a. gel electrophoresis
TRANSCRIPT
Agarose Gel Electrophoresis
PRESENTED BY:-
AVINASH SHARMA
PALB-3235
power supply
gel box
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES1
Contents:-
Definition of agarose gel electophoresis
Procedure of agarose gel electrophoresis
Application of gel electrophoresis
Problem in agarose gel electrophoresis
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES2
What is agarose?• it is a polymer of D,L galactose and it is a heteropolysaccharide extracted
from red algae seaweed.
• Low electro endosmosis and good for working.• Agarose has large range of separation but low resolving power.
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES3
Agarose Gel ElectrophoresisGel electrophoresis is a procedure that separates molecules on the basis of size and/or charge and/or shape under the influence of an electrical field. It is based on the principle of “ohms law equation” and”power equation”.
Gel electrophoresis is a widely used technique for the separation of nucleic acids and proteins. Agarose gel electrophoresis is routinely used for the separation of DNA.
e will be using agarose gel electrophoresis to determine the presence and size of PCR products. + -
Power
DNAHO2
+
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES4
Procedures of agarose gel electrophoresis:-
An agarose gel is prepared by combining agarose powder and a buffer solution.
Agarose�
Buffer�
Flask for boiling�
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES5
Casting trayGel combs
Power supply
Gel tank Cover
Electrical leads
Electrophoresis Equipment
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES6
Seal the edges of the casting tray and put in the combs. Place the casting tray on a level surface. None of the gel combs should be touching the surface of the casting tray.
Preparing the Casting Tray
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES8
Agarose Buffer Solution
Combine the agarose powder and buffer solution. Use a flask that is several times larger than the volume of buffer.
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES9
Agarose is insoluble at room temperature (left).The agarose solution is boiled until clear (right).
Gently swirl the solution periodically when heating to allow all the grains of agarose to dissolve. ***Be careful when boiling - the agarose solution may become superheated and may boil violently if it has been heated too long in a microwave oven.
Melting the Agarose
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES10
Allow the agarose solution to cool slightly (~60ºC) and then carefully pour the melted agarose solution into the casting tray. Avoid air bubbles.
Pouring the gel
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES11
Each of the gel combs should be submerged in the melted agarose solution.
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES12
When cooled, the agarose polymerizes, forming a flexible gel. It should appear lighter in color when completely cooled (30-45 minutes). Carefully remove the combs and tape. 26/04/14
PGS 504{0+1} BASIC CONCEPTS IN LABORATORY TECHNIQUES
13
Place the gel in the electrophoresis chamber.
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES14
buffer
Add enough electrophoresis buffer to cover the gel to a depth of at least 1 mm. Make sure each well is filled with buffer.
Cathode(negative)
Anode(positive)
wells
DNA
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES15
Loading Buffer: • Bromophenol Blue (for color) • Glycerol (for weight)
Sample Preparation
Mix the samples of DNA with the sample loading buffer ( tracking dye). This allows the samples to be seen when loading onto the gel, and increases the density of the samples, causing them to sink into the gel wells.
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES16
Loading the Gel
Carefully place the pipette tip over a well and gently expel the sample. The sample should sink into the well. Be careful not to puncture the gel with the pipette tip.
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES17
Place the cover on the electrophoresis chamber, connecting the electrical leads. Connect the electrical leads to the power supply. Be sure the leads are attached correctly - DNA migrates toward the anode (red). When the power is turned on, bubbles should form on the electrodes in the electrophoresis chamber.
Running the Gel
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES18
wells Bromophenol Blue
Cathode(-)
Anode(+)
Gel
After the current is applied, make sure the Gel is running in the correct direction. Bromophenol blue will run in the same direction as the DNA.
DNA(-)
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES19
100 200 300
1,650
1,000
500
850
650
400
12,000 bp
5,000
2,000
DNA Ladder Standard
Inclusion of a DNA ladder (DNAs of know sizes) on the gel makes it easy to determine the sizes of unknown DNAs.
-
+
DNAmigration
bromophenol blue
Note: bromophenol blue migrates at approximately the same rate as a 300 bp DNA molecule
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES20
Staining the Gel
***CAUTION! Ethidium bromide is a powerful mutagen and is moderately toxic. Gloves should be worn at all times.
• Ethidium bromide fluorescent dye and binds to the DNA under UV light allowing the visualization of DNA on a Gel.
• Ethidium bromide can be added to the gel .
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES21
Staining the Gel
• Place the gel in the staining tray containing warm diluted stain.• Allow the gel to stain for 25-30 minutes.• To remove excess stain, allow the gel to destain in water.• Replace water several times for efficient destain.
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES22
Ethidium Bromide requires an ultraviolet light source to visualize
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES23
Factors affecting the DNA in gel elctrophoresis:-
• Agarose concentration:-↑ conc. Of agarose then migration larger fragments of dna and vice versa so for
larger fragments of dna agarose conc.=0.7% and smaller fragments of dna agarose conc.=1.5%.• Voltage:-Voltage increases then migration of larger fragments of dna will be more than
the smaller dna.so voltage should not be more than 5volts/cm.• Electrophoresis buffer:-TAE and TBE buffers are used in the gel electrophoresis because it establish
the pH8.0 and provide ions for the migration of dna and ionic strength are different.• Effects of ethidium bromide:-It is a carcinogenic mutagen and fluorescent dye which are used in the
visualizing the dna when agarose gel are placed in the UV transilluminator then dye are free in the gel then dye binds with dna then it produce orange light for visualizing the dna.
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES24
Application of electrophoresis
• To separate fragments from each other• To determine the sizes of fragments• To determine the presence or amount of DNA• To analyze restriction digestion products prior to southern transfer or
of RNA prior to northern transfer.• Analysis of PCR product.
26/04/14PGS 504{0+1} BASIC CONCEPTS IN LABORATORY
TECHNIQUES25
Proplems in agarose gel electrophoresis:-• Smearing
•voltage too high for large fragments•too much DNA
• Gel melts•voltage too high•ionic strength too low
• Poor resolution•wrong [agarose]•small bands are fuzzy – the gel run may have been too long at too low a voltage, allowing diffusion of the DNA and broadening of the band