DEVELOPMENT OF NEUTRON RADIOSCOPY AT THE SLOWPOKE-2 FACILITY AT RMC FOR THE INSPECTION OF CF1 88 FLIGHT CONTROL SURFACES DÉVELOPEMENT DE LA RADIOSCOPIE NEUTRONIQUE AVEC LE &ACTEUR SLOWPOKE-2 AU CMR POUR L'INSPECTION DES SURFACES DE CONTROL DE VOL DU FC188 A Thesis Submitted To the Faculty o f the Royal Military College of Canada BY Thomas Robert Chalovich In Partial Fulfilrnent of the Requirements for the Degree of Master of Engineering in Chernical and Materials May 2000 O This thesis may be used within the Department of National Defence, but copyright for open publication remains the property of the author.

Upload: others

Post on 19-May-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

DEVELOPMENT OF NEUTRON RADIOSCOPY AT THE

SLOWPOKE-2 FACILITY AT RMC FOR THE INSPECTION OF

CF1 88 FLIGHT CONTROL SURFACES

DÉVELOPEMENT DE LA RADIOSCOPIE NEUTRONIQUE AVEC LE &ACTEUR

SLOWPOKE-2 AU CMR POUR L'INSPECTION DES SURFACES DE CONTROL DE VOL

DU FC188

A Thesis Submitted

To the Faculty o f the Royal Military College of Canada

BY

Thomas Robert Chalovich

In Partial Fulfilrnent of the Requirements for the Degree of

Master of Engineering in Chernical and Materials

May 2000

O This thesis may be used within the Department o f National Defence, but copyright for open publication remains the property of the author.

Page 2: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

National tibrary u*m of Canada Bibliothèque nationale du Canada

Acquisitions and Acquisitions et Bibliographie Services services bibliographiques 395 WeUington Street 395. rue Wellington OttawaON KlAON4 OnawaON K 1 A W Canada Canada

The author has granted a non- L'auteur a accordé une licence non exclusive licence aUowing the exclusive permettant à la National Library of Canada to Bibliothèque nationale du Canada de reproduce, loan, distribute or seil reproduire, prêter, distribuer ou copies of this thesis in microform, vendre des copies de cette thèse sous paper or electronic formats. la forme de microfiche/nlm, de

reproduction sur papier ou sur format électronique.

The author retains ownership of the L'auteur conserve la propriété du copynghî in this thesis. Neither the droit d'auteur qui protège cette thèse. thesis nor substantial extracts fiom it Ni la thèse ni des extraits substantiels may be printed or otherwise de celle-ci ne doivent être imprimés reproduced without the author' s ou autrement reproduits sans son permission. autorisation.

Page 3: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

ACKNO WLEDGMENTS

1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major

W.J. Lewis, for their guidance, advice and assistance during the course of this research. for

without their help this thesis would not have been possible.

1 would also like to thank Ms K.S. Nielson, Director of the SLOWPOKE-2 Facility, for

her assistance and patience throughout this project. 1 am grateful to Mr. Orv Francescone who

instnicted me on the technical and practical aspects of neutron radiography, giving me a solid

foundation for this research. Further thanks goes to Leslie Chalovich, for being my initial editor

and for holding down the "fort" while 1 was pre-occupied with this joumey.

Page 4: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

ABSTRACT

Chalovich, Thomas Robert. M. Eng (Chernical and Materials). Royal Military College of Canada. May 2000. Development of Neutron Radioscopy at the SLOWPOKE-2 Facility at RMC for the Inspection of CF1 88 Flight Control Surfaces.

Supervisors: Dr. L.G.1 Bennett Major W.J. Lewis

An ongoing investigation has identified a problem with the flight control surfaces on the

CF1 88 Hornet aircraft in that they suffer from water ingress into the composite layers and the

aluminium honeycomb core causing hydration and wat er entrapment , respectively . Neutron

radiography is being used to indicate the presence of water in these components. However, the

low neutron flux at the image plane at the SLOWPOKE-2 Facility at the Royal Military College

(RMC) causes relatively long exposure tirnes for neutron radiography. As a faster alternative,

neutron radioscopy, which is the combination of charged coupled device (CCD) or analogue

camera technology with a neutron source to produce a digital image on a persona1 computer, has

been investigated. With the advent of high quality and efficient CCD carneras, improved

software enhancement techniques and powerfûl cornputers, it is now possible to combine these

technologies to shorten the exposure times relative to neutron radiography for some applications

that do not require high resolution.

A methodical and intensive study was conducted to detennine the CCD camera's

characteristics and response to the elhination of light leaks, changes to carnera temperature,

varying exposure t h e , different diffuser characteristics and variable neutron flux levels. The

Page 5: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

availability of two research reactors with neutron beams (SLOWPOKE-2 at RMC and the

Breazeale Nuclear Reactor at Pemsylvania State University) allowed for testing at different

neutron f lues which further validated the response of the carnera system. Further testing of the

system was carried out to confirm its ability to detect water in a honeycomb test piece. Final

testing required imaging a CF188 rudder with known water ingress to detemine the camera's

ability to image the presence of water in a realistic situation. Finally, a cost comparison between

neutron radiography and neutron radioscopy was made.

Analysis of the neutron radioscopy (camera) system installed at the SLOWPOKE-2

Facility at RMC has shown that it is capable of producing equivalent neutron images of flight

control surfaces at significantly shorter exposure times. In fact, by analysing the digital image as

a histogram of pixel count vs intensity value, a clear indication of the presence of moisture was

made evident. Moreover, comparison to higher neutron intensities confhned that the

SLOWPOKE-2 Facility has the capability to perfom adequate neutron radioscopy. As a result of

this investigation, it is recommended that neutron radioscopy be used as the initial inspection

method for the CF 1 88 flight control surfaces, followed by neutron radiography when warranted.

Page 6: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Chalovich, Thomas Robert. M. Eng (Chernical and Materfals). Collège militaire royal du Canada. May 2000. Dévelopement de la Radioscopie Neutronique avec le Réacteur SLOWPOKE-2 au CMR pour l'Inspection des Surfaces de Control de Vol du FC188

Superviseurs: Dr. L.G.1 Bennett Major W.J. Lewis

Une recherche continue a identifié un problème avec les surfaces de commande de vol

sur l'avion CF 18 8 Hornet c'est-à-dire qu'elles absorbent l'eau dans ses composantes en

composites et le noyau de nid d'abeilles d'aluminium causant de l'hydration et de l'occlusion,

respectivement. La radiographie neutronique est employée pour indiquer la présence d'eau dans

ces composantes. Cependant, le bas flux de neutron au plan d'image en service au SLOWPOKE-

2 du Collège Militaire Royale (CMR) entraîne des temps relativement longs d'exposition pour la

radiographie neutronique. La radioscopie neutronique comprend la combinaison de la

technologie analogique d'appareil-photo avec une source de neutron et le dispositif couplé chargé

(CCD) pour produire une image digitale sur un ordinateur personnel. Avec l'arrivée d'appareils-

photo et de CCD de haute qualité, les techniques améliorées de perfectionnement de logiciel et

les ordinateurs puissants, il est maintenant possible de combiner ces technologies pour raccourcir

le temps d'exposition aux neutrons pour certaines applications qui ne requiert pas Ia de haute

résolution. Une étude méthodique et intensive a été entreprise pour déterminer les

caractéristiques de de l'appareil-photo CCD et la réponse à l'élimination des fuites légères, les

changements de température de l'appareil-photo, le temps variable d'exposition, les différentes

Page 7: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

vii

caractéristiques du diffuseur et les niveaux variables du flux de neutron. La disponibilité de deux

réacteurs pour des recherches scientifiques avec des faisceaux neutroniques (SLOWPOKE-2 au

CMR et le Breazeale à l'université de 1' Etat de la Pennsylvanie) a permis des tests à différentes

intensités du flux de réacteur ce qui a validé la réponse du système d'appareil-photo. Davantage

de tests du système ont été effectué pour confirmer la détection d'eau dans un morceau d'essai de

nid d'abeilles. Le test final comprenait l'imagerie d'un gouvernail de direction CF188 avec une

absorption d'eau connue afin d'évaluer la capacité de l'appareil-photo de détecter la présence

d'eau dans une situation réaliste. En conclusion, une comparaison des coûts entre la radiographie

neutronique et la radioscopie neutronique ont été faits. L'analyse du système de radioscopie

neutronique (appareil-photo) installé au seMce SLOWPOKE-2 du CMR a démontré qu'il peut

produire des images satistaisantes de piéces d'avion à des tamps dlexposition beaucoup plus

courts.

Page 8: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

... Vlll

TABLE OF CONTENTS

Page

.................................................................................. ABSTRACT iv

.......................................................................... LIST OF FIGURES xi

... ........................................................................... LIST OF TABLES ml

........................................ LIST OF SYMBOLS AND ABBREVIATIONS xiv

................................................................................... CHAPTER 1 1

1 . Introduction ............................................................................... 1 ............................................................... 1.1 Non-Destructive Testing 1

1.2 Radiography .............................................................................. 1 ......................................................... 1.3 Status of Neutron Radioscopy 4

...................................................................... 1.4 Goal of this Thesis 6

................................................................................... CHAPTER 2 8

2 . Principles of Radioscopy ............................................................... 8 2.1 Neutron Sources ........................................................................ -8

.................................................................... 2.2 Neutron Attenuation 9 ........................................................................ 2.3 Imaging Devices 11

................................................................. 2.3.1 Scintillator Screens 11 ....................................................................... 2.4 Camera Assembly 12

2.4.1 Charged Coupled Devices (CCD) Camera ...................................... 13 2.5 Image Formation ........................................................................ 17 2.5.1 Hurnan Perspective ................................................................. 17 2.5.2 Scintillator Screen .................................................................. 18 2.5.3 CCD Camera ........................................................................ -20

2.6 Image Error .............................................................................. 22 . . 2.6.1 Statistical Error ..................................................................... -22 2.6.2 Systematic and Random Error ..................................................... 24

2.7 Digital Image ............................................................................ 26

Page 9: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

................................................................................. 2.7.1 Format 27 ................................................................. 2.7.2 Image Enhancement 29

...................................................................... 2.8 System Resolution 38 2.8.1 Total Unsharpness ................................................................... 38 2.8.2 Measuring Total Unsharpness ..................................................... 40

.............................................................. 2.8.3 Resolution ftom MTF 41 2.8.4 Sensitivity Indicator (SI) ........................................................... 45 2.9 Image Represent ation ................................................................... 47

............................................................................ 2.9.1 Histogram -47 2.9.2 Intensity Difference Method ....................................................... 49

.................................................................................. CHAPTER 3 51

3 . Experimental Procedure ............................................................... 51 3.1 Equipment .............................................................................. 51 3.2 Experiments ............................................................................. 54 3.2.1 Installation and Optimisation of the CCD Camera System at RMC ......... 56 3.2.2 Camera Parameters ........................................................... 57 3.2.3 Resolution Measurement ..................................................... 59 3.2.4 Water Ingress Simulation ......................................................... 60 . .....................................................................* 3 2.5 CF 1 8 8 Rudder 64

................................................................ 3.2.6 Comparison to Film -66

................................................................................... CHAPTER 4 67

................................................................. . 4 Results and Discussion 67 ................................. 4.1 Installation and Optimisation of Diffuser at RMC 67

........................................................... 4.1.1 Reduction of Light Leaks 67

.......................................................... 4.1.2 Configuration Parameters 70 ...................................................................... 4.2 Carnera Parameters 72

............................................... 4.2.1 Dark Image - Temperature Change 72 4.2.2 Dark Image - Diffuser Changes ................................................... 74 4.2.3 Unified Images - Neutron Flux Effects .......................................... 76 4.2.4 Dynamic Range .................................................................... -79

................................................................. 4.3 Resolution Calcuiation 80 4.3.1 System Resolution Calculations .................................................. 81 . . . 4.3.2 System Sensitiwty ................................................................ 84 4.4 Water Ingress Simulation ............................................................. 85 4.4.1 WaterIngress ........................................................................ 85 4.4.2 Sïmulated Water .................................................................... -102 4.5 CF188 Rudder ........................................................................... 108 4.6 ComparisonofRadioscopy to Radiography ......................................... 118 4.6.1 Resolution ........................................................................... 118 4.6.2 Time and Cost Study ................................................................ 120

Page 10: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

................................................................................... CHAPTER 5 125

................................................................................ 5 . Conclusion 125 ............................................................................ 5.1 Introduction 125

.............................................................. 5.2 Installation o f Hardware 125 ........................................................... 5.3 Charactensation of System 126

.............................................. 5.4 Development of Neutron Radioscopy 128 5.5 S u m m r y ................................................................................ 130

................................................................................... CHAPTER 6 131

......................................................................... 6 . Recomrnendat ions -131

APPENDICES

A Characteristics of the SLOWPOKE-2 and Breazeale Reactors .................... A- 1 ..................................................................... B Image Specincations B-1

....................................... C System Optimisation and Camera Parameters C-1 ............................................. D Error Calculation of Intensity Difference D- 1

......................................................... E Enhanced Images of Test Piece E- 1 ....................................... F Neutron Radioscopy Images of CF 1 88 Rudder F- 1

....................................................... G Cornparison of Nomialised Data G- 1

Page 11: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

LIST OF FIGURES

Page

........................................................... Radiography Configuration 2 ............................................................ Radioscop y Configuration 3

............... Mass Attenuation Coefficients as a Function of Atomic Number 10 ................................................................... Single Pixel of CCD 14

..................................................................... CCD Architectures 16 .................................... Radiography and Radioscopy Image Formation 19 . . . . .................................................................... Digitisation Process 21

.............................................................................. Grey Scale -28 ..................................................... Image Enhancement Techniques 29

...................... Example of a Contrast and a Gamma Conversion Function 31 ............................................................... Geometric Unsharpness 39

................................................ 2.10 Ut Measurement for GD Knife.edge 41 ....................................................... 2.1 1 Example of Frequency per mm 42 ..................................................... 2.12 Sample of ERF and LSF Graphs 43

............................................................................. MTF Graph -44 ................................................... Modified ASTM E545-75 Type A 46

................................................................ Example of Histograrn -48 .................................................................... Histogram Example 49

..................................................... Beam Stop and Diffiser at RMC 52 ............................................................ Diffuser used at Penn State 53

................................................. Camera Assembly and Control Box 54 ........................................................... Flow Chart of Experiments 55

............................................ Example of Dynamic Range Histogram 59 .................................................... Construction of a CF1 88 Rudder 60

...................................................... Test Piece with Water at RMC 62 ........................................................... Test Piece with PE at RMC 63

..................................................... Test Piece with PE at Penn State 64 Film Layout on CF1 88 Rudder ....................................................... 65

................................................. Full Intensity Scale of Optimisation 68 .................. Enlarged Histogram Before and M e r Light Leak Reductions 68

................................................ Statistical Parameters of Dark Image 70 ...................................................... Effects of Camera Temperature 73

................... Peak intensity Values of Dark Images for Different Difisers 75 ............... Peak Intensity Values of Unified Images at Various Flux Values 77

Page 12: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

xii

............................ 4.7 Dynamic Range of S ystem at Various Power Settings 79 ................................ 4.8 MTF Graph for 4-Minute Exposure taken at RMC 82

4.9 LinePair Gauge ........................................................................ 83 ................... 4.10 Effects of Varying Exposure Time on a Single Ce11 of Water 86

....................... 4.11 Intensity Difference of a Single Ce11 at 0.14 rnL of Water 89 ............................... 4.12 Effects of Varying Exposure Time for Three Cells 91

......................... 4.13 Intensity Difference of Three Cells at 0.16 mL of Water 93 ................................ 4.14 Effects of Camera Gain on a Single Ce11 of Water 95

.................................. 4.15 Effects of Camera Gain on Three Cells of Water 96 ................................................ 4.16 Fout Minute Exposure of Test Piece 98

............................................. 4.17 Results fkom Equalisation of Histogram 99 .............................................................. 4.18 Results f?om Erode Filter 100 .............................................................. 4.19 Results fkom J Flat Filter 101

........................... 4.20 Results of Varying Flux Levels for a Single Cell of PE 103 ............................ 4.2 1 Results of Varying Flux Levels for Three Ceils of PE 105

.............................................. 4.22 Exposure 2 of Rudder ffom Penn State 108 ......................... 4.23 Sequence of Exposure 5 Images taken at RMC (10 kW) 110

..................................... 4.24 Histogram of Exposure 5 nom RMC (10 kW) 111 ................... 4.25 Sequence ofExposure 5 Images taken at Penn State (10 kW) 112

............................... 4.26 Histogram of Exposure 5 nom Penn State (10 kW) 113 .................. 4.27 Sequence of Exposure 5 Images taken at Penn State (100 kW) 114

.............................. 4.28 Histogram of Exposure 5 from Penn State (1 00 kW) 115 4.29 Best Results of Exposure 5 fiom (a) RMC at 10 kW, @) Penn State

............... at 10 kW, and (c) Penn State 100 kW, with image enhancements 116 .............................. 4.30 Norrnalised Graph at Two Different Power Settings 117

................................................................ 4.3 1 Resolution Cornparison 119 .................................................. 4.32 Tirne Study Flow Chart for Rudder 123

Page 13: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

LIST OF TABLES

Page

..................................................... 2.1 Designation for Sensitivity Levels 46

........................................................... 4.1 Statistical Data of 30 Images 71

...................................................... 4.2 Configuration Image Parameters 71 .

.......................................... 4.3 Resolution Variables for the NRS at RMC 81

.............................. 4.4 Hole Values fiom Modified ASTM E545-75 Type A 84

............................................... 4.5 Interna1 Component s of CF 1 88 Rudder 109

................................................ 4.6 Resolution Variable for Film at RMC 118

..................... 4.7 Cornparison of Capital Cost of Radioscopy and Radiography 121

........................................................................... 4.8 Cost per Image 122

.............. 4.9 Operational Cost and Inspection of Radiography and Radioscopy 124

Page 14: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

LIST OF SYMBOLS AND ABBREVATIONS

AECL

ASTM

ATESS

BNR

CCD

ERF

m

FWHM

HFRNR

IFD

IP

1 Q 1

I(V

LPG

LSF

U A )

MTF

Atomic Energy o f Canada Lirnited

Arnerican Society for Testing and Materials

Aerospace and Teiecornmunicat ions Engineering Support

Squadron

Breazeale Nuclear Reactor

Charged Coupled Device

Edge Response Function

Fast Fourier Transfonn

Full Width Half Maximum

High Frame Tate Neutron Radiography

Image File Directories

Imaging Plates

Image Quality lndicators

Light fiom an object

Line Pair Gauge

Line Spread Function

Energy distribution

Modulation Trans fer Function

Page 15: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

NAA

NBT

NCA

NDE

NDT

NRS

PE

Penn State

QE

QETE

RESC

Ri.WC

RMS or 00

SI

SD

T m

TRIGA

Neutron Activation Analysis

Neutron Beam Tube

Neutron Camera Assembly

Non-Destructive Evaluatior

Non-Destnictive Test ing

Neutron Radioscopy System

Po lyet hy lene

Pennsylvania State University

Quantum E fficiency

Quality Engineering and Test Establishment

Radiation Science and Engineering Centre

Royal Military CoUege

Root Mean Square

Sensitivity Indicator

Standard Deviation

Tagged image File Format

Training Research and Isotope production reactor by

General Atomics

Total Unsharpness

Geometric Unsharpness

Inherent Unsharpness

Neutron Flux

Page 16: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

CHAPTER 1

1.1 Non-Destructive Testing

Non-destructive testing (NDT) and non-destructive evaluation (NDE) are the

generic tems for material evaluation methods used to assess the quality of a part or piece

of material without impairhg its hture usefihess. Non-destructive testing has become a

highly developed and well-utilized science within the a*ame and power plant

manufacturing industries. Modem aircraft with their complex structures, multiple and

unpredict able load paths, 10 w engineering safety factors and extended li fe times, combine

to produce a requirement for non-destructive inspections. The materials now introduced

into the manufacturing process, which can create lighter and stronger components, have

made the inspection of these components more complex and difficult. There are six

standard NDT techniques: visual inspections, liquid penetrant inspection, magnetic

particle inspection, ultrasonic testing, electromagnet ic eddy current and radiograp hy. It is

the last one, radiography, that is the focus of this thesis.

Radiography is a NDT technique in which an image can be taken of the interna1

characteristics of a part by utilizing either gamma, X-ray or neutron radiation. The layout

of a radiographie procedure is straightfonvard as indicated in Figure 1.1. An object is

Page 17: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

placed in the path of a beam of radiation (X-ray, gamma or neutron) and a recording

medium is placed behind the object to be inspected. Some radiation will be absorbed or

scattered by the object, but most radiation will travel through and impinge on the

recording medium, producing a latent image. When the recording medium is processed,

the object will be visible. Intenial or material defects of the object will be indicated on the

recording medium as different densities relative to the background density. For gamma or

X-ray radiography, the source is either a radioisotope gamma source or an X-ray tube

while the recording medium is X or gamma ray sensitive or light-sensitive f i l a For

neutron radiography, the neutron source is a nuclear reactor, a radioisotope or an

accelerator (with a neutron producing target) while the recording medium is a converter

screen used to capture the neutrons and transfer the latent image onto X or gamma ray

sensitive or light-sensitive film The preferred neutron source is a nuclear reactor because

of the large neutron intensities available (which reduces exposure tirne) uniess other

conditions are paramount, such as location or any unique requirement of the object to be

inspected (e.g., intact aircraft at an airfield).'

l maging Plane

I Object

Figure 1.1 - Radiography Configuration

Page 18: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Neutrons interact with the nuclei of atoms rather than with the orbital electrons, as

do X or gamma radiation, and therefore, each elemental nuclide has its own characteristic

neutron cross section, showing no pattern in cornparison to the increasingly uniform

attenuation with miss nurnber for X and gamma radiation (see Section 2.2, Figure 2.3). In

fact, there is a slight reverse or complementary trend towards light elements for neutrons.

The major advantage of neutron radiography is its ability to reveal light elements such as

hydrogen contained in materials such as corrosion products and water.

With neutron radioscopy, the same source is used as with neutron radiography, but

the recording medium is different. It c m be any combination of a scintillation screen, a

photosensitive screen, an image intensifier, an analogue video camera, or a charged

coupled device (CCD) camera. A typical layout for radioscopy with a scintillation screen

and CCD camera is shown in Figure 1 .S.

Neutron Source

Aperture /

Scintillation Scryen

I

Ob ject _.: . ....... _.:.:.

Lens & A CCD Carnera

Figure 1.2 - Radioscopy Configuration

Page 19: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

1.3 Status of Neutron Radiosco~v

Neutron radioscopy utilising a CCD camera is a relatively new technology as it

was fia applied to neutron radiography in 1990.' Advancements in CCD cameras and

cooling technology have made the use of CCD cameras both affordable and practical, and

therefore, their use has proliferated in the last five or six years. Although the CCD camera

is a new application in regards to neutron radiography, the development of a CCD system

has been thoroughly in~esti~ated.' As well, the CCD camera performance has been

analysed and the effects of c hanging the camera temperature, the exposure time and the

background radiation levels have been well documented.' Nurnerous investigations into

different material compounds that gWe the scintillation screens their basic characteristics

have been carried out to determine the matenal having the largest target for neutrons, the

best resolution and the bnghtest light output.sv 6 p As well, screens that respond to thermal,

epithermal and fast neutrons have also been investigated to determine their practicality.8

The majority of facilities that have installed a CCD camera system for neutron radioscopy

have done so for research purposes.

One of the largest area of research for neutron radioscopy utilising CCD camera

has been in the field of three-dimensional tomography, as the digital format lends itself

toward this form of compiled imagery. A practical use for tomography bas been the

detection of corrosion in turbine blades of jet engines. 9

It has been suggested that neutron radiography could be used for many aircraft

applications; specifically, it has been suggested that aluminium, honeycomb and

composite structures could be inspected for corrosion, moisture and adhesive defects. l0

Page 20: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Research has indicated that neutron radiography has a good sensitivity to detect corrosion

in aircraft grade aluminium. ' l Neutron radiography has been used to inspect corrosion in

KC-135 and DC-9 wingsl' and water ingress in CF188 flight control surfaces. l3 The

McClellan Air Force Base had the facilities to inspect complete aircrafi (FIA- 18 and F-

1 1 1) using neutron radiography and radioscopy (with Vidicon tube cameras). Although,

McClellan Air Force Base has been closed and the facility to inspect aircraft has been

decommissioned, the TRIGA (Training Research and Isotope production reactor by

General Atomics) reactor facility, at tbis location, with its five beam ports has the

capability to inspect large individual aircraft components.

At present, the practical use of neutron radioscopy using a CCD camera system has

been limited to only a few aircrafi applications. For example, one such application has

been the use of neutron radioscopy to investigate corrosion inside Iap joints of a KC 135

aircraft. '" In the recent past, a concern was raised over the airworthiness of the CF1 88

Homet, as initial indications pointed to possible problems that may exist in some of the

structures made fiom honeycornb construction. The Canadian Forces had aircrafi 188733

cornpletely inspected at McClellan Air Force Base with both real-time radioscopy and

conventional (film) radiography using both X-rays and neutrons.15 The results of the

inspection revealed 93 anomalies that included moisture ingress, ce11 corrosion, damaged

honeycomb core, foreign object material, voids and repaired areas. The problem areas

tended to be associated with components that could be removed fiom the aircrafi relatively

easily. Afier this inspection, the major emphasis then focused on how the components

Page 21: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

that indicated potential problems could be inspected off the aircraft in-house rather than by

the very costly method of a complete aircrafi inspection at such a facility.

After an initial investigation to detemine which NDT technique would be suitable,

the Royal Military College W C ) , in conjunction with ATESS and QETE, was tasked to

do a comparison test between through transmission ultrasonic testing and neutron

radiography. While obtaining commissioning information, a small sample of CF 1 88 flight

control surfaces was inspected, which included thirteen rudders, eight ailerons, eight

trailing edge flaps and eight inboard and outboard leading flaps, l6 and the results of th&

test indicated that neutron radiography was supenor in detecting water ingress. The

SLOWPOKE-2 Facility (Safe Low Power c(K)ntical Experirnent) at RMC was

commissioned and then tasked to inspect a total of 10 sets of aircraft flight control

s d a c e s using neutron radiography. As well, due to the low neutron flux at the image

plane in the RMC Facility, a studyl' was completed in which faster neutron radiography

imaging rnethods and a quantitative measure of the water trapped in the honeycomb were

investigated to aid in the inspection of these components.

1.4 Goal of this Thesis

The goal of this thesis was to install, characterise and develop a near reakirne

radioscopy system at the neutron radiography system at RMC in order to expedite the

inspection of CF1 88 aircraft composite flight control surfaces for water ingress.

Installing, characterishg and developing a near real-the radioscopy system

requires a knowledge of the production and behaviour of neutrons, the operation

characteristics of a CCD carnera and scintillation screen, the formation of an image, the

Page 22: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

accumulation effects of system error, the application of digital enhancements and the

calculation of system resolution.

M e r these concepts were understood, the camera system had to be tested in a

methodical way to determine how changing the basic parameters of the system affected

the image qualit y. Light leaks, camera temperature, exposure time, diffuser c haract erist ics

and neutron flux levels were al1 varied in different experiments to understand their

interdependency and their effects on the camera system. However, these tests and

experiments did not indicate how the system would perfom in detecting water ingress in a

honeycomb structure.

A honeycomb and composite test piece with varying amounts of water and high-

density polyethylene (PE) was used to discover the system's sensitivity to detect small

amounts of water. This test piece was also used to determine the optimum exposure tirne

and the digital enhancement technique for the system.

Finally, a CF188 nidder with known water ingress problems was imaged to

validate al1 of the previous experiments and to test the performance of the camera system

in a realistic situation. This component was used to provide a cornparison with film-based

neutron radiography.

Page 23: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

PRLNCIPLES OF RADIOSCOPY

2.1 Neutron Sources

The three recognised neutron sources available for neutron radiography or

radioscopy are a radioactive source, an accelerator, and a nuclear reactor." These

sources offer a wide range of neutron intensities, operational complexity and special

features. The advantages and disadvantages of the these sources are well documentedg

and, depending on the requirements, it is generally accepted that a nuclear reactor is the

b a t neutron source with respect to neutron flux (4) (defineci as the number of neutrons

passing through a unit area per second).20 Two nuclear reactor facilities were used to

explore the parameters associated with radioscopy throughout this study. The first was

the SLOWPOKE-2 Facility at Royal Military College (RMC) in Kingston, Ontario,

Canada and the second was the Breazeale Nuclear Reactor (BNR) at Pemsylvania State

University (Penn State), Penns ylvania, United States. The characteristics of both

reactors used in this study are descnbed in Appendix A.

Page 24: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

2.2 Neutron Attenuation

Attenuation of absorbed neutrons is exponential and is dependent on both the

neutron energy and the target material. The attenuation of neutrons in a beam follows

an exponential law:"

= L~WP)P'

where:

1 = emergent radiation intensity, I,, = incident radiation intensity, p = linear attenuation coefficient (cm-'), p = density, and x = material thickness

A cornparison of the mass attenuation for most of the elements for thermal

neutrons and X-radiation (Figure 2.1) shows that, for neutrons, each nucIide has a

varying attenuation which has no pattern or relationship to density or other physical or

chernical properties. In fact, there is a slight reverse or complernentary trend towards

light elements for neutrons. Neutron attenuation varies drastically between

neighbouring elements, and therefore, a single neutron radiograph can easily show

many different materials.

For most neutron radiography, the energy range used is between 0.0 1 to 0.3 eV

and the neutrons are referred to as thermal neutrons since their average energy is equal

to that of the material in the moderator. "

Page 25: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure 2.1 - Mass Attenuation Coefficients as a Function of Atomic Number

For neutrons, it is more wnvenient to have a relationship between the

coefficient and the cross section of the material. The linear attenuation coefficient can

where:

N = number of nuclei per cm3 of attenuating material q = total cross section (cm2) equal to the sum of the absorption (aa) and

scattering (a,) cross sections

The nuclear cross section is a vaiue that indicates the probability that an

interaction will take place between a neutron and one or more nuclei in the target

material. The cross section can be considerd as a measure of an effective target area

with units of square metres or barns (1 O"* m2).24 There are severai types of cross

sections, but the two of principal interest for neutron radiography are the absorption and

scattering cross sections, with the total cross section being their sum. The absorption

Page 26: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

and scattering cross sections are the probabilities, respectively, that a neutron is

absorbed or scattered by the nuclei in the target rnaterial.

2.3 Imarring; Devices

For neutron radioscopy, a method is required to attenuate the neutrons and,

through a reaction, produce a sufficient amount of light to detect them with an

electronic or photosemit ive imaging s ystem. There are several methods used for

detecting neutrons to fonn an image, but the three most common rnethods used are

image intemifiers, imaging plates (IP) and scintillation screens of which the latter one

is used in this study. "

Originally, the term scintillation was applied to an optically clear homogeneous

slice of a single crystal of a rnaterial that fluoresces when irradiated by X-rays. The

terrn scintillation now takes on a more generic meaning and refers to al1 materials that

fluoresces whether clear, opaque, homogeneous, non-homogenous or constnicted fkom

a vast composition of elements. A typical screen is composed of a substrate, binding

matrix and phosphorescent rnaterial. The substrate can be constructed of plastic, glass

or aluminium. The substrate of choice for neutron radiography is aluminium as

attenuation to neutrons for this material is minimal, making it vimially transparent and

thus allowing a maximum number of neutrons to collide with the phosphorescent

rnaterial. The binding matrk should also be transparent to neutrons to eliminate further

neutron attenuation. There are several different compositions of phosphorescent

Page 27: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

materials that have been tested for resolut ion and int ensity , with evaluat ions indicat k g

that Ti-based scintillation screens showed clear superiority? The mechanism for

producing light fiom neutron interactions is similar for most screens. This mechanism

will be discussed as it specifically applies to a 6~i-based scintillation screen having the

composition of 6 ~ i ~ : ~ n ~ : ~ u (NE 427 @) as used in this study.

Thermal neutrons interact with Lithium due its large cross section of 9 IO barns

to produce an alpha particle (4~e") and tritium ( 3 ~ daughter products.'7 The energy

fi-om the alpha particle is deposited into zinc sulfide (ZnS), an efficient phosphor that

produces visible light. The copper elenient acts as a wave length shifier to produce

light in the yellow-green region, which has an average wavelength of 525 na For an

optimised screen thickness, the 6 ~ ~ : ~ n ~ : ~ u screen can have a minimum spatial

resolution of 200 The characteristics of the 6 ~ i ~ : ~ ~ : ~ g screen is identical to

those of the 6 ~ i ~ : ~ n ~ : ~ u screen except that it has a average wavelength of 440 nm

caused by the silver (Ag) element. The 6 ~ i ~ : ~ n ~ : ~ g screen has a light output per

stopped neutron of 8.5 x 104 photonslneutron and a maximum detection efficiency of

15% for a given thi~kness.'~ For comparable screen parameters, it can be assumed that

the 6 ~ i ~ : ~ n ~ : ~ u screen has similar characteristics.

2.4 Camera Assemblv

The electronic medium used for recording an illuminated image fi-om an image

intensifier or scintillation screen is either a Vidicon or a SIT video camera or a CCD

video camera, of which the latter is used in this study.

Page 28: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

2.4.1 Charged Coupled Device (CCD) Camera

CCD cameras are the most prominent devices used in neutron radioscopy due

to their high resolution, high quantum efficiency, wide spectnim response, low noise

and linear characteristics. The primary component of the carnera, fiom which the name

of the canera is derived, is the CCD silicon chip. The lem, electronic equipment and

woling unit support the CCD chip. The lem focuses and gathers incident light ont0 the

chip while the electronic equipment transfers the intensity values and location fkom the

CCD chip - either by an analogue or digital signal - to an output device.

The CCD chip is made fkom a matrix of individual collection sites. These sites

are called pixels. A single pixel can range f?om 6 pn to 30pm in size and, when

grouped together into a square array, the format can range from a 5 12 x 5 12 matrix to a

4096 to 4096 mat ri^.^' Pixel and matrix size determines the overall physical size of the

array and the area coverage possible.

Growing a thin layer of silicon dioxide ont0 the base silicon chip and then

depositing a conductive polysilicon gate structure over the oxide foms a collection site.

Applying a positive voltage to the gate creates a depletion zone between the silicon

dioxide and silicon interface, Figure 2.2.

Page 29: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Incoming Light

1 1 1 1 Elecrrical Connection

Figure 2.2 - Singie Pixel of CCD

A depletion zone or space charge zone is an area where charges are rare;

therefore, a charge void has been created in the material. In the depletion zone,

incoming photons are converted into electrons (called photoelectrons) and stored in the

depletion zone, also known as a potential well. Incident light projected on a collection

site for a fixed time allows an accumulation of photoelectrons into the potential we1lO3l

If an image in the f o m of incident light is projected onto a CCD chip, then each

individual potential well will accumulate photoelectrons according to the intensity of

the light hitting each collection site. At the end of the integration time, the image is

transfmed f?om the CCD chip. The process of charged coupling is used to move the

stored charges in the potential wells. The photoelectrons associated with a potential

well are called a charge packet During charge coupling, charge packets at a collection

site are transfmed fiom site to site by rnanipulating the gate potentials. The separation

of individual charge packets is preserved during the hansfer process. Three diEerent

methods of charge coupling are utilised, each haWig advantages and d i~advan ta~es .~~

Page 30: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

To remove an image fiom a CCD chip, each charge packet must be transferred

to a serial register. There are three possible CCD architectures, each having a different

configuration for t r a n s f e g the charge packets to the seriai register, as shown in

Figure 2.3. A Full Frame CCD moves a row of charged packets into the serial register

and the packets are moved one at a time to the output amplifier. Downloading al1 the

charge packets to the amplifier takes a considerable amount of tirne, typically up to

twenty seconds. Altemately, a Fnune Transfer CCD chip is divided into two areas.

The first area is the normal configuration of a CCD, while the second area is masked

nom the incident light. The charge packets are transferred to the masked area, thereby

making the CCD chip ready for another exposure. A row of charge packets can then be

loaded into the serial register independent of the actions being taken by the other CCD

area. This configuration is faster at downloading the image, but this increase of speed

cornes at a significant cost. Either the resolution is reduced, due to half of the area

available, or the camera becomes more expensive to purchase, due to the increased size

of the CCD chip. Thirdly, the Interline Transfer CCD has small vertical areas masked

fiom the incident light. The charge packets are t r a n s f d to these masked areas

simultaneously and then loaded by rows into the serial register. This architecture has

fast d o m transfer thes , but the resolution is reduced and the masked areas may

interfere with picture resolution. The CCD camera used in this thesis was configured

with a full fiame transfer CCD chip.

Page 31: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Output Serial Clocks Output Seriai Clocks Output Amplifia AmpUer I

vvv vvv Amplfier

b

Dlrcctlon of Direction of ~ a r a l ~ d shirt ~ a r a l l ~ r shut

RiI1 RPme CCD Frime M u CCD -fer CCD

Figure 23 - CCD Architectures

2.4.1.1 CCD Chip Pdormance

The CCD chip's performance c m be defined in t m s of its dynamic range,

linearity, Quantum Efficiency (QE) and transfer efficiency. The dynarnic range of a

CCD chip is the ratio of the full saturation charge to the system noise and is expressed

in units of electrons. System noise is dependent on the quality of the system and can

vary fiom a few electrons for high performance systems to several thousand for

inexpensive video equipment. The saturation charge is detemiined by the well

capacity, which is related linearly to pixel size.

Linearity is the relationship between the amount of illumination on a single

pixel and the response to that ill~mination.~~ Linearity is lirnited to a working range of

intensities. At low illumination, threshold phenornena reduce linearity significantly

and, at high illumination, saturation of the potential well degrades the response.

Page 32: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Quantum efficiency is the measure of the effectiveness on an imager to produce

electronic charge fiom incidence photons. QE is dehed as the average nurnber of

detected photons divided by the average number of incident photons. CCDs have

efficiency ranges fiom 40% to 60%, compareci to photographie emulsions whose

efficiencies are presently 2% to 4%." The sensitivity of a CCD is limited by the gating

structure, which interferes and absorbs some of the incoming light. One solution to

eliminate this effect and increase the QE for a given CCD is to thin the silicun on the

backside of the CCD chip and have the incident light corne nom this side. A CCD chip

that has been thinned is refmed to as a backside iiluminated CCD. The CCD camera

used in this thesis is a backside illuminated CCD chip for the reasons stated above.

Transfer Efficiency is the rneasure of effectiveness of a CCD to iransfer charge

packets fiom one register to another. Only very small transfer inefficiencies cm be

tolerated due to the large quantities of charge packets that pass through the large sized

arrays. CCD registers have a typical transfer efficiency of 0.999990." h w

temperatures ( las than -1 00°C) and small nunber of charges (less than 1 O00 electrons)

adversely affect the transfer efficiency.

2.5 lmaae Formation

Visible light is electromagnetic radiation that is perceived by humans. Light is

quantified by the spectral energy distribution L(k), a fhction of wavelength (A) that

range fkom 350 to 780 nm. Light received ~ o m an object can be defïned by either the

Page 33: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

reflectivity or transmissivity of the object p(h) and the incident energy distribution UA)

and express4 as:

I(V = ~ ( h ) x UA) (2.3)

The range of illumination over which the human eye c m operate is roughly 1 to 10'' or

10 orders of magnitude.

The human visual system is not stnctly the structure of the eye, but a

combination of receptor (eye) and interpreter (brain and visual experiences). Human

perception is sensitive to luminance contrast rather than the absolute luminance.

Bnghtness (also refmed to as apparent brightness) of an object is the perceived

luminance and is dependent on the luminance of the surrounding are~i.'~ Humans are

only capable of discriminating between 64 levels of shades of grey; given large number

of grey shades, a human cm no longer distinguish between the different shades." At

fewer than 16 shades of grey, the contours between the grey shades become a hindrance

in analysing an image. It will be shown in Section 2.7.1 that the CCD camera cm

quanti@ many more levels of grey than is possible with the human eye.

2.5.2 Scintillator Screen

The image formation fkom neutron radioscopy is fundarnentally different from

neutron radiograp h y. For radiography, after the neutrons interact (b y either absorption

or scatter) with an object, they are absorbed by the conversion screen and a conversion

electron fiom the conversion screen produces a latent image on the film. When the

negative fihn is processed, the area associated with the location of the object is less

exposed, translating to a lower film density. The radiographer sees this area as having a

Page 34: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

lighter shade of grey than the surrounding area. In radioscopy, neutrons interact with

an object in the same manner with fewer neutrons hitting the scintillating screen.

Therefore the screen is not exposed in this area, but the remaining area of the

scintillating screen will be luminous at a wavelength of approximately 500 to 575 m.

The radiographer will view the area associated with the matter as a darker shade of grey

compared to the surrounding area. Essentially, a positive image, rather than a negative

one, is directly obtauied. Figure 2.4 is a drawing iIlustrating the results of radiography

(using film) and radioscop y (using a scintillating screen).

rp Radiography

S cre en Processed Film

Radioscopy

Scintillating Screen Compiaer Screen

Figure 2.4 - Radiography and Radioscopy Image Formation

Page 35: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

2.5.3 CCD Carnera

An image presented to a sensor is made up of a continuous spectnim of

fiequencies and amplitudes of energy. A process interprets these energies so that the

image can be represented in a form that can be used by the sensor. For a CCD carnera,

digitisation is the method of mapping a continuous function into a discrete matrix with

a f i t e number of eiement~.'~ Digitisation of a continuous image wnstitutes an

enonnous loss of information, since the information conceming the grey values is

reduced fkom an infinite to a finite number of points. This Iarge loss of information

relative to the original object is one source of error in the system.

Digitisation of an image includes three steps as shown in Figure 2.5. These

steps include image formation, sampling and limitation to a finite window. Image

formation Unplies that the illumination intensity is collected over a specific area of the

sensor rather than at an exact location. The area is dependent on the size of the sensor.

For a CCD camera, the illumination intensity is integrated over each individual pixel.

Sampling takes the intensity values of the given area and assigns an average value of

intensity to grid points that outline this area of interest. The intensity values associated

with the area are then removed, leaving only grid point values. Sampling occurs in a

CCD when the illumination intensity is reduced to a single point, which is associated

with a single pixel. Sampling is physically done in the CCD when the collected

photons are converte- to an electrical signal. In th-, the image is still infinite in

size, as the original image is a continuous fûnction. The last process is called "limiting

to a finite window", and allows the image to be represented as a 2D matrix. In practical

Page 36: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

ternis, the CCD array acts as the f i t e window and the pixel size lùnits the resolution

of the image.

Image Formation :

Sampling :

Limitation J

A Quant ization . ,'

Figure 2.5 - Digitisation Process

Once the image is digitised, the pixels are related to continuous grey scale

values. Due to the limitations of a cornputer, this continuous grey scale must be

mapped ont0 a lirnited nurnber of discrete grey values. This process is called

quantization. Quantization is a mapping process, but tends to be confusecl by users as it

is often used when referring to the bit size of an image. Although there are several

quantization scales, only two will be of interest to this process. The fint is an eight bit

image, which has 2* = 256 grey scale, and sixteen bit image which has 216 = 65536 grey

scale. The latter is prirnarily used as this large grey scale produces b e r intensity

levels, which produce intensity values that closer represent the original image intensity

values.

Page 37: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

2.6

not

Image Error

Any measurement process introduces a probability that the value generated may

represent the actual item being measured. Two important classes of error exist,

statistical and systematic =or, into which al1 errors associated with neutron radioscopy

can be divided.

2.6.1 Statistical Error

Statistical error descnbes the scattering of measured values if the same

measurement is repeated. A measurement of the width of the distribution of a function

gives the statistical error while the centroid gives the mean measured value. An

accumulation of data indicates probability and confidence levels.

In neutron radioscopy, the measured quantity at a certain point in the image

plane is the luminescence. Originally, the observed radiance is charactensed by a

probability density function, p@, which is the probability of observing a grey value

(g). Quantization transfonns the radiance values to a discrete value and therefore,

statistically, the data must not be handled as a continuous but as a discrete fiinction, &.

A measured quantity governed by a random process, as is the case for observing grey

values, g, is referred to as a random variable.

Two basic parameters describing a random variable are the mean and variance

value. The mean value for discrete finite values is defïned as:

where p = mean grey value, and g, = grey value

Page 38: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Since a finite number of measurements are taken, the determination of the mean

remains an estimate with uncertainty, as it depends on the form of the probability

density function, in other words, the type of random process. Many processes with

contïnuous random variables (before quantization) can be adequately described by the

nomal or Gaussian probability distribution.

The variance 2 is a measure of the extent to which the measured values deviate

fiom the mean value. The variance for discrete values with probabilities that are al1

equal is define. as: 39

Given a normal distribution, there is an expectation that a single point will lie in

the confidence interval. A confidence interval is a part or whole value of a variance

that is added and subtracted to the mean value. Examples of the three most used

confidence intervals are:

p + l c + 68.27%

+ 20 + 95.45% C L - f 30 + 99.73%

It is only the middle confidence interval that is used throughout this report. The percent

value is referred to as the confidence level and implies that a person is confident by X

percent that a random value will fa11 within a given range. For example, using two

variances either side of the rnean, a point will have a value between the upper variance

and the lower variance 95.45 % of the time. The level of significance is the probability

that a given solution to a problem is wrong. The level of significance is directly related

Page 39: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

to testing a statistical hypothesis and, d e r extensive tests of the hypothesis, a

probability is calculated indicating the potential failure of that hypothesis. A table

indicating the varied values for confidence levels and level of significance will indicate

the number of times the hypothesis must be tested4'

2.6.2 Systematic and Random Error

Systematic error occurs when the mean value of a measurement is beyond

statistical error margins of a true value. A precise but inaccurate measurement is

encomtered when the statistical error is low, but the systematic emor is hi&. Items that

cause significant systematic mors for radioscopy are electronic noise, dark current,

thermal noise, blooming and radiation.

Electronic noise associated with a CCD camera include noise fkom the

amplifier, the CCD (fixed pattern noise) and f?om charge transfer. Al1 of these types of

noise tend to be very small and are consistent for a given CCD camera. An amplifier

converts the photoelectrons to voltage values after leaving the register. The noise

associated with the amplifier tends to be very small and not significant. Fixed pattern

noise is due to the mors associated with CCD chip fabrication as the QE and charge

collection volume for each collection site varies. Unifoxm illumination of a CCD chip

causes site to site non-uniformity in collecting charges. Consistent and carehl

manufacturing will make variations in the site response small, thus making the error

negligible. The error associated with the charge transferring is a small residual charge

left in a site when the charge packets are moved to the register. Current CCD's have

Page 40: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

achieved charge transfer efficiencies greater than 0.99999 and therefore it is reasonable

to assume that al1 charges are transferred and that this noise source is negligible.

Dark current is a leakage of charge into a site due to thermal agitation of the

crystalline lattice of the image dete~tor.~' If the dark current is not rernoved by reading

the element, it will accumulate and evenîdly saturate the pixel. Dark current is

dependent on the integration time and temperature of the CCD. At room temperature, a

standard CCD will saturate in only a few seconds. Dark cwent halves in value for

about every 6OC &op in temperature and are negligible at -1 00°C. " The magnitude of

this error is limited by the hardware specifications. This noise will be depicted as a

broad intensity spectnim fiom one pixel count to several hundred per site. The limiting

factor of dark current is that the quantity of dark current varies between sites, therefore

producing a non-unifonn image.

Thermal noise is a random component of the signal and is dependent on the

mean value of the dark ~urrent.'~ If a single site or pixel is measured over a single

exposure taken in total darkness and compared to a mean value determined from a large

number of identical exposures, the single value will differ slightly fiom the mean value.

The standard deviation or Root Mean Square (RMS) of total electrons accumulated for

a given site is represented as:

where a, = RMS, and Nt = number of total thexmal charges created

Bloorning is an observable effect due to light saturation of a single or multiple

sites. A site subjected to excessively strong illumination can accumulate photoelectron

Page 41: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

charges that exceed the well capacity and leak into adjacent sites. The characteristic

image of over saturation is a bright spot with a halo effect surrounding this spot. In

addition, the number of charges accumulated in a saturated well can be such that its

contents cannot be emptied in one or more tramfers. A trail starting at the saturated

point appears in the direction of the transfer of rows and is the signature of blooming.

Cosmic and background radiation also introducw noise to an image. The site is

not a filter and therefore cannot distinguish between visible light and other fonns of

energy. The CCD is susceptible to cosmic radiation over long exposure times (100-500

seconds) at low temperatUres." A CCD camera operating in a nuclear reactor

environment will be subjected to higher background noise due to the presence of

gamma and neutron radiation. Radiation affects individual sites and can increase the

photoelectron amount significantly. Gamma radiation is visible on an image as one to

several pixels that have been saturated and therefore have a high value of intensity.

Neutron radiation will produce a line of saîurated pixels in an image. Noise due to

radiation is difficult to reduce given the relatively high levels of radiation encountered

at the camera location and, therefore, it is necessary to place the CCD camera in a

properly shielded location near the bearn.

2.7 Digital Image

With the advent of fast and large capacity cornputers and the development of

scanners and CCD carneras, the concept of digital images has become common. A

digital image diffas significantly fiom conventional film, in that a digital image is an

Page 42: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

image constructecl £iom nurnbers and can be manipulateci in real time to adjust a poor

image or reveal latent information.

2.7.1 Format

AAer the camera produces the image, it can be foxmatted into several different

data types or classes. The data type or class indicates the number of bits used to

represent the pixel intensity value in an image and is referred to as its pixel depth or bits

per pixel. The term "bit" stands for binary digit and is the fundamental placeholder in a

binary system of numbers. A combination of bits equates to a number in the base 1 0

system. Eight bits equal one byte and the byte is the foundational unit when r e f e g

to digitd cornputers. The classes for image formats include:

Byte or Grey Scale 8 - 1 byte per pixel, 2* = 256, values in the image will range fiom O to 255;

Short integer - 2 bytes per pixel, values range fiom -32768,32768;

Unsigned 16 or Grey scale 16 - 2 bytes per pixel, values range fiom O to 65535;

Long Integer or Grey Scale 32 - 4 bytes per pixel, values fiom -232, z3*- 1 ;

Colour 16 - 2 bytes per pixel, with 5 bits assigned to the intensity of each of these colours, red, green & blue.

Colour 24 or RGB 24 - 4 bytes per pixel, with 8 bits assigned to the intensity of each of these colours, red, green, bIue.

There are several formats in which the image may be saved to a amputer

storage medium. A requirement exists so that no data are lost when saved to a file;

therefore, the file format must not have any built-in data compression. To be

Page 43: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

transportable to different types of software, a recognised standard format is required. A

file format called Tagged Image File Format (TIFF) is one such format meeting al1 the

requirements. A TIFF file format is comprised of three basic components:

1. An image file header which indicates the order and version of TIFF file and the location of the first IFD;

2. The image's pixel information which indicates location and intensity value; and

3. The Image File Directones (IFD) which are a collection of descriptors that indicate the characteristics of the image.

There are more details signifiant to the TIFF file format including further

explanations on the three components making up this file fomat.''

The image is represented as a 3D matrix compnsed of values that indicate the x

and y location and the intensity level for every pixel at the location. The intensity level

depends on the data type or class. An image format of grey scale 16 produces intensity

values that correspond fiom black, havkg a value of zero, to white, having a value of

65535, as indicated in Figure 2.6.

Intensity Values

Black White

Figure 2.6 - Grey Scale

The location of the pixels on the image correlate to the location of the pixels

fiom the CCD camera. Therefore the image size is identical to the nurnber of pixels in

Page 44: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

the CCD chip; a CCD with 5 12 x 5 12 pixels will produce an image that has 5 12 x 5 12

pixels and be approximately 512 kilobytes in size and will require half a megabyte of

storage space.

2.7.2 Image Enhancernent

The goal of image enhancement is not to increase inherent information in the

data, but to accentuate certain features for subsequent analysis or image display. Three

basic methods or classifications associated with dancing an image are m o d i m g the

intensity level, applying a spatial filter and manipulating the image fiequency (as

indicated in Figure 2.7). No universal grouping or classification has been officially

recognised, as other persons have chosen to categorise these methods differently.46

Image Enhancement I

htensity Level Spatial Filters Frequency Filtering

t

Brightness

Gamma Correction High P a s Fiat Fielding

Thrrsholding m i 0 Flat

Histograms J Flat

Figure 2.7 - Image Enhancernent Techniques

Page 45: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

2.7.2.1 Intensitv Level

M o d i m g the intensity can enhance the overall appearance of an image by

changing the brightness, the contrast, the Gamma correction factor, the thresholding

values or the histognim shape (refmed to as histogram equalisation). uitensity is

related to the amount of light within a given image exposure and the distinction

between the shades of dark and light grey.47

Brightness (also referred to as apparent brightness) of an object is the perceived

luminance and is dependent on the luminance of the mounding area4* Luminance is

a measurable quantity that is not dependent or related to the object's surrounding

luminance. Mathematically, increasing the brightness of an image requires that each

pixel be universally increased to a higher value. Increasing the brightness requires that

a constant numerical value be added to the value of each pixel - the greater the value,

the larger the increase in brightness. Lowering the brightness requires that a constant

value be subtracted fiom each pixel value. In mathematical terms, if a given pixel

f?om an image is defined as Il, then increasing the brightness of that pixel is as follows:

I I f = II + Constant (2-7)

Contrast indicates the degree of difference between the brightest and darkest

components of an image. An image has poor contrast if the pixel values are within a

small range; conversely, an image with good contrast will have a large range associated

with the pixel's values. Changing the contrast of an image requires that each pixel

value be scaied by a contrast value, which redistributes the intensity values over a wider

or narrower range. Changing the wntrast is accomplished by multiplying each pixel

Page 46: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

value by a constant, which by definition is a simple linear conversion. The formula for

changing the cuntrast of a single pixel (II) is as follows:

1,' = Il x Constant (2-8)

This method can also expressed as a h c t i o n and graphed as indicated in Figure 2.8.

Contrast and Gamma F unctbns

(Black) O 8192 16384 24576 32768 40960 49152 57344 65536 (White)

Original lnterrsity Value

Figure 2.8 - Example of a Contrast and a Gamma Conversion Function

Applying a Gamma correction factor is a non-linear method of changing the

contrast. The Gamma correction factor is used to enhance the contrast in either the

darkest or lightest area of the image. This is accomplished by applying pre-detennined

shaped non-linear curves to the existing pixel values. These curves skew the

conversion of the pixel values so that a dark image cari be lightened to enhance the

cuntrast in the darker areas or a light image can be darkened to enhance the contrast in a

Page 47: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

lighter area as shown in Figure 2.8. The characteristics of the curves can be linear,

logarithmic, exponential or several other predetermined shapes.

Thresholding is a technique used to segment an image by its intensity values in

order to extract important features fiom that image. Thresholding requires the user to

indicate the upper and lower bounds of the intensity values of interest, while the

rernaining intensity values are ignored. Thresholding can be used to separate out the

intensities associated with a specific feature such as neutron absorption by the water.

Histogram equalisation is used to alter an image with poor visual characteristics,

by altering the associated histogram of that image. A nomal histograrn will either span

the full range (fiom O to 65k) or a large range of intensities. When an image has a

compressed range of intensities, histogram equalisation can be utilised to expand the

histogram completely by applying one of the many available equalisation functions.

Some types of equalisation functions include linear, bell-shaped (a greater

concentration of intensities are Iocated at the centre of the histogram), logarithmic,

exponential or a best-fit function. The best-fit function stretches the histogram to

achieve the best possible contrast di~tribution.~' The steps utilised for histogram

equalisation are as f01lows:~~

a. Compute the histogram of the image,

b. Integrate the histogram to form the cumulative distribution,

c. Scale the cumulative distribution to a mapping, which has values between O

and 65536, and

d. Apply the resultant mapping to the image.

Page 48: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

2.7.2.2 Svatial Filters

Filtering operations reduce or increase the rate of change that occurs in the

intensity transitions within an image. Areas in which there are sudden or rapid changes

in intensity appear as hard edges in an image; convesely, areas where there are gradual

changes produce sof€ edges. Spatial filters detect and modiw the rate of change of

intensities throughout an image. A filter can increase the intensity differences on sofi

edges to make a feature appear sharper, or d u c e the rate of change of the intensity on

a hard edge in order to smooth and soften a feature.

Spatial filtering techniques are divided into three categories. The h t two

filtering techniques are called convolution (linear filters) and non-convolution (non-

linear) filters. The final filtering technique does not fit into either of these categories

and is classified under the heading of Mathematical in Figure 2.7. Convolution and

non-convolution filters alter the original image by examining and processing the data

around small regions, called pixel "neighbourhoods." A neighbourhood is a square

region of image pixels, typically 3x3,5x5, or 7x7 in size.

2.7.2.2.1 Convolution Filters

There are several different convolution filters available to the user; al1 of these

types of filters multiply the values within a neighbourhood by a matrix of filtering

coefficients (integer values) which is referred :O a s a kemel. Kernels are the same size

as the neighbourhood. The multiplieci results (neighbour pixel value times kernel

value) are summed and divided by the sum of the kemel values. The different b d s of

kernels and their size will produce different results to an image. An example of the

convolution method is as follows:

Page 49: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

A 3 x 3 neighbourhood of an image is displayed below:

A sample kernel applied to the complete image could have the values:

The new intensity value located at the centre of the matrix would be:

The values in the kemel dictate the type of convolution filter. There are

different types of Convolution filters available and their application is dictated by the

difficulties associated with the image.

2.7.2.2.2 Non-Convolution Filters

There are several different non-convolution (non-linear) filters available to the

user. Non-convolution filters use pixel neighbourhoods and kernels, but unlike the

convolution filters, no predetemined values are used in the kernel; rather the intensity

values around the pixel of interest in the image are used in the kernel. Either the

statistical methods or mathematical formulas are applied to the kernel and

neighbouiliood values so that the pixel of interest may be modified. There are several

different kinds of non-convolution filtm available, but only Median and Erode filters

are applicable in this study.

Applying a Median filter will smooth an image by mod img pixel values that

Vary significantly fkom their surroundings and is accomplished by replacing the centre

Page 50: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

pixel in a neighbourhood with the median value of the neighbourhood. Although

median filtering will soften an image, it generally preserves the edges in an image.

This filter is particularly effective at removing random, high-impulse noise from an

image (e.g., spots or points that vary significantly fkom the background).

The Erosion filter is a morphological filter that changes the shape of objects in

an image by reduculg (eroding) the boundaries of bright objects, and enlarging the

boundaries of dark abjects.'' The Erosion filter is used to d u c e or elirninate small,

bright objects in the background of an image.

2.7.2.2.3 Mathematical Techniques

The Mathematical filter method does not fit into either dassification

(convolution or non-convolution filters), but is a f o m of spatial filter. This spatial

filter mathematically combines either a single image or several images to a master

image. Each pixel value in the master image is altered by a pixel value fiom another

image at the same location.

Multiplying the master image by itself several times is a simple mathematical

technique, and has the effect of changing the intensity of an overly dark image to a

lighter image depending on the multiplication factor. This filter amplifies the original

image's intensiv, the effect is similar to increasing the gain of the camera, without the

introduction of electronic noise usually associated with large gain increases. This

technique also amplifies the intensity of the original noise in the image.

The next four mathematical techniques are similar and use reference images

(Flat fielding, Ratio flat, J flat and Optical density). An image of total darkness

(referred to as a dark image), an image of interest (referred to as the chta image) and

Page 51: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

an image with a constant value of brightness (referred to as a unified image) are

mathematicaiiy mixed by a specified ratio. A dark image is an image where there was

no neutron interaction with the scintillation screen. The unified image is an image of

the scintillation screen that has been illuminated by neutron interaction without an

object present between the screen and the neutron source and therefore the scintillation

screen would be illuminated unifonnly, if the neutron beam were equaily distributed

over the area of the screen. Image ratios are used in a variety of applications to correct

or calibrate image data.

Flat fielding is generally used to correct for defects in an imaging detector array

and for non-uniform lighting. Such non-uniformity is really a "multiplicative noise"

problem and therefore requins a division operation to correct for this type of defect.

This correction method is more accurate than simply subtracting the background. Flat

fielding is calculated £iom the f o l l o ~ i n ~ : ' ~

Flat fielding = Constant * ((Data image - Dark image)Nniform image) (2.10)

Another variation of Flat fielding called Ratio Bat has also been utilised with some

success. The large différence is that the constant is replaced with a mean dark image.

The formula for Ratio flat is as follows:

Ratio 8 at = mean (dark image) * (Data image - Dark image)/Dark image (2.1 1 )

Page 52: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

A simpla Flat fielding c m be utilised by eliminating the bnght image and

multiplying the result with a constant value. This method will be referred to as J flat

and calculateci by:

J flat = Constant * Data image/Dark image

Optical density calculations are more rnathematically intensive and produce

more accurate flat fielding. A base 10 logarithm is performed on the data after the ratio

of images has been taken. The formula for this method is as follows:

Optical density = Log 1 0 * Dark image/(Data image - Dark image) (2.13)

2.7.2.3 Frwuency Filtering

Image interference in the form of a reguiar pattern across an image can be

difficult to remove using spatial filters. The best method for removing periodic or

coherent noise is by converting the image f?om the spatial domain to the fkequency

domain and then editing out the noise and converting the image back to the original

foxmat. Coherent noise is a stable, reproducible, well-defined pattern that is not time

dependent? Although there are several types of fkequency filtering methods available,

the best known is Fast Fourier Transforms (FFT). Neutron radioscopy does not

produce an image having the characteristics that require the application of FFT and

therefore furtha discussion of this method is not required.

Page 53: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

2.8 Svstem Resolution

Spatial resolution is generally quantified in terms of the smallest separation at

which two points can be disthguished as separate entities. The fiml resolution is an

accumulated resolution fkom d l components of the system. Some components will

have greater effects in reducing the totr! remlution. !f pixel size is the limiting factor,

the resolution is related to the srnallest item that can be resolved and not necessarily the

smallest feature. A feature considerably smaller than a single pixel can affect the

pixel's contrast relative to the surrounding pixels and is known as "partial volume

effect."" Direct analysis of a test sample only determines the total unsharpness of the

system and not necessarily the resolution of the system. System sensitivity is

quantified in terms of the system's ability to detect the smallest standard discontinuity

in any given sensitivity indicator (SI) that is obsenmble in an image.''

2.8.1 Total Unsharpness

Radioscopie images are essentially a "shadow picture" where several factors

contribute to the blurred detail. The total unsharpness of a system is defined as:

U: = cl: + U: (2.14)

where:

Ut = total unsharpness U, = geometric unsharpness Ui = inherent unsharpness

Geometric unsharpness is related to the physical characteristics of the neutron

facility and the object to image plane relationship. Geometric unsharpness is the

Page 54: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

bliirring of the image at an object's edge. In some cases, geomeûic unsharpness will be

the limiting factor of a system's resolution. Geometric parameters required for

unsharpness calculaiions are illustrateci in Figure 2.9. The geometnc unsharpness has

unie of length and is calculated as follows:

If (L >> t), then geometric unsharpness can be written as:

where:

Ug = geometric unsharpness, D = neutron source size or collimator aperture size, t = object thiclcness or object to image distance, and L = source to image distance.

Image Plane

Neutron Object A us t

4 - Ue t

Figure 2.9 - Geometric Unsharpness

uiherent unsharpness is the product of al1 the imaging components contributing

to the unsharpness of an image. Inherent unsharpness cannot be rneasured directlx

rather the total is rneasured for an image and then the contribution of the geometric

unsharpness is subtracted.

Page 55: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

2.8.2 Measuring Total Unsharpness

An image taken of a gauge made fiom a neutron absorbing material with

v-g size holes c m be analysed to detennine the smallest hole visible. The smallest

size hole will roughly indicate the resolution or more accurately the system's

unsharpness. Similarly, a gauge with varying width slots can also be used in the same

rnanner.

Another method of deterrnining the total systern unsharpness (-Ut) can be

measured directly fiom the Edge Response Function (EW) graph as shown in Figure

2-10? An ERF graph is produced by taking an image of a very thin (25 pm thick)

gadoiinium (Gd) or cadmium (Cd) sheet placed with the largest area perpendicular to

the neutron beam. The next step is to take a line density reading across the image of the

Gd or Cd sheet and the background. The line density values were measured utilising a

function in the image software for this research. Plotting the normalised densities

against distance produces a graph with a shape sirnilar to Figure 2.10. A curve fit of the

data will produce an ERF, which corresponds to the line in Figure 2.10.

Page 56: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

0.0 0.4 0.8 1.2 1.6 2.0 2.4 Distance from ecige (mm)--->

Figure 2.10 - measurement from Gd Me-edge

2.8.3 Resolution fiom MTF

A Modulation Transfer Function (MW) is a more general method compareci to

the SI or other foms of Image Quality Indicators (IQI) and has been used for assessing

the spatial resolution of an optical imaging device." MTF provides a fast, reliable and

repmducible means of indicating the resolution of a system and is capable of

monitoring any changes in the optical system. The MTF of a system is the ratio of the

magnitude of the system's output amplitude to the magnitude of the system's input

amplitude as a function of spatial fiequency.58 The measurement of the MTF can be

made by imaging either ao object with altemathg O to 100% contrasting bars, a thin

open slit between 100% contrast material or a knife edge of 100% contrast material. A

knife-edge is the most convenient object to comtruct and image for neutron radioscopy

and was used to calculate the resolution of the CCD system at RMC. The resuit fiom

Page 57: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

an MTF analysis is a graph indicating MTF values against fkquency (cycles per mm).

The resolution of the system is indicated by the fiequency value, which is the number

of line pairs in a given distance. A line pair is two equally thick lines that are highly

contrastai. The larger the fkquency number, the thinner the cycle is and the h e r the

resolution of the system. An example of a frequency with 2 cycles is shown in Figure

2.11. A system with a resolution of 2 cycledmm (Figure 2.1 1) would have a single

cycle width of 0.5 mm and a resolution of 0.25 mm (half of a single cycle).

Two Cycles per mm

Figure 2.11 - Example of Frequency per mm

The first step in producing an MTF graph is to create an ERF for the system as

previously describecl, which will produce a graph similar to Figure 2.12.

Differentiating the ERF will produce another fùnction called a Line Spread Function

(LSF) and, when the LSF is plotteci, it would look sirnilar to Figure 2.12.

Page 58: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Graph of ERF Graph of LSF derviveci fmm ERF

Figure 2.12 - Samples of ERF and LSF Graphs

Mathematicall y, an MTF is the magnitude of the optical tramfer function, which

is a nomalised Fourier transform of the LSF. Calculation of the MTF fiom the

nomalised transforrn of the LSF is as follows:

is the Fourier transforrn of the magnitude of the

system's output amplitude (LSF), and

.Cao

ILSF(x)d ir the to the Fourier transfomi of the magnitude of the - aD

system's input amplitude

Page 59: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

An approxirnate solution to this equation, which is used to calculate the system

resolution at RMC (Section 4.3) is: 59

where p = measured pixels per mm, f = fiequency per mm, and Ut = total unsharpness of the system

Varying the value of the fiequency, f, the MTF values are cornputed and are

plotîed on a graph a s shown in Figure 2.13. A value of 10% for a MTF is normally

chosen and the fiequency is then detennined from the graph. A value of 10Y0

modulation indicates a 10% contrast difference between each half cycle.

A direct method of calculating the MTF of a system can be accomplished by

taking an image of a line pair gauge. A line pair gauge is constnicted so that several

fiequencies per mm are simulated with material having a 100% contrast- The image of

the line gauge is read and, when the altemathg contrasting material cannot be

distinguished, then the spacing associated with that group is the MTF value. A line pair

gauge was used at Penn State to determine the MTF for CCD camera system.

Figure 2.13 - MTF graph

Page 60: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

2.8 -4 Sensitivity Indicator (SI)

No recognised standard resolution gauge or Sensitivity Indicator (SI) is

available for neutron radioscopy, but several methods have been adopted fiom neutron

radiography for measuring the total unsharpness of a system.

A modified Type A SI, fkom the American Society for Testing and Materials

(ASTM) 545-75 standard can be used to determine the smallest discontinuity or

porosity that the system can image. The goal of this A S T M sensitivity indicator is to

provide visual and quantitative data that could indicate the size of defect and the

maximum thickness of material for a given hole size. To adapt this ASTM standard to

radioscopy, the step wedge with the holes was thickened and the holes were enlarged,

Figure 2.14. A table indicaihg the R value fkom the ASTM and the new hole sizes and

wedge thiclmess are presented in Table 2.1. The sensitivity level reported is the largest

consecutive value of R that is visible in the image of the SI. An R value was calculated

for the CCD system at RMC and at Penn State using the Type A SI.

Page 61: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure 2.14 - Modified ASTM E545-75 Type A

Table 2.1 - Designation for Sensitivity

1 o f R 1 Discontinuity 1 Absorber Step 1

Levels 1 Value 1 Size o f Hole Thicbess of

Page 62: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

2.9 Image Rmresentation

Depending on the method used to analyse a digital image, there may be a

requirement to present the image in other formats. In general, a digital image can be

analysed by a subjective or comparative method. The subjective method requires an

expert to analyse the image and give an opinion on the contents. The results fiom a

subjective analysis are not reproducible as the opinion is based on a person's

experience. The subjective method will not be discussed M e r . The comparative

method requires that the information fkom an image be presented as a histogram grapha

or a graph containing intensity difference values.

2.9.1 Histogram

A digital image is constmcted fkom a 3D matrix comprishg of values indicating

the x and y location and intensity level for every pixel location in an image. A

histogram is a graph that compiles the intensity data from an image into a particula.

format. By convention, the y-axis corresponds to the arnount of pixels per intensity

interval and the x-axis is the intensity scale, f?om O to a maximum value. The x-axis is

divided into sequentially grouped intensity intervals. An intensity interval can have a

range of values fiom one to a maximum of 256. An exarnple of a histogram is shown

with an explanation of its characteristics in Figure 2.15. Due to the large range of pixel

counts encountered, fkom 1 to 262 144, at times only a section of the histogram will be

s h o w e.g., the intensity may range fkom O to 4096 (black end of the scale), where the

intensity value 4096 is the cut-off point. There are still data beyond the cut-off point,

Page 63: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

but due to a very low pixel count, anywhere £kom 0.02% (524 pixels) to 6% (1 5729

pixels) of the total number of pixels, it is deemed not significant enough to be included

in the histogram.

Histogram Graph

This point has 1000 pixels with an intensity value of 51 2

lntensity Scale

Figure 2,15 - Example of a Histograrn

A histogram is a powerful tool, as inîrhsic information can be identified more

easily and made more readable than by visually analysing an image. Subtle variations

in intensities are eliminated when presented as an image. Limitations to the image can

be due to both the cornputer monitor, which is capable of showing 256 levels of

intensity, and the human visual system which is sensitive to only 64 levels of intensity.

Histograms can represent the complete intensity scale. Unfortunately, a histogram can

only indicate quantitative information and can simultaneousIy represent two different

images as s h o w in Figure 2.16.~' Both images (Figure 2.16-a and Figure 2.16-c)

produce the same histogram, yet look totally different; therefore, the picture associated

Page 64: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

with a histograrn shouid be reviewed to understand Mly the characteristics of the

histogram. The nurnber of pixels that correspond to the black intensity are the same for

each picture. This situation produces the left peak on the histograrn (Figure 2. la).

The number of white intensity pixels are also the sanie for both pictures; hence, the Ieft

peak in Figure 2.16b. If there were only two true intensities, black and white, there

would be only two lines at each end of the scale. Since these are not true black and

white images, the peaks are located more centrally and not at the ends. The peaks have

a width which indicate intensity variations around a centrally dominant intensity value.

Histogram

Intensity

Figure 2.16 - Histogram Example

2.9.2 Intensity Difference Method

For comparative analysis, the image must not be digitaily enhanced as simple

enhancements such as changing the contrast, brightness and gamma values oui improve

a poor image or degrade a good image.

Utilising the mean square criteria, which produces an average or s u m of squares

of the error between two images, is also not applicable.62 The mean square criterion

Page 65: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

requires that al1 images be analysed agauist a standard or master image. No standard

image is available for comparison and, therefore, this technique is not applicable.

Normalising the image data allows comparison of images without the previous

disadvantages. Normalising, or what will be referred to as intensity diEerence, is

accomplished by comparing the average intensity of the background of an image

against the average intensity of a specific area or region of interest, i.e.,

intensity Difference (%) = [(avmge background intensity - average area of interest intensity)/ average

background intensityJgl 00 (2.19)

Utilising this method allows comparison between areas of interest on the same

image or similar areas of interest on other images. Intensity difference is a valid

method for comparing images, as the human eye is sensitive to perceived luminance

contrast rather than the absolute luminance valued3 In other words, the eye is more

sensitive in spotting an item if there is a large difference in contrast between two items,

in this case, the background and the area of interest. The difference in contrast between

two areas is the fundamental principle behind neutron attenuation; therefore, intensity

difference is an ideal mathematical method for comparing areas of interest or images.64

Page 66: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

CHAPTER 3

EXPERIMENTAL PROCEDURE

The first neutron source used for neutron radioscopy for this thesis was the

Neutron Radioscopy System (NRS) installed in the SLOWPOKE-2 Facility at RMC. The

research reactor in this Facility is capable of producing a maximum thermal power of 20

kW (kilowatts), and is regulated by the neutron flux measured at the reactor core. For

neutron radiographylradioscopy, the reactor presently operates at half power for a nominal

neutron flux outside the reactor core of 5.0 x 10' n/cm2 S. The flux f?om the core produces

2 to 3 x 1 o4 dan2 s at the image plane.65 The Neutron Beam Tube (NBT) is circular in

cross section, tilted 8.S0 fiom vertical and has a measured LAI of approximately 1 0 0 . ~ ~

The second neutron source used by the author was the Breazeale Nuclear Reactor at

(BNR) Pemsylvania State University (referred to as Penn State). This facility is capable

of producing a maximum power of 1 MW. For the experiments, the reactor was set to 1,

10, and 100 kW which equate to a neutron flux level at the image plane of 3.0 x 104

n/cm2s (calculated), 3.0 x 10' n/cm2 s (calculated), and 3 .O x 106 n/crn2s (measured),6'

respectively. Flux values for 1 and 10 kW were calculated assurning a linear relationship

between power setting and measured flux value. The beam tube has a round shape, is

horizontal and has an UD ratio of 155.

Page 67: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

The section of the NRS at RMC called the rniddle beam stop has a reenterent shape

that not only contains and stops the neutron beam, but also acts as a difhser for the CCD

camera system as shown in Figure 3.1. An aluminium and rubber beflows structure

comects the scintillation screen to the diffuser making a light tight enclosure. For the

experiment at Penn State, the scintillation screen and CCD camera were properly oriented

by a portable aluminium rectaogular and trapezoidai enclosure that was dimensionally

identical to the diffuser at RMC, Figure 3.2.

Scinl Screen

Beamstop

CC0 Carnera 8 Zoom Lens

Diffuser

Bottam Beamstop

Figure 3.1 - Beam Stop and Diffuser at RMC

Page 68: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Top View

Neutrons - /

Scintillation Screen

Camera

Figure 3.2 - Diffuser Used at Penn State

Except for the diffuser, the same camera systern was used at each site. The

equipment consisted of a 431.8 mm (17 in.) x 431.8 mm (17 in) aluminium backed

6 ~ i ~ : ~ n ~ : ~ u scintillating screen, an Apogee AP7 CCD camera with a SITe502 CCD

sensor, a front surface mkor with adjusting mechanism, a Vivitar AF-SC 28-70 mm, F3.5

zoom lens, a red laser source (used for focusing the lens), an Apogee video capture board

and a 200 Mhz personal cornputer running Windows 98. The SITe502 CCD is a back lit,

full h n e CCD chip with an anay size of 5 12 x 5 12 with a pixel size of 24p. The pixels

have a maximum well capacity of 296000 electrons and a dark curent rate of 0.8

electrons/pixeVsecond at -33°C. The CCD camera, cooling unit, laser, &or and mirror

adjuster are housed in a single unit as indicated in Figure 3.3. The CCD canera is

thennoelectrically cooled to an absolute temperature of -50°C, which corresponds to a

-30°C working temperature. Image acquisition was accomplished with ''Image Pro Plus"

software and the image analysis was completed with "IP lab Spectnim -Scientific Imaging

Software for Macintosh". All image analysis was done on an Apple Macintosh Cornputer,

Mode1 8500, as recomrnended by the software vendor.

Page 69: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure 3 3 - Camera Assembly and Control Box

There were six different sets of experiments used to analyse the potential

capability of the CCD camera system. The first three sets of experirnents detexmined the

characteristics of the system when no object was present in the neutron beam. The next

two sets of experiments indicated the performance of the camera when a component was

in the neutron beam. The last experiment compared neutron radiography with neutron

radioscopy. A flowchart of the major headings and sub headings of the various

experiments is shown in Figure 3.4.

Initially, software restrictions forced the CCD camera to be cooled to -15°C rather

thm -30°C and reduced the camera resolution to 5 1 1 x 5 1 1 vice 5 12 x 5 12. Throughout

al1 the experiments, the exposures were increased by one-minute intervals except for the

first one, which was only increased by 30 seconds. In Appendix B are indicated camera

temperatures, camera resolution and qmtity of images taken for each experimental

Page 70: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

section. The normal operaing temperature for the camera was -30°C with a resolution of

512 x 512.

Installation and Optimisation of System at RMC Reduction of Ligbt Leaks

Configuration Parameters

1

Camera Parmeters Dark Image - Temperanne Change Dark Image - Diffuser Changes

Unified Images - Neutron Flux Effects Dynarnic Range

1 Resolution

Systern Resolution Calcuiations System Sensitivity il

r 7

Water Ingress Simuhtion

Water inpress Water Ingress - Single Cells Waîer Ingress - Three Cells Wata Ingress - Camera Gain

Water Ingress - Digital Enhancement

Simuiated Water Simuiated Water - Single CeIl Simuiated Water - Three Cells

l Comparison of Radioscopy to Rndiography Resolution

C m I Figure 3.4 - Flow Chart of Experiments

Page 71: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

3 -2.1 Installation and Optimisation of the CCD Camera S ystem at RMC

3.2.1.1 Reduction of Li& Leaks

The CCD camera, bellows and scintillation screen were installed, and the camera

was visually lined up to the scintillation screen. As part of the optimisation processes, the

insides of the bellows and screen assembly were painted flat black to reduce light

reflections. Silicon sealant was applied to any crevasses or seams, which were by-

products of the manufacruring process. Opaque duct tape was applied to areas that could

not be pemianently sealed due to the disassembly and removal requirements of the

bellows. Not al1 fonns of duct tape are opaque and testing for opaqueness was mandatory

before use.

The reactor remained off and five images of various exposure times (referred to as

dark images) were taken before any changes where made to the bellows and screen

assembly. Light leaks in the system were eliminated by trial and error, until no M e r

improvement could be obtained. Another series of images were taken to compare its

histogram against the histogram of the original images in order to quanti@ the

improvement.

3.2.1.2 Confimation Parameters

Due to the configuration of the camera system, the bellows and scintillating screen

are removable as the scintillation screen occupies the location required for the vacuum

cassette used in radiography. To veri@ that light leaks have not been induced due to the

bellows installation, a method for detennining the quality of the system is required so that

consistent performance fkom the camera systern is achieved. After optimisation, 30 dark

Page 72: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

images of a 2-minute exposure were taken to determine statistically the lowest pixel count

value of the centroid and the largest FWHM intensity values of the curve. These two

values indicate the minimum acceptable performance of the camera system. When

beginning to use the CCD camera, an image was taken with a 2-minute exposute and its

peak value and FWHM intensity values were compared against the statistical values

calculated in this section. An image not meeting these criteria uidicated a light leak or a

problem with the CCD carnera. A 2-minute exposure was chosen so that a reasonably

reproducible image could be produced in a relatively short exposure time.

3.2.2 Camera Parameters

The following experiments detennined the characteristics of the camera and

scintillation screen. Altering the diffuser, changing the temperature of the camera, varying

the exposure tirne and varying the flux level al1 affect the performance of the camera

system and the quality of the image. The above parameters were altered when the system

was not exposed to neutrons (called a dark image) and when the system was exposed to

neutrons (called a unified image). A unified image is a standard term used in digital

enhancement techniques and will be used throughout this work to indicate when no object

is placed between the neutron bearn and the scintillation screen so that the neutrons

uniformly illuminate the screen. If intensity values fiom a dark image and a unified image

are combined, the resdting histogram will show the dynamic range of the system.

Page 73: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

3.2.2.1 Dark Image

With the reactor off, the temperature, di&er and exposure times were varied so

that the performance of the camera could be andysed. The temperature of the çamera was

altered fiom -1 5°C to -30°C, for varied exposure times.

For the experiments involving the different diffusers, the camera temperature and

resolution were held constant at the standard values, while the type of d i h e r and the

exposure times varied. The first series of exposures did not involve the diffûser. The

camera was placed in a photographic dark room and the lem covered. These exposures

represented the absolute darkest Mage that the carnera could record. The second series of

exposures were taken with the portable diffber used at Penn State reactor, while the last

series of exposures were taken with the fked diffuser at RMC.

3.2.2.2 Unified Image

Varyhg the neutron flux and the exposure time indicated the response of the

scintillation screen and the camera. With RMC's reactor at half power, a series of images

were taken with varied exposure times. At Penn State, images were taken when the

reactor was set at 1, 10 and 100 kW.

3.2.2.3 Dvnamic Range

The dynamic range was calcuiated by combining the data fkom the histograms of

the dark image and the unified image which produces an envelope indicating the size of

the dynamic intensity range for a given exposure, as shown in Figure 3 .S. The range is

calculated by subtracting the hvo peak centroid values, the k t peak centroid value from

the dark image and the second peak centroid value fiom the unified image. This span is

Page 74: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

the dynamic intensity range of the system for a given neutron flux and exposure the

(Figure 3.5). When a component is placed in the neutron beam, there will be a range of

pixel intensities that range fiom a minimum value that corresponds to a dark image to a

maximum value that corresponds to a uninecl image.

Eumple of Dynuiilc Range C.lwht&n

LHS - Peak œmid 80000

mOaO

60000 - C

s - 2 p-

30000

2MXK) of udfied image I

lOOaO

O l a4 i2M r i 3 6 1782 aW8 23M 2560 2816 3072 3328 3584 3840 4096

Figure 3.5 - Example of Dynamic Range Histogram

3.2 -3 Resolution Measurernents

The system resolution was analysed for varying neutron flux and exposure times.

At the NRS at RMC, the MTF and a modified ASTM 545-75 Type A indicator were used

to determine resolution and sensitivity. Images of the two indicators were taken at various

exposure times varying fiom 30 seconds to 10 minutes. At Penn State, a Line Pair Gauge

Page 75: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

(LPG) and a modified ASTM 545-75 type A indicator were imaged with the reactor set at

10 and 100 kW.

3.2.4 Water Ingress Simulation

3.2.4. i Test Piece

A s m d section of a CF188 rudder, measuring 14 cm x 14 cm was used as a test

piece for a series of experiments. The CF1 88 rudder is a composite construction that

consists of two outer skuis of carbon graphitdepoxy made fkom several plies of carbon

graphite material (1- to 3-mm thick) attached to an duminium honeycornb core by a FM-

300 adhe~ive,~* as shown in Figure 3.6. The top skin was removed so that the honeycomb

could be accessed. Afier the cells of the honeycomb were filled with distilled water, the

top was held in place with clear tape. The test piece was not as thick as the original rudder

as some honeycomb had been removed when the top skin was cut fiom the test piece. It is

estimated that the removal of the top surface reduced the height of the honeycomb by

approxirnately 0.5 cm ( 1 3% reduction).

Figure 3.6 - Construction of a CF188 Rudder

Page 76: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

3.2.4.2 Procedure

For a single exposure tirne, a specific area of interest was extracted fiom the

original image. Areas associated with a set of single cells and a set of three cells were

analped. Thresholding was applied to the new image and statistical data such as mean,

minimum and maximum and standard deviation were calculated for each ce11 or cells

containhg water or High Density Polyethylene (PE). The intensity difference was then

calculated for each ceIl or cells containing water or PE.

The intensity difference for a representative volume of water was calculated for

either one or three cells for exposure tirnes that varied f h n 0.5 to 10 minutes. These

intensity difference values were then plotted against the exposure time. The amount of

error for these graphs was calculated fiom the standard deviation values for each ce11 or

cells. The optimum exposure t h e for the system was determined from the graphs for the

intensity difference of the representative volume of water versus the exposure time.

3.2.4.3 Water Inmess

All images of water ingress were taken at the NRS at RMC and accurately

simulate ingress into a flight control surface. The results fiom this experiment were used

to detemine the optimum exposure tune, minimum sensitivity, best camera gain setting

and best digital enhancement method. The effects of camera gain were not explored in

previous sections, as ampl img a "dark image" or "unified image" does not indicate the

effects that the gain will have on an image of an object. To simulate water ingress in a

fligbt control, distilled water was inserted into the test piece in the pattern and quantity

indicated in Figure 3.7. The test piece was imaged at various exposure times.

Page 77: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Further images of the test piece (Figure 3.7) were taken using a three-minute

exposure and changing the gain fkom O to 16, by increments of four. From past

experhents, increasing the gain incrementally by a small nurnber (a factor of one or two)

did not rnake a noticeable diffaence in the results. A factor of four is easily divided into

the maximum gain value and clearly showed the trend of the data for increased gain.

Utilizing the optimum exposure time and gain setthg for the test piece, the

exposures were subjected to a large number of enhancement techniques. The goal was to

create the best picture from a given exposure. The best image would include as many of

the features and as much detail as is lmown from the original object The details fiom the

original object were well understood, being a test piece with known characteristics that

were either implanted or were naniral for this structure. The enhanced images were

judged subjectively to indicate the best enhancement techniques.

Section of CF-18 Rudder PerpendiculYtaSlirhCeofRudder

Figure 3.7 - Test Piece with Water at RMC

Page 78: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

3.2.4.4 Simulated Water

To compare the effects of changing the neutron flux on the test piece, water wuld

not be used as the test piece sat horizontally at RMC and vatically at Penn State. High

Density PolyethyIene (PE) was substituted for water. Al1 images taken at RMC had PE

and water inserted into the honeycomb of the test piece in the pattern and quantity

indicated in Figure 3.8. The results fkom the water were not used in this experirnent, but

the water was added to the test piece to take advantage of the

more data. The test piece with PE and water Iayout

for various exposure times.

Section of CF-18 Rudder Perpendicular to the Su- of R d e r

opportunity for collecting

in Figure 3 -8 was irnaged

Wter (mL) in a single ceil

.O 28.0 24.0 22.0 20.0 18.0 10.0 14.0 120 10.0 8.0 6.0 4.0 2.0

t

PE (mm) u-d- ,-,-: ,- in a single ceIl

'-- 18.0 20.0 10.0 14.0 120 10.0 Q.0 8.0 7.0 0.0 5.0 4.0 3.0 2.0

Aluminum Honeycom b

Upper and lower safaces mrnoved for danty

Figure 3.8 - Test Piece with PE at

Page 79: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

AU images taken at Penn State had PE inserted into the honeycomb of the test

piece in the pattern and quantity indicated in Figure 3.9. With the reactor set at 10 and

100 kW, various exposures were taken at each power setthg.

Section of CF-18 Rudder Perpendiwlar to aie Surface of Rudâer

Upper and lowef surfaces removed for ciarity

Figure 3.9 - Test Piece with PE at Penn State

3.2.5 CF188 Rudder

A CF188 Rudder, senal

3 1 1 -NRT (revision 1) utilising

NRS is capable of mapping al1

radioscop y

8 positions

0241, was inspected as per technique MC- 188-

instead of radiography techniques. The RMC

of the rudder, as s h o w in Figure 3.10. At the

Page 80: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Penn State reactor, the same images were taken as per the technique, less exposures 1, 7

and 8. The beam tube at Penn State produced a round beam that did not cover the

complete area of the scuitillating screen, as shown in Figure 3.10 for exposure 5.

Therefore, taking images 1, 7 and 8 was not feasible. Image 1 is of the root of the rudder,

containing a large interna1 aluminium structure. Image 3, 7 and 8 are of the trailing edge,

located at the inboard end of the aircraft, and contain small amounts of honeycomb

structure. At1 three images are of less significance for water ingress inspection.

CF-18 Rudder

~ a d i o ~ r a ~ h k Film Positions 17in. xl4in.

Figure 3.10 - Fîim Layout on CF188 Rudder

Page 81: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

3.2.6 Comparkon to Film

3.2.6.1 Resolution

The resolution of an image using radioscopy, SR film (the slowest, highest

resolution) and CX film (the fastest, lowest resolution) was derived fiom the MTF

technique and the modified ASTM E454-75 Type A SI.

3.2.6.2 Time and Cost Study

The cost of doing either technique can be divided into capitd costs and cost per

exposure. To simplifi calculations, al1 costs were assumed to occur initially and therefore

"the value of money" was not used. The concept of " t h e value of money" implies that

money has a value that varies depending on when it is received or disbursed. Capital cost

did not inchde items that are considerd part of the Facility, such as the reactor, the bearn

stops, the positioning systern and the radiation meters (installed, not portable). These

items are cornmon and essential for both neutron techniques and therefore were not

included in the cost study.

A CF 188 rudder was neutron radiographed at RMC as per technique MC- 188-3 1 1-

NRT (revision 1) and the time to complete the activities was recorded. These timings

were compared to the time requird to comp1ete neutron radioscopy using the sarne

procedure.

Page 82: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

CHAPTER 4

RESULTS AND DISCUSSION

4.1 Installation and Optimisation of Diffiiser at RMC

4.1.1 Reduction of Light Leaks

Cornparing the histognims of the systern before and after optimising the RMC

diffuser (as described in Section 3.2.1.1) produced simila. trends for any given exposure.

A graph of a two-minute exposure illustrating the effects of improving the system is

s h o w in Figure 4.1. Two mmplete sets of exposure data fYom 30 seconds to four

minutes, collected before and after optimisation are, in Appendix C, Figures C. 1 and C.2.

The graph in Figure 4.1 shows the complete intensity grey scale fkom O to 65536. The

identical data are plotted in Figure 4.2 with the upper intensity scale value limited to

4096, to reveal the shape of the two curves.

Page 83: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure 4.1 - Full Intensity Scale of Optimisation

-- - -- -

Figure 4.2 - Enlarged Histogram Before and After Light Leak Reductions

Page 84: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

The goal in optimising the system was to eliminate al1 light leaks and hence

obtain the darkest image possible. A dark image would produce a large thin peak with

an intensity value close to zero (black) i.e., to the left of the graph. Figure 4.1

dramatically shows that both peaks are situated to the far left of the graph with the peak

after optimisation taller and namower.

Without optimisation, the number of pixels at the peak centroid is relatively srnall

at 79,635, which was ody 30.5% of the total number of available pixels. The total

possible pixel count fiom a 5 1 1 x 5 1 1 array is 26 1 12 1 pixels. As well, the peak centroid

had an intensity value of 1799. Moreover, the non-optimised cuve is skewed to the right

of the peak centroid as shown in Figure 4.2. Full Width of the curve at Half Maximum

(FWHM)~' is a measurement of curve width. The non-optimised curve has a F W

intensity value of 656. The shape indicates that a large number of intensities are being

detected.

The peak centmid of the optimised cuve has a relatively large pixel count

(2 1 1,626) or 8 1 .O% of total pixels available which is a 62.4% increase in pixel count

relative to the non-optimised curve. The intensity value of the peak for the optimised

c w e is shifted left relative to the peak of the non-optimised curve &om 1799 to 1285

which is a 28.6% irnprovement in peak intensity value. Moreover, the optimised curve is

evenly distributed around the peak cenîroid. The optimised curve is wnsiderably thinner

than the non-optimised curve, with a FWHM intensity value of 280, a 57.3 % decrease in

cuve width.

Page 85: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

The above analyses indicate that the work completed for optimisation has

improved the relative darkness of the system. This improvement is indicated by an

increase in the height of the peak centroid, shifhg of the peak centroid intensity value to

the left (darker intensity values), and narrowing of the curve (FWHM value reduced).

4.1.2 Configuration Parameters

Thirty dark images with a 2-minute exposure were taken for statistical purposes to

ensure consistency between radioscopy sessions (Section 3.2.1 2). M y the results from

the thuty images are graphed in Figure 4.3, as the original data are not shown for clarity.

The graph shows the minimum acceptable values corresponding to two standard

deviations (SD) from the average cuve (Section 2.6. l), indicating the envelope for an

expected configuration image.

Figure 4 3 - Statisticai Parameters of Dark Image

Page 86: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

The statistical data fkom the thirty images are shown in Table 4.1. The tolerance

values were calcdated for one standard deviation. The pixel count was the only variable

that had a significant amount of error, which was calculated to be approximately 6 %.

This variation indicates the statistical nature and magnitude of the error associated with

the system. Taking thirty images produces a 90% probability that a new image will

produce sirnilar results with a 75% confidence leve170. Peak and FWHM values

indicating the limits for a configuration image are shown in Table 4.2. Future dark

images having peak and FWHM values falling within the statistical parameters indicated

in Table 4.2 would be considered valid and meet the required standard of darkness.

Table 4.1 - Statistical Data of 30 Images 1 Peak Values I LHS -FWHM I RHS -FWHM 1

Average intensity value

Table 4.2 - Configuration Image Parameters

Average Pixel count

Peak Value - Lower 2 Standard Deviaîion values

Pixel Count 1 Intensity

>=985 19

Average Intensity d u e

F m - Outer 2 Standard Deviation values

Pixel Count 1 LHS - 1 RHS -

Average Intensity value

Average Pixel count

Value

<= 1205

Average Pixel count

>=55315

Intensity Value

>=Il81

Intensity Value

<=1219

Page 87: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

4.2 Camera Parameters

4.2.1 Dark Image - Temperature Change

The effects of camera temperature were analysed by cornparhg the data taken at

two camera temperatures, - 1 SOC and -30°C (Section 3.2.2.1). Comparing the two

temperatures produced similar trends for any given exposure; thus a graph of a 2-minute

exposure, for example, can be used to illustrate the effects of lowering the camera

temperature, Figure 4.4. Note that the intensity interval was changed fiom 256 (Figure

4.2) to 16, which produced a graph with greater detail. Therefore, the size and values

associated with the -1 5°C c w e in Figure 4.4 do not correspond to the previous curves at

this temperature (Figure 4.2). Two complete sets of exposure data collected at -1 5°C are

shown in Annex C, Figure C.3 and at -30°C are shown in Annex C, Figure C.4.

Page 88: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure 4.4 - Effects of Camera Temperature

At the camera temperature of -l5OC, the peak centroid pixel count was relatively

small at 22,8 15 +1369 (6%) which was only 8.7% of the total nurnber of pixels available.

The intensity value associated at the peak centroid was 1397. The -lS°C graph is almost

symmetrical around its peak value and has a FWHM intensity value of 156. n i e shape of

this curve indicates that a large number of intensities are being detected.

Lowering the camera temperature to -30°C produced a graph with a peak centroid

pixel count of 108,170 k6490 (6%) pixels which was 58.6% of the total pixels available

and was an increase in value of approximately four times that of the -15°C curve. The

location of the peak centroid had an intensity value of 1205, which was a 13.7%

Page 89: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Unprovernent in peak centroid intensity value. The "FWHM" for the -30°C curve has an

intensity value width of 34, producing a 78.2% decrease in curve width at the FWElM

value. The analysis has shown that, by reducing the temperature of the CCD camera, the

exposures becarne darker due to the increased peak centroid pixel value, the FWHM was

reduced (narrower curve) and the peak centroid intensity was shifted towards the darker

scale. The results obtained by l o w e ~ g the camera temperature were similar to the

reçults obtained by optimising the carnera system (elirninating light leaks). Reducing the

temperature of the camera reduces the quantity of dark current, which in turn reduces the

noise associated with the image.

A set of 11 images was taken without a diffuser and then a set of images was

taken using the tùted diffuser at RMC and the portable diffuser at Penn State (Section

3.2.2.1). The histograms associated without a di-, fixed diffuser and portable

diffuser are shown in Appendix C, Figures C S to C.7, respectively. Although the camera

had a slightly lower resolution of 5 1 1 x 5 1 1 for the images taken at RMC, the cornparison

is still valid. This slightly lower resolution will not affect the results when comparing

them to the other two histograms. The peak intensity value for each image fiom al1 three

histograms was plotted on a graph of intensity and exposure time in Figure 4.5. The

results fiom this graph were used to determine the effects of the diffuser on the camera

system.

Page 90: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Comparlron benneen DWu+ers Udng Peak htenslty Values 30 Second to 10 Utnute Exposure

@ 3 4 C . 5 1 2 ~ 5 1 2 ~ * r a u 1 I y i n t e n e i o l 1 6

Figure 4.5 - Peak Intensity Values of Dark Images for Different Dinusers

In general, as the exposure time increases for any given diffuser, the intensity

value increases, going fiom dark to lighter. Also the images without a diffuser produced

the darkest values for a given exposure time, followed by the portable diffuser used at

Penn State and the fixed diffiser used at RMC. The error associated with quantization of

intensity values is very low and, when combined with a relatively large intensity interval

of 16, the e m r becomes insignificant. It is assumed therefore, that the error associated

with the data in Figure 4.5 is less than 1%.

Page 91: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

The reasons that dl three curves increase in intensity value, as exposure time

increased, is due to the image being integrated over tirne and the accumulation of dark

current. The images without the diffuser behaved as expected, having the darkest images

of al1 three diffusers, as no surfaces for the light to reflect are available, which would

cause an increase in image intensity. Producing an image with an intensity value of zero

is not obtainable due to system noise and dark current. The lowest the camera system can

achieve is a peak intensity value of 1 140 (Section 4.1.1). The scintillation screen, which

is part of the diffuser and is perpendicular to the camera's field of vision, causes some

reflection of light; hence, the increase in intensity values. The bellows assembly attached

to the RMC diffuser most likely contributes to the still higher intensity values. The

bellows assembly has different absorption and reflective properties than the sides of the

diffuser and the probability of a very small light leak in the bellows assembly is higher,

causing the highest peak intensity values.

4.2.3 Unified Image - Neutron Flux Effects

A number of images were taken with the two diffusers without an object in place

and at four different reactor power settings (Section 3.2.2.2). At RMC, the images were

taken at half power (10 kW) and at Penn State the images were taken at 1, 10 and 100

kW. The RMC power settings and the 1 kW settuigs were sirnilar in flux magnitude at

the image plane so that the screen output could be compared. These series of images

were taken to investigate the response of the saeen to higher neutron interaction and to

Page 92: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

determine how varying the Iength of the exposure affects the image at these higher flux

values. n i e exposure times were 0.5, 1, 2, 3, 4 and 5 minutes. The histogram of the

unified images fkom RMC (10 kW) and Penn State at 1, 10 and 100 kW are shown in

Appendix C, Figures C.8 to C. 1 1, respectively. The peak intensity values for the

different exposure times are graphed in Figure 4.6. The results f?om this graph were used

to determine the effects of neutron and screen interaction on the camera system.

Peak hlensity Yakes from Unitïed hrrges for OHkrent Ffux VaIues at Vulous Expouire Thes

@ r )C . 512x 512fusaUim imemityintmaîoi 16

- - - - - - -- --

Figure 4.6 - Peak Intensity Vaiues of Unified Images at Various Fiux Values

As the exposure time increases, the intensity of the peaks ïncreases, hence the

positive slope of the lines (Figure 4.6). This trend is similar to the trend associated with

Page 93: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

the varying exposure &es of the dark images (Section 4.2.2). Three variables contribute

to this effect, dark current, screen illumination and integration of the image, with

integration being the dominant effect among the three. As well, as power increases (Le.,

flux increases), the peak intensity values increase, as indicated by the increase in the

slope of the line for each power level. Note the large increase in the slope for the 100 kW

value for Penn State. The exception being that the line associated with the data fiom

RMC (at a flux of 2 to 3 x 104 n/cm2 s) was above the Penn State 1 kW values (at a flux

of 3.0 x lo4 n/cm2 s). There are three possibilities f ~ r this discrepancy. For Penn State, it

was assumed that the flux was linear fiom the 100 kW value. As well, the power setting

at Penn State can vary by 5% and still be within normal operating parameters. For RMC,

the flux could be greater than previously measwed.

Figure 4.6 also shows that the half power setting at RMC (10 kW) and the 1kW

power level produced very similar results and therefore the characteristics of the RMC

system were reproducible.

Reviewing tfie images at the highest power setting at Penn State, the image

became visible when the peak intensity value was greater than approximately 10,000.

Therefore, the images at the lOOkW power setting above one-minute exposure would be

visible without image enhancements. Note that, at the 100 kW power setting with a 5-

minute exposure, the image becornes saîurated. At the power setting of 10 kW, images

exposed for approximately 7 to 8 minutes would start to becorne visible without image

enhancement. Extrapolating the data fiom the RMC NRS, it would take approximately

500 minutes to reach saturation and 69 minutes for the intensity value to be greater than

Page 94: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

10,000. Therefore, image enhancanent would be required for any reasonable exposure

time at RMC.

The dynamic range was calculated (Section 3.2.2.3) and graphed for different

exposures times taken at RMC at 10 kW (Appendix C, Figure C.12) and at Penn State at

1kW (Appendix C, Figure C.13), at 10 kW (Appendix C, Figure C.14) and at 100 kW

(Appendix C, Figure C. 15). Since the trend was sirnilar for different exposure times for a

given power setting, a representative graph showing only the values of the dynamic range

was fompiled for a 3-minute exposure time fiom each power seîting (Figure 4.7).

Dynamlc Rang. m m Peak Centroid vakwas tor Duk and U n W lmrges

Reactot Power Setthgr

Figure 4.7 - Dynamic Range of System at Various Power Settings

Page 95: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

The dynamic range of the system increaz;ed dramatically when the flux was

increased. The dynamic range at 1 kW was extremely nmow with an htensity value of

337. The dynamic range at RMC had an intensity value of 963. Even though the flux

levels for both these systems was assumecl to be similar, the dynarnic range at Penn State

at 1 kW was 65% smaller, indicating that the calculated flux value for RMC at 10 kW

(assuming linear relationship) might be approximately 8 x 104 n/cm2 s instead of the

rneasured value of 2 to 3 x 104 n/crn2 S. The dynamic range for 10 and 100 kW were

3,584 and 36,196, respectively. As previously stated, intensity values above 10,000 are

visible without image enhancements and therefore it is expected that a 3 min exposure at

100 kW would contain not only a visible image but have a full range of intensities.

Having a large dynamic range indicates the potential to record small intensity differences

between two objects. Alternately, having a small dynamic range wouid mean that the

intensity difference between two objects could not be identified.

4.3 Resolution Calculation

The system resolution was calculated using the MTF method and a Iine pair gauge

(LPG). A modified ASTM E545-75 Type A was used to detemiine the sensitivity of the

system (Section 3.2.3).

Page 96: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

4.3.1 S ystem Resolution Cdculations

At RMC, the image plane was extended to the fully down position and al1

exposures of a knife-edge gadolinium foil were taken. At this position the pixel size was

calculated to be 0.73 rnm/pixel or inversely 1.37 pixels1mm. The analytical software

recorded the densities across the Gd foil and an Edge Response Function (ERF) graph

was created f?om these densities. The Ut (total unsharpness) was measured fkom the ERF

graphs corresponding to a 0.5, 4 and 10-minute exposure. The Ut values and calculated

cycle/mm values with absolute and relative mors for the three exposure times are shown

in Table 4.3.

Table 4 3 - Resolution Variables for the NRS at RMC

. 10 minute 1 1.37 1 1.50 + 0.50 (33%) 1 0.44 (+ 0.2 11 -0. 12j

Exposure time 30 second 4 minute

Using the pixeVmrn and Ut values for the 4 minute exposure tirne, a MTF graph

was calculated and is shown in Figure 4.8.

pixeVrnm 1.37 1.37

ut (mm) 1.35 L- 0.15 (1 1%) 1.40 4 0.60 (42%)

Cycles1rnm at MTF Values of 10 % 0.50 (+ 0.041 - 0.07) 0.47 (+ 0.341 -0.14)

Page 97: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure 4.8 - MTF Graph for 4-Minute Exposure taken at RMC

The resolution does not seem to be time dependent, as the resolution values

measured in Table 4.3 are within the calculated error. At the 30-second exposure time,

the intensity values between the background and gadolinium foi1 did not Vary

significantly. This lack of intensity range makes this exposure time unacceptable. The

10-minute exposure time had a large intensity variation with considerable system noise,

which produceci the largest Ut and smallest resolution measurernent. At the optimum

exposure time of 4 minutes, a system resolution c m be derived from Figure 4.8 that has a

range of 0.8 1 to 0.33 cycles/mm with an average resolution of 0.47 cycles/mm at a 10%

MTF. These values translate into a range that varies f?om 1.23 mm/cycle to 3.33

mdcycle with an average of value of 2.13 mm/cycle. Dividing these values in half, the

Page 98: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

system resolution would range fÏom 0.61 mm to 1.66 mm with an average system

resolution of 1 .O6 mm.

System resolution was also measured at Penn State using a LPG. The images of

the line pair gauge were taken at two diEerent power settings and for varying exposure

times. A typical image of the LPG taken at 100 kW for a 5-minute exposure is shown in

Figure 4.9.

Figure 4.9 - Line Pair Gauge

The system resolution value is directly read from the image, and the resolution

value is determined by the smallest set of altemating dark and light lines. The resolution

value fÏom the LPG was neither power setting nor exposure time dependent. The system

resolution was measured at 0.656 Ip/mm or 1.5 m d i p as seen in Figure 4.9. The unit

line pair (lp) is identical to the unit cycle used in the MTF method.

Page 99: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

At Penn State, the system resolution was higher (0.656 Iplmm) than the average

system resolution at RMC (0.47 cycles/~nm). The LPG seemed insensitive to changes in

power setting and exposure time. As well, the MTF method was also insensitive to

different exposure times.

Exarnining the images of the ASTM E54S-75 Type A standard taken at RMC and

at Penn State for various exposure tirnes produced the following results as indicated in

Table 4.4. Note that the power level at Penn State (1 kW) comparable to RMC (1 O kW)

was not taken.

Table 4.4 - Hole Values from Modified ASTM E545-75 Tme A

Exposure Time

I 10 I 7 I 11 Note: Vdue from RMC (10 kW) are not M y comparable to Aues from Penn State (10 kW).

~ 6 b e r of Holes Observed at RMC 1 Penn State 1 Penn State

I

The holes visible from the RMC images produced an averaged value of 6 out of

-

(minu tes) 1 (10kw) 1 (10 kW)

16 compared to the images taken at the two higher power Ievels at Penn State, both of

(100 kW)

which had a consistent value of 1 1 out of 16. The flux intensity is directly proportional to

Page 100: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

the amount of holes visible in the sensitivity ùidicator. Using the ASTM E545-75

standard," images taken at RMC had a R6 rating and Penn State images had a RI1

rating.

4.4 Water Ingiress Simulation

Without an object in the neutron beam, the camera's charactexistics did not

indicate the system's performance. Using the NRS at RMC, an experiment was carried

out using distilled water and a honeycomb test piece, which simulated water ingress into

a fiight control surface. These two items were used to determine the camera's

capabilities at a low flux level. To determine and compare the effects of different flux

levels on the camera system, an experiment using PE (hi@ density Polyethylene) in the

same test piece was used at both RMC and Penn State.

4.4.1 Water Ingress

Al1 water ingress images were taken at RMC with the reactor at half power, 10

kW. To analyse the best exposure tirne, two tests were made: the k t for single celIs of

water and the second for three adjacent cells of water. In boîh tests, the amount of water

in each ce11 or set of cells was varied. The results fiom the experiments were used to

determine the optimum exposure t h e , minimum sensitivity, optimum camera gain

setting and best digital enhancement method obtainable at RMC.

Page 101: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

4.4.1.1 Water Ingres - Single Ce11

Filling single cells with incremental voIumes of distilled water simulated the most

strenuous situation to be encomtered during radioscopy of a flight control surface, due to

the smali quantity of water and small surface area being presented. Varying the mount

of water also gave a good indication of the sensitivity of the radioscopy system for

detecting water ingress in a single cell. The intensity difference denned in Section 2.9.2

was calculated for each ceIl for a given exposure time and then graphed in Figure 4.10. If

the camera system is sensitive enough to record small variations in neutron absorption

and scatter, then the intensity difference should hcrease with increased volumes of water

for a given exposure time. As well, integrating an image over time should cause the

intensity difference to increase as the exposure time increases.

-- - ro mn

-9 min a min 7 m n

6 mm - 5 mn -4 min 4 - 3 min -2 min -1 min 4 -30 a

O i O 0.02 0.04 0.06 0.08 0.1 û.12 0.14 0.16 0.18 0.2 0.22 0.24 0.26 0.28 0.3

Volume of Water in Ceil (ml)

Figure 4.10 - Effects of Varying Exposure Tirne on a Single Ceïi of Water

Page 102: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Error Calculations For clarity, no error bars are shown in Figure 4.10. Since the

cell intensity is an average value for the cell, the error for a ce11 would be the value + two

standard deviations. Since thresholding (Section 2.7.2.1) was used to detennine the

average intensity values, the error associated with each ceil is relatively small. For mal1

volumes of water, the error is approximately 1% and for large volumes of water,

approximately 15%. As part of the standard thresholding practice, thresholding was not

applied to the background area and therefore system noise increased the error associated

with the background intensities. The error associated with background intensities can be

as high as 60%. Using standard =or calculations produces a relative error of

approximately 1 1% to 5 1 % of the intensity difference fiom the mallest volume and

shortest exposure t h e to the largest volume and longest exposures. For example, at a 5-

minute exposure for 0.04 ml of water, a relative error of 36% was calculated; therefore,

the intensity difference value with errors wodd be 19% f 6% (36%). An error calculation

of this example is included in Appendix D.

G r a ~ h Features In general, as the exposure tirne and the volume of water

increases so does the intensity difference. The intensity difference curve for a given

exposure is positively sloped, indicating a correlation between the volume of water in a

ce11 and the intensity difference. The shortest exposure time has the lowest intensity

Page 103: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

difference values while the longest exposure time has the largest intensity difference

values.

The ce11 that contained 0.02 rnL of water produced poor results with exposures

greater than 4 minutes. There was such a small quantity of water in the ce11 that it did not

cover the ce11 completely, making it difficult to threshold this area properly. For a given

exposure time the data values tend to oscillate between the measured volumes of water.

The oscillation is well within the error band and is the result of thresholding and data

pmcessing and does not seern to be related to any physical phenomena related to the

system.

There seems to be two bands of data - the 30 second to 2 minute exposures and 3

to 10 minute exposures. This second band tends to be much wider in intensity difference

values. Analysing the raw data of the intensities the for 2-minute and 3-minute exposure

times, the intensity of the water in the cells increased linearly, but the difference in the

empty ce11 background intensities increased dramatically. The empty ce11 measurement is

the intensity value fkom the scintillation screen after absorbing and scattering neutrons

through the test piece's skins, adhesive and honeycomb. During short exposure times

(less than 3 minutes), the test piece is acting a s a filter and lowering the light intensity,

causing the first band of data, until more neutrons are available to pass through and be

detected, causing the second band of data.

From Figure 4.10, because of the sunilar behaviour of al1 volumes of water, any

volume would give sùnilar results therefore, the intensity difference value for a chosen,

and possibly representative volume of water (0.14 mL) was plotted against exposure time

Page 104: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

and the results are s h o w in Figure 4.1 1. The mor for each ce11 was calculated fiom two

standard deviations as previously describeci. The error value ranged from 12% for a 30-

second exposure to 47% for the 10-minute exposure as shown in Figure 4.1 1. A sample

calculation of the error is included in Appendix D. Figure 4.1 1 will be used to detennine

the optimal exposure time.

intensity Drtferecrce kr a Single Cell of W m r atû.l4rnLOemsdtanDataib<enarFWC

Figure 4.11 - Intensity Dinerence of a Single Ceii at 0.14 mL of Water

In general, the intensity difference increases as the exposure tirne increases. An

increase in intensity difference of about 20% occurs fiom 0.5 to 4-6 minutes of exposure

tirne, while about 5% of intensity d i f fmce occurs nom 4-6 to 10 minutes of exposure

time. From the MTF resolution calculations, a 10% difference in intensity is the

Page 105: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

minimum threshold for human vision and therefore, the optimum exposure time for a

single ce11 of water would be 4 or 5 minutes. The camera system has the sensitivity and

capability of responding to a difference of 0.02 mL of distilled water and capable of

responding to a minimum volume of 0.02 mL in a single cell.

4.4.1.2 Water Ingress - Three Cells

Three cells containing distilled water simulate a typical water ingress situation in

a flight control surface. Varying the amount of water for each set of three cells gave a

good indication of the response of the radioscopy systern for large volumes of water. The

intensity difference was calculated for each group of cells for a given exposure tirne and

then graphed in Figure 4.12. It is expected that, as the exposure time and water volume

increase, so will the intensity difference value.

Page 106: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

3 Celh of W8?ar vs htsnsity Dtffenncs al VuylnO Expouire Times -iakriatwcatwPower

Dy-

O am a- aos am 0.1 a12 a i 4 ais ais 0 2 ozz 428 428 0 3 0.34 o s Volume of W i b r (mL)

- 10 min -9 mtn

-8 mm -7 mm 6 min -5 min -4 mm -3 min

-2 mm - 1 min

-30 s

Figure 4.12 - Effect of Varying the Exposure Time for Three Cells

The characteristics of Figure 4.12 are simila. to the single ce11 graph, Figure 4.10.

Using standard error calculations produced a relative error of approximately 4%

(corresponding to 0.02 rnL of water at a 30-second exposure) to 40% (corresponding to

0.34 mL at 10 minute exposure) of the intensity difference. The relative error was

smaller for three cells of water compareci to the relative error for the single cells. The

increased surface area of the three cells makes the detection of water easier and produces

more accurate resuits, thereby causing a reduction in relative error.

The cells that contained 0.02 mL of water produced good results with all

exposures times. The oscillation between volumes of water was also reduced, another

benefit fiom the increased area of the cells.

Page 107: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

As in pnor experiments, there were two bands of data - the 30 second to 1 minute

and the 2-10 minute exposures. This second band was much wider in its intemity

différence values. Sunilar to the single ce11 of water, the difference in background

intensities increased dramaticall y between the one-minute and two-minute exposure tirne,

while the intensity of the cells increased linearl y.

As the exposure tirne and volume of water increased, the intensity difierence

value increased. The intensity difference curves for a given exposure are positively

sloped, indicating a correlation between the amount of water in a ce11 and the intensity

diffaence. In general, the intensity difference values are lower for the three cells

compard to the intensity difference values for a single ce11 with the same exposure time.

It was easier to identiw several cells of water due to the increased surface area, even with

slightly lower intensity difference values. It is assurned that the effects of neutron

scattering caused the lowering of the intensity difference.

The intensity difference value for a representative volume of water (0.16 mL) was

plotted against exposure time and the results are shown in Figure 4.13. The emor for each

ce11 was calculated fiom two standard deviations as previously described. The error

ranged fkorn 7% for a 30-second exposure to 38% for the 10-minute exposure. Figure

4.13 will be used to determine the optimal exposure t h e for three cells of water.

Page 108: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

htnsity Diffemtnce for 3 CeUs at0.16mLDaiadIrom DotPtîknat FIMC

Figure 4.13 - intensity DWerence of Three Cells at 0.16 mL of Water

The data has been presented in the same manner as in Figure 4.11. In general, the

intensity difference value increases as the exposure time increases. An increase in

intensity difference of about 20% occurs fiom 0.5 to 5 minutes of exposure tirne, while

about 5% of intensity difference occur fkom 6 to 10 minutes of exposure time. The error

associated with the last four exposures (7 - 10 minutes) is rather large due to the

increased noise of the image background, which is caused by the integration of the image.

The three longest exposures produced the largest intensity difference values but they also

produced the largest amount of image noise, which might cancel the benefits of the

increased contrast. If a 10% reduction in intensity difference were assumed, as this

Page 109: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

amount of contrast is not detectable, then the optimum exposure for three cells of water

would be either 4 or 5 minutes. Since three ceils increase the contrasted area on the

image, it is easier to accept less than optimum intensity difference values in order to keep

the exposure time to a minimum.

4.4.1 -3 Water hgress - Camera Gain

To determine the effects of altering the digital gain of the camera, a . analysis was

conducted in the same manner as the tests descnbed in the Sections 4.4.1.1 and 4.4.1.2.

The gain was increased fiom a value of 1 to 16 by increments of four. From past

experiments, increasing the gain incrernentally by a srnall nurnber (a factor of one or two)

did not make a noticeable difference in the final results. Both single cells of water and

three cells of water were analysed as described in the previous sections. To determine the

effect of a gain increase, a single exposure tirne of 3 minutes was selected.

4.4.1.3.1 Water ingress - Camera Gain - Single Cells

Single cells in the test piece were incrementally filled with distilled water and

imaged at different camera gain settings. The intensity difference was calculated for each

ce11 for a given exposure time and then graphed (Figure 4.14).

Page 110: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

O J

O 0.05 O. 1 0.15 0 2 0.2s 0.3

Volume of Wabr In Cell (mL)

-4xGain 8 x Gain -m- 12 x Gain +Or Gain -16 x Gain

Figure 4.14 - Effects of Camera Gain on a Singie Celi of Water

The relative error for the above data ranges fiom 21% for the srnaIl volumes of

water and increases to 35% for the larger volumes of water. Al1 curves, except for the

largest gain of 16, were within each other's error band and therefore statistically, they are

identical. A large decrease in intensity difference was recordecl when the gain was set to

16. This decrease in intensity ciifference was produced fiom the increase in analogue

noise, which caused saturation of the exposure, making it difficuit to threshold a single

ce11 of water.

When the large relative error is taken into account, the benefits associated with

increasing the gain for a single ce11 are negligible and increasing the gain to maximum

limits had a detrimental effect.

Page 111: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

4.4.1.3.2 Water hgress Camera Gain - Three Cells

Three cells of water in the test piece were imaged at different camera gain

settings. The intensity difference was calculated for each ce11 for a given exposure tirne

and then graphed (Figure 4.15).

O L O 0.05 O. 1 0.15 02 0.25 0.3 0.35 0.4

Volume of Waîer In Ceüs (mL)

Figure 4.15 -Effects of Camera Gain on Three Celis of Water

The relative ermr for zero gain to a factor of 12 ranged fiom 18% to 26%. The

relative error for the Mages at a gain of 16 ranged fiom 45% to 57%. This large error is

causeci by image saturation. For three cells, the effect of increasing the gain up to a

setting of 12 is negligible, producing neither benefit nor detriment to the intensity

Page 112: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

difference for three cells. This conclusion is based on the close grouping of the curves in

Figure 4.15.

Overall, there is no visible benefit fiom increasing the digital gain of the carnera

as the results have shown that the intensity difference values for both the single and three

cells of water are negligible when the relative error is considered.

4.4.1.4 Water I n ~ e s s - Digital Enhancernent

Previous analysis (Section 4.4.1.1 ) has shown that the optimum exposure time for

the test piece was 4 to 6 minutes; therefore, that image of the test piece will be digitally

enhanced using several methods and then the results will be subjectively analysed. Since

digital enhancernent alters the original information, analysing histogram or intensity

clifferences proved futile in deteminhg the best image. The method used in detennining

the optimum image after enhancements will be a subjective cornparison of hown

features from the test piece to identifiable objects in the image. The image will be

analysed for overall image clarity, amount of background noise (random white spots),

presence of the honeycomb pattern, presence of water, and variations in intensity due to

volume changes in water. The following types or methods of enhancernent techniques

were anal ysed:

Contrast, gamma and brigbtness, Equalisation of histogram, Erode filter, J Flat filter, Median filter, Mu1 tipl ying images, Ratio flat, and Optical density method.

Page 113: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

A11 eight enhancement techniques were applied and the results are presented in

Appendix E. Due to the grey scale limitations of the printer and size of image, the

images in Appendix E have poorer resolution on paper than on a computer monitor. Only

the original image of the test piece and the best-enhanced images will be discussed. The

best enhancement methods producing the best images when compared to the original

image of a 4-minute exposure (Figure 4.16) were the equalisation of histogram (Figure

4.17), the Erode filter (Figure 4.18) and the J Flat filter (Figure 4.19).

a) 4 Minute Exposure b) 4 Minute Exposure w/Contrast Changes

Figure 4.16 - Four-minute Exposure of Test Piece

The original image is black as shown in Figure 4.16a and, with contrast

manipulation (Figure 4.16b) the test piece becomes visible. The minimum water volume

Page 114: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

visible is 0.06 mL for the single cells and 0.04 mL for the group of three cells. There is a

considenble arnount of noise in the image, making fine details difficult to see.

a) Equalisation of Histogram b) Equalisation of Histograrn W/ Contrast Changes

Figure 4.17 - Results from Equalisation of Histogram

When the original image is enhanced by the equalisation of the histogram

technique, the image has a significant arnount of noise (in the f o m of dark grey areas)

(Figure 4.17a) but, when contrast is applied (Figure 4.17b), the noise is reduced and the

honeycomb pattern is apparent. The minimum volume of water visible is 0.04 mL for the

single cell and 0.02 mL for the three cells. A significant amount of contrast enhancement

was required to reduce the noise and this contnst washed out the background of the

image leaving only the outline of the test piece.

Page 115: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

a) Erode Filter b) Erode Filter w/Contrast

Figure 4.18 - Results from Erode Filter

Applying the erode filter does not lighten the image as seen in Figure 4.18a and

therefore contrasting was applied to produce a visible image in Figure 4.18b. Al1

volumes of water were detectable and the noise is reduced considerably. The outline of

the test piece is recognisable, but the honeycomb pattern is not visible. This is the best

enhancement technique for detecting water in the structure. as it is capable of showing

the various volumes of water.

Page 116: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

a) J Flat Filter b) J Flat Filter w/Contrast

Figure 4.19 - Results from J Flat Filter

Utilising the J Flat Filter produces a visible image without applying contrast

enhancements as seen in Figure 4.19a. Both images tend to show a large amount of

noise. In Figure 4.19b, the minimum amount of water visible for a single ceil is 0.04 mL

and for the three cells is 0.03 mL.

The equalisation and erode filter are the easiest of the three enhancement

techniques to implement as they are usually integrated in a software package while the J

flat filter requires that another image be taken and added to the original. Of the three

techniques, the erode filter is the best enhancement technique for water detection.

The results from the above experiments were used to determine that:

Page 117: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

a the optimum exposure time for the test piece containing water is 4-6 minutes;

b. the minimum volume of water detectable is 0.02 mL;

c. the minimum detectable incremental volume of water is 0.02 mi,;

d. the optimum carnera gain setting is zero; and

e. the best digital enhancement method for water detection is the erode filter.

4.4.2 Simulated Water

To determine the effects of different flux levels, the same test piece was used but

PE replaced the water. Images were taken at RMC at half reactor power (1 0 kW), and at

Pem State at 10 kW and 100kW. The images were processed in the same manner as in

Section 4.4.1 which produced graphs with intensity difference values for one ce11 and

three cells of PE. Only the data for the 10 mm PE will be analysed to determine optimum

exposure time and effecr of neutron flux changes. Similar to the other sections, single

cells and grouped three cells will be analysed.

4.4.2.1 Simulated Water - Single Cell

The results fiom calculating the intensity difference for a representative height of

PE (10 mm) are shown in Figure 4.20. The calculation of the relative error for Figure

4.20 was calculated in the manner as described in Appendix D. Inserting the relative

error onto the graph made it difficult to read and, therefore, the average relative mor is

indicated in the legend.

Page 118: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

rtne (min)

Figure 4.20 - Results of Varying Flux Levels for a Shgle Ceii of PE

In general, al1 three curves increase with exposure time. The curve nom the RMC

data increases constantly until becoming asymptotic at the 9-minute exposure time. The

beginning of the RMC curve increased about 37% for the first 5 minutes and then raised

o d y 13% for the remaining 5 minutes, making a total intensity difference increase of

approximately 50%. Accounting for image noise and the lack of ability to differentiate

between 10% contras4 an optimum exposure time would be 6 minutes.

The c w e fiom the Penn State data (10 kW) increased rapidly and became

asymptotic at a 5-minute exposure time with the dope of the cuve increasing again at the

10-minute exposure time. The beginning of the Penn State (10 kW) curve increased

Page 119: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

about 45% for the £kt 5 minutes and then raised only 1% or 2% for the remainùig 4

minutes (the 1s t data point is ignored) making a total intensity difference increase of

approximately 47%. Accounting for image noise and the 10% contrast factor, an

optimum exposure thne would be 3 minutes.

The cuve fiom the Penn State data (100 kW) increased rapidly and becornes

asymptotic at the 4-minute exposure tirne. Applying the same logic as before, (noise and

10% factor) an optimum exposure tirne would be a 2-minute exposure.

Both lower fiux values (RMC and Penn State (10 kW)) produced the largest

ïnmease in intensity difference over the 5 minute range. The data fiom W C produced a

37% increase in intensity difference while the data fiom Penn State (10 kW) produced a

45% difference over 5 minutes, cornpared to data fkom Penn State (100 kW) which

produced an approximately 20% increase.

Two trends are evident due to the increasing flux level. First, the optimum

exposure time for the flux level at RMC (2 to 3 x 104 n/cm2 s) would be 6 minutes, but

decreases to 3 minutes for the 10 kW power setting (3.0 x 105 n/cm2 s) and decreases

M e r to 2 minutes for the 100 kW power seîthg (3.0 x 1 o6 n/cm2 s); therefore, as flux

inmeases, optimum time decreases. Second, the intemity difference value for a given

exposure time increased with increasing flux values.

Page 120: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

4.4.2.2 Simulateci Water - Three Cells

The r d t s from calculating the intensity difference for a representative height of

PE (10 mm) are shown in Figure 4.2 1. The method for calculating the relative mor and

its placement is identical to the graph from the previous section (4.4.2.1).

hmnsity Difbr9n~9 for Varying Flux Values (hgQpdlOmm(pr3CaIsdPE)

-- -

Figure 4.21 - Results of Varying Flux Levels for Three Cells of PE

As previously observed, al1 of the curves inaease as the exposure tirne increases.

As well, the curves are becoming asymptotic with the longest exposure tirne. As seen

previously (in Figure 4.20), the intensity difference increases with exposure the. The

poor fit of the curve to the Penn State (100 kW) data is due to the large variations of the

intensity difference values caused by the large mor (39.3%) associated with these data.

Page 121: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

For a given curve, the intensity difference values are less than the same curve from the

single ceil graph in Figure 4.20.

The curve fiom the RMC data increases constantly, but does not appear to

becorne asymptotic at the 10-minute exposure tirne. The beginning of the RMC curve

increased about 26% for the first 5 minutes and then rose only 12% for the remaining 5

minutes, making a total intensity difference increase of approximately 3 8%. Accounting

for image noise and the lack of ability to differentiate between 10% contrast, an optimum

expotaire time would be 7 minutes.

The curve from the Penn State data (10 kW) increased rapidly and became

asymptotic at a 5-minute exposure time with the cuve increasing in slope again at the 10-

minute exposure tirne. The second nse in the curve is from fitting a third order

polynomial c w e to the data. This conclusion is derived fiom the fact that only the Penn

State (10 kW) cuves (for the single ce11 and three cells) have this characteristic and,

therefore, this anomaly should be ignored. The beginning of the Penn state (10 kW)

cuve increased about 45% for the f3st 5 minutes and then rose only 1% or 2% for the

remaining 4 minutes (the 1s t data point is ignored) making a total intensity difference

increase of approximately 47%. These values are the same as the single ce11 PE for this

power setting. Accounting for image noise and the 10% contrast factor, an optimum

exposure t h e would be 3 minutes.

The curve fiom the Penn State data (100 kW) increased rapidly and becomes

asymptotic at the Cminute exposure time. Applying the same logic as before (noise and

10% factor), an optimum exposure time would be 2 minutes.

Page 122: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

The lower flux value fiom Penn State (IO kW) produced the largest increase in

intensity difference over the 5 minute range. Compared to the data fkom RMC which

showed approximately 26% increase in intensity difference, the data fiom Penn State (10

kW) showed approximately 45% ciifference and that data fiom Penn State (100 kW) had

approximately 22% increase.

The optimum exposure time for the flux level at RMC (2 to 3 x lo4 n/cm2 s)

would be 7 minutes, but decreases to 3 minutes for the 10 kW power setting (3.0 x 10'

n/cm2 s) at Penn State and decreases M e r to 2 minutes for the 100 kW power setting

(3 .O x 1 o6 n/cm2 s). Therefore, as flux increases, the optimum time decreases.

There was a good correlation between optimum exposure times for the single ce11

and the three cells for a given flux level. The small increase of one minute to the

optimum exposure at the lowest flux level for three cells of PE compared to a single ce11

is not significant and is within the experimental error. Since the target is larger than the

single cell, reducing the exposure time to one minute would not degrade the image

contrast significantly and therefore an optimum exposure t h e for the three cells would

be 6 minutes.

Overall, as the flux increases, the intensity difference, which is the contrast

between the PE and the test piece, increased significantly, for both the single and three

cells. The optimum exposure tirne decreased as flux values increased £kom seven minutes

to 2 minutes. The calculation of the optimum exposure time was used to understand the

effects of the camera system for varying flux levels. These results will not be used to

determine the optimum exposure time at RMC as PE only simulated the effects of water

Page 123: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

ingress. In practical terrns, as the flux values increased, the easier it was to see the pieces

of PE which simulated water ingress. Logically then, water shoutd also be easy to see at

these higher flux levels.

4.5 CF188 Rudder

To understand the results of the neutron radioscopy performed on the CF188

nidder, a basic knowledge of the interna1 structure is required. A print of a neutron

radioscopy image of a serviceable section of a rudder with its components labelled is

shown in Figure 4.22 (Note that more detail is observed in the actual computer monitor

image). In Table 4.5 is listed the components corresponding to the capital letters in

Figure 4.22. Al1 images from this section are shown in a Iarger format in Appendix F.

Figure 4.22 - Exposure 2 of Rudder frorn Penn State

Page 124: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Since the neutron beam at Penn State did not completely cover the entire area of

Table 4.5 - Interna1 Components of CF188 Rudder

the scintillation screen, d l RMC images were cropped so that the images can be

Letter A B C D E

compared. Only the data inside the circle were used in the histograms as the data

Component Al-um Honeycomb pattern fiom FM300 adhesive Porous Adhesive - Connects honeycomb to main spar Main Spar - Aluminium material 1 0/32 Steel Anchor Nuts (four in total) 0.050 inch Aluminium Sheet - Fairing support

associated with the dark area outside the circle were excluded fiom the histogrms. Al1

images are presented in a format similar to Figure 4.22 and contain similar intemal

components as indicated in Table 4.5.

Figure 4.23 shows a series of images taken at RMC of Exposure 5 taken at a

4-minute exposure. The first image (a) is without any data manipulation and the second

image (b) had the conîrast and brightness manipulated to make the original image visible.

The 1st image (c) has had a 2 x 2 erosion filter applied to the original image and then the

brightness and contrast were manipulated.

Page 125: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

(a) no enhancements (b) basic enhancemats (c) advanced enhancements

Figure 4.23- Sequence of Exposure 5 Images taken nt RMC (10 kW)

Although not visible, in Figure 4.23% there are some white spots indicating

gamma contamination or system noise. Appendix F, Figure F.2, clearly shows the

evidence of system noise. Figure 4.23b has very little definition, as some of the

individual components shown in Figure 4.22 are not clearly distinguishable, but water is

indicated, as is the hinge assembly. Water ingress and its location were previously known

liom neutron radiography. In Figure 4.23 b and c, water ingress is visible and the pattern

of water is identical to known images of this area. Figure 4 . 2 3 ~ also indicates clearer

details such as the main spar, porous adhesive and the anchor nuts.

The histogram associated with the original image (Figure 4.23a) of exposure 5

taken at RMC is shown in Figure 4.24. Thresholding was applied to the image to identiS

the features related to the curve and are indicated in Figure 4.24.

Page 126: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

O

1024 1536 2048 2560 3072 3584 4096 inœnsltll Value

Figure 4.24 - Histogram of Exposure 5 from RMC (10 kW)

The intensity values of the graph in Figure 4.24 are in the low thousands and

therefore the original image is totally black as previously seen in Figure 4.23a. The

intensity values have a small range, from 1536 to approximately 3500, which is only 3%

of the total number of intensity values available. This small intensity range produced an

image with a small dynarnic range, which reduced image contrast between components as

shown in Figure 4.23b.

Figure 4.25 shows a series of images of Exposure 5 taken at Penn State with the

reactor at 1 0 kW and an exposure time of 2 minutes. The original image (a) is without

any data manipulation and the second image (b) had the contrast and brighmess

Page 127: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

manipulated to make the original image visible. Due to the increased flux levels, no

enhancement filters were requred.

(a) no enhancements (b) basic enhancements

Figure 4.25 - Sequence of Exposure 5 images taken nt P ~ M State (10 kW)

Again in the original image, Figure 4.25% no details are visible without some

contrast and brightness alterations appIied to the image. Al1 the major components are

visible in the enhanced image as seen in Figure 4.2% There are visible intensity

differences fiom the main spar, a porous adhesive area and water. The honeycomb

pattem is dso visible in both areas, with and without water.

The applicable histogram of Figure 4.25% with labels indicating the major

components of the rudder, is presented in Figure 4.26.

Page 128: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Penn StibeData8t10 kW-Exp5of Rudder @ -30 C, 512 x 512 d U h , Wi inrsiel-16

Figure 4.26 - EIistogram of Exposure 5 from Penn State (10 kW)

The intensity values of the graph in Figure 4.26 are in the low thousands and

therefore the original image is totally dark as previously seen. The histogram covers 4%

of the available range of intensity values, fiom 1 536 to approxùnately 4 1 20, producing an

image with a larger dynamic range than the image fkom W. This larger range

indicates that the intemal components should produce visible density differences on an

image, which c m be seen in Figure 4.2%

An image of Exposure 5 taken at Penn State with the reactor at 100 kW and an

exposure time of 1 minute is shown in Figure 4.27a and b. At this power setting? the flux

produced was two orders of magnitude greater than available at RMC. Due to the higher

Page 129: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

flux values, the only enhancement required was a small change to the contrast and

brightness to produce the image in Figure 4.2%.

(a) no enhancements (b) with basic enhancements

Figure 4.27 - Sequence of Exposure 5 images taken at Penn State (100 kW)

Unlike the previous undtered images, the rudder components are visible and,

adjusting only the brightness and contrast, a good image of Exposure 5 can be seen in

Figure 4.27b. Al1 intemal components show good dennition and dynarnic intensity

range. The honeycomb pattern c m be clearly seen in a larger image in Appendix F,

Figure F.8.

The applicable histogram for Figure 4.27a, with labels indicating the major

components of the rudder, is presented in Figure 4.28.

Page 130: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure 4-28 -Eistogram of Exposure 5 from Penn State (100 kW)

This histogram covers 14% of the total available intensity scale, fiom

approximately 3000 to 12000. Since the histogram spans past the intensity value of

10,000, the image should be visible as can be seen in Figure 4.2%. The large dynarnic

range of this histogram should translate to significant differences in intensity values for

given interna1 components of the rudder, which can also be seen in Figure 4.27b.

The image produced at a 100 kW power setting (Figure 4.27) has the best

definition and the largest range of intensities compared to the other two images taken at

RMC and Penn State at 10 kW. Al1 three images fiom RMC (10 kW) and Penn State at

10 and 100 kW show the intemal components of the rudder water ingress, Figure 4.29.

Page 131: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Due to the clarity of the images produced fkom the Penn State reactor at 100 kW, these

images can be used as a cornparison standard.

Figure 4.29 - Best Results of Exposure 5 from (a) RMC at 10 kW, (b) Penn State at 10 kW, and (c) Penn State 100 kW, with image enhancements

Plotting nomalised histograms of Exposure 5 at RMC (1 0 kW) with an exposure

time of 4 minutes and Exposure 5 at Penn State (100 kW) with an exposure time of 1

minute produced results shown in Figure 4.30. Both sets of data were normalised to the

centroid of the honeycomb peak. In Appendix G, the images for each rudder exposure

were correlateci to the applicable histograms, similar to Figure 4.30.

Page 132: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

--+ +- Horieycomb

1 .

- RMC -min - Pem State 100 kW EYP 5 1 min

0.6 -

a6 - 1 1 Hqe 1-r ~ d h ~ 3 ~ ~

r\ A

Anchor NuLs a4 8

0.2 *

0 --- O Q 1 0.2 a3 a4 a5 a6 o. 7 as 0.9 1

Nomrlltad hbmrfly Vilue

Figure 4.30 - Normaiised Graph at Two Different Power Settings

Normalising the data allowed a direct comparison between the two graphs and

compensateci for the large magnitudes of pixel count and intensities. The goal in

normalising the data was to confirm that both images produced sirnilar shapes and

features that were related to the intemal components of the nidder. Although the

histogram graphs are not identical, the shapes are similar and features on the graph do

relate to specific items in the images. The diffaence between the two graphs shows that

the histogram fkom the image taken at Penn State uses a larger percentage of the intensity

scale. The curve of the histogram fiom the image taken at RMC is compressed on the lefi

side of the peak and elongated on the nght side of the peak relative to the other

histogram. Overall, the image taken at RMC contains al1 the required data to produce an

Page 133: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

image that is capable of indicatuig the honeycomb pattern, main spar, porous adhesive,

anchor nuts, duminium structure and water ingress.

4.6.1 Resolution

The resolution of CX and SR radiography film was detemüned utilising the MTF

technique and then these results were compared to the resolution calculated for the

radioscopy system. The CX film was exposed for 18 minutes and the SR film for 135

minutes. The 10% MTF values were calculated as descnbed in Section 2.8.3. The Ut

values and calculated cyclelmm values, with absolute mors for both types of film, are

shown in Table 4.6.

r I I

SR- 135 1 39.3 1 1 0.1 1 + O.OU-.O1 (1 8Yo/9%) 1 170 (+ 201 - 20)

Table 4.6 - Resolution Variables for Film at RMC

These resolution calcdations indicate that both films have very similar resolution

Exposure t h e 1 PixeYmm CX - 18 minutes 1 39.3 1

values when the error is taken into account, with the SR film having a smaller resolution

variation than the CX film. The resolution of SR film cdcdated fkom half a cycle is

ut (mm) 0.12 + 0.02 (16%)

Cycles/mm at MTF Values of 10 % 160 (+ 28/ - 20)

Page 134: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

0.0029 +0.0001/- 0.0003 mm and the resolution for CX film is 0.003 1 +0.0003/- 0.0005

mm.

Comparing the resolutions of radioscopy and radiography against expowe time

produces the results shown in Figure 4.3 1.

Worklng Rem Won 8t RMC of NRS

Figure 431 - Resolution Comparison

The largest resolution area associated with the Weutron Radiography at RMC"

indicates the estimated range available for film. The maximum value of the range was

calculated by using the full cycle resolution value for CX film, and the minhum value of

the range was calculated by using the smallest half cycle resolution value £kom the SR

film. The exposure time was calculated for "Neutron Radiography at M C " by

Page 135: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

arnalgamating the exposure times for both the CX and SR film, which created a range of

possible exposure thes .

The resolution calculated for both films used the maximum and minimum half

cycle values and the exposure times are actual times used at RMC for a given type of

film. The exposure tirne for the 'heutron radioscopy at RMC" indicates a range of

exposure times (fiom 30 seconds to 10 minutes) rather than a single optimum exposure

time. Graphing the optimum exposure time would make the area associated with film

look even larger and therefore, a range of possible exposure times was shown. Figure

4.3 1 shows the general inverse relationship between resolution and exposure tirne. It also

shows that radioscopy has fast exposures at the cost of poorer resolution, while film has

very fine resolution at the cost of long exposure times.

The results fkom the modifiai ASTM E545-75 Type A were not significant as the

SR film showed al1 12 holes while the CX film showed 11 holes. The holes were too

large in the standard to be useful for the SR film. For film, the SI for the ASTM E545-91

should be used.

4.6.2 Time and Cost Study

For a cost cornparison between neutron radiography and radioscopy at RMC,

capital cos& will not include items that are now considered part of the Facility such as the

reactor, the beam stops, positionhg system and permanently installed radiation meters.

These items are not included, as both techniques require thern for inspection of flight

Page 136: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

control surfaces using a neutron beam. The capital costs for both techniques are shown in

Table 4.7.

Table 4.7 - Cornparison of Capital Cost of Radioscopy and Radiography

- -- - -

The estimate of total cost for both techniques show that radioscopy is

approximately 23% greater for the initial capital purchase. Capital costs are usually a

single cost, while consumable costs are small but ongoing.

For consumables and labour, the mst per image is indicated in Table 4.8. The

operating costs of the reactor (Le., fuel usage, operator salary, etc) was not included in

this analysis as the reactor is used simultaneously for the application of NAA.

Radiography 1 Cost

Cassettes & Screens(2) Film Processor Vacuum Pumps (2) BP1,SI Densitometer Dark Room Oight, timers, door etc) Total

lbdioscop y 1 Cost (Cdn dollars)

2 x 5930 16 000 2 x 300 4000 2000 2000

$36 460

CCD Camera System Image software Scintillation Screen Bellows Cornputer Volt meter

Total

(Cd. dollars) 32 920 4000 5000 710 2000 300

$44 930

Page 137: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Table 4.8 - Cost per Image Radiography Cost per image

CX Film (50 shcas/box )

AA Film (50 shects/bo~)

Developer Fixer Density Strip Misc. (~nvclopes, ~IOVCS,

Costhnage (CX)/(AA)

Cost per image (Cdn dollars)

As indicated in Table 4.8, radioscopy is about 6 to 10 t h e s less in cost per image

Radioscopy

($240/50) = 4.80 ($583/50) = 1 1 -66 ($66/100') = 0.66 ($36/1001) = 0.36 ($1 56/500L) = 0.3 1 Estimate = 0.50

$6.63/ 13.49

Labour costhnage (CWAA) Total codimage (CX/AA)

due its simpler image storage method and shorter exposure time. The final cost analysis

Image storage (np disk)

Printing image ciara)

Costhage

is related to a time study that was based on an actual practice and as close as possible

Note: 1 - Estirnated number of images 2 - Labour wst at %38/hour for a 18 minute exposure (CX) or 25 minute acposure (AA) 3 - Labour cost at S38hour for a 4 minute expomre

$1 1.40/ 15.83 $18.03129.32

replicates the real activities associate- with that component. Figure 4.32 incorporates

two flow charts, showing the steps required in taking a single image of a rudder using

Labour cost/imap;e Total costhmage

radiography and radioscopy, and the time required for each step.

$2.53 $3.13

Page 138: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure 432 - T h e Study Flow Chart for Rudder

It takes eight images and two position changes to radiograph a CF188 rudder.

Using the flow chart for radiography, a total of approximately 204 minutes (using an 18-

minute exposure time) is required to radiograph a rudder. It takes approximately 76

minutes (using a 4-minute exposure tirne) to image a rudder using radioscopy. Using

radioscopy, there was a 2/3 reduction in person hours compared to radiography and a

78% reduction in reactor time ([(18 minutes * 8) - (4 minutes*8)]/ (18 minutes * 8) * LOO

= 78%). The method used to determine the total time and cost to radiograph a rudder was

extrapolated for the remainuig CF188 flight control surfaces. The tirne and cost

appreciation for each component is shown in Table 4.9, based on usage of CX film.

Page 139: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Table 4.9 - Operational Cost and Inspection T h e of Radiography and Radioscopy 1

1 Flight Control Surface 1 Operational Inspection Time 1 Operational cost ( labour + 1

Although neutron radioscopy is iaitially more expensive than neutron radiography

(23%), the cost per image is less expensive (83%). A tirne study has indicated that when

radioscopy is used to inspect a rudder, the total inspection time is reduced by

approximately 60%, translating into a total cost savhgs of 70% relative to neutron

radiography. These savings can be extrapolated to a complete inspection of al1 twelve of

the flight control surfaces, as indicated at the bottom of Table 4.9. Considering the

higher capital cost of radioscopy (Table 4.7) with its lower operating cost (Table 4.9) the

break even point is at about three sets of aircrafk components. Moreover, neutron

radioscopy requires about 78% less reactor time to image a nidder than neutron

radiography, fkîher reducing the cost of this technique. It is the reduction in labour and

reactor time that is significant, making radioscopy economically favourable.

-

Inboard Leading Edge Flap* Outboard L&g Edge Flap* Trailinp; Edge Flap* Ailerons* Horizontal Stab* Rudder

Total for half set Total set (complet e aircraf t)

Note: * - extrapolated inspection and cost estimates, # - based on CX film usage

(mulutes) ~adiographf

33 1 129 454 179 708 204 2005 4010

consumables) Radioscopy

123 49 166 67 206 76 74 1 1482

~adiogra~hf $295.85 $1 14.86 $406.9 1 $159.79 $634.1 O $182.26 $1793.76 $3587.52

Radioscopy $85.70 $34.03 $1 15.79 $46.63 $181.47 $52.93 $5 16.70 $1033.40

Page 140: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

CHAPTER 5

CONCLUSION

5.1 Introduction

The Neutron Radiography System (NRS) installed on the SLO WPOKE-2 Facility

at RMC has the ability to inspect CF1 88 aircrafi flight controI surfàces for water ingress.

The low flux value of 2 to 3 x 104 n/cm2 s at the image plane necessitates using a fast film

in order to have a reasonable exposure t h e to produce a image with a sufficient

background density. A fast film (CX) has been adapted to reduce exposure times to 18 to

25 minutes fiom the standard film (SR) which takes more than 2 hours. A CCD camera

and scintillation screen combination could replace the film and produce images with lower

exposure times without degrading the ability to detect water ingress into the aircraft flight

control surface. The goal of this thesis was to install. characterise and develop a near real-

tirne radioscopy system on the NRS at RMC in order to expedite the inspection of

composite aircraft flight control surfaces.

5.2 Installation of Hardware

The first phase of the study was the installation of the hardware, which included

the CCD camera, mirror, lem, scintillation screen, iight-tight enclosure and image

processing software. This phase was a relatively straight fonvard endeavour, as the

Page 141: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

technology tends to be an assembly of off-the-shelf cornponents and requires the user to

become familiar with the operation of the software and hardware.

Characterisat ion of the Svstem

The second phase was the characterisation of the CCD camera system to detemine

t s capabilities and performance. Images taken to establish the system's capabilities did

not have a component or object between the neutron bearn and the scintillation screen so

that only the camera and scintillation screen were analysed. Parameters that affected the

performance of the CCD camera were the relative darkness of the difiùser, the camera

temperature, the quantity of neutron flux, the exposure time of an image and the

resolution.

When "dark images" were taken, the darkest images were produced by a capped

camera lem (i.e., no diffuser) for a given exposure tirne, followed by the portable difiser

used at P ~ M State and the h e d diffuser used at RMC. Producing an image with an

intensity value of zero is not obtainable due to system noise and dark current. The lowest

the carnera system could achieve was a peak intensity value of 1140, which was taken

without a diffuser and at a 30 second exposure tirne.

The camera temperature had a large effect on the system's performance. For a dark

image, as the temperature of the camera decreased, the image became darker and became

more centralised around the peak. At -1 5 OC, the peak intensity value was recorded as

1397 and, when the temperature was lowered to -3 0 OC, the peak intensity value dropped

to 1205, for a 2-minute exposure t h e . This drop in peak intensity values is directly

proportional to a reduction in dark current.

Page 142: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

The neutron flux was also varied by using two different reactors and power levels

and, as the neutron flux increased, the overail intensity of the unified image also increased,

producing a larger dynamic intensity range. The unified images taken at RMC at 10 kW

produced peak intensity values korn approximately 1400 to 4000 depending on the

exposure time. The dynamic range produced fiom the intensity values fiom the dark

image and unified image varied fiom approxirnately 200 to 2000. A saturated image was

acquired at Penn State at 100 kW (3.0 x 106 n/cm2 s at the image plane) nom a 5-minute

exposure.

In general as the exposure time for an image increased for either situation (dark

image or unified image), the intensity value increased and became less centralised around

its peak. Long exposure times produced images with a large range of intensity values with

additional system noise; conversely, very short exposure t h e s produced Mages that

lacked detail due to low contrast conditions.

To measure the resolution, the MTF method was used at M C , and a line pair

gauge was used at Penn State. To measure the system's sensitivity, a modified ASTM

E545-75 Type A indicator was used. For exposure times ranging from 30 seconds to 10

minutes at RMC, a measured resolution of 0.47 (+0.34/-0.14) cycles/mm at 10% MTF was

calculated. The line pair gauge recorded a resolution of 0.656 Ip/mm for both power

settings and al1 exposure times at Penn State. At RMC, the measurable spatial resolution

for the CCD system was 1 .O6 mm, which is half of the d c y c l e measurement.

In practical terms, this series of analyses showed that the dark image should be as

black as possible; therefore, every effort should be made to eliminate light leaks, reduce

the camera temperature, and reduce the absorption and reflective surfaces in the diffuser.

Page 143: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

The analysis has also indicated that the unified image should be as bright as possible;

therefore, the greater the neutron flux, the brighter the screen, recognising that CCD pixel

well saturation is a h i t i ng factor.

Ideally, the further the intensity values of the dark image are nom the unified

image, the greater the dynamic range of the systea A large dynamic range translates into

a larger intensity separation between dfierent materials with different neutron absorption

and scattering properties.

A method of revealing the intensity values of observed indications in the

radioscopy image was devised. A histogram was made of the image such that the

dynamic range is indicated between the dark and unified image. Within this range,

indications such as moisture and components of the object cm be displayed.

5.4 Develo~rnent of Neutron Radiosco~v Svstem

Once the NRS was characterised, it was then tested by irnaging test pieces and then

a flight control surface. A honeycomb test piece with water inserted into it was used to

detennine the optimum exposure t h e , camera gain setting and digital enhancement

technique. Using the best possible camera settings determined fkom the test piece, a

CF188 rudder was then imaged at three different flux values. The images taken at the

higher flux values at Penn State were compared to the images taken at RMC to detennine

the quality of these images.

Images taken of the nidder were transformed into histograms and normalised. This

normalisation around the peak centroid allowed the same image of the rudder to be

compared at different flux levels. Thresholding was applied to the images, which allowed

Page 144: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

the different characteristics on the histogram to be identified. It was discovered that water

has a characteristic location on the histogram for all flux levels.

The optimum exposure time for radioscopy at RMC was detennined by

comparing the intensity diffaence (cell intensity value subtracted fiom the background

intensity value) for varying exposure tirnes. This cornparison was completed for a set of

single cells and a set of three cells, with each cell or cells having varying volumes of

water. For the water indication, a 4 to 6 minute exposure time for both the single ce11 and

the three celk produced the best image, having good contrast between the cell and

background.

Using intensity dflerence to determine the effects of camera gain, a 3-minute

exposure t h e showed that no measurable benefit was produced, therefore, it was

concluded that the gain should left at the zero setting.

A large number of enhancement techniques were applied to images of the

honeycomb test piece taken at a four-minute exposure time with no camera gain. Of al1

the enhancement techniques applied, about t en techniques produced reasonable images.

Of the ten, three techniques produced very good images, these being equalisation of the

histogram, a J flat filter and an erode fiher. Of the three, the erode filter produced an

image that reduced the noise considerably and indicated the various volumes of water for

single cells and three cells.

The images of the CF188 rudder taken at RMC were compared to images taken at

the Penn State facility that had flux levels at the image plane two orders of magnitude

larger. Images produced at the Penn State facility with the same camera system showed

larger intensity dflerences between intemal components and better quality images overall,

Page 145: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

at a quarter of the exposure time compared to images taken at RMC. Enhanced images of

the CF188 rudder taken at RMC have very little system noise, show all internal

coqonents, indicate some contrast differences between internal components, and clearly

indicate water ingress.

Analysing the CCD camera based neutron radioscopy system installed at the

SLO WPOKE-2 Facility at the RMC has shown that it is:

1. a viable technology that is capable of producing quality neutron images at a

very low neutron flux level;

2. a capable technology that can produce images at a shorter exposure time

relative to film;

3. a sensitive technology, capable of detecting small amounts of water;

4. a cost effective technology relative to film based systems; and

5. a viable technology for inspecting CF1 88 flight control surfaces.

Therefore, the SLOWPOKE-2 Facility at RMC has a neutron radioscopy system

that has met and exceeded al1 performance objectives and is now ready to supplement the

film based inspections carried out on CF188 flight control surfaces.

Page 146: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

CHAPTER 6

RECOMMENDATIONS

Recommendat ions

The following are the recomrnendations for the CCD camera-based neutron radioscop y

systern installed at the SLOWPOKE-2 Facility at the M C , based on the work completed

for this thesis.

1. The radioscopy system should be used as an initial screening inspection of the

CF188 flight control surfaces so that water ingress can be located. Then (the more

costly) neutron radiography can be used in these locations that have indicated

problems, so that higher resolution and permanent images of the area can be taken.

Using neutron radioscopy in this fashion eliminates the expensive process of

taking radiographie images of areas that have no indication of water ingress.

2. An investigation into applying pattern recognition or neural network techniques on

the histograms, taken from the neutron images of flight control surfaces, to analyse

automatically the area on the histogram associated with water ingress should be

undert aken.

3. An increase in flux at the image plane would provide a slightly lower exposure

tirne, less requirernent for enhancement and, most importantly, a larger dynamic

range which would reveal the presence of water in the histogram. With the

Page 147: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

SLOWPOKW-2 operating at full power (20 kW), the tlux should be about halfof

Penn State at 10 kW. It is recommended that the flux at the image plane be

measured and the shielding required be assessed regarding the feasibility to

perfonn neutron radioscopy at full power (20 kW) at RMC.

Page 148: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

REFERENCES

Standard Practices for Thermal Radiomwhv of Materials, American Society for Testing and Materials, ASTM €748-95, pg. 2,1996.

Kobayasi, H., Neurron Radiography, 3, Kluwer, pg. 42 1, 1990.

Lange, KI., Leeflang, H.P., Mark af, J.F.W., Neutron Radiosco~v usinn Electronic Cameras with Zoom F Ca~abiliw and Hi& Detectivity, 6 European Conference on Non Destructive Testing, pg. 8 17-82 1, 1994.

Casali, P., Chiro, R , Partemi, Neutron Imagine Usinp; a Low-Cost Cooled CCD Camera, Nuclear Energy in Central Europe: Present and Perspectives, Portoroz, Slovenia, pg. 566575, June 1993.

Brenizer, J.S.,Berger, H., Stebbings, C.T., Gillies, G.T., Performance Charactenstics of Scintillators for use in an Electronic Neutron Imaginn Svstem for Neutron Radiom~hy, Science Instruments, Vol. 68, No. 9, pg. 337 1-3379, September 1997.

Berger, EL, Polanslq, D., Brenizer, J.. Bueno, C., Electronic Neutron Imaaq, Indusnial Quality Inc Final Report, Report Number N68936-93-C-0274, January 1994.

Czirr, J.B., MacGillivray, G.M, MacGillivray, RR, Seddon, P.J., Performance and Characteristics of a New Scintillator, Nuclear Instnunents and Methods in Physics Research, A 424, pg. 15-19, 1999

Brenizer, J.S., Berger, iX, Gillies, G.T., Mengers, P., Stebbings, C.T., Polansky, D., Rogerson, D.J., Develo~ment of a New Electronic Neutron Imaging System, Nuclear Instruments and Methods in Physics Research, pg. 9- 13, March, 1999.

Lanza, RC., The Use of Neutron Tomoerraphv Techniques for the Detection of Corrosion D a m e in A i r d Structures, Amencan Society of Mechanical Engineers, Aerospace Division. AD-04. Vol. 47, pg. 253-259, 198 1.

10. Balasko M, Svab E, Neutron Radioeraphy in Research and Developinent, Nukieonika 39, No 1-2, Pg. 3-22, 1994.

1 1. Leeflang, KP., Markgaf, J.F.W., Detection of Corrosion on Airmafi Component by Neutron Rad iom~hy , Non-destructive Testing Evaluation, Vol. 1 1, pg. 1 15- 128, 1994.

12. Dupuis, J.P., Huriet, J.R, Industrial Inspection of Aeronautic Samples - First step with the DIANA, 5 World Conference on Neutron Radiography, pg. 657-664, June 1996.

13. Bickerton, M., Deveio~ment of Im~roved Techniaues for the Neutron Radioai-aphv of CF188 Flinht Control Surfaces, Department of Chemistry and Chernical Engineering, Royal Military ColIege of Canada, Kingston Ontario, 1 998.

14. Lanza, RC., Shi, S., McFarland, E.W, A Cooled CCD Based Neutron lmaging Svçtern for Low Fluence Neutron Sources, E E E Transactions on Nuclear Science. Vol. 43, No. 3, June 1996.

1 5. Non-destructive Ins~ection Reuort Canadian Forces CF- 1 8A, United States Air Force Sacramento Air Logistics Cen ter, McClellan Aïr Force Base, California. 1995.

Page 149: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

16. DCIEM Report No 98-TM-44, Progress in the Non-destructive Evaiuation of CF-18 Comwsite Flight Controls for Water ïngress and Related Damage, DCIEM Air Vehicle Research Detachment, Ottawa, Ontario. June 1998.

17. Bickerton, M., Develo~ment of Improved Techniaues for the Neutron Radiomphv of CF188 Flinht Control Surfaces, Department of Chemistry and Chernical Engineering, Royal Military CoUege of Canada, Kingston Ontario, 1998.

18. Standard Practices for Thermal Radioerarihv of Materials, American Society for Testing and Materials, ASTM E748-95, pg. 2,1996.

19. Whittemore, W., Neutron Sources and Facilities - Present Status and Future, 6th World Conference on Neutron Radiogaphy, Japan, May 1999.

20. MacGiIlivary, G.M., Neutron Radiorzraphv Princi~les and Practices, Course notes fiom Neutron Radiography course given at M C 15-19 Feb, section 2.5, pg.11. 1999.

2 1. Standard Practices for Thermal Radionra~hv of Materials, Amencan Society for Testing and Materials, ASTM E748-95, pg. 8,1996.

22. Balasko, M., Svab, E., Neutron Radiographv in Reasearch and Develo~ment, Nukeonika 39, No. 1-3, pg. 7, 1994.

23. Standard Practices for Thermal Radiographv of Materials, American Society for Testing and Materials, ASTM E748-95, pg. 9,1996.

24. Standard Practices for Thermal Radio41.aohv of Materials, Amencan Society for Testing and Materials, ASTM E748-95, pg. 7,1996.

25. Standard Guide for Radiosco~y, American Society for Testing and Materials, ASTM E1000-92. pg. 396- 397,1996.

26. Brenizer, J.S., Berger, H., Stebbings, C.T., Gillies, G.T., Performance Characteristics of Scintillators for use in an Electronic Neutron ha-gin& Svstem for Neutron Radiomphy, American Institute of Physics, Rev. Science Instruments, Vol. 68, No. 9, September, pg. 3371-3379, 1997.

27. Berger, K, Polansky, D., Brenizer, J., Bueno, C., Electronic Neutron Imaninq, Industrial Quality Inc Final Report, Report Number N68936-934-0274, pg.4, January 1994.

28. Pleinert, H., Lehmann, E., Komer, S., Design of a New CCD- Camera Neutron Radiofsaphv Detector, Nuclear Instruments and Methods in Physics Research, Elsevier Science Publishers B.V. A 399. pg. 386, 1997.

29. Lanza, R, Shi, S., McFarland, E., A Cooled CCD Based Neutron Imaging System for Low Fluence Neutron Sources, IEEE Transactions on Nuclear Science, Vol. 43, No. 3, pg. 1347, June 1996.

30. Hi& Performance Digital haaine, Optikon Briefing Notes f?om Neutron Radiography Course given at RMC 1 5- 19 Feb 1999.

31. Healey, G., Kondepudy, R, Radiometric CCD Camera Calibration and Noise Estimation, IEEE Transactions on Pattern Analysis and Machine Intelligence, Vol. 16, No. 3, pg. 267-268, March 1994.

32. Buil, C., CCD Astronomv Construction and Use of an Astronomical CCD Camera, William Bell Inc, Virginia, pg.4-8, 199 1.

Page 150: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

33. Buil, C., CCD Astronomv Construction and Use of an Astronomicd CCD Camera, Wiliam Bell Inc, Virginia, pg.34, 199 1.

34. Buil, C., CCD Astronomv Construction and Use of an Astronomical CCD Camera, William Beil Inc, Vuginia, pg. 26, 199 1.

35. Buil, C., CCD Astronomv Construction and Use of an Astronomical CCD Camera, W i a m Bell h c , Virginia pg. 30, 199 1.

3 6. Jain, A, Fundamentals of Digital image Processing, Prentice Hall, Englewood C l m , NJ, pg. 5 1, 1989.

37. Buil, C., CCD Astronomv Construction and Use of an Astronomicd CCD Camera, Wfiam BeU Inc, Vxghia, pg. l69,;99 1.

3 8. Jahne, B., Digital Image Processinn. Conceuts, Alnorithms, and Scientifk Au~lications, Springer, New York, Pg. 21 1, 1997.

39. Spiegel, M., Probabilitv and Statistics, McGraw-Hill Book Company, pg. 78, 1975.

40. Avallone, A., Baumeister, T., Mark's Standard Handbook for Mechanical Engineers, McGrall-Hill. New York, pg. 17-21, 1978.

41. Buil, C., CCD Astronomy - Construction and Uses of an Astronomical CCD Camera, Willmann-BeU. Virginia, pg. 37,199 1.

42. MCD IHandbook, Andor Technology Training Notes, Pg. 53, Jan 98

43. Euil, C., CCD Astronomv - Construction and Uses of an Astronomical CCD Camera. Willmann-Bell. Virginia, pg. 43, 199 1

44. MCD Handbook, Andor Technology Training Notes, Pg. 82. January 1998.

45. Borne, J., Processine Digital Images, Department of Mathematics and Cornputer Science, Royal Military College of Canada, Kingston, Ontario, Canada, pg. 1 1, 1996.

46. Jain, A., Fundamentals of Digital Image Processinq, Prentice Hall, Englewood C l f i , NI, 1989.

47. Russ, J., The Image Processine Handbook, CRC Press, A m Arbor, pg. 34, 1992.

48. Jain, A., Fundamentals of Digital I m a ~ e Processinq, Prentice Hall, Englewood CliEs, NJ, pg. 51, 1989.

49. image-Pro Plus Ver 4 for Windows Reference Guide, Media Cybemetics, pg. 1-19, 1998.

50. Image-Pro Plus Ver 4 for Windows Reference Guide, Media Cybemetics, pg. 1-169, 1998.

51. Motomura, Y., Yoshii, K., Sonoda, IL, Application of Momholonicd Filters to Noise Suupression in Neutron Radiogra~hv Image, 5& World Conference on Neutron Radiography, Berlin, June 17-20, 1996.

52. Buil, C., CCD Astronomy, Wilùnann-BeU Inc, Richmond, Virginia pg. 253, 1991

53. Miday, D., Resnick, R, Physics, Parts 1 & 2, John Wiley & Sons Inc, Toronto, pg. 999, 1978.

54. Standard Guide for Comuuter T o m o m n h ~ (CT) Imagine;, American Society for Testing and Materials, ASTM E1441-95, pg. 712, 1997.

Page 151: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

55. Standard Tenninologv for Non-destructive Examination, American Society for Testing and Materiais, ASTM El3 1697b, pg. 636, 1997.

56. Panchal, C., Sinha, A., Bhawa, B., Shyan, A., S~ ivasan , M., Eduat ion of ksolvinn Power of Neutron Radioscouv System fiom edge Remonse Function Measurements, Insight, Vol. 39, No. 2, February, pg. 98, 1997.

57. Tobin, K., Brenizer, J., Mait, J., Modulation Transfer Function Technique for Real Time Radioscoriic Svstern Characterization. Ap~lied &tics, Vol. 28, No. 23, Decernber. 1989.

58. Gaskill, D., Linear Systems. Fourier Transforms and Outics, Wiley, New York. 1978.

59. Panchal, C., Sinha, A., Bhawa, B., Shyan, A., Srhivasan, M., Evaluation of Resolwlg Power of Neutron Radioscopv Svstem fiorn edne Remonse Function Measurements, Insight, Vol. 39. No. 2. February, pg. 99, 1997.

60. Russ, C., The Image Processinn Handbook, CRC Press, Ann Arbor, pg. 14, 1992.

61. Jahne, B., Digital Image Processinn, Conceuts, Alnorithms. and Scientific Applications, Springer, New York, Pg. 85, 1997.

62. Jain, A, Fundamentais of Digital Image Processinq, Prentice Hall, Englewood Clifi, NJ, pg. 59, 1989.

63. Jain, A., Fundamentals of Digital h a n e Processing, Prentice Hall, Englewood Ciiffs, NJ, pg. 51, 1989.

64. Standard Guide for Radiosco~y, ASTM E 1000-92 (Reproved 1 W6), pg. 405, 1996.

65. Bickerton, M., Develo~ment of Imuroved Techniques for the Neutron Radiograuhy of CF188 Flight Control Surfaces, Department of Chemistry and Chernical Engineering, Royal Military College of Canada, Kingston Ontario, pg.77, 1998.

66. Lewis, W.J., B e ~ e t t , L.G.I., Kirby, C.T., Enhancement to Neutron Radiolow Facilitv at the SLOWPOKE- 2 Facilitv at RMC, 5 World Conference on Neutron radiography, Berlin, pg. 425, 1996.

67. Brenizer, J.S., Hughes, D.E., Bryan, M.E., Gould, R., Flinchbaugh, T.L, Sears, C.F., Recent Improvements to the Pemsylvania State University's Neutron Radiom~hv Facility, 6th World Conference on Neutron Radiography, Japan, May 1999.

68. Lepine, B., McRae, K., Progress in the Non-Destmctive Evaluation of CF-18 Composite Flight Controls For Water ingress and Related Damage, DCIEM Report No. 9 8 -TM-44, pg. 3, 199 8

69. Halliday, D., Resick, R, Phvsics part 1 and 2, John Wiley and Son, Pg. 867, 1978.

70. Availone, A., Baumeister, T., Mark's Standard Handbook for Mechanical Ennineers, McGrall-Hill, NewYork. Pg. 17-2 1, 1978.

7 1. Standard Method for Determinin2 Image Oualitv in Thermal Neutron Radiorrra~hv Testing, ASTM E 545- 75, American Society for Testing and Matenals, PA, USA, pg. 600, 1975.

Page 152: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Avallone A, Baurneist er T, Mark's Standard Handbook for Mechanical Enaineers, McGraU-Hill, NewYork, 1978.

Balasko M, Svab E, Neutron Radioera~hy in Research and Develooment, Nukeonika 39, NO. 1-2, 1994.

Berger H, Polansky D, Brenizer J, Bueno C, Electronic Neutron Irnaeing, Industrial Quality Inc Final Report, Report Number N68936-93-C-0274, January, 1994.

Bickerton M, Development of Iqroved Techniques for the Neutron Radiogra* of CF 1 88 Flieht Control Surfaces, Deparûnent of Chemistry and Chernical Engineering, Royal Military College of Canada, Kingston Ontario, 1998.

B o u e J, Processine. Digital Images, Department of Mathematics and Cornputer Science, Royal Military College of Canada, b g s t o n , Ontario, Canada, 1996.

Brenizer J.S, Berger H, Gillies G.T, Mengers P, Stebbings C.T, Polansky D, Rogerson D.J, Develo~ment of a New Electronic Neutron Imag;in~ System, University of Virginia, Nuclear Instruments and Methods in Physics Research, March, 1999.

Brenizer J. S, Berger H, Stebbings C.T. Gillies G.T, Performance Characteristics of Scintillators for use in an Electronic Neutron Ima d e . SY stem for Neutron Radiographv, -

Arnerican Institute of Physics, Rev. Science Instruments, Vol. 68, No. 9, September, 1997.

Brenizer J.S, Hughes D.E, Bryan M.E, Gould R, Flinchbaugh T.L,Sears C.F, Recent Im~rovements to the Pennsvlvania State University's Neutron Radioma~hv Facilitv, 6th World Conference on Neutron Radiography, Japan, May 1999.

Buil C, CCD Astronomv - Construction and Uses of an Astronomical CCD Carnera, Wilhnn-Bell. Virginia, 199 1.

Casalî, P. Chiro, R Partemi, Neutron Imagine Usine. a Lo w-Con Cooled CCD Carnera, Nuclear Energy in Central Europe: Present and Perspectives, Portoroz, Slovenia, June 1993.

10. Czirr J.B, MacGillivray G.M, MacGillivray R.R Seddon P.J, Performance and Characteristics of a New Scintillator, Nuclear Instnunents and Methods in Physics Research, A 424, 1999

Page 153: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

1 1. DCIEM Report No 98-TM-44, Promess in the Non-destructive Evaluation of CF- 18 Cornosite Flight Controls for Water Ingress and Related Damage, DCIEM Air Vehicle Research Detachment, Ottawa, Ontario, June 1998.

12. Dupuis J.P, Huriet J.R, Industrial Iwection of Aeronautic Sarnples - First ster, with the DIANA, 5 World Confêrence on Neutron Radiography, June 1996.

13. Gaskill D, Linear Svsterns. Fourier Transforrns and O~tics, Wiley, New York, 1978.

14. Haiiday D, Resnick R, Phvsics. Parts 1 & 2, John Wiley & Sons Inc, Toronto, 1978.

15. Healey G, Kondepudy R, Radiometric CCD Carnera Calibration and Noise Estimation, IEEE Transactions on Pattern Analysis and Machine Intelligence, Vol. 16, No. 3, March 1994.

16. H i ~ h Perfomiance Digital Imaging, Optikon Briefing Notes from Neutron Radiography Course given at M C 15- 19 Feb 1 999.

17. Image-Pro Plus Ver 4 for Windows Reference Guide, Media Cybemetics, 1998.

18. Jahne B, Digital Image Processing, Concepts. Al~orithms. and Scientïfic A~~lications, Springer, New York, 1997.

19. Jain A, Fundamentals of Digital Image Processing, Prentice Hall Englewood Cliffs, NJ, 1989.

20. Kobayasi H, Neutron Radioaa~hv, 3, Kluwer, 1990.

2 1. Lange H.1, Leeflang H. P, Markgraf J.F, Neutron Radiosco~v using Electronic Cameras with Zoom Ca~ability and High Detectivity, 6th European Conference on Non Destructive Testing, 1994.

22. Lanza R, Shi S, McFarland E, A Cooled CCD Based Neutron Imaging System for Low Fluence Neutron Sources, IEEE Transactions on Nuclear Science, Vol. 43, No. 3, June 1996.

23. Leeflang H.P, Markgaf J.F. W, Detection of Corrosion on Aircrafl Component by Neutron Radioera~h~, Non-destructive Test ing Evaluation, Vol. 1 1, 1 994.

24. Lepine B, McRae K, Proaess in the Non-Destructive Evaluation of CF- 1 8 Com~osite Flight COntrols For Water ingress and Related Damaee, DCIEM Report No. 98 -TM-44, 1998.

25. Lewis W.J, Bennett L.G.1, Kirby C.T, Enhancement to Neutron Radiolow Facility at the SLO WPOKE-2 Facilitv at RMC, 5 World Conference on Neutron radiography , Berlin, 1996.

Page 154: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

26. MacGillivary G.M, Neutron Radioeraphv Princip les and Practices, Course notes kom Neutron Radiography course given at RMC 15- 19 Feb, section 2.5, 1 999.

27. MCD Handbook, Andor Technology Training Notes, January 1998.

28. Motomura Y, Yoshii K, Sonoda K, Application of Momholotrical Filters to Noise Sumression in Neutron Radiowaphv Image, 5th World Conference on Neutron Radiography, Berlin, June 1 7-20, 1 996.

29. Non-destructive inspection - Report Canadian Forces CF- 1 8A, United States Air Force Sacramento Air Logistics Center, McClellan Air Force Base, California, 1995.

30. Panchal C, Sinha A, Bhawa B, Shyan A, Srhivasan M. Evaluation of Resolvine Power of Neutron Radioscoov Svstem fkom E d ~ e Remonse Function Measurements, Insight, Vol. 39, No. 2, Febmary, 1997.

3 1. Pleinert H, Lehmann E, Korner S, Design of a New CCD- Camera Neutron Radioaa~hv Detector, Nuclear Instniments and Methods in Physics Research, Elsevier Science Publishers B.V, A 399, 1997.

32. Russ J, The Image Processine; Handbook, CRC Press, Ann Arbor, 1992.

33. Spiegel M, Probabilitv and Statistics, McGraw-Hill Book Company, 1975.

34. Standard Guide for Radioscooy, Amencan Society for Testing and Materials, ASTM E 1000-92, 1996.

35. Standard Guide for Comouted Tomowa~hv (CT) Imaeing, American Society for Testing and Materials, ASTM 1441-95, 1997.

36. Standard Method for Detennininrr Image Oualitv in Thermal Neutron Radiomaphy Testing, ASTM E 745-75, American Society for Testing and Materials, PA, USA, 1975.

3 7. Standard Terrninolom - - for Non-destructive Examinations, American Society for Test ing and Materials, ASTM E1316-97b, 1997.

38. Tobin K, Brenizer J, Mait J, ModuIation Transfer Function Techniaue for Real Time Radiosco~ic Svstem Characterization, Applied Optics, Vol. 28, No. 23, December, 1989.

39. Whittemore W, Neutron Sources and Facilities - Present Status and Future, 6th World Conference on Neutron Radiography, Japan, May 1999.

Page 155: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

APPENDICES

Page 156: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

APPENDIX A

CHARACTERISTICS OF THE SLOWPOKE AND BEAZEALE REACTOR

A. 1 SLOWPOKE-2 Facility

The SLOWPOKE-2 Facility (Safe Low Power c(K)ritical Experirnent) is a small

light - water, pool reactor designed by the Atomic Energy of Canada Limited (AECL) and

is capable of producing a maximum thermal power of 20 kW. Onginally it was designed

for Neutron Activation Analysis (NAA), but has been modified extensively to facilit ate

neutron radiography and neutron radioscopy. The major sections of the SLO WPOKE-2

Facility include the reactor and the Neutron Radiology System (NRS) (Figure A. 1). The

NRS coosists of a Neutron Bearn Tube (NBT) and a Neutron Carnera Assembly (NCA).

The terrn NRS will be used, but only in reference to the image plane (scintillation screen),

bellows, diffuser and carnera assembly.

The reactor core is located at the bottom of a seaied cylindrical aluminium

container suspended in a stainless steel-lined reactor pool that is 5.87 meters deep and

2.46 meters in diameter. The reactor core, located inside the critical assembly, is

manufactured fiom a zirconium alloy cage containing 198 fuel elements and is 4.4 m

below the surface of the water.' Both the pool and reactor container are filled with

' Bickerton, M., Develo~ment of lmproved Technimes for the Neutron Radioaa~hv of CF188

Flinht Control Surfaces, Deparunent of Chemistry and Chernical Engineering, Royal Military College of

Canada, Kingston Ontario, pg. 12, 1998.

Page 157: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

deionized light water that acts as a codant, radiation shield and, in the case of the critical

assernbly, as a moderator. Each element, formed &om Zircaloy-4 tubing, is packed with

pellets of sintered uranium dioxide enriched to 19.89% in Uranium 235. The core is

located within a beryllium annulus and berylliurn plate reflectors located above and below

the annulus. The berylliurn acts as a moderator and neutron reflector as well as the

housing of five inner radiation sites. A small thermal column of heavy water was installed

in place of an outer irradiation site, located between the beryIlium annulus and the reactor

container wall adjacent to the bottom end of the NBT (Figure A. 1 ). At half reactor power,

the flux measured at the location adjacent to the column of heavy water is 1.4 x 10" n/cm2

s vs 5.1 x 10'' n/cm2 s measured at the reactor container wall at a different radial location.'

Lewis, W.J., Andrews, W.S., Bennett, L.G.I., Beeley P.A., Measurernent in Sumort of a Neutron

Radioma~hv Facilitv for the SLOWPOKE- 2 at RMC, Nuclear lnstnunents and Methods in Physics

Research, Elsevier Science Publishers B.V. A299, pg. 430-433, 1990.

Page 158: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure A.1 - SLOWPOKE-2 Facilit. at RMC

Page 159: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

The NBT is neutrally buoyant in the reactor pool and has a divergent beam

with a measured L/D of approximately 100 starting fiom an aperture of 5 cm in diameter

and diverging to a beam of 60 cm in diameter at the lower beam stop. ' It is positioned

tangentially to the reactor container and, when the bottom end is swung against the reactor

container with a hand pumped hydraulic actuator, it is tilted 8.5' from the vertical. This

action places the iLluminator (a graphite block) adjacent to the centre line of the hiel cage

and almost in contact with the exterior of the reactor container. Neutrons passing through

the annula berylliurn refiector and the heavy water thermal co1um.n (which are both

located inside the reactor container) then pass into the illuminator which redirects the

thermal neutrons through the aperture and up the NBT. The action of swinging the NBT

away fiom the reactor container to the vertical position provides a shuttering fùnction.

The NCA consists of semi-conformal shielding, a positioning system, an image

plane, a belbws and three beam stops. The semi-conforma1 shielding is constnicted fiom

two layers of borated polyethylene and is attached to an actuated frame, which in tuni is

attached to the middle beam stop. The shielding is lowered and confoms to a part under

inspection even if it protrudes fiom the perimeter of the middle beam stop. The

positioning system is an aluminium structure Iocated between the bottom and middle

beam stop. A cornputer program is used to control an X-fiame and Y-fiame assembly, so

that the part being radiographed can be incrementally moved in either the x or y direction.

At the image plane, an aluminium fiame holds either a vacuum cassette with a gadolinium

3 Lewis, W.J., Bennett, L.G.I., Kirby, C.T., Enhancernent to Neutron Radiolo~v Facilitv at the

SLOWPOKE-2 Facility at RMC, 5 World Conference on Neutron radiography, Berlin, pg. 425, 1996.

Page 160: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

converter screen and film for neutron radiography, or a scintillation screen, for neutron

radioscopy. The image plane resides above the positionhg system and has the capability

of moving vertically (z-direction) to accommodate part movement and varying part

thickness. The neutron flux at the image plane has been measured using the thin gold foi1

method and was determined to be 2 to 3 x 104 dcm' S.^ The bellows is only installed for

neutron radioscopy and produces a light tight area between the scintillation screen and

middle beam stop. The lower beam stop is positioned above the NBT and has an angled

square hole 44 cm x 44 cm to match the beam output of the NBT. The middle beam stop

has a similar 44 cm x 44 cm square hole opposite to the lower beam stop that tapers to a

12.7 cm x 12.7 cm square hole at the top of the rniddle beam stop. The tapered hole in the

middle beam is a diffuser used to accommodate the CCD camera. The upper beam stop

fits on top of the middle beam stop, covering the small square hole and housing the CCD

A.2 Breazeale Reactor at Pennsylvania State University

The Breazeale Reactor at the Penn State University Radiation Science and

Engineering Centre (RESC) is a medium sized, light water, pool reactor and is capable of

Bickerton. M, Development of Improved Techniques for the Neutron Radionrauhv of CF188

Flierht Control Surfaces, Departmen t of C hemistry and Chernical Engineering, Royal Military CoUege of

Canada, Kingston Ontario, pg.77, 1998.

Bennett, L.G.L. Lewis, W.J., MacGiUivray, G.M., Francescone, O., Enhancement to Neutron

Radiolom Svstem on the SLOWPOKE-2 Facilitv at RMC, 6th World Conference on Neutron radiography,

Japan, 1999

Page 161: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

producing a maximum power of 1 MW. The major sections of the Penn State Breazeale

Nuclear Reactor (BNR) include the reactor, beam hibe and shielding as indicated in Figure

A.2.

Figure A2 - Top View of BNR

Uriginally, the reactor had a plate-type fuel core, but in the mid-sixties the reactor

core was converted to a TRIGA design. The reactor can produce a steady state operation

Page 162: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

fiom 100 W to 1000 kW, with a capability to pulse the power up to 2000 MW for

approximately 10 miUi~econds.~ The reactor core connects to a bridge that runs above the

reactor pool that allows the reactor core to move in lateral, longitudinal and rotational

directions. The three-axis movement permits the reactor core to be placed tangentially

against a heavy water tank that connects to the NBT, reducing the gamma content of the

neutron beam. The reactor fiel is composed of zirconium hydride mixed with uranium-

235. The heavy water tank is a right cylinder radially aligned with the #4 beam port and is

6 1 cm in diameter and 30.5 cm thick. The fiel rods are placed in a fiame assembly and

are in direct contact with the reactor pool water, which acts as a coolant, radiation shield

and moderator.

There are seven beam ports. but only #7 and #4 beam ports are in operation. The

#7 beam port is used for neutron gauging, while the #4 beam port has a beam tube

installed in it and is used for neutron radiography. The beam tube is a horizontal tapered

collimator that is 297 cm in length and has a diameter of 12.7 cm at the heavy water tank

and expands to 19.05 cm at exit of the beam tube. A gadolinium aperture and bismuth

filter is set at the entrance to the bearn tube adjacent to the heavy water tank. The beam

tube has an LID ratio of 34 and, at the end of the beam tube a flux of 3.0 x 106 dcm2 s,

was measured with the gold foi1 method. '

ïhe Penn State Breazcale Reactor: A facilitv of the Past. Present. and Future of the Pennsvivania

State University, Penn State Department of Publications, pg. 12.

7 Brenizer. J.S., Hughes, D.E., Bryan, M.E., Gould, R. Flinchbaugh, T.L., Sears, C.F., Recent

Improvements to the Pemsylvania State Universitv's Neutron Radiomhy Facility, 6th World Conference

on Neutron radiography, Japan, 1999

Page 163: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

The shielding is a fully enclosed room rneasuring 3.8 m by 2.4 m and constxucted

fiom concrete-filled steel boxes sumounding the exit of the beam tube. This back wall of

the structure acts a beam stop and houses the camera, i n t e n s w g screen, support

equipment and the article being radiographed. For this research, a light tight diffiser,

which was constructed so as to simulate dimensionally the middle beam stop fkom the

NRS at the SLOWPOKE-2 Facility, held a CCD carnera and scintillation screen at the

proper positions inside this shielded r w a When the diffuser was properly aligned, the

neutron beam illuminated a 35 cm diameter circle on the scintillation screen.

Page 164: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

APPENDIX B

IMAGE SPECIFICATIONS

B. 1 Installation and Optimisation of System at RMC Reduction of Light Leaks

-Hoc, 2 sets of 30 second to 4 minutes exposure, 5 1 1 x 5 1 1 resolution. Statistical Results

-30°C, 30 images at 4 minute exposure, 5 1 1 x 5 1 1 resolution.

B .2 Camera Paramet ers Dark Image - Temperature Change

RMC, -lS°C, 30 second to 4 minute exposures, 5 12 x 5 12 resolution. RMC, -30°C, 30 second to 10 minute exposures, 5 12 x 5 12 resolution.

Dark Image - Diffuser Changes No Diffuser, -30°C, 30 second to 10 minute exposures, 512 x 5 12

resolution. D f i s e r at RMC, -30°C, 30 second to 10 minute exposures, 51 1 x 51 1

resolution. Diffuser used at Penn State -30°C, 30 second to 10 minute exposures, 5 12 x

5 12 resolution.

Unified Images - Neutron Flux Effects RMC (10 kW), -30°C, 30 second to 10 minute exposures, 5 12 x 512

resolution. Penn State (1 kW) -30°C, 30 second to 5 minute exposures, 5 12 x 5 12

resolution. Penn State (10 kW) -30°C, 30 second to 10 minute exposures, 512 x 512

resolution. Penn State (100 kW), -30°C, 30 second to 5 minute exposures, 512 x 512

resolut ion.

Dynamic Range Combination of data fiom dark images and unified images taken at RMC

(10 kW), Penn State (1 kW, 10 kW, 100 kW).

B.3 Resolut ion System Resolution Calculations

RMC (10 kW), -30°C, 30 second to 10 minute exposures, 5 12 x 5 12 resolut ion.

Penn State (10 kW),-30°C, 30 second to 10 minute exposures, 5 12 x 5 12 resolut ion.

Page 165: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Penn State (100 kW), -30°C, 30 second to 5 minute exposures, 512 x 512 resolution.

S ystem sensitivity RMC (10 kW), -30°C, 30 second to 10 minute exposures, 512 x 512

resolution. Pem State (10 kW),-30°C, 30 second to 10 minute exposures, 512 x 512

resolut ion. Penn State (100 kW), -30°C, 30 second to 5 minute exposures, 512 x 512

resolut ion.

B.4 Water Ingress Simulation

Wat er Ingress Water Ingress - Single Cells

RMC (10 kW), -30°C, 30 second to 10 minute exposures, 512 x 512 resolut ion.

Water Ingress - Three CeUs RMC (1 0 kW), -30°C, 30 second to 10 minute exposures, 5 12 x 5 12

resolution.

Water Ingress - Camera Gain Camera Gain - Single Cells

RMC (10 kW), -30°C, 3 minute exposures, 5 12 x 5 12 resolution, at gain settings 0,4, 8, 12, 16.

Camera Gain - Three Cells RMC (10 kW), -30°C, 3 minute exposures, 512 x 512 resolution, at

gain settings 0,4, 8, 12, 16.

Water Ingress - Digital Enhancement RMC (1 0 kW), -30°C, 4 rninute exposures, 5 12 x 5 12 resolution

with digital enhancements applied.

S imulated Water Simulated Water - Single Ce11

RMC (10 kW), -30°C, 30 second to 10 minute exposures, 5 12 x 5 12 resolution.

Penn State (10 kW), -30°C, 30 second to 10 minute exposures, 5 12 x 5 12 resolution.

Penn State (100 kW), -30°C, 30 second to 5 minute exposures, 5 12 x 5 12 resolution.

Page 166: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Simulated Water - Thfee Cells RMC (10 kW), -30°C, 30 second to 10 minute exposures, 5 12 x 5 12

resolution. Penn State (10 kW), -3OoC, 30 second to 10 minute exposures, 512

x 5 12 resolution. Penn State (100 kW), -30°C, 30 second to 5 minute exposures, 512

x 5 12 resolution.

B.5 CF188 Rudder RMC rudder exposure number 1 to 7 less S), -30°C, 2,4,6 minute exposures, 512

x 5 12 resolution, rudder exposure number 5, -30°C, 30 second to 10 minute exposures, 5 12 x 5 12 resolution.

Penn State (10 kW for rudder exposure number 2 to 6, less 5), -30°C, 2, 4, 6 minute exposures, 512 x 512 resolution, expoçure number 5, -30°C, 30 second to 1 0 minute exposures, 5 1 2 x 5 12 resolution.

Penn State (100 kW for rudder exposure number 2 to 6), -30% 1, 2, 3 miaute exposures, 5 1 2 x 5 12 resolution

B .6 Comarison of Radiosco~y to Radio~aphy Resolution

RMC (10 kW) Two film shots (CX at 18 minutes and SR at 2 hours and 15 minutes)

Page 167: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

APPENDE C

SYSTEM OPTIMISATION AND CAMERA PARAMETERS

C. 1 Installation and Optimisation of Systern

Figure C. 1 is a histogram before the light leaks were reduced at RMC while Figure C.2 is a histogram after the light leaks were reduced.

Figure C.1- Before Light Leaks Reduced

O 256 512 ïôû lû24 1280 1536 1792 aY8 PO, 258) aBi6 3072 3328 35ô4 3840 4096

hdsrulty Value

Figure C.2 - After Light Leaks Reduced

Page 168: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

C.2 Camera Parameters

C.2.1 Dark Image - Temperature Change

The following graphs are of dark images at two different camera temperatures (-15°C -30°C). Resolution, intensity interval and diffher type were held constant, while the exposure time was varied.

Figure C 3 - Dark Image at -15OC

- - - -

Figure C.4 - Dark Images at -30C0

Page 169: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

C.2.2 Dark Image - Diffuser Changes

The following are graphs of dark images with camera temperature, resolution, and intensity interval held constant, while varying the exposure tirne and the diffuser type.

Figure CS - Dark Images taken with No Dinuser

l I '.

-10 min I

Figure C.6 - Dark Images taken with diffuser used at Penn State

Page 170: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure C.7 - Dark Images taken with Diffuser used at RMC

C.2.3 Unifies Image - Neutron Flux Effects

Several images were taken at four different power settings; at RMC, the images were taken at hdf power (10 kW) and at Penn State the images were taken at 1, !O and 100 kW.

- - - - --

Figure C.8 - Unified Images taken at RMC at 10 k W

Page 171: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure C.9 - Unified Images taken nt Penn State at 1 kW

-2 min

-3 mfn

- a O

0 loooo! - -8 min

Figure C.10 -Unified Images taken at Penn State at 10 kW

Page 172: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

- -

Figure C.11 - Unified Images Taken at Penn State at 100 k W

C.2.4 Dynamic Range

The following graphs show the dynamic envelope of the carnera system for RMC (10kW) and Penn State (1 kW, 10 kW and 100 kW) by plotting the FWHM value for the dark and unified images for the indicated exposure tirnes. The dynamic range in the text was calculated using peak centroid values, but the following graphs indicate dynarnic range using FWHM values. The dynamic range is similar in magnitude for both methods.

Figure C.12 - Dynamic Range of System from Images taken at RMC

Page 173: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure C.13 - Dynamic Range of System from Images taken nt Penn State (1kW)

Figure C.14 - Dynamic Range of System from Images taken at Penn State (10 kW)

Page 174: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure C.15 - Dynamic Range of System from Images taken at Penn State (100kW)

Page 175: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

APPENDIX D

ERROR CALCULATION OF LNTENSITY DIFFERENCE

D. 1 Error Theorv

Standard error calculàt ions were used throughout the report. The basic de finitions of the two types of error, absolute and relative uncertainty will be given. Then the basic method for adding and multiplying mors will be shown and an error calculation will be developed for intensity difference formula.

The absolute uncertainty of the measurement is an estimate of the range in which the measurement could fa11 due to the limited accuracy of the measurement technique, e.g., 0.52 + 0.02. Where 0.52 is the best estimate and 0.02 is the absolute uncertainty. The relative uncertainty is a measure of the quality of the measurement and is fiequently stated as a percentage uncertaint y emor. Relative uncertaint y or ftact ional uncertaint y is represented as M x or in the exarnple above relative error = 0.02/0.52 x 100 = 3.8% where âx = 0.02 and x = 0.52.

The uncertainty in a sum or difference is the sum of the absolute uncertainties, i.e., if C = A + B then AC = A A + AB. The percentage relative uncertainty in a quotient or a product is the sum of the relative uncertainties, Le., Z = AB then AZ/Z = M A + D I B .

The formula for intensity difference is as follows:

ID (%) =(C-B)/B x 100 Where:

ID = Intensity Difference C = Cell Intensity Value B = Background Intensity Value

D.2 Error Calculation

The following is the formula for the error calculation for intensity difference.

The uncertainty error for Z.

Where Z = (C-B) is

The relative error for ID.

Where ID = (C-B)/B = 233 is

Page 176: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

AZ is substituted to obtain the relative error for intensity difference.

D.3 Samde Calculation of Intensitv Difference

A sample calculation of the intensity difference for a single ce11 of water, with a volume of 0.04 mi taken at a m u t e exposure will be carried out. The data for this calculation is presented in Table D. 1.

sure

The absolute value (AC) for the ce11 can be calculated by the following:

MaXLmum value (3 129) - mean value (2336.2) = positive error value = 792.8 Minimum value (22 14) - mean value (23 3 6.2) = negative error value = - 1 22.2

A large intensity value (relative to the background intensity value) is associated with the noise of the system; hence, the reason for the large value in the maximum colurnn in Table D.1. Using the maximum intensity value in an error calculation causes a very large calculated error for the system. One solution to make the error more realistic is not to use the maximum value but to use the minimum error for both the negative and positive value (i.e., 2336 * 122.2).

Another method for calculating the AC value is to use the standard deviation value. Using the standard deviation produces an error value of:

2 x Standard deviation = 2 x 215.6 = AC = 431.2

Using the standard deviation produces a value that is less consewative then using the minimum value, but more conservative than using the maximum value and therefore will be used for calculating the AC value in this Appendix.

The minimum intensity value for the background will be used for the absolute error calculation, as the threshold technique was not used on the background and therefore the maximum value tends to be very large. The calculation for hB is as follows:

Page 177: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Minimum value (2363) - mean value (2880.5 1) = negative error value = -5 17.5 1

Therefore: AE3 = * 517.51

The calculated the relative error for a 5-minute exposure of a 0.04 mL ceii is as follows:

D.4 Error Calculation for Intensity Difference of a Single Cell

The relative error was calculated for a single volume of water (0.14 mL) for exposure times f?om 0.5 to 10 niinutes as described in the above paragraphs. The relative errors for the various exposure times are shown in Table D.2.

Table ater

Averaging the relative errors produced an average relative error of 29%, which was used in Figure 4.1 1.

Page 178: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

APPENDIX E

ENHANCED MAGES OF TEST PECE

2 Minute Exposure 2 Minute Exposure wlcontrast Changes

4 Minute Exposure 4 Minute Exposure wlcontrast Changes

Page 179: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Equalization of H istogram

Erode Filter

Equalization of Histogram w l Contrast Changes

Erode Filter wfcontrast

Page 180: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

J Flat Filter J Flat Filter wlcontrast

M edian Filter M edian Filter wlcontrast

Page 181: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

M ultiplying Image M ultiplying image

wIContrast Changes

Ratio Flat Ratio Flat

wlcontrast Changes

Page 182: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Ratio Optical Ratio O ptical wlcontrast Changes

Page 183: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

APPENDIX F

NEUTRON RADIOSCOPY IMAGES OF CF 188 RUDDER

The following images were taken at RMC (10 kW) and Penn State (10, 100 kW) of a CF188 Rudder for an exposure time that is indicated in figure caption for the image.

Figure F.1- Exposure 2,3-minute exposure time taken at Penn State at 100 k W

Page 184: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure F.2 - Exposure 5, Cminute exposure time taken at RMC

Figure F.3 - Exposure 5,4-minute exposure time taken at RMC with contrast

Page 185: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure F.4 - Exposure 5,4-minute exposure time taken at RMC w/ Erode & contrast

Figure F.5 - Exposure 5, Zminute exposure time taken at Penn State, 10 kW

Page 186: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure F.6 - Exposure 5, Zminute exposure time taken at Penn State, 10 kW with contrast

Figure F.7 - Exposure 5,l-minute exposure time taken at Penn State, 100 kW

Page 187: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure F.8 - Exposure 5,l-minute exposure time taken at Penn State, 100 kW with contrast

Page 188: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

APPENDIX G

COMPARISON O F NORMALISED DATA

Figure G 1 - Exposure 2 of Rudder, taken at Penn State (100 kW), 4-minute Exposure

Nomlised Data of Exp 2 of Rudder O -3OC. 512 x 512 m8drilm-1. W m ~ ~ l y rilarol-13

Normilized htansïty Value

- -

Figure G2 - Normalised Histogram of Exposure 2 at Two Different Power Settings

Page 189: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure G3 - Exposure 3 of Rudder, taken at Penn State (100 kW), 4-minute Exposure

- N o m l i D m ot Exp 3 of Rudder

O Ja C. 512 x 512 mscirrrm. rirms~ty nlwiai -16

-- --

Figure 6 4 - Normaliseci Histogram of Exposure 3 at Two Different Power Settings

Page 190: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure G5 - Exposure 4 of Rudder, taken at Penn State (100 kW), Cminute Exposure

--

N o m i a l i d Daia of Exp 4 of Rudder O -J) C. 512 x 512 mcUm. Wenscy 1ds91-16

-- -WC EV a n - Pem Çcae 100 kW Exp clmn,

\

O O 1 0 2 O 3 O 4 O 5 O 6 O 7 O a 0 9 1

Nomalüed htensity Value

-- .- ---

Figure 6 6 - Normalised Histogram of Exposure 4 at Two Different Power Settings

Page 191: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure G7 - Exposure 5 of Rudder, taken at Penn State (100 kW), 4-minute Exposure

-- Nonnalkgd Data of Ex(, 5 of Rudder

O -30 C. 512 I 5.12 resdim. n m a y inaisi -16

Normal id htensity Value

Figure GS - Norrnalised Histogram of Exposure 5 at Two Different Power Settings

Page 192: FORcollectionscanada.gc.ca/obj/s4/f2/dsk1/tape3/PQDD... · ACKNO WLEDGMENTS 1 wisb to express my sincere gratitude and appreciation to Dr. L.G.1 Bennett and Major W.J. Lewis, for

Figure G9 - Exposure 6 of Rudder, taken at Penn State (100 kW), Cminute Exposure

Figure GlO - Normalised Histogram of Exposure 6 at Two Different Power Settings