5/24/2013 1 In modern cataract surgery, freedom from spectacles is becoming more and more important. To select a patient for Multifocal IOL, One Diopter of Astigmatism is the maximum permissible amount.

Upload: lethuy

Post on 24-May-2018

259 views

Category:

Documents


5 download

TRANSCRIPT

Page 1: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

1

• In modern cataract surgery, freedom from

spectacles is becoming more and more

important.

• To select a patient for Multifocal IOL, One

Diopter of Astigmatism is the maximum

permissible amount.

Page 2: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

2

History of Toric and multifocal

IOLs

El-Massry 2013

History of Toric IOLs

• Shimizu et al 1992: PMMA Toric IOLs.

• 1994 the first Toric foldable, through

3.2mm.

• Since 1996 advancements in Toric IOLs to

improve rotational stability which is the

commonest compliciation in Toric IOLs

Page 3: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

3

PMMA Hydrophobic Acrylic

1949 1994

Sir Harold Ridley

The Ridley IOL

1983 2005

Aspheric Optic

UV and Blue Light Filtering

Multifocal Design

2006

Toric Silicone

2007

Single Piece Acrylic

2001

Aspheric Multifocal

IOLs

Alcon Advances in IOL Technology

2008

ReSTOR® Aspheric +3

IOL

2010

Current Time Period Advances

El-Massry 2013

Ten Brands of toric IOLS are

present in the market

1- Acrysof (Alcon) : (Hydropobic).

2- AF-1 toric (Hoya): (Hydrophobic with PMMA

haptic)

3- Acricomfort ( Zeiss) ( Hydrophilic).

4- Lentis Tplus (Ocuolentis) (Hydrophiclic).

5- Light adjusatable (silicone with PMMA).

Page 4: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

4

Ten Brands of toric IOLS are

present in the market

6- Microsil/torica (Human optics).

7- Morcher 89 A ( Silicone with PMMA

haptic)

8- Staar surgical (Silicone)

9- T flex ( Rayner) ( Hydrophilic)

10- Tecnis Toric ( Abott) ( Hydrophobic)

Four Brands of Toric Multifocal

IOLs are present in the market:

- Acrysof Restor Toric IOL (Hydrophobic).

-Acrylisa Toric (Zeiss) , (Hydropilic).

- M flex (Rayner) , (Hydrophilic) .

- Lentis- M plus (Hydrophilic).

Page 5: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

5

AcrySof® IQ Toric

El-Massry

No astigmatism 1.0 D astigmatism 2.0 D astigmatism

Quality of vision is deteriorated considerably by astigmatism

El-Massry

Page 6: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

6

AcrySof® IQ Toric IOL

El-Massry

›Enhanced image quality: -decrease spherical HOA. -Increased CS

Best rotational stability

Corrects variable degrees of astigmatism

AcrySof® IQ Toric IOL Specifications

Model Number SN6AT3 SN6AT4 SN6AT5 SN6AT6 SN6AT7 SN6AT8 SN6AT9

IOL Plane 1.5 D 2.25 D 3.00 D 3.75 D 4.50 D 5.25 D 6.00 D

Optic Diameter 6.0 mm

Overall Length 13.0 mm

Optic Type Biconvex Toric Aspheric Optic

IOL Powers +6.0 D to 30.0 D

Haptic Angulation 0 degrees (planar)

Haptic Configuration STABLEFORCE® modified L haptic lenses

Suggest A-Constant 119.0†

Refractive Index 1.55

Light Filtration UV and Blue Light

†Provided as guideline only

El-Massry

Page 7: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

7

•Reduction of residual refractive cylinder.

•Improved uncorrected distance visual acuity

•Increased spectacle-independent distance vision

Precise Astigmatism Correction

El-Massry

Reduction of Residual Refractive Cylinder

› 63% of patients achieved ≤0.50 diopters of residual refractive cylinder1. › 87% of patients achieved ≤1.00 diopters of residual refractive cylinder1. †† AcrySof® Single-Piece IOL (SA60AT)

1. Based on unilateral clinical study results (Models SA60T3, SA60T4, SA60T5). See package insert.

El-Massry

Page 8: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

8

Rotational Stability

less than 4º average rotation 6 months after Implantation1.

› 81.1% of patients were ≤5º of intended axis1. › 97.1% of patients were ≤10º of intended axis1.

1. Based on unilateral clinical study results (Models SA60T3, SA60T4, SA60T5). See package insert.

El-Massry

Rotational Stability, continue

2. Alcon Data On File. 3. Linnola RJ, Sund M, Ylönen R, Pihlajaniemi T. Adhesion of soluble fibronectin, lamin, and

collagen type IV to intraocular lens materials. J Cataract Refract Surg. 1999;25:1486-1491.

› STABLE FORCE® lens haptics keep IOL

highly stable and centered in capsular bag2.

› Flexible haptic design provides

optimal placement in capsular bag, regardless of size2.

› AcrySof® lens materials binds to fibronectin.

» Ensures adhesion to anterior/ posterior capsule3

El-Massry

Three factors

Page 9: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

9

AcrySof® IQ ReSTOR® +3 IOL

El-Massry

2005

AcrySof® ReSTOR® +4 Add IOL

2007

AcrySof® IQ ReSTOR® Aspheric +4 Add IOL

AcrySof® IQ ReSTOR® IOL

Background 2008

AcrySof® IQ ReSTOR® Aspheric +3 Add IOL

El-Massry

Page 10: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

10

• Utilizes diffraction to simultaneously create a near and distance focal point independent of pupil size.

Diffractive IOLs

El-Massry

Apodized Diffractive Structure

• 3.6 mm diameter Apodized diffractive region

• Precise reduction in step heights from 1.3 microns to 0.2 microns

• Decreasing steps heights direct more light energy to the distance focal point

• gradual energy blend between powers

• Width between steps determine the add power

• Step heights control the light distribution

El-Massry

Page 11: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

11

Theoretical Light Energy Allocation

Source: 5. Data on file. Alcon, Inc. R&D Technical Report

El-Massry

Overall Spectacle Wear

78%

20%

2% 0%

20%

40%

60%

80%

100%

Never Sometimes Always

% o

f Su

bje

cts

N =138

6 MONTHS POSTOP

ReSTOR +3

Source: AcrySof® IQ ReSTOR® IOL clinical data on file (models SN6AD1 and SN6AD3). Fort Worth, TX: Alcon Laboratories, Inc.

El-Massry

Page 12: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

12

Measured Pinhole Images through Modified ISO Model Eye* to Simulate Night-time Car Headlight Halo

* 0.2 micron spherical aberration at 6 mm entrance pupil

15 arc min

2 arc min

(5 mm Pupil, distance focus, simulate car headlight 18 degree off-axis at 360 meter)

ReSTOR® IOL1 ReZoom Tecnis ZM900

ReSTOR® Aspheric IOL (4.0 add)

ReSTOR® Aspheric IOL (3.0 add)

Crystalens

Source: Data on file. Alcon, Inc. R&D Technical Report 1. Non-Aspheric +4 ReSTOR

El-Massry

Patient Quality of Life

0

1

2

3

4

Before Cataract Surgery

After Bilateral Surgery

Completely Satisfied

Not at all Satisfied

OVERALL PATIENT SATISFACTION WITH (WITHOUT GLASSES)

ReSTOR +3 IOL

Source: AcrySof® IQ ReSTOR® IOL clinical data on file (models SN6AD1 and SN6AD3). Fort Worth, TX: Alcon Laboratories, Inc.

El-Massry

Page 13: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

13

AcrySof® IQ ReSTOR® Toric

El-Massry

AcrySof® IQ ReSTOR® Multifocal Toric IOL

• Single surgical procedure for presbyopia

correction and corneal astigmatism

management

• When used in conjunction with the online

calculator enables the surgeon to greatly

reduce residual cylinder, eliminating the

need for additional surgical procedures.

El-Massry

Page 14: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

14

Anterior Apodized Diffractive Aspheric Surface

Same design as current AcrySof® IQ ReSTOR® +3.0D Add Multifocal IOL (Model SN6AD1)

9 apodized diffractive steps for +3.0D add power and balanced light energy management

Negative 0.1 micron spherical aberration factor corrects for the positive spherical aberration of the cornea

AcrySof® IQ ReSTOR® Multifocal Toric IOL Product Evolution

Posterior Toric Lens Surface

Same design as current AcrySof® Toric IOLs

Posterior toric axis marks

Posterior toric surface

Allows the lens to correct pre-existing corneal astigmatism

El-Massry

Model Numbers

Product

Characteristic

SND1T2 * SND1T3

SND1T4

SND1T5

Optic Type Biconvex Apodized Diffractive Aspheric Toric

Material Ultraviolet and blue light filtering Acrylate / Methacrylate Copolymer

Optic Diameter (mm) 6.0

Overall Length (mm) 13.0

Haptic Angle 0°

IOL Powers

(spherical equivalent

diopters)

Current availability is 16.0 to 25.0 Diopters (0.5D increments)

IOL Cylinder Powers (IOL

Plane) 1.00 D 1.50 D 2.25 D 3.00 D

IOL Cylinder Powers (Corneal Plane)†

0.68 D 1.03 D 1.55 D 2.06 D

Add Power +3.0 D

Add Power at Spectacle

Plane +2.5 D

Suggested A-Constant 118.9

†Based on average pseudophakic human eye .

AcrySof® IQ ReSTOR® Multifocal Toric IOL (Model SND1TT) Product Characteristic Summary Table

El-Massry

Page 15: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

15

Day 3

Page 16: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

16

Calculator Output Screen Summary:

• Appropriate SND1TT IOL model

– Check calculated models with effective Dpt and

Cyl range availability

• Spherical equivalent lens power

• Optimal axis location of the IOL

• Anticipated Residual Astigmatism

AcrySof® IQ ReSTOR® Multifocal Toric IOL Online Calculator Summary

El-Massry

Page 17: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

17

DAY 1

Source: Data on file. Alcon, Inc. R&D Technical Report

Theoretical Total Energy Balance

El-Massry

Page 18: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming

5/24/2013

18

Key Considerations For Optimal Outcomes

• Marking the eye:

– Prior to surgery mark the operative eye with at least two

reference points (e.g. three o’clock and nine o’clock positions)

while the patient is sitting upright to prevent cyclotorsion.

• Temporal incisions are recommended.

• Target the lowest possible residual astigmatism.

Patients with significant postoperative astigmatism

>1.0D may not achieve optimal visual outcomes.

El-Massry

Review Article

Toric and Multifocal toric IOLs

Journal of cataract and refractive surgery

April 2013

Page 19: acrysof® Iq Restor® Toric - Esoirsesoirs.com/files/ESOIRS2013/ESOIRS2013-presentations/021001.pdf · 5/24/2013 1 •In modern cataract surgery, freedom from spectacles is becoming