7/31/2019 American Ecotech Series B Analyzers.sflb
1/3
TheEC9800
ambient
air
analyzers
are
intelligent
analyzers
with
state
of
the
art
features,
outstanding
gas
measurement
sensitivityand reliability,anda lowcostofownership.Theanalyzers in this familyprovidemeasurementsofozone,
carbonmonoxide,sulfurdioxide,andnitrogenoxidesandareallUSEPAEquivalentMethodcertified.
TheEC9800BSeriesanalyzersrequireexternalpumpsgivingthesystem integratorthepossibilityofusingmorethan
onepumpperanalyzer.Byunbundlingmanyofthesupportinghardware itemsandsimplifyingtheoperationofthese
analyzerstheEC9800BseriesprovidesoutstandingperformanceinlowcostbuthighperformanceOEMtypeanalyzers.
Similarly, the choice of calibration techniques (e.g. external zero/span valves, internal zero/span check) can be
optimized for thesystemallowingmore flexibility for the sys
temintegrator.
FEATURESINCLUDED
ON
ALL
B
SERIES
ANALYZERS
:
Selectable ppm, ppb, mg/m3 or ug/m3 corrected for
temperatureandpressurevariations.
Gaspressure, temperaturemeasurement and compensa
tion for accurate concentrationmeasurements under all
conditions.
Sampleflowratesareestablishedbybuiltincriticalorifice
andcontrolledbyanexternalsamplepump.
Softwarealgorithmshandleall internaladjustments,performdiagnostics, indicateerrors,displaystatus, linearize
andmakecalculationsofconcentration.Theonlyoperatorfunctionsareroutinemaintenanceofpneumaticsand
periodiccalibrationoftheunit.
Kalmanadaptive
filter
for
continuously
optimized
noise
and
response
time.
Menu driven large alphanumeric display provides a complete display of gasmeasurements, instrument status,
setupparametersanddatatrendingdisplay.
Eventlogforautomaticrecordingofsignificantmaintenance,calibration,anddiagnosticevents.
Allanalyzertemperaturesdisplayedonasinglemenuscreen.
EC9800analyzersfeatureinternalFlashROMdatastorageforinexcessof135daysof5minuteaverages.Datacan
bedownloadedusingtheanalyzersRS232portorremotelyusingamodem.Fullremotecontrolofeachanalyzer
overmultidropRS232serialportwithaPC.FirmwareeasilyupdatedutilizingUSBorRS232port.
Internal lithiumbatterybackupofconfigurationparametersensuresthatnormaloperation isrestoredevenafter
apowerinterruption.
Folddownfrontpanelprovideseasyaccesstoexpendableitemssuchasparticulatefiltersandcharcoalscrubbers,
andpower
control
switches.
Commonchassis,electronics,andsoftware.Slideoffchassiscoverprovideseasyaccesstoallinternalconstruction.
Optionalmanualoperatorcontrolledorautomaticinternalzero/spancheck.
Optionalexternalzerospanvalves.
Lowcarbonfootprintandutilizes12VDCinternalelectronics.
USEPAEquivalentreferencemethodcertification.
Special versionsof theseanalyzersareavailable formeasuring carbondioxide,background traceprecursor levelsof
carbonmonoxide,nitrogenoxides and sulfurdioxide,NOy andammonia,high concentrationnitrogenoxides, sulfur
dioxidesandcarbonmonoxide.
EECOTECHCOTECHEC9800BSEC9800BSERIESERIESGGASASAANALYZERSNALYZERS
-
7/31/2019 American Ecotech Series B Analyzers.sflb
2/3
OZONEANALYZEREC9810BTheEC9810BOzoneAnalyzercombinesmicroprocessorcontrolwith
ultraviolet(UV)
photometry
to
measure
O3
in
the
range
of
050
ppb
or020ppmwithadetectionlimitof0.5ppb.
TheEC9810BusestheprinciplethatcertainwavelengthsofUV light
areabsorbedbyozone(BeerLambertRelationship).Amercuryvapor
lampisusedtoprojectabeamof215nmlightthroughaglasssam
plecellthatalternatelycontainsthemeasurementsampleandthena
reference sample that is selectively scrubbed of ozone. The differ
ence inUV lightattenuationbetweenthesampleand thereference
gas streamsprovidesanaccurateand reliablemeasurementof the
ozoneconcentration.
FEATURES: Ozone measurement with photometric dualchannel technique
providesreliable,nearlydriftfreemeasurement.
Theuseofasmallglassmeasurementcellboostsdataqualityand
performance.
Withonlyonecell,lowflowandnomirrors,cleaningfrequencyis
lessthan thatneeded forothersystems,whileexpendables life
timeisextended.
Photometer allows for manual operator controlled calibration
zero/span sequence or a timed 24hour cycle at an operator
specifiedtimeorremotelycontrolledcalibrationbyexternalcon
tactclosure. Fullfivepointcalibration.
CARBONMONOXIDEANALYZEREC9830BTheEC9830BCarbonMonoxideAnalyzerutilizesNDIRGasFilterCorrelationphotometryandmicroprocessorcontrol tomeasureCO in the
rangeof0200ppmwithadetectionlimitof50ppb.
The EC9830 rotating gas filter correlation wheel has two gas filled
cells.OnecontainsCOwhichremovesallIRenergyintheCOabsorp
tionwavelengths thereby creating a reference beam. The other cell
containsnitrogenwhichallowstheCOabsorptionspectralbandstobe
measuredwiththesamplegas.ThedifferenceinIRenergysensedby
the solid state detector duringmeasurement and reference cycle is
proportionaltotheCOpresentinthemeasurementcell.Thisratiomet
ric techniqueeliminatesmost sourcesofdrift (i.e. interference from
CO2andH2O)typicalofotherlesssophisticatedinfraredanalyzers.
TheEC9830B
is
typically
used
with
external
calibration
gases
for
auto
maticzero/spanchecks.Highpressurespanvalvesareavailablewhen
usingmoderatelypressurizedgassources.
FEATURES: Long life infrared(IR)sourceandadvancedgasfilterwheelallows
fortroublefreeoperationandminimummaintenance.
Builtinzeroairscrubberensureszerostabilitywithnoneedforan
externalscrubber.
Temperaturestabilizedfilterandgasfilterwheelensureminimum
spandrift.
Customdesignednafiondryervirtuallyeliminatesinferenceof
watervapororcarbondioxideduringmeasurement.
NITROGENOXIDESANALYZEREC9841BTheEC9841BNitrogenOxidesAnalyzerutilizesmicroprocessor c
troland
chemiluminescence
detection
to
measure
NO,
NO2
and
N
intherangesof050ppband020ppmwithadetectionlimitof
ppb.
TheEC9841Bprincipleofmeasurementisbaseduponthereaction
theNOmoleculewithan internal sourceofozone inanevacua
reaction cell that results in theemissionof light.The EC9841B
singlechannelinstrumentthatmeasurestheNOinasamplegast
isalternatelypassedthroughoraroundacatalyticconvertertoc
verttheNO2toNO.Themeasurementoftheuntreatedsamplep
vides a NO value and themeasurement of the converted sam
provides theNOxmeasurementwithNO2 calculated as the dif
encebetweenthetwomeasurements.
FEATURES:
Fastresponseandlownoiseachievedthroughauniquelowp
surereactioncelldesign.
Autozero routine allows the analyzer to periodically check
correct for background illumination, virtually eliminating z
drift.
Low sample flow rate extends molybdenum converter
320ml/min.
Theuseofelectropolishedstainlesssteelreactioncellresult
decreasedmaintenanceandgreatlyimprovedlongevity.
SULPHURDIOXIDESANALYZEREC9850BThe EC9850B Sulphur Dioxide Analyzer combines microproces
controlwithfluorescencedetectiontomeasureSO2intherange
050ppband020ppmwithadetectionlimitof0.5ppb.
TheEC9850dependson the fluorescent radiationproducedbyS
moleculeswhenexcitedbyUV radiation.TheexcitationU.V. ra
tionismeasuredbyareferencedetectorandthefluorescencera
tion (SO2) ismeasured by a photomultiplier tube (PMT). The t
measurements are ratioed in a classical dualchannel technique
minimizetheeffectsofvariationinlampintensity,opticalcontam
tionandcommonPMTdriftcharacteristics.Ahydrocarbonscrubb
system,containingnoconsumablematerials,also removes inter
inghydrocarbonspriortotheambientsamplebeingmeasured.
FEATURES: AcustomdesignedzincUVsourceproducesasinglespectral
at214nmallowingtheanalyzertobefreefromwatervapor
terference,eveninveryhighhumidityconditions.
Optionalinternalzero/spancheckdevicewhichutilizesaconst
temperaturepermeationtubetogenerateaknownSO2span
concentration.
-