LARGE GROUPMANUFACTURERS UPDATE
SATURDAY – APRIL 14, 20128:00 A.M. – 10:00 A.M.
LARGE GROUPMANUFACTURERS UPDATE
SATURDAY – APRIL 14, 20128:00 A.M. – 10:00 A.M.
TopicsNew BSC Model NU-481
Use and Service of DC ECM MotorsNew Slanted A2 – NU-475, 477, 480
Miscellaneous Service Updates
TopicsNew BSC Model NU-481
Use and Service of DC ECM MotorsNew Slanted A2 – NU-475, 477, 480
Miscellaneous Service Updates
NUAIRE WEBSITENUAIRE WEBSITEReceiving TB’s and have access to Website’s secure area?
Let us know and provide current contact info if you need access.
NUAIRE WEBSITENUAIRE WEBSITE
Website – Support Manuals Technical Articles Technical Bulletins Service Companies Training (On-line Registration) Service Procedures Warranty Information Cannot log on – Call Service
for help 800-328-3352
NU-475, NU-477, NU-480/481A2 SLANTED FRONT
NU-475, NU-477, NU-480/481A2 SLANTED FRONT
TouchLink Control (NU-480/481)
NU-475 (NU-425 Controls)
NU-477 (NU-437 Controls)
60 fpm Down-flow
105 fpm Inflow
DC-ECM Motor
3” Constricted Inflow Method
Permanent Pressure Plenum
The CellGard HD (Hazardous Drug) ES (Energy Saver) Class II, Type A2 Biological Safety Cabinet offers personnel, product and environmental protection for handling of hazardous particulate drugs and powders
CELLGARD™ HD NU-481 ESCELLGARD™ HD NU-481 ES
CELLGARD™ HD NU-481 ESAIR FLOW
CELLGARD™ HD NU-481 ESAIR FLOW
Clean Oval Canister HEPA Filtered Air
Clean HEPA Filtered Air
Room Air
Contaminated Work Surface Air
The Ease of Bag-In / Bag-Out Remove Work Surface Remove paper catch grill over each oval HEPA prefilter Seal contaminated interior of each oval HEPA prefilter Remove four (4) fasteners per each oval HEPA prefilter Bag-Out used oval HEPA prefilter and replace
CELLGARD™ HD NU-481 ESCELLGARD™ HD NU-481 ES
CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER
INTEGRITY TESTING
CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER
INTEGRITY TESTING
PAO Wand
Filter Gard
CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER
INTEGRITY TESTING
CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER
INTEGRITY TESTING Set the NU-481 into calibration mode
Go to Sensor Setup / Calibrate Downflow Sensor / Blower Duty Cycle
Set Blower Duty cycle to the RPM at the current calibrated airflow velocity
Keep blower at this RPM (constant) throughout testing process
Remove the work tray
CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER
INTEGRITY TESTING
CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER
INTEGRITY TESTINGINTERNAL REFERENCE CALCULATIONS
Model Size *Supply Area (ft2)(m2) Inflow Area (ft2)(m2) 8” Access
Inflow Area (ft2)(m2) 10” Access
400 8.53 (.242) 2.38 (.221) 3.22 (.299)
500 10.74 (.304) 3.15 (.293) 4.05 (.376)
600 12.95 (.367) 3.92 (.364) 4.89 (.454)
*Measured 4 Inches (102mm) above the bottom edge of the window
Laskin Nozzle Concentration Formula
# Nozzles x 135 CFM x 100 ug/L ChallengeDownflow (CFM) + Exhaust (CFM) Concentration (ug/L)
# Nozzles x 229 CMH x 100 ug/L ChallengeDownflow (CMH) + Exhaust CMF) Concentration (ug/L)
=
=
CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER
INTEGRITY TESTING
CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER
INTEGRITY TESTING Filter Gard is used to direct the
PAO from the wand into one area of prefilter at a time
Filter Gard can be used for initial certification and after 1st annual certification
Replace HEPEX Gard (B-481-8691) Fiberglass rods can be re-used
CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER
INTEGRITY TESTING
CELLGARD™ HD NU-481 ESPRIMARY HEPA FILTER
INTEGRITY TESTING Move PAO Wand throughout filter, rotating the filter guard
Leak is detected if a sustained leak rate greater than .005 is detected at the challenge port on top of the cabinet
ECM 5.0 Genteq Motor
1450 RPM
10” Forward Curved Fan (was 9”)
PWM (DC signal) varies the motor speed Motor Control, Monitors RPM and torque which delivers
constant air volume as HEPA Filters load
NU-437, NU-440, NU-475, NU-477, NU-480, NU-481 (A2 units)
NU-427 (B1), NU-430, NU-435 (B2) - DC ECM Motor is standard Optional on NU-425 (A2)
ELECTRONICALLY COMMUTATED MOTOR
ELECTRONICALLY COMMUTATED MOTOR
DC ECM CONTROL SYSTEM A2DC ECM CONTROL SYSTEM A2
NU-425 – ES (Energy Saver) - Optional & Field Upgradeable
NU-475, NU-629 – ES (Energy Saver) – Standard
425 TYPE DC SPEED CONTROLLEROPTIONAL ON NU-425
425 TYPE DC SPEED CONTROLLEROPTIONAL ON NU-425
12 Volt DC Power supply
Automatic Constant Airflow volumeThru motor rpm feedback
Manual Airflow volume adjustment (POT)
On-board jumper positions for correct cabinet size
Speed Cont. locks in setting after 1 Minute idle or certifier sets via push button & LED turns off
425 TYPE DC MOTORSPEED CONTROLLER SYSTEM
425 TYPE DC MOTORSPEED CONTROLLER SYSTEM
425 TYPE DC MOTORSPEED CONTROLLER SYSTEM
425 TYPE DC MOTORSPEED CONTROLLER SYSTEM
SPEED CONTROLLER JUMPERSSPEED CONTROLLER JUMPERS
3’ 0 0
4’ 1 0
5’ 0 1
6’ 1 1
J2 J3
0 = Not Jumped1 = Jumped
GENERAL AIRFLOW BOARD (GAF)GENERAL AIRFLOW BOARD (GAF)
NU-437, NU-477 (A2) – ES (Energy) -Saver Standard
GAF Also used in NU-813 , NU-126 , NU-619 - Change Station & All Isolators models
AC Motor Output: Older Series NU-437
AC Impeller Output: Older Series NU-813; NU-126; NU-619 and Isolators
DC ECM Motor Output: Std. on Current NU-437, NU-477
DC EC Impeller Output: Std. on All Isolators (S-10^) NR/NTE/PR; NU-813, NU-126, NU-619
New Software Release (Rev 1.5.0001), STB0201 Corrects increase motor speed (PWM signal) due to site
specific events, (i.e. power fluctuations, frequent on/off cycles) STB0202: RF Upgrade Installation Instructions
GAF CONTOR BOARDMULTIPLE BLOWER CONTROLS
GAF CONTOR BOARDMULTIPLE BLOWER CONTROLS
GAF MAIN CONTOR BOARD LAYOUTGAF MAIN CONTOR BOARD LAYOUT
AC Blower Output
AC Blower POT.
Board ACInput
Connection
EC #1VDC T.P.’s
EC Connector & POT #1
EC Connector and POT #2
EC #2VDC T.P.’s
DC ECM Blower 4 PinMotor Control
Connector
AC T.P.’s
GAF AC/D MOTOR / IMPELLERAIRFLOW ADJUSTMENTS
GAF AC/D MOTOR / IMPELLERAIRFLOW ADJUSTMENTS
AC motor – Airflow adjusted via the POT located under heat-sink
DC ECM motors - Airflow adjusted via sequence of control panel keys actions. Cal mode = press & hold Alarm Silence key, hit FL light on, light off key;Increase Airflow = UV key and decrease = FL light key
DC EC impellers utilize another set of potentiometers,on GAF board, Std. all Isolators (S-10^)
GAF Control board EC Impeller ConnectionsGAF Control board EC Impeller Connections
NU-437 (SERIES 60^) / 477 (A2)AIRFLOW CALIBRATION
NU-437 (SERIES 60^) / 477 (A2)AIRFLOW CALIBRATION
The DC-ECM System works like this:
1. AC voltage is applied directly to the DC motor’s control module
2. The motor control module (Microprocessor) converts the AC to DC for Internal electronics and to power the motor
3. In addition, there is a second electrical cable that goes from the Main Control Board’s speed controller DC output to the motor. This generates a pulse-width-modulation (PWM) 2.5 – 8 VDC control signal and returns an RPM feedback signal from the motor to the speed controller CKT.
GAF CONTROL BOARD JUMPERSDC ECM CONTROL SIGNALS
GAF CONTROL BOARD JUMPERSDC ECM CONTROL SIGNALS
Nite Care Setback
GAF BOARD AC OR DCJUMPER WIRE
GAF BOARD AC OR DCJUMPER WIRE
Cut Jumper Wire for AC Motor
NU-437 / 477 (A2)CALIBRATION AIRFLOW
NU-437 / 477 (A2)CALIBRATION AIRFLOW
Once the airflow calibration is complete – calibrate the FlowGard Airflow Monitor
1. Press and hold ( ) and ( ) arrow keys simultaneously for 3 seconds until the center green LED blinks.
2. Allow the Unit to run for a minimum of 2 minutes.
3. Press RESET Key to enter the nominal airflow value; the Green LED will stop blinking and display normally.
This sets the high/low alarm limits +/-0.13 w.g. from nominal. Reference the operation manual for the NU-437/477 for more in-depthinformation and calibration procedures. If you require help, or havequestions, contact NuAire’s Technical Support Depart.
BSCC TOUCHLINK AIRFLOWCONTROL SYSTEM
BSCC TOUCHLINK AIRFLOWCONTROL SYSTEM
BSCC CONTROL BOARD BSCC CONTROL BOARD
DC ECM Output: NU-440 (S-60^); NU-480, NU-481 (A2)NU-430/435 (B2) (S – 60^);
NU-427 (B1) (S-60^)
Note: Control Board is Backwards Compatible to all TouchLinks cabinets
TOUCHLINK SOFTWARE UPDATES TOUCHLINK SOFTWARE UPDATES
New Revision Notifications are sent out to all service organizations
New Revisions are available free of charge
Use RF484 to order latest revision and cable connections
Latest TouchLink Software 61N.138N. (June 2011) Selection for AC or DC Motor Screen Saver display update Air Flow Sensor – Comm/Sensor error messages Down Flow Sensor - Averaging time/Alarm time adjustable Filter Life available in the Calibration/Service Menu Electrical power harmonic distortion compensating software
NEW TOUCHLINKCALIBRATION CARDNEW TOUCHLINK
CALIBRATION CARDTouchLink - Calibration and Reset ProceduresThe TouchLink System is a Liquid Crystal Display (LCD) that is menu driven and thus you need to enter specific menus to accomplish the maintenance tasks required. Refer to the appropriate O & M Manual for additional in depth information on a specific model. This is to be used as a reference guide, there may be some variances due to software revision levels.
From the main display hit the Menu ICON to advance to other displays. Hit the Menu ICON to return to previous displays
Cabinet Type/Motor Type- Press MENU from the main display- Select CALIBRATION/SERVICE, enter password 9876, and press ENTER- Select CABINET TYPE- Select the type of cabinet; i.e. A2, B1, B2. Types NOT ending with a letters (D,E,G) are for 115 VAC systems.- Select the model of cabinet- Hit SAVE- Press MENU- Press SET MOTOR TYPE (If available)- Select AC or DC MOTOR-Press SAVE
Find Sensors- Used for both calibration and sensor replacement- Press MENU from the main display- Go into CALIBRATION /SERVICE, enter password 9876, press ENTER- Select SENSOR SETUP- Select FIND DOWNFLOW SENSOR- Power down cabinet- Remove the 4 pin exhaust probe connector from the main control board- Power up and tap displays “Power Loss Alert”- Select FIND NEW DOWNFLOW SENSOR- Press LEARN- Acknowledge that down flow sensor is found. If not, repeat previous steps.- Power down cabinet- Reconnect 4 pin exhaust probe connector back on main control board
CLASS II TYPE A2CANOPY EXHAUST TRANSITIONS
CLASS II TYPE A2CANOPY EXHAUST TRANSITIONS
CANOPY EXHAUST TRANSITIONSCANOPY EXHAUST TRANSITIONS
Traditional Fixed Flow NU-906 / 907 / 908 (NU-475 / 477 / 480 / 481) NU-916 / 917 / 918 (NU-425 / 437 / 440) NU-926 / 927 / 928 (NU-407 / 602)
New Variable Flow NU-911 (NU-475 / 477 / 480 / 481) NU-921 (NU-425 / 437 / 440)
VARIABLE FLOW CANOPY (VFC)VARIABLE FLOW CANOPY (VFC)
Low profile (5” housing, 8½” overall)
Built-in low airflow alarm
Variable air gap range (25 to 250cfm)
VFC FUNCTION / OPTIONSVFC FUNCTION / OPTIONS
Note: VFC Setup adjustment should be made with cabinet set atnominal airflows
1. Adjust slide plates and/or exhaust volume desired so average velocity measured in the center of the (3) test points of either side is between 150fpm (.76mps) and 550fpm (2.8mps)
2. Verify containment with the capture of smoke
3. Verify low exhaust alarm with loss of smoke capture with in 15 seconds
VFC SETUP INSTRUCTIONSVFC SETUP INSTRUCTIONS
VFC INTERFACE / SERVICEVFC INTERFACE / SERVICE
Alarm contacts
Audible alarm silence contacts
Front panel access for filter testing
Decon options
NU-911/921 VFC TRANSITIONSNU-911/921 VFC TRANSITIONS
NU-602NU-602
NU-602 NEW LIFT SYSTEMNU-602 NEW LIFT SYSTEM
NU-602 NEW LIFT SYSTEMNU-602 NEW LIFT SYSTEM
RF Kit to convert older series actuators
Re-Zeros every time unit is fully retracted
NU-602NU-602
NU-612 / 617 / 619SMALL ANIMAL
CAGE CHANGING STATIONS
NU-612 / 617 / 619SMALL ANIMAL
CAGE CHANGING STATIONS
NU-612 / 617 / 619 PreventativeMaintenance Check List
NU-612 / 617 / 619 PreventativeMaintenance Check List
NU-612 / 617 / 619PREVENTATIVE MAINTENANCE
CASTER INSPECTIONREF. STB0185 / STB0214
NU-612 / 617 / 619PREVENTATIVE MAINTENANCE
CASTER INSPECTIONREF. STB0185 / STB0214
NU-612/617CASTER FAILURES
REF. STB0185
NU-612/617CASTER FAILURES
REF. STB0185
NU-612 / 617 / 619P.M.C.I.
REF. STB0185 / STB0214
NU-612 / 617 / 619P.M.C.I.
REF. STB0185 / STB0214
PHARMAGARDPHARMAGARD
PR-797 – Inspect outer door hinge bolts for wear during normal certification – replace as needed – STB0191
NuAire provides replacement bolts at No Charge –RF525 – keep some in your stock
OUTER DOOR HINGE BOLTSOUTER DOOR HINGE BOLTS
Inter-chamber door not staying up – RF 508/509/510 Available at N/C, replace during normal certification for improved latching
PR797 Series 3 – 5RF508
NR797 Series 3 – 7RF509 (LH); RF510 (RH)
NTE797 Series 3 – 5RF509 (LH); RF510 (RH)
Current inter-chamber dooris counter balanced
INTER-CHAMBER DOORINTER-CHAMBER DOOR
NU-797 SHARPS PLUG RETROFIT& SHARPS UPGRADE
NU-797 SHARPS PLUG RETROFIT& SHARPS UPGRADE
NU-430 SERIES 60^NEW B2 FEATURESNU-430 SERIES 60^NEW B2 FEATURES
Airflow Measurement Procedure Modified Down-flow Performed First Inflow taken with DIM and Supply Blower ON
DC ECM Motor
Permanent Pressure Plenum
Exhaust monitored via Pressure Grid
DC ECM Motor
Permanent Pressure Plenum
Exhaust monitored via Pressure grid
NU-427 SERIES 60^NEW B1 FEATURESNU-427 SERIES 60^NEW B1 FEATURES
PRESSURE SWITCH RETROFITVACS REPLACEMENT INSTALL.PRESSURE SWITCH RETROFIT
VACS REPLACEMENT INSTALL.
PTB0170 TouchLink Software Loading Procedure
STB0136 Verify B1/B2 Exhaust Probe Cal. (S 20 – 49)
STB0185 NU-612/617 Preventive Maintenance
STB0187 Convert NU-425 Series 20^ to DC ECM Motor
STB0191 PR797 Isolator, Outer Door Hinge Bolt (PM)
STB0192 TouchLink, Master Reset Procedure
STB0197 GAF Control Board Usage/Calibration
STB0201/202 GAF Module Software Update (RF541)
TECHNICAL BULLETINS / RF KITSTECHNICAL BULLETINS / RF KITS
STB0204 DC ECM Speed Control Replacement (425 type)
STB0208 TouchLink Software Version 60N.136N
STB0211 TouchLink Software Version 61N.138N
STB0214 Preventative Maintenance Check List
STB0215 Pressure SW-VACS Replacement Installation
STB0216 797 Sharps Retrofit
STB0219 619 Caster Upgrade
TECHNICAL BULLETINS / RF KITSTECHNICAL BULLETINS / RF KITS