a practical guide to running the scanning electron
TRANSCRIPT
-
7/21/2019 A Practical Guide to Running the Scanning Electron
1/61
APractical
Guide
To
Running
the
ScanningElectronMicroscope
WilliamS.Monroe
-
7/21/2019 A Practical Guide to Running the Scanning Electron
2/61
Sections
I. SampleIntroduction
II. DataAcquisition
III. SampleExtraction
-
7/21/2019 A Practical Guide to Running the Scanning Electron
3/61
-
7/21/2019 A Practical Guide to Running the Scanning Electron
4/61
LogintoSEMLogBook(monitoron
theleft)
-
7/21/2019 A Practical Guide to Running the Scanning Electron
5/61
Loginto
SEM
Log
Book
Loginbywritinganewrow. Informationrequiredforloggingin:
Date,UserName,Advisor,Sample/Project,TimeIn,
SampleClean?,
Stage
at
Zero?,
ChamberPressure,GunPressure,andEmissionCurrent
WheretofindChamberPressure,
GunPressure,
and
Emission
Current
-
7/21/2019 A Practical Guide to Running the Scanning Electron
6/61
SampleIntroduction:
Turnon
chamber
camera
1:Selectchambercamera(lowerrightquadrant)
2:clickpausebuttoninthetoolbarto
activatechamber
camera
feed
-
7/21/2019 A Practical Guide to Running the Scanning Electron
7/61
SampleIntroduction:
Makesure
that
stage
is
in
the
zero
position
1:Makesurethestageislowered,
Movemouse
inside
chamber
camera
videofeedandclickscrollbuttonon
mouseanddragdown
2:Makesurethestageiscentred(afterstageislowered),
PressCTRL+0(thatisazero,nottheletterO)
-
7/21/2019 A Practical Guide to Running the Scanning Electron
8/61
-
7/21/2019 A Practical Guide to Running the Scanning Electron
9/61
SampleIntroduction:
Ventthe
chamber
Ifstage
is
lowered
and
centred:
VentthechamberbypressingtheVentButton
-
7/21/2019 A Practical Guide to Running the Scanning Electron
10/61
SampleIntroduction:
Ventingthe
chamber
Duringventing,
the
chamber
portion
of
the
SEMiconwillchangefromgreentoorange.
-
7/21/2019 A Practical Guide to Running the Scanning Electron
11/61
SampleIntroduction:
Thechamber
is
vented
Oncethe
chamber
is
vented,
the
chamber
portionoftheSEMiconwillchangefrom
orangetodarkgray.
-
7/21/2019 A Practical Guide to Running the Scanning Electron
12/61
SampleIntroduction:
Openthe
chamber
Openthe
chamber
by
pulling
on
themetalbaronthebottomof
thestage.CAUTION:Donotleanonanypartofthe
instrumentorapplypressureinotherareas
than
the
metal
bar
on
the
bottom
of
the
stage. Doingsocouldharmthevacuum.
-
7/21/2019 A Practical Guide to Running the Scanning Electron
13/61
SampleIntroduction:
Loadyour
sample
Wear
clean
gloves!!!CAUTION:
You
MUST
wear
clean
gloves
when
handling
and
loading
your
sample.
-
7/21/2019 A Practical Guide to Running the Scanning Electron
14/61
SampleIntroduction:
Loadyour
sample
Thisisthestage,youwillseenumbered
holes.
Placeyour
sample
to
sit
in
one
of
theseholes.
CAUTION:Thestageisdelicate,donotforce
anything!
Slot
4
is
broken
because
of
a
user
exertingtoomuchforce. Toomuchpressure
maypermanentlydamagethestage.
-
7/21/2019 A Practical Guide to Running the Scanning Electron
15/61
SampleIntroduction:
Loadyour
sample
Exampleofaplacedsample
-
7/21/2019 A Practical Guide to Running the Scanning Electron
16/61
SampleIntroduction:
Closethe
chamber
Ifyour
sample
is
too
large,
STOP! Reopenthestageand
removeyoursample. Contact
WilliamMonroeandhewillload
ashorterstageforyou.
Watchthechambercamera.
Slowlyandgentlyclosethestage.
Makesurethatyoursampleisnot
goingtohitanyofthedetectors.
-
7/21/2019 A Practical Guide to Running the Scanning Electron
17/61
SampleIntroduction:
Pumpthe
chamber
Withyoursamplesafelyloadedandthe
chamberclosed,
press
the
pump
button
-
7/21/2019 A Practical Guide to Running the Scanning Electron
18/61
SampleIntroduction:
Pumpingthe
chamber
Theiconwillchangetoorangeduringpumping
-
7/21/2019 A Practical Guide to Running the Scanning Electron
19/61
SampleIntroduction:
Turningthe
beam
on
Theiconwillchangetogreenoncepumpingis
complete
Turnontheelectronbeam
usingtheBeamOnbutton
-
7/21/2019 A Practical Guide to Running the Scanning Electron
20/61
SampleIntroduction:
Viewingthe
microscope
image
Selecttheupperleftquadrantbyclickingonit
Then,activatethe
SecondaryElectronDetector
byunpausingthatwindow
-
7/21/2019 A Practical Guide to Running the Scanning Electron
21/61
SampleIntroduction:
SecondaryElectron
Detector
Output
Youshouldnowbeseeingalivefeedfromthe
SecondaryElectronDetectorintheupperleft
quadrant
-
7/21/2019 A Practical Guide to Running the Scanning Electron
22/61
SampleIntroduction:
Setyour
accelerating
voltage
and
spot
size
Setyouracceleratingvoltageandspotsizetoyour
desiredsetting.(Voltage=20kVandSpotSize=4.0
is
a
good
starting
point)
-
7/21/2019 A Practical Guide to Running the Scanning Electron
23/61
SampleIntroduction:
Setyour
accelerating
voltage
and
spot
size
Setyouracceleratingvoltageandspotsizetoyour
desiredsetting.(Voltage=20kVandSpotSize=4.0
is
a
good
starting
point)
-
7/21/2019 A Practical Guide to Running the Scanning Electron
24/61
SampleIntroduction:
Brightnessand
Contrast
Whenthelivefeedstarts,youmayneedtoadjust
thebrightnessandcontrastforabetterview. You
can
do
this
manually
or
simply
use
the
Auto
ContrastBrightnessButton
-
7/21/2019 A Practical Guide to Running the Scanning Electron
25/61
SampleIntroduction:
Scanningspeed
Iftheimageisupdatingslowly,adjustthescanning
speed. 300nsisanappropriateworkingspeed. The
rabbit
icon
will
set
the
speed
to
300
ns.
-
7/21/2019 A Practical Guide to Running the Scanning Electron
26/61
SampleIntroduction:
Findyour
sample
Youcanmovethesamplebydoubleclickinginthis
window. Thesamplewillberecentredwhereyou
double
click.
Youwillwantthechambercameraunpausedtoensure
yoursampleisnotgoingtoramthedetectorsduring
translation.
-
7/21/2019 A Practical Guide to Running the Scanning Electron
27/61
-
7/21/2019 A Practical Guide to Running the Scanning Electron
28/61
SampleIntroduction:
Increasemagnification
to
~200x
UsethelargeMagnificationknobto
changethemagnificationto
approximately200x
-
7/21/2019 A Practical Guide to Running the Scanning Electron
29/61
SampleIntroduction:
Focusat
@
~200x
UsetheFocusknobstobringyour
sampleintofocus. Mostofthetime,
youwillonlyneedtousetheFine
knob
-
7/21/2019 A Practical Guide to Running the Scanning Electron
30/61
SampleIntroduction:
Increasemagnification
to
~3000x
UsethelargeMagnificationknobto
changethemagnificationto
approximately3000x
-
7/21/2019 A Practical Guide to Running the Scanning Electron
31/61
SampleIntroduction:
Focusat
@
~3000x
UsetheFocusknobstobringyour
sampleintofocus. Mostofthetime,
youwillonlyneedtousetheFine
knob
-
7/21/2019 A Practical Guide to Running the Scanning Electron
32/61
SampleIntroduction:
Linkthe
Zheight
to
the
working
distance
Onceyoursampleisinfocusatabout
3000x,linktheZheighttotheworking
distanceby
clicking
on
the
icon
with
the
redquestionmark.
-
7/21/2019 A Practical Guide to Running the Scanning Electron
33/61
SampleIntroduction:
Raisethe
stage
Makesurethechambercamera(lowerrightquadrant)is
notpaused
-
7/21/2019 A Practical Guide to Running the Scanning Electron
34/61
-
7/21/2019 A Practical Guide to Running the Scanning Electron
35/61
SampleIntroduction:
Raisethe
stage
Raisethe
stage
by
clicking
the
scroll
wheel
on
the
mousewhilehoveringoverthechambercamerafeed
andthendraggingthemousecursorup
-
7/21/2019 A Practical Guide to Running the Scanning Electron
36/61
SampleIntroduction:
Raisethe
stage
Lockthe
Zand
Tstage
position
parameters
by
pushing
thepinbuttonscorrespondingtoZandT
-
7/21/2019 A Practical Guide to Running the Scanning Electron
37/61
DonewithSampleIntroduction
Youarenowreadytoacquiredata!
-
7/21/2019 A Practical Guide to Running the Scanning Electron
38/61
DataAcquisition
-
7/21/2019 A Practical Guide to Running the Scanning Electron
39/61
DataAcquisition:
FullScreen
F5willtoggletheactivatedquadranttofull
screenandback
-
7/21/2019 A Practical Guide to Running the Scanning Electron
40/61
DataAcquisition:
Focusat
High
Magnification
Forthesharpestimage,itisbesttofocus
atahighermagnificationthanyouare
acquiringdata. (ForExample,fortaking
imagesat3000x,focusat10000x)
-
7/21/2019 A Practical Guide to Running the Scanning Electron
41/61
DataAcquisition:
Stigmate
Inadditiontofocus,togetthesharpest
imageyoumayneedtoadjustthe
stigmation usingtheXandYknobs
labeled Stigmator
-
7/21/2019 A Practical Guide to Running the Scanning Electron
42/61
DataAcquisition:
AdjustingContrast
and
Brightness
Adjustbrightnessandcontrastpriortotakinganimageforbestresults
-
7/21/2019 A Practical Guide to Running the Scanning Electron
43/61
DataAcquisition:
Takeapicture
Clickthe
camera
icon
to
acquire
an
image
Thiswillautomaticallyslowscanningrate,pause
acquisitionat
the
end
of
ascan,
and
prompt
you
tosavetheimage.
-
7/21/2019 A Practical Guide to Running the Scanning Electron
44/61
DataAcquisition:
SaveAs
Saveyourimageintoasubdirectoryofthe
SharedData folderonthedesktop
Givethefileauniquenameincludingsampletype,magnification,anddetector
typeusedforexample,PipeCrossSection_6000x_SE.tif
Alwayssaveas.tif,andunlessyouhaveannotatedtheimageorneedcolor,the
8bittype
is
fine.
A i i i
-
7/21/2019 A Practical Guide to Running the Scanning Electron
45/61
DataAcquisition:
BackscatterElectron
Activatingtheupperrightquadrantwilldisplay
datafromtheBackscatterElectronDetector
-
7/21/2019 A Practical Guide to Running the Scanning Electron
46/61
DonewithDataAcquisition
-
7/21/2019 A Practical Guide to Running the Scanning Electron
47/61
SampleExtraction
S l E t ti
-
7/21/2019 A Practical Guide to Running the Scanning Electron
48/61
SampleExtraction:
Turnthe
Beam
Off
Turnoff
the
beam
by
clicking
the
Beam
On
button
S l E t ti
-
7/21/2019 A Practical Guide to Running the Scanning Electron
49/61
SampleExtraction:
Unlockthe
stage
Selectthe
Navigation
Tab
UnlocktheZandTstagepositionparametersbypushing
thepinbuttonscorrespondingtoZandT
Sample Extraction:
-
7/21/2019 A Practical Guide to Running the Scanning Electron
50/61
SampleExtraction:
Lowerthe
stage
Makesure
the
chamber
camerais active
Lowerthestage byclickingthescrollwheelon
themouse
while
hovering
over
the
chamber
camerafeedandthendraggingthemouse
cursordown
Sample Extraction:
-
7/21/2019 A Practical Guide to Running the Scanning Electron
51/61
SampleExtraction:
Zerothe
stage
Makesure
the
chamber
camerais active
Withthestagelowered,pressCTRL0tocentre
thestage
CAUTION:Ifyounoticethatmovingthestage
willcauseaproblem,youcanstopthecurrent
motionbypressingtheEsckeyonthe
keyboard
Sample Extraction:
-
7/21/2019 A Practical Guide to Running the Scanning Electron
52/61
SampleExtraction:
Ventingthe
chamber
Whilethe
chamber
is
venting,thechambericon
willturnorange
-
7/21/2019 A Practical Guide to Running the Scanning Electron
53/61
Sample Extraction:
-
7/21/2019 A Practical Guide to Running the Scanning Electron
54/61
SampleExtraction:
Openthe
chamber
Openthechamberbypullingon
themetalbaronthebottomof
thestage.CAUTION:Donotleanonanypartofthe
instrumentorapplypressureinotherareas
thanthemetalbaronthebottomofthe
stage.
Doing
so
could
harm
the
vacuum.
Sample Extaction:
-
7/21/2019 A Practical Guide to Running the Scanning Electron
55/61
SampleExtaction:
Removeyour
sample
Wearcleangloves!!!
CAUTION:
You
MUST
wear
clean
gloves
when
handling
and
loading
your
sample.
Sample Extraction:
-
7/21/2019 A Practical Guide to Running the Scanning Electron
56/61
SampleExtraction:
Removeyour
sample
Exampleofsampleneedingtobe
removed
CAUTION:Thestageisdelicate,donotforce
anything! Slot4isbrokenbecauseofauser
exerting
too
much
force.
Too
much
pressure
maypermanentlydamagethestage.
Sample Extraction:
-
7/21/2019 A Practical Guide to Running the Scanning Electron
57/61
SampleExtraction:
Loadyour
sample
Thestageshouldbeclearofsamplewhen
youaredonewithyoursession.
Sample Extraction:
-
7/21/2019 A Practical Guide to Running the Scanning Electron
58/61
SampleExtraction:
Closethe
chamber
Watchthechambercamera.
Slowlyand
gently
close
the
stage.
Makesurethatthestageisnotgoing
tohitanyofthedetectors.
Sample Extraction:
-
7/21/2019 A Practical Guide to Running the Scanning Electron
59/61
SampleExtraction:
Pumpthe
chamber
PumpthechamberbypressingthePump
button
Thechambericonwillturnorangeasthe
chamberis
pumped
to
high
vacuum.
Makesurethechamberpumpsallthewayto
greenbeforeyouleave!
-
7/21/2019 A Practical Guide to Running the Scanning Electron
60/61
Log
out
of
SEM
Log
BookLogoutofSEMLogbookbyfinishingtherowyoustarted
previously. Informationrequiredforloggingout:
TimeOut,EngineerTime,SampleRemoved,PumpModeHigh
Vacuum,StagePumpedtoGreen,UserCommentsandComplaints,andAccountNumberifithaschanged
Afteryouhaveenteredallyourdatasavethespreadsheet.
-
7/21/2019 A Practical Guide to Running the Scanning Electron
61/61
Data
Transfer TheSEMcomputersarenotonthenetwork
anddo
not
have
up
to
date
virus
protection!
UsetheprovidedUSBFlashDrivestotransfer
yourdatafromtheleftSEMcomputertothe
networkcomputer
in
the
lab.
Atthenetworkcomputerinthelabyoucan
transfer
your
data
via
the
network
or
on
to
yourpersonaldatastoragedevice.