ColourVision BodyGuard TM Pain Management Pump Programmed Intermient Bolus (PIB) User Reference Guide Refer to Operang Manual for full operang instrucons. Users must have undertaken training before operang this device. Screen informaon/sequences may vary with different soſtware versions and local pump configuraon. Always follow screen prompts and before pressing keys to proceed, ensure selecons made correspond with what is required. Doc Ref: QUG26/CV545 & 575 PIB.QUGv1/Dec2015 With the keypad lock acvated, the user can STOP and START an infusion, and with the infusion running use the INFO key to review infusion status. Users cannot (for example) power off, rate change, deliver a clinician acvated bolus or go to the main menu with the keypad lock acvated. Keypad Lock With the infusion/protocol running, press the INFO key repeatedly to display: Press 1 - Shiſt totals: elapsed and operaon me, boluses aempted/delivered, bolus volume and total volume. Press 2 - Current infusion: volume remaining, volume infused, mL/hour rate and total infused. Press 3 - Bolus summary: paent aempted/delivered. Bolus dose, doses given, lockout me. Clinician bolus. Press 4 - Protocol informaon (#1): protocol tle, bag volume, current rate and maximum rate. Press 5 - Protocol informaon (#2): protocol tle, paent bolus, maximum bolus, lockout me and hour limit. Press 6 - Baery level: graph format of baery status (%). Press 7 - Time and date: event log me and date. Press 8 - Pump idenficaon: pump name, serial number, pump producon date and soſtware version. Press 9 - Bolus Remaining: difference between bolus mls delivered from requested. Accessing Protocol/Infusion Summary Notes When an ALERT is acvated: The infusion connues, LED indicator light remains green, intermient beeps are heard approx. every 3/4 minutes and a screen message alternates with the infusion running screen. When an ALARM is acvated: The infusion stops, the LED indicator light turns red, a connuous audible alarm acvates and a screen message displays indicang the cause. Press STOP/NO to mute alarms. Follow screen prompts to manage alerts and alarms. Alert, Alarms and Troubleshoong Low Baery Connect to Charger Alert: Approximately 30 minutes of baery life remaining. Connect to mains power. End Baery Connect to Charger Baery power is depleted. Connect to mains power immediately. Place pump in charger or use the baery cable to connect to the DC socket. Down Occlusion Check set/access The set or access device is occluded (below pump). Restart Pump Switch off & on Error xx A technical/Internal malfuncon has occurred. Follow screen prompts. If the problem cannot be recfied, remove from paent use. Near End Alert: Bag nearly empty, approximately 5mL volume remaining. Prepare new bag, if applicable. End of Infusion Current infusion has completed. Either disconnue use or change bag to start a new protocol. Pump Paused Too Long Pump has been leſt in the STOP state for more than 2 minutes with no key presses. Press START/OK to resume, press STOP/NO to connue pause or press and hold down STOP/NO for menu. Paent Bolus available in: 00:05 Auto Bolus available in: 00:30 Screen prompt: When an Auto bolus is being delivered, the pump beeps intermiently and a screen message displays. Air/Up Occlusion Check for air/occlusion Air in set or the set is occluded between the infusion bag and pump. Bolus Handset Disconnected. Paent Bolus Unavailable Alert: Bolus handset disconnecon. Re-connect handset to pump. Missing Set Key Load Set & Close Door The set is loaded incorrectly or a non-proprietary set is loaded. Door Open Close Door The pump door has opened during infusion. Check set is loaded correctly, close door and re-start infusion. Tel: 01253 206700 | Fax: 01253 896648 | Email: [email protected] | cmemedical.co.uk Screen Descripon

Upload: others

Post on 22-Jun-2020

9 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Alert, Alarms and Troubleshooting ColourVision ......Doc Ref: QUG26/CV545 & 575 PIB.QUGv1/Dec2015 what is required. With the keypad lock activated, the user can STOP and START an infusion,

ColourVision BodyGuardTM

Pain Management PumpProgrammed Intermittent Bolus (PIB) User Reference Guide

• Refer to Operating Manual for full operating instructions.• Users must have undertaken training before operating this device.• Screen information/sequences may vary with different software versions and local pump configuration.• Always follow screen prompts and before pressing keys to proceed, ensure selections made correspond with

what is required.Doc

Ref

: QU

G26

/CV

545

& 5

75 P

IB.Q

UG

v1/D

ec20

15

With the keypad lock activated, the user can STOP and START an infusion, and with the infusion running use the INFO key to review infusion status.Users cannot (for example) power off, rate change, deliver a clinician activated bolus or go to the main menu with the keypad lock activated.

Keypad Lock

With the infusion/protocol running, press the INFO key repeatedly to display:Press 1 - Shift totals: elapsed and operation time, boluses attempted/delivered, bolus volume and total volume.Press 2 - Current infusion: volume remaining, volume infused, mL/hour rate and total infused.Press 3 - Bolus summary: patient attempted/delivered. Bolus dose, doses given, lockout time. Clinician bolus.Press 4 - Protocol information (#1): protocol title, bag volume, current rate and maximum rate.Press 5 - Protocol information (#2): protocol title, patient bolus, maximum bolus, lockout time and hour limit. Press 6 - Battery level: graph format of battery status (%).Press 7 - Time and date: event log time and date.Press 8 - Pump identification: pump name, serial number, pump production date and software version.Press 9 - Bolus Remaining: difference between bolus mls delivered from requested.

Accessing Protocol/Infusion Summary

Notes

When an ALERT is activated:The infusion continues, LED indicator light remains green, intermittent beeps are heard approx. every 3/4 minutes and a screen message alternates with the infusion running screen.When an ALARM is activated:The infusion stops, the LED indicator light turns red, a continuous audible alarm activates and a screen message displays indicating the cause.

Press STOP/NO to mute alarms. Follow screen prompts to manage alerts and alarms.

Alert, Alarms and Troubleshooting

Low BatteryConnect to Charger

Alert: Approximately 30 minutes of battery life remaining.

Connect to mains power.

End BatteryConnect to Charger

Battery power is depleted.

Connect to mains power immediately. Place pump in charger or use the battery cable to connect to the DC socket.

Down OcclusionCheck set/access The set or access device is occluded (below pump).

Restart PumpSwitch off & onError xx

A technical/Internal malfunction has occurred.

Follow screen prompts. If the problem cannot be rectified, remove from patient use.

Near EndAlert: Bag nearly empty, approximately 5mL volume remaining.

Prepare new bag, if applicable.

End of InfusionCurrent infusion has completed.

Either discontinue use or change bag to start a new protocol.

Pump Paused Too Long

Pump has been left in the STOP state for more than 2 minutes with no key presses.

Press START/OK to resume, press STOP/NO to continue pause or press and hold down STOP/NO for menu.

Patient Bolus available in: 00:05Auto Bolus available in: 00:30

Screen prompt: When an Auto bolus is being delivered, the pump beeps intermittently and a screen message displays.

Air/Up OcclusionCheck for air/occlusion Air in set or the set is occluded between the infusion bag and pump.

Bolus Handset

Disconnected. Patient Bolus Unavailable

Alert: Bolus handset disconnection.

Re-connect handset to pump.

Missing Set KeyLoad Set & Close Door The set is loaded incorrectly or a non-proprietary set is loaded.

Door OpenClose Door

The pump door has opened during infusion.

Check set is loaded correctly, close door and re-start infusion.

Tel: 01253 206700 | Fax: 01253 896648 | Email: [email protected] | cmemedical.co.uk

Screen Description

Page 2: Alert, Alarms and Troubleshooting ColourVision ......Doc Ref: QUG26/CV545 & 575 PIB.QUGv1/Dec2015 what is required. With the keypad lock activated, the user can STOP and START an infusion,

Remember to de-acti vate and acti vate keypad lock if necessary.1. With the infusion in progress, enter the new rate using the numerical keypad, press START/OK.2. Enter the relevant access code using the numerical keypad, press START/YES.3. Check that the new basal mL/hour rate displays on the infusion running screen.

1. Prepare bag and set as per local policy. Load administrati on set into the pump.2. To power on press and hold down ON/OFF key, pump setti ngs display then the screen displays:

3. To go to main menu,

press STOP/NO.

Enter code,

press START/OK.

Ensure the set is disconnected, press

START/OK.

On completi on, the next screen displays.

6. Enter a numerical ID number (opti onal),

press START/OK.

7. Select protocol (use scroll keys if necessary),

press START/OK.

4. Select prime,

press START/OK.

Enter code,

press START/OK.

5. To select protocol,

press START/OK.

Enter code,

press START/OK.

8. To confi rm selecti on of auto bolus protocol,

press PRIME/BOLUS.

10. When INFO is pressed, the user can change setti ngs:

• Use numerical keypad to select the opti on required, press START/OK then follow screen prompts.

• When ready, press STOP/NO to return to the review protocol setti ngs screen.

12. Select opti on required and follow screen prompts.• Press START/OK to give auto bolus and commence protocol.Press INFO to delay auto-bolus, follow screen prompts to commence protocol.• Press PRIME/BOLUS to deliver a loading dose (CAB), follow screen prompts,

enter access code to commence protocol.

11. Att ach the administrati on set to the correct pati ent access port (unclamp if necessary).

9. Review protocol setti ngs.To start protocol press START/OK.To change, press INFO.

Remember to de-acti vate and acti vate keypad lock if necessary.1. Press PRIME/BOLUS.2. Select either “Change Bolus Dose” or “Change Lockout Time”, press START/OK.3. Enter values required using the numerical keypad, press START/OK.4. Enter relevant code using the numerical keypad, press START/OK.5. Check change complete on infusion running screen.

Remember to de-acti vate and acti vate keypad lock if necessary.Ensure set is clamped/disconnected from pati ent`s access device.• If end infusion alarm has acti vated the keypad lock is automati cally removed. To confi rm end of infusion and

mute alarm, press STOP/NO.To record VTBI and VI, press INFO.• Follow screen prompts to return to main menu, complete bag change, re-connect/unclamp set and resume

the protocol.Note: When a protocol is resumed and the protocol summary displays, the new VTBI and the total of all volume infused to date out of previous bag(s) will display.

Remember to de-acti vate and acti vate keypad lock if necessary.1. To stop the infusion, press STOP/NO.2. Press PRIME/BOLUS.3. Enter Clinician Bolus access code using the numerical keypad, press START/OK.4. Follow on-screen prompt to enter the bolus dose required and to commence delivery, press START/OK Press

STOP/NO at any ti me to stop delivery.On completi on of Clinician Bolus delivery, the infusion recommences.

Basal Rate Change During Delivery (Rate Titrati on)

Infusion Screen Running

Changing Bolus Dose and/or Lock Out Time

Bag Change

Clinician Bolus (Designated Users Only)

Start a New Infusion

Actual pressureThe bar displays real ti me in-line pressure. Units of measure, start value, and peak value are subject to the pump pressure setti ngs. The bar is divided to 3 parts. Actual pressure is displayed as red fi ll. When the pressure rises from one segment to the next a beep will sound. If the pressure reaches the end value (1100 mmHg in the above example) the pressure alarm will acti vate.

Title bar - type of protocol Protocol lett er and name

Protocol rate

Protocol parameters (example shows bolus and lockout ti mes for pati ent and PIB bolus)

Title bar - ti me, batt ery status

Actual pressure bar