advanced multi-purpose calibrator - cert-trak · russian gost-180°c -50°c 0.18°c 0.19°c 0.23°c...

12
s ignal ISO 9001 Manufacturer SS-CP-2380-US Specification Sheet Laboratory standard Laboratory reference level bench top instrument for electrical signals, handheld calibrators, and testing of instrumentation Output: DC current 0-100 mA DC voltage 0 to 100 V DC mV 0 to 100 mV TC simulation -10 to 75 mV Resistance 5 to 4000 ohm (RTD simulation) Input: DC mV -10 to 75 mV (TC temperature conversion) Resistance 0 to 4000 ohm (RTD temperature conversion) Pressure modules Pressure reference capability increases the flexibility of the instrument with the addition of external pressure modules. Accuracies up to 0.01% Temperature reference 13 different Thermocouple types, 14 different RTD types Accuracies of up to 0.025°C Fast RTD simulation Works with all pulsed transmitters Fully remote controllable With a parallel instrument bus IEEE-488 and RS232 serial interface The JOFRA AMC900 not only offers temperature and pressure calibration features, it includes sourcing of current and voltage and measurement of milli-volts, and resistance. This flexibility means that you need only one unit for calibration or verification. This standard gives you the confidence you need for your test instruments as well as in your field calibrations. Upgrade your laboratory or workshop to a higher level of accuracy and flexibility with the AMC900. The JOFRA AMC900 has the ability to store up to 9 setpoints for each output range. The setpoints can be selected manually or automatically stepped at timed intervals. The AMC900 also features an easy-to-operate front panel user interface and computer control through an RS232 or IEEE-488 interface for automated production testing. The unit sources DC voltage and current for multifunction workload coverage, enabling calibration of data loggers, strip chart recorders, multi-meters, handheld calibrators, and other industrial instruments. This is the cost-effective solution for your multiple calibration needs. The AMC900 bench calibrator is your solution for high precision signal, temperature, and pressure calibrations. This multi-purpose unit offers laboratory grade accuracy at an industrial calibrator price. The standard is delivered pre-loaded with 13 thermocouple curves and 14 RTD curves. You can also program curves to suit your individual needs. With the optional exter- nal pressure modules, you can tailor the system to address your calibration needs. PRODUCT DESCRIPTION AMC900 Advanced Multi-purpose Calibrator Multi-purpose calibrator offering laboratory features and accuracies.

Upload: others

Post on 08-Jul-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

s i g n a l

ISO 9001 Manufacturer

SS-CP-2380-US

Specification Sheet

Laboratory standard Laboratory reference level bench top instrument for electrical signals, handheld calibrators, and testing of instrumentation

Output: DC current 0-100 mA DC voltage 0 to 100 V DC mV 0 to 100 mV TC simulation -10 to 75 mV Resistance 5 to 4000 ohm (RTD simulation)

Input: DC mV -10 to 75 mV (TC temperature conversion)Resistance 0 to 4000 ohm (RTD temperature conversion)Pressure modules

Pressure reference capabilityincreases the fl exibility of the instrument with the addition of external pressure modules. Accuracies up to 0.01%

Temperature reference13 different Thermocouple types,14 different RTD typesAccuracies of up to 0.025°C

Fast RTD simulationWorks with all pulsed transmitters

Fully remote controllable With a parallel instrument bus IEEE-488 and RS232 serial interface

The JOFRA AMC900 not only offers temperature and pressure calibration features, it includes sourcing of current and voltage and measurement of milli-volts, and resistance. This fl exibility means that you need only one unit for calibration or verifi cation. This standard gives you the confi dence you need for your test instruments as well as in your fi eld calibrations. Upgrade your laboratory or workshop to a higher level of accuracy and fl exibility with the AMC900.The JOFRA AMC900 has the ability to store up to 9 setpoints for each output range. The setpoints can be selected manually or automatically stepped at timed intervals. The AMC900 also features an easy-to-operate front panel user interface and computer control through an RS232 or IEEE-488 interface for automated production testing. The unit sources DC voltage and current for multifunction workload coverage, enabling calibration of data loggers, strip chart recorders, multi-meters, handheld calibrators, and other industrial instruments. This is the cost-effective solution for your multiple calibration needs.

The AMC900 bench calibrator is your solution for high precision signal, temperature, and pressure calibrations. This multi-purpose unit offers laboratory grade accuracy at an industrial calibrator price. The standard is delivered pre-loaded with 13 thermocouple curves and 14 RTD curves. You can also program curves to suit your individual needs. With the optional exter-nal pressure modules, you can tailor the system to address your calibration needs.

PRODUCT DESCRIPTION

AMC900Advanced

Multi-purpose

Calibrator

Multi-purpose calibrator offering laboratory features and accuracies.

SS-CP-2380-US

Specification SheetSS-CP-2380-US

Specification Sheet

New calibration lab with one unit

The AMC900 is the single unit that can make the greatest change to your instrument shop or calibration lab. This calibrator offers the ability to perform calibrations and verifi cations of signal, temperature, and pressure instrumentation with one device. You can source current and source/read voltage and resistance. Simulate virtually any temperature sensor made today. Combined with a good digital multimeter and you have a complete electrical process calibration lab.

Saving you time

The AMC900 is fl exible. If you need to store test points for particular calibrations, you can save up to 9 setpoints along with dedicated dwell times to allow for easy repeat calibrations. AMC900 has also installed many of the curves for today’s wide range of temperature sensors so that you can spend time calibrating instruments and not programming the calibrator or referencing look up tables. Your pressure calibration needs are met with a vast array of modules that communicate directly with the AMC900. Additionally, you have the option of using an RS232 or IEEE-488 interface to easily automate your calibrations. Just use HyperTerminal, Visual Basic, or any other software using an ASCII interface.

AMC900 as your temperature reference

Use the AMC900 to maintain a high level of confi dence in all your handheld temperature indicators. Check the electrical side against the AMC900 and the probes in a JOFRA dry-block calibrator - using the AMC900 as the external temperature reference.

The AMC900 is a versatile performer. In addition to the 14 RTD curves and 13 T/C curves are factory-installed on the unit, you may have an occasional need to add a custom curve for a specific sensor. The user interface allows you to add the constants to calibrate these sensors.

Save money by maintaining all your temperature calibration equipment yourself. The AMC900 can be a standard for your dry-block calibrators. The unit is confi gured to accept reference probe inputs for temperature verifi cation. (Compatible with all

JOFRA STS100 sensors, see more at page 9). You can calibrate any of our dry-block calibrators using a reference probe and the AMC900. As a temperature reference you can also verify temperature generation devices such as circulating baths, other dry-block calibrators, and ovens.

In the lab or in the shop

The AMC900 is designed to meet calibration needs in a laboratory environment or as a reference standard for instrument shops. The adaptability of this calibrator allows it

to be utilized in a lab to be the standard for signal, temperature, and pressure calibrations. Where you are responsible for the operation of a wide range of equipment, such as in an instrument shop, you can bench- test instruments before installation and you can save money by maintaining the traceability and performance of all your test instruments in operating order with this effi cient calibrator.

AMC900 as pressure reference

Use your current pressure source, add a JOFRA APM pressure module to the JOFRA AMC900 calibrator and start your pressure calibration (see more at page 10 + 11).Each module communicates directly with the calibrator to ensure proper information transfer and accurate calibrations. The AMC900 operates as a pressure standard with a wide range of pressure modules covering pressure ranges from vacuum to 10,000 psi / 700 bar. Modules are available in gauge, differential, absolute, and vacuum ranges. Pressure can be displayed in a wide range of engineering units with up to 0.01% full scale accuracy.

The modules are designed so that they plug into the front panel and c o m m u n i c a t e directly with the AMC900 and the handheld mu l t i f un c t i o n calibrator JOFRA ASC300.

JOFRA AMC900 APPLICATIONS

CALIB

RATIO

N INSTRUM

ENTS

MENUESC

AUTOSTEP

SWITCH

TEST

- 2 - - 3 -

SS-CP-2380-US

Specification SheetSS-CP-2380-US

Specification Sheet

Resistance INPUTAccuracy

OUTPUTAccuracy

Resolution Allowable current 1 90 days 12 months 90 days 12 months

5 to 400 ohm 0.01 ohm 1 to 10 mA ±0.025 ohm ±0.03 ohm

5 to 4000 ohm 0.1 ohm 250 µA to 1 mA ±0.25 ohm ±0.3 ohm

0 to 400 ohm 0.001 ohm 1 mA ±35 ppm + 0.003 ohm

±40 ppm + 0.003 ohm

401 to 4001 ohm 0.01 ohm 0.1 mA ±35 ppm + 0.03 ohm

±40 ppm + 0.03 ohm

Absolute uncertainty, tcal ±5°C / 9°F 1. Continuously variable from 0 to 4 Kohm (Output). Current is fixed to 1 and 0.1 mA in input

RTD and thermistor

TEMPERATURErange

INPUTaccuracy ±

OUTPUT 1

accuracy ±

from °C / °F

to °C / °F

90 days°C / °F

12 months°C / °F

90 days°C / °F

12 months°C / °F

Pt100alpha 3916

-200°C -190°C 0.006°C 0.006°C 0.06°C 0.07°C

-190°C -80°C 0.011°C 0.012°C 0.06°C 0.08°C

-80°C 0°C 0.014°C 0.015°C 0.06°C 0.08°C

0°C 100°C 0.018°C 0.019°C 0.06°C 0.08°C

100°C 260°C 0.025°C 0.028°C 0.07°C 0.08°C

260°C 300°C 0.026°C 0.029°C 0.07°C 0.08°C

300°C 400°C 0.031°C 0.034°C 0.07°C 0.09°C

400°C 600°C 0.040°C 0.045°C 0.08°C 0.09°C

600°C 630°C 0.042°C 0.047°C 0.08°C 0.09°C

-328°F -310°F 0.011°F 0.011°F 0.11°F 0.13°F

-310°F -112°F 0.020°F 0.022°F 0.11°F 0.14°F

-112°F 32°F 0.025°F 0.027°F 0.11°F 0.14°F

32°F 212°F 0.032°F 0.034°F 0.11°F 0.14°F

212°F 500°F 00.045°F 0.050°F 0.13°F 0.14°F

500°F 572°F 0.047°F 0.052°F 0.13°F 0.14°F

572°F 752°F 0.056°F 0.061°F 0.13°F 0.16°F

752°F 1112°F 0.072°F 0.081°F 0.14°F 0.16°F

1112°F 1166°F 0.076°F 0.085°F 0.14°F 0.16°F

Cu100Russian Gost

-180°C -50°C 0.18°C 0.19°C 0.23°C 0.24°C

-50°C 200°C 0.14°C 0.15°C 0.18°C 0.20°C

-292°F -58°F 0.32°F 0.34°F 0.41°F 0.43°F

-58°F 392°F 0.25°F 0.27°F 0.32°F 0.36°F

Pt200alpha 385

-200°C -80°C 0.031°C 0.032°C 0.31°C 0.38°C

-80°C -0°C 0.035°C 0.036°C 0.32°C 0.38°C

0°C 100°C 0.039°C 0.041°C 0.33°C 0.39°C

100°C 260°C 0.042°C 0.045°C 0.33°C 0.39°C

260°C 300°C 0.050°C 0.053°C 0.36°C 0.43°C

300°C 400°C 0.053°C 0.057°C 0.36°C 0.43°C

400°C 600°C 0.070°C 0.075°C 0.42°C 0.50°C

600°C 630°C 0.071°C 0.076°C 0.42°C 0.50°C

-328°F -112°F 0.056°F 0.058°F 0.56°F 0.68°F

-112°F 32°F 0.063°F 0.065°F 0.58°F 0.68°F

32°F 212°F 0.070°F 0.074°F 0.59°F 0.70°F

212°F 500°F 0.076°F 0.081°F 0.59°F 0.70°F

500°F 572°F 0.090°F 0.095°F 0.65°F 0.77°F

572°F 752°F 0.095°F 0.103°F 0.65°F 0.77°F

752°F 1112°F 0.126°F 0.135°F 0.76°F 0.90°F

1112°F 1166°F 0.128°F 0.137°F 0.76°F 0.90°F

Absolute uncertainty, tcal ±5°C / 9°F 1. 2-wire output 2. Based on MINCO application aid no. 18.

°F

°C

°F°C

°F

°C

- 2 - - 3 -

SS-CP-2380-US

Specification SheetSS-CP-2380-US

Specification Sheet

RTD and Thermistor

TEMPERATURErange

INPUTaccuracy ±

OUTPUT 1

accuracy ±

from to 90 days 12 months 90 days 12 months

Pt500alpha 385

-200°C -80°C 0.016°C 0.017°C 0.13°C 0.15°C

-80°C -0°C 0.019°C 0.020°C 0.13°C 0.15°C

0°C 100°C 0.023°C 0.025°C 0.13°C 0.16°C

100°C 260°C 0.030°C 0.033°C 0.14°C 0.17°C

260°C 300°C 0.032°C 0.035°C 0.14°C 0.17°C

300°C 400°C 0.037°C 0.041°C 0.15°C 0.18°C

400°C 600°C 0.047°C 0.052°C 0.16°C 0.19°C

600°C 630°C 0.048°C 0.076°C 0.16°C 0.19°C

-328°F -112°F 0.029°F 0.031°F 0.23°F 0.27°F

-112°F 32°F 0.034°F 0.036°F 0.23°F 0.27°F

32°F 212°F 0.041°F 0.045°F 0.23°F 0.29°F

212°F 500°F 0.054°F 0.059°F 0.25°F 0.31°F

500°F 572°F 0.058°F 0.063°F 0.25°F 0.31°F

572°F 752°F 0.067°F 0.074°F 0.27°F 0.32°F

752°F 1112°F 0.085°F 0.094°F 0.29°F 0.34°F

1112°F 1166°F 0.086°F 0.137°F 0.29°F 0.34°F

Pt10alpha 385

-200°C -0°C 0.11°C 0.11°C 0.75°C 0.90°C

-0°C 400°C 0.17°C 0.18°C 0.75°C 0.90°C

400°C 800°C 0.17°C 0.18°C 0.85°C 1.00°C

-328°F 32°F 0.20°F 0.20°F 1.35°F 1.62°F

32°F 752°F 0.31°F 0.32°F 1.35°F 1.62°F

752°F 1472°F 0.31°F 0.32°F 1.53°F 1.80°F

Pt50alpha 385

-200°C 400°C 0.045°C 0.047°C 0.15°C 0.18°C

400°C 800°C 0.065°C 0.071°C 0.17°C 0.20°C

-328°F 752°F 0.08°F 0.08°F 0.27°F 0.32°F

752°F 1472°F 0.12°F 0.13°F 0.31°F 0.36°F

Cu50Russian Gost

-180°C -50°C 0.19°C 0.20°C 0.3°C 0.31°C

-50°C 200°C 0.15°C 0.16°C 0.25°C 0.26°C

-292°F -58°F 0.34°F 0.36°F 0.54°F 0.56°F

-58°F 392°F 0.27°F 0.29°F 0.45°F 0.47°F

Pt100alpha 385

-200°C -80°C 0.011°C 0.012°C 0.06°C 0.07°C

-80°C 0°C 0.018°C 0.020°C 0.08°C 0.10°C

0°C 100°C 0.018°C 0.020°C 0.08°C 0.10°C

100°C 300°C 0.027°C 0.030°C 0.07°C 0.09°C

300°C 400°C 0.031°C 0.035°C 0.07°C 0.09°C

400°C 630°C 0.042°C 0.047°C 0.08°C 0.09°C

630°C 800°C 0.050°C 0.057°C 0.08°C 0.10°C

-328°F -112°F 0.023°F 0.022°F 0.11°F 0.13°F

-112°F 32°F 0.032°F 0.036°F 0.14°F 0.18°F

32°F 212°F 0.032°F 0.036°F 0.14°F 0.18°F

212°F 572°F 0.049°F 0.054°F 0.13°F 0.16°F

572°F 752°F 0.056°F 0.063°F 0.13°F 0.16°F

752°F 1166°F 0.076°F 0.085°F 0.14°F 0.16°F

1166°F 1472°F 0.090°F 0.103°F 0.14°F 0.18°F

Pt100alpha 3926

-200°C -80°C 0.011°C 0.011°C 0.06°C 0.07°C

-80°C 0°C 0.014°C 0.015°C 0.06°C 0.07°C

0°C 100°C 0.018°C 0.019°C 0.06°C 0.08°C

100°C 300°C 0.026°C 0.029°C 0.07°C 0.08°C

300°C 400°C 0.031°C 0.034°C 0.07°C 0.09°C

400°C 630°C 0.041°C 0.046°C 0.08°C 0.09°C

-328°F -112°F 0.020°F 0.020°F 0.11°F 0.13°F

-112°F 32°F 0.025°F 0.027°F 0.11°F 0.13°F

32°F 212°F 0.032°F 0.034°F 0.11°F 0.14°F

212°F 572°F 0.047°F 0.052°F 0.13°F 0.14°F

572°F 752°F 0.056°F 0.061°F 0.13°F 0.16°F

752°F 1166°F 0.074°F 0.083°F 0.14°F 0.16°FAbsolute uncertainty, tcal ±5°C / 9°F 1. 2-wire output 2. Based on MINCO application aid no. 18.

°C

°F

°C

°C

°C

°C

°C

°C

°F

°F

°F

°F

°F

- 4 - - 5 -

SS-CP-2380-US

Specification SheetSS-CP-2380-US

Specification Sheet

DC Voltage OUTPUT 1

Accuracy

Resolution Maximum burden 2

Stability - 24 hours±1°C / 1.8°F

90 days 12 months

0 to 100 mV 1 µV 10 mA ±5 ppm + 2 µV ±25 ppm + 3 µV ±30 ppm + 3 µV

0 to 1 V 10 µV 10 mA ±4 ppm + 20 µV ±25 ppm + 30 µV ±30 ppm + 20 µV

0 to 10 V 100 µV 10 mA ±4 ppm + 200 µV ±25 ppm + 200 µV ±30 ppm + 200 µV

0 to 100 V 1 mV 1 mA ±5 ppm + 1 mV ±25 ppm + 2 mV ±30 ppm + 2 mV

Thermocouple output

-10 to 75 mV 1 µV 10 mV ±5 ppm + 2 µV ±25 ppm + 3 µV ±30 ppm + 3 µV

Absolute Uncertainty, tcal ±5°C / 9°F 1. All outputs are positive only 2. Remote sensing is not provided. Output resistance is < 1 ohm Noise: 0 to 100 mV: 1 µV p-p from 0.1 to 10 Hz, 6 µV rms from 10 Hz to 10 KHzNoise: 0 to 1 V: 10 µV p-p from 0.1 to 10 Hz, 60 µV rms from 10 Hz to 10 KHzNoise: 0 to 10 V: 100 µV p-p from 0.1 to 10 Hz, 600 µV rms from 10 Hz to 10 KHzNoise: 0 to 100 V: 10 ppm + 1 mV p-p from 0.1 to 10 Hz, 20 mV rms from 10 Hz to 10 KHz

RTD and Thermistor

TEMPERATURErange

INPUTaccuracy ±

OUTPUT 1

accuracy ±

from to 90 days 12 months 90 days 12 months

Pt1000alpha 385

-200°C -80°C 0.011°C 0.012°C 0.06°C 0.07°C

-80°C -0°C 0.014°C 0.015°C 0.06°C 0.08°C

0°C 100°C 0.019°C 0.020°C 0.07°C 0.08°C

100°C 260°C 0.025°C 0.028°C 0.07°C 0.08°C

260°C 300°C 0.027°C 0.030°C 0.07°C 0.09°C

300°C 400°C 0.030°C 0.034°C 0.07°C 0.09°C

400°C 600°C 0.041°C 0.045°C 0.08°C 0.09°C

600°C 630°C 0.042°C 0.047°C 0.08°C 0.09°C

-328°F -112°F 0.020°F 0.022°F 0.11°F 0.13°F

-112°F 32°F 0.025°F 0.027°F 0.11°F 0.14°F

32°F 212°F 0.034°F 0.036°F 0.13°F 0.14°F

212°F 500°F 0.045°F 0.050°F 0.13°F 0.14°F

500°F 572°F 0.049°F 0.054°F 0.13°F 0.16°F

572°F 752°F 0.054°F 0.061°F 0.13°F 0.16°F

752°F 1112°F 0.074°F 0.081°F 0.14°F 0.16°F

1112°F 1166°F 0.076°F 0.085°F 0.14°F 0.16°F

Ni 120alpha 385

-80°C 0°C 0.009°C 0.010°C 0.04°C 0.05°C

0°C 100°C 0.010°C 0.011°C 0.04°C 0.04°C

100°C 260°C 0.011°C 0.012°C 0.03°C 0.03°C

-112°F 32°F 0.016°F 0.018°F 0.07°F 0.09°F

32°F 212°F 0.018°F 0.020°F 0.07°F 0.07°F

212°F 500°F 0.020°F 0.022°F 0.05°F 0.05°F

Cu10alpha 4272

-100°C 260°C 0.067°C 0.069°C 0.63°C 0.75°C

-148°F 500°F 0.121°F 0.124°F 1.13°F 1.35°F

YSI 400 15°C 50°C 0.005°C 0.007°C 0.005°C 0.007°C

59°F 122°F 0.009°F 0.013°F 0.009°F 0.013°F

Pt25SPRT

-200°C 660°C 0.05°C 0.06°C - -

-328°F 1220°F 0.09°F 0.11°F - -

Absolute uncertainty, tcal ±5°C / 9°F 1. 2-wire Output 2. Based on MINCO application aid no. 18.

°F

°C

°C

°F

°C°F°C°F

°F°C

- 4 - - 5 -

SS-CP-2380-US

Specification SheetSS-CP-2380-US

Specification Sheet

Thermocouple TEMPERATURErange

INPUT and OUTPUTaccuracy ± 1

from to 90 days 12 months

B 600°C 800°C 0.32°C 0.36°C

800°C 1000°C 0.29°C 0.29°C

1000°C 1550°C 0.30°C 0.30°C

1550°C 1820°C 0.34°C 0.35°C

1112°F 1472°F 0.58°F 0.65°F

1472°F 1832°F 0.52°F 0.52°F

1832°F 2822°F 0.54°F 0.54°F

2822°F 3308°F 0.61°F 0.63°F

C 0°C 150°C 0.15°C 0.20°C

150°C 650°C 0.11°C 0.16°C

650°C 1000°C 0.13°C 0.21°C

1000°C 1800°C 0.28°C 0.40°C

1800°C 2316°C 0.53°C 0.74°C

32°F 302°F 0.27°F 0.36°F

302°F 1202°F 0.20°F 0.29°F

1202°F 1832°F 0.23°F 0.38°F

1832°F 3272°F 0.50°F 0.72°F

3272°F 4200°F 0.95°F 1.33°F

E -250°C -100°C 0.28°C 0.40°C

-100°C -25°C 0.06°C 0.08°C

-25°C 350°C 0.04°C 0.05°C

350°C 650°C 0.04°C 0.06°C

650°C 1000°C 0.06°C 0.11°C

-418°F -148°F 0.50°F 0.72°F

-148°F -13°F 0.11°F 0.14°F

-13°F 662°F 0.07°F 0.09°F

662°F 1202°F 0.07°F 0.11°F

1202°F 1832°F 0.11°F 0.20°F

J -210°C -100°C 0.10°C 0.17°C

-100°C -30°C 0.08°C 0.10°C

-30°C 150°C 0.04°C 0.06°C

150°C 760°C 0.04°C 0.07°C

760°C 1200°C 0.08°C 0.13°C

-346°F -148°F 0.18°F 0.28°F

-148°F -22°F 0.14°F 0.18°F

-22°F 302°F 0.07°F 0.11°F

302°F 1400°F 0.07°F 0.13°F

1400°F 2192°F 0.14°F 0.23°F

K -200°C -100°C 0.15°C 0.23°C

-100°C -25°C 0.09°C 0.22°C

-25°C 120°C 0.04°C 0.06°C

120°C 1000°C 0.09°C 0.16°C

1000°C 1372°C 0.20°C 0.30°C

-328°F -148°F 0.27°F 0.41°F

-148°F -13°F 0.16°F 0.40°F

-13°F 248°F 0.07°F 0.11°F

248°F 1832°F 0.16°F 0.29°F

1832°F 2501°F 0.36°F 0.54°F

L 200°C -100°C 0.27°C 0.27°C

-100°C -800°C 0.16°C 0.16°C

800°C 900°C 0.07°C 0.07°C

392°F -148°F 0.49°F 0.49°F

-148°F 1472°F 0.29°F 0.29°F

1472°F 1652°F 0.13°F 0.13°F

Absolute uncertainty, tcal ±5°C / 9°F 1. Does not include thermocouple wire error and Cold Junction compensation.

°F

°C

°F

°C

°F

°F

°C

°F

°C

°F

°C

°C

- 6 - - 7 -

SS-CP-2380-US

Specification SheetSS-CP-2380-US

Specification Sheet

Thermocouple TEMPERATURErange

INPUT and OUTPUTaccuracy ± 1

from to 90 days 12 months

N -200°C -100°C 0.23°C 0.30°C

-100°C -25°C 0.10°C 0.14°C

-25°C 120°C 0.06°C 0.09°C

120°C 410°C 0.04°C 0.08°C

410°C 1300°C 0.11°C 0.17°C

-328°F -148°F 0.41°F 0.54°F

-148°F -13°F 0.18°F 0.25°F

-13°F 248°F 0.11°F 0.16°F

248°F 770°F 0.07°F 0.14°F

770°F 2372°F 0.20°F 0.31°F

R 0°C 250°C 0.48°C 0.48°C

250°C 400°C 0.24°C 0.25°C

400°C 1000°C 0.21°C 0.23°C

1000°C 1750°C 0.20°C 0.30°C

32°F 482°F 0.86°F 0.86°F

482°F 752°F 0.43°F 0.45°F

752°F 1832°F 0.38°F 0.41°F

1832°F 3182°F 0.36°F 0.54°F

S 0°C 250°C 0.46°C 0.46°C

250°C 1000°C 0.26°C 0.26°C

1000°C 1400°C 0.20°C 0.27°C

1400°C 1750°C 0.25°C 0.36°C

32°F 482°F 0.83°F 0.83°F

482°F 1832°F 0.47°F 0.47°F

1832°F 2552°F 0.36°F 0.49°F

2552°F 3182°F 0.45°F 0.65°F

T -250°C -150°C 0.41°C 0.53°C

-150°C 0°C 0.08°C 0.14°C

0°C 120°C 0.03°C 0.06°C

120°C 400°C 0.02°C 0.04°C

-418°F -238°F 0.74°F 0.95°F

-238°F 32°F 0.14°F 0.25°F

32°F 248°F 0.05°F 0.11°F

248°F 752°F 0.04°F 0.07°F

U -200°C 0°C 0.46°C 0.46°C

0°C 600°C 0.17°C 0.17°C

-328°F 32°F 0.83°F 0.83°F

32°F 1112°F 0.31°F 0.31°F

BP Russian Gost

0°C 200°C 0.32°C 0.32°C

200°C 600°C 0.22°C 0.22°C

600°C 800°C 0.29°C 0.06°C

800°C 1600°C 0.35°C 0.16°C

1600°C 2000°C 0.47°C 0.30°C

2000°C 2500°C 0.57°C 0.70°C

32°F 392°F 0.58°F 0.58°F

392°F 1112°F 0.40°F 0.40°F

1112°F 1472°F 0.52°F 0.54°F

1472°F 2912°F 0.63°F 0.65°F

2912°F 3632°F 0.85°F 0.86°F

3632°F 4532°F 1.03°F 1.26°F

XK Russian Gost

-200°C -100°C 0.12°C 0.12°C

-100°C 300°C 0.02°C 0.03°C

300°C 800°C 0.09°C 0.10°C

328°F -148°F 0.22°F 0.22°F

-148°F 572°F 0.04°F 0.05°F

572°F 1472°F 0.16°F 0.18°F

Absolute uncertainty, tcal ±5°C / 9°F 1. Does not include thermocouple wire error and Cold Junction compensation.

°F

°F

°C

°F

°C

°F

°C

°F

°C

°F

°C

°C

°C

°F

- 6 - - 7 -

SS-CP-2380-US

Specification SheetSS-CP-2380-US

Specification Sheet

DC current OUTPUT 1

Accuracy

Resolution Maximum compliance

voltage

Maximum inductive

load

90 days 12 months

0 to 100 mA 1 mA 10V 100 mH ±85 ppm + 2 µA ±100 ppm + 2 µA

Absolute uncertainty, tcal ±5°C / 9°F 1. All outputs are positive only.Noise: 2 µA p-p from 0.1 to 10 Hz, 20 µA from 10 Hz to 10 KHz

Mains specifi cations

Selectable ............................. 115 V(90-132), 230 V(197-264)Frequency .............................................................. 47 - 63 HzPower consumption (max.) ........................................... 15 VA

Warm up and settling time

Warm up - maximum ............................................. 30 minutesSettling time ................................................ Less than 5 sec.

Ambient temperature

Operating ........................................... 0 to 50°C / 32 to 122°FCalibration (tcal)................................. 18 to 28°C / 64 to 82°FStorage.............................................-20 to 70°C / -4 to 158°FTemperature coeffi cient (outside tcal ±5°C / ±9°F):Celsius......................................... 10% of specifi cation per °CFahrenheit .................................. 5.5% of specifi cation per °F

Relative humidity

Operating ............................................< 80% to 30°C / 86°F............................................................< 70% to 40°C / 104°F............................................................< 40% to 50°C / 122°FStorage............................................... <95% non-condensing

Altitude

Operating .............................................< 3,050 m / 10,000 ftNon-operating ..................................... < 12,200 m / 40,000 ft

Standard interface

Serial ...........................................................................RS 232Parallel ........................................................ IEEE-488 (GPIB)

CE-conformity

EN61326: 1997/A1:1998

Safety

EN 61010 Second, ANSI/ISA-S82.01-1994CAN/CSA-C22.2 No. 1010.1-92, NRTL

ThermocoupleCold junction

Temperature range Accuracy ±

min max 12 months

CJC compensation 18°C / 64°F 28°C / 83°F 0.1°C / 0.18°F

CJC outside above 0.05°C/°C / 0.05°F/°F

Isolation

Analog low isolation .........................................................20 V

Instrument dimensions

AMC900 W×HxD: ........ 300×120×220 mm / 11.5×4.7×8.9 in.

Instrument weight

AMC900 ............................................................... 2.3 kg / 5 lb

Pressure specifi cations

The JOFRA AMC900 can display pressure from the modules in any of the below mentioned engineering units.

Range:................................................ Determined by moduleAccuracy: ........................................... Determined by moduleResolution: .................................................................. 5 digits

psi .......................................................pound per square inchinH2O4°C.............................................inches of water at 4°CinH2O20°C........................................ inches of water at 20°CcmH2O4°C.................................. centimeters of water at 4°CcmH2O20°C ............................. centimeters of water at 20°CBAR..................................................................................barsmBAR ........................................................................millibarsKPAL..................................................................... kilopascalsinHG 0°C......................................... inches of mercury at 0°CmmHG 0°C ................................ millimeter of mercury at 0°CKg/cm2 ...............................kilograms per square centimeter

SPECIFICATIONS AMC900

- 8 - - 9 -

SS-CP-2380-US

Specification SheetSS-CP-2380-US

Specification Sheet

JOFRA industrial temperature reference probes are based on more than 50 years of industrial temperature sensor manufacturing experience. The main requirement of a reference probe is stability: The less the probe drifts, the lower the measurement uncertainty.

The JOFRA STS100A industrial temperature reference probes are built to last. All JOFRA Secondary Temperature Standard probes are economical and offer fast response times, low immersion depths, compact physical sizes, and specifi ed low drift rates: even at high temperatures. These are all important considerations when selecting a reference probe.

In addition to straight probes, AMETEK offers a 90° angled version specifically developed for use with dry-block temperature calibrators.

This probe allows the user to have both the sensor-under-test and the reference probe in the thermowell at the same time: even if the sensors have a connection or a transmitter head.

All probes are subjected to a long approval process. This includes mechanical stress reduction of the entire assembly as well as aging the sensor element itself. The purpose of aging the sensor is to remove the initial drift. The procedure involves cycling the sensor to 650°C / 1202°F a number of times and monitoring the drift. Finally all sensors are exposed to maximum temperature for 16 hours and again monitored for drift. To be accepted for fi nal calibration and certifi cation, the probe must meet our minimum tolerance.

For more details about the JOFRA STS-100 series please see specifi cation sheet: SS-CP-2290

Dimensions

Ref. mm inch

A 250 9.84

350 13.78

500 19.69

B 4 0.16

C 165 6.50

D 4 0.16

E 42 1.65

F 12 0.47

G 134 5.28

H 6 0.24

JOFRATM

STS SeriesSuperior Temperature Standard Probes

Temperature range

All probes ............................... -150 to 650°C / -238 to 1202°F

Accuracy

Hysteresis1) @0°C / 32°F................................ 0.01°C / 0.02°FLong term stability 2) @0°C / 32°F .......typ. 0.014°C / 0.025°FRepeatability 1) .........................................0.002°C / 0.0036°F1) when used in the range -80 to 650°C / -112 to 1202°F.2) when exposed to 650°C / 1202°F for 100 h. Stability will depend on actual use of the sensor.

Sensing element

Type................................................................................Pt100Nominal resistance@0°C / 32°F................................... 100 ΩLength ............................................................. 40 mm / 1.6 inTemperature coeffi cient ........................... α100=0.00385 1/°C

Minimum immersion depth

70 mm / 2.8 in

Self-heating effect

0.06°C/mW / 0.108°F/mW

Response time

τ0.5 (50%)................................................................ 8 secondsτ0.9 (90%).............................................................. 26 secondsLiquid in motion v=0.4m/s.

Electrical connections

Cable ................................................................4 wire + shieldConnection ..................................................LEMO goldplated

Insulation resistance

@ 23°C / 73°F ........................................................ 100 Gohm@ 650°C / 1202°F ....................................................70 Mohm

Outer tube

Inconel 600

Operating conditions

(Probe, connection, and cable) ................ Max. 70°C / 158°FStorage temperature ....................... -20 to 60oC / -4 to 140oFHumidity ............................................................ 0 to 90% RHProtection class (connectors) ......................DIN 40050 IP-50

Shipping dimensions

Straight probes (including aluminum case):L x W x H..................740 x 140 x 135 mm / 29.1 x 5.5 x 5.3 in90° angled probe (including carrying case):L x W x H.....................220 x 250 x 60 mm / 8.7 x 9.8 x 2.4 inShipping weight including packingStraight probes................................................. 1.9 kg / 4.2 lb90° angled probe................................................ 550 g / 1.2 lb

Standard delivery

• JOFRA STS-100 A probe• Cable - according to order number• Accredited certifi cate, points: -40, -20, 0, 50, 100, 200, 320, 450 and 650°C (-40, -4, 32, 122, 212, 392, 608, 842 and 1202°F).• 90° angled probe: Plastic carrying case with foam insert• Straight probes: Aluminium case with foam insert• User guide

SPECIFICATIONS STS SERIES

- 8 - - 9 -

SS-CP-2380-US

Specification SheetSS-CP-2380-US

Specification Sheet

Resolution: .......5 digits (20 to 100 ppm depending on range)Dimensions: ...................... 2.4 x 1.8 x 3 in / 60 x 45 x 75 mm Weight: ..............................................................0.77 lb / 350 gCable length: .......................................................24 in / 60 cm Pressure connections: .................................. 1/4” NPT femaleConversion adapter to:.................................. 1/4” BSP female Electrical connection:............................... Lemo, 5-pole, maleCalibration certificate: ..................................... NIST traceableWarm up:...................................10 minutes to rated accuracyReading rate:.................................................... 50 per secondDisplay rate ASC300 and AMC900: .................. 2 per secondCompliance - CE standards:........... EN 50081-1, EN 50082-1 EN 50081-2, EN 50082-2

The JOFRA APM S series are industrial pressure modules with good accuracies. The modules are designed so that the cable is integrated into the module housing, and the overall profile allows for easy storage of multiple modules.

Pressure Reference Over- Burst Reference 12 month Range Type psi bar port*2 port*2 pressure*3 pressure*4 accuracy*5 accuracy*6 Part number

1 psi*1 Gauge 1.0000 0.0689 DRY DRY x 5 x 10 0.15% 0.30% APM001PNSG 5 psi*1 Gauge 5.0000 0.3447 DRY DRY x 5 x 10 0.05% 0.10% APM005PNSG 15 psi*1 Gauge 15.000 1.0342 DRY N/A x 4 x 6 0.025% 0.05% APM015PNSG 1 psi Gauge 1.0000 0.0689 316L SS DRY x 4 x 6 0.075% 0.15% APM001PGSG 5 psi Gauge 5.0000 0.3449 316L SS DRY x 2 x 3 0.025% 0.05% APM005PGSG 15 psi Gauge 15.000 1.0342 316L SS N/A x 2 x 3 0.025% 0.05% APM015PGSG 30 psi Gauge 30.000 2.0684 316L SS N/A x 2 x 3 0.025% 0.05% APM030PGSG 100 psi Gauge 100.00 6.8948 316L SS N/A x 2 x 3 0.025% 0.05% APM100PGSG 200 psi Gauge 200.00 13.790 316L SS N/A x 2.5 x 3.7 0.025% 0.05% APM200PGSG 300 psi Gauge 300.00 20.684 316L SS N/A x 1.6 x 2.5 0.025% 0.05% APM300PGSG 500 psi Gauge 500.00 34.486 316L SS N/A x 2 x 3 0.025% 0.05% APM500PGSG 1000 psi Gauge 1000.0 68.948 316L SS N/A x 2 x 3 0.025% 0.05% APM01KPGSG 2000 psi Gauge 2000.0 137.90 316L SS N/A x 2.5 x 3.7 0.025% 0.05% APM02KPGSG 3000 psi Gauge 3000.0 206.84 316L SS N/A x 1.6 x 2.5 0.025% 0.05% APM03KPGSG 5000 psi Gauge 5000.0 344.76 316L SS N/A x 1.5 x 2 0.025% 0.05% APM05KPGSG 10000 psi Gauge 10000 689.48 316L SS N/A x 1.2 x 2 0.025% 0.05% APM10KPGSG ±15 psi Combination ±15.000 ±1.0342 316L SS N/A x 2 x 3 0.035%*12 0.07% APM015PCSG -15 to 200 psi Combination -015.00 -1.0342 316L SS N/A x 2.5 x 3 0.025%*13 0.05% APM200PCSG +200.00 +13.790 5 psi*10 Differential 5.0000 0.3449 316L SS DRY x 2 x 3 0.035% 0.07% APM005PDSG 100 psi*11 Differential 100.00 6.8948 316L SS DRY x 2 x 3 0.025% 0.05% APM100PDSG 5 psi Absolute 5.0000 0.3448 316L SS N/A x 6 x 9 0.035% 0.07% APM005PASG 15 psi Absolute 15.000 1.0342 316L SS N/A x 2 x 3 0.025% 0.05% APM015PASG 30 psi Absolute 30.000 2.0684 316L SS N/A x 2 x 3 0.025% 0.05% APM030PASG

1 bar Gauge 14.503 1.0000 316L SS N/A x 2 x 3 0.025% 0.05% APM001BGSG 4 bar Gauge 58.012 4.0000 316L SS N/A x 1.2 x 2.5 0.025% 0.05% APM004BGSG 6 bar Gauge 87.018 6.0000 316L SS N/A x 2.3 x 3.5 0.025% 0.05% APM006BGSG 10 bar Gauge 145.03 10.000 316L SS N/A x 2 x 3 0.025% 0.05% APM010BGSG 16 bar Gauge 232.05 16.000 316L SS N/A x 2.1 x 3.2 0.025% 0.05% APM016BGSG 21 bar Gauge 304.56 21.000 316L SS N/A x 1.6 x 2.4 0.025% 0.05% APM021BGSG 25 bar Gauge 362.58 25.000 316L SS N/A x 2.7 x 4.1 0.025% 0.05% APM025BGSG 40 bar Gauge 580.12 40.000 316L SS N/A x 1.7 x 2.5 0.025% 0.05% APM040BGSG 60 bar Gauge 870.18 60.000 316L SS N/A x 2.2 x 3.4 0.025% 0.05% APM060BGSG 100 bar Gauge 1450.3 100.00 316L SS N/A x 3.4 x 5 0.025% 0.05% APM100BGSG 160 bar Gauge 2320.5 160.00 316L SS N/A x 2.1 x 3.2 0.025% 0.05% APM160BGSG 200 bar Gauge 2900.6 200.00 316L SS N/A x 1.7 x 2.5 0.025% 0.05% APM200BGSG 250 bar Gauge 3625.8 250.00 316L SS N/A x 2 x 2.7 0.025% 0.05% APM250BGSG 400 bar Gauge 5801.2 400.00 316L SS N/A x 1.3 x 1.7 0.025% 0.05% APM400BGSG 600 bar Gauge 8701.8 600.00 316L SS N/A x 1.4 x 2.2 0.025% 0.05% APM600BGSG 700 bar Gauge 10152 700.00 316L SS N/A x 1.2 x 2 0.025% 0.05% APM700BGSG ±1 bar Combination ±14.503 ±1.0000 316L SS N/A x 2 x 3 0.035%*14 0.07% APM001BCSG -1 bar Vacuum -14.503 -1.0000 316L SS DRY 2 bar 3 bar 0.035% 0.07% APM001BVSG 7 bar Absolute 101.52 7.0000 316L SS N/A x 2 x 3 0.025% 0.05% APM007BASG

7 kPa Gauge 1.0152 0.0700 316L SS DRY x 4 x 6 0.10% 0.20% APM007KGSG 35 kPa Gauge 5.0761 0.3500 316L SS DRY x 2 x 3 0.035% 0.07% APM035KGSG 70 kPa Gauge 10.152 0.7000 316L SS N/A x 2 x 3 0.035% 0.07% APM070KGSG 160 kPa Gauge 23.206 1.6000 316L SS N/A x 2 x 3 0.025% 0.05% APM160KGSG 200 kPa Gauge 29.006 2.0000 316L SS N/A x 2.1 x 3.1 0.025% 0.05% APM200KGSG 250 kPa Gauge 36.259 2.5000 316L SS N/A x 2 x 3 0.025% 0.05% APM250KGSG 350 kPa Gauge 50.761 3.5000 316L SS N/A x 2 x 3 0.025% 0.05% APM350KGSG

-29 inHg Vacuum -14.243 -1.0342 316L SS DRY 60 inHg 90 inHg 0.035% 0.07% APM029HVSG

10 inH2O*1 Gauge 0.3612 0.0249 DRY DRY x 5 x 10 0.20% 0.30% APM010WGSG

JOFRATM

APM SeriesAdvanced Pressure Modules Series S

- 10 - - 11 -

SS-CP-2380-US

Specification SheetSS-CP-2380-US

Specification Sheet

The JOFRA APM H series are high accuracy laboratory units: 0.01% of F.S. The outstanding performance makes these modules perfect for use as an electronic pressure reference at the top of your calibration hierarchy. These modules are easy to work with and easy to transport.

Pressure Reference Over- Reapeat- 6 month Range Type psi bar port*2 port*2 pressure ability*7 Accuracy*8 Part number 15 psi*9 Gauge 15.0000 1.03427 DRY DRY x 1.5 0.003% 0.010% APM015PGHG 50 psi Gauge 50.0000 3.44756 DRY DRY x 1.5 0.003% 0.010% APM050PGHG 100 psi Gauge 100.0000 6.89512 DRY DRY x 1.5 0.003% 0.010% APM100PGHG 500 psi Gauge 500.000 34.4756 DRY DRY x 1.5 0.003% 0.010% APM500PGHG 1000 psi Gauge 1000.00 68.9513 DRY DRY x 1.5 0.003% 0.010% APM01KPGHG 15 psi*9 Absolute 15.0000 1.03427 DRY N/A x 1.5 0.003% 0.010% APM015PAHG 50 psi Absolute 50.0000 3.44756 DRY N/A x 1.5 0.003% 0.010% APM050PAHG 100 psi Absolute 100.000 6.89512 DRY N/A x 1.5 0.003% 0.010% APM100PAHG 500 psi Absolute 500.000 34.4756 DRY N/A x 1.5 0.003% 0.010% APM500PAHG 1000 psi Absolute 1000.00 68.9513 DRY N/A x 1.5 0.003% 0.010% APM01KPAHG 3000 psi Absolute 3000.00 206.854 DRY N/A x 1.25 0.003% 0.010% APM03KPAHG

Note *1 Non isolated transducer - dry air or non-corrosive gas compatible media only Note *2 Media compatability on pressure or reference port side of the module - DRY indicates dry air or non-corrosive gas as compatible media. 316L SS indicates media compatible with type 316 stainless steel Note *3 Applying steady pressure up to the specifi ed overpressure will not destroy or change performance of the module Note *4 Applying burst pressure (short pressure spikes) above the specifi ed limit can destroy the module Note *5 Reference accuracy is expressed as % F.S. and includes only liniarity, hysteresis, and repeatability at laboratory conditions. Note *6 12 month accuracy - expressed as % F.S. Includes liniarity, repeatability, hysteresis, and temperature compensation 0 to 50°C (32 to 122°F). Note *7 The closeness of agreement between independent test results obtained under laboratory conditions. Note *8 6 month accuracy - expressed as % F.S. Includes liniarity, repeatability, hysteresis, and temperature compensation 15 to 45°C (59 to 113°F). Note *9 All ranges less than 30 psi (2 bar) specify physical orientation of use. Note *10 5 psi module: Line pressure max. 10 psiNote *11 100 psi module: Line pressure max. 200 psiNote *12 F.S. = 30 psiNote *13 F.S. = 215 psiNote *14 F.S. = 2 bar

APM ORDER NUMBER STRUCTURENumber Description

SPK-APM-001 APM H communication cable to JOFRA APC, ASC300 and AMC900SPK-APM-002 Communication kit for PC Pressure port fi tting:12-90135 APM H to 1/8” in male NPTT-786 APM S to 1/4” in female BSP60R115 APM S to 1/4” in male BSP

For connection to various pumps, please see the individual pump descriptions.

Resolution: ...........6 digits (2 to 10 ppm depending on range)Dimensions: ................ 3.9 x 2.18 x 2.18 in / 99 x 55 x 55 mm Weight: .................................................................. 1 lb / 450 g Pressure connections: ............................ 7/16-20 SAE femaleConversion adapter to:...................1/8” NPT and BSP female Electrical connection:................................9 pole D-sub, maleCommunication cable: ......module to ASC300/AMC900/APCCable length: .......................................................24 in / 60 cm Calibration certifi cate: ..................................... NIST traceableWarm up:...................................10 minutes to rated accuracyReading rate:.................................................... 50 per secondDisplay rate ASC300 and AMC900: .................. 2 per secondCompliance - CE standards:........... EN 50081-1, EN 50082-1

EN 50081-2, EN 50082-2

Order No. Description

Pressure module - 1st thru 3rd charactersAPM Advanced pressure module Pressure range - 4th thru 6th characters Engineering unit - 7th character P psi B bar K KPa W inH2O H inHg Pressure reference - 8th character G Gauge N Gauge (Non-isolated transducer) A Absolute D Differential V Vacuum C Combined vacuum and pressure Accuracy family - 9th character S Standard industrial accuracy H High laboratory accuracy Certifi cate - 10th character G NIST traceable calibration certifi cate H Accredited calibration certifi cate

APM 015PGSG Sample order number

APM ACCESSORIES

JOFRATM

APM SeriesAdvanced Pressure Modules Series H

- 10 - - 11 -

t e m p e r a t u r e

s i g n a l

s o f t w a r e

p r e s s u r e

CALIBRATION INSTRUMENTS

AMETEK Test & Calibration Instruments • Florida, USA (Western Hemisphere)Tel: +1 727-536-7831 • Tel: +1 800-527-9999 • [email protected]

AMETEK Denmark A/S • Denmark (Europe and the Middle East)Tel: +45 4816 8000 • [email protected]

AMETEK GmbH • Germany (Germany only)Tel: +49 2159 91360 • [email protected]

AMETEK Singapore Pte. Ltd. • Singapore (Asia)Tel: +65 6 484 2388 • [email protected]

AMETEK Inc. Beijing Representative Office • China (China only)Tel: +86 10 85262111• [email protected]

Information within this document is subject to change without notice.©2005, by AMETEK, Inc. All rights reserved.

Pub code SS-CP-2380-US Issue 0504

AMETEK Calibration Instruments

offers a complete range of calibration equipment for pressure, temperature,

and signal - including software.

JOFRA Temperature standardsPortable precision thermometer. Dry-block calibrators: 4 series, more than 20 models - featuring speed, portability, accuracy, and advanced

documenting functions.

M&G Primary pressure standardsPneumatic fl oating-ball or hydraulic piston deadweight testers - easy-to-use with accuracies up to 0.015% of

reading.

JOFRA Pressure standardsConvenient electronic systems ranging from -1 to 700 bar (25 inHg to 10,000 psi) - multiple choices of pres-sure ranges, pumps, and accuracies, fully temperature-compensated for problem-free and accurate fi eld use.

JOFRA Signal calibrationProcess signal measurement and simulation for easy control loop cali-bration and measurement tasks - from handheld fi eld instruments for multi or single signals to laboratory refe-

rence level bench top instruments.

...because calibration is a matter of confi dence

AMETEK is a leading global manufacturer of electrical and electromechanical products for niche markets. Listed on the New York Stock Exchange (AME) since 1930. AMETEK’s annual sales exceed $1 billion. Operations are in North America, Europe, and Asia, with about one third of sales to markets outside the United States.

www.ametekcalibration.com www.jofra.com

JOFRA AMC900 - Advanced Multi-purpose Calibrator

Order No. Description Base model number (1st thru 6th characters)AMC900 JOFRA AMC900 Benchtop calibrator Power supply - (7th thru 9th characters) 115 115 VAC, 50/60 Hz 220 230 VAC, 50/60 Hz Mains power cable type (10th character) A EUROPEAN, 230 V B USA/CANADA, 115 V C UK, 240 V D SOUTH AFRICA, 220 V E ITALY, 220 F AUSTRALIA, 240 V G DENMARK, 230 V H SWITZERLAND, 220 V I ISRAEL, 230 V Certifi cate (11th character) G NIST traceable certifi cate (standard) H Accredited certifi cate

AMC900 115 B G Sample order number JOFRA AMC900 for 115 VAC, NIST traceable certifi cate.

ORDERING INFORMATION

ACCESSORIES

- 12 -

Part No. Description

SPK-AMC-001 LEMO to banana bushings with 1m / 3 ft. cableSPK-AMC-002 Rack/panel mount kitSPK-AMC-003 LEMO connector (kit)SPK-AMC-004 Manual for JOFRA AMC900121985 Extension cable for Pt100 sensor 5 m121983 Extension cable for type K 5 m122523 Extension cable for type N 5 m124012 Alu-carrying case120519 Thermocouple male plug - type Cu-Cu (White)120518 Thermocouple male plug - type R-S (Green)120517 Thermocouple male plug - type K (Yellow)120516 Thermocouple male plug - type J (Black)120515 Thermocouple male plug - type T (Blue)120514 Thermocouple male plug - type N (Orange)2206011 Wire adapter - type K2206012 Wire adapter - type T104203 Test lead set105366 RS-232 cable for the AMC900 unit