acs800 manual for laboratory works

16
Tallinn Technical University Department of Electrical Drives and Power Electronics Valery Vodovozov, Dmitri Vinnikov, Tanel Jalakas ACS800 Manual for Laboratory Works Tallinn 2007 

Upload: gedeus

Post on 13-Jul-2015

35 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 1/16

 

Tallinn Technical University Department of Electrical Drives and Power Electronics 

Valery Vodovozov, Dmitri Vinnikov, Tanel Jalakas

ACS800Manual for Laboratory Works

Tallinn 2007 

Page 2: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 2/16

 

 

Content

Preface .........................................................................................................................................2Safety Warnings ..........................................................................................................................2

Laboratory Setup Description......................................................................................................2

Laboratory Works Instructions....................................................................................................4

1 Commissioning the Power Converter...................................................................................4

1.1 Remote control the power converter from console .......................................................4

1.2 Control the power converter from the control panel .....................................................5

1.3 Data observing from the control panel ..........................................................................6

1.4 Taking measurements....................................................................................................6

Individual report contents....................................................................................................7

2 Power Converter Characteristics ..........................................................................................8

2.1 Voltage/frequency characteristics .................................................................................82.2 Critical speeds control ...................................................................................................8

2.3. IR-compensation...........................................................................................................8

2.4 Output characteristics ....................................................................................................9

Individual report contents....................................................................................................9

3 Computer Examining of Power Converter .........................................................................10

3.1 Acquaintance user interface of DriveWindow ............................................................10

3.2 Control the drive..........................................................................................................10

3.3 Monitoring the drive acceleration................................................................................11

3.4 Saving and analyzing the results .................................................................................11

Individual report contents..................................................................................................12

Using Digital Oscilloscope TPS2000 Series .............................................................................131. Viewing a Signal ...............................................................................................................13

2. Taking Quick Measurements.............................................................................................13

3. Taking Cursor Measurements............................................................................................13

4. Taking Automatic Measurements......................................................................................14

Initial Settings of ACS800 for Laboratory Works ....................................................................15

Page 3: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 3/16

 

  2

Preface 

The manual contains the instructions which the students must follow carrying out laboratory

works. The reader is expected to know electronic components, the standard electrical wiring, and

electrical schematic symbols. The brochure agrees with the curriculum and contents of 

“Advanced Course on Power Electronics”. Read and understand the safety warnings prior to start

working with the laboratory setup.

Safety Warnings 1.  Always remember that the laboratory setup is a dangerous device. Never apply the main

power and never run the motor if it may cause a danger or injury.

2.  Before energize the laboratory setup, make sure that all equipment and measuring

instruments are suitable for operation throughout the speed, voltage, and current range

provided by the drives.

3.  To avoid electric shock, never open the covers and do not touch with cables or motors

when the main power is applied or shafts rotate.

4.  Do not control the motors by the device disconnecting; instead, use the control panels’

keys or console buttons.5.  Do not run motors over the rated speed and do not allow their long-term overloading.

6.  Remember that the STOP key on the control panel has no effect if the control location is

not set to LOCAL; to stop the drive in this mode press the STOP button on the console.

7.  Remember that pressing the STOP key or the STOP button neither causes an emergency

braking of the motor nor instantly disconnects the drive from dangerous potential. Press

the EMERGENCY switch on the console to stall the drive in danger.

8.  Remember that the maximum allowed number of power-ups is 5 in 10 minutes; this

exceeding may damage the capacitors and switching devices.

Laboratory Setup Description 

The laboratory setup includes the two motor drives ACS800 series: the testing drive and the

loading drive. Each drive has the same structure and consists of:

•  induction motor

•  power converter

•  remote console

•  cabinet, housing, measuring, and cabling equipment

The motor shafts of the drives are mechanically coupled to each other to provide joint rotation.

Both power converters are wall-mountable low-harmonic units supplied the motors. Eachincludes the line-side active rectifier and the motor-side inverter connected via the dc link.

The typical converter cabinet contains:

•  power converter

•  cooling fan under the top cover

•  control drive panel mounted on the front cover

•  heat sink on the back side

•  connection box under the bottom cover

Page 4: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 4/16

 

  3

The main circuit diagram of a single motor drive is shown in Fig. 1. Each converter involves two

modules coupled via dc link. Both the line-side rectifier and the motor-side inverter have six

IGBTs with freewheeling diodes. In the motoring mode, the line-side rectifier passes the three-

phase ac voltage into the intermediate dc link, which further supplies the motor-side inverter

running the motor. Instead, in the braking mode, the motor-side inverter passes energy back fromthe motor to the supply lines through the dc link and the line-side rectifier. The line filter

suppresses the voltage and current harmonics. Electrolytic capacitors in the dc link serve for

energy buffering. By default, the converter dc link voltage reaches the peak value of the line-to-

line voltage and if necessary can be set also higher. The IGBT gating is based on the space

vector modulation principle further used for the scalar converter adjustment or direct torque

control (DTC). To calculate the switch turn on/off instants and fault protect, the line currents and

the dc link voltage are measured automatically by the converter sensors.

There are three sources for the drive control:

•  REM – remote (external) control from the console coupled via input/output connections

(I/O) or fieldbus interface

•  LOC [1] – local control from the control panel

•  LOC [2] – local control from the computer DriveWindow software

Each source provides a user interface both the line-side and motor-side modules with essential

controls and parameter settings for the application programs.

Fig. 1

M

Page 5: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 5/16

 

  4

Laboratory Works Instructions 

1 Commissioning the Power Converter

Objective: Commissioning is a system start-up that involves testing and adjusting to assure

proper functioning and understanding the principle of operation. The work aim is to learn how to

start, stop, change the rotation direction, adjust the converter from the console and the controlpanel, and perform several ways to take the signals observing and measuring as well.

1.1 Remote control the power converter from console

•  Check the laboratory setup. Make sure that the loading drive is not powered.

•  Learn about the console using Fig. 2 and familiarize with the control panel in Fig. 3.

•  Power on the testing drive using the Main switch . Obtain information from the controlpanel: panel identification data, drive identification data, and actual signals.

•  Press the ACT key on the control panel and explore the status row displayed there:

Module ID Control status Direction Reference Drive status

1 – motor-side

2 – line-side

 L – local

non or R– remote→ forward

← reverse

 Hz

 I – running

0 – stopped

non – disabled

•  Ensure the drive is in remote control mode (no L visible in control status); otherwise,press LOC/REM key to change between the local and remote control.

Fig. 2

Page 6: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 6/16

 

  5

•  To run the motor press the Start   button on the console. To adjust the speed, turn the

Speed reference  knob. To change the direction of the motor shaft rotation, use the

Reverse switch. To turn off the motor press the Stop button.

•  Power off the laboratory setup.

1.2 Control the power converter from the control panel

•  Ensure the laboratory setup is not powered and the motors not rotate.

•  Observe the tested drive ratings before the further work by reading and recording the data

from converter and motor nameplates.

•  Power on the testing drive.

•  Ensure the motor-side inverter is currently being controlled; otherwise, press the DRIVE  and ENTER keys, scroll to ID number 1, and press the ENTER and ACT keys to verify

the change.

•  If a fault occurs during the control, display **FAULT** or **WARNING** by the ACT  key, and press the RESET key.

•  Learn about the following general parameter setting procedure of the control panel:

o  to enter the Parameter Mode press the PAR key

o  to scroll the parameter groups press the PAGEUP and PAGEDN keys

o  to scroll parameters within a group use the UP and DOWN keys

o  to activate the setting of a new value press the ENTER key

o  change the parameter value by the UP  and DOWN  keys or fast change by the

PAGEUP and PAGEDN keys

o  to fix your choice press the ENTER key

•  Follow this procedure, enter START-UP DATA (parameter group 99), then set given

 APPLICATION MACRO and  MOTOR CONTROL MODE , and further enter the motor

data from the motor nameplate:   MOTOR NOMINAL VOLTAGE , CURRENT ,

FREQUENCY , SPEED, POWER.•  To change from Remote  to Local control  press the LOC/REM  key ( L visible on the

display).

No Name Function 1 LCD  Display, 4 lines2 ACT  Actual Signals Mode3 PAR  Parameter Mode4 FUNC  Function Mode5 DRIVE  Drive Selection Mode

6 PAGEUP  Large step up7 UP  Small step up8 ENTER  Set the choice9 PAGEDN  Large step down10 DOWN  Small step down11 LOC/REM  Local/Remote control12 RESET  Fault reset13 REF  Reference setting14 START  Run the motor15 FORWARD  Forward direction16 REVERSE  Reverse direction17 STOP  Stop the motor

Fig. 3

Page 7: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 7/16

 

  6

•  To set or change the frequency Reference press the REF key, then press the UP/DOWN  

to slow reference change or the PGUP/PGDOWN keys to fast change. Press the ENTER  

key to save the new reference in the memory until the power switch-off or further

changing.

•  To run the motor, press the START  key on the control panel. To change the motor

direction of rotation, press the REVERSE or FORWARD key. To stop the motor press the

STOP key.

1.3 Data observing from the control panel

•  Run the drive in remote control mode to the given reference value.

•  Press the ACT key to enter the Actual Signals Mode (parameter group 01) where the LCD  

displays three signals at a time. In the table below, some possible signals are listed:

Parameter Name Description

01.03 FREQUENC[Y] Inverter output frequency, Hz

01.04 CU[RRENT] Motor current, A

01.06 POWER Motor power, % of nominal

01.07 DC VOLTA[GE] Intermediate link voltage, VDC

01.08 MAINS VO[LTAGE] Supply rms voltage, V

01.09 MOTOR VO[LTAGE] Motor rms voltage, V

01.10 PP TEMPE[RATURE] Temperature of heat sink, C

01.37 MOTOR TE[MPERATURE] Estimated motor temperature, C

•  Read and record sequentially the motor-side inverter values from this list. To change a

signal, move the blinking cursor by the UP/DOWN keys and press the ENTER key in the

selected row, than select an actual signal by the UP/DOWN keys and press the ENTER  

key again.•  To obtain the line-side rectifier values, unacceptable in the motor-side wiring, press the

DRIVE  key, scroll to  ID number 2, and press the ACT  key. Using the preceding

procedure, read and record the rectifier values accordingly the next table:

Parameter Name Description

01.05 FREQUENCY  Line frequency, Hz

01.06 LINE CURRENT  Line current, A

01.07 REACTIVE POWER Reactive power, kVAr

01.08 POWER Line-side power, kW

01.09 POWER Input power, % of nominal

01.10 DC VOLTAGE  Intermediate link voltage, VDC

01.11 MAINS VOLTAGE  Supply rms voltage, V

01.27 COSFII  Power factor

1.4 Taking measurements

•  Employ the oscilloscope to view and measure ac voltages and currents using threedifferent ways.

•  Firstly, use the oscilloscope graticule to make a quick, visual estimate of a waveform,

count the major and minor divisions, and calculate the peak-to-peak voltage. Then, recordthem and sketch the on-screen waveforms.

Page 8: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 8/16

 

  7

•  Secondly, switch on the amplitude and time cursors, move them along the measured

trace, read their numeric values from the display readouts, and record them.

•  Thirdly, take automatic measurements of the data pointed in the above tables, which are

more accurate and complete than the two preceding ones.

•  Attach the multimeter, measure, and record the same currents and voltages from its

readouts.

•  Stop and power off the drive.

Individual report contents

•  functional circuit and specification of the laboratory setup

•  converters and motors data including company, type, standard, degree of protection aswell as the electrical and mechanical ratings

•  measuring instruments data in the same scope as above

•  the given reference value of the testing drive

•  table of the signals, measured using the control panel, oscilloscope, and multimeter

  scaled diagrams of the current and voltage waveforms sketched from the oscilloscopetraces

•  conclusions concerning the results estimation, comparison, and explanation

Page 9: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 9/16

 

  8

2 Power Converter Characteristics

Objective: Characteristics help to understand converter performance in the loading and adjusting

processes. The work aim is to observe and explain converter characteristics as well as to build

the corresponding diagrams.

2.1 Voltage/frequency characteristics

Voltage/frequency characteristic is the trace of a converter output voltage versus frequency. It

shows the converter linearity and the frequency and voltage range. When operated above a motor

field weakening point, the converter cannot reach the full required output voltage thus reducing

the load capacity.

•  Power on the testing drive. Set the zero reference frequency and define the motor voltage,

current, and power as the actual signals. Record the obtained frequency, voltage, and

power into the protocol table.

•  Run the drive. Smoothly changing the reference in the range 0…70 Hz, fill the measured

values into the protocol again.

•  Using the records, build the voltage/frequency and power/frequency diagram. Find the

field weakening border and the maximum motor voltage in the diagram.

•  Return the reference frequency to zero and stop the drive.

2.2 Critical speeds control

The critical speeds functions are useful for applications where it is required to avoid certain

motor speeds or speed bands because of mechanical resonance or supply oscillation problems.

  To activate the functions set the parameter 25.01 CRIT SPEED SELECT to ON . •  Define the minimum and maximum limits for the critical speed band by setting the

parameters 25.02 CRIT SPEED 1 LOW  and 25.03 CRIT SPEED 1 HIGH  to a given

values (10…40 and 60…90 Hz accordantly).

•  Run the drive. Smoothly changing the reference in the range 0…70 Hz, fill the measured

values into the protocol and build the voltage/frequency and power/frequency diagram.

Show the critical speed band in the diagram.

•  Return the reference frequency to zero, stop the drive, and set the parameter 25.01 to

OFF .

2.3. IR-compensation

When IR-compensation is activated, the converter gives extra voltage boost to the motor at low

frequencies. It is useful in applications that require high breakaway torque. Since voltage cannot

be fed to the converter at 0 Hz, full IR-compensation starts around the slip frequency.

•  To define the relative output voltage boost at zero speed, set the parameter 26.03 IR-

COMPENSATION to a given value (10…30% of the motor nominal voltage).

•  Run the drive. Smoothly changing the reference in the range 0…70 Hz, fill the measured

values into the protocol and build the voltage/frequency and power/frequency diagram

again. Show the IR-compensation band in the diagram.

•  To define a frequency at which the IR-compensation reaches full value, set the parameter

26.04 IR STEP-UP FREQ to a given value (5...25 Hz) and repeat the experiment. Showthe new IR-compensation band in the diagram.

Page 10: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 10/16

 

  9

•  Return the reference frequency and the parameters 26.03 and 26.04 to zero and stop the

drive.

2.4 Output characteristics

Output characteristics collect a set of dependences of a converter output voltage versus current.

They describe the loading possibilities and steady properties of a converter.

•  Run the testing drive to a given reference frequency (20...30 Hz).

•  Power on the loading drive, set the reference torque to zero, and run it.

•  Record the loading torque and testing frequency, voltage, current, and power into the

protocol table.

•  Smoothly raising the loading torque from zero to the nominal current of the testing drive,fill the same measured values into the protocol again. Unload the drive. Using the

records, build the output diagram for the given frequency.

•  Set other given reference frequency of the testing drive. Then, changing the loading

torque from zero to the motor nominal current fill the measured values into the protocol

and build the output diagram for the new frequency. Further, repeat the experiment in the

frequency range 0 to 70 Hz.

•  Set the previous value of IR-compensation and repeat the experiment in the low-frequency region.

•  Unload and remove IR-compensation, stop and power off the drives.

Individual report contents

•  tables and diagrams of voltage/frequency characteristics without and with the critical

speed bands and IR-compensation

  table and diagrams of output characteristics with IR-compensation band•  conclusions concerning the results estimation, comparison, and explanation

Page 11: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 11/16

 

  10

3 Computer Examining of Power Converter

Objective: To maintain and learn ABB drives, the DriveWindow soft tool is intended. The work 

aim is to master the converter remote control, its signal monitoring, working with parameters,

and graphical trending.

3.1 Acquaintance user interface of DriveWindow

The user interface shown in Fig. 4 includes the standard Windows  areas handling from the

keyboard and mouse by user editing, moving, dragging, and resizing.

•  Power on the testing drive.

•  Connect DriveWindow  to the drive by choosing   ABB SMP string in the Select OPC 

Server dialog box

•  Find the main parts of the user interface:

o  title bar

o  menu bar

o  toolbars including Standard  toolbar, Monitor  toolbar, Logger  toolbar, and Drive  

panel at the bottom

o  status bar along the bottom

o  window area divided by four panels: Browse tree and Item sets on top as well as

Trend settings and Trend display below

3.2 Control the drive

To control the drive, either the Drive  panel or the Drive  menu can be used. The status bar

describes user actions.

•  Select the drive in the Browse tree pane.

•  Take the control by toggling the Take/Release Control button.

•  Examine the status image, name, and address of the currently controlled drive.

Fig. 4

Page 12: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 12/16

 

  11

•  Reset the fault and clear the fault logger if an error occurs.

•  Assign given reference frequency (30...70 Hz) into the edit field and click the Set 

Reference button or press Enter  to send it to the drive.

•  Run the drive, and then stop it.

•  Change the motor direction of rotation and run it again.

•  Execute the coast stop.

•  Release the control and ensure the drive is adjustable from the control panel.

3.3 Monitoring the drive acceleration

Control monitoring can be executed using the Monitor toolbar or the Monitor menu. Monitor runs

and collects data by reading cyclically the items from the drive in real time.

•  Take the control.

•  To learn the drive acceleration, choose Parameters in the Browse tree panel and set the

parameter 22.02 ACCEL TIME 1 to a given value (5...15 s) in the Item sets pane.

  In theTrend Settings 

 pane, set X-Axis Length above the given acceleration time.•  To add the motor voltage, current, and frequency for monitoring, choose 01 ACTUAL

SIGNALS in the Browse tree pane and drag the required strings into the Trend Settings  

pane. The numeric images appear in front of the added items. The Delete button lets to

remove unnecessary items from monitoring.

•  To start monitoring, click the Start  or  Continue Monitoring button in the Monitor toolbar

and right away run the drive to the given reference frequency.

•  Wait until the running comes to the end and press the Stop Monitoring button. Then stop

the drive also.

•  To scale the trends swing, use Adapt Y-Axis from the Axis submenu or Autoscale from the

Scaling submenu of the Monitor menu.

•  When viewing the stopped or paused trends, click the drawing area to display the graphcursor, which measures the items.

3.4 Saving and analyzing the results

There are several ways to save trends. The currently displayed trends can be saved from File  

menu into a DriveWindow graph file, exported to a text file, copied to the clipboard, or printed.

•  To process the data, select Export in the Graph submenu and assign a name of the created

.txt  file. Later, open the file by Microsoft Excel , recalculate per-unit data into Volts,

Amperes, and Hertz, and build the required diagrams.

•  To save the trends for continuation the work, select Save As  command and name and

comment the new .dwt  file. Saved graphs are actually incomplete offline workspaces,

which contain just enough information to restore the graphs in the trend display. In the

future, by double clicking or selecting Open command in the Graph submenu the saved

file may be restored.

•  To use the trends in other applications, first copy them into the clipboard using Copy 

Graph command in Edit menu and then paste in the required software.

•  To print the trends when the current printer is unavailable, use Microsoft Office Document Image Writer. 

•  After successful saving the monitor can be cleared by the Clear button and the parameter

22.02 ACCEL TIME 1 returned to 3 s.

Page 13: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 13/16

 

  12

Individual report contents

•  scaled diagrams of voltage, current, and frequency timing traces

•  mutual voltage/frequency diagram built using the voltage and frequency data

•  output voltage/current diagram built using the voltage and current data

•  conclusions concerning the results estimation and explanation

Page 14: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 14/16

 

  13

Using Digital Oscilloscope TPS2000 Series 

1. Viewing a Signal

To quickly display a signal of unknown amplitude or frequency, follow these steps:

•  Power on the oscilloscope.•  To connect the probe, align the slot in the probe connector with the key on the CH 1

BNC , push, and twist to the right to lock the probe in place.

•  Push the CH 1 MENU button.

•  Choose Probe.Voltage.Attenuation.500X and set the switch on the probe to the same

value.

•  Connect the Channel 1 probe tip to the signal and the reference lead to the circuit

reference point.

•  Push the AUTOSET button.

•  To view two signals, connect the second probe to CH 2 BNC and use the CH 2 MENU  

button.

The oscilloscope sets the vertical, horizontal, and triggering controls automatically.

2. Taking Quick Measurements

To measure a time and amplitude, follow these steps:

•  Turn the SEC/DIV knob to display the starting edge of the waveform.

•  At the chosen channel, turn the VOLTS/DIV and VERTICAL POSITION knobs to set thewaveform amplitude to about five divisions.

•  Push the CH 1 MENU button.

•  Choose Volts/Div.Fine.

•  Turn the VOLTS/DIV knob to set the waveform amplitude to exactly five divisions.

•  Turn the VERTICAL POSITION knob to center the waveform; position the baseline of the

waveform 2…3 divisions below the center graticule.

Use readouts in the screen bottom area to know the vertical scale factors of the channels and the

main time base setting.

3. Taking Cursor Measurements

Cursors quickly take time and amplitude measurements on a waveform. To measure a time span,

follow these steps:

•  Push the CURSOR button to display the Cursor Menu .

•  Choose Type.Time.

•  Choose the channel, for example Source.CH1.

•  Choose the Cursor 1 option button to display the first cursor location.

•  Turn the multipurpose knob to change the cursor position.

•  Choose the Cursor 2 option button to display the second cursor location.

•  Turn the multipurpose knob to change the cursor position.

•  Time cursors display  ∆t  (time difference between the cursors), 1/  ∆t (frequency),  ∆V (or

 ∆ I , voltage/current difference), and the times at cursors relative to the trigger in theDisplay Menu .

Page 15: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 15/16

 

  14

To measure amplitude, follow these steps:

•  If the Cursor Menu is not displayed, push the CURSOR button.

•  Choose Type.Amplitude.

•  Choose a channel if necessary.

•  Choose the Cursor 1 option button to display the first cursor location.

•  Turn the multipurpose knob to change the cursor position.•  Choose the Cursor 2 option button to display the second cursor location.

•  Turn the multipurpose knob to change the cursor position.

Amplitude cursors display  ∆V  (or  ∆ I , voltage/current difference between the cursors), and the

signal values in the Display Menu .

4. Taking Automatic Measurements

There are eleven types of measurements available: Freq – frequency, Period – the time of the

first cycle,  Mean – the arithmetic mean amplitude over the entire period, Pk-Pk – the absolute

difference between the maximum and minimum peaks of the waveform, Cyc RMS – rms value of the first complete cycle,  Min and Max – the minimum and the maximum values of the last 2500

points, Rise Time and Fall Time – durations of the first rising and falling edges, Pos Width and

 Neg Width – the first positive and the first negative pulses widths at the waveform 50% level.

The oscilloscope can display up to five measurements at a time. To take a measurement, follow

these steps:

•  Push the MEASURE button to see the Measure Menu .

•  Choose the channel, for example Source.CH1.

•  Push another option button and choose the measurement type, for example, Type.Freq.

•  Push the  Back  option button to return to the Measure Menu  and display the selected

measurement.

Page 16: ACS800 Manual for Laboratory Works

5/12/2018 ACS800 Manual for Laboratory Works - slidepdf.com

http://slidepdf.com/reader/full/acs800-manual-for-laboratory-works 16/16

 

  15

Initial Settings of ACS800 for Laboratory Works 

Index Value

10.01 DI1,2

10.02 DI6,5

10.03 REQUEST 11.01 REF1(REF2*)

11.05 70

16.01 YES

Index Value

16.02 OPEN 

20.01 -70 (-1500*)

20.02 70 (1500*)22.02 3

25.01 OFF 

25.02 0

Index Value

26.03 0

26.04 OFF 

99.02 HAND/AUTO** (T-CTRL*)

99.04 SCALAR

99.10 NO (ID MAGN*)

* Loading drive

** Application macro User 2