Absolute Multiline-Technology
Applications and new developments
![Page 2: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/2.jpg)
Etalon AG Founded 2004 as a spin-off of PTB in Braunschweig / Germany
Specialized in Interferometric measurement and calibration of machines and structures
Worldwide customer base in industry and research
Today 20 people, 70 % Master / PhD
2016 subsidiary in the US (Seattle) established
![Page 3: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/3.jpg)
LaserTRACER-NG
Calibration of CMM and machine tools with
sub-micron accuracy
LaserTRACER-MT
Calibration of 5 axismachine tools: Accurate,
complete and fast
Absolute Multiline Technology
Absolute multi-channelinterferometry for
monitoring anddeformation analysis
Etalons product line
![Page 4: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/4.jpg)
Absolute Multiline®-Technology
Absolute interferometer
Central unit with up to 88 independent channels
Measurement uncertainty (95%) : 0.5 µm/m
Measurement length up to 20 m
Simple measurement channel consisting only of telecom fiber, collimator and triple reflector (no electrical systems at detector)
Almost unlimited fiber length possible (several kilometers)
Eye save infrared radiation
Metrological stability by gas absorption cell
Usable in vacuum and cryogenic applications
No real-time capability !
![Page 5: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/5.jpg)
Classical “frequency scanning interferometry”
Frequency scanning laser
Reference length: LR
Measurement length: LM
Zeit
Intensity
Intensity
Zeit
Laser is modulating its wavelength and is generating interferences on measurement arm and reference arm.
Number of fringe counts depends on the measured distance
Phase of interference is evaluated
Main restrictions:
Motion of target is degrading the measurement accuracy
Reference length must be known and long term stable
R
M
R
M
L
L
Phase
Phase
![Page 6: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/6.jpg)
3. Integration of an optical signal distribution
Distribution of the signal to multiple measurement lines
Simultaneous measurements on all channels possible
Telcom Technology
Laser 1
Reference length: LR
Meas. length: LM1
Gas
cell
Extensions “Absolute Multiline Technology”
2. Second Laser with inversed scanning direction
Motion of target during the shot can be measured -> Motion of target does not degrade measurement accuracy
1. Integration gas absorption cell
• Stable wave length reference
Reference length can be determined
Source of metrological traceability
Laser 2
Meas. length: LM2
Meas. length: LM100
…..
![Page 7: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/7.jpg)
Relative deviation between 0.2 m and 20 m
Relative standard deviation = 0.17 ppm
![Page 8: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/8.jpg)
Monitoring of a vibrating target
300Hz
5 μm
900Hz600Hz
5.5 μm @ 300 Hz
![Page 9: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/9.jpg)
-1,0-0,9-0,8-0,7-0,6-0,5-0,4-0,3-0,2-0,10,00,10,20,30,40,50,60,70,80,91,0
0 200 400 600 800 1000 1200 1400 1600
Deviation to reference interferometerDeviation [µm]
Measured distance [mm]
![Page 10: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/10.jpg)
Application : Deformations and vibration on mechanical systems
![Page 11: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/11.jpg)
Application: Geometrical control of rigs and fixtures
![Page 12: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/12.jpg)
Application: monitoring of machine deformations(a) while machine development and testing(b) constantly integrated in machine structure
![Page 13: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/13.jpg)
40 m
Application: Alignment and monitoring of Telecope structures
![Page 14: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/14.jpg)
LCLS Variable Gap Undulator
Application: Deformation measurement on ondulator (SLAC)
![Page 15: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/15.jpg)
Application: Alignment and monitoring of components in vacuum/ cryogenic/radioactive environment
![Page 16: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/16.jpg)
productionF&E departmentserver room inspection room
Application: Facility wide metrology system for central traceability in all areas
![Page 17: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/17.jpg)
First Installation of Absolute Multiline Technology For CMM and machine tool monitoringat the Nuclear Advanced Manufacturing Research Center (NARMC)
Absolute interferometer technology for monitoring entire production machinery
Geometry of all machines under control (CMM and MT)
Reliability of production increased
Quality inspection of parts can be done on machine tool
![Page 18: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/18.jpg)
Permanently mountedbeam launchers
Example: Monitoring of a bridge type machine
![Page 19: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/19.jpg)
Satellites with multiple measurement beams
![Page 20: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/20.jpg)
Beam launcherIn protected satellite
![Page 21: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/21.jpg)
Characteristic 1st generation 2nd generation
Sampling electronics custom build µTCA with custom buildanalogue RTM
Sampling rate 2.7 MHz 125 MHz
Data transfer USB 2 10 Gbit Ethernet
Uncertainty U(95%)= 0.5 ppm U(95%)= 0.3 ppm
Channels per extension board
16 12
Maximum distance 20 m > 30 m
Maximum measurement channels
88 124
Measurement andEvaluation
sequential parallel
Features of the 2nd generation Absolute Multiline System
![Page 22: Absolute Multiline-Technology Applications and new developments · 2018. 11. 19. · Absolute Multiline®-Technology Absolute interferometer Central unit with up to 88 independent](https://reader035.vdocuments.us/reader035/viewer/2022081619/60f91da94f066728b13f0315/html5/thumbnails/22.jpg)
Summary
Etalons Absolute Multiline Technology offers unique possibilities for monitoring and alignment tasks in physics and precision engineering
Various applications have been identified and presented
Facility wide networks of interferometer channels including TCP/IP controlled switches are possible
2nd generation will introduce new performance specifications