advanced gpc part 1 – gpc and viscometry
DESCRIPTION
Advanced GPC Part 1 – GPC and Viscometry. Introduction. The GPC experiment with a single concentration detector is called conventional GPC This is by far the most common form of GPC However there are some limitations to this technique - PowerPoint PPT PresentationTRANSCRIPT
Advanced GPC Part 1 – GPC and Viscometry
2
Introduction
The GPC experiment with a single concentration detector is called conventional GPC
This is by far the most common form of GPC
However there are some limitations to this technique
Recently, developments in detector technology have made viscometers more widely available
These detectors avoid some of the problems associated with conventional GPC
This presentation outlines GPC viscometry as an analysis methodology
3
Gel permeation chromatography separates polymers on the basis of size in solution
Separation occurs through the partitioning of polymer molecules into the pore structure of beads packed in a column
Re-cap - Gel Permeation Chromatography (GPC)
4
Conventional GPC
Calibrate the column by chromatographing a number of narrow standard polymers of known molecular weight, correlating MW with molecular size
For unknown samples slice the peak into components of weight Mi and height/area Ni, sum to determine molecular weight averages
5
Column separates on basis of molecular size NOT molecular weight
two different polymers will interact differently with solvent
At any molecular weight, the two polymers will have different sizes in solution
Molecular weights from conventional GPC are dependent on a comparison in size between the standards and the sample
The result – practically speaking the majority of conventional GPC experiments give the wrong results!
Viscometers get round this problem…
Limitations with Conventional GPC
6
Viscosity of Polymers
All polymers increase the viscosity of solutions by increasing the resistance to flow
Different types of polymers have differing viscosities depending on the interactions with the solvent
Viscometers are used to determine intrinsic viscosity, IV or [ŋ]
Intrinsic viscosity can be though of as the inverse of the molar density
At any given MW, a high IV means the sample is a large diffuse molecule, a small IV means a compact, dense molecule
7
Intrinsic Viscosity
8
Effect of Solvent and Temperature on Intrinsic Viscosity
Polystyrene
Solvent affects the intrinsic viscosity of polymers by altering how well solvated they are
Large changes occur in solvents of different polarities
Temperature has less of an effect
9
So Why do Viscometry? – The Universal Calibration
Ref : Grubisic, Rempp, Benoit, J. Polym. Sci., Part B, Polym. Lett., 5:753 (1967)
If a calibration of size versus retention time could be generated then one true calibration would hold for all sample types
Hydrodynamic volume = [] M
A Universal Calibration plot of log[]M versus RT holds true for all polymer types
Can use measured intrinsic viscosity and retention time to get accurate molecular weights
10
Accurate Molecular Weights
As a result of using the viscometer, a universal calibration can be set up that gives the same calibration line regardless of the type of standards employed
The chemistry of the sample is also unimportant – the column is separating on size and that is the parameter we have calibrated
Therefore the GPC/viscometer experiment will give accurate molecular weights for any samples regardless of their or the standard’s chemistry assuming that pure SEC takes place
We are still doing chromatography – the column must be calibrated
11
Comparisons of Conventional and Universal Calibrations
Conventional calibrations are offset due to differences in the molecular size of polystyrene and polyethylene
Universal calibrations account for the offset to the calibrations overlay
Discrepancy at low molecular weight is due to a conformation change in polyethylene
12
The Mark-Houwink Plot
A Mark-Houwink plot of log IV versus log M should give a straight line as long as the Universal Calibration is obeyed (i.e no interactions occur)
K and alpha vary between different solvents and polymers
Alpha is an indication of the shape of the polymer in solution
IV
13
The PL-BV 400 Series
14
Viscometer Operation
T
15
A viscometer that measures specific viscosity
A concentration detector that tells us how much material is eluting from the column
Can be any type that gives a response proportional to concentration
Typically a differential refractive index detector is used
DRI detector response proportional to concentration
Operation identical to conventional GPC, determines the concentration of material eluting from a GPC column
RIsignal = KRI (dn/dc) C
Measuring Intrinsic Viscosity - What do we need?…
16
Calibration with a series of narrow standards of known Mp and concentration
Calculate detector constant (Kvisc) using one standard for which IV is known
For the remainder of the standards, calculate [ from the viscometer response
Plot log M[ versus retention time to generate the Universal Calibration
For unknown sample, for each slice across the distribution determine [ from the viscometer, and then convert to molecular weight via the Universal Calibration curve
GPC/Viscometry Experimentation
17
Typical Chromatograms
18
Analysis
19
Columns: 2 x PLgel 5µm MIXED-C Eluent: TetrahydrofuranFlow rate: 1.0 ml/min Temperature: 40˚CDetector: PL-GPC 50 Plus differential refractive index, PL-BV 400RT viscometer
Example chromatograms of one sample
Analysis of Poly(styrene-co-butadiene)
20
Only small differences in the MWD of the two samples
21
The Mark-Houwink plots indicate the materials are structurally similar
22
Columns: 2 x PLgel 5µm MIXED-D Eluent: TetrahydrofuranFlow rate: 1.0 ml/min Temperature: 40˚CDetector: PL-GPC 50 Plus differential refractive index, PL-BV 400RT viscometer
Example chromatograms of one sample
Analysis of Polylactide and Poly(lactide-co-glycolide)
23
The copolymer (red) has a considerably lower molecular weight than the homopolymer (blue)
24
Structurally the co-polymer is very different to the homopolymer across the molecular weight range
25
Columns: 3 x PLgel 10µm MIXED-B Eluent: Dimethyl sulphoxide + 0.1% lithium bromideFlow rate: 1.0 ml/min Temperature: 50˚CDetector: PL-GPC 50 Plus differential refractive index, PL-BV 400RT viscometer
Example chromatograms of one sample
Analysis of Cornflour
26
Large differences in the MWD of the two samples
27
Large differences in the Mark-Houwink plot indicate the samples are structurally dissimilar
28
Conventional GPC has limitations in that the results obtained are purely comparative
The situation can be remedied by adding a viscometer to the system
The viscometer allows calibrations of retention time as a function of molecular size to be generate
This give accurate molecular weight information regardless of the type of standards used in the analysis
The Mark-Houwink plot allows the change in density of the polymers as a function of molecular weight to be analysed
Summary