Accutron MRInstructions for use
All brands and trademarks referred to in this manual are the property of the respective owners.
Hauptstraße 255D-66128 Saarbrücken
Telephone +49 (0)681/970 17-0Telefax +49 (0)681/970 17-20
http://www.medtron.com
© 2008 All rights reserved
0535These instructions for use, Accutron MR, version 2 GB,
apply to the high-pressure injector Accutron MR.
05/2008
Contents1 IntroductionRegarding these instructions for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6Manufacturer's plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8General conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2 Safety instructionsProtection of people and property . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
3 Product descriptionGeneral features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13Items supplied and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14Consumables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Connections on the control panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16Controls mounted on the control panel . . . . . . . . . . . . . . . . . . . . . . . . . . .17The touch screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18The functions of the touch screen buttons . . . . . . . . . . . . . . . . . . . . . . . .20On/Off switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22Connection for the charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22Castors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22Remote control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22Handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
4 Commissioning the injectorCharging the battery cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23Switching ON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24Stand-by function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25Display state of charge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Setting the touch screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26Setting the touch screen brighter or darker . . . . . . . . . . . . . . . . . . . . . . . .26Selecting a language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27Calibrating touch screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27Setting stand-by time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28Setting idling speed of the pistons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28Setting filling speed of the pistons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Preparing injector head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30Moving the pistons back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Inserting the syringes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32Connecting the tube system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5 Operating the injectorFilling injector with contrast medium and NaCl . . . . . . . . . . . . . . . . . . . . . .35
Move both pistons forward . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36Venting the tube system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Changing the maximum operating pressure . . . . . . . . . . . . . . . . . . . . . . . .43Selecting an injection profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Changing an injection profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Storing changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Overwriting an existing profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47Entering a profile name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48Saving changes under a new profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49Creating a new profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Adding a new phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52Deleting a phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Performing an injection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54Maintaining vein access with the KVO function . . . . . . . . . . . . . . . . . . . . .55Starting the injection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56Interrupting the injection on the remote control . . . . . . . . . . . . . . . . . . . . .57Interrupting the injection on the injector . . . . . . . . . . . . . . . . . . . . . . . . . . 57Injection end . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Replenishing injector with contrast medium/NaCl . . . . . . . . . . . . . . . . . . . . 59Enter filling volume and fill . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60Venting after replenishing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
Post-examination work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64
6 AnnexWaste disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65Safety inspections and maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65Cleaning and storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65System messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
1 IntroductionThis chapter contains some preliminary remarks on the use of the injector Acc-utron MR with Easy Loading Syringe (ELS) as well as explanations on the structure of these instructions for use and on the use of symbols and text markings.
Regarding these instructions for useThese instructions for use explain how to correctly commission and operate the Accutron MR for your application.
MEDTRON wants you to be able to use your Accutron MR safely, correctly and efficiently. To this end, it is essential that you read these instructions for use thoroughly before you use the injector. They contain important instructions designed to assist you in avoiding risks and, at the same time, increasing both the reliability and service life of the injector and its accessories.
For your own safety you should read the section Safety instructions. Comply exactly with all the instructions to ensure that you do not put yourself, other employees or patients at risk and prevent damage to the injector.
If you have any questions concerning the handling of the Accutron MR not answered by these operating instructions, please contact:
Hauptstraße 255 D-66128 Saarbrücken
Telephone +49 (0)6 81/970 17-0 Telefax +49 (0)6 81/970 17-20
Accutron MR 05/2008 5
6
Regarding these instructions for use 1 Introduction
Intended useThe injector Accutron MR is intended exclusively for the injection of contrast media and physiological saline solution into patients for examinations involv-ing magnetic resonance imaging (MRI). Any use above and beyond this scope is deemed as inappropriate. Never use the Accutron MR for other purposes!
Manufacturer's plateThe manufacturer's plate of the Accutron MR bearing the model name and serial number is located on the rear of the operating unit. You will require this information if you contact the MEDTRON Customer Service, e.g. if you want to order accessories or spare parts.
Enter the model name and serial number here so that you have them at hand when required:
Model name: ____________________________________________ Serial number: ___________________________________________
2006
05/2008 Accutron MR
1 Introduction Regarding these instructions for use
ContentsThis manual describes the commissioning and operation of the high-pressure injector Accutron MR. It consists of the following chapters:– The chapter Safety instructions contains important safety instructions
which are essential to observe for the safety of people and the system.– The chapter Product description introduces you to the injector with its
functions and controls.– The chapter Commissioning the injector describes how to set the touch
screen to suit your needs, how to prepare the injector heads and connect the tube system.
– The chapter Operating the injector explains how to fill the contrast medium and saline solution, how to deal with the injection profiles and how to perform the injection.
– In the Annex you will find notes on inspection and maintenance work, cleaning and storage, a table with system messages and errors and their remedy as well as the technical specifications of the injector and the declaration of conformity.
– The Index helps you to quickly find the desired text passages you want.
Accutron MR 05/2008 7
8
Regarding these instructions for use 1 Introduction
Symbols usedYou will find the following two symbols identifying important safety instructions at several points in this manual:
The following symbol indicates tips and particularities which facilitate the oper-ation of the device.
General conventions
Bold type print
Designations and lettering of menus, menu items and buttons are printed in bold type.
Italic type print
Cross references to other chapters or text passages are printed in italics.
Inverted commas
Device functions are placed in "inverted commas".
Control steps
Control steps are identified as follows:• Make sure that the piston is moved back.
• Push the syringe into the recipient until it engages audibly.
• Remove the protective cap from the syringe.
Listings
Listings are marked as follows:– Variant A– Variant B– Variant C
CAUTION! This symbol indicates risks which could involve fatal or serious personal injuries or severe damage to property.
ATTENTION! This symbol highlights risks which could involve material damage.
Note! If the unit has been out of service for a prolonged period, charge the battery cells before starting it up again.
05/2008 Accutron MR
2 Safety instructions
– Do not connect the patient to the system until you have vented the syringes and the tube system.
– Vent the syringe and the tube system every time after filling the injector with contrast medium or saline solution.
– Make sure before every injection that there are no air inclusions in the syringes and the connected tubes.
– Only start an injection when the syringes are in the lowered position (injection position).
– Make sure before every injection that the injection parameters shown on the display do not endanger the patient.
– Maintenance and repair work may only be performed by the MEDTRON Customer Service or persons authorised by MEDTRON. Authorised persons will receive the documents necessary for the maintenance and safety inspections from MEDTRON.
– Make sure that the injector is no longer operated when the display of the injector has only failed slightly or fails completely.
Read the following safety instructions thoroughly and adhere to them closely. They are intended to protect your own safety and the safety of colleagues and patients as well as to avoid damage to the Accutron MR and its accessories.
Protection of people and propertyCAUTION! Risk of fatal or serious injuries due to air embolisms!
CAUTION! Risk of uncontrolled supply of contrast medium due to hydrostatic pressure compensation! Make sure that the patient and the syringes are at the same level.
Accutron MR 05/2008 9
10
Protection of people and property 2 Safety instructions
– Replace all disposable articles with new ones after they have been used once. Otherwise, you jeopardise your patients' health.
– Do not remove the protective caps of the syringes and the tube ends until you have made the connections.
– Only use consumables and accessories which MEDTRON recommends.
– Make sure that the packings of the disposable articles do not show any signs of damage.
– Make sure that the 'use-by' date of the disposable articles has not been passed.
CAUTION! Risk of infection!
CAUTION! Danger from electric current!
– Make sure that no liquid can enter the injector and the peripheral equipment. Should this happen, switch off the device immediately, and secure it from being switched on again.
– Make sure that the electric connection cables are intact and cannot be bent or squeezed. If you discover any damage, pull out the mains plug and secure the remote control from being switched on again and do not continue to use the charger.
ATTENTION! Risk of malfunctions of the Accutron MR! Only use syringes and tubes recommended by MEDTRON.
05/2008 Accutron MR
2 Safety instructions Electromagnetic compatibility
Electromagnetic compatibilityThe injector is subject to special safety precautions as regards EMC. Installation and commissioning are performed by trained staff authorised by MEDTRON.
The injector satisfies the requirements of the standard DIN EN 60601-1-2 for medical equipment in terms of the emitted interference and interfer-ence immunity.
You can request the document Tables on EMC Rev00 from MEDTRON.
ATTENTION! Portable and mobile HF communications equipment may affect the injector. Make sure that the injector is not positioned near such equipment.
ATTENTION! The injector may only be used with the accessories described in the instructions for use. Operation of the injector with other accessories or equipment may result in increased emitted interference or a reduction in the interference immunity.
Accutron MR 05/2008 11
12
Electromagnetic compatibility 2 Safety instructions
05/2008 Accutron MR
3 Product description
General featuresThe Accutron MR is designed for exactly dosed injections of contrast medium (CM) and physiological saline solution (NaCl). You can store the injection parameters you want– Delay time– Volume,– Concentration– Flow rate and– Injection time
in up to 80 profiles and retrieve them at any time. The reproducibility of the injection parameters is therefore reliable and easy. A profile can consist of a maximum of 6 individually programmable injection phases which are per-formed automatically after the program has been started. With the parameter "Concentration", you can, if required, inject contrast medium and NaCl at the same time.
The Accutron MR has two injection units which can be controlled independ-ently of each other and so the contrast medium can be positioned variably and accurately in the examination area with NaCl, depending on the application. This prevents tailings of the contrast medium which previously made delimita-tion of the examination area more difficult. The function "Keep-Vein-Open" (KVO) injects 0.5 ml of NaCl every two min-utes and therefore ensures that access to the vein is retained.
Thanks to the use of microprocessor-controlled electric motors, you can also control the piston speeds during idling and when filling the syringe manually via the keyboard.
The injector runs independently of the mains network as it is wireless and is operated with battery cells. By means of the remote control you can control the high-pressure injector from the control room.
Accutron MR 05/2008 13
14
Items supplied and accessories 3 Product description
Items supplied and accessoriesThe system of the Accutron MR comprises the following elements:
Equipment– Mobile and wireless injector with swivelling injections unit and recipients for
holding syringes, control panel with touch screen, battery cells in the unit base and separate charger, instructions for use
– Touch screen remote control with power supply unit and transmitter, data transmission via radio, instructions for use
Consumables– Disposable syringe 200 ml– MR tube system– Patient tube
Assemblies
1 Injector base with battery cells and connector for charger
2 Receiving aerial for remote control
3 Column stand
4 Bottle holders for contrast medium and NaCl bottles
5 Control panel with touch screen
6 Swivelling injection unit for contrast medium
7 Recipient for holding the syringe for contrast medium
8 Recipient for holding the syringe for NaCl
9 Swivelling injection unit for NaCl
10 On/Off switch
11 Four smooth-running castors, the front two with locking brakes
05/2008 Accutron MR
3 Product description Assemblies
Accutron MR 05/2008 15
16
Controls 3 Product description
ControlsThe following contains a description of the controls of the Accutron MR.
Connections on the control panelThere are three connections on the rear of the control panel which are irrele-vant for the operation of the injector.– Two connections for the LWL cables to the limit switches of the pistons.
The cables always remain connected.– A service connector for the MEDTRON Customer Service
05/2008 Accutron MR
3 Product description Controls
Controls mounted on the control panelThe control panel of the Accutron MR consists of buttons on a keypad and a central touch screen, a combination of display and entry field.
No. Symbol Designation Function1 Left-hand button group for manual piston movements for CM
Piston forward While pushing the button the piston moves forward
Increase piston speed
Increase piston speed on each push of the button
Piston back While pushing the button the piston moves back
2 Touch screen To display and enter injection parameters
3 Right-hand button group for manual piston movements for NaCl
4 Interrupt injection To interrupt the injection
5
Pilot light
– lights up green when the injector is switched on – flashes green during the injection – lights up red in the event of a fault
6 Start injection To start the injection
7 Wake-up button To activate the display
Accutron MR 05/2008 17
18
Controls 3 Product description
The touch screenThe touch screen displays messages and graphics like a normal monitor. In addition, you can make entries and call up functions by touching the buttons displayed. If a button cannot be used at the moment, it is greyed.
Touch the symbol you want lightly with your finger tip. Depending on the but-ton, you can mark or select a value or call up a function and a new window will open.
In addition to entering commands, the main window of the touch screen also serves to display the process information currently set. The profile name is located in the text field in the top left and the parameters of the injection profile currently selected are in the table underneath.
ATTENTION! Damage to the touch screen possible! Do not strike or hit the touch screen. Do not use any items such as pencils or tools. Only touch the buttons lightly with your fingers.
Prof. Nr. 5
Phase Delays
Volumeml
Conc.%
Flowml/s
Inj. times
1
2
3
20
20
20
10
10
10 100
65
100 6.0
6.0
6.0 0
0
0
Not airfree
180 178
05/2008 Accutron MR
3 Product description Controls
The column headers for the injection parameters at the top of the display describe the following parameters for each phase:
During the injection the touch screen also shows status information on the cur-rent injection in two windows. With the Info button you can switch between the windows.
Phase Numbering of the individual phases
Delay (s) Breaks between the injection phases in seconds
Volume (ml) The amount of fluid, in millilitres, to be injected
Concentration (%)
Concentration of the injection media:– 100% means only contrast medium– 40% means 40% contrast medium, 60% NaCl– 0% means only NaCl
Flow (ml/s) Injection speed in millilitres per second
Inj. time (s) Injection time in seconds
The remaining contrast medium in millilitres is shown below the green syringe symbolThe remaining NaCl in millilitres is shown below the blue syringe symbol
The battery symbol shows the current state of charge of the battery cells.
Injected volume CM Contrast medium injected so far
Injected volume NaCl NaCl injected so far
Phase time (s) Duration of the current injection phase
Delay (s) Breaks between the injection phases
Scan delay (s) Delay between injection start and scanner start (can only be set on the remote control)
Injection time Counts the time from pressing the start button
KVO Indicates whether the KVO function (Keep Vein Open) is activated
Accutron MR 05/2008 19
20
Controls 3 Product description
Example of dialogue messages:
You answer dialogue messages by touching the relevant button. In the example shown above you can answer the question "CM-syringe air-free?" either using the button with the green symbol for Confirm or the button with the red cross for No/Reject.
The functions of the touch screen buttons
Symbol Designation Function
Call up profile To call up an existing profile
Save profile To save a profile
Confirm To confirm inquiries and changes
Reject To reject unstored entries
Back To scroll backwards within a menu
Forward To scroll forwards within a menu
Keep Vein Open To activate and deactivate the KVO function in order to maintain access to the vein
Call up filling menu To call up the filling menu
Fill syringes To enter the filling volume in the filling menu
Piston forward To move the piston forward in the filling menuTo scroll upwards within a list
CM-syringe airfree?
Venting
05/2008 Accutron MR
3 Product description Controls
Piston back To move the piston back in the filling menuTo scroll downwards within a list
Add phase To add a new phase to an injection profile
Delete phase To delete a phase
Interrupt injection To temporarily stop the injection
Set system To open the System settings menu
Calibrate touch screen To calibrate the touch screen
Save system settings To save changed system settings
Info To change between the two windows with the injection data
Activate To prepare the injector for the injection
Continue To continue the injection after an interruption with the HOLD button
NaCl volume To select preset filling volumes of NaCl: 50, 100, 200 ml
CM volume To select preset filing volumes of contrast medium: 20, 65, 100 ml
Leave menu To leave the menu back to the main window
Airfree To confirm the previously performed venting operation
Symbol Designation Function
Accutron MR 05/2008 21
22
Controls 3 Product description
On/Off switchThe Accutron MR has an On/Off switch on the injector base. When you have switched on the injector with this button, the integrated lamp of the switch comes on and the pilot light on the control panel lights up green. By pressing the button again, you switch the injector off again and both lights go off.
Connection for the chargerA socket is located on the side of the injector base. Here you can connect the connecting cable to the charger when you want to charge the battery cells.
CastorsYou can conveniently move the injector using the smooth-running castors to the location you want. When it is in the correct position, secure the injector using the two locking castors. To this end press the locking lever downwards with your foot. When you want to release the castors, raise the locking lever slightly with your foot.
Remote controlThe injector can be remotely controlled using the remote control. The data are transferred to the injector by radio. You will find more details in the instructions for use of the touch screen remote control.
HandleThere is a connecting bar between the two swivelling injection units. With this bar you lower the injector heads backwards into the injection position after establishing all the injection parameters. To prepare a new injection, pull the upper injector section on the connecting bar forwards into the vertical position.
Note! Secure the injector position using its locking castors before you swivel the injector head forwards or backwards.
Note! Secure the injector position using its locking castors before you swivel the injector heads forwards or backwards.
05/2008 Accutron MR
4 Commissioning the injectorIn this chapter you learn how to charge the battery cells, set the touch screen, insert the syringes into the injector head and connect the tube system.
Charging the battery cellsPrior to initial commissioning, you must charge the battery cells accommo-dated in the injector base.
• Make sure that the injector is switched off.
• Insert the connector for the charger into the socket on the injector base and connect the charger to the mains network.
The charging operation starts when the red LED on the charger comes on. After some time the yellow LED comes on. The light comes on green when the battery cells are fully charged.
• As soon as the battery cells are fully charged, disconnect the charger from the mains network and pull the connector of the charger out of the socket on the injector base.
The injector is now ready to be switched on, see next section.
Note! If the unit has been out of service for a prolonged period, charge the battery cells before starting it up again.
Note! The battery cells are maintenance-free. They are replaced every two years or when necessary exclusively by the MEDTRON Customer Service.
ATTENTION! The injector must be switched off before the battery cells are charged.
CAUTION! Do not charge the unit in the MR room. The injector must be at least 2.5 m away from the patients when the battery cells are being charged.
Note! The charging time depends on the state of charge of the battery cells. If they are completely discharged, charging takes about 8 hours.
Accutron MR 05/2008 23
24
Switching ON 4 Commissioning the injector
Switching ONBefore being switched on, the injector must be charged and disconnected from the charger.• Press the On/Off switch on the injector base to switch the injector on.
The integrated lamp of the switch comes on and the pilot light on the control panel lights up green.
Self-testAfter being switched on, the unit automatically performs a self-test. As soon as the self-test is completed, you briefly see the display with the MEDTRON logo:
Then the touch screen immediately shows the main window from where you can call up the submenus.
Accutron MR
Prof. Nr. 5
Phase Delays
Volumeml
Conc.%
Flowml/s
Inj. times
1
2
3
20
20
20
10
10
10 100
65
100 6.0
6.0
6.0 0
0
0
Not airfree
180 178
05/2008 Accutron MR
4 Commissioning the injector Switching ON
Stand-by functionIf you have not pressed a button for a preset period, the display switches off automatically. This saves energy and avoids any disruptive electromagnetic radiation during the scan operation. You can reactivate the deactivated display at any time by pressing the Stop button or the Wake-up button.
Display state of chargeIn addition to the battery symbol in the main window, the injector features also a more exact display of the charge state. When the injector is switched on, you can read off the charge of the battery cells in the System settings menu.
The System settings 1/2 window displays the state of charge in percent.
If the state of charge is less than 10 %, you must charge the battery cells. In this case charge the battery cells fully using the charger, see Charging battery cells on page 23.
If you have not yet set the touch screen to your operating habits, you should make some basic settings now. For this purpose remain in the System set-tings menu.
If the touch screen is already adapted to your needs, you can continue with the section Preparing injector head on page 30.
Note! If you reactivate the display during injection, the injection will be interrupted automatically.
• Open System settings menu.
System settings 1/2
State of charge 90 %
Accutron MR 05/2008 25
26
Setting the touch screen 4 Commissioning the injector
Setting the touch screenIn order to be able to use the touch screen easily and without any problems, you should set it at the outset to suit your operating habits and the environmen-tal conditions. The System settings menu consists of two windows where you can make the following settings:– Set touch screen brighter and darker– Select language for the display texts– Adjust touch screen to the tactile characteristics of the operator– Set stand-by time for the injector– Set idling speed of the pistons– Set filling speed of the pistons
In the dialogue window System settings 1/2 the state of charge is displayed in percent:
Setting the touch screen brighter or darkerAdapt the brightness of the screen to your vision or to different ambient lighting conditions. For this purpose:• Touch the button of the slide controller in the top of the dialogue window.
Pull the slider to the left to make the display brighter. Pull the slider to the right to make the display darker.
• Open System settings menu.
System settings 1/2
State of charge 90 %
05/2008 Accutron MR
4 Commissioning the injector Setting the touch screen
The display changes immediately. As soon as you have set the brightness you want:
Selecting a languageIn this dialogue window you can select the language in which the texts in the touch screen are to be displayed. The following can be selected– German– French– Portuguese– English– Spanish– Italian
• Touch button with the relevant flag.
Calibrating touch screenEveryone operates a touch screen in their own particular way. A right-handed person, for example, does not position his finger at the same place as a left-handed person. The size and angle of the fingers also play a role. To ensure that the touch screen correctly adopts your entries and commands, you must first calibrate it to your touch habits.
To be able to calibrate the touch screen, you must first log into the system.
The dialogue window Login is displayed. You are requested to select your name.• Touch list entry User.• Enter the identification 1001 on the number buttons.
• Touch OK button.
The calibration window is displayed. Cross-hairs are shown in the centre of the window. Here you are requested in a german text:
"Bitte drücken Sie fest auf die Fadenkreuze. Bleiben Sie solange auf den Fadenkreuzen, bis diese verschwinden."
Which means: "Please press the cross-hairs firmly. Remain on the cross-hairs until they disappear."
• Save the brightness settings.
• Touch cross-hairs in the dialogue window System settings.
Accutron MR 05/2008 27
28
Setting the touch screen 4 Commissioning the injector
• Keep your finger pressed on the cross-hairs in the centre of the screen until they disappear.
The cross-hairs are gradually displayed in each corner of the screen.• Press each cross-hair until it disappears.
To be able to make the further settings, you must open the dialogue window System settings 2/2:
The dialogue window System settings 2/2 is displayed:
Setting stand-by timeHere you can determine after how many seconds the injector is to switch into the stand-by mode when it is not being used or during the injection.
Setting idling speed of the pistonsHere you can set at what speed the empty pistons are to be moved. 1 ... 10 ml/s can be set.
Setting filling speed of the pistonsHere you can set at what speed the liquids are to be filled. 1 ... 5 ml/s can be set.
• Save calibration.
• Open System settings 2/2.
System settings 2/2
Stand-by time
Idling speed
Filling speed
60 s
7 ml/s
4 ml/s
05/2008 Accutron MR
4 Commissioning the injector Setting the touch screen
Make all settings as follows:• Touch the relevant button with your finger.
A pop-up window opens. The value currently set and the entry range are displayed. • Overwrite the marked value using the numeric keypad with a new value.
If you have accidentally entered the wrong number, you can delete it with the DEL key.
The pop-up window closes.
When you have made all the settings, you can now either:
or
When all settings have been made, you can now insert the syringe. For this purpose read the next section Preparing injector head on page 30.
• Confirm the new value.
• change to the window System settings 1/2
• quit the dialogue window System settings to display the main window.
Accutron MR 05/2008 29
30
Preparing injector head 4 Commissioning the injector
Preparing injector head
Moving the pistons backOn delivery the pistons are retracted - however, it may happen that, prior to the routine change of the syringes, the pistons are advanced. In order to be able to insert a new syringe, you must first move the pistons back. For this purpose:
The dialogue message with the confirmation prompt "Move piston CM back?" is displayed.
Note! You can only insert the syringes into the injector head when the pistons of the injector are fully moved back.
CAUTION! Risk of infection! Do not remove the protective caps of the syringes until you have made the connections. Make sure that the packings of the disposable articles do not show any signs of damage.
• Call up filling menu in the main window.
• Touch left button for the CM piston.
• Confirm the message.
• Touch right button for the NaCl piston.
Move piston CM back?
05/2008 Accutron MR
4 Commissioning the injector Preparing injector head
The dialogue message with the confirmation prompt "Move piston NaCl back?" is displayed.
Both pistons move back into their lowest position.
The piston is retracted when it is right at the bottom of the syringe, the syringe is completely green/blue and the remaining volume displayed is 200 ml (see diagram).
• Confirm the message.
• Return to the main window.
Move piston NaCl back?
Prof. No. 5
Rem.vol. CM 200 ml
Rem.vol. NaCl 200 mlNot airfree
Accutron MR 05/2008 31
32
Preparing injector head 4 Commissioning the injector
Inserting the syringes• Make sure that both pistons are moved back.
• Push the syringe from the front into the respective recipient until it engages audibly.
Then you can connect the tube system, see next section.
Note! The left-hand piston is intended for the injection of contrast medium, the right-hand piston for NaCl solution.
05/2008 Accutron MR
4 Commissioning the injector Connecting the tube system
Connecting the tube system
The MR tube system has four valves which function as follows:– When you fill the syringe CM with contrast medium, valve A1 is opened and
valve A2 closed.– When you inject contrast medium out of the syringe CM, valve A1 is closed
and valve A2 opened.
The valves B1 and B2 function in the same way.
The MR tube is connected to the syringes as follows:• Remove the protective caps from the syringe and tube end.
• Screw the MR tube with 1/4 to 1/2 turn onto the syringe.
CAUTION! Risk of infection! Do not remove the protective caps of the tube ends until immediately before making the connections. Make sure that the packings of the disposable articles do not show any signs of damage.
Note! The tube marked with a green stripe is intended for the contrast medium. The tube section between valve A2 and the Y-piece is very short in order to save contrast medium.
ATTENTION! When screwing on the MR tube, you may damage the Luer-Lock thread and therefore make the syringe and tube system unusable. Therefore do not exceed one 1/2 turn (180°)
A1A2
B2
B1
syringe NaCl
to the contrast medium bottle
connection patient tube
to the NaCl bottle
syringe CM
Accutron MR 05/2008 33
34
Connecting the tube system 4 Commissioning the injector
• Repeat steps for the NaCl tube.
In the next step connect the MR tube to the bottle.• Hang the contrast medium bottle and NaCl bottle on the two bottle holders.
• Pull the protective cap off the pin of the drip chamber and press it into the contrast medium bottle (infusion level).
• Open valve cap on the drip chamber.
• Briefly press drip chamber together twice to fill it roughly half-full with contrast medium.
• Repeat steps for the NaCl drip chamber.
• Connect the patient tube.
The injector is now ready to be filled with contrast medium and NaCl.
05/2008 Accutron MR
5 Operating the injectorThis chapter describes how to operate the injector manually, how to work with injection profiles and how to carry out an injection.
Filling injector with contrast medium and NaCl
Both syringes are inserted, left for contrast medium, right for NaCl, the MR-tube system is connected, the injector is switched on. The main display is shown.
In order to fill the syringes, the pistons must first be moved forward.
Note! Also fill the injector with contrast medium first and then with the saline solution. In this way you save contrast medium when venting the tube system.
CAUTION! Risk of fatal or serious injuries due to air embolisms! Make sure that the patient is not yet connected to the system!
• Call up the filling menu in the main window.
Prof. No. 5
Phase Delays
Volumeml
Conc.%
Flowml/s
Inj. times
1
2
3
20
20
20
10
10
10 100
65
100 6.0
6.0
6.0 0
0
0
Not airfree
200 200
Accutron MR 05/2008 35
36
Filling injector with contrast medium and NaCl 5 Operating the injector
The filling menu is displayed.
Move both pistons forward
The dialogue message with the confirmation prompt "Move piston CM for-ward?" is displayed.
Note! Wenn the pistons are fully moved back, the correspondent filling button isn't active and thus marked grey.
• Touch left button for the CM piston.
• Confirm the message.
Prof. No. 5
Rem.vol. CM 200 ml
Rem.vol. NaCl 200 mlNot airfree
Move piston CM forward?
05/2008 Accutron MR
5 Operating the injector Filling injector with contrast medium and NaCl
The dialogue message with the confirmation prompt "Move piston NaCl for-ward?" is displayed.
Both pistons of the syringes move forwards into their uppermost position, the remaining volume displayed becomes lower on both sides, the two symbols for the syringes become white.
When the pistons are fully extended, the filling menu is shown like this:
• Touch right button for the NaCl piston.
• Confirm the message.
Note! When advancing the pistons, the valves may cause a low hissing sound – this is normal.
Move piston NaCl forward?
Prof. No. 5
Rem.vol. CM 0 ml
Rem.vol. NaCl 0 mlNot airfree
Accutron MR 05/2008 37
38
Filling injector with contrast medium and NaCl 5 Operating the injector
Enter filling volume for contrast medium and fill
A dialogue window with number buttons is displayed.
The range which you can enter, e.g. 0.0 ... 200.0, is in the top right of the dia-logue window Enter filling volume and fill. The current (maximum) value is already specified. You can change the value by entering the value you want on the numpad. If you have accidentally entered a wrong number, you can delete it using the DEL button.• Enter the filling volume you want.
The dialogue window Enter filling volume and fill is closed. The piston of the syringe for contrast medium moves back.
Enter filling volume for NaCl and fill
A dialogue window with number buttons is displayed.
The range which you can enter, e.g. 0.0 ... 200.0, is in the top right of the dia-logue window Enter filling volume and fill. The current (maximum) value is already specified. You can change the value by entering the value you want on the numpad. If you have accidentally entered a wrong number, you can delete it using the DEL button.• Enter the filling volume you want.
The dialogue window Enter filling volume and fill is closed. The piston of the syringe for NaCl moves back.
• In the filling menu touch the Fill button in order to enter the filling volume for contrast medium.
• Confirm entry.
• Touch the blue Fill button in the filling menu to enter the filling volume for NaCl.
• Confirm entry.
05/2008 Accutron MR
5 Operating the injector Filling injector with contrast medium and NaCl
While the pistons draw in contrast medium and NaCl, you see the following display:
• Then connect the patient tube to the MR tube.
• When the filling procedure is finished, return to the main window.
CAUTION! Risk of fatal or serious injuries due to air embolisms! At the latest now you must connect the patient tube to the MR tube so that the entire tube system right to the patient is completely vented.
Prof. No. 5
Rem.vol. CM 102 ml
Rem.vol. NaCl 135 mlNot airfree
Accutron MR 05/2008 39
40
Filling injector with contrast medium and NaCl 5 Operating the injector
Venting the tube systemThe syringes and the tube system still contain air. You must vent the system using the buttons for the manual piston movements.
• Make sure that the patient tube is connected to the MR tube.
• Swivel the two injection units into the vertical position.
• Keep venting the CM side until there are no more air inclusions just after the Y-piece.
• Vent the entire tube system incl. the patient tube with the NaCl piston.
• Keep venting until there are no more air inclusions in the entire tube system.
During venting the injection volume actually available appears in the window Rem. vol.(ml).
Note! Firstly, vent the contrast medium side up to the Y-piece and only then the entire system including the patient tube with the piston for NaCl. In this way you save contrast medium.
•
In the left-hand button group for the manual piston movements press the button Piston forward and keep it depressed to advance the CM piston.
• To gradually increase the piston speed, also press this adjacent button.
• In the right-hand button group for the manual piston movements press the button Piston forward and keep it depressed to advance the NaCl piston.
• To gradually increase the piston speed, also press this adjacent button.
CAUTION! Risk of fatal or serious injuries due to air embolisms! There must be no more air inclusions in the entire tube system.
05/2008 Accutron MR
5 Operating the injector Filling injector with contrast medium and NaCl
The dialogue message with the confirmation prompt "CM-syringe airfree?" is displayed.
• On completion of venting touch the now active button Airfree or the text field Not airfree.
• Confirm completed venting.
Prof. No. 5
Phase Delays
Volumeml
Conc.%
Flowml/s
Inj. times
1
2
3
20
20
20
10
10
10 100
65
100 6.0
6.0
6.0 0
0
0
Not airfree
197 195
CM-syringe airfree?
Venting
Accutron MR 05/2008 41
42
Filling injector with contrast medium and NaCl 5 Operating the injector
The dialogue message with the confirmation prompt "NaCl-syringe airfree?" is displayed.
The parameters of the current injection profile are displayed. The text field indicates "Airfree" in green. The Activate touch button is highlighted in colour and therefore active.• Press injection unit downwards into the injection position.
The injector is now ready for the injection.
• Confirm completed venting.
NaCl-syringe airfree?
Venting
05/2008 Accutron MR
5 Operating the injector Changing the maximum operating pressure
Changing the maximum operating pressureYou can set the desired injection pressure. The pressure actually reached dur-ing injection depends on a number of ambient variables such as the viscosity of the contrast medium, needle diameter, set flow rate etc.
If the set maximum working pressure is reached during the injection, the injec-tion is interrupted.
The maximum operating pressure can only be changed on the remote control supplied with the Accutron MR. Please read the separate instructions for use to see how to proceed for this purpose.
CAUTION! Risk to the patient! Make sure that the maximum pressure set does not endanger the patient.
Accutron MR 05/2008 43
44
Selecting an injection profile 5 Operating the injector
Selecting an injection profileYou can create with the injector up to 80 injection profiles and store them under a profile number. Depending on the application, you call up the profile you want with the injection parameters saved and you can immediately start the injection. For this purpose proceed as follows:
The dialogue window with the injection profiles already available is displayed.
Using the arrow buttons, you can scroll to the right and left and in this way view all profile names.• Touch the syringe symbol of the profile you want to use.
In this way you have selected a profile. The parameters of this profile (here Profile No. 7) are in the main window.
• Call up profile.
Prof. No. 7
Prof. No. 1
Prof. No. 2
Prof. No. 3
Prof. No. 4
Prof. No. 5
Prof. No. 6
Prof. No. 7
Prof. No. 8
Prof. No. 9
Prof. No. 10
05/2008 Accutron MR
5 Operating the injector Changing an injection profile
Changing an injection profileYou can change the parameters of an injection profile as required.
The touch screen displays the injection profile to be changed.
Example: The profile to be changed is displayed. You want to change the volume of a phase.• Touch the relevant value with your finger.
A pop-up window opens. The value currently set and the entry range are displayed.• Overwrite the marked value using the numeric keypad with a new value.
If you have accidentally entered the wrong number, you can delete it with the DEL key.
You edit the other values, such as delay, concentration and flow in the same way.
• Confirm the new value.
Note! Take note before entering the concentration:
– 100% means contrast medium only – 40% means 40% contrast medium, 60% NaCl – 0% means NaCl only
Prof. No. 7
Phase Delays
Volumeml
Conc.%
Flowml/s
Inj. times
1
2
3
20
20
20
10
10
10 100
65
100 6.0
6.0
6.0 0
0
0
Not airfree
180 178
Accutron MR 05/2008 45
46
Changing an injection profile 5 Operating the injector
The system displays the changed value in the main window. The profile name has changed automatically. Prof. No. 0 is now shown in the main window.
Storing changesIf you have made changes to the parameters of an injection profile, you must store the profile. However, after a change the original name of the profile is no longer on the screen but Prof.Nr. 0.
Note! Some values have a reciprocal effect on one another. If you change the volume value or the flow, the injection time of this phase changes automatically. If you change the injection time, the flow is changed accordingly.
CAUTION! Risk of serious injuries! Before entering the flow (up to 10 ml/s possible!), check the constitution of the patient to be examined.
ATTENTION! Risk of data loss! Save changes to parameters under a profile number. Otherwise the changes are lost when you switch off the injector.
• Save profile.
05/2008 Accutron MR
5 Operating the injector Changing an injection profile
The dialogue window with the injection profiles already available is displayed.
You can scroll to the right and left using the arrow buttons and in this way view all the profile names.
You have two possibilities:– overwriting an existing profile, see next paragraph, or– saving the changes under a new profile, see next paragraph but one.
Overwriting an existing profileIf you want to overwrite an already existing profile,• touch the appropriate syringe symbol.
The writing in the syringe symbol is green and the name of the selected profile number is now in the text field for the profile name.
Prof. No. 7
Prof. No. 1
Prof. No. 2
Prof. No. 3
Prof. No. 4
Prof. No. 5
Prof. No. 6
Prof. No. 7
Accutron MR 05/2008 47
48
Changing an injection profile 5 Operating the injector
Entering a profile nameIf you want to change the name of this profile:• touch the dark-blue text field.
The touch screen displays the input mask for the profile name.
You can enter the new profile name using the keyboard. You can delete indi-vidual letters with the DEL button or delete the entire name with the Delete button. You can write capital letters with the SHIFT button.
• Enter a new profile name.
• Touch the OK button to confirm the entry.
The changed profile is stored and the main window is displayed.
• Save the new name for the profile.
new name
Delete
ESC
05/2008 Accutron MR
5 Operating the injector Changing an injection profile
Saving changes under a new profileIf you want to save the changed profile under a new profile,• select the empty syringe symbol.
The marking in the syringe symbol is green and is also shown at the top in the dark-blue text field.• Touch the dark-blue text field.
The touch screen displays the input mask for the profile name, see preceding section.• Enter the name for the new profile with the keyboard.
• Touch the OK button to confirm the entry.
The new entry is now in the profile list.
The main window displays the newly created profile.
• Save the new profile with the new profile name.
new name
Prof. No. 1
Prof. No. 2
Prof. No. 3
Prof. No. 4
Prof. No. 5
Prof. No. 6
Prof. No. 7
new name
Accutron MR 05/2008 49
50
Changing an injection profile 5 Operating the injector
Creating a new profileIf you want to create a completely new profile, you must first select the relevant syringe symbol. For this purpose:
The list with the profiles is displayed.
You can scroll to the right and left with the arrow buttons and in this way view all the syringe symbols.• Choose the empty syringe symbol of the new profile and touch it.
In the main window you have opened an empty profile with only one phase which you can now fill with the injection parameters you want.• Enter all the profile data and parameters you want.
• Select the syringe symbol you want again.
• Touch dark-blue text field to enter the profile name.
The touch screen displays the input mask for the profile name, see page 48.• Enter the name for the new profile with the keyboard.
• Call up profile in the main window.
• Save profile.
-
Prof. No. 1
Prof. No. 2
Prof. No. 3
Prof. No. 4
Prof. No. 5
Prof. No. 6
-
05/2008 Accutron MR
5 Operating the injector Changing an injection profile
• Touch the OK button to confirm the entry.
The main window displays the newly created profile.
• Save the new profile with the new profile name.
Accutron MR 05/2008 51
52
Changing an injection profile 5 Operating the injector
Adding a new phaseYou can add a new phase (a maximum of 6 are possible) to an existing profile with, for example, 2 phases. The main window of the relevant profile is displayed.
• Touch parameter fields of the new phase.
• Enter the individual values in the input mask using the number buttons.
The main window of the profile is displayed.
• Touch the appropriate syringe symbol.
The lettering in the syringe symbol is green and the profile number is in the text field at the top.• Enter profile name as described on page 48.
• Touch the OK button to confirm the entry.
The changed profile is stored and displayed in the main window.
• Add a phase.
• Confirm the new value.
• Save the changes.
• Save the profile under the profile name selected.
05/2008 Accutron MR
5 Operating the injector Changing an injection profile
Deleting a phaseYou can completely delete the last phase from a profile. Proceed as follows:
• Touch the appropriate syringe symbol.
The lettering in the syringe symbol is green and the profile number is in the text field at the top.• Enter profile name as described on page 48.
• Touch the OK button to confirm the entry.
The changed profile is stored and the main window is displayed.
• Delete phase in the main window of the correspondent profile.
• Save the changed profile.
• Save the profile under the profile name selected.
Accutron MR 05/2008 53
54
Performing an injection 5 Operating the injector
Performing an injection
– Do not connect the patient to the system until you have vented the syringes and the tube system.
– Vent the syringes and the tube system every time after filling the injector with contrast medium or saline solution.
– Make sure before every injection that there are no air inclusions in the syringes and the connected tubes.
– Only start an injection when the syringes are in the lowered position (injection position).
– Make sure before every injection that the injection parameters shown on the display do not endanger the patient.
• Swivel the injection units out of the vertical position backwards into the lowest position (injection position):
This injection position ensures that air bubbles rise to the piston and are not injected.• Make sure again that the MR tube and the patient tube systems do not
contain any air inclusions.
• Remove protective cap from the patient tube and connect the tube to the patient cannula.
The injector is vented, the Activate button on the touch screen is active.
CAUTION! Risk of fatal or serious injuries due to air embolisms!
CAUTION! Risk of uncontrolled supply of contrast medium due to hydrostatic pressure compensation! Make sure that the patient and the syringes are at the same level.
05/2008 Accutron MR
5 Operating the injector Performing an injection
Maintaining vein access with the KVO functionBefore performing the injection, you can now establish whether you want to activate the Keep-Vein-Open function (KVO).The KVO function ensures that access to the vein is retained during prolonged examinations.
If you activate this function, the injector injects 0.5 ml of saline solution every 2 minutes.
– When the KVO function is activated, the writing in the button is green.– If the KVO function is deactivated, the writing in the button is grey.
• If you want, activate KVO.
Note! If the NaCl volume falls below the minimum of an injection volume with the KVO function activated, the KVO function is switched off and in, in addition, a warning appears in the display.
Accutron MR 05/2008 55
56
Performing an injection 5 Operating the injector
Starting the injection
Prof. No. 7
Injector readyKVO on
Ph. Delays
Vol.ml
Conc.%
Flowml/s
Times
1
2
3
20
20
20
10
10
10
100
65
100
6.0
6.0
6.0 0
0
0
0.0 s
The following message is displayed:
• Touch the Activate button.
• Using the Info button, you can change to a second window with more injection data.
Prof. No. 7
Injected volume - CM
Injected volume - NaCl
Phase time
Delay
Injection time
0
0
0.0
20.0
86.7
Injector readyKVO on
0.0 s
05/2008 Accutron MR
5 Operating the injector Performing an injection
The LEDs of the start and stop button on the control panel come on.
The injection starts.
The pilot light on the control panel flashes green.
After the start of the injection, you can follow the individual phases and the details of already injected and remaining contrast medium, phase time, delay, injection time and scan delay on the display of the remote control. The stop-watch in the top right begins to run when you press the start button.
Interrupting the injection on the remote controlYou can interrupt the injection at any time by touching the HOLD button on the remote control. In the confirmation prompt then displayed you have two pos-sibilities:– Touch the End button if you want to interrupt the injection completely, – Touch the Continue button if you want to continue the interrupted injection.
The unit then continues the injection at the point where it was interrupted.
Interrupting the injection on the injectorIf the display is deactivated, you can stop the injection at any time by
or by
The injector stops the injection.
The injection data window is displayed, and you can either continue the injec-tion or stop it completely.
• Press the start button on the control panel of the remote control or on the injector.
Note! The touch screen on the injector switches off on expiry of the set stand-by time and you can follow the course of the injection on the remote control.
• pressing the Stop button on the control panel of the injector
• pressing the Wake-up button on the control panel of the injector.
Accutron MR 05/2008 57
58
Performing an injection 5 Operating the injector
Injection endThe program stops after passing through all the phases. All time values of the injection count backwards to 0.The injected volume counts forwards.
The message "Injection finished" is displayed.
• The main window is displayed in the touch screen.
• When the examination is over, disconnect the patient tube from the cannula and dispose of the cannula.
If you want to perform further injections, change the patient tube, prepare a new injection, replenish contrast medium and vent, as described above.
If you want to stop working with the injector, read and follow the instructions in the chapter Post-examination work on page 64.
Note! The program run stops when all the phases have been passed through. If the KVO function has been activated, NaCl is injected from this time onwards until you confirm the program end.
• Confirm message.
• If necessary, wake up touch screen on the injector.
CAUTION! Risk of infection! Replace the patient tube with a new one after it has been used once. Otherwise, you jeopardise your patients' health.
Observe the applicable hygiene regulations!
05/2008 Accutron MR
5 Operating the injector Replenishing injector with contrast medium/NaCl
Replenishing injector with contrast medium/NaClWhen the contrast medium or NaCl filling of the syringes is coming to an end and you want to perform more injections or if the message "CM-volume too low" appears, top up the syringes.
How to proceed is described here taking the example of contrast medium.• Confirm message "CM-volume too low".
The main window of the touch screen is displayed.• Remove the patient tube.
The filling menu is displayed.
CAUTION! Risk of fatal or serious injuries due to air embolisms! Make sure that the patient is not yet connected to the system!
• Touch the Fill button in the main window.
Prof. No. 5
Rem.vol. CM 28 ml
Rem.vol. NaCl 33 mlNot airfree
Accutron MR 05/2008 59
60
Replenishing injector with contrast medium/NaCl 5 Operating the injector
Enter filling volume and fill
The dialogue window Enter filling volume and fill with number buttons is dis-played.
The range with the possible values you can enter, e.g. 0.0 … 172.0, is in the top right of the dialogue window. The current (maximum) value is already specified, in this example that is 172 ml as 28 ml are available as the remaining volume. You can change the value by entering the value you want on the numeric keypad. If you have accidentally entered the wrong number, you can delete it with the DEL button.• Enter the filling volume you want.
The dialogue window Enter filling volume and fill is closed. The piston of the syringe moves back. While the piston draws in the contrast medium, you see the following display:
• Touch the green Fill button in the filling menu to enter the filling volume for contrast medium.
• After entry, confirm the value you want.
• When the filling procedure is finished, return to the main window.
Prof. No. 5
Rem.vol. CM 172 ml
Rem.vol. NaCl 33 mlNot airfree
05/2008 Accutron MR
5 Operating the injector Replenishing injector with contrast medium/NaCl
Venting after replenishing• Then connect the patient tube to the MR tube.
At present, the syringe and the tube system still contain air. You must vent the system using the buttons for the manual piston movements.
• Make sure that the patient tube is connected to the MR tube.
• Keep venting CM side until there are no more air inclusions downstream of the Y-piece.
With the NaCl side vent the entire tube system including the patient tube.
• Keep venting until there are no more air inclusions in the entire tube system.
CAUTION! Risk of fatal or serious injuries due to air embolisms! At the latest now you must connect the patient tube to the MR tube so that the entire tube system right to the patient is completely vented.
Note! First vent the contrast medium side up to the Y-piece and only then the entire system including the patient tube with the piston for NaCl. In this way you save contrast medium.
•
In the left-hand button group for the manual piston movements press the button Piston forward and keep it depressed to advance the CM piston.
• To gradually increase the piston speed, also press this adjacent button.
•
In the right-hand button group for the manual piston movements press the button Piston forward and keep it depressed to advance the NaCl piston.
• To gradually increase the piston speed, also press this adjacent button.
CAUTION! Risk of fatal or serious injuries due to air embolisms! There must be no more air inclusions in the entire tube system.
Accutron MR 05/2008 61
62
Replenishing injector with contrast medium/NaCl 5 Operating the injector
During venting the injection volume actually still available appears in the win-dow Rem.vol.(ml).
The dialogue message with the confirmation prompt "CM-syringe airfree?" is displayed.
• On completion of venting touch the now active button Airfree or the text field Not airfree.
• Confirm completed venting.
Prof. No. 7
Phase Delays
Volumeml
Conc.%
Flowml/s
Inj. times
1
2
3
20
20
20
10
10
10 100
65
100 6.0
6.0
6.0 0
0
0
Not airfree
197 29
CM-syringe airfree?
Venting
05/2008 Accutron MR
5 Operating the injector Replenishing injector with contrast medium/NaCl
The dialogue message with the confirmation prompt "NaCl-syringe airfree?" is displayed.
The parameters of the current injection profile are displayed. The text field indicates "Airfree" in green. The Activate button is highlighted in colour and therefore active.
The injector is now ready for the injection.
• Confirm completed venting.
NaCl-syringe airfree?
Venting
Accutron MR 05/2008 63
64
Post-examination work 5 Operating the injector
Post-examination work• When the examination is over, disconnect the patient tube from the cannula
and dispose of the cannula.
• Remove the syringe with tube system. This is possible at every piston position.
• Properly dispose of all disposable articles in accordance with the applicable provisions.
• In the filling menu, move both pistons back, see page 30.
• Press the On/Off switch to switch the injector off.
The green LED of the switch and the pilot light on the control panel go off.• Connect the injector to the charger in order to charge the battery cells for
use the next day.
CAUTION! Risk of infection! Replace the patient tube with a new one after it has been used once. Otherwise, you jeopardise your patients' health.
Observe the applicable hygiene regulations!
Note! To attach a new syringe, the piston has to be in the lowest position.
05/2008 Accutron MR
6 Annex
Waste disposal
Safety inspections and maintenance
MEDTRON prescribes annual maintenance and a safety inspection for the injector. The maintenance work is performed by staff trained and authorised by MEDTRON. Regular maintenance ensures the functionability of the injec-tor.
Safety Inspections and maintenance work should be certified and documented in the medical equipment logbook.
Cleaning and storageRemove any disposable articles from the injector immediately after use and dispose of them properly. Disposable articles must not be sterilised and re-used. Remove all dirt caused by contrast medium with warm water before the dirt dries in.
To disinfect the Accutron MR, we recommend Mikrozid Liquid from Schülke & Mayr GmbH, Germany. Do not use any aggressive disinfectant to disinfect the injector.
In the case of prolonged breaks in operation, store the Accutron MR in a safe place where it is protected against dust and moisture.
Do not put old devices to the normal domestic waste! MEDTRON undertakes to take back its old devices as well as electric and electronic accessories, such as chargers, remote controls and similar. The disposable articles are to be disposed of in a proper manner.
CAUTION! Risk of injury to operator and patient! Maintenance and repair work may only be performed by the MEDTRON Customer Service or people authorised by MEDTRON.
ATTENTION! Risk of malfunctions of the Accutron MR! Do not immerse the injector in water! Do not use any aggressive cleaning agents or solvents. Warm water and a mild soap are sufficient.
Accutron MR 05/2008 65
66
System messages 6 Annex
System messagesThe following table contains system messages which indicate either faults in the injector or incorrect operation. Using the following table, try to eliminate the cause of the fault in question. If the cause remains unclear and the suggestions for remedy prove unsuc-cessful, please contact the MEDTRON Customer Service.
Message/Error Possible cause and remedy
CM VOLUME TOO LOW
There is too little contrast medium in the syringe to implement the injection profile.Acknowledge message and replenish contrast medium, see page 59.
NACL VOLUME TOO LOW
There is too little NaCl in the syringe to implement the injection profile.Acknowledge message and replenish NaCl, see page 59.
No operation possible
The value for the volume or flow is missing in at least one phase. Enter missing values.
CONFIGURATION ON THE REMOTE CONTROL
This message appears while settings are being made on the touch screen of the remote control.
At the same time:CONFIGURATION ON THE REMOTE CONTROLCONFIGURATION ON THE INJECTOR
The injector and remote control were operated at the same time.Press the stop button on the injector.Never operate units at the same time.
The unit was not switched off correctly! Please remove the syringe and activate the OK button. The piston moves back!
05/2008 Accutron MR
6 Annex Technical data
Technical data
The injector satisfies the requirements of the standard DIN EN 60601-1-2 for medical equipment in terms of the emitted interference and interference immu-nity.
Flow rate per unit 0.1 ml/s ... 10 ml/s, increment 0.1 ml/s
Injection volume Contrast medium: 200 ml Saline solution: 200 ml
Partial volume which can be pre-selected
1 ml ... 200 ml, increment 1 ml
Max. injection pressure 21 bar
Filling speed 1 ml ... 5 ml
KVO parameters 0.5 ml of saline solution every 2 minutes.
Injection profiles 80
Injection phases 1 ... 6
Accuracy The accuracy of the contrast medium volume injected is 1% of the syringe volume (only applies to the empty syringes recommended by MEDTRON)
Charger Mascot Type 2040 with medical approval
Input voltage charger 100 ... 240 V~, 50 ... 60 Hz
Power consumption charg-er
< 100 VA
Protection class injector device with internal power supply
Device type CF, identified by
Dimensions (H × W × D) 1180 × 470 × 620 mm in the injection position
Weight 37 kg
Ambient conditions during operation
– operating temperature 10 °C ... 40 °C – relative humidity 30 % ... 75 % – air pressure 700 hPa ... 1060 hPa
Ambient conditions for storage and transport
– storage temperature - 20 °C ... 60 °C – relative humidity up to 75 % – air pressure 700 hPa ... 1060 hPa
Rechargeable batteries 24 V / 17 Ah
Operating time when com-pletely charged
Fully charged batteries permit at least 100 injec-tions of 30 ml at 2.0 ml/s
Radio communication Bluetooth version 2.0
Bluetooth class 1
Frequency range 2.402 ... 2.48 GHz
Accutron MR 05/2008 67
68
Technical data 6 Annex
05/2008 Accutron MR
Index
Index
AAir inclusions 40, 61Ambient light 26
BBattery cells 14
charging 23charging time 23
Battery symbol 19Bluetooth class 67Bottle holders 14Breaks in operation 65Brightness
setting 26
CCalibrating touch screen 27Calibration window 27Castors 14, 22Charger 23Cleaning and storage 65Column stand 14Concentration 19
changing 45Connections 16Contrast medium
filling 35replenishing 59
Contrast medium sideventing 61
Control panel 17Cross-hairs 27
DDelay 19, 46Disinfection 65Disposable articles
disposing of 64Drip chamber 34
EElectromagnetic compatibility 11EMC 11
FFadenkreuze 27Fault messages 66Filling speed 67
setting 28Flow 19, 46
changing 45Frequency range 67
HHandle 22
IIdling speed
setting 28Inj. time 19Injected volume 19Injection
end 58performing 54phases 67position 54pressure 67profiles 67starting 56unit 14
Injection profilestoring changes 46
Injectorcommissioning 23switching ON 24
Inspection and maintenance work 65Intended use 6Items supplied 14
KKeep-Vein-Open function 55KVO 19KVO parameters 67
LLanguage
selecting 27Login 27
Accutron MR 05/2008 69
70
Index
MMain window 18Maintenance work 65Manufacturer’s plate 6Medical equipment logbook 65Model name 6
NNaCl
filling 35replenishing 59
NaCl sideventing 40, 61
OOn/Off switch 14, 22Operating
habits 26pressure 43time 67
Optional accessories 14
PParameters
changing 45Phase 19
adding new 52Phase time 19Pistons
move forward 36moving back 30
Pressurechanging 43
Profilechanging 45selecting 44
Protectionclass 67of people 9of property 9
RRadio communication 67Receiving aerial 14Rechargeable batteries 67Recipient 14Remote control 22Repair work 65
SSafety instructions 9Scan delay 19Serial number 6State of charge
displaying 25Storage 65Switching ON 24Syringe
inserting 32venting 40
Syringe symbol 44, 50System settings 26, 28
TTouch screen 18
calibrating 27
VVenting the syringe 40Volume 19
05/2008 Accutron MR
Index
Accutron MR 05/2008 71