Read carefully the instructions published in this manual before the first use of the level meter. Keep the manual at a safe place. The manufacturer reserves the right to implement changes without prior notice.
INSTRUCTION MANUALprůmyslová elektronika
Ultrasonic level meters Ulm – 70
Firmware: v.2.0
Table of conTenTs
1 . Measuring principle .................................................................................................................3
2 . Range of applications ..............................................................................................................3
3 . Variants of sensors ..................................................................................................................3
4 . Dimensional drawings .............................................................................................................4
5 . Installation and putting into operation ..................................................................................4
6. Installation instructions ..........................................................................................................5
7. Electrical connection ...............................................................................................................9
8. Examples of ULM-70 connection ............................................................................................10
8.1 . Wiring diagram of the level meter with current output and PDU unit .................................10
8.2 . WiringdiagramofthelevelmeterwithcurrentoutputandPDUunit .............................. in versions for explosive areas ..........................................................................................11 8 .3 . WiringdiagramofthelevelmeterwithcurrentoutputandMGUunit ................................13
8 .4 . WiringdiagramofthelevelmeterwithModbus/RS485andMGUunit ...........................13
8 .5 . WiringdiagramofthelevelmeterwithModbus/RS485andPDUunit ............................13
9 . Set-up elements .......................................................................................................................14
10. Status signalization ................................................................................................................14
11 . Operation and setting ............................................................................................................15
10.1 . Basicconfiguration ..........................................................................................................15
10.2 . Servicesettings ...............................................................................................................18
10 .3 . Additionalfunctions .........................................................................................................20
12. HART® communication protocol ...........................................................................................22
13. Modbus® communication protocol ........................................................................................23
14 . Order code ..............................................................................................................................23
15 . Accessories ............................................................................................................................23
16 . Safety, protection, compatibility and explosion proof .......................................................24
17 . Use, manipulation and maintenance ....................................................................................24
18 . Marking of labels ....................................................................................................................25
19 . Menu structure .......................................................................................................................27
20 . Specifications .........................................................................................................................28
21 . Packing, shipping and storage .............................................................................................30
ULM–70 © Dinel, s.r.o.3
1 . Measuring principle
UltrasoniclevelmeterULM® is a compact measuring device consisting of two parts - main level me-ter(thebodywithmeasuringelectronics)anddisplaymodule.Usingtheelectroacousticconverter,the level meters transmit the sequence of ultrasonic pulses that spread towards the surface level. Theconverterrecuperatesreflectedacousticwavesthataresubsequentlyprocessedintheelec-tronicmodule.Theintelligentevaluationblockfiltersoutinterferingsignals,comparesthecleanedreceivedsignalwith the false reflectionmap (e.g. frommixers, ladders, reinforcementetc.)andselectsasuitablereflection(echo).Basedontheperiodduringwhichtheindividualpulsesspreadtowardsthesurfacelevelandbackandbasedonthemeasuredtemperatureinthetank,theinstantdistancetothesurfaceleveliscalculated.Accordingtothelevelheight,thelevelmeteroutputissetandthemeasuredvalueisdisplayedonthedisplay.
2 . range of applicaTions
Forcontinuousnon-contactlevelmeasurementofliquids(watersolutions,seweragewater,etc.),mashandpastematerials(sediments,sticks,resinsetc.)inclosedoropenvessels,sumps,reser-voirsandopenchannels.Incasethelevelofbulk-solidmaterialsismeasured,themeasurementrange is reduced.The levelmeterscancontinuouslymeasure levelsofbulk-solidmaterialswitha low concentration of dust particles. Consult the manufacturer on recommended use of the level meterforbulk-solidmaterials.
3 . VarianTs of sensors
All operations described in this instruction manual have to be carried out only by trained personnel or an accredited person. Warranty and post warranty service must be exclusively carried out by the manufacturer.
Improper use, installation or set-up of the level meter can result in crashes in the application (overfilling of the tank or damage of system components).
The manufacturer is not responsible for improper use, losses of work caused by either direct or indirect damage, and for expenses incurred during installation or use of the level meter.
safeTy
ULM –70_ –02 Measuring range from 0.15m to 2m, plasticPVDFtransmitter,mechanicalconnectionwiththreadG1".
ULM –70_–06 Measuring range from 0.25m to 6m, plasticPVDFtransmitter,mechanicalconnectionwiththreadG1½".
ULM –70_–10 Measuring range from 0.4m to 10m, plasticPVDFtransmitter,mechanical connection with thread G2¼".
ULM–70_–20 Measuring range from 0.5m to 20m, plasticPVDFtransmitter,mechanicalconnectionwithaluminiumalloyflange.
4© Dinel, s.r.o. ULM–70
4 . DiMensional Drawings
ULM–70_–02
Pg11 cable gland
Ground terminal
variant ULM-70 with protective conductor
ULM–70_–06
ULM–70_–20
ULM–70_–10
This procedure includes the following three steps.
• insTallaTion
• elecTric connecTion
• seTTing
5 . insTallaTion anD puTTing inTo operaTion
ULM–70 © Dinel, s.r.o.5
Fig. 2: Level meter dead zone
m – Dead zone (blind zone, blocking distance)
6 . InstallatIon InstructIons• Install the level meter in the vertical positionintotheupperlidofthetankorreservoirusingaweld-
ingflange,afasteningnutoraflangesothatthelevelmeteraxiscanbeperpendiculartothesurfacelevelofthemeasuredliquid(Fig.1).
• Themin.dimensional parameters to install the level meterintoalidoraceilingofatankaregiveninFig.3.
• When installing in an open channel(reservoir,drainetc.),installthelevelmeterontoabracketascloseaspossibletotheexpectedmax.level.
• Inconnectionwiththemeasurementprinciple,nosig-nals reflectedintheareaimmediatelyunderthelevelmetercanbeevaluated.The zone(Fig.2)determinesthe min. distance possible between the level meterand the highest surface level. The min. distances to themediumaregiveninthechapter"Specifications".
• It is necessary to install the levelmeter so that thebin level cannot interfere with the dead zone when filleduptothemaximum.Ifthemeasuredlevelinter-fereswiththedeadzone,thelevelmeterwillnotworkproperly.
Fig. 1: Recommended installation in the tank
Fig. 3: Installation distance from the tank wall
• If themaximumsurface level in the tank interfereswith thedeadzone, the levelmeterhastobemountedintoahigherinstallation neck.Inthisway,thetankcanbefillednearlyuptothemaximumvolume.Theinnernecksurfacehastobeevenandsmooth(withoutedgesandweldedjoints);theinneredgeshouldberoundedwheretheultrasonicwaveleavesthepipe.Theneckdiametershouldbeaslargeaspossiblebuttheneckheightshouldbeaslowaspossible.RecommendeddimensionsoftheinputneckaregiveninFig.4.
ULM–70–02 ; 10 d > 1/12 c (min. 200 mm)
ULM–70–06 d > 1/8 c (min. 200 mm)
ULM–70–20 d > 1/10 c(min. 200 mm)
Fig. 4: Possible installation of the installation neck
a – Neck heightb – Neck widthm – Dead zone (blind zone, blocking dist.)
ULM–70–02 ; 06 a < 3 bb > 100 mm
ULM–70–10 a < 1,5 bb > 100 mm
ULM–70–20 a < 1,5 bb > 150 mm
d – Distance from the tank wall
c – Maximum reach of the level meter
6© Dinel, s.r.o. ULM–70
• Iftheemittedacousticsignalofthelevelmeterisaffectedbynearobjects(roughnessonwallsofthetank,variouspartitions,mixersetc.),itisnecessarytomapfalsereflectionsbyactivatingthemode"TEACHING".Incaseofinstalledmixers,itisnecessarytoputthemixerstopositionunderthelevelmeter(directthemixerpaddletotheultrasonicsignalbeam)(Fig.5and6).
• Donotinstallthelevelmeterinorabovethefillingpoint(Fig.7).
Fig. 5: False echo from obstacles in the tank Fig. 6: False echo from the mixer paddle
Fig. 7: Level meter installation outside the influence of filling
• Incasethelevelofbulksolidsismeasured,themeasurementrangeisreducedDuetoabsorp-tionofacousticwavesbyabulkmedium,short-eningofthemeasuringrangeoccursbyupto50% depending on the grain size. We therefore recommend selecting a level meter with greater range than the maximum range of measuring the medium. It is also appropriate to use a di-rectionalhorn(seeimage8),whichreducestheshorteningofthemeasuringrange,becauseitbetterconcentratesacousticenergywhilepre-serving the same beam angle, and improvesthe sensitivity when receiving the reflectedecho. We recommend to consult the use with the manufacturer.
Fig. 8: Level meter installation in silo or hopper
hornadapter
ULM–70 © Dinel, s.r.o.7
!
!
• Scattering or attenuation of theultrasonic signal can result if the surfacelevelhasbeenmoderate-ly stirred or rippled (byamixer,coming liquid etc.). It can resultin reduction of the measurement rangeorunreliablefunctionofthelevelmeter(Fig.10).
• Rotatingmixer blades can cause thatthesurfaceisstirred,whichre-sultsinfalsereflectionsoftheultra-sonic signal from the surface level and unreliable operation of thelevelmeter(Fig.13).(obr.11).For a rippled or swirling level, you can use the directional horn to elimi-nate scattering of the ultrasonic signal.
• During filling, mixing and other processes, foam can arise on the surface level of the measured liq-uid. The foam considerably absorbs the ultrasonic sig-nal which might cause malfunction of the level meter (Fig.9).Forsuchcases, it isnecessary tosetup"SENSI-TIVITY"modeto"high"orcontactthemanufacturerifneed. Incaseofathinlayeroffoam,itisalsopossibletousethedirectionalhornforimprovingreceiptofthereflectedecho.
Fig. 9: Foam on the surface
Fig. 10: Moderately stirred surface
Fig. 11: Intensely stirred surface
• If the level sensor is mounted to bottlenecks and places with ba-rriers,ornearunevenwallsor thefilling area, where the transmissi-on signal could be distorted, werecommend using a guide tube(acoustichorn).The tubemustbemade from a single material with a smooth inner surface (see image 12a,12b).Theminimumtubedia-metermusthavethedimension"b"according to image 4 on page 5. After installing, you must performthe procedure "LEARNING". Werecommend consulting with the manufacture on the construction of theguidetube.
Obr. 12a: Short guide tube installation
ventilation holes
sleeves for mounting
Obr. 12b: Total guide tube installation
8© Dinel, s.r.o. ULM–70
• Thelevelmetermustnotbeinstalledinplaceswithdirectso-lar radiationandmustbeprotectedagainstweathereffects.
• If the installation in places with direct solar radiation is inevi-table,itisnecessarytomountashielding coverabovethelevelmeter(Fig.13).
• Itissuitabletorunthecableunderacablebushing(obliquelydown inslack)according toFig.14 topreventpenetration of humidity.Thentherainandcondensingwatercanflowofffreely.
• The cable bushing and connector have to be sufficiently tightenedtopreventpenetrationofhumidity.
• To lower theminimum distance to themeasuredmedium,a reflection boardmadefromsolid,evenandsmoothmate-rialcanbeinstalledtothelevelmeter.Thenthetankcanbefillednearlyuptothemaximumheight.Thesolutionissuit-ableforopentanksandreservoirs(Fig.15).
H + L = mrtvá zóna
Fig. 13: Solar radiation shielding cover
Fig. 14: Prevention to avoid intrusion of humidity
Fig. 15: Reflection board
ULM–70 © Dinel, s.r.o.9
7 . elecTrical connecTionThelevelmeter isconnectedtoconsequential (evaluating)devicewithasuitablecablewith theouterdiameterof6to8mmusingscrewterminalslocatedunderthedisplaymodule.Therecom-mendedcrosssectionofcoresforthecurrentversion2x0,5÷0,75mm2 and for the version with Modbuscommunication2x2x0,25mm2 (twistedpair,shielded).Pluspole(+U)isconnectedtotheterminal(+),minuspole(0V)totheterminal(-)andtheshielding(onlyforshieldedcables)totheterminal( ).Commu-nicationwiresAandBofthelineRS-485(forversion"M"-Modbus)areconnectedtotheterminalsAandB.Procedure to connect the cable to the level meter:
1. Unscrew the nut of the upper transparent lid.
2. Take the upper edge of the display module andtakeitoutcarefullybymildswingingup.
3. If you cannot grasp the module, you can usea small screwdriver. Insert it as far as the seam and usefromseveralsidestoslightlyliftthemodule.
4. Release the cableoutlet and thread the strippedsupplycablein.
5. Connectthecabletothescrewterminalsaccord-ingtothediagraminFig.17or18.Firmlytightentheterminalsandthecableoutlet.
6. If the level meter with Modbus is involved as aterminalforRS-485,werecommend(toavoidre-flectionsontheline)toconnect120Ωterminationresistor.This is donebymovinga small lever oftheswitchmarked120ΩtotheONposition.Onthelevelmetersconnected to the lineRS-485asanintermediatedevice, the termination resistorsarenotconnected(switchremainsoff).
7. Insert the displaymodule back into the head sothattheconnectorisproperlyconnected.
8. Slidesiliconesealonthethreadofthelevelmeterbody,thentightenthenutoftheupperlid.Connectthecabletoconsequentialdevice.
display unit connector
Terminal block
metal clip
Electrical connection must be done in de-energized state!
With regard to possible occurrence of electrostatic charge on non-conductive parts of the level meter, all level meters for explosive spaces (ULM–70Xi type) must be grounded. It will be done using a screw placed on the head of the level meter under the cable outlet.
Fig. 17: Inside view of screw terminals of the level meter with current output ULM-70_-_-_-I
Fig. 16: Wiring diagram of the level meter with current output ULM-70 _-_-_- I
+ A B ISGND
ON
120
display unit connector
Terminal block
metal clip
120Ω switch
Fig. 18: Wiring diagram of the level meter with Modbus ULM-70_-_-_-M
Fig. 19: Inside view of screw terminals of the level meter with Modbus ULM-70_-_-_-M
10© Dinel, s.r.o. ULM–70
The supply voltage source should be preferably realized as a stabilized power supply unit with safe voltage from 18 to 36 V DC (18 ÷ 30 V DC for Xi version), which can be a part of the evaluation or display device.
In case of strong electromagnetic interferences (EMI), parallel cable ducting with power lines, or when cable length exceeds 30 m we recommended to use shielded cable.
It is also necessary to design and take measures to reduce the effects of static electricity to a safe level in the wiring.
Installation in explosive atmospheres needs to be carried out in compliance with CSN EN 60079-14 (Electrical installations for explosive gaseous atmospheres – Part 14: Electrical installations in dangerous areas other than mining) and possibly also in compliance with other standards relating to the area concerned.
8 . exaMples of ulM-70 connecTion
PDU-420-W (input 4 ... 20 mA )
(0V) -
ULM-70N-_ _-_-I
User manual - METER PDU-4xx-P
Figure 4.7. Terminals description (relay outputs, device with current output)
Figure 4.8. Terminals description (OC-type outputs)
Figure 4.9. Connection of 2-wire current converters
11
1 2 5 6 7 8 9
GND DATA+DATA-
RS - 485
+ Uo -
+24V +5%, -10% Imax = 100mA
10 11 12 13 14 15
R1
3 4
optional ACTIVEcurrent output
- +
R2
n.c. n.c. n.c.n.c.
GND
0/1 - 5V, 0/2 - 10V
0/4-20mA
++
(optional) (optional)
Powersupply
(depending on version)
1631
1732
1833
1934
2035
1 2 5 6 7 8 9
GND DATA+DATA-
RS - 485
+ Uo -
+24V +5%, -10% Imax = 100mA
10 11 12 13 14 15
3 4
OC1 OC2 - + + -
OC1 ÷ OC4: Umax = 30V DC,
Imax = 30mA, Pmax = 100mW
+ -
- +
OC3
OC4(optional)
(optional)
n.c.n.c.
GND
0/1 - 5V, 0/2 - 10V
0/4-20mA
++(optional)(optional)
Powersupply
(depending on version)
1631
1732
1833
1934
2035
5 6 24V DC
- +
+ -
+-
+ -
internally connected
1631
1732
1833
1934
2035
1631
1732
1833
1934
2035
ULM-70N-_ _-_-I
connectedinternally
PDU-4xx-P (input 4 ... 20 mA )
8 .1 . Wiring diagram of the level meter With current output and pdu unit
(+U) + (+U) + (0V) -
connectedinternally
User manual - METER SRP-N118
a) b)
Figure 4.2. Recommended dimensions of cable strippinga) for big connectors (1 to 6), b) for small connectors (7 to 13, 15 to 17)
Figure 4.3. Method of connecting cables to the clamping connectors
All connections must be made while power supply is disconnected !
Figure 4.4. Terminals description (relay outputs)
9
!
+ Uo -n.c
.
R1 R2
GND
0/1 - 5V0/2 - 10V
0/4 - 20mA
+
+
GND
GN
D
DA
TA
+
151617 12 11 10 9 8 71314 123456
DA
TA
-
RS - 485
(option) (option)
Po
wer
su
pp
ly(d
ep
end
ing
on
ve
rsio
n)
24V (+5%, -10%)
Imax = 100 mA
PE
n.c
.
5 mm
ma
x.
1.6
mm6 mm
ma
x.
2 m
m
Connection of PDU-420-W is valid for firmware version 6.00 or higher. The older versions (up to version 5.99), the level meter output +U is connected to the terminal 7 and the output 0V to the terminal 10.
ULM–70 © Dinel, s.r.o.11
8 .2 . Wiring diagram of the level meter With current output and pdu unit in versions for explosive areas
(+U) +
(0 V) –
ULM-70Xi-_ _-_-I
IRU-420-I
explosive area
User manual - METER SRP-N118
a) b)
Figure 4.2. Recommended dimensions of cable strippinga) for big connectors (1 to 6), b) for small connectors (7 to 13, 15 to 17)
Figure 4.3. Method of connecting cables to the clamping connectors
All connections must be made while power supply is disconnected !
Figure 4.4. Terminals description (relay outputs)
9
!
+ Uo -n.c
.
R1 R2
GND
0/1 - 5V0/2 - 10V
0/4 - 20mA
+
+
GND
GN
D
DA
TA
+
151617 12 11 10 9 8 71314 123456
DA
TA
-
RS - 485
(option) (option)
Po
wer
su
pp
ly(d
ep
end
ing
on
ve
rsio
n)
24V (+5%, -10%)
Imax = 100 mA
PE
n.c
.
5 mm
max.
1.6
mm6 mm
ma
x. 2
mm
PDU-420-W-230V (input 4 ... 20 mA )
Connection of PDU-420-W is valid for firmware version 6.00 or higher. The older versions (up to version 5.99), the terminal 9 of the IRU unit is connected to the terminal 10 of the PDU unit and the terminal 10 of the IRU unit is connected to the terminal 11 of the PDU unit.
non-explosive area
User manual - METER SRP-N118
a) b)
Figure 4.2. Recommended dimensions of cable strippinga) for big connectors (1 to 6), b) for small connectors (7 to 13, 15 to 17)
Figure 4.3. Method of connecting cables to the clamping connectors
All connections must be made while power supply is disconnected !
Figure 4.4. Terminals description (relay outputs)
9
!
+ Uo -n.c
.
R1 R2
GND
0/1 - 5V0/2 - 10V
0/4 - 20mA
+
+
GNDG
ND
DA
TA
+151617 12 11 10 9 8 71314 123456
DA
TA
-
RS - 485
(option) (option)
Po
wer
su
pp
ly(d
ep
end
ing
on
ve
rsio
n)
24V (+5%, -10%)
Imax = 100 mA
PE
n.c
.
5 mm
ma
x.
1.6
mm6 mm
ma
x.
2 m
m
LN
230 V AC
12© Dinel, s.r.o. ULM–70
User manual - METER SRP-94
Double numeration means, that depending on device version, particular terminal can be marked according to the top or bottom number.
Figure 4.6. Terminals description (relay outputs)
Figure 4.7. Terminals description (OC-type outputs)
11
GND DATA+
DATA-
RS - 485
+ Uo -
+24V +5%, -10% Imax = 100mA
1 2 5 6 7 8 9
10 11 12 13 14 15
3 4
31 32 33 34 35
OC1 OC2 - +
OC1 ÷ OC4: Umax = 30V DC,
Imax = 30mA, Pmax = 100mW
OC3
OC4
+ - + -
+ -
(optional) (optional) (optional)
(optional)
Power
supply
(depending on version)
n.c. n.c.
GND
0/1 - 5V, 0/2 - 10V
0/4-20mA
++
Power
supply
(depending on version)
GND DATA+
DATA-
RS - 485
+ Uo -
+24V +5%, -10% Imax = 100mA
1 2 5 6 7 8 9
10 11 12 13 14 15
3 4
31 32 33 34 35
R1 R2 R3
R4
(optional) (optional) (optional)
(optional)
n.c. n.c.
GND
0/1 - 5V, 0/2 - 10V
0/4-20mA
++
1631
(+U) +
(0 V) –
ULM-70Xi-_ _-_-I
IRU-420-I
PDU-4xx-P-230V (input 4 ... 20 mA )
User manual - METER SRP-94
Double numeration means, that depending on device version, particular terminal can be marked according to the top or bottom number.
Figure 4.6. Terminals description (relay outputs)
Figure 4.7. Terminals description (OC-type outputs)
11
GND DATA+
DATA-
RS - 485
+ Uo -
+24V +5%, -10% Imax = 100mA
1 2 5 6 7 8 9
10 11 12 13 14 15
3 4
31 32 33 34 35
OC1 OC2 - +
OC1 ÷ OC4: Umax = 30V DC,
Imax = 30mA, Pmax = 100mW
OC3
OC4
+ - + -
+ -
(optional) (optional) (optional)
(optional)
Power
supply
(depending on version)
n.c. n.c.
GND
0/1 - 5V, 0/2 - 10V
0/4-20mA
++
Power
supply
(depending on version)
GND DATA+
DATA-
RS - 485
+ Uo -
+24V +5%, -10% Imax = 100mA
1 2 5 6 7 8 9
10 11 12 13 14 15
3 4
31 32 33 34 35
R1 R2 R3
R4
(optional) (optional) (optional)
(optional)
n.c. n.c.
GND
0/1 - 5V, 0/2 - 10V
0/4-20mA
++
1631
explosive area
non-explosive area
User manual - METER PDU-4xx-P
Figure 4.7. Terminals description (relay outputs, device with current output)
Figure 4.8. Terminals description (OC-type outputs)
Figure 4.9. Connection of 2-wire current converters
11
1 2 5 6 7 8 9
GND DATA+DATA-
RS - 485
+ Uo -
+24V +5%, -10% Imax = 100mA
10 11 12 13 14 15
R1
3 4
optional ACTIVEcurrent output
- +
R2
n.c. n.c. n.c.n.c.
GND
0/1 - 5V, 0/2 - 10V
0/4-20mA
++
(optional) (optional)
Powersupply
(depending on version)
1631
1732
1833
1934
2035
1 2 5 6 7 8 9
GND DATA+DATA-
RS - 485
+ Uo -
+24V +5%, -10% Imax = 100mA
10 11 12 13 14 15
3 4
OC1 OC2 - + + -
OC1 ÷ OC4: Umax = 30V DC,
Imax = 30mA, Pmax = 100mW
+ -
- +
OC3
OC4(optional)
(optional)
n.c.n.c.
GND
0/1 - 5V, 0/2 - 10V
0/4-20mA
++(optional)(optional)
Powersupply
(depending on version)
1631
1732
1833
1934
2035
5 6 24V DC
- +
+ -
+-
+ -
internally connected
1631
1732
1833
1934
2035
1631
1732
1833
1934
2035
User manual - METER SRP-N118
a) b)
Figure 4.2. Recommended dimensions of cable strippinga) for big connectors (1 to 6), b) for small connectors (7 to 13, 15 to 17)
Figure 4.3. Method of connecting cables to the clamping connectors
All connections must be made while power supply is disconnected !
Figure 4.4. Terminals description (relay outputs)
9
!
+ Uo -n.c
.
R1 R2
GND
0/1 - 5V0/2 - 10V
0/4 - 20mA
+
+
GND
GN
D
DA
TA
+
151617 12 11 10 9 8 71314 123456
DA
TA
-
RS - 485
(option) (option)
Po
wer
su
pp
ly(d
ep
end
ing
on
ve
rsio
n)
24V (+5%, -10%)
Imax = 100 mA
PE
n.c
.
5 mm
max.
1.6
mm6 mm
ma
x. 2
mm
LN
230 V AC
ULM–70 © Dinel, s.r.o.13
(+U) +
(0 V) -
ULM-70N-_ _-_-I
MGU-800 (with module IUI4)
8 .3 . Wiring diagram of the level meter With current output and mgu unit
12 13 14 18 19
A B +_
S
RS 485 - Master+24V DC
Imax=120 mA
(A) Y
E
(B) G
N
(0V)
WH
(+U)
BN
Cab.
Sh.
ULM-70N-_ _-_-M
(A) YE(B) GN
(0V) WH(+U) BN
Cab.Sh.
MGU-800 (without additional modules)
8 .4 . Wiring diagram of the level meter With modbus / rs485 and mgu unit
8 .5 . Wiring diagram of the level meter With modbus / rs485 and pdu unit
ULM-70N-_ _-_-M
14© Dinel, s.r.o. ULM–70
ESC OK
www.dinel.cz
Ultrasonic Level Meter ULM-70
ESC OK
1745T E mm
10 . sTaTus signalizaTion
1) symbol appears in the lower left corner of the display
display function
"NO ECHO" Lighting intermittently – the level meter is not able to receive echo for a long time. Incorrect installation of the level meter
"DEAD ZONE" Lighting intermittently – the measured level is in the "dead zone" of the level meter or the ultrasonic converter is dirty.
"NO PASSWORD" It will appear in the item "MENU" – the level meter is protected using a pass-word against unauthorised setting. Enter the correct password (see p. 19).
Symbol "T" 1) Lighting permanently – "TEACHING" mode activation.
Symbol "E" 1) Lighting intermittently – correct echo receiving (of the reflected signal) from the measured surface level.
Symbol
ESC OK
www.dinel.cz
Ultrasonic Level Meter ULM-70
ESC OK
1745T E mm 1)
Lighting permanently – level meter is locked against unauthorized settings by a password. You must enter the correct password to unlock it (see page 19).
Level meter locked
Ultrasonic transmitter
Upper lid nutCable gland
Control buttons
OLED display Label
9 . seT-up eleMenTs
Button
• Set-upmodeaccess• Confirmationofselectediteminthemenu• Movethecursorintheline• Savingofset-updata
Button
• Moveinthemenu• Change of values
Button
• Cancelling of carried out changes• Shiftonelevelup
Echo receiving
UnitsTeachingmode activation
Displayof measured values
Set-up elements
Fig. 20: Full view of the level meter
ESC OK
www.dinel.cz
Ultrasonic Level Meter ULM-70
ESC OK
1745T E mm
ULM–70 © Dinel, s.r.o.15
START ING
Dinel®
11 . operaTion anD seTTing
Setthelevelmeterusing3buttonsplacedonthedisplaymodule(seeChapterSet-upelements).After5min.ofinactivity,thelevelmeterautomaticallyreturnsbacktothemeasurementmode.Ifthepassword is active, the levelmeterwill be also locked.The val-uesthathavenotbeenconfirmedusingthebutton willnotbesaved!After themeter is locked, youcannot change the setting!Whenyouattempttoedit,thewords"NOPASSWORD"willappearonthedisplay.Howtounlockthelevelmeterisgivenonpage16.Afterconnectionofthesupplyvoltagetothelevelmeterthedisplayshowsthelogo"Dinel"andthetext"Starting"(approx.15s).Then,thelevelmetergoestothemeasuringmodeandthedisplayshowsthe current measured value.
MIN LEVEL and MAX LEVEL
Youcan freelydefine theminimum / maximum distance from the front surface of the level meter(item"LEVEL"forcurrents4/20mA).The"DISPLAY"isintendedtosetthevaluedisplayedonthedisplay.Settingtheunitsisdoneinthe"UNITS".
Ifinthebottomofthedisplayappears(whenenteringthevalues)theinscription"OUTOFLIMITS",thevaluespecifiedfortheitem"LEVEL"isoutsidethemeasuringrangeofthelevelmeter.Iftheinscription"SPANTOOSMALL"isshown,itmustbespecifiedalargerspanbetweenMinandMaxvalues.Formoreinformation,seechapter"Specifications".Thedecimalpointpositionoftheitem'LEVEL'isfirmlyset(accordingtotheselectedunits),intheitem"DISPLAY"itisfreelyadjustable
11 .1 . basic configurationAfterthefirststartofthelevelmeteritisnecessarytoperformthebasicconfiguration(settingofthemeasuringrange,choiceofunitsandpossibly damping).The settingsareaccessible in thebasicmenubypressing the"BASICSETTINGS".
UN ITS
LEVEL : mm
DISPLAY: %
TEMPERATURE °C
MIN LEVELMAX LEVELUN ITSDAMP INGSENS IT IV ITYTEACH ING
MIN ACTUAL LEVEL0 2 8 7 0 mm
OUTPUT : 0 4 . 0 0 mA
LEVEL : 0 6 0 0 0 mm
DISPLAY: 0 0 0 0 0 %
MAX ACTUAL LEVEL0 2 8 7 0 mm
OUTPUT : 2 0 . 0 0 mA
LEVEL : 0 0 2 5 0 mm
DISPLAY: 0 0 1 0 0 %
ACTUAL LEVEL: Actual distance to levelOUTPUT: current 4 mA / 20 mALEVEL: Definition of the min / max levelDISPLAY: The value showed on the displayMIN LEVEL
MAX LEVELUN ITSDAMP INGSENS IT IV ITYTEACH ING
BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDLANGUAGEINFO
16© Dinel, s.r.o. ULM–70
UNITS
LEVEL : mm
DISPLAY: %
TEMPERATURE °C
DAMP ING
05
UNITS
Levelmetercanprocessandconvertalargenumberofdifferentphysical values. The setting is doneintheitem"UNITS".
LEVEL: Unit selection (mm, cm, m, in, ft)DISPLAY: The unit showed on the display (%, mm, cm, m, in, ft, l, hl, m3, gal, bbl, mA)TEMPERATURE: Temperature unit (°C, °F)
DAMPING
Settingthe response time of the measurements. Thefunctionisusefulforsuppressinglevelfluc-tuations,wavesandrapidchangesofthelevel.The reaction time will depend on the exponential function.Dampingwithadefineddelayinsecondsrepresentsthetimewhenexponentialreaches2/3ofitsmaximumvalue.
The damping time can be set in the interval from 0 to 99 s.
1. To enter to the menu press thesamebuttontoselect"BASICSETTINGS".Then,usingand select"UNITS".
2. Nowthemenuitem"UNITS"isshown. Bypressingthe and buttonmakethesettingsofindividual items.
3. Bypressing buttonsave thedata.Bynextpressesof thebutton leave the menu. The level meter returns to measurement mode.
1. To enter to the menu press thesamebuttontoselect"BASICSETTINGS".Then,usingand select"DAMPING".
2. Nowthemenuitem"DAMPING"isshown. Bypressingthe and buttonmakethesettingsof individual items.
3. Bypressing buttonsave thedata.Bynextpressesof thebutton leave the menu. The level meter returns to measurement mode.
1. To enter to the menu press thesamebuttontoselect"BASICSETTINGS". Then,usingand select "MINLEVEL"or"MAXLEVEL".
2. Nowitisshowntheitem"MINLEVEL"("MAXLEVEL").Bypressing and set the output cur-rent"OUTPUT",thedistanceforthedefinedcurrent"LEVEL"thevalueonthedisplay "DISPLAY".
3. Bypressing buttonsave thedata.Bynextpressesof thebutton leave the menu. The level meter returns to measurement mode.
MIN LEVELMAX LEVELUN ITSDAMP INGSENS IT IV ITYTEACH ING
MIN LEVELMAX LEVELUN ITSDAMP INGSENS IT IV ITYTEACH ING
ULM–70 © Dinel, s.r.o.17
You can set the sensitivity in three degrees: LOW – MEDIUM – HIGH.
SENSITIVITY
Thesettingisdefinedinthreestepsofthelevelmetersensitivity.
"LOW" – Low sensitivity in case of surrounding interferences affecting the measurement. "MEDIUM" – Medium sensitivity (suitable for most applications)."HIGH" – Enhanced sensitivity for measured mediums partly absorbing the ultrasonic signal (bulk solids, foams)
If there are no above obstacles in the tank, it is not necessary to start this mode.
TEACHING
The mode serves for suppressing false reflections resulting fromreflectionof theultrasonicsignalfromroughnessesonwallsofthetank,variouspartitions,mixersorotherobstacles.Thesensorstartingthismodedetectsfalsereflectionsandsavetheminthememory.Thenthesefalsereflectionswillnotaffectthesubsequentmeasurement(theyaremasked).
Before starting the mode it is necessary to empty the tank as much as possible (prefer-ably completely).
1. To enter to the menu press thesamebuttontoselect"SERVICE".Then,using andselect"TEACHING".
2. Nowitisshowntheitem"TEACHING".Bypressing setthevalue"LEVELDISTANCE"(dis-tanceto the level)–supposeddistancefromthefaceof thesensor to themediumlevel. If thedistancetothelevelisnotpreciselyknown,enteravalueratherlower(inthetolerancefieldasshowninFig.19).
3. Afterenteringthe"SETLEVELDISTANCE"bypressing buttonthesystemstarts"teaching"(falsereflectionmapping).Duringthemapping,thedisplayshowsflashingsign"RUNNING".
4. Themappingoffalseechoescanbecompletedwhenyouseetheinscription"PressOKtostop"andyoupress .
5. Theprocedureiscompletelyfinishedwhenyoucanseetheinscription"DONE".Itisthenpos-sibletoexitthemenubyrepeatedpressingthebutton .
SENS IT IV ITY
MED IUM
TEACH ING
SET LEVEL D ISTANCE :
0 6 0 0 0 m m
START
MIN LEVELMAX LEVELUN ITSDAMP INGSENS IT IV ITYTEACH ING
MIN LEVELMAX LEVELUN ITSDAMP INGSENS IT IV ITYTEACH ING
TEACH ING
SET LEVEL D ISTANCE :
0 6 0 0 0 m m
RUNN ING
TEACH ING
SET LEVEL D ISTANCE :
0 6 0 0 0 m m
DONEPRESS ESC TO EX IT
The mode "TEACHING" will stop automati-cally after ca. 1000 measurements.
TEACH ING
SET LEVEL D ISTANCE :
0 6 0 0 0 m m
RUNN INGPRESS OK TO STOP
If during the scanning of the tank in the bottom of the display appears the dialog "press OK to stop" (see figure) the level meter already found no further obstacles and "TEACHING" mode may be terminated. If it is not terminated, the level meter is still ready for the possible presence of obstacles (e.g. paddles of the agitator). Once it registers a further obstacle, the dialogue disappears and the obstacle is erased. This process may be repeated up to 1000 cycles. After this the "TEACHING" mode is automatically stopped.
18© Dinel, s.r.o. ULM–70
Incaseofinstalledmixers,itisnecessary to posi-tion the mixers under the level meter (direct the mixerbladetotheultrasonicsignalbeam).
Note:Iftherearesignificantobstaclesintheupperhalf of the tank,multiple false reflections can occur especially in closed tanks. In such casesit isnecessary to reduce the level in the tankasmuchsepossibletocorrectlymaskthesepossiblemultiplefalsereflections.
1/2 of Tank
Obstacles
Tolerance zone for "LEVEL DISTANCE" value
MEDIUM TEMPERATURE
The level meter is equipped with automatic temperature compensation. If for instance in the tank there is a difference of 10°Cbetween the temperature of themeasuredmaterial (me-dium) and the temperature at themounting site of the levelmeter (see themode "DIAGNOS-TICS, page 20), the measuring accuracy will be reduced by around 1% of the set range. Ifthis function is activated, this temperature difference can be compensated. If in the tank (openchannel) isabigdifferencebetweenthetemperatureof themeasuredmedium(liquid)andtem-perature in theplaceof installationof theULM(seemode, "DIAGNOSTICS"page20), it isad-vised to improve the precision of the measurement by the zone temperature compensation. Otherwise,thismodeisnot necessary to run.
MEDIUM TEMPERATUREFA ILURE MODEHARTFACTORY DEFAULTRESET
MED IUM TEMPERATURE
YES
+20°C
Inactive compensation (initial state), the word "NO" appears on the display.See the "UNITS" menu for temperature unit selection (°C or °F).
Afterstartofthezone temperature compensationmodeit isnecessarytosetthetemperatureof the surface of the medium. The level meter then calculates the average value from the medium temperature and the temperature at the installation place of the level meter. With such an average temperatureitcountsincalculatingofthevelocityofacousticwavespropagationandfordetermina-tion of the level position.
Leve
l Dist
ance
NO ECHO: Current in case of echo lossDEAD ZONE: Dead zone currentThe values can be set in three steps: 3.75 mA, 22 mA and LAST (last measured data).
FAILURE MODE
It defines the output current of the level meter when the measured medium level is in the dead zone ("DEADZONE")oroutsidethemeasurementrangeincaseofecholoss("NOECHO").
FA ILURE MODE
NO ECHO: 3 .75 mA
DEAD ZONE : 22 mA
11 .2 . services settingsIn the supplemented configuration, you can set parameters ofsensitivity, mapping of false reflections, temperature differencecompensation, behaviour in case of fault conditions or HART® communication.Here,youcansetthesensorintotheinitialstateorresetitaswell.Thesettingsareaccessibleinthebasicmenuundertheitem"SERVICE".
BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDINFO
MED IUM TEMPERATUREFA ILURE MODEHARTFACTORY DEFAULTRESET
Fig. 21: Level distance zone
ULM–70 © Dinel, s.r.o.19
HART
HART® mode(pointtopoint,multidrop)andmultidropmodeaddress setting.Upto15unitscanbeconnectedtoonetwo-wiredcableinthemultidropmode.
In case of the address "00", the point to point mode is enabled. The range from "01" to "15" is reserved for addresses in the multidrop mode.
After you press the button "RUNNING" will be displayed for about 3 sec. After the initial values are set, "DONE" will be appear on the display.
FACTORY DEFAULT
To reset the initial values ofthelevelmetersetbythemanufacturer,pressthebutton (see the Factorydefaulttable,p.28).
RESET
Complete restartofthelevelmeter.Thesameeffecthasalsoashort-timeinterruptionofthesupplyvoltage.Toenabletheresetting,pressthebutton .
FACTORY DEFAULT
ARE YOU SURE?
FACTORY DEFAULT
RUNN ING
FACTORY DEFAULT
DONEPRESS ESC TO EX IT
RESET
ARE YOU SURE?During the restart process, "RUNNING" will be displayed. Then the level meter will be automatically turned off and on.
MODBUS
This itemispartofamenuwithModbusoutput levelmeterULM-70_-_-_-M.Modbusmode isintendedforthesettingsofthelevelModbusaddresses,baudrateandparitysettings.
MODBUS
ADDRESS : 001
BAUD RATE : 9600
FORMAT: 8N1
ADDRESS: 1 to 247 (default 1) BAUD RATE: 4800, 9600, 19200 (default 9600) FORMAT : 8N1, 8O1, 8E1, 8N2 (default 8N1)
data:
parita: N – non parityO – odd parityE – even parity
8 – number of bits
FORMAT: number of stop bits: 12
MEDIUM TEMPERATUREFA ILURE MODEHARTFACTORY DEFAULTRESET
HART
00
POLL ING ADDRESS
MED IUM TEMPERATUREFA ILURE MODEMODBUSFACTORY DEFAULTRESET
MED IUM TEMPERATUREFA ILURE MODEMODBUSFACTORY DEFAULTRESET
MED IUM TEMPERATUREFA ILURE MODEMODBUSFACTORY DEFAULTRESET
20© Dinel, s.r.o. ULM–70
DIAGNOSTICS
Itcontainsinformationabouttheactualtemperatureinsidethetank(or about the compensated temperature) "TEMPERATURE" andcurrent flowing through the loop "CURRENT". If the temperaturecompensation("MEDIUMTEMPERATURE")isactivated,thecor-rectedtemperatureisdisplayed.
DISTANCE TO LEVELTEMPERATURECURRENT
TEMPERATURE
+23°C
CURRENT
17 .30 mA
The temperature is measured inside the tank where the level meter is installed.If the temperature of the measured medium is different, we recommend you to carry out the temperature compensation "MEDIUM TEM-PERATURE" because of accuracy (see p. 15). Then the displayed temperature is an average value from the temperature set in the "MEDIUM TEMPERATURE" and the actual temperature measured by the sensor.
DISTANCE TO LEVELTEMPERATURECURRENT
CLONE SETTINGS
This mode is intended for copyingofthelevelmeter(ULM–70body) configuration into the display module (DM–70) andback.Thedisplaymodulecanthenberemovedfromthelevelmeterbodyandputintoanotherlevelmeterandmaketherethesettingstransfer(cloning).The"CLONESETTINGS"modetransfersalldata,excludingsettingofthe"Teaching"andHART®.
BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDINFO
CLONE SETT INGS
NOW CLON ING
CLONE SETT INGS
SENSOR D ISPLAYMODULE
D ISPLAYMODULE
SENSOR
→→
1. Press toenterthemenuandselecttheitem"CLONESETTINGS".Copyingofthesettingsfromthebodyofthelevelmetertodisplaymoduleisdonebyselecting"SENSOR→ DISPLAYMODULE".TotransferthesettingsfromthedisplaymoduletoanotherlevelmeterselecttheitemDISPLAYMODULE→ SENSOR.
2. Theselectedmodestartsbypressingbutton Duringtransmissionthedisplayshows"NOWCLONING".
3. Aftercompletingtheprocessinthemiddleofthescreendisplays"DONE".Itisthenpossibletoleavethemenuandthemodebypressingthebutton .
CLONE SETT INGS
SENSOR D ISPLAYMODULE
D ISPLAYMODULE
SENSOR
→→
ARE YOU SURE?
11 .3 . additional functionsAdditionalfunctionsincludemodestodisplaytemperatureinthetankortofindouttheactualflowingcurrentintheloop.Besides,tolockmodificationsusingapasswordandinformationaboutthelevelmeterversion.Allofthefunctionsareaccessiblefromthemainmenu.
BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDLANGUAGEINFO
D ISTANCE TO LEVEL
01220 mm
DISTANCE TO LEVELTEMPERATURECURRENT
ULM–70 © Dinel, s.r.o.21
CLONE SETT INGS
WARN ING–CLON ING IS NOT POSS IBLE
WRONG SENSOR TYPE
PRESS ESC TO EX IT
CLONE SETT INGS
WARN ING–CLON ING IS NOT POSS IBLE
NO SETT INGS SAVED
PRESS ESC TO EX IT
Incompatible type of level meter. Transfer of the settings can be realized only with the same type of level meter (e.g. ULM–70–02 → ULM–70-02, ULM–70–10 → ULM–70–10) and with the firmware version 2.0 and later.
PASSWORD
You can lock the level meter data against unauthorized editing. Afteractivatingthepasswordthedatamayberead,butcannotbeedited.Ifyoutrytoeditthesettings(withouttruepassword)thedisplayshows"NOPASSWORD".Thepasswordcanbeany5-digitnumericcombination.Thecom-bination of numbers 00000 is reserved for disabling the pass-word.
1. Usethebuttons and inthemenu"PASSWORD"toselectthemode"ENTER"forenteringthepasswordorthemode"CHANGE"forchangingthepassword(whenactivated,thewordsaredisplayedinversely).Pressthebutton onceagaintoconfirmtheselection.Youcanchangethepasswordonlywhenthelevelmeterisunlocked.Otherwise,thewords"NOPASSWORD"willbedisplayed.
2. Nowyoucaneditthepassword.Theactualediteditemisdisplayedinversely.Pressthebuttontomovetothenextposition(clockwisedirection),button servestochangethevalues(0...9).
3. Aftertheoperationiscompleted,confirmtheediteddatabypressingthebutton .
Display of status information to confirm data:"YES" – correctly edited password"NO" – incorrectly edited password"OK" – the password saved (only in case of "CHANGE")The password is automatically hidden after it is edited or changed ("00000" will appear). To deactivate the password, edit the nu-merical combination "00000" in the mode "CHANGE".
TEACH ING
SET LEVEL D ISTANCE :
0 6 0 0 0 m m
START
NO PASSWORD
PASSWORD
ENTER
0 0 0 0 0
PASSWORD
CHANGE
0 0 0 0 0
The data set is not stored into the display module (DM–70). The transfer can not be done. It is necessary to repeat the procedure of the copying the settings in the mode "CLONE SETTINGS".
The level meter with activated password will be automatically locked after 5 minutes of inactivity or after 5 min. from switching to measuring mode. Locking of level meter is indicated in the lower left corner of the screen by the letter "L".
If the password is lost, contact the manufacturer.
BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDLANGUAGEINFO
22© Dinel, s.r.o. ULM–70
INFO
Informationaboutthetype,serialnumberandproductiondateofthelevelmeter(type,serialnumber–SNandfirmwareversion–SW).
SENSOR
ULM-70N -06 - I HART
SN: 140025
SW: 2 .0
D ISPLAY MODULE
DM-70
SN: 140041
SW: 4 .1
SENSORD ISPLAY MODULE
SENSORD ISPLAY MODULE
BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDLANGUAGEINFO
SENSOR
ULM-70N -06 -M MODBUS
SN: 140025
SW: 2 .0
LANGUAGE
Settingthelanguageofdisplaymenu.
BAS IC SETT INGSSERV ICED IAGNOST ICCLONE SETT INGSPASSWORDLANGUAGEINFO
LANGUAGE
ENGL ISH
You can set three kinds of language:ČESKY – ENGLISH – по русски
12 . proTokol HarT®
Universal communication interface for data communication of peripheral devices with the level meter.Datatransmissionrunsthroughthesamelineasthe4÷20mAcurrentloopwithoutimpact on analog communication. Forsettingthelevelmeterandcollectionofmeasureddata,itisnecessarytohaveavailableaHARTcommunicator,bywhichitispossibletocommunicatedirectlywiththelevelmeter,orusingit,tomediatecommunicationwithaperipheraldevice,seeimage 22.
≈ 250 R
Obr. 22: Typical hardware configuration with HART
ULM–70_-_ _-_-I
RS232 / USB
HART®
communicator
Power supply unit
ULM–70 © Dinel, s.r.o.23
13 . proTokol MoDbus®
14 . orDer coDe
Standard – incl. in the price of the level sensor Optional – for extra charge
• 1pcofSeal(forULM–70_–02–I,06–I)• free-to-download programme Basic Scada
Level(fortheModbusversion)
• FixingnutsG1"andG1½"andG2 ¼• HornadapterST–G1,
STG1,5andST–G2,25• for versionModbusconvertorURC-485
02 – 0,15 ... 2 m06 – 0,25 ... 6 m10 – 0,4 ... 10 m20 – 0,5 ... 20 m
ULM – 70 –
Output type: I – current outputM – RS-485 (Modbus RTU)
– – –
Process connection: G – threadF – flange
Electrical connection: B – short cable outletH – outlet for protective conductor
– Set-up elements: D – basic version with displayL – without display, metal lid
Performance: N – Normal – usable in non-explosive areas onlyXi – Ex. proof – suitable for explosive areas
Maximum reach:
15 . accessories
Obr. 23: Typical hardware configuration with Modbus®
DatacommunicationtakesplacealongaserieslineofastandardRS-485withprotocolModbusRTU.Alistofrelevantvariablesisprovidedinaseparateannex.
Tosetupthelevelmeterandcollectmeasureddata,youcanusethesoftwareapplication"BasicSCADAlevel",whichisfreelyavailableattheWebsitewww.dinel.cz.ConnectingthelevelmetertoaperipheraldevicecanbeperformedusingaconverterURC-485,seeimage23.
RS485 / USB
convertor URC-485
ULM–70_-_ _-_-M
Power supply unit
24© Dinel, s.r.o. ULM–70
It is forbidden to perform any modifications or interventions into the ULM–70 level meter without manufacturer's approval. Potential repairs must be carried out by the manufacturer or by a manufacturer authorized service organization only.
Installation, commissioning, operation and maintenance of the ULM–70 level meter has to be carried out in accordance with this instruction manual; the provisions of regulations in force regarding the installation of electrical equipment have to be adhered to.
Installation in areas with potentially explosive atmospheres must be carried out in accordance with standard EN 60079-14 (Electrical apparatus for explosive gas atmospheres - Part 14: Electrical installations in hazardous areas other than mines) and respectively in accordance with other standards that apply to a given area.
16 . safeTy, proTecTion, coMpaTibiliTy anD explosion proof
The levelmeterULM–70 isequippedwithprotectionagainst reversepolarityandoutputcurrentoverload.ProtectionagainstdangerouscontactissecuredbylowsafetyvoltagethatcomplieswithEN332000-4-41.ElectromagneticcompatibilityaccordingtoEN55022/B,EN61326/Z1andEN61000-4-2to6.ExplosionproofofULM–70Xitypecomplieswiththefollowingstandards:EN60079-0:2007;EN60079-11:2007;EN60079-26:2007andexaminedbyFTZÚ-AO210Ostrava-RadvanicecertificateNo.:FTZÚ09ATEX0277X.
Special conditions for safe use ULM–70Xi:
ThedeviceisdesignedforconnectiontotheisolatingrepeaterIRU-420.Whentheotherapprovedsupplyunitisused,whoseoutputparameterssatisfyabovementionedoutputparameters,itisnec-essary tohaveagalvanicseparationor, ifsupplyunitwithoutgalvanicseparation isused (Zenerbarriers),itisnecessaryprovidepotentialequalizationbetweensensorandpointofbarrierearthing.Forapplicationinzone0thepresentexplosiveatmospheres-mixtureofairwithflammablegases,vapourormistsmustcomply:0,8bar<p<1,1bar.Thedevicemustbeinstalledinsuchaway,topreventmechanicaldamageofsensorface.Itisnecessarycarriedoutearthingbyscrewwhichisplaced on head of level meter.
17 . use, ManipulaTion anD MainTenance
The levelmeterdoesnot requireanypersonnel for itsoperation.Follow-updisplayingdevice isusedtoinformthetechnologicalentityoperatingpersonnelonthemeasuredsubstancelevelheightduring the operation.
Maintenanceofthisequipmentconsistsinverificationofintegrityofthelevelmeterandofthesup-plycable.Dependingonthecharacterofthesubstancemeasured,werecommendtoverifyatleastonceperyeartheclarityoftheultrasoundtransduceremittingfieldandtocleanit,respectively.Incaseanyvisibledefectsarediscovered, themanufactureror resellerof thisequipmentmustbecontactedimmediately.
The device must be installed to prevent tensile overload rope electrode level meter.
The device must be installed in such a way, to prevent mechanical damage of sensor face.
ULM–70 © Dinel, s.r.o.25
18 . Marking of labels
Symbolofproducer:logoDinel®
Internetaddress:www.dinel.czLevelmetertype:ULM–70N–__–_–I–_–_Serialnumber:Ser.No.:xxxxx–(fromtheleft:productionyear,serialprodutionnumber)Supplyvoltage:Ui = 18...36V=Outputcurrentrange:I =4÷20mAAmbienttemperaturerange: ta=-30...+__°C(see.Temperature range according to type)Protectionclass:IP67Compliancemark: Electro-wastetake-backsystemmark:
ULM-70N-20-F-I-B-D
U = 18 36 V...
I = 4 20 mA...t = -30 ... +60 °C
a
ULM-70N-10-G-I-B-D
U = 18 36 V...
I = 4 20 mA...t = -30 ... +60 °C
a
ULM-70N-06-G-I-B-D
U = 18 36 V...
I = 4 20 mA...t = -30 ... +70 °Ca
ULM-70N-02-G-I-B-D
U = 18 36 V...
I = 4 20 mA...t = -30 ... +70 °Ca
IP67
IP67
IP67
IP67
www.dinel.cz
DinelMade in Czech Republic
www.dinel.cz
DinelMade in Czech Republic
www.dinel.cz
DinelMade in Czech Republic
www.dinel.cz
DinelMade in Czech Republic
Ser. No.: ______
Ser. No.: ______
Ser. No.: ______
Ser. No.: ______
Varianty značení:
ExampleoflabelfortypeofULM-70N-02-G-B-D
Temperature range according to type:
02 ta = -30 ... +70°C06 ta = -30 ... +70°C10 ta = -30 ... +60°C20 ta = -30 ... +60°C
Labels for type of ULM-70N-_ _-_-I-_-_:
Clasification non-explosive performance:
02 II1/2GExiaIIBT5Ga/Gb
06 II1/2GExiaIIBT5Ga/Gb
10 II1/2GExiaIIAT5Ga/Gb20 II2GExiaIIAT5Gb
Labels for type of ULM-70Xi-_ _-_-I-_-_:
ULM-70Xi-02-G-I-B-D
IP67
U = 30 Vi
t = -30 ... + 70 °Ca
II 1/2 G Ex ia IIB T5 Ga/Gb
P = 0,99 W, I = 132 mAi i
C = 370 nF, L = 0,9 mHi i
FTZÚ 09 ATEX 0277X
1026
I = 4 20 mA...
ULM-70Xi-06-G-I-B-D
IP67
1026
ULM-70Xi-10-G-I-B-D
IP67
1026
ULM-70Xi-20-F-I-B-D
IP67
1026
U = 30 Vi
t = -30 ... + 70 °Ca
P = 0,99 W, I = 132 mAi i
C = 370 nF, L = 0,9 mHi i
I = 4 20 mA...
U = 30 Vi
t = -30 ... + 60 °Ca
P = 0,99 W, I = 132 mAi i
C = 370 nF, L = 0,9 mHi i
I = 4 20 mA...
U = 30 Vi
t = -30 ... + 60 °Ca
P = 0,99 W, I = 132 mAi i
C = 370 nF, L = 0,9 mHi i
I = 4 20 mA...
www.dinel.cz
DinelMade in Czech Republic
www.dinel.cz
DinelMade in Czech Republic
www.dinel.cz
DinelMade in Czech Republic
www.dinel.cz
DinelMade in Czech Republic
II 1/2 G Ex ia IIB T5 Ga/Gb
FTZÚ 09 ATEX 0277X
II 1/2 G Ex ia II T5A Ga/Gb
FTZÚ 09 ATEX 0277X
II G Ex ia II T52 A Gb
FTZÚ 09 ATEX 0277X
Varianty značení:
Ser. No.: ______
Ser. No.: ______
Ser. No.: ______
Ser. No.: ______
ExampleoflabelfortypeofULM-70Xi-02-G-I-B-D
Symbolofproducer:logoDinel®
Internetaddress:www.dinel.czLevelmetertype:ULM–70Xi–__–_–I–_–_Serialnumber:Ser.No.:xxxxx– (fromtheleft:productionyear,serialprodutionnumber)Outputcurrentrange: I =4...20mAMax.internalvalues:Ui = 30V=,Ii=132mA;Pi =0,99W;Ci=370nF;Li=0,9mHAmbienttemperaturerange: ta=-30...+__°C(viz.Teplotnírozsahdletypu)Labelofnon-explosivedevice: ,Performance:II_GExiaII_T5_/_Numberofcertificateofintrinsicallysafety:FTZÚ09ATEX0277XProtectionclass:IP67Compliancemark: ,No.ofauthorizedpersonexaminingcontrolofsystemquality:1026Electro-wastetake-backsystemmark:
Temperature range according to type:
02 ta = -30 ... +70°C06 ta = -30 ... +70°C10 ta = -30 ... +60°C20 ta = -30 ... +60°C
26© Dinel, s.r.o. ULM–70
ULM-70N-20-F-M-B-D
ULM-70N-10-G-M-B-D
ULM-70N-06-G-M-B-D
ULM-70N-02-G-M-B-D
IP67
IP67
IP67
IP67
U = 18 36 V...
RS-485 (Modbus RTU)
t = -30 ... +70 °Ca
U = 18 36 V...
RS-485 (Modbus RTU)
t = -30 ... +70 °Ca
U = 18 36 V...
RS-485 (Modbus RTU)
t = -30 ... +60 °Ca
U = 18 36 V...
RS-485 (Modbus RTU)
t = -30 ... +60 °Ca
www.dinel.cz
DinelMade in Czech Republic
www.dinel.cz
DinelMade in Czech Republic
www.dinel.cz
DinelMade in Czech Republic
www.dinel.cz
DinelMade in Czech Republic
Ser. No.: ______
Ser. No.: ______
Ser. No.: ______
Ser. No.: ______
Varianty značení:
Labels for type of ULM-70N-_ _-_-M-_-_:
ExampleoflabelfortypeofULM-70N-10-G-M-B-D
Symbolofproducer:logoDinel®
Internetaddress:www.dinel.czLevelmetertype:ULM–70N–__–_–I–_–_Serialnumber:Ser.No.:xxxxx–(fromtheleft:productionyear,serialprodutionnumber)Supplyvoltage: U= 18...36V=Dataoutput:RS-485(ModbusRTU)Ambienttemperaturerange: ta=-30...+__°C(see.Temperature range according to type)Protectionclass:IP67Compliancemark: Electro-wastetake-backsystemmark:
Real label size is 70x20mm.
Temperature range according to type:
02 ta = -30 ... +70°C06 ta = -30 ... +70°C10 ta = -30 ... +60°C20 ta = -30 ... +60°C
ULM–70 © Dinel, s.r.o.27
19 .
Men
u s
Tru
cTu
reM
INM
AX
05
00
SE
NS
OR
GR
LM
-70
N-1
0-I
HA
RT
EL
EC
TR
OD
E:
10
00
mm
SN
: 1
40
01
5
SW
: 2
.1
DIS
PL
AY
MO
DU
LE
DM
-70
SN
: 1
40
08
5
SW
: 4
.1
SE
NS
OR
DIS
PL
AY
MO
DU
LE
BA
SIC
SE
TT
ING
SS
ER
VIC
ED
IAG
NO
ST
ICC
LO
NE
SE
TT
ING
SP
AS
SW
OR
DL
AN
GU
AG
EIN
FO
BA
SIC
SE
TT
ING
SS
ER
VIC
ED
IAG
NO
ST
ICC
LO
NE
SE
TT
ING
SP
AS
SW
OR
DL
AN
GU
AG
EIN
FO
BA
SIC
SE
TT
ING
SS
ER
VIC
ED
IAG
NO
ST
ICC
LO
NE
SE
TT
ING
SP
AS
SW
OR
DL
AN
GU
AG
EIN
FO
BA
SIC
SE
TT
ING
SS
ER
VIC
ED
IAG
NO
ST
ICC
LO
NE
SE
TT
ING
SP
AS
SW
OR
DL
AN
GU
AG
EIN
FO
BA
SIC
SE
TT
ING
SS
ER
VIC
ED
IAG
NO
ST
ICC
LO
NE
SE
TT
ING
SP
AS
SW
OR
DL
AN
GU
AG
EIN
FO
BA
SIC
SE
TT
ING
SS
ER
VIC
ED
IAG
NO
ST
ICC
LO
NE
SE
TT
ING
SP
AS
SW
OR
DL
AN
GU
AG
EIN
FO
BA
SIC
SE
TT
ING
SS
ER
VIC
ED
IAG
NO
ST
ICC
LO
NE
SE
TT
ING
SP
AS
SW
OR
DL
AN
GU
AG
EIN
FO
MIN
LE
VE
LM
AX
LE
VE
LU
NIT
SD
AM
PIN
GS
EN
SIT
IVIT
YT
EA
CH
ING
AC
TU
AL
02
87
0m
m
OU
TP
UT
:0
4.0
0m
A
LE
VE
L:
06
00
0m
m
DIS
PL
AY
:0
00
00
%
AC
TU
AL
02
87
0m
m
OU
TP
UT
:2
0.0
0m
A
LE
VE
L:
00
25
0m
m
DIS
PL
AY
:0
01
00
%
UN
ITS
LE
VE
L:
mm
DIS
PL
AY
:%
DA
MP
ING
ME
DIU
M T
EM
PE
RA
TU
RE
FA
ILU
RE
MO
DE
HA
RT
FA
CT
OR
Y D
EF
AU
LT
RE
SE
T
TE
AC
HIN
G
TA
NK
MU
ST
BE
EM
PT
Y
ST
AR
T
FA
ILU
RE
MO
DE
NO
EC
HO
: 4
.00
mA
HA
RT
RE
SE
T AR
E Y
OU
SU
RE
?
FA
CT
OR
Y D
EF
AU
LT
AR
E Y
OU
SU
RE
?
DIS
TA
NC
E T
O L
EV
EL
01
22
0 m
m
DIS
TA
NC
E T
O L
EV
EL
CU
RR
EN
TT
EM
PE
RA
TU
RE
+2
3°
C
CL
ON
E S
ET
TIN
G
SE
NS
OR
DIS
PL
AY
MO
DU
LE
DIS
PL
AY
MO
DU
LE
SE
NS
OR
→ →
PA
SS
WO
RD
EN
TE
R
00
00
0
LA
NG
UA
GE
EN
GL
ISH
PO
LL
ING
AD
DR
ES
S
SE
NS
ITIV
ITY
SE
NS
ITIV
ITY
: M
ED
IUM
1 2
3
4 5
6!
7!
8!
MA
TE
RIA
L:
LIQ
UID
ME
DIU
M T
EM
PE
RA
TU
RE
YE
S
+2
0°
C
CU
RR
EN
T 17
.30
mA
28© Dinel, s.r.o. ULM–70
TecHnical specificaTions – leVel MeTer
Measuringrange1)ULM–70_–02ULM–70_–06ULM–70_–10ULM–70_–20
0.15 ... 2 m0.25...6m0.4 ... 10 m0.5 ... 20 m
Adjustablemeasuringrange(SPAN) Min.200mm
Supplyvoltage ULM–70N–__ULM–70Xi–__
18...36VDC18 ... 30 V DC
Output ULM–70_–__-_-IULM–70_–__-_-M
4...20mA(Limitvalues3.9...20.5mA),HART®
RS-485withprotocolModbusRTU
Current consumption ULM–70_–__-_-IULM–70_–__-_-M
4...20mA/Max.22mAMax.20mA
ResolutionULM–70_–02;10ULM–70_–06ULM–70_–20
<1mm<2mm<2.5mm
Accuracy(withinthetotalrange) 0.15 %
Temperature error Max.0.04%/K
Operatingfrequency
ULM–70_–02ULM–70_–06ULM–70_–10ULM–70_–20
120kHz75kHz50kHz30kHz
Beamwidth(-3dB)ULM–70_–02;10ULM–70_–06ULM–70_–20
10°14°12°
Ambienttemperaturerange ULM–70_–02;06ULM–70_–10;20
-30...+70°C-30...+60°C
Short-timetemperaturestressresistance +90°C/1hour
Max.operationoverpressure(ontransmissionsurface) 0.1MPa
Sensitivity 3steps(low–medium–high)
Damping 0 ... 99 s
Measuringperiod 1 ... 4 s
Risetime ca. 30 s
AdditionaltechnicaldataforExproof 2)–Max.internalvalues Ui=30VDC;Ii=132mA;Pi=0.99W;Ci=370nF;Li=0.9mH
Failure indication(echoloss,levelindeadzone3),internalfailure)
Adjustableinmodes:3.75mA;22mA;Lastmeasuredvalue
Maximalresistanceofcurrentoutputload(U=24VDC) Rmax=270Ω4)
Mechanicalconnection
ULM–70_–02ULM–70_–06ULM–70_–10ULM–70_–20
ScrewingwiththreadG1''ScrewingwiththreadG1½''ScrewingwiththreadG2¼''Aluminiumalloyflange
1) In case the level of bulk-solid materials is measured, the measurement range is reduced.2) Allowed pressure range in the zone 0: 80 ... 110 kPa.3) Dead zone = Blind zone = Blocking distance4) Including 250Ω resistor in case of HART® connection.
20 . specificaTions
ULM–70 © Dinel, s.r.o.29
facTory DefaulT
ULM–70_–02 ULM–70_–06 ULM–70_–10 ULM–70_–20
MINLEVEL(Distancetomin.level) 2 000 6000 10 000 20 000
MAXLEVEL(Distancetomax.level) 150 250 400 500
UNITS mm;%;°C mm;%;°C mm;%;°C mm;%;°C
DAMPING 2 5 10 10
SENSITIVITY MEDIUM MEDIUM MEDIUM MEDIUM
MEDIUMTEMPERATURE NO NO NO NO
FAILUREMODE–NOECHO 3.75mA 3.75mA 3.75mA 3.75mA
FAILUREMODE–DEADZONE1) 22mA 22mA 22mA 22mA
POOLINGADDRESS(HART®) 00 00 00 00
PASSWORD Nopassword Nopassword Nopassword Nopassword1) Dead zone = Blind zone = Blocking distance
area classificaTion (according to EN 60079-10 and EN 60079-14)
ULM–70N–__ Performance for non-explosive areas
ULM–70Xi–02ULM–70Xi–06
Explosiveproof–suitableforexplosiveareas(combustiblegasesorvapours)II1/2GExiaIIBT5Ga/GbwithIsolatingrepeater(IRU–420),thewholelevelmeter
–zone1,frontheadpart–zone0
ULM–70Xi–10Explosiveproof–suitableforexplosiveareas(combustiblegasesorvapours)
II1/2GExiaIIAT5Ga/GbwithIsolatingrepeater(IRU–420),thewholelevelmeter –zone1,frontheadpart–zone0
ULM–70Xi–20 Explosiveproof–suitableforexplosiveareas(combustiblegasesorvapours)II2GExiaIIAT5GbwithIsolatingrepeater(IRU–420),thewholelevelmeter–zone1
TecHnical specificaTions – Display MoDule
Displaytype MatrixOLED
Resolution 128x64pixel
Characterheight/Numberofdigitsmeasuredvalue 9mm/5Digits
Displaycolour Yellow
Buttons Membraneswitchpanel
Ambienttemperaturerange -30...+70°C
Weight 46g
TecHnical specificaTions – leVel MeTer
Protection class IP67
Weight
ULM–70_–02ULM–70_–06ULM–70_–10ULM–70_–20
0.3kg0.4kg0.6kg3.1kg
30© Dinel, s.r.o. ULM–70
useD syMbols
Toensuremaximumsafetyofcontrolprocesses,wehavedefinedthefollowingsafetyinstructionsandinformation.Eachinstructionislabeledwiththeappropriatepictogram.
Alert, warning, danger
Thissymbol informsyouaboutparticularly important instructionsfor installationandop-erationofequipmentordangeroussituations thatmayoccurduring the installationandoperation.Notobservingtheseinstructionsmaycausedisturbance,damageordestructionofequipmentormaycauseinjury
Information
Thissymbolindicatesparticularlyimportantcharacteristicsofthedevice.
Note
Thissymbolindicateshelpfuladditionalinformation.
21 . packaging, sHipping anD sTorage
ThedeviceDLM–35ispackagedinapolyethylenebag,andtheentireconsignmentisplacedintoacardboardbox.Asuitablefillermaterialisusedinthecardboardboxtopreventmechanicaldamageduring transport.
Removethedevicefromthepackagingonlyjustbeforeusing,therebyprotectingitfrompotentialdamage.
Aforwardingcompanywillbeusedtoshipgoodstothecustomer.Uponprioragreement,orderedgoodscanbepickedupinpersonatcompanyheadquarters.Whenreceiving,pleasechecktoseethattheconsignmentiscompleteandmatchestheorder,ortoseeifanydamagehasoccurredtothepackaginganddeviceduringtransport.Donotuseadeviceclearlydamagedduringtransport,butrathercontactthemanufacturerinordertoresolvethesituation.
If thedeviceistobefurthershipped, itmustbewrappedin itsoriginalpackagingandprotectedagainst impact and weather conditions.
Storethedeviceinitsoriginalpackagingindryareascoveredfromweatherconditions,withhumid-ityofupto85%withouteffectsofchemicallyactivesubstances.Thestoragetemperaturerangeis-20°Cto+60°C.
Level meters of variants ULM–53_– 02, 06, 10 are fitted with protective caps to prevent damage to the ultrasonic transducer. Remove the cover prior to putting into operation.
Dinel, s.r.o.U Tescomy 249
760 01 ZlínCzech Republic
Tel.: +420 577 002 003Fax: +420 577 002 007E-mail: sale@dinel.cz
www.dinel.cz
The lastest version of this instruction manual can be found at www.dinel.czVersion:
industrial electronics
3/2016
This project was realised under financial support from the funds of the state budget through the "Department of Trade and Industry".