Lab Scan
Version 1.0 | 990487 - Y9500 | September 2016
INSTRUCTION FOR USE
22 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
SAFETY INSTRUCTIONS 3
INTENDED USE 4
TRANSPORT AND PACKAGING 4
IDENTIFICATION OF PARTS AND ACCESSORIES 5
USE OF EXTERNAL EQUIPMENT 7
HARDWARE SETUP 7
FIRST USE 12
OPERATION 13
ACCESSORIES 21
CARE OF THE UNIT 43
TROUBLESHOOTING 44
DISPOSAL OF THE UNIT 46
WARRANTY AND REPAIR 46
MAINTENANCE AND SERVICE 46
TECHNICAL SPECIFICATIONS 47
Prior to use, carefully read these instructions
CONTENTS
33Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
CAUTION
• Follow the instructions in order to use the unit correctly and safely. Keep these instructions within easy reach of the device.• Network/UBS/Power connection may not be plugged out when the machine is ON.• Install the scanner on an appropriate, level and stable working surface, free from vibrations.• Only qualified dental personnel should use the scanner.• Prior to use, make sure that the scanner is working correctly. The unit should be recalibrated once it has been moved or repositioned• The Aadva Lab Scan is designed for use on dental models only and should not be used for anything else
other than its intended purpose. • Do not touch the device while the tilt and swivel unit is in motion. This may cause scan malfunctions, accidental squeezing of the fingers or damage
to the scanning unit itself.• Do not look directly into the light emitted by the scanning device. This may cause physical discomfort.• The unit should not be handled with wet hands. This may cause equipment malfunction.• The unit should not be exposed to direct sunlight, dust, wet environment or placed near a heater.• It is expressly prohibited to open, disassemble or remodel the unit, or use the device in a manner not
covered by these instructions, as this will lead to the termination of the warranty, guarantee and liability of the manufacturer.
• Touching/handling the sensor head may cause the unit to malfunction. • Additionally installed software can impair the operation of the system.• No additional software can be installed on the computer of the Aadva Lab Scan. The system settings on the computer of the Aadva Lab Scan may not be changed. The exceptions to this rule are the date, time, time zone and country settings.
The internet connection (minimum 128Kbs speed and ideally about 520Kbs) should be connected directly to the PC via a RJ45 crossover cable. The internet connection must be dedicated for the functionality of the scanner (including system upgrades).
WARNING • Do not use the unit if the user has a pacemaker. This device may interfere with the pacemaker.• Do not look direct into the LED light source. • Do not plug / unplug the device with wet hands. This may cause an electric shock.• If the unit should begin to smoke, emit an unusual odour, deform or discolour during scanning, calibration or storage, discontinue use immediately. • Avoid sudden or strong impact to the device. This may cause the unit to breakdown or lead to errors in the scanning function of the device. • Keep the unit away from flammable materials.• Do not store or operate the scanning device in areas where ceramic dust is produced. This may interfere with the scanning functionality and/or cause damage to the scanning device. • Do not place the scanner where liquids may accidentally get poured on it. Do not place any liquid close
to the scanner.• Do not place or store objects on top of the scanner. • Do not use the accompanying USB license dongles for data storage. This may interfere with the correct licensing and functioning of the Aadva Lab Scan.
SAFETY INSTRUCTIONS
Aadva® Lab Scan
44 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
The Aadva Lab Scan is a structured light scanner for creating digital 3D dental models by scanning dental models which can then be used in the (CAD/CAM) design, manufacture and production of dental prosthesis. The device uses blue LED light and has a scan time of 980 milliseconds, making it one of the fastest dental scanners available. The included software also allows the user to directly upload edited 3D models for CAD/CAM processing and manufacture (for detailed indications see page 4, 5, 6).
The PC component of the AadvaScanner constitutes an integral part of the AadvaScanner and may not be used for any purposes other than for the functionality of the AadvaScanner itself. No components (whether hardware and/or software) and/or other applications may be changed, added or removed.
GC Europe n.v. and/or the GC Europe distributors and/or service partners approved and appointed by GCE are responsible for delivering and installing patches, fixes, service packs and software upgrades related to the installed software in order to guarantee the proper functioning of the AadvaScanner and in accordance with the maintenance contract concluded if the latter is applicable.
GC Europe n.v. is not deemed to have knowledge of or to take account of the specific use that will be made by the customer of the purchased goods, and thus GC Europe n.v. cannot be held liable for this. The customer is expected to use the purchased good in accordance with its intended use.
INTENDED USE
Prior to setting up and assembling the Aadva Lab Scan, please make sure that all parts and components are included in your delivery (refer to the accompanying delivery note). Also, check for any visible signs of damage which the device may have incurred during transport.
CONDITIONNEMENT
The Aadva Lab Scan is shipped in a standard Euro pallet. For the safe disposal and recycling of the packaging, please ensure that all packing materials are properly disposed of according to current recycling directives. In case the device has to be returned, please use the original packing material for re-packaging, as only this would ensure safe transport.
CONTENTS Quantity 1. Aadva Lab Scan 1 2. Workstation PC 1 3. Mouse 1 4. Keyboard 1 5. Power Cord 3 6. USB Cable Type A/B 1 7. LAN Cable 1
8. Hardware Dongle 1 9. Software Dongle 1 10. Model Holder Low 1 11. Model Holder High 1 12. Single Die Holder 1 13. Calibration Set 1 14. Software and documentation 1
TRANSPORT AND PACKAGING
55Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
IDENTIFICATION OF PARTS AND ACCESSORIES
Correct and safe operation of the Aadva Lab Scan is only possible with original parts and accessories delivered by GC Europe and with the indicated compliant external equipment. For any use of other parts or accessories or non-compliant external equipment the user assumes the risks.
1. Aadva Lab Scan
4. ON/OFF Switch and Power Inlet on Back Panel of Aadva
Lab Scan
2. Aadva Lab Scan Power Cord
5. LAN Port on Back Panel of Aadva Lab Scan
3. Aadva Lab Scan Sensor Head
6. USB Port on Back Panel of Aadva Lab Scan
LIFTING AND CARRYING
The most practical method of moving and placing the Aadva Lab Scan will be by handling manually. Be aware of your personal limits or seek for assistance, taking account the weights of the scanner and the PC indicated in the specifications.
Aadva® Lab Scan
66 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
13. Hardware Dongle 14. Model Holder Low / High 15. Keyboard
Lab Scan from GC
INSTRUCTION FOR USE
Version 1.2 | V4902 | January 2016
16. Calibration set 17. Single Die Holder 18. Documentation and Software
7. Workstation 8. Workstation Power Cord 9. Mouse
10. USB Cable Type A/B 11. LAN Cable 12. Software Dongle
77Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
The Aadva Lab Scan requires the use of external equipment for proper operation. These devices must be compliant with the following standards: DIN/EN/IEC 60950, UL 60950 and CAN/CSA-C22.2NO. 60950-00. Our software is compatible with Dell and HP monitors. The customer is required to provide the following external equipment:
1. Monitor 2. Monitor Power Cord 3. VGA Monitor Cable
USE OF EXTERNAL EQUIPMENT
1. Install the Aadva Lab Scan on a flat and stable place, in a medical room or laboratory. The Aadva Lab Scan should not be located in the vicinity of patients.
The scanner will require a level surface area measuring approximately 400 x 300 mm. 2. Connect the power cord to the main power outlet.
3. Connect the LAN cable to the network plug on the Aadva Lab Scan. Then, insert Type B connector of the USB Cable into the USB port on the back panel
of the Aadva Lab Scan. 4. Install the monitor and workstation on a flat and stable place, near the Aadva Lab Scan.
HARDWARE SETUP
Aadva® Lab Scan
88 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
5. Connect the power cords of the monitor and the workstation to the main power outlet (110-240V AC).
6. Connect the LAN cable to the network plug on the workstation. Then, insert the Type A connector of the USB Cable into a USB port on the workstation.
USB Cable Type A/B
99Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
7. Connect the VGA cable to the VGA ports on the monitor and the workstation.
LAN Cable
Aadva® Lab Scan
1010 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
8. Insert the software and hardware dongle into available USB ports on the workstation (preferably at the back of the PC in order to avoid any damage).
Hardware Dongle
Software Dongle
1111Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
External LAN
10. Turn on the workstation and monitor. Once prompted, choose USER 1. No password is needed.
11. Do not pull out any of the connection cables while the device is still on.
9. External LAN (internet, network, …)
Aadva® Lab Scan
1212 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn on and the GC logo on the front panel will light up.
After a few seconds, the blue light from the LED in the sensor head will be visible. Switch on the PC.
2. Once the workstation and monitor are switched on, check to see if the LEDs on the license dongles blink a few times.
3. Start the Dental manager software via the application shortcut on the desktop.
4. Once the Aadva Lab Scan and the scanning software are running correctly, the Aadva Lab Scan must be calibrated again. This may be necessary if the Aadva Lab Scan has been transported in a horizontal position or if the scanner has undergone rough transport. The Aadva Lab Scan may have to be recalibrated even if the scan software has not indicated that calibration is necessary.
5. Make sure the calibration plate is clean before starting the different calibration steps. Use a dry and soft tissue to clean.
FIRST USE
1313Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
This section covers the basic operation of the Aadva Lab Scan and the scanning software. For a detailed description of all the features of the scanning software, please refer to the software documentation (http://wiki.exocad.com/). This document is permanently available on the web.
1. Start the dental manger software via the application shortcut on the desktop.
2. Once the application starts, the user has to define a scan profile, or select from a list of stored profiles.
3. To define a scan profile, the following information has to be entered:
a. Customer ID and Nameb. Patient ID and Namec. Reconstruction Type d. Technician ID and Name (optional)
The first three settings have to be entered, otherwise the Save button is deactivated and the user cannot start with the scan.
OPERATION
Aadva® Lab Scan
1414 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
4. Click on the Edit button in the Customer section or select an existing customer from the drop-down menu.
5. Select an existing customer or the Default customer. Alternatively, click on the New Entry button to define a new customer. Click on Save once all details have been entered. Then, click on the Close button to exit the window.
6. You can define the patient information under the Job section. Click on the Edit button to define a new patient or select an existing patient in the drop-down menu.
7. Select an existing patient or the Default patient. Alternatively, click on the New Entry button to define a new patient. Click on Save once all details have been entered. Then, click on the Close button to exit the window.
1515Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
8. Now, you can select the type of reconstruction for which the scan is to be made. To do this, first select the relevant position/s in the dental model and click on the corresponding number/s.
9. Once a position has been selected, define the reconstruction type from the list which appears. Click on the Save button to proceed.
For this operational guide, a scan of tooth 16 for an anatomic crown reconstruction type will be used. For a full listing of various scans and reconstruction type, please refer to the software documentation (http://wiki.exocad.com/).
Aadva® Lab Scan
1616 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
10. After entering the necessary information for customer, patient and reconstruction type, you can also enter the technician information. This information is not mandatory, but can be useful in identifying various scan profiles.
Click on the Edit button to start defining the technician information, or select an existing technician from the drop-down menu.
11. Select an existing technician or the Default technician. Alternatively, click on the New Entry button to define a new technician. Click on Save once all details have been entered. Then, click on the Close button to exit the window.
12. Now that the information for customer, patient and reconstruction type has been entered, the scanner software will allow you to save the profile.Click on the Save button to save this scan profile.
13. After a scan profile has been defined, you can then continue with performing a scan or single scan. Click on the appropriate buttons to start scan.
1717Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
14. The scan software will now connect to the Aadva Lab Scan. An image of what the sensor head finds on the scanner is displayed immediately.
If there are red areas on the dental model, you can adjust the exposure time using the adjustment setting in the lower left hand section of the sensor head image (marked by a white margin in the image below).
Select the upper box to adjust the exposure time for measurements
Select the lower box to adjust the exposure time for reference points
After selecting the exposure setting, simply hold the mouse over the image and use the scroll wheel on the mouse to make adjustments. Moving the wheel up increases the exposure time, moving it down decreases it.
To see the exact exposure time in milliseconds, click and drag the image of the wheel instead.
15. Besides adjusting the exposure setting, you can click on the dental model directly and use the scroll wheel on the mouse to zoom in and out of the object.
Aadva® Lab Scan
1818 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
16. After adjusting the exposure and zoom levels, you should have an image of the dental model with all relevant positions clearly visible and without red markings, similar to the image below.
17. You are now ready to start the preparation scan. Click on the Next button to start the scan.
18. After cropping and deleting various regions from the preparation scan, you will have an image of the dental model free of any objects or parts not relevant for the current scan job, similar to the image below (keep only the relevant parts of the scan).
The scanner can now proceed with a more detailed scan.Click on the Next button to continue.
1919Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
19. The Aadva Lab Scan can make several scans, depending on the angle and the level of detail which is required for the reconstruction type. For the anatomic crown reconstruction of tooth 16, the digital image has been cropped and trimmed down to the single tooth.
20. If the scanned image contains incomplete data (e.g. holes in the model surface), the scan software allows you the option to select and close any holes in the image.
Aadva® Lab Scan
2020 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
21. Once the dental model has been scanned sufficiently, the user is given the option to save and close the scan wizard or to start the restoration design.
22. After scanning and/or designing, the dental technician has the option to upload the scan for further CAD/CAM processing.
2121Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
1. SINGLE DIE HOLDER
1. Start the dental manger software via the application shortcut on the desktop.
2. How to use : - Apply an amount of putty in the central part of the holder without covering any of the reference points.- Fix the die or other scanable object firmly in the putty reassuring that the object will not move
during scanning.
2. MULTI DIE HOLDER 5 UNITS
1. Indication : scanning of up to 5 dies / abutments of different patients
2. How to use : - apply an amount of putty in the 5 pits of the holder without covering any of the reference points.- Fix the dies or scanable objects firmly in the holder reassuring that this object will not move during
scanning.
ACCESSORIES
Aadva® Lab Scan
2222 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
- In Aadva Manager select ‘Multiedie mode’ to visualize the holder and create the project as usual in Exocad.
Aadva LAB SCAN GC 22 Version 1.21 V4903l J u l y 2016
1) Indication : scanning of up to 5 dies / abutments of different patients
2) How to use : -‐ apply an amount of putty in the 5 pits of the holder without covering any of the reference points.
-‐ Fix the dies or scanable objects firmly in the holder reassuring
that this object will not move during scanning.
-‐ In Aadva Manager select ‘Multiedie mode’ to visualize the holder and create the project as usual in Exocad.
3) Multi die holder 10 units :
UnknownFormatted: Font:12 ptUnknownFormatted: Font:12 pt
2323Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva® Lab Scan
Aadva LAB SCAN GC 23
Aadva® Lab Scan
1) Indication : scanning of up to 10 dies / abutments (different patients)
2) How to use :-‐ apply an amount of putty into the 10 pits of the holder
without covering any of the reference points.
-‐ Fix the dies or scanable objects firmly into the holder reassuring that these objects will not move during scanning
-‐ In Aadva Manager select ‘Multidie mode’ to visualize the holderd and create the project as usual in Exocad.
4) Model holder low :
UnknownFormatted: Font:16 pt
UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))
UnknownFormatted: Font:12 ptUnknownFormatted: Font:12 pt
3. MULTI DIE HOLDER 10 UNITS
1. Indication : scanning of up to 10 dies / abutments (different patients)
2. How to use :- apply an amount of putty into the 10 pits of the holder without covering any of the reference points. - Fix the dies or scanable objects firmly into the holder reassuring that these objects will not move during
scanning
- In Aadva Manager select ‘Multidie mode’ to visualize the holderd and create the project as usual in Exocad.
2424
Aadva LAB SCAN GC 24 Version 1.21 V4903l J u l y 2016
1) Indication : scanning of dental models
2) How to use : -‐ fix the model firmly and as much to the back as possible
-‐ Screw the hexagonal scanning dot onto the model holder in
any desired position so that the reference points are situated just above the occlusal plane.
-‐ Execute the scan as usual in Exocad.
5) Model holder high :
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))
UnknownFormatted: Font:12 pt
Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
4. MODEL HOLDER LOW
1. Indication : scanning of dental models
2. How to use : - fix the model firmly and as much to the back as possible
- Screw the hexagonal scanning dot onto the model holder in any desired position so that the reference points are situated just above the occlusal plane.
- Execute the scan as usual in Exocad.
2525
Aadva LAB SCAN GC 24 Version 1.21 V4903l J u l y 2016
1) Indication : scanning of dental models
2) How to use : -‐ fix the model firmly and as much to the back as possible
-‐ Screw the hexagonal scanning dot onto the model holder in
any desired position so that the reference points are situated just above the occlusal plane.
-‐ Execute the scan as usual in Exocad.
5) Model holder high :
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))
UnknownFormatted: Font:12 pt
Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva® Lab Scan
5. MODEL HOLDER HIGH
1. Indication : scanning of dental models
2. How to use : - fix the model firmly and as much to the back as possible
- Screw the hexagonal scanning dot onto the model holder in any desired position so that the reference points are situated just above the occlusal plane.
- Execute the scan as usual in Exocad.
2626 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
6. MODEL PLATE WITH PUTTY
1. Indication : scanning of dental models or other dental objects that, due to their size, do not fit in the model holder low or model holder high.
2. How to use : - Fix the model with some putty in such a way that it is central to the camera of the scanner and that the
reference points of the hexagonal scanning dot are also captured by this camera. Execute the scan as usual in Exocad.
Aadva LAB SCAN GC 26 Version 1.21 V4903l J u l y 2016
1) Indication : scanning of dental models or other dental objects that, due to their size, do not fit in the model holder low or model holder high.
2) How to use : -‐ Fix the model with some putty in such a way that it is central to the camera of the scanner and that the reference points of the hexagonal scanning dot are also captured by this camera. Execute the scan as usual in Exocad.
7) Impression model holder :
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))
2727Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva® Lab Scan
7. IMPRESSION MODEL HOLDER
1. Indication : Scanning of physical dental impressions
2. How to use : - Assemble the impression model holder onto a regular model holder low by removing the anterior retention
screws and using the posterior retention screws to fix the impression holder.
- Insert the impression by means of the impression tray handle into the holder and fix with the black fixation screw on top of the impression holder. Apply scan spray if necessary to scan silicon impression.
Remove antorior retention pins
Fix impression holder with posterior retention pins
Fix retention pins into the retention holes of the impression holder
2828 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
- After having created the Exocad project, open Aadva Scanner with ‘Impression scan’, as indicated in image below.
- Make sure to select ‘Impression’ in the Scan Wizard.
Aadva LAB SCAN GC 28 Version 1.21 V4903l J u l y 2016
-‐ After having created the Exocad project, open Aadva Scanner with ‘Impression scan’, as indicated in image below.
-‐ Make sure to select ‘Impression’ in the Scan Wizard.
8) Wax up frame :
UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))
UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))
UnknownFormatted: Font:Bold, Underline
Aadva LAB SCAN GC 28 Version 1.21 V4903l J u l y 2016
-‐ After having created the Exocad project, open Aadva Scanner with ‘Impression scan’, as indicated in image below.
-‐ Make sure to select ‘Impression’ in the Scan Wizard.
8) Wax up frame :
UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))
UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))
UnknownFormatted: Font:Bold, Underline
2929Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva® Lab Scan
8. WAX UP FRAME
1. Indication : Scanning (both sides) of dental objects, like diagnostic wax-ups, dentures, provisional crowns, etc.
2. How to use : - Before using the wax up frame it is necessary to do the wax up frame calibration process in ‘Kiosk mode’ :
1) Open ‘Kiosk mode’ in Aadva Manager
Aadva LAB SCAN GC 29
Aadva® Lab Scan
1) Indications : Scanning (both sides) of dental objects, like diagnostic wax-‐ups, dentures, provisional crowns, etc.
2) How to use : -‐ Before using the wax up frame it is necessary to do the wax up frame calibration process in ‘Kiosk mode’ :
1) Open ‘Kiosk mode’ in Aadva Manager
2) Select ‘Waxup frame calibration’ and fill in the serial number of
the waxup frame holder (situated on the bottom)
UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))
UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))
Minale, Mario 23/6/16 16:40Formatted: Dutch
Aadva LAB SCAN GC 29
Aadva® Lab Scan
1) Indications : Scanning (both sides) of dental objects, like diagnostic wax-‐ups, dentures, provisional crowns, etc.
2) How to use : -‐ Before using the wax up frame it is necessary to do the wax up frame calibration process in ‘Kiosk mode’ :
1) Open ‘Kiosk mode’ in Aadva Manager
2) Select ‘Waxup frame calibration’ and fill in the serial number of
the waxup frame holder (situated on the bottom)
UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))
UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))
Minale, Mario 23/6/16 16:40Formatted: Dutch
3030 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
2) Select ‘Waxup frame calibration’ and fill in the serial number of the waxup frame holder (situated on the bottom)
3) Scan the waxup frame one sided
4) Check the reference points of the scan and click ‘OK’
Aadva LAB SCAN GC 30 Version 1.21 V4903l J u l y 2016
3) Scan the waxup frame one sided
4) Check the reference points of the scan and click ‘OK’
5) Rotate the upper part of the wax up frame, so that the arrows are again facing each other.
Aadva LAB SCAN GC 30 Version 1.21 V4903l J u l y 2016
3) Scan the waxup frame one sided
4) Check the reference points of the scan and click ‘OK’
5) Rotate the upper part of the wax up frame, so that the arrows are again facing each other.
Aadva LAB SCAN GC 30 Version 1.21 V4903l J u l y 2016
3) Scan the waxup frame one sided
4) Check the reference points of the scan and click ‘OK’
5) Rotate the upper part of the wax up frame, so that the arrows are again facing each other.
3131Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva® Lab Scan
5) Rotate the upper part of the wax up frame, so that the arrows are again facing each other.
6) Check the alignment of the reference points and dlick ‘OK’
7) Approve component if alignment is ok. The waxup frame is now calibrated and will be recognized when scanning.
- When creating the project, it is only possible to choose a job definition in ‘Digital copy milling’ as shown below.
Aadva LAB SCAN GC 31
Aadva® Lab Scan
6) Check the alignment of the reference points and dlick ‘OK’
7) Approve component if alignment is ok. The waxup frame is now calibrated and will be recognized when scanning.
-‐ When creating the project, it is only possible to choose a job definition in ‘Digital copy milling’ as shown below.
-‐ Before scanning check the box ‘Scan waxup both way’ in order to
generate the additional scan of the object in the waxup frame.
Aadva LAB SCAN GC 31
Aadva® Lab Scan
6) Check the alignment of the reference points and dlick ‘OK’
7) Approve component if alignment is ok. The waxup frame is now calibrated and will be recognized when scanning.
-‐ When creating the project, it is only possible to choose a job definition in ‘Digital copy milling’ as shown below.
-‐ Before scanning check the box ‘Scan waxup both way’ in order to
generate the additional scan of the object in the waxup frame.
Aadva LAB SCAN GC 31
Aadva® Lab Scan
6) Check the alignment of the reference points and dlick ‘OK’
7) Approve component if alignment is ok. The waxup frame is now calibrated and will be recognized when scanning.
-‐ When creating the project, it is only possible to choose a job definition in ‘Digital copy milling’ as shown below.
-‐ Before scanning check the box ‘Scan waxup both way’ in order to
generate the additional scan of the object in the waxup frame.
3232 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
- Before scanning check the box ‘Scan waxup both way’ in order to generate the additional scan of the object in the waxup frame.
- When executing the ‘Waxup scan in holder’ the dental object needs to be well fixed with the metal pins, so it will not move during scanning.
- After having scanned the upper part, the software will ask to rotate the holder in order to scan the lower part.
Aadva LAB SCAN GC 32 Version 1.21 V4903l J u l y 2016
-‐ When executing the ‘Waxup scan in holder’ the dental object needs to be well fixed with the metal pins, so it will not move during scanning.
-‐ After having scanned the upper part, the software will ask to rotate the holder in order to scan the lower part.
Aadva LAB SCAN GC 32
Version 1.21 V4903l J u l y 2016
-‐ When executing the ‘Waxup scan in holder’ the dental object needs to be well fixed with the metal pins, so it will not move during scanning.
-‐ After having scanned the upper part, the software will ask to rotate the holder in order to scan the lower part.
Aadva LAB SCAN GC 32 Version 1.21 V4903l J u l y 2016
-‐ When executing the ‘Waxup scan in holder’ the dental object needs to be well fixed with the metal pins, so it will not move during scanning.
-‐ After having scanned the upper part, the software will ask to rotate the holder in order to scan the lower part.
3333Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva® Lab Scan
9. SCANFLAGS
1. Indication : Automated detection of implant brand, type, diameter and position. Please note that scan flags are implant specific. Make sure you select the proper scan flag and diameter. For the complete list of implants supported please refer to the price list.
2. How to use : - Choose the dedicated scanflag for the specific implant interface and fix it with an appropriate screw into
the model analogue. Do not use the same scanflag in two different positions.
- Open the scan server with scan and check in the scan flow if the first scan to be done is ‘Implant position scan ….’
- If not all scanflags are detected during the scan, the software will indicate this with a warning : ‘Scan adapter number mismatch : found yyy Expected xxx’,. In this case an add scan needs to be done in order to detect all scanflags.
Aadva LAB SCAN GC 34 Version 1.21 V4903l J u l y 2016
1) 2) How to use : -‐ Choose the dedicated scanflag for the specific
implant interface and fix it with an appropriate screw into the model analogue. Do not use the same scanflag in two different positions.
-‐ Open the scan server with scan and check in the scan flow if the
first scan to be done is ‘Implant position scan ….’ -‐ If not all scanflags are detected during the scan, the software will
indicate this with a warning : ‘Scan adapter number mismatch : found yyy Expected xxx’,. In this case an add scan needs to be done in order to detect all scanflags.
10) Scanbodies
UnknownFormatted: Font:Bold, Underline
3434
Aadva LAB SCAN GC 35
Aadva® Lab Scan
1) Indication : -‐ Detection of an implant position Please Note that scan bodies are implant specific. Make sure you select the proper scan body and diameter. For the complete list of implants supported please refer to the price list
2) How to use : -‐ Before scanning a scanbody, make sure that the specific shape is in the implant library. -‐ Fix the scanbody with an appropriate screw into the model analogue. -‐ Open the scan server with ‘Scan Optical Abutment’ and check if the first scan in the scan flow is ‘Scan Abutment scan’
-‐ For further progress of the design proces, follow the Exocad
training syllabus
Articulator frame calibration You will receive the following items (the baseplate in the frame will depend on your articulator type)
UnknownFormatted: Font:Bold, Underline
UnknownFormatted: Font:14 pt, Kern at 18 ptUnknownFormatted: Font:14 pt, Kern at 18 pt
Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
10. SCANBODIES
1. Indication : Detection of an implant position Please Note that scan bodies are implant specific. Make sure you select the proper scan body and diameter. For the complete list of implants supported please refer to the price list.
2. How to use : - Before scanning a scanbody, make sure that the specific shape is in the implant library.- Fix the scanbody with an appropriate screw into the model analogue.- Open the scan server with ‘Scan Optical Abutment’ and check if the first scan in the scan flow is ‘Scan
Abutment scan’
- For further progress of the design proces, follow the Exocad training syllabus
Aadva LAB SCAN GC 35
Aadva® Lab Scan
1) Indication : -‐ Detection of an implant position Please Note that scan bodies are implant specific. Make sure you select the proper scan body and diameter. For the complete list of implants supported please refer to the price list
2) How to use : -‐ Before scanning a scanbody, make sure that the specific shape is in the implant library. -‐ Fix the scanbody with an appropriate screw into the model analogue. -‐ Open the scan server with ‘Scan Optical Abutment’ and check if the first scan in the scan flow is ‘Scan Abutment scan’
-‐ For further progress of the design proces, follow the Exocad
training syllabus
Articulator frame calibration You will receive the following items (the baseplate in the frame will depend on your articulator type)
UnknownFormatted: Font:Bold, Underline
UnknownFormatted: Font:14 pt, Kern at 18 ptUnknownFormatted: Font:14 pt, Kern at 18 pt
ARTICULATOR FRAME CALIBRATION
You will receive the following items (the baseplate in the frame will depend on your articulator type)
LOWER articulator frame (SAM) UPPER articulator frame (SAM)
3535Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva® Lab Scan
UPPER and LOWER calibration piece
Baseplate for Adesso split Baseplate for Artex
1. The first step is to fix the articulator pieces into the articulator (with the matching base plate)
In order to use the new articulator functions you will have to do an articulator calibration first.
3636 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
This calibration has to be performed only once!
2. Open the ‘Kiosk’ mode by clicking on the new GC_KIOSK icon that you will find in the Aadva Manager.
3. Once the Kiosk software is open select ‘Articulator frame calibration’ and enter in ‘Case ID:’ a name or number under which you would like the file to be saved. Then continue with ‘Start’.
Aadva LAB SCAN GC 38 Version 1.21 V4903l J u l y 2016
3. Once the Kiosk software is open select ‘Articulator frame calibration’ and enter in ‘Case ID:’ a name or number under which you would like the file to be saved. Then continue with ‘Start’.
4. Follow the instructions on the screen.
a) First scan will be the empty LOWER b) Scan the LOWER frame with the articulator frame. Enter the frame calibration object inserted. in the scanner and click on ‘Ok’.
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
Aadva LAB SCAN GC 38 Version 1.21 V4903l J u l y 2016
3. Once the Kiosk software is open select ‘Articulator frame calibration’ and enter in ‘Case ID:’ a name or number under which you would like the file to be saved. Then continue with ‘Start’.
4. Follow the instructions on the screen.
a) First scan will be the empty LOWER b) Scan the LOWER frame with the articulator frame. Enter the frame calibration object inserted. in the scanner and click on ‘Ok’.
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
3737Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva® Lab Scan
Aadva LAB SCAN GC 39
Aadva® Lab Scan
c) Scan UPPER frame with calibration d) Scan the empty UPPER frame object
e) Scan the bottom of the articulator f) If satisfied with the result base plate. Fix the calibration continue by clicking on the green object (with articulator plate) on check mark. a model holder.
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
Aadva LAB SCAN GC 39
Aadva® Lab Scan
c) Scan UPPER frame with calibration d) Scan the empty UPPER frame object
e) Scan the bottom of the articulator f) If satisfied with the result base plate. Fix the calibration continue by clicking on the green object (with articulator plate) on check mark. a model holder.
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
Aadva LAB SCAN GC 39
Aadva® Lab Scan
c) Scan UPPER frame with calibration d) Scan the empty UPPER frame object
e) Scan the bottom of the articulator f) If satisfied with the result base plate. Fix the calibration continue by clicking on the green object (with articulator plate) on check mark. a model holder.
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
Aadva LAB SCAN GC 39
Aadva® Lab Scan
c) Scan UPPER frame with calibration d) Scan the empty UPPER frame object
e) Scan the bottom of the articulator f) If satisfied with the result base plate. Fix the calibration continue by clicking on the green object (with articulator plate) on check mark. a model holder.
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
Aadva LAB SCAN GC 39
Aadva® Lab Scan
c) Scan UPPER frame with calibration d) Scan the empty UPPER frame object
e) Scan the bottom of the articulator f) If satisfied with the result base plate. Fix the calibration continue by clicking on the green object (with articulator plate) on check mark. a model holder.
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
Aadva LAB SCAN GC 39
Aadva® Lab Scan
c) Scan UPPER frame with calibration d) Scan the empty UPPER frame object
e) Scan the bottom of the articulator f) If satisfied with the result base plate. Fix the calibration continue by clicking on the green object (with articulator plate) on check mark. a model holder.
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
4. Follow the instructions on the screen.
a) First scan will be the empty LOWER articulator frame. Enter the frame in the scanner and click on ‘Ok’.
b) Scan the LOWER frame with the calibration object inserted.
c) Scan UPPER frame with calibration object
d) Scan the empty UPPER frame
e) Scan the bottom of the articulator object (with articulator plate) on a model holder.
f) If satisfied with the result check mark.
3838 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva LAB SCAN GC 40 Version 1.21 V4903l J u l y 2016
g) If the base plate has been h) The file will be saved on C: Drive calculated by the system choose in the Kiosk-‐data. the articulator type and click on ‘Export’.
5. Now you can use these articulator frames in the Aadva software to do the articulator scan.
SEND FILE VIA DSP
1) Once the design is finished and saved open DentalDB and click on GC_DSP
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
Tomelleri, Marco 22/6/16 16:02Formatted: English
Aadva LAB SCAN GC 40 Version 1.21 V4903l J u l y 2016
g) If the base plate has been h) The file will be saved on C: Drive calculated by the system choose in the Kiosk-‐data. the articulator type and click on ‘Export’.
5. Now you can use these articulator frames in the Aadva software to do the articulator scan.
SEND FILE VIA DSP
1) Once the design is finished and saved open DentalDB and click on GC_DSP
UnknownFormatted: Font:14 pt, Kern at 18 pt
UnknownFormatted: Font:14 pt, Kern at 18 pt
Tomelleri, Marco 22/6/16 16:02Formatted: English
5. Now you can use these articulator frames in the Aadva software to do the articulator scan.
g) If the base plate has been system choose the articulator type and click on ‘Export’.
h) The file will be saved on C: Drive calculated by the in the Kiosk-data.
3939Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva® Lab Scan
SEND FILE VIA DSP
1) Once the design is finished and saved open DentalDB and click on GC_DSP
Aadva LAB SCAN GC 41
Aadva® Lab Scan
2) Enter username and password and check ‘I accept the license agreement’.
Then click on Upload.
3) The file will now be uploaded to the DSP account.
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
Aadva LAB SCAN GC 41
Aadva® Lab Scan
2) Enter username and password and check ‘I accept the license agreement’.
Then click on Upload.
3) The file will now be uploaded to the DSP account.
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
2) Enter username and password and check ‘I accept the license agreement’. Then click on Upload.
4040 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
3) The file will now be uploaded to the DSP account.
4) After the upload is finished the DSP account will open automatically
Aadva LAB SCAN GC 42 Version 1.21 V4903l J u l y 2016
4) After the upload is finished the DSP account will open automatically
5) Open the Tab ‘Home’. You can find the uploaded case under ‘My Cases’.
6) Click on the case to open further information/options.
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
Aadva LAB SCAN GC 42 Version 1.21 V4903l J u l y 2016
4) After the upload is finished the DSP account will open automatically
5) Open the Tab ‘Home’. You can find the uploaded case under ‘My Cases’.
6) Click on the case to open further information/options.
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
5) Open the Tab ‘Home’. You can find the uploaded case under ‘My Cases’.
Aadva LAB SCAN GC 42 Version 1.21 V4903l J u l y 2016
4) After the upload is finished the DSP account will open automatically
5) Open the Tab ‘Home’. You can find the uploaded case under ‘My Cases’.
6) Click on the case to open further information/options.
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
4141Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva® Lab Scan
6) Click on the case to open further information/options.
Aadva LAB SCAN GC 43
Aadva® Lab Scan
7) Click on the icon GC Factory and GC Milling to send the file to the GC milling center.
8) Confirm your choice by clicking on ‘Yes’
9) You can find all your sent cases under the Tab ‘Transactions’.
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
Tomelleri, Marco 22/6/16 16:05Formatted: English
7) Click on the icon GC Factory and GC Milling to send the file to the GC milling center.
Aadva LAB SCAN GC 43
Aadva® Lab Scan
7) Click on the icon GC Factory and GC Milling to send the file to the GC milling center.
8) Confirm your choice by clicking on ‘Yes’
9) You can find all your sent cases under the Tab ‘Transactions’.
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
Tomelleri, Marco 22/6/16 16:05Formatted: English
8) Confirm your choice by clicking on ‘Yes’
Aadva LAB SCAN GC 43
Aadva® Lab Scan
7) Click on the icon GC Factory and GC Milling to send the file to the GC milling center.
8) Confirm your choice by clicking on ‘Yes’
9) You can find all your sent cases under the Tab ‘Transactions’.
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
UnknownFormatted: Font:12 pt
Tomelleri, Marco 22/6/16 16:05Formatted: English
9) You can find all your sent cases under the Tab ‘Transactions’.
4242 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
1. Keep the device dust-free and covered when not in use.
2. The LED in the scan head is not susceptible to damage caused by prolonged use. The Aadva Lab Scan can be kept running if necessary, but the device can be turned off as an energy-
saving measure.
3. Unplug the device and remove any objects in the scanning tray prior to routine cleaning and maintenance of the unit.
4. Do not attempt to clean the sensor head inside the casing of the scanner, as this may damage the device.
5. Use a soft cloth moistened with a small amount of water or other detergents used for cleaning computers. Wipe the cloth on exposed surfaces of the Aadva Lab Scan to remove dust or other particles. Do not use
cleaning sprays on the surface of the scanner.
6. The dental clamps and trays can also be cleaned using a cloth moistened with a small amount of water.
7. While cleaning the device with a moistened cloth, make sure that the capacitors or other electrical parts of the Aadva Lab Scan do not come into contact with any moisture.
8. Do not place any objects on top of or above the scanner, as this may interfere with the scanning functionality.
Make sure that no liquids could accidentally spill on the device.
CARE OF THE UNIT
4343Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva® Lab Scan
If any difficulty with the Aadva Lab Scan unit is experienced, please check below for possible causes of the problem and suggested corrective actions before seeking further advice or repair from an authorized GC Dealer or Service Center.
Problem Check Corrective Action
When the Aadva Lab Scan is switched ON, no light appears.
Is the power cord correctly attached to the Aadva Lab Scan and to the power outlet?
Switch the device OFF. Unplug and plug the power cord correctly into the Aadva Lab Scan and the power outlet. Then switch the device ON again and wait for a few seconds.
The Aadva Lab Scan application does not start. The following error message appears:
Is the software dongle (page 7) available and correctly inserted at the USP port of the workstation?
Make sure that the software dongle (page 7) is attached. Also, make sure that your license is currently valid. (how?)
The Aadva Lab Scan software is running, but the scan job does not start (or the scanner software indicates that no Aadva Lab Scan was found). The following error message appears:
Is the hardware dongle (page 7) available and correctly inserted at the USP port of the workstation?
Make sure that the hardware dongle (page 7) is attached. Also, make sure that your license is currently valid.
TROUBLESHOOTING
4444 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
An error occurs during the preparation scan:
The scan wizard shows a red button, implying that there is no connection to the device.
Is the USB cable connected to the Aadva Lab Scan and the workstation?
Make sure that the USB cable has been attached correctly. The Type B end of the USB cable connects to the Aadva Lab Scan, while the Type A connector is inserted into the workstation.
Is the Aadva Scan switched on? Check to see if the Aadva Lab Scan is emitting a blue light. Otherwise, turn the unit on.
During the scan, an error message appears during scan data triangulation:
Has the device completed scanning the dental model?
Adjust the view of the model. If necessary, trim or crop any parts of the initial scan which are not relevant for the reconstruction type.
The device has completed the scan but the scanned object appears incomplete.
Has the device completed scanning the dental model?
Adjust the view of the model. If necessary, trim or crop any parts of the initial scan which are not relevant for the reconstruction type.
The power supply is interrupted during the scanning procedure.
Restart the scanner. If necessary reload the scan profile or setup a new one.
4545Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva® Lab Scan
Your product is marked with this symbol for an environmentally responsible recycling/disposal of your product. Do not discard it with normal waste. The black line under the crossed-out waste bin symbol signifies that this product has been put into circulation after 13.08.2005 (see Directive 2012/19/EU and EN 50419:2005).
Please be aware that this product has been submitted to the European Directive 2012/19/EU (WEEE) and the laws applicable in your country for environmentally responsible recycling/disposal. Please ask GC Europe or your GC dealer when and where your product will have to be disposed of.
DISPOSAL OF THE UNIT
Your Aadva Lab Scan has a warranty period of two years from the date of purchase.
In case of a breakdown during this period, repair will be carried out free of charge provided that the unit has been used under normal conditions and according to the instructions for use. In case of damage under these conditions stop using the instrument immediately and contact GC Europe (contact details see last page of this instructions for use).
WARRANTY AND REPAIR
Only GC Europe may perform the maintenance, service and repair. In case of damage please contact GC Europe (contact details see front page of this instructions for use).
Aadva Lab Scanner Rating Plate
MAINTENANCE AND SERVICE
4646 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016
Aadva Lab Scanner
Name/Model Aadva Lab Scan
Light source LED (Blue)
Resolution 2 megapixels
Number of cameras 2
Positioning unit 2 axis (tilt & swivel)
Accuracy 6 microns
Scan time < 60 seconds
Scan accuracy < 10 µm
Supply voltage, frequency range and current 110/220 V CA, 50-60 Hz, 3,5/2 A
Dimensions (w x h x l) 400 x 600 x 300 mm
WeightScanner 27 kg
PC 20 kg
Environmental conditions in use
For indoor use only
Temperature 5°- 40 °C (41-104°F)
Altitude up to 2000 m or
Barometric pressure 700 hPa – 1060 hPa
Relative Humidity 30-75%
Main supply voltage fluctuation +/- 10%
Pollution degree 2
Temperature -10°- +40 °C (14-104°F)
Environmental conditions at transport and storage
Altitude up to 2000 m or
Barometric pressure 700 hPa – 1060 hPa
Relative Humidity 30-75%
Protection class XP00
StandardsEN 61010-1
EN 61326-1
Regulatory Conformity
RoHS2
WEEE
Note: The specifications of the unit may be subject to change without notice.
TECHNICAL SPECIFICATIONS
9904
87 -
Y95
00
Manufactured & distributed by: GC Europe N.V.Researchpark Haasrode-Leuven 1240Interleuvenlaan 33B-3001 LeuvenTel: +32.16.74.10.00Fax: +32.16.40.02.14info@gceurope.comwww.gceurope.com
Distributed by:GC UNITED KINGDOM Ltd.Coopers CourtNewport Pagnell UK-Bucks. MK16 8JSTel. +44.1908.218.999 Fax. +44.1908.218.900 info@uk.gceurope.comhttp://uk.gceurope.com
The Aadva® Lab Scan instructions for use can be downloaded at the following web address http://www.gceurope.com or can be requested at no charge at following contacts: digitalservices@gceurope.comPhone 0032 16742600