acm619 service manual (s)

Upload: sn-dayanidhi

Post on 06-Apr-2018

291 views

Category:

Documents


31 download

TRANSCRIPT

  • 8/2/2019 Acm619 Service Manual (s)

    1/14

    ACM619 Service Manual

    I ACM619 Anesthetic Workstation

    AIR

    AIRN2O

    N2OO2

    O2

    CYLINDER

    AIRN2O AIR

    CYLINDER

    N2O

    CYLINDER

    O2

    PIPE LINEPIPE LINE

    kPax100kPax100kPax100kPax100kPax100kPax100

    PIPE LINE

    O2

    PCV

    IPPV

    O2+

    O2 PpeakkPa

    f/min

    PEEPkPa

    MVL

    0

    %

    1

    VOL5

    4

    3

    0

    VOL%

    4

    5

    2

    13

    2

    300

    600

    900

    1200

    1500

    Fig.6102535 Front1

  • 8/2/2019 Acm619 Service Manual (s)

    2/14

    II Gas Part

    1 Operation Principle

  • 8/2/2019 Acm619 Service Manual (s)

    3/14

    1. O2 CylinderOptional 2.O2 Pipeline 3.N2O CylinderOptional 4.N2O Pipeline 5.AIR

    CylinderOptional6.AIR Pipeline 7 Pressure gauge0-1000kPa 8 Pressure gauge0-25000kPa

    9. pressure regulatoroutput 400kPa10. pressure regulatoroutput 400kPa 11. pressure regulatoroutput

    400kPa 12 filter 13. unidirectional valve 14. relief valve840kPa 15. alarm shift valve

    120kPa-200kPa 16. alarm 17. gas storage cylinder

    18. pressure reducing I valveoutput 270kPa 19. N2O driven regulate valve 20. ventilator pressure reducing I

    valve

    Output 25010kPa21. proportion valve 22. electromagnetic valve 23. ventilator pressure reducing II valve

    output 505kPa 24. expiration unidirectional valve 25. atmosphere 26. relief valve6kPa 27. linage

    flowmeter (0.1-2L/min;2.5-10L/min) 28. support circuit 29. evaporator0.2%-5% 30. relief valve38kPa

    31. O2 flush valve (25L/min-75L/min) 32. Venturi valve33. Inspiration unidirectional valve 34. Expiration

    unidirectional valve 35.O2 concentration sensor 36.flow sensor 37. pressure sample interface 38.

    patient 39. absorber1000mL2 40. manual/ mechanical ventilator control 41. APL valve

    0-6.5kpa 42.30mm exhaust outlet 43. Manual bag 44. Bellows 45. relief valve (0.2kPa-0.3kPa)

    46.30mm exhaust outlet

    Operation Principle compendium

    ACM619 Anesthesia Machine has O2, N2O and Air three gases and according to CE

    standards, it has two air supply ways, Serial number is 135 is the Cylinder inlet(back-up),

    and the serial number246 is the Pipeline inlet (common usage), the pressure ofPipeline gas

    should set as 280kPa~ 600kPa.After the gas into the machine, it will be through the filter (12)

    to be filtrated the impurities and grains, and the Pressure gauge (78)will display the gas pressure,

    cylinder gas1,3,5is reduced to 400 kpa bypressure regulator(9,10,11), connecting with the relief

    valve(14), then paralleling with the pipeline gas by unidirectional valve (13)

    The gas supply failure alarm system is made of alarm shift valve for oxygen15,alarm

    unit (16) , gas storage cylinder(17) and unidirectional valve(13), When the gas pressure is higher than

    200 kPa, the gas will be through the unidirectional valve (13) into the gas storage cylinder(17), until

    the gas storage cylinders pressure is the same with the gas pressure. If the pressure controlled shift valve

    (broken line) is on the position 1(left), the gas is closed. When the gas pressure is less than 200kpa, the

    pressure controlled (broken line) shift valve is on the position 2(right), the gas of the gas storage cylinder

    flows through the shift valve and blows the alarm unit. The alarm time is not less than 7 seconds.

    one part of the O2 gas enters into the ventilator, then through the primary pressure reducing

    valve I (20), reducing the pressure to 25010kPa, the main gas enters into the

    bellows(44)through proportion valve(21) to supply dynamic for breathing. The auxiliary gas

    passes through electromagnetic valve (22) and secondary pressure reducing valve (23) , enters into

    expiration valve (24), meanwhile, the expiration valve connects with atmosphere (25) and relief valve

    (26). When inspiration, ventilator controls the proportion valve (21), and turns on in accordance with the

    set tidal volume, at the same time the electromagnetic valve (22) is connecting, and the gas pressure is

    reduced to 505kPa through the secondary pressure reducing valve. The controlling

    expiration valve 24 is closed, the main gas is separated with the atmosphere and the driven

    gas is produced to press bellows to inspiriting. When expiration, proportion valve is closed,

  • 8/2/2019 Acm619 Service Manual (s)

    4/14

    meanwhile, electromagnetic valve (22) is closed, too. The gas from auxiliary gas channel is released by

    electromagnetic valve, then expiration unidirectional valve opens, the driven gas is released to

    atmosphere by expiration unidirectional valve. If accident happens during inspiration, say, too much gas

    supply, pressure increasing, in order to protect the patient, relief valve 26 can automatically open to

    reduce the pressure when pressure grows to 5.5~6kPa.

    O2 is reduced to 270kPa by primary pressure reducer 18 and enters into the linkage

    flowmeter (27). At the same time, the pressure reduced O2 controls the N2Os in an out by N2O driven

    adjusting valve. When the O2 pressure is lower than 200kPa~120kPa, N2O is cut off; when the O2

    pressure is higher than 200kPa~120kPa, the N2O enters into the linkage flowmeter (27). And the air

    source also enters into the linkage flowmeter(27) by primary pressure reducing valve (18)

    The pressure reduced O2, N2O, Air is mixed in the linkage flowmeter 27, then passes

    through by-pass valve (28) and enters into the vaporizer; after mixing with the anesthestic agent in

    accordance with the set proportion, the mixed gas enters into the circle absorber through Venturi valve(32)

    (it is consisted of the elements from 33 to 43). At the same time, the pressure reduced O2

    also enters into the Venturi valve by O2 flush valve (31), and then into the circle absorber.

    The working procedures of the circle absorber (it is consisted of the elements from 33

    to 43).

    Setting the switch of Manual/ mechanical to mechanical, and the ventilator works to

    pressure the bellows to let the gas entering into the circle absorber, and then this gas passes

    to the patient 38 through unidirectional valve (33). when expiration, pressure of bellows releases,

    pressure of patient lung release CO2 by expiration unidirectional valve (34) and make a chemical reaction

    with sodalime which is in the absorber (39). if the too much gas compensation quantity or other reasons

    cause the end expiratory pressure higher than the set pressure of spill valve 0.2 kPa ~0.3 kPa, the

    spill valve opens to release the gas Setting the switch of Manual/ mechanical to manual,

    anaesthetist uses the manual bag to supply gas to patient, and make the gas exchange.At the

    same time, in order to protect the patient, you should set the safety pressure of APL valve.

    When the pressure is higher than the set value, exhaust gas will be exhausted from exhaust

    outlet (42)

    2 Common Malfunction1Circle absorber leakage

    First to test the vaporizers connecting to see whether it leaks.

  • 8/2/2019 Acm619 Service Manual (s)

    5/14

    Checking the leakage of circle absorber

    Under the manual condition, using a silicone tube to connect the inspiratory outlet andexpiratory outlet.

    Opening the flow meter slowly to increase of the pressure of the absorbers gauge to 3KPa,

    then close the flow meter

  • 8/2/2019 Acm619 Service Manual (s)

    6/14

    Observing the pressure gauge for 2~3 minutes. If the indicator of the pressure gauge doesnt

    drop, it expresses that the circle absorber doesnt leak. Testing the folding bag and spill

    valve of the bellows assembly, ff the circle absorber leaks, please check whether the

    absorber is tightened.

    2Spill valves adjustmentDissembling the bellows assembly

  • 8/2/2019 Acm619 Service Manual (s)

    7/14

    Unscrew the three screws and take down the cover, adjust the place of the pressure nut to change the

    pressure value.

    Rotatethe nut anticlockwise to increase the spill valves pressure value; rotate nut clockwise to reduce

    the pressure value. After adjustment, assemble the bellows again. After the machine works well, observe

    the airwaypressure gauge on the circle absorber, and the end-expiratory pressure should be 0.2~0.3KPa.

    Screw

    Pressure adjustment

    nut

  • 8/2/2019 Acm619 Service Manual (s)

    8/14

    Electric Part

    1.Operation Principle

    Anesthesia machine wiring diagram

  • 8/2/2019 Acm619 Service Manual (s)

    9/14

    ib. diagram, the left side power supply220V or 110V via X1entrance ventilator, through magnetism ring L1, L1

    for shield interfere comes from of power supply. Live wire (L), Null Line (N), in series 8amp. fuses, connect with

    switch S1, earth wire(E) connect with the machine shell. From of S1 lead two line, L and N. One joint 2A slow fuse

    via capacitance C1, C2 which fixed on the machine shell connect with the switch U1, the another one joint 2A quick

    fuse connect X2, X3, X4 together with E which one on the shell.

    Power switch U1 output two wires, Earth wire GND and 24V, connect with electrical outlet X2 on the control board

    A1.

    Control board A1

    Electrical outlet X7 output 5 lines, D shape outlet which designed for signal interference shielding, two for oxygen

    concentration, two for manual switch, five for coupler test board.

    Oxygen concentration Manual switch Coupler test board

    1 2 1 2 1 2 3 4 5

    GND Signal GND +5V Signal Signal Signal GND

    Electrical outlet X8, three lines connect to buzzer board.

    1 2 3

    +5V GND Control board

    Electrical outlet 5, connect to Proportional Valve Controller:

    1 2 3 4

    Control board GND +12V +24V

    Gray Black Red Purple

    Proportional valve controller output:

    Black Red

    On-off +12V

    Electrical outlet 4, connect to electromagnetic valve

    1 2

    Control board +12V

    Electrical outlet X11, 26 core cable connect to the LED display board A2.

    1 2 3 4 5 6

    +5V +5V GND GND PVC setting VCV setting

    7 8 9 10

    Inspiration pressure setting I:E setting Respiratory rate setting TV setting

    11 12 13 14

    Confirm button Optical encoder OC1 Optical encoder OC2 Display board D2 Chip

    select signal

    15 16 17 18

    VCV Lights Display Board D1 Chip

    select signal

    PCV Light SCLOCK Bus signal

  • 8/2/2019 Acm619 Service Manual (s)

    10/14

    19

    SDA Bus signal

    Line 20~26 Empty.

    Electrical outlet X6

    1 2 3GND D1 Serial ports TXD end D1 Serial ports TXD end

    Electrical outlet X10

    1 2 3 4 5 6 7

    -12V Empty Empty DGND 5VD EGND 12V

    Electrical outlet X11

    1 2 3 4

    GND Pressure analog output Flow analogy signal output Flow analogy signal

    output

    LCD Display

    LED Display board A2

    12V Input

    Inverter

    High-voltage output

    Cable connected to upper board.

    26 core cable X2 connect with A1X1

    6 core cable X3 connect with shuttle button.

    10 core X2 connect with upper board X4

    Connect with the membrane board

  • 8/2/2019 Acm619 Service Manual (s)

    11/14

    Electrical outlet X36 core

    1 2 3 4 5 6

    GND Confirm button GND shuttle button OC1 shuttle button OC2 +5V

    Electrical outlet X2 connect with electronic board electrical outlet X4 (10 core)

    1 2 3 4 5STBY Indicator light STBY MENU Mute RESET

    setting setting setting setting

    6 7 8 9 10

    LCD display setting Confirm Shuttle button OC1setting Shuttle button OC2 setting GND

    Electrical outlet X4 (19 core)

    1 2 3 4 5

    STBY Indicator light STBY setting MENU setting Mute setting RESET setting

    6 7 8 9 10

    LCD display setting TV Setting Respiratory rate Setting I:E Setting Inspiration pressure setting

    11 12 13 14 15

    PCV function IPPV function PCV indicator light IPPV indicator light Empty

    16 17 18 19

    Empty Empty GND GND

    Electronic board

    Electrical outlet X6, serial interface, used for machine debugging.

    The others tie-in please refers to the circuit board instructions.

  • 8/2/2019 Acm619 Service Manual (s)

    12/14

    2Parts testingProportional valve

    Yellow/Yellow Yellow-green/Yellow together put in the right side jack.

    Down

    Up

    ENTER

    1Connected proportional valve and the controller as the indication above;2Turn on ventilator, long press ENTER for 5 seconds to enter setup menu;3Into the INP (input) menu, by pressing the down button to INP = 0 ~ 5V, the display interface to the END option,

    press ENTER exit.

    4Into the OUT (output) menu, then enter the UALU (body type) menu, by pressing the down button to set to 2833,the display interface to the END option, press ENTER exit;

    5Into the OUT (output) menu, then enter the ADJ (adjust coil) menu, by pressing the down button to LO (low) =250mA, so that HI (high) = 400mA, the display interface to the END option, press ENTER exit;

    6Into the CUT (screen) menu, by pressing the down button to make CUT value = 0, the display section to the CUToption, press ENTER exit;

    7Into the CUT (screen) menu, by pressing the down button to make CUT value = 0, the display section to the CUToption, press ENTER exit;

    8Other menu options do not do changeThe display interface to the END option, press ENTER to return control window.

    Pressure settings:

    Preheated for 5 minutes after boot, anesthesia machine work for the state, anesthesia ventilator control panel DIP

    switch S1, 4,3,2 segment is set to on, 1 is set to off, then anesthesia ventilator tidal volume window adjusted value for

    the pressure, the digital pressure gauge to the pressure sampling port, debug as follows:

    1) To the pressure sampling port on the air, anesthesia ventilator display 2.00 ~ 2.02, or adjust control panel

    potentiometer R4;

  • 8/2/2019 Acm619 Service Manual (s)

    13/14

    2) Measurement of the container through the sampling port pressure to 7kPa, anesthesia ventilator showed a value of

    8.98 ~ 9.02, or adjust control panel board potentiometer R3;

    Repeatedly adjust R3, R4, to meet the requirements of airway pressure display.

    Adjustment is completed, the control panel DIP switch S1 of a segment set to on.

    3Fault:Tidal volume are not accurate (gas line problem need to exclusion before electrical test) as a pressure sensor,

    two pressure sampling tubes often stagnant water, pressure will affect the transmission of tidal volume are

    not accurate, so first check whether the sampling tube does have water, and clear it. Analysis carried out

    after confirmation.

    Three problems lead the Tidal volume not accurate: 1) pressure sensor failure; 2) the proportional valve

    controllers problem; 3) The proportional valve is not working properly;

    Pressure sensor test:

    N12

    After the boot to be working properly, measuring 3 feet N12 voltage, consistent with Table 1, differential pressure

    sensor to determine the normal sampling:

  • 8/2/2019 Acm619 Service Manual (s)

    14/14

    No. TVml Voltagemax

    1 250 2.53

    2 350 2.71

    3 500 2.87

    4 650 2.95

    5 800 3.046 1000 3.14

    Table 1

    Proportional Valve Controller Detection:

    Output current value

    Measure input voltage

    Start working properly, set the different tidal volume, simultaneous detection of the input voltage proportional valve

    controller, the controller shows the current value of observation, compared with, if found in the following table (Table

    2), the control panel (A1) and proportional valve is working properly.

    No. TV Input Voltage Current mA

    1 250 1.58 298

    2 350 1.84 305

    3 500 2.21 316

    4 650 2.56 327

    5 800 2.95 338

    6 1000 3.55 355

    Table 2

    Black screen, black screen of Death;2.Check the back of LCD panels before and after the 220V inverter and +12 V, power supply socket X1 PC board

    situation.

    3. membrane board key failure:First test the use of all buttons, the screen is normal operation; further determined that under fault phenomenon

    mask the problem, the PC and control board failures.

    PC board controls: buttons, "home screen", "menu", "confirm", "", "", "STBY", "Mute 2min", "reset." All

    screen display.

    Control Panel Control: Key "PVC", "VCV", "suction pressure", "smoking calls more than", "respiratory rate",

    "tidal volume", "confirm."

    By checking the LED display X2, X4, X1 socket, to confirm cause of the malfunction.