Extractables Evaluation of a Single-Use Bioprocess Bag
According to Published Recommendations
April 2016
Copyright Chemic Laboratories Inc., 20161
Study Objective
An independent evaluation comparing various SUS extractable study parameters .
To generate data sufficient to evaluate the merit of solvents, times, analytical techniques, and procedures.
To evaluate and optimize analytical methodologies with a consideration toward extract compatibility.
Copyright Chemic Laboratories Inc., 20162
Timeline (2015 Activities)
Copyright Chemic Laboratories Inc., 20163
Solvent
Time/Temperature
T=0 (≤ 30 min)
at 25°C
24 hours
at 40°C
21 days at 40°C
30 days at 40°C
70 days at 40°C
BP
OG
50% EtOH Yes Yes Yes Yes Yes
1% PS-80 Yes Yes Yes Yes Yes
5M NaCl Yes Yes Yes Yes Yes
0.5N NaOH Yes Yes Yes Yes Yes
0.1M H3PO4 Yes Yes Yes Yes Yes
USP H2O Yes Yes Yes Yes Yes
HEX Yes Yes Yes Yes Yes
100% EtOH - - - - Yes
100% IPA - - - - Yes
0.01M HCl - - - - Yes
Study Design
SolventInstrumental Analysis
HPLC-DAD/MSGC/MS HS-GC/MS ICP/MS1 TOC pH NVR1
APCI +/- ESI +/-50% EtOH Yes Yes Yes Yes Yes - Yes -1% PS-80 Yes Yes Yes Yes Yes - Yes -5M NaCl Yes Yes Yes Yes Yes Yes Yes -
0.5N NaOH Yes Yes Yes Yes Yes Yes Yes -0.1M H3PO4 Yes Yes Yes Yes Yes Yes Yes -
USP H2O Yes Yes Yes Yes Yes Yes Yes YesHEX Yes Yes Yes - - - - Yes
100% EtOH1 Yes Yes Yes - Yes - - -100% IPA1 Yes Yes Yes - Yes - - -
0.01M HCl1 Yes Yes Yes - Yes - Yes -
38 extracts370 unique assays
1Only the final (70 day) time point was assessed.
Copyright Chemic Laboratories Inc., 20164
Study Design
Commercially available 1-liter 2D bioprocess bags were γ-irradiated at 45-55 kGy and all associated components (ports, tubing, etc.) were removed and heat sealed with an internal wetted surface area of 726 cm2. Each test sample was filled with 121 mL of extraction solvent in order to achieve a 6:1 (cm2/mL) surface area: volume ratio.
Copyright Chemic Laboratories Inc., 20165
Study Design
Samples were prepared for each solvent & each time point, placed in secondary containment within an incubator. The study was staged in reverse within 5 weeks of γ-irradiation.
At termination the sample extracts and controls were placed into inert containers for analytical testing.
Copyright Chemic Laboratories Inc., 20166
Solvent
% Solvent Loss (w/w)
24 hours at 40°C
21 days at 40°C
30 days at 40°C
70 days at 40°C
50% EtOH 0.12% 2.6% 3.7% 8.7%
1% PS-80 0.093% 2.0% 2.6% 6.0%
5M NaCl 0.058% 1.1% 1.4% 3.3%
0.5N NaOH 0.078% 2.0% 2.4% 5.4%
0.1M H3PO4 0.080% 2.0% 2.7% 5.2%
USP H2O 0.087% 1.9% 2.6% 5.9%
HEX 0.0% 1.8% 2.1% 5.2%
100% EtOH - - - 12%
100% IPA - - - 0.0%
0.01M HCl - - - 6.0%
Solvent Loss
By ensuring the integrity of the sealing procedure and using secondary containment, solvent loss was not a significant concern.
Copyright Chemic Laboratories Inc., 20167
pH Results
Copyright Chemic Laboratories Inc., 20168
NVR Results
DescriptionExtraction
SolventMass Difference (initial vs. final)
NVR (μg/cm2)
Method Control USP H2O -0.0003 NA
Sample Extract USP H2O 0.0003 0.93
Method Control Hexane 0.0001 NA
Sample Extract Hexane 0.0982 300
70 day USP Water & Hexane extracts were transferred to appropriately cleaned vessels and evaporated to dryness.
Copyright Chemic Laboratories Inc., 20169
TOC Results Notables
Devitrification of quartz and catalyst poisoning due to sodium content of extract solutions.
Chemic has invested in 2 new state of the art TOC units, a combustion unit and a UV persulfate unit to resolve this issue.
Copyright Chemic Laboratories Inc., 201610
Crystallization of glass
Catalyst surface contaminant
TOC Results
DescriptionExtraction
Solvent
Results (μg/cm2)
T=0 (≤ 30 min)
at 25°C
24 hours at 40°C
21 days at 40°C
30 days at 40°C
70 days at 40°C
Sample Extract 5 M NaCl < LOQ < LOQ < LOQ < LOQ < LOQSample Extract 0.5 N NaOH < LOQ < LOQ < LOQ < LOQ < LOQSample Extract 0.1 M H3PO4 < LOQ < LOQ < LOQ < LOQ < LOQSample Extract USP H2O 0.034 0.13 0.23 0.61 0.72
Quantitation was performed using external standardization. Linearity, precision, system suitability and internal QC met predetermined specifications.
The instrumental LOQ = 200 ppb (≈ 0.03 μg/cm2). Salt and base extract matrix effects resulted in a 1000X LOQ increase.
Copyright Chemic Laboratories Inc., 201611
ICP/MS Results
70 day extracts were assayed for 30 metals via ICP/MS. Linearity, precision, system suitability and internal QC met predetermined specifications.
Extraction Solvent Analyte Concentration (μg/cm2)
50% EtOH ND
1% PS-80 ND
5M NaCl Copper 2.2
0.5N NaOH Copper 0.06
0.1M H3PO4 ND
USP H2O ND
100% EtOH ND
100% IPA ND
0.01M HCl ND
Copyright Chemic Laboratories Inc., 201612
Qualitative Results
Solvent
Number of extractable analytes across all time points(Number of analytes unique to extraction solvent)
HPLC-DAD/MS DI-GC/MS HS-GC/MS
50% EtOH 9 (2) 4 (0) 14 (12)1% PS-80 14 (11) 2 (0) 2 (2)5M NaCl 0 (0) 0 (0) 1 (1)
0.5N NaOH 10 (4) 1 (0) 0 (0)
0.1M H3PO4 1 (0) 0 (0) 0 (0)
USP H2O 0 (0) 0 (0) 0 (0)
HEX 9 (6) 21 (8) - -100% EtOH 12 (3) 12 (1) - -100% IPA 8 (0) 16 (1) - -
0.01M HCl 1 (0) 0 (0) - -
Copyright Chemic Laboratories Inc., 2016
All Organic solvents (mixtures) isolated a min. of 1 unique analyte whereas Aqueous solutions were not always successful.
13
HPLC-DAD/MS System Suitability
Criteria Parameter Specification Result
BPOG Criteria
Standards BPA and Irganox 1010 -
Limit of
DetectionBPA, S:N ≥ 3:1 4.7:1-12.6:1
Precision 1 ppm BPA, RSD ≤ 20% (n=6) 0.067-2.6%
Spike Recovery
1ppm BPA in H2O;
80-120% recovery101-102%
1ppm BPA in 50% EtOH;
80-120% recovery100-101%
Chemic Criteria
Linearity R2 ≥ 0.995
2,4-di-tert-butylphenol 0.993
Irganox 245 1.00
Irganox 259 1.00
Irganox 1010 1.00
Irganox 1076 1.00
Irgafos 168 1.00
BPA 1.00
Irganox 1010 1.00
Limit of
QuantitationBPA, S:N ≥ 10:1, 80-120% recovery 21.7:1-48.3:1, 96.2-101%
Accuracy1 ppm BPA check std, 80-120%
recovery93.8-102%
Copyright Chemic Laboratories Inc., 201614
HPLC-DAD/MS Results Notables
The results indicated the presence of extractable peaks in all solvents with the exception of 5M NaCl and USP H2O.
Linearity was run for eight common antioxidants establishing the instrumental range of the method from 0.1 to 2.5 μg/mL.
Copyright Chemic Laboratories Inc., 201615
HPLC-DAD/MS ResultsExtraction Solvent Analyte
Concentration (μg/cm2) (number of analytes)
30 Minutes 24 Hours 21 Days 30 Days 70 Days
50% EtOH
2, 4-Di-tert-butylphenol (DAD; 280nm)
ND 0.092 0.25 0.87 0.44
bDtBPP (DAD; 280nm) ND 0.032 0.05 0.08 0.07
Total unknowns ND 0.033 (1) 0.32 (5) 0.17 (2) 0.65 (7)
1% PS-80
2, 4-Di-tert-butylphenol (DAD; 280nm)
ND 0.26 0.61 0.62 1.3
bDtBPP (ESI-) ND ND 1.1 1.3 0.14
Total unknowns 2.8 (1) 11.7 (4) 4.1 (3) 21 (6) 7.0 (6)
5 M NaCl None detected ND ND ND ND ND
0.5 N NaOH
2, 4-Di-tert-butylphenol (DAD; 280nm)
ND ND ND 0.58 0.10
Total unknowns ND 0.23 (1) 4.2 (2) 7.4 (6) 3.3 (7)
0.1 M H3PO4 Total unknowns ND ND ND 0.04 (1) NDUSP H2O None detected ND ND ND ND ND
HexanebDtBPP (ESI-) ND ND ND 2.1 0.90
Total unknowns 4.8 (2) 7.1 (2) 11 (4) 3.0 (2) 3.6 (3)
100% EtOH
2, 4-Di-tert-butylphenol (DAD; 280nm)
0.32
bDtBPP (DAD; 280nm) 0.11
Irgafos 168 phosphate (APCI+) 0.41
Total unknowns 3.5 (9)
100% IPA
2, 4-Di-tert-butylphenol (DAD; 280nm)
0.46
bDtBPP (DAD; 280nm) 0.15
Irgafos 168 phosphate (APCI+) 0.47
Total unknowns 2.6 (5)0.01 M HCl Total unknowns 0.16 (1)
Copyright Chemic Laboratories Inc., 201616
DI-GC/MS System Suitability
Criteria Parameter Specification Result
BPOG Criteria
Standards n-Octane and BHT -
Limit of
DetectionBHT, S:N ≥ 3:1 2.0:1-3.6:1
Precision 1 ppm BHT, RSD ≤ 20% (n=6) 4.7% (n=6)
Spike Recovery1 ppm BPA in H2O; 80-120% recovery 96.8-107%
1 ppm BPA in 1% PS-80; 80-120% recovery 96.6-106%
Chemic Criteria
Standards BHT -
Linearity R2 ≥ 0.995 BHT 1.00
Limit of
QuantitationBHT, S:N ≥ 10:1, 80-120% recovery ~3.7:1-4.8:1, 90.6-94.1
Accuracy1 ppm BHT check (bracketing) standard,
80-120% recovery84.5-116%
Copyright Chemic Laboratories Inc., 201617
DI-GC/MS Results Notables
The data demonstrated the presence of extractable peaks in all solvents with the exception of 5M NaCl, 0.1M H3PO4, USP H2O and 0.01M HCl.
n-Octane was initially evaluated and response was erratic due to its volatility.
An extraction surrogate is added to monitor the efficiency of the liquid-liquid exchange procedure. The behavior of the surrogate may not be representative of all “unknowns”.
Copyright Chemic Laboratories Inc., 201618
DI-GC/MS ResultsExtraction Solvent Analyte
Concentration (μg/cm2) (number of analytes)
30 Minutes 24 Hours 21 Days 30 Days 70 Days
50% EtOH
1,3-Di-tert-butylbenzene 0.06 0.19 0.15 0.14 0.34
2,4-Di-tert-butylphenol ND 0.027 0.10 0.50 0.35
Total siloxanes ND ND ND 0.03 (1) NDTotal hydrocarbons ND ND ND 0.05 (1) ND
1% PS-801,3-Di-tert-butylbenzene 0.02 0.06 0.08 0.11 0.11
2,4-Di-tert-butylphenol ND 0.04 0.05 0.20 0.44
5 M NaCl None detected ND ND ND ND ND
0.5 N NaOH 2,4-Di-tert-butylphenol ND 0.04 0.05 0.05 0.02
0.1 M H3PO4 None detected ND ND ND ND NDUSP H2O None detected ND ND ND ND ND
Hexane*
1,3-Di-tert-butylbenzene 1.6 2.6 2.3 3.0 3.52,4-Di-tert-butylphenol 0.06 0.05 0.64 0.21 0.36
Total siloxanes 0.22 (2) 0.06 (1) 0.63 (2) 0.24 (2) 0.56 (2)Total hydrocarbons 0.35 (8) 2.9 (16) 3.9 (17) 3.2 (17) 3.7 (17)
100% EtOH*
1,3-Di-tert-butylbenzene 3.3
2,4-Di-tert-butylphenol 0.48
Total siloxanes 0.22 (1)
Total hydrocarbons 0.36 (9)
100% IPA*
1,3-Di-tert-butylbenzene 2.8
2,4-Di-tert-butylphenol 0.75
Total siloxanes 0.19 (1)
Total hydrocarbons 0.96 (13)
0.01 M HCl None detected ND
*Samples directly injectedCopyright Chemic Laboratories Inc., 201619
Headspace-GC/MS System Suitability
Criteria Parameter Specification Result
BPOG Criteria
Standards Methanol, MEK and D4 -
Limit of
DetectionMEK, S:N ≥ 3:1 3.4:1
Precision 1 ppm MEK, RSD ≤ 20% (n=6) 7.9-11% (n=6)
Spike
Recovery
1 ppm MEK in H2O;
70-130% recovery76.4-108%
1 ppm MEK in 1% PS-80;
70-130% recovery96.2-114%
Chemic Criteria
Limit of
DetectionD4, S:N ≥ 3:1 4.4:1-5.6:1
Accuracy1 ppm MEK check (bracketing)
standard, 70-130% recovery75.0-118%
Spike
Recovery
1 ppm D4 in H2O; 70-130% recovery 96.6%
1 ppm D4 in 1% PS-80;
70-130% recovery243%
Copyright Chemic Laboratories Inc., 201620
HS-GC/MS System Suitability Notables
Methanol was demonstrated to be unsuitable.Residual methanol in the ethanol further compounded the difficulty in interpretation.
MEK and D4 spiked at 2.5 μg/mL were well resolved but displayed significant differences in relative response factors.
Toluene-d8 is not water soluble and is may not be appropriate without added organic solvent modifier.
System suitability criteria was met for MEK, however D4 recovery was not met in PS-80 due to co-elution.
Copyright Chemic Laboratories Inc., 201621
HS-GC/MS Results
Extraction Solvent AnalyteConcentration (μg/cm2) (number of analytes)
30 Minutes 24 Hours 21 Days 30 Days 70 Days
50% EtOH
Methylcyclopentane ND ND ND ND 0.63
Cyclohexane ND ND ND ND 0.18
Diethoxyethane 0.04 ND ND 0.02 0.34
Triethoxyborate ND ND ND ND 2.1
Octene ND ND ND ND 0.10
Octamethylcyclotetrasiloxane ND ND ND ND 0.23
Decamethylcyclopentasiloxane ND ND ND ND 0.04
1,2-Di-tert-butylbenzene 0.02 0.06 0.03 0.03 0.41
Total unknowns ND ND 0.02 (1) ND 0.59 (5)
1% PS-80Octamethylcyclotetrasiloxane 0.18 ND ND ND ND
1,3-Di-tert-butylbenzene 0.02 0.08 0.09 0.09 0.08
5 M NaCl tert-Butanol ND ND 0.03 ND 0.03
0.5 N NaOH None detected ND ND ND ND ND0.1 M H3PO4 None detected ND ND ND ND ND
USP H2O None detected ND ND ND ND ND
Extraction ConditionNumber of Extractable Analytes
DetectedRepresentative Compound Classes
3 grams dry film incubated at 40°C
for 1 hour29
Linear, cyclic and branched alkanesLinear and cyclic siloxanes
AlcoholsKetonesSilanols
3 grams dry film incubated at 80°C
for 1 hour45
Copyright Chemic Laboratories Inc., 201622
General Study Conclusions
Although limited in scope this study provides:
-Additional information regarding aggressiveness of thesolvents employed.
-Insight into the variability that could be encountered.
- Analytical challenges to be overcome associatedwith solvent-instrument compatibility.
Copyright Chemic Laboratories Inc., 201623
General Study Conclusions cont.
-Additional sample handling and analytical methodoptimization is required to ensure robust evaluation(s).
-Significant investment is required to conduct studiesto the level of rigor necessary to obtain usable data.
Copyright Chemic Laboratories Inc., 201624
Key Facility Requirements
Large chambers are required for a single study
“Typical” TOC combustion units are not sufficient based on Na+ concentrations.
All MS platforms must have dual ionization sources and ability to switch polarity +/-.
A heat sealer is required to mitigate solvent loss from films.
Copyright Chemic Laboratories Inc., 201625
Special Thanks To:
Study plan reviewers(members of: BPOG, BPSA & USP and Industry SMEs)
Test material supplier
Chemic Laboratories’ analytical staff
All others who have contributed to ongoing discussions
For an expanded version of todays presentationvisit: http://www.chemiclabs.com/publications.php