MR for quantitative fat imaging

Pernilla Peterson1, Sven Månsson1

1Medical Radiation Physics, Lund University, Malmö, Sweden

WHY FAT QUANTIFICATION?

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 2

• Obesity and diabetes

• Fat accumulation

– Liver, pancreas, myocardium

MR-image based fat quantification

THE BASIC PRINCIPLE

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 3

1. Separation of MR-signals from fat and

water

2. Quantification of the amount of fat relative

the amount of water

SUMMARY

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 4

• Fat / water separation

• Fat quantification

– T1-relaxation

– T2-relaxation

– Proton density

• Simple strategies

SUMMARY

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 5

• Fat / water separation

• Fat quantification

– T1-relaxation

– T2-relaxation

– Proton density

• Simple strategies

FAT / WATER SEPARATION

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 6

W F

FAT / WATER SEPARATION

Chemical shift

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 7

FAT / WATER SEPARATION

Chemical shift

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 8

Total signal

Water

Fat

Time after excitation (ms)

FAT / WATER SEPARATION

2-point Dixon

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 9

FWSIP

FWSIP

SUMMARY

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 10

• Fat / water separation

• Fat quantification

– T1-relaxation

– T2-relaxation

– Proton density

• Simple strategies

FAT QUANTIFICATION

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 11

=

0

50

100

W F

F

+

SOURCES OF IMAGE CONTRAST IN MR

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 12

T1-relaxation T2*-relaxation Proton density

SOURCES OF IMAGE CONTRAST IN MR

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 13

T1-relaxation T2*-relaxation Proton density

Resonance

frequency

FAT QUANTIFICATION

T1-relaxation

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 14

0 0.5 1 1.5 2Repetition time (s)

|Sig

nal|

Water

Fat

0 50 1000

20

40

60

80

100

Actual FF (%)

Estim

ate

d F

F (

%)

Identity line

Esimated FF

FAT QUANTIFICATION

T2*-relaxation

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 15

Time after excitation

Ma

gn

itu

de

sig

na

l

0 0.05 0.1 0.15 0.2Echo time

Ma

gn

itu

de

sig

na

l

Water

Fat

FAT QUANTIFICATION

Proton density

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 16

Water voxel Fat voxel

SUMMARY

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 17

• Fat / water separation

• Fat quantification

– T1-relaxation

– T2-relaxation

– Proton density

• Simple strategies

SUMMARY

Pernilla Peterson, pernilla.peterson@med.lu.se

Medical Radiation Physics, Lund University, Malmö 18

• Simple strategies to avoid bias

– Small flip angle or long repetition time

– A priori T2-values for fat and water

– Estimate proton density for water and

vegetable oils