alignment methods developed for the validation of the …at 401 measurements • nobody allowed in...

18
Alignment methods developed for the validation of the thermal and mechanical behavior of the Two Beam Test Modules for the CLIC project Hélène MAINAUD DURAND, on behalf of the CLIC pre-alignment team

Upload: others

Post on 08-Sep-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Alignment met hods developed f or t he val idat ion of t he t hermal and mechanical behavior of t he Two Beam Test Modules f or t he CLIC project Hélène MAINAUD DURAND, on behalf of the CLIC pre-alignment team

Page 2: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Agenda

• Int roduct ion • Short range st rategy of pre-alignment on module components • Preparat ion of the thermo-mechanical tests • Methods of measurements developed • Results obtained

Out line

Thermal & mechanical behavior of TBTM: alignment methods developed & misalignments observed

Page 3: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Int roduct ion CLIC project

Page 4: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Int roduct ion CLIC module

Components:

Quadrupole

PETS

Acce le rat ing St ructure (AS)

Modules = key component of CLIC • Length = 2 m • 20 000 modules

Page 5: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Int roduct ion CLIC module

2 Two Beam Test Modules (TBTM) under installat ion:

• Validat ion of the assembly & integrat ion of the components • Validat ion of the short range st rategy of pre-alignment • A bet t er underst anding of t hermal and mechanical behavior of a

module under dif f erent operat ing modes

Misalignment s observed

Page 6: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Short range st rategy of pre-alignment St rategy

Pre-alignment of several components on 1 support

2 cradles per girders “snake conf igurat ion”

cWPS & actuators coupled to each cradle

Page 7: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Short range st rategy of pre-alignment Process of pre-alignment

Fiducial isat ion

Pre-al ignment on girder

Inst al lat ion in t unnel

Page 8: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Preparat ion of thermo-mechanical test s Context of measurements

Tunnel environment reproduced by air condit ioning & vent ilat ion system:

• T© f rom 20 ©C to 40 ©C • Longitudinal air speed: 0.3 to

0.8m/s

All component s equipped with e lect ric heat ers t o reproduce the power dissipat ion est imated in t he module

Cooling syst em int egrat ed in RF component s (not in t he DB quadrupole )

1st ope rat ing scenario simulat ed: - Zero posit ion - Power-up of DB quadrupole - Unloaded condit ions (RF but no beam) - Loaded condit ions (RF + beam)

Finit e Element Analysis prepared with ANSYS to simulate displacement s (but ready during the measurement s process)

Page 9: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Posit ion of t he fiducials w.r.t V-shaped support axis

Posit ion of t he fiducials w.r.t mechanical axis of t he component

Measurement s using the lase r t racker AT 401

Preparat ion of thermo-mechanical tests

Page 10: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Preparat ion of thermo-mechanical test s Zero posit ion of the components

Page 11: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Methods of measurements Means of measurements

cWPS coupled to the cradle

Permanent monitoring of t he posit ion of t he girde r (mean axis of V-shaped support ), within a few µm

AT 401 measurement s

Posit ion of t he component s on the girde r

Page 12: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Methods of measurements Means of measurements

AT 401 measurements

• Nobody allowed in the room during the measurements

• Use of a heavy t ripod for a bet ter stabilit y • Durat ion of a stat ion < 45’ • Warm-up of AT401 4 hours before the

f irst measurements • All measurements performed back side

and f ront side, integrated over 5’’. • Fiducials located on the girder considered

as references during best -f it s (as measured by CMM)

Page 13: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Results : misalignments observed Misalignments of components

How to understand the data?

init ial posit ion of t he mean axis of girder

current posit ion of t he mean axis of girder

current posit ion of t he mean axis of component w.r.t mean axis of girder

Page 14: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Results : misalignments observed Misalignments of components

During the DBQ heat ing: vert ical displacements of ~40μm are obse rved, as no cooling system is associat ed with t he DBQ (for a T© of 35 ©C) The unloaded cycle has been pe rformed twice : t his gives a good idea of t he repeat abilit y, be t t e r t han 10 μm and confirms the precision of measurement s

Page 15: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Results : misalignments observed Misalignments of components

No radial displacements visible, which seems logical: a T© increase should not misalign radially the reference axis of the component as it is symmet ric. Very good repeatabilit y: this conf irms the precision of measurements

Page 16: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Results : misalignments observed Misalignments of components

Misalignments during the f irst phase are very small, which is logical as there is no heat ing on the MB side. The cooling system plays it s role during the unloaded and loaded phases: misalignments < 20 μm

Page 17: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Results : misalignments observed Misalignments of components

No radial misalignments observed as the component is symmet ric.

Page 18: Alignment methods developed for the validation of the …AT 401 measurements • Nobody allowed in the room during the measurements • Use of a heavy tripod for a better stability

Summary

• Heaters added to the components, cooling system to RF components

• T©, cWPS, AT401 measurements performed

• FEM carried out

• Have a be t t e r knowledge of t he rmal expansion of t he cradle (where sensors are locat ed) w.r.t girde r

• Increase t he number of st at ions pe r side • Use simult aneously 4 AT 401 measuring the posit ion

of t he component s • To equip all t he component s with fiducials • To automat ize t he process of measurement s

One first duty cycle (machine st art -up) simulat ed

Improvement of t he methods of measurement s needed

• Very small impact of such cycle on the alignment of component s t hat are cooled down

• Good mat ching be tween simulat ions and measurement s

• At the limit of t he uncert ainty of measurement s of AT401

Would anybody have an experience in t hat ?