    APractical

    Guide

    To

    Running

    the

    ScanningElectronMicroscope

    WilliamS.Monroe

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    2/61

    Sections

    I. SampleIntroduction

    II. DataAcquisition

    III. SampleExtraction

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    3/61

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    4/61

    LogintoSEMLogBook(monitoron

    theleft)

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    5/61

    Loginto

    SEM

    Log

    Book

    Loginbywritinganewrow. Informationrequiredforloggingin:

    Date,UserName,Advisor,Sample/Project,TimeIn,

    SampleClean?,

    Stage

    at

    Zero?,

    ChamberPressure,GunPressure,andEmissionCurrent

    WheretofindChamberPressure,

    GunPressure,

    and

    Emission

    Current

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    6/61

    SampleIntroduction:

    Turnon

    chamber

    camera

    1:Selectchambercamera(lowerrightquadrant)

    2:clickpausebuttoninthetoolbarto

    activatechamber

    camera

    feed

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    7/61

    SampleIntroduction:

    Makesure

    that

    stage

    is

    in

    the

    zero

    position

    1:Makesurethestageislowered,

    Movemouse

    inside

    chamber

    camera

    videofeedandclickscrollbuttonon

    mouseanddragdown

    2:Makesurethestageiscentred(afterstageislowered),

    PressCTRL+0(thatisazero,nottheletterO)

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    8/61

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    9/61

    SampleIntroduction:

    Ventthe

    chamber

    Ifstage

    is

    lowered

    and

    centred:

    VentthechamberbypressingtheVentButton

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    10/61

    SampleIntroduction:

    Ventingthe

    chamber

    Duringventing,

    the

    chamber

    portion

    of

    the

    SEMiconwillchangefromgreentoorange.

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    11/61

    SampleIntroduction:

    Thechamber

    is

    vented

    Oncethe

    chamber

    is

    vented,

    the

    chamber

    portionoftheSEMiconwillchangefrom

    orangetodarkgray.

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    12/61

    SampleIntroduction:

    Openthe

    chamber

    Openthe

    chamber

    by

    pulling

    on

    themetalbaronthebottomof

    thestage.CAUTION:Donotleanonanypartofthe

    instrumentorapplypressureinotherareas

    than

    the

    metal

    bar

    on

    the

    bottom

    of

    the

    stage. Doingsocouldharmthevacuum.

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    13/61

    SampleIntroduction:

    Loadyour

    sample

    Wear

    clean

    gloves!!!CAUTION:

    You

    MUST

    wear

    clean

    gloves

    when

    handling

    and

    loading

    your

    sample.

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    14/61

    SampleIntroduction:

    Loadyour

    sample

    Thisisthestage,youwillseenumbered

    holes.

    Placeyour

    sample

    to

    sit

    in

    one

    of

    theseholes.

    CAUTION:Thestageisdelicate,donotforce

    anything!

    Slot

    4

    is

    broken

    because

    of

    a

    user

    exertingtoomuchforce. Toomuchpressure

    maypermanentlydamagethestage.

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    15/61

    SampleIntroduction:

    Loadyour

    sample

    Exampleofaplacedsample

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    16/61

    SampleIntroduction:

    Closethe

    chamber

    Ifyour

    sample

    is

    too

    large,

    STOP! Reopenthestageand

    removeyoursample. Contact

    WilliamMonroeandhewillload

    ashorterstageforyou.

    Watchthechambercamera.

    Slowlyandgentlyclosethestage.

    Makesurethatyoursampleisnot

    goingtohitanyofthedetectors.

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    17/61

    SampleIntroduction:

    Pumpthe

    chamber

    Withyoursamplesafelyloadedandthe

    chamberclosed,

    press

    the

    pump

    button

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    18/61

    SampleIntroduction:

    Pumpingthe

    chamber

    Theiconwillchangetoorangeduringpumping

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    19/61

    SampleIntroduction:

    Turningthe

    beam

    on

    Theiconwillchangetogreenoncepumpingis

    complete

    Turnontheelectronbeam

    usingtheBeamOnbutton

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    20/61

    SampleIntroduction:

    Viewingthe

    microscope

    image

    Selecttheupperleftquadrantbyclickingonit

    Then,activatethe

    SecondaryElectronDetector

    byunpausingthatwindow

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    21/61

    SampleIntroduction:

    SecondaryElectron

    Detector

    Output

    Youshouldnowbeseeingalivefeedfromthe

    SecondaryElectronDetectorintheupperleft

    quadrant

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    22/61

    SampleIntroduction:

    Setyour

    accelerating

    voltage

    and

    spot

    size

    Setyouracceleratingvoltageandspotsizetoyour

    desiredsetting.(Voltage=20kVandSpotSize=4.0

    is

    a

    good

    starting

    point)

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    23/61

    SampleIntroduction:

    Setyour

    accelerating

    voltage

    and

    spot

    size

    Setyouracceleratingvoltageandspotsizetoyour

    desiredsetting.(Voltage=20kVandSpotSize=4.0

    is

    a

    good

    starting

    point)

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    24/61

    SampleIntroduction:

    Brightnessand

    Contrast

    Whenthelivefeedstarts,youmayneedtoadjust

    thebrightnessandcontrastforabetterview. You

    can

    do

    this

    manually

    or

    simply

    use

    the

    Auto

    ContrastBrightnessButton

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    25/61

    SampleIntroduction:

    Scanningspeed

    Iftheimageisupdatingslowly,adjustthescanning

    speed. 300nsisanappropriateworkingspeed. The

    rabbit

    icon

    will

    set

    the

    speed

    to

    300

    ns.

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    26/61

    SampleIntroduction:

    Findyour

    sample

    Youcanmovethesamplebydoubleclickinginthis

    window. Thesamplewillberecentredwhereyou

    double

    click.

    Youwillwantthechambercameraunpausedtoensure

    yoursampleisnotgoingtoramthedetectorsduring

    translation.

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    27/61

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    28/61

    SampleIntroduction:

    Increasemagnification

    to

    ~200x

    UsethelargeMagnificationknobto

    changethemagnificationto

    approximately200x

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    29/61

    SampleIntroduction:

    Focusat

    @

    ~200x

    UsetheFocusknobstobringyour

    sampleintofocus. Mostofthetime,

    youwillonlyneedtousetheFine

    knob

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    30/61

    SampleIntroduction:

    Increasemagnification

    to

    ~3000x

    UsethelargeMagnificationknobto

    changethemagnificationto

    approximately3000x

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    31/61

    SampleIntroduction:

    Focusat

    @

    ~3000x

    UsetheFocusknobstobringyour

    sampleintofocus. Mostofthetime,

    youwillonlyneedtousetheFine

    knob

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    32/61

    SampleIntroduction:

    Linkthe

    Zheight

    to

    the

    working

    distance

    Onceyoursampleisinfocusatabout

    3000x,linktheZheighttotheworking

    distanceby

    clicking

    on

    the

    icon

    with

    the

    redquestionmark.

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    33/61

    SampleIntroduction:

    Raisethe

    stage

    Makesurethechambercamera(lowerrightquadrant)is

    notpaused

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    34/61

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    35/61

    SampleIntroduction:

    Raisethe

    stage

    Raisethe

    stage

    by

    clicking

    the

    scroll

    wheel

    on

    the

    mousewhilehoveringoverthechambercamerafeed

    andthendraggingthemousecursorup

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    36/61

    SampleIntroduction:

    Raisethe

    stage

    Lockthe

    Zand

    Tstage

    position

    parameters

    by

    pushing

    thepinbuttonscorrespondingtoZandT

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    37/61

    DonewithSampleIntroduction

    Youarenowreadytoacquiredata!

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    38/61

    DataAcquisition

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    39/61

    DataAcquisition:

    FullScreen

    F5willtoggletheactivatedquadranttofull

    screenandback

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    40/61

    DataAcquisition:

    Focusat

    High

    Magnification

    Forthesharpestimage,itisbesttofocus

    atahighermagnificationthanyouare

    acquiringdata. (ForExample,fortaking

    imagesat3000x,focusat10000x)

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    41/61

    DataAcquisition:

    Stigmate

    Inadditiontofocus,togetthesharpest

    imageyoumayneedtoadjustthe

    stigmation usingtheXandYknobs

    labeled Stigmator

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    42/61

    DataAcquisition:

    AdjustingContrast

    and

    Brightness

    Adjustbrightnessandcontrastpriortotakinganimageforbestresults

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    43/61

    DataAcquisition:

    Takeapicture

    Clickthe

    camera

    icon

    to

    acquire

    an

    image

    Thiswillautomaticallyslowscanningrate,pause

    acquisitionat

    the

    end

    of

    ascan,

    and

    prompt

    you

    tosavetheimage.

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    44/61

    DataAcquisition:

    SaveAs

    Saveyourimageintoasubdirectoryofthe

    SharedData folderonthedesktop

    Givethefileauniquenameincludingsampletype,magnification,anddetector

    typeusedforexample,PipeCrossSection_6000x_SE.tif

    Alwayssaveas.tif,andunlessyouhaveannotatedtheimageorneedcolor,the

    8bittype

    is

    fine.

    A i i i

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    45/61

    DataAcquisition:

    BackscatterElectron

    Activatingtheupperrightquadrantwilldisplay

    datafromtheBackscatterElectronDetector

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    46/61

    DonewithDataAcquisition

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    47/61

    SampleExtraction

    S l E t ti

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    48/61

    SampleExtraction:

    Turnthe

    Beam

    Off

    Turnoff

    the

    beam

    by

    clicking

    the

    Beam

    On

    button

    S l E t ti

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    49/61

    SampleExtraction:

    Unlockthe

    stage

    Selectthe

    Navigation

    Tab

    UnlocktheZandTstagepositionparametersbypushing

    thepinbuttonscorrespondingtoZandT

    Sample Extraction:

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    50/61

    SampleExtraction:

    Lowerthe

    stage

    Makesure

    the

    chamber

    camerais active

    Lowerthestage byclickingthescrollwheelon

    themouse

    while

    hovering

    over

    the

    chamber

    camerafeedandthendraggingthemouse

    cursordown

    Sample Extraction:

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    51/61

    SampleExtraction:

    Zerothe

    stage

    Makesure

    the

    chamber

    camerais active

    Withthestagelowered,pressCTRL0tocentre

    thestage

    CAUTION:Ifyounoticethatmovingthestage

    willcauseaproblem,youcanstopthecurrent

    motionbypressingtheEsckeyonthe

    keyboard

    Sample Extraction:

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    52/61

    SampleExtraction:

    Ventingthe

    chamber

    Whilethe

    chamber

    is

    venting,thechambericon

    willturnorange

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    53/61

    Sample Extraction:

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    54/61

    SampleExtraction:

    Openthe

    chamber

    Openthechamberbypullingon

    themetalbaronthebottomof

    thestage.CAUTION:Donotleanonanypartofthe

    instrumentorapplypressureinotherareas

    thanthemetalbaronthebottomofthe

    stage.

    Doing

    so

    could

    harm

    the

    vacuum.

    Sample Extaction:

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    55/61

    SampleExtaction:

    Removeyour

    sample

    Wearcleangloves!!!

    CAUTION:

    You

    MUST

    wear

    clean

    gloves

    when

    handling

    and

    loading

    your

    sample.

    Sample Extraction:

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    56/61

    SampleExtraction:

    Removeyour

    sample

    Exampleofsampleneedingtobe

    removed

    CAUTION:Thestageisdelicate,donotforce

    anything! Slot4isbrokenbecauseofauser

    exerting

    too

    much

    force.

    Too

    much

    pressure

    maypermanentlydamagethestage.

    Sample Extraction:

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    57/61

    SampleExtraction:

    Loadyour

    sample

    Thestageshouldbeclearofsamplewhen

    youaredonewithyoursession.

    Sample Extraction:

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    58/61

    SampleExtraction:

    Closethe

    chamber

    Watchthechambercamera.

    Slowlyand

    gently

    close

    the

    stage.

    Makesurethatthestageisnotgoing

    tohitanyofthedetectors.

    Sample Extraction:

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    59/61

    SampleExtraction:

    Pumpthe

    chamber

    PumpthechamberbypressingthePump

    button

    Thechambericonwillturnorangeasthe

    chamberis

    pumped

    to

    high

    vacuum.

    Makesurethechamberpumpsallthewayto

    greenbeforeyouleave!

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    60/61

    Log

    out

    of

    SEM

    Log

    BookLogoutofSEMLogbookbyfinishingtherowyoustarted

    previously. Informationrequiredforloggingout:

    TimeOut,EngineerTime,SampleRemoved,PumpModeHigh

    Vacuum,StagePumpedtoGreen,UserCommentsandComplaints,andAccountNumberifithaschanged

    Afteryouhaveenteredallyourdatasavethespreadsheet.

  • 7/21/2019 A Practical Guide to Running the Scanning Electron

    61/61

    Data

    Transfer TheSEMcomputersarenotonthenetwork

    anddo

    not

    have

    up

    to

    date

    virus

    protection!

    UsetheprovidedUSBFlashDrivestotransfer

    yourdatafromtheleftSEMcomputertothe

    networkcomputer

    in

    the

    lab.

    Atthenetworkcomputerinthelabyoucan

    transfer

    your

    data

    via

    the

    network

    or

    on

    to

    yourpersonaldatastoragedevice.