F2 Series
GLOSSMETER
USER's MANUAL
ISO-2813RoHS
2 / 60 /85 GLOSS0
Vers
ion
20171215
-
Made in China
Add.: 264 Huangling, Zhangban Town, Taiwanese Investment Zone,
Quanzhou, Fujian 362123,
China
Tel.: +86-595 22391041 Fax: +86-595 22391048
E-mail: sales@ksj.cn
sales@hidea.cn
Website:www.ksj.cn
www.glossmeter.cn
Website:www.hidea.hk E-mail: sales@hidea.hk
Hidea Co., Ltd.Product Developer:
KSJ Photoelectrical Instruments Co., Ltd.Manufacturer:
Hidea
www.hidea.hk
m.en.ksj.cn
-
1.
Precautions
Important Notes
Brief Introduction
Specifications
Main Functions
Construction
Measurement
Menu Operation
Using Software
Trouble shooting
Warranty
Packing List
2
3
4
5
7
9
13
17
26
35
39
40
...........................
...........................
...........................
...........................
...........................
...........................
...........................
...........................
...........................
...........................
...........................
...........................
CONTENTS
-
©2014 KSJ instruments. All rights reserved.
KSJ and Hidea are either registered trademarks in China and other countries.
All other trademarks are property of their respective owners.
PRECAUTIONS
2.
-
IMPORTANT NOTES
Short for Micro Glossmeter
A set of measuring angles (Single: X, Double: XX, treble: XXX)
Product Series Code (First grade, Second generation)
Example MG268-F2 means that the product belong to F2 series with treble measuringangles.
Note: XXX are Arabic numerals 2=20 ,6=60 ,8=85 .
Model in production including:
This manual is based on the model MG268-F2, some descriptions may be different withyours, please pay attention to them.
MG268-F2 MG26-F2 MG68-F2 MG6-F2
The definition of haze is quoted from ASTM D4039.
Gloss quantity is denoted by Gs(X), X=20 ,60 or 85 . GU is short for Gloss Unit.
e.g. Gs(60 )=78GU, it says a measurement at incidence angle of 60 is 78 Gloss Unit.
When we mention a reflective index, a specified measuring angle should be related to.
3.
-
4.
All kinds of coating and finishing surfaces, such as paints, varnishes,
printing ink, etc.
Decorative materials, such as marble, granite, polishing brick, ceramic
tile and so on.
The other kinds of materials and objects, such as plastic, woodenware,
paper, etc.
F2 series meters are designed and manufactured according to ISO2813 and GB9754.
They are also compatible with the standards of ASTM , ASTM , GB/T13891,
GB/T7706, and GB/T 8807. The technical parameters conform with JJG696. The gloss
values can be traced to the those of the National Primary Standards in SIMT in the
P.R.China. The definition of haze is equivalent to that of ASTM D4039.
D523 D2457
Thanks for your choice to purchase the F2 Series Meter!
Main Applications:
BRIEF INTRODUCTION
-
5.
Specifications Unit F2 Series Glossmeter
Measuring Range
Resolution
Reproducibility
Repeatability
GU
GU
GU
GU
GU
GU
+10 +40
Gs 20
Gs 60
Gs 85
0.0 119.9
0.0 119.9
0.0 119.9
Gs 20
Gs 60
Gs 85
120 2000
120 1000
120 160
0.1 1
0.2
0.2
0.5
0.2
0.5
Zero-point Deviation
Operating Temperature
Deviation 1.5 1.5%
SPECIFICATIONS
-
6.
Specifications Unit F2 Series Glossmeter
-10 +60
Gs 20
Gs 60
Gs 85
9 10
9 16
5 39
Storage Temperature
Relative Humidity Up to 85%, no moisture condensation, for operation
Measuring Spot
g 400 ( with a battery and a holder)
155 48 75 ( without a holder)mm3
mm2
One AA-sized battery, 1.2V 1.5V,Rechargeable or alkaline.Power Supply
Consumption
Volume
Weight
It's different to battery types. An alkaline battery
may support 20000 measurements or more.
SPECIFICATIONS
-
7.
One key press to get three measurements of treble measuring angles. The result will be
automatically stored into memory.
Optional angle sets. You can let all three angles valid or only choose one or two angles.
Self-diagnosis and self-calibration.This function help to find some faults, and so you can
deal with them in time.
Friendly display interface. In measurement mode, current values, averages and other
information are synchronously shown on the Display Window. The clear characters and
simple menus make operation more easily.
Presettable group size. A number between 1 to 99 can be chosen for you.
Large data storing memory. In three-angle mode, more than 10000 measurements or
999 groups can be allowed to stored in F2 series meters.
Inquiry for data stored. Through menu operation you can browse the data in memory.
Haze values can be calculated and shown on the screen.
You can choose a measurement unit , GU (Gloss Unit) or % (Reflective Index).
MAIN FUNCTIONS
F2 series meters have the main functions as below:
-
8.
Deleting defective measurements. By pressing Measuring Button for two seconds or
more, the measurement displaying on the Display Window will be deleted. This function
is only effective in current group.
USB Adapter make it easier that a meter communicates with a PC.
Practical and friendly software. KSJ Data Operator is not only friendly but practical. It
can help you to upload the data from the , preset the measuring conditions, make
a printout, perform data processing and export the data to an Excel document.
Low power consumption. need only a 1.2v to 1.5v DC power supply.
You can choose a piece of AA battery. An alkaline battery will
support more than 20000 measurements. The power for communicating is taken from
a PC.
Warning for low battery and full memory.
Automatic shutoff. If no operation will turned off .
meter
F2 series meters
alkaline or rechargeable
in a preset time, the meter
MAIN FUNCTIONS
Note: Since this series meters are neither multilingual nor time settable, the releventfunctions of the software are invalid.
-
Screen
Front Shell
Holder Bat. Chamber Cover
Rear Shell
Menu Wheel
9.
Whole Device (Model MG268-F2)
CONSTRUCTION
-
10.
Rear ShellMeasuring Button
Mark For LocationIncidence Direction Front Shell
RS232 Port
Label
Measurement Aperture
Main Device (Model MG268-F2)
CONSTRUCTION
-
Holder
Calibrating Standard
Gloss Value Label
11.
Holder With Calibrating Standard
CONSTRUCTION
-
12.
Battery in Position
Battery Chamber Cover
Front Shell
Rear Shell
Battery With Positive Pole Up!
AA or LR6, Alkaline or Rechargeable.
CONSTRUCTION
-
13.
P l e a s e1 . C l e a n S T D2 . S e t H o l d e r
MEASUREMENT
Turn on the meterLay the main device into the holder, and push down the Menu Wheel t .The infos of model and series number will be shown in the Display Window, meanwhilethe meter performs self-diagnose and self-calibration. After passing this procedure thedisplay should turn to measuring mode, the meter would be ready for operation. Or elseit shows as below:
o turn on a meter
Notes:1) Sometimes a great change of environment temperature might cause a self-calibration
failed. In this case, the meter needs some time to establish a balance of temperaturebefore calibrating.
2) If a main device don't be lay into the holder, it can also be turned on by pushing downthe Menu Wheel without self-diagnose and self-calibration.
The messages prompt you to confirm that:1) The calibrating Standard (STD) is clean;2) The main device is set into the holder properly.
Please check and calibrate the meter manually !(Refer to page 21 and page 35~36.)
-
14.
MEASUREMENT
Measuring and measurement interface
008/03/05 means that the current measuement isthe third of five measurements in the eighth group.Angle mode is shown in the first column, the valuesof a current measurement in the second and theaverages in the third.
Remove the holder ement, then push the Measuring Button, and a measurement willbe completed
within 2 seconds. Meanwhile the result in Screen as below and storedinto the memory automatically.
and lay the Measur Aperture just against the position that youwant to measure
will be shown the
008/03/0520
!" 87.4 87.5
6 " 91.9 91.885" 98.8 98.7
0
Note:
Turn off the meter
The meter will be turned off automatically by a preset time. See page 25.
Note:
After turning off the meter will be in a sleeping state and consume little power. Even soit is strongly recommended to remove the battery when a meter be left unused for a longtime, e.g. a moth or longer.
-
15.
Deleting measurements
Note: Only the measurements in a current group can be deleted one by one.
To delete a current measurement, e.g. a defective measurement, please push down theMeasuring Button for more than 2 seconds.
MEASUREMENT
Notes for measuring
The holder removed fro m the main device should be kept in a safe and clean place toprevent from damage or pollution. If not in use, please lay the device into the holder.
While measuring keep environment light from directly irradiating into theMeasurement Aperture.
pleaseIf necessary, especially under strong light, shade it with a
piece of light-tight cloth.
A big temperature difference between environment and the meter would badly affectmeasuring. In this case please wait for some time till the temperature get to a balanceand then calibrate the meter again.
When moisture condensation occurs, a measurement may not be gotten correctly .
-
16.
MEASUREMENT
Pay attention to keep Calibrating Standard and Measurement Aperture clean. Any dirt,such as fingermarks, dust, steam and filth, should be cleaned gently with lens paperor a piece of soft cloth dipped in absolute ethanol.
When a battery runs out, the following message will be shown. Please replace the oldbattery with a new one.
Memory is Full
If the memory is fully stored by data, a message will be shown as below on the Screen.In this case measuring operations will be valid, but the results can't be stored into thememory any more. It is better to upload the data to a PC and clear the memory.
Low Power
If measurement operation last a long time, for example an hour or more, it's necessaryto recalibrate the meter.
-
17.
MENU OPERATION
Menu interface
Note: To get back to the measurement interface please pexcept some specials.
ress the Measuring Button. It isvalid for most interfaces,
# Enter the next submenu
} Return to the prior menu
Option selecting
Turn up or down the Menu Wheel to select the options. Push down the Menu Wheel forentering your option selected. The following describes what the symbols mean.
On the Interface, push the Menu Wheel to enter the main menu..measurement
Main Menu
Data
Cal ibration
Setup
Information
*
#
#
#
}
]
* *} # >Selected item Current setup
[ ] hided items before/behind the mark
-
18.
MENU OPERATION
Main Menu Option
Data menu and its submenu
Data *#
Da ta Br ow se
Gr oup Num be r
0 0 1
Dele te Al l D a ta ?
Ye s # }No
Data
Browse
Delete Al l
*}
#
#
This is a headline showing that it is a main menu interface. Select this option and pushdown the Menu Wheel to return to measurement interface.
-
19.
MENU OPERATION
submenuBrowse
In data menu select browse option and push down the Menu Wheel to enter the browsesubmenu. Turn up and down to set the group number that you want to browse andthen push down the Menu Wheel to enter the browse interface. Here you can browseall the measurements in this group by turning the Menu Wheel.
Note:
GRP001:01/0520
!" 16.5 17.6
6 " 49.4 51.785" 81.5 84.9
0
1)!
GRP001:01/05 means the first measurement in a group that is named as GRP001and preset to involve five measurements. represents the average of this group.
2) Turn up and down the Menu Wheel to browse data. Press the Measuring Button toback to the last interface for another operation.
3) Push down the Menu Wheel, then turn the Menu Wheel to change the group one byone in order. Push down the Menu Wheel again to browse values.
-
20.
submenuDelete All
Enter the submenu and turn the Menu Wheel to select to clear all data in memorypermanently. Or else select
OK
NO to exit.
Calibration Menu and its submenu
Calibrate the meter immediately!
MENU OPERATION
Calibration *#
Cal ibration
Modify Values
Cal ibrate
}
*
Modify Values20" 0087.56 " 0092.085" 0098.7
0
*}
#
#
-
21.
MENU OPERATION
submenuModify Values
optionCalibrate
Note:
Note:
1) Before calibrating be sure that the Calibrating Standard is clean!2) You can also calibrate the meter with Verifying Standard that is available for purchase.
2) When you have chosen Reflection in Unit submenu, this submenuwill be invalid! Seepage 24.
1) The values here are Standard values which measuring is based on, you can modifythem only when the standard has been verified by an authoritative organization.
Enter the submenu, turn the Menu Wheel to select a value and push down the MenuWheel to confirm, then , turn the MenuWheel to change the value, and finally push down the Menu Wheel to exit. the sameoperation for other values.
push down the Menu Wheel to shift cursor right
In menu, turn the Menu Wheel to select and pushdown the Menu Wheel to perform it. The calibrating will be carry out automatically, theresult will be shown on the Screen, and it will return to the measurement interface.
Calibration Calibrate option
-
22.
MENU OPERATION
Setup menu and its submenu
Data Num be r pe r .
gr oup
N = 0 5
Angle Mode
20
6
" , 60 "
" , 85 "
" , 60 " ,85 "
0
20
*}
[
> #
#
Setup *#
Unit
Gloss Un it
Re fle cta n ce
*}
> #
#Standby Time
Off
30s
60s
120s
}
>
]
#
#
*#
Se tup
Angle Mode
Haze
Uni t
Sh andb y Tim e
G roup
*}
]
#
#
#
#
-
23.
MENU OPERATION
submenuAngle Mode
submenuHaze
meter with three angles has six angle modes as below,while the meters with one or two angles have their different submenus:Angle Mode submenu of the
Push down the Menu Wheel to switch to the function. When is on, a hash symbolshows at the right of option.
Haze
Haze
Select an option and push down the Menu Wheel to confirm..
Then it will go back to theSetup menu automatically
Angle Mod e
20
6
" , 60 "
" , 85 "
" , 60 " ,85 "
0
20
An gle Mode
2 0
6
"
"
85 "
0
1) Haze option will be able only when one of the angle modes that includes both 20 and60 have been selected, meanwhile in Unit menu is valid.Gloss Unit
2) Haze data are shown at the last line on the Screen.
Note:
*}
[
#
#
*}
]
#
#
-
submenuUnit
Enter Unit submenu, turn the Menu Wheel to select the Unit option, then push down theMenu Wheel to confirm, and it will go back to the Setup menu automatically.
Note:
In Reflectance mode, the standard values modifying is unavailable! See page 21.
008/03/0520
!" 87.4 87.5
6 " 91.9 91.80
H 4.5 4.3
GRP001:01/0520
!" 16.5 17.6
6 " 49.4 51.70H 32.9 34.1
Setup
A n gle Mode
H a ze
U n it
*}
]
#
> #
Angle Mode
20 " ,60 "
" , 60 " ,85 "20
*}
#
#
24.
MENU OPERATION
-
25.
MENU OPERATION
submenuGroup
Note:
Caution:
The optional value must be between 1 to 99.
When you select , the meter wouldn`t be turn off till the battery is used out.OFF
It shows a Hardware Number and a SoftwareNumber of a meter. Any key to exit.
submenuStandby TimeThis menu is for setting a no-operation time interval over which shutdown automatically. In the menu select an option and push down the Menu Wheel toconfirm. And t .
the meter will be
hen it will go back to the Setup menu.
Information interfaceHardware Number
F27139001
Software Number
F20114021
Enter the menu, turn the Menu Wheel to set a number of measurements in a group andconfirm, then it will return to the measurement interface.
-
26.
USING SOFTWARE
Brief introduction
Software installation and uninstall
The software is on the CD coming with the . The procedure of intallation and unin-stall is the same as a normal software, please follow the prompts.
meter
KSJ Data Operator is a software for data processing. We can use it to upload the data thatis stored in the memory of the meter to a PC, then process them, export them as an Efile. We can also operate on a PC to set the measuring conditions for the and clearits memory.
xcelmeter
Note:
Please pay attention to backup the file of KDO.mdb in the install directory termly toprotect your data from missing.
To use KSJ Data Operator, you need:
an IBM compatible PC with a CD-ROM.Microsoft Windows 7 or later versions.
-
27.
USING SOFTWARE
How to connect an F2 series meter with a PC
There is a cable provided with the meter in the package. That is USB Adapter used for themeter to connect with a PC to upload data. To use the USB Adapter, please do as follows:
Pre-install the driver for the USB Adapter
Using the USB Adapter
Open the directory of indin the CD and
double click it to install the driver.
After pre-installing the driver, please plug theUSB Adaptor to a USB port on the PCand the other end to the meter. The USB Adapter should get ready to work.
Note:
To uninstall the driver, please run PL2303_DriverInstaller.exe again, or enter theControl Panel / Uninstall a program to get it uninstalled.
-
28.
Software interface
Upload interface (default)
USING SOFTWARE
Title bar
Menu bar
Zone for uploading
Data area
Founction keys
-
29.
Search interface
USING SOFTWARE
Title bar
Menu bar
Data area
Founction keys
-
30.
Menu commands
Setup menu
Meter Settings menu
The current active set of reference values is highlighted in red color. You can selectany one else and apply it as an active one.
This group of commands are used to set the measuring conditions for a glossmeter ina .communication state
USING SOFTWARE
Language
Reference
Angle ModeThere are six options to be selected including single, double and treble angles.
Here you are allowed to make a new set of reference values which can be used asquality control values to decide which average is passed or failed and other results.You can also edit or delete an existed set of reference values.
This software provides multilingual user interfaces, you can choose one available.'s
Note:
Group SizeTo preset a group size, optional numbers among 1- 99.
-
Note: It is very important to keep the verified values of a Calibrating Standard .Therefore, don't modify those values at will.
correct
31.
Verified Values
Standby Time
Clear MemoryThis is for clearing all the data stored in the memory of a glossmeter permanently.And so it is better to reconsider before clearing.
This command is for presetting a no-operation time interval for which the meter willbe shut down automatically. There are four options to be selected including "off". Ifyou choose "off" option, the meter will run till a battery is used up.
This is for modifying the verified values of a Calibrating Standard in the Holder.
USING SOFTWARE
Language and Time Syn
B are unavailable for this .oth of the functions series meters
Note: Since the series meters are neither multilingual nor time settable, the relevantfunctions of the software are invalid.
-
32.
USING SOFTWARE
Help menu
Function Keys and others
Contents
Select All
Inverse
DeleteTo delete selected data shown in the data area permanently.
If you have selected some rows of data in the data area, make an inverse selectionby pressing this key.
To select all data shown in the data area.
About KSJ Data OperatorAbout the copyright and version of KSJ Data Operator and other information.
To show the contents of help
ExportTo export data selected in the data area to an Excel file. When no data is selected,the all data shown will be exported.
Note: When a group is selected by ,the measurements in the group will shown in Original Data area.
checking the box at the beginning of the group
-
33.
USING SOFTWARE
Data Search
Connect and Upload keys
Connect a glossmeter to a PC with a USB Adapter, turn on the glossmeter and runKSJ Data Operator, then click Connect key to set up a communication, finally pressUpload key to do data uploading.
When you click Search key or Search tag, a dialog will pop up. Choose a meter typeand a Meter ID in the dropdown lists, and check the search conditions, by uploadingtime and / or by group name. Finally click Search key to find designated data.
Note:
Note: Online Measurement function is unavailable for this series meters.
1) Data is uploaded group by group with prompt tone.2) Before uploading a communication must be set up, so that KSJ Data Operator
can identify a meter type and get a meter ID.3) KSJ Data Operator can receive and manage more than one KSJ's glossmeters.
-
34.
USING SOFTWARE
Group Data area
In this area Group Data are shown. The items are Group Name, Group Size, Mea-surement Angle, Average, Standard Deviation, Pass/Failed, Max value, Min value,Measurement Time and Upload Time.
Original Data area
When you check a box at the beginning of a group, the original measurements inthis group will be shown in Original Data area. The items are group Name, Numa-rical order, values at a measurement angle and their Deviations.
Note:1) The determination of P/F is based on the active reference set which defines the
Max values and the Min values. If an average exceeds the range defined, it willbe noted with fail , inversely with pass .
2) Since this series meters get a measurement with no timestamp, the measurementtime is unavailable.
Note: A deviation is a result coming from comparing a measurement value and areference value.
-
35.
P l e a s e1 . C l e a n S T D .2 . S e t H o l d e r .
TROUBLE SHOOTING
Errors in self-diagnose and self- calibration
When the meter is turned on with the Holder, thewarning message (on right hand side) may comeout. That means something is wrong.
Please check and settle the troubles in turn:
ny dirt, such as dust
fingermarks, steam and filth, is found, it should be cleaned gently with lens paper or a
piece of soft cloth dipped with absolute ethanol ( for best results, add a same volume
absolute ether !).
self-diagnose and self- calibration
1. Clean the Calibrating Standard.
Observe the surface of the Calibrating Standard against light. If a
Some abrasion and damage on the surface will change its verified
values. If so, please contact the manufacturer to deal with.
2. Set the main unit into the Holder.
Only when the main unit lay in the Holder properly, the meter can pass the procedure
of
Clean STD.
Set Holder.
-
36.
Moisture condensation
Corrosion caused by a bad battery
When the meter is taken from a cold location to a hot environment, moisture conden-sation occurs easily. If so don't use the meter any more and wait for the moisture toevaporate. See above item also.
A bad or over-discharged battery will produce corrosive liquid to damage the electricpoles and elements. When it happens, please remove the battery in time and clean thepoles with a cotton bud dipped with absolute ethanol. If the meter don't work well, pleasecontact us for repair.
TROUBLE SHOOTING
Besides the two of the above problems, a big difference of environment temperaturemay cause a self-calibration failed. In this case, you may take the meter out of a roomor closing to a hot source, or it may be a great weather change. To reduce the effectthe meter needs some time to get a balance before recalibrating. For calibration seepage 21 also.
-
37.
TROUBLE SHOOTING
Problems with communication
It is all right to use a USB port to carry out a communication. Problems uoccur because :
The meter doesn't connect with PC properly. e.g. a poor contact and so on.The meter is not in a turn-on state.No data is in the meter's memory.
sually
Failing to communicate through a USB port on a PC
And please be sure that the driver of the USB adapter has been preinstalled properly.
-
38.
How to verify your glossmeter
A glossmeter is a kind of metrical instruments. It is important and necessary to keep itsaccuracy. For this reason the meter must be verified in a period of time, usually a year.Then, we recommend you to verify the glossmeter as follow:
Send it to a public organization which is authorised by the government and ableto perform the gloss verifying.Send it to a manufacturer that produces glossmeters.Purchase certified standards to verify the glossmeter by yourself.
Note:
TROUBLE SHOOTING
We also provide some models of Verifying Standards as below.
MG268S-H, high gloss.MG268S-M, medium gloss.MG268S-L, low gloss.
-
39.
Limited Warranty And Technical Support
The limited warranty period is one year commencing on the date of purchase.
We provide spare parts anytime for distributors and customers.
The inquiries on the gloss measurement may be directed to sales@ksj.cn
We also provide glossmeter customizing for a special application.
Note:
If your
meter needs service, carry it to the local reseller or contact us at http://www.ksj.cn
for repair.
The meter should not be taken down at will in order to avoid precision destroying
and maintaining difficulty.
If the breakdown is caused by taking-down or inappropriate use, we reserve the
rights not to guarantee it free in repair.
WARRANTY
-
40.
Holder With a Calibrating Standard
USB Adapter
KSJ Data Operator
Carrying case
User's Manual
Tracable Certificate
1 Unit
1 Piece
1 CD
1 Piece
1 Piece
1 Piece
PACKING LIST
One unit of Main Device, the model as below:
MG68-F2 MG6-F2
MG268-F2 MG26-F2
/ColorImageDict > /JPEG2000ColorACSImageDict > /JPEG2000ColorImageDict > /AntiAliasGrayImages false /CropGrayImages true /GrayImageMinResolution 300 /GrayImageMinResolutionPolicy /OK /DownsampleGrayImages true /GrayImageDownsampleType /Bicubic /GrayImageResolution 300 /GrayImageDepth -1 /GrayImageMinDownsampleDepth 2 /GrayImageDownsampleThreshold 1.50000 /EncodeGrayImages true /GrayImageFilter /DCTEncode /AutoFilterGrayImages true /GrayImageAutoFilterStrategy /JPEG /GrayACSImageDict > /GrayImageDict > /JPEG2000GrayACSImageDict > /JPEG2000GrayImageDict > /AntiAliasMonoImages false /CropMonoImages true /MonoImageMinResolution 1200 /MonoImageMinResolutionPolicy /OK /DownsampleMonoImages true /MonoImageDownsampleType /Bicubic /MonoImageResolution 1200 /MonoImageDepth -1 /MonoImageDownsampleThreshold 1.50000 /EncodeMonoImages true /MonoImageFilter /CCITTFaxEncode /MonoImageDict > /AllowPSXObjects false /CheckCompliance [ /None ] /PDFX1aCheck false /PDFX3Check false /PDFXCompliantPDFOnly false /PDFXNoTrimBoxError true /PDFXTrimBoxToMediaBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXSetBleedBoxToMediaBox true /PDFXBleedBoxToTrimBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXOutputIntentProfile () /PDFXOutputConditionIdentifier () /PDFXOutputCondition () /PDFXRegistryName () /PDFXTrapped /False
/CreateJDFFile false /Description > /Namespace [ (Adobe) (Common) (1.0) ] /OtherNamespaces [ > /FormElements false /GenerateStructure false /IncludeBookmarks false /IncludeHyperlinks false /IncludeInteractive false /IncludeLayers false /IncludeProfiles false /MultimediaHandling /UseObjectSettings /Namespace [ (Adobe) (CreativeSuite) (2.0) ] /PDFXOutputIntentProfileSelector /DocumentCMYK /PreserveEditing true /UntaggedCMYKHandling /LeaveUntagged /UntaggedRGBHandling /UseDocumentProfile /UseDocumentBleed false >> ]>> setdistillerparams> setpagedevice