ω 10,000 Ω 0.008 % ±0.002 % FS 0.01 % ±0.003 % FS 0.01 ΩFrequency 7) 0.5 ... 10,000 Hz 8) 50,000 Hz 0.01 Hz 0.01 Hz 0.001 Hz
10,000 ... 20,000 Hz 8) 50,000 Hz 0.1 Hz 0.1 Hz 0.001 Hz20,000 ... 30,000) Hz 9) 50,000 Hz 1 Hz 1 Hz 0.001 Hz30,000 ... 50,000) Hz 9) 50,000 Hz 20 Hz 20 Hz 0.001 Hz
Pulses 10) 1 ... 999,999 999,999 N/A N/A 11) One year specifications with temperature effect: 0.001 % of reading * |t – tc for t : -10 °C ≤ t ≤ 19 °C and 23 °C ≤ t ≤ 50 °C and tc = 20 °C2) Maximum input voltage: DC ±100 V 3) Input impedance: > 100 MΩ4) Input impedance: 0.5 MΩ5) Maximum input current: ±120 mA Input impedance: < 20 Ω
6) Measure current: < 200 µA7) Maximum input voltage: ±100 V input impedance: > 100 MΩ Minimum amplitude of square wave: 1.5 V p-p @ 50 kHz, 0.7 V p-p @ 5 Hz Configurable duty cycle from 10 % up to 90 % with minimum amplitude of 5 V p-p8) For both frequency inputs simultaneously (IN A + IN B)9) For only one frequency input (IN A or IN B) in the same time10) Amplitude: 1 ... 80 V, frequency: 0.5 ... 20 Hz
Temperature Input type Precision up to
Accuracy up to
Resolution up to
RTD Pt100 α = 0.00385 °C-1 0.05 °C 0.06 °C 0.01 °CPt100 α = 0.003916 °C-1 0.05 °C 0.06 °C 0.01 °CPt100 JIS α = 0.003916 °C-1 0.05 °C 0.06 °C 0.01 °CPt100 α = 0.003902 °C-1 0.05 °C 0.06 °C 0.01 °CPt100 α = 0.003926 °C-1 0.05 °C 0.06 °C 0.01 °CPt100 α = 0.003923 °C-1 0.05 °C 0.06 °C 0.01 °CPt200 α = 0.00385 °C-1 0.05 °C 0.06 °C 0.01 °CPt500 α = 0.00385 °C-1 0.05 °C 0.06 °C 0.01 °CPt1000 α = 0.00385 °C-1 0.05 °C 0.06 °C 0.01 °CPt1000 α = 0.003916 °C-1 0.05 °C 0.06 °C 0.01 °CCu10 α = 0.0042 °C-1 0.23 °C 0.28 °C 0.1 °CCu100 α = 0.0042 °C-1 0.06 °C 0.07 °C 0.01 °CNi100 α = 0.00617°C-1 0.04 °C 0.05 °C 0.01 °CNi120 α = 0.00672°C-1 0.04 °C 0.05 °C 0.01 °C
Temperature Input type Precision% of rdg ±% FS
Accuracy% of rdg ±% FS
Linear error up to
Resolution up to
TC Type J1), K1), T1), F1), R, S, B1), U1), L1), N, E, M, C1) 0.008 % ±0.002 % FS 0.01 % ±0.003 % FS 0.03 °C 0.01 °C
1) Precision and Accuracy of the e.m.f. values For measurements with internal cold junction compensation: cold junction error = 0.15 °C Maximum input voltage: DC ±100 V Input Impedance: > 100 MΩ Temperature effect: 0.001 % of reading * |t – tc| for t : -10 °C ≤ t ≤ 19 °C and 23 °C ≤ t ≤ 50 °C and tc = 20 °C One year specifications
Page 5 of 8WIKA data sheet CT 92.35 ∙ 03/2015
Electrical / temperature output signal
Electrical signal Measuring range Full scale Precision% of rdg ±% FS
Accuracy% of rdg ±% FS Max. resolution
DC Voltage 1) 0 ... 100 mV 2) 100 mV 0.01 % ±0.003 % FS 0.015 % ±0.003 % FS 0.0001 mV0 ... 2 V 3) 2 V 0.01 % ±0.003 % FS 0.015 % ±0.002 % FS 0.000001 V
0 ... 20 V 3) 20 V 0.015 % ±0.003 % FS 0.02 % ±0.003 % FS 0.00001 VDC Current 4) 0 ... 20 mA 5) 100 mA 0.02 % ±0.003 % FS 0.025 % ±0.003 % FS 0.0001 mAResistance 4) 0 ... 400 Ω 400 Ω 0.008 % ±0.003 % FS 0.01 % ±0.002 % FS 0.001 Ω
0 ... 10,000 Ω 10,000 Ω 0.008 % ±0.002 % FS 0.01 % ±0.001 % FS 0.01 ΩFrequency 0.5 ... 50,000 Hz 50,000 Hz 0.1 Hz 0.1 Hz 0.001 HzPulses 6) 1 ... 999,999 999,999 N/A N/A 11) One year specifications with temperature effect: 0.001 % output * |t – tc| for t : -10 °C ≤ t ≤ 19 °C and 23 °C ≤ t ≤ 50 °C and tc = 20 °C2) Output impedance = 10 Ω - Rlmin > 1 kΩ 3) Output impedance < 30 mΩ - Rlmin > 1 kΩ
4) One year specifications with temperature effect: 0.002 % output * |t – tc| for t : -10 °C ≤ t ≤ 19 °C and 23 °C ≤ t ≤ 50 °C and tc = 20 °C5) Output impedance > 100 MΩ - Rlmax < 750 Ω6) Amplitude: 0.1 ... 15 Vrms, frequency: 0.5 ... 200 Hz
Temperature Input type Precision up to
Accuracy up to
Resolution up to
RTD Pt100 α = 0.00385 °C-1 0.05 °C 0.06 °C 0.01 °CPt100 α = 0.003916 °C-1 0.05 °C 0.06 °C 0.01 °CPt100 JIS α = 0.003916 °C-1 0.05 °C 0.06 °C 0.01 °CPt100 α = 0.003902 °C-1 0.05 °C 0.06 °C 0.01 °CPt100 α = 0.003926 °C-1 0.05 °C 0.06 °C 0.01 °CPt100 α = 0.003923 °C-1 0.05 °C 0.06 °C 0.01 °CPt200 α = 0.00385 °C-1 0.05 °C 0.06 °C 0.01 °CPt500 α = 0.00385 °C-1 0.05 °C 0.06 °C 0.01 °CPt1000 α = 0.00385 °C-1 0.05 °C 0.06 °C 0.01 °CPt1000 α = 0.003916 °C-1 0.05 °C 0.06 °C 0.01 °CCu10 α = 0.0042 °C-1 0.23 °C 0.28 °C 0.1 °CCu100 α = 0.0042 °C-1 0.06 °C 0.07 °C 0.01 °CNi100 α = 0.00617°C-1 0.04 °C 0.05 °C 0.01 °CNi120 α = 0.00672°C-1 0.04 °C 0.05 °C 0.01 °C
Temperature Input type Precision% of rdg ±% FS
Accuracy% of rdg ±% FS
Linear error up to
Resolution up to
TC Type J1), K1), T1), F1), R, S, B1), U1), L1), N, E, M, C1) 0.008 % ±0.002 % FS 0.01 % ±0.003 % FS 0.03 °C 0.01 °C
1) Precision and accuracy of the e.m.f. values For measurements with internal cold junction compensation: cold junction error = 0.15 °C Maximum input voltage: DC ±100 V Input impedance: > 100 MΩ Temperature effect: 0.001 % of reading * |t – tc| for t : -10 °C ≤ t ≤ 19 °C and 23 °C ≤ t ≤ 50 °C and tc = 20 °C One year specifications
Page 6 of 8 WIKA data sheet CT 92.35 ∙ 03/2015
2) Pressure-vacuum pneumatic signals panel
Pressure generation - Manual pump
Pressure generation Up to 21 bar
Vacuum generation Down to -0,9 bar
Pressure release Push button for venting
Adjustment Precision pressure and vacuum volumetric regulator, for small adjustments, sensitivity 0.1 mbar
Output connection Two output plug-in connectors
Regulated pressure
Input pressure 7 bar air supply from external source
Output pressure Adjustable 0 ... 6 bar
Adjustment Precision pressure regulator
Output connection Two output plug-in connectors
Pressure presets
Input pressure 7 bar air supply from external source
Output pressure Adjustable 0 ... 2 bar
Adjustment Two precision pressure regulator
Output preset Three-way pneumatic push button selector for three pressure points quick calibration
Output connection Four output plug-in connectors
Electric vacuum pump
Residual vacuum Down to 2 mbar residual
Adjustment Manual regulator with switch push buttons for coarse / fine vacuum adjustment
Output connection Four output plug-in connectors
System mains switch and AC sockets
Input AC 230 V ±10 %, 50 Hz
Switch rating AC 230 V, 32A
Output Four sockets connected to mains and available with socket to DIN/VDE standards
Power The total power is limited by the system mains switch rating
Page 7 of 8WIKA data sheet CT 92.35 ∙ 03/2015
Dimensions in mm
WIKA Italia Srl & C. Sas Via G. Marconi, 820020 Arese (Milano)/ItaliaTel. +39 02 93861-1Fax +39 02 93861-74info@wika.itwww.wika.it
© 2015 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.The specifications given in this document represent the state of engineering at the time of publishing.We reserve the right to make modifications to the specifications and materials.
Page 8 of 8 WIKA data sheet CT 92.35 ∙ 03/2015
03/2
015
EN
Mechanical The system CBS-SC9000 housing is composed by one 19’’ module cabinet. Steel structure, finely finished In light grey RAL 7032; mimic panels engraved and colored on anodised aluminum plates
Dimensions
■ Front 553 mm ■ Depth 500 mm ■ Height 766.5 mm
Weight 25 kg net (packed 35kg)
Electrical safety
The system CBS-SC9000 is featured for the electrical safety, including:
■ Approved mains switch ■ Earth leakage circuit breaker to standard IEC-1010-
1 for personnel and fire protection by detecting an earth leakage fault current in excess of 30 mA and disconnecting the main supply within 200 ms
■ ‘Double isolation’ protection in connecting the external mains supply to the system main switch, by completely enclosing live connections in an insulating case
■ Protective transparent covers over live connections inside the equipment
Power requirement
■ Main supply AC 230 V ±10 % 50/60 Hz ■ Main switch rating: 32A ■ System operation: 6.5 kVA (maximum) ■ AC sockets rating: 16 A
Operating environment
■ Specified for 0°C ... 50°C ■ Specified to 80% r.h. at 35°C, not condensing ■ Storage environment: -40° ... 70°C
Supply connections
■ Main power cable gland 16 mm diameter for 3-wire 4 mm2 cross section cable. Located at the rear panel
Standard certification
■ Certificate of conformity. ■ Calibration reports
Certification on request
■ Certificate of calibration, against reference masters, traceable to national and international standards (E.A.).
■ Official certificates issued by accredited laboratories of the Italian national service (ACCREDIA) membership of the E.A. (European Accreditation Laboratory).
Standard supply
■ One plug for each socket ■ Kit of cables and pressure plugs/tubing
Technical documentation
Two sets of the Installation, operating and maintenance manual.