a fast optimization approach for treatment planning of ......fluence-map optimization algorithm and...

13
RESEARCH Open Access A fast optimization approach for treatment planning of volumetric modulated arc therapy Hui Yan, Jian-Rong Dai * and Ye-Xiong Li * Abstract Background: Volumetric modulated arc therapy (VMAT) is widely used in clinical practice. It not only significantly reduces treatment time, but also produces high-quality treatment plans. Current optimization approaches heavily rely on stochastic algorithms which are time-consuming and less repeatable. In this study, a novel approach is proposed to provide a high-efficient optimization algorithm for VMAT treatment planning. Methods: A progressive sampling strategy is employed for beam arrangement of VMAT planning. The initial beams with equal-space are added to the plan in a coarse sampling resolution. Fluence-map optimization and leaf- sequencing are performed for these beams. Then, the coefficients of fluence-maps optimization algorithm are adjusted according to the known fluence maps of these beams. In the next round the sampling resolution is doubled and more beams are added. This process continues until the total number of beams arrived. The performance of VMAT optimization algorithm was evaluated using three clinical cases and compared to those of a commercial planning system. Results: The dosimetric quality of VMAT plans is equal to or better than the corresponding IMRT plans for three clinical cases. The maximum dose to critical organs is reduced considerably for VMAT plans comparing to those of IMRT plans, especially in the head and neck case. The total number of segments and monitor units are reduced for VMAT plans. For three clinical cases, VMAT optimization takes < 5 min accomplished using proposed approach and is 34 times less than that of the commercial system. Conclusions: The proposed VMAT optimization algorithm is able to produce high-quality VMAT plans efficiently and consistently. It presents a new way to accelerate current optimization process of VMAT planning. Keywords: Volumetric modulated arc therapy, Fluence-map optimization, Leaf-sequencing Background VMAT is widely used in cancer treatment in radiation on- cology departments due to its high efficiency in treatment delivery [1]. Compared to conventional IMRT, VMAT de- livers radiation dose while MLC leaf, dose rate and gantry move simultaneously [2]. It uses less treatment time and total monitor units (MU) compared to conventional IMRT technique [3]. The predecessor of modern VMAT is intensity modulated arc therapy (IMAT) which was first developed by Cedric Yu in 1995 [4, 5]. It was motivated from the idea of delivering plans with a large number of gantry positions. The fluence map of the beam is pre-calculated and decomposed to several apertures. These apertures is then delivered at a given gantry pos- ition by multiple arcs. IMAT requires more arcs which causes extended treatment time. Direct aperture optimization (DAO) was thus proposed to handle the complexity of VMAT optimization using stochastic ap- proach. The stochastic approach is computationally inten- sive and time-consuming [68]. Later, Otto presented an iterative algorithm for VMAT optimization [9, 10]. This algorithm employs progressive sampling strategy and aperture-based algorithm for VMAT optimization where high-quality dose plan can be achieved by a single arc. * Correspondence: [email protected]; [email protected] Department of Radiation Oncology, National Cancer Center/Cancer Hospital, Chinese Academy of Medical Sciences, Peking Union Medical College, Beijing 100021, China © The Author(s). 2018 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. Yan et al. Radiation Oncology (2018) 13:101 https://doi.org/10.1186/s13014-018-1050-x

Upload: others

Post on 03-Jan-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

RESEARCH Open Access

A fast optimization approach for treatmentplanning of volumetric modulated arctherapyHui Yan, Jian-Rong Dai* and Ye-Xiong Li*

Abstract

Background: Volumetric modulated arc therapy (VMAT) is widely used in clinical practice. It not only significantlyreduces treatment time, but also produces high-quality treatment plans. Current optimization approaches heavilyrely on stochastic algorithms which are time-consuming and less repeatable. In this study, a novel approach isproposed to provide a high-efficient optimization algorithm for VMAT treatment planning.

Methods: A progressive sampling strategy is employed for beam arrangement of VMAT planning. The initial beamswith equal-space are added to the plan in a coarse sampling resolution. Fluence-map optimization and leaf-sequencing are performed for these beams. Then, the coefficients of fluence-maps optimization algorithm areadjusted according to the known fluence maps of these beams. In the next round the sampling resolution isdoubled and more beams are added. This process continues until the total number of beams arrived. Theperformance of VMAT optimization algorithm was evaluated using three clinical cases and compared to those of acommercial planning system.

Results: The dosimetric quality of VMAT plans is equal to or better than the corresponding IMRT plans for threeclinical cases. The maximum dose to critical organs is reduced considerably for VMAT plans comparing to those ofIMRT plans, especially in the head and neck case. The total number of segments and monitor units are reduced forVMAT plans. For three clinical cases, VMAT optimization takes < 5 min accomplished using proposed approach andis 3–4 times less than that of the commercial system.

Conclusions: The proposed VMAT optimization algorithm is able to produce high-quality VMAT plans efficientlyand consistently. It presents a new way to accelerate current optimization process of VMAT planning.

Keywords: Volumetric modulated arc therapy, Fluence-map optimization, Leaf-sequencing

BackgroundVMAT is widely used in cancer treatment in radiation on-cology departments due to its high efficiency in treatmentdelivery [1]. Compared to conventional IMRT, VMAT de-livers radiation dose while MLC leaf, dose rate and gantrymove simultaneously [2]. It uses less treatment time andtotal monitor units (MU) compared to conventionalIMRT technique [3]. The predecessor of modern VMATis intensity modulated arc therapy (IMAT) which was firstdeveloped by Cedric Yu in 1995 [4, 5]. It was motivated

from the idea of delivering plans with a large number ofgantry positions. The fluence map of the beam ispre-calculated and decomposed to several apertures.These apertures is then delivered at a given gantry pos-ition by multiple arcs. IMAT requires more arcs whichcauses extended treatment time. Direct apertureoptimization (DAO) was thus proposed to handle thecomplexity of VMAT optimization using stochastic ap-proach. The stochastic approach is computationally inten-sive and time-consuming [6–8]. Later, Otto presented aniterative algorithm for VMAT optimization [9, 10]. Thisalgorithm employs progressive sampling strategy andaperture-based algorithm for VMAT optimization wherehigh-quality dose plan can be achieved by a single arc.

* Correspondence: [email protected]; [email protected] of Radiation Oncology, National Cancer Center/Cancer Hospital,Chinese Academy of Medical Sciences, Peking Union Medical College, Beijing100021, China

© The Author(s). 2018 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, andreproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link tothe Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver(http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Yan et al. Radiation Oncology (2018) 13:101 https://doi.org/10.1186/s13014-018-1050-x

Page 2: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

This technique is successfully adopted in commercialtreatment planning system and has been applied to a widerange of treatment sites [11].Compared to IMRT, VMAT presents a complex

optimization problem because of the significantly increasednumber of plan parameters such as gantry angle, MLC leafposition, dose rate, etc. [12, 13]. Various approaches wereproposed attempting to solve this problem. Most of themare evolved from approaches originally developed for IMRT[14–22]. Currently, most of commercial treatment planningsystems provide VMAT optimization function and heavilyrely on DAO algorithm. RapidArc (Varian Medical System,Palo Alto, CA, USA) employs a progressive sampling strat-egy and simulated annealing-based DAO algorithm forVMAT planning [9]. Due to the nature of stochastic ap-proach, the optimization process is time-consuming andthe optimization result is seldom repeatable. As reportedthe maximal DVH variation could be up to 2% for OARsusing RapidArc [10]. SmartArc (Philips Healthcare, Inc.,Thornton, CO, USA) utilizes an IMRT plan (consisting ofequal-space fields) to initialize the arc plan. The fluencemaps of IMRT plan are then segmented and redistributedto their neighboring angles. A local gradient-basedoptimization approach is used to fine-tune these aperturesto meet the mechanical constraints for actual delivery [16,17]. In this approach, The fluence maps of all beams are ac-tually derived from few static beams and not fully deter-mined by fluence map optimization (FMO) algorithm.The purpose of this study is to develop a high-efficient

optimization approach for VMAT planning. The remain-der of this paper is organized as follow. The progressivesampling strategy is first introduced and the optimizationprocess is described. Then, two important components,fluence-map optimization algorithm and leaf-sequencingalgorithm, are explained in brief. These algorithms weredeveloped on an in-house developed treatment planningsystem. Three typical clinical cases (head and neck, lung,prostate) were evaluated on both in-house developed andcommercial treatment planning systems. The dosimetricquality, delivery efficiency, and running time of VMATplans were then assessed. Finally, the advantage and disad-vantage of this approach are discussed.

MethodsProgressive sampling strategyMost of VMAT optimization algorithms model Linacsource as a series of static gantry positions. The MLCpositions and MU setting is then determined at eachgantry position. Provided the complexity of VMAToptimization, to limit the scale of problem a progressivesampling strategy is employed in this study. It starts witha coarse sampling of gantry positions, then move to finesampling of gantry positions. An example is demon-strated as shown in Fig. 1. At the beginning of VMAT

optimization a coarse sampling of the gantry positions isused to model the gantry rotation range. The initial 5beams with equal-space are chosen for the firstoptimization stage. The MLC positions and MUs for theinitial 5 beams are achieved by the fluence-mapoptimization and leaf-sequencing algorithms. Next, thesampling resolution is doubled. And another 5 beamsare added to the plan and optimized. The sampling reso-lution is continuously increased and more beams areadded to the plan until the total number of beams is ar-rived. As shown in Fig. 1, the number of new beams in 5optimization stages is 5, 10, 20, 40, and 60, while thecorresponding beam spacing is 72°, 36°, 18°, 9°, and 6°.A VMAT optimization approach was developed based on

the existing fluence-map optimization and leaf-sequencingalgorithms provided by an in-house developed treatmentplanning system [23]. The flowchart of VMAT planningproposed in this study is presented in Fig. 2. First, the basicplan parameters such as couch angle and arc range are setby planner. An initial plan consisting of few beams withequal-space is created and their fluence maps are generatedby FMO algorithm. These resulting fluence maps are thenprocessed by the leaf sequencing algorithm, and the optimalMLC positions and MUs are determined. Next, the coeffi-cients of FMO algorithm are adjusted based on plan object-ive and known fluence maps of these beams. In the nextround, the sampling resolution is increased and morebeams are added to the plan. The optimization processcontinues until the maximal number of beams is arrived.

Fluence-map optimizationA gradient-based optimization approach, fast monotonicdescent (FMD) algorithm is employed for VMAToptimization. Due to the nature of gradient-based algo-rithm, the global minimum of objective function can befound quickly. The goal is to find minimum of objectivefunction subject to non-negative fluences for all pencilbeams. A non-synchronous updating scheme is usedwhich allows only one pencil beam adjusted at a time.The detail of classic FMD algorithm is described in ref-erences [23–25]. To accommodate the progressive sam-pling strategy, FMD algorithm was modified to accountfor the varied number of beams in multiple optimizationstages. The formulation of the modified FMD algorithmis described in Appendix 1. In brief, new beams areadded to the plan in each optimization stage and theirfluence maps are optimized with the fluence maps of oldbeams known. Here, xnew denotes the fluence map ofnew beam which is unsolved in current stage and xold

denotes the fluence map of old beam which is solved inprevious stage. For a given voxel vijk, its dose is calcu-lated by the sum of dose contributed from xnew and xold.For solving xnew the coefficients of FMD algorithm is ad-justed based the planning dose of target and the dose

Yan et al. Radiation Oncology (2018) 13:101 Page 2 of 13

Page 3: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

contribution from xold. As fluence map of xnew known, itis processed by a leaf sequencing algorithm and a uni-form fluence map for a single aperture is obtained. Theuniform fluence map replaces the original fluence mapof xnew and is used as the final value of xnew.

Leaf sequencingSeveral leaf sequencing algorithms were developed inthe past years such as powers of 2, linear programming,and graphics searching [26–29]. These algorithms

decompose fluence map into several apertures with uni-form fluences. Given the dynamic feature of beams inVMAT, the leaf sequencing algorithm has to deal withmore mechanical constraints related to gantry speed, leafspeed, and dose rate than those of static beams. Mixedinteger linear programming (MILP) was successfully ap-plied in many industrial applications due to its capabilityin handling large-scale optimization problems [30, 31].In this study, a MILP model was developed using a com-mercial optimization software package (IBM ILOG

Fig. 1 The illustration of progressive sampling scheme. a The first beam set with 5 control points (beam spacing 72°). b The second beam setwith additional 5 control points (beam spacing 36°). c The third beam set with additional 10 control points (beam spacing 18°). d The fourthbeam set with additional 20 control points (beam spacing 9°). e The fifth beam set with additional 20 control points (beam spacing 6°)

Fig. 2 The flowchart of the proposed optimization algorithm for VMAT treatment planning

Yan et al. Radiation Oncology (2018) 13:101 Page 3 of 13

Page 4: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

CPLEX). It assumes that the fluence map of a beam at agiven control point is a uniform fluence map for a singleaperture. The aperture of the new beam is determinedsubjected to the apertures of the two old beams closet toit (one is at front side and one is at end side). As the ori-ginal flence map of xnew is non-uniform, it is necessaryto convert it to a uniform fluence by a leaf-sequencingalgorithm with certain constraints. These constraints aremostly related to mechanical limitation of Linac andMLC and described in Appendix 2. Due to the nature ofprogressive sampling strategy, the beams solved in theearlier stages have larger apertures and weights than thebeams solved in the latter stages. This is because thenumber of beams is less in the earlier stages and morein the latter stages. The MILP model is described inAppendix 2. The goal of this model is to search for anoptimal solution, uniform fluence map, from anon-uniform fluence map under certain constraints. TheMILP problem is solved using branch-and-bound tech-nique provided by CPLEX software. As an example offluence map consisting of 30×30 bixels, the total numberof constraints is ~ 1800 and the total number of inde-pendent variable is ~ 2000. The processing time is ap-proximately 0.5 s per beam.

Plan evaluationThree clinical cases were selected for evaluation in-cluding head and neck, prostate, and lung. The pa-tient CT images and contours were exported fromPinnacle (Philips Healthcare, Inc., Thornton, CO,USA). Then they are segmented into regions of inter-est (ROIs) for plan optimization using an in-housedeveloped treatment planning system. In this study,the simulated leaf width is 0.5 cm and the leaf step is0.5 cm, which resulted in beamlet size of 0.5×0.5 cm.The range of leaf speed is 0.5–1 cm per degree be-cause the leaf speed is 3–6 cm per second and gantryspeed is 6 degree per second. The prescription dosefor PTV and dose constraints for OARs are the samefor both IMRT and VMAT plans for each case. A sin-gle dynamic arc plan consisting of 180 beams and 2°spacing was created and fluence maps were optimizedby the proposed VMAT optimization algorithm onthe in-house developed system. The numbers ofbeams in five optimization stages are 15, 30, 60, 120,and 180, while the corresponding beam spacing is24°, 12°, 6°, 3°, and 2°. A corresponding IMRT planconsisting of 9 equal-space fields was created and flu-ence maps were optimized by FMD algorithm on thein-house developed system. The dosimetric quality ofplans is quantified by conformity index (CI), homo-geneity index (HI), quality score (QS) [32], andweighted root mean-square error [23], which are de-fined as below.

CI = VTV/V95%, where VTV is the target volume andV95% is the volume corresponding to 95% dose prescrip-tion. If V95% = VTV, CI =1 which is perfect.HI = D5%/D95%, where D5% and D95% are doses received

at least 5 and 95% target volume respectively. if D5% =D95%, HI =1 which is perfect.Quality score (QS) quantitatively measures the differ-

ence between the dose plan and the dose constraint of

all anatomical structures. QS ¼ Xj

j ðMj−C jÞC j

j if object-

ive is violated by plan dose, where Cj is the j-thdose-volume constraint, Mj is the corresponding plandose. If no objective is violated, QS = 0 which is perfect.Weighted root mean-square error (WE) measures the

difference between the plan dose and prescribed dose ofall voxels and represents the value of objective function.

WE ¼ ð

Xi; j;k∈VTV

wijkðPijk−DijkÞ2 þX

i; j;k∈VCO

wijkðPijk−DijkÞ2 þX

i; j;k∈VNT

wijkðPijk−DijkÞ2

N Þ

12

,where Pijk, Dijk, and Wijk are defined in Eq. 1, and N isthe total number of voxels belonging to TV, CO andNT. For a plan which Pijk =Dijk for all voxels, WE = 0which is perfect.The dose distribution of plans is evaluated based on

cross-sectional dose distribution and dose-volume histo-gram. The total number of segments and monitor unit forboth plans are recorded. Additionally, estimated arc deliv-ery times and the computation time of plan optimizationare recorded. For demonstration purpose, few importantROIs are used in plan optimization. For head and neckcase, they are PTV, spinal cord, left and right parotids andmandible. For prostate case, they are PTV, left and rightfemur heads, bladder and rectum. For lung case, they arePTV, left and right lungs, spinal cord and heart. All testsare performed on a DELL Optiplex N9010 computer,equipped with Intel(R) i7–3770 CPU and 72GB RAM.Corresponding to the VMAT plans made on in-house de-

veloped system, three VMAT plans made on Pinnacle plan-ning system were assessed. They were made by experiencedplanners and approved for clinical treatment. These plansare VMAT plans consisting of 2 full arcs with 4° spacingand 6 MV beams computed with convolution superpositionalgorithm. The dosimetric and delivery statistics of theseclinically approved plans are compared to those of IMRTand VMAT plans made on the in-house developed system.Note that the optimization algorithm, the dose calculationengine, the dose-volume constraints and many others ofclinically approved plans are different from those of plansmade on the in-house developed system. The metric, WS, isnot calculated for the clinically approved VMAT plans be-cause the weight specification in Pinnacle system is differ-ent. For distinguishing clinically approved VMAT plansfrom the plans made on the in-house developed system,VMAT plans made on pinnacle system are represented byVMAT_P.

Yan et al. Radiation Oncology (2018) 13:101 Page 4 of 13

Page 5: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

ResultsThe fluence map and plan dose distribution achieved byprogressive sampling strategy is demonstrated in Fig. 3.For each plan, the subplots at the top are the segmentsof all beams while the subplots at the bottom are thedose distribution in axial view. The grey level of segmentindicates the magnitude of its weight. White representsthe highest value (255) and black represents the lowestvalue (0). The corresponding beam arrangements areshown in Fig. 1. In the earlier stages, the values of flu-ence maps of new beams are larger as there are fewerbeams. As more beams included, the values of fluencemap of new beams are smaller. The plan dose qualityimproves quickly in the earlier stages (as shown in Fig.3b and c), and gradually saturates in latter stages (asshown in Fig. 3d and e.For the head and neck case, the objective is to main-

tain uniform dose (60 Gy) to the PTV, while minimizethe dose to the parotid gland (V30Gy < 50%) andmaximize the sparing of the brainstem (Dmax< 54 Gy),the larynx (Dmax<40Gy), and the mandible (Dmax<60 Gy) as much as possible. Dose distributions for IMRTand VMAT plans are shown in Fig. 4. DVHs for IMRTand VMAT plans are compared as shown in Fig. 5. TheVMAT plan is better than the corresponding IMRTplan. The coverage of PTV is similar for both VMATand IMRT plans. The OAR sparing from VMAT plan isbetter for brainstem and mandible, but is similar for lar-ynx and parotids (left and right). The max dose to

brainstem is reduced by 10 Gy. Dose distribution ofVMAT plan is more uniform with fewer hot and coldspots. The QS, WE, CI, and HI for both plans are shownin Table 1. The WE is reduced by 10% for VMAT plan,while QS, CI and HI are similar for both plans. The totalnumber of segments and monitor units for both plansare shown in Table 1. The number of segment is similarfor both plans. The total MU for VMAT plan is reducedby 40%. The computation time is 40 s for IMRT planand 312 s for VMAT plan as shown in Table 1. The esti-mated delivery time is between 96 and 192 s based on aVarian Trilogy machine with variable dose rate (300–600MUs per minute). VMAT_P plan show better plan qual-ity and fewer MUs than VMAT plan, but the number ofsegments and optimization time are higher.For the prostate case, the objective is to maintain uni-

form dose (76 Gy) to the PTV while keep the rectal doseat V50 Gy < 50% and maximize sparing of bladder(V50 Gy < 50%) and femoral head (V50 Gy < 5%). Dosedistributions for IMRT and VMAT plans are shown inFig. 6. DVHs for IMRT and VMAT plans are compared asshown in Fig. 7. The VMAT plan is comparable to the cor-responding IMRT plan. The coverage of PTV is similarbetween VMAT and IMRT plans. For VMAT plan, OARsparing is better for rectum, femoral heads (left and right)while similar for bladder. Dose distribution of VMAT planis more uniform and has less high dose region in normaltissue. The QS, WE, CI, and HI for both plans are shownin Table 1. The WE is reduced by 12% for VMAT plan,

Fig. 3 The segments and dose distributions of VMAT plans for five beam sets with the increasing numbers of beams. a VMAT plan with 5equal-space beams (72° spacing). b VMAT plan with 10 equal-space beams (36° spacing). c VMAT plan with 20 equal-space beams (36° spacing).d VMAT plan with 40 equal-space beams (9° spacing). e VMAT plan with 60 equal-space beams (6° spacing)

Yan et al. Radiation Oncology (2018) 13:101 Page 5 of 13

Page 6: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

while QS, CI, and HI are similar for both plans. The totalnumber of segments and monitor units for both plans areshown in Table 1. The number of segment is similar forboth plans. The total MU for VMAT plan is reduced by16%. The computation time is 36 s for IMRT plan and294 s for VMAT plan as shown in Table 1. The estimateddelivery time is between 118 s and 236 s based on a VarianTrilogy machine with variable dose rate (300–600 MUsper minute). VMAT_P plan show better plan quality andfewer MUs than VMAT plan, but the number of segmentsand optimization time are higher.For the lung case, the objective is to maintain uniform

dose (60 Gy) to the PTV while keep the lung dose atV20 Gy < 25% and maximize the sparing of cord (Dmax<

45 Gy), heart (V30Gy < 40%, V40Gy < 30%) and esophagus(V50Gy < 50%). Dose distributions for IMRT and VMATplans are shown in Fig. 8. DVHs for IMRT and VMATplans are compared as shown in Fig. 9. The dose coverageof PTV is similar in both plans. For VMAT plan, dose toheart, esophagus, and cord are slightly higher than those ofIMRT plan, while dose to lung is similar. The Dose distri-bution of the VMAT plan is more uniform and focused onPTV, however the dose distribution of IMRT plan is spreadalong anterior-posterior direction. The QS, WE, CI, and HIfor both plans are shown in Table 1. The WE, QS, CI, andHI are similar for both plans. The total number of segmentsand monitor units for both plans are shown in Table 1. Thenumber of segment for VMAT plan is reduced by 16%.

Fig. 5 The DVH of head-and-neck case in IMRT plan and VMAT plan

Fig. 4 The transaxial plan dose distribution of head-and-neck case in (a) IMRT plan and (b) VMAT plan

Yan et al. Radiation Oncology (2018) 13:101 Page 6 of 13

Page 7: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

The total MU is similar for both plans. The computationtime is 32 s for IMRT plan and 216 s for VMAT plan asshown in Table 1. The estimated VMAT delivery times arebetween 116 s and 232 s based on a general linear acceler-ator with variable dose rate (300–600 MUs per minute).VMAT_P plan show better plan quality and few MUs thanVMAT plan, but the optimization time is higher.

DiscussionsThe results of clinical case study show that plan qualityof VMAT plans is similar to or better than that of trad-itional IMRT plans. The VMAT plans, in many cases,are able to achieve better OAR sparing compared toIMRT plans. VMAT plan has the potential to tailor highvolume low dose regions because of its flexibility tospread out the dose at many beam angles. This characterreduces the need of specific structures defined to removehot spots, thus reduces overall planning effort. Plansconsisting of more than 2 arcs are not reported in thiswork, but it is a viable option in clinical practice. Wealso tested optimizing all beams in multiple arcs whichhas slightly improved dosimetric quality. These data arenot included in this paper. The dose quality of clinicallyapproved plans made on pinnacle planning system isbetter. This may be caused by the differences of

optimization algorithm, dose calculation engine,dose-volume constraints between pinnacle system andin-house developed system. However, the optimizationand delivery time of VMAT plans achieved by our ap-proach is less. It is promising to implement this ap-proach in commercial planning system to accelerate thecurrent process of VMAT optimization.The selection of beam orientations in initial stages of

optimization would affect the final dose considerably.However it could be minimized with increasing numberof beams. It was reported that optimizing beam orienta-tions is most valuable for a small numbers of beams (≤5)and the gain diminishes rapidly for higher numbers ofbeams (≥15) [33]. In several pioneering studies of VMAToptimization it showed 10–15 beams with equal-spacewould be a better choice in initial stage of optimization[9, 10, 17]. We also tested our algorithm with the differ-ent initial beam orientations in the initial stage. It wasfound that for a higher number of beams (≥15) in theinitial stage the final value of objective function is lessaffected. In the proposed algorithm the number ofbeams in the initial stage is set to 15, while this value is10 in RapidArc of Eclipse (Varian Medical System, PaloAlto, CA, USA) and 15 in SmartArc of Pinnacle (PhilipsHealthcare, Inc., Thornton, CO, USA).

Table 1 Comparison of performance between VMAT plan and IMRT plan

Treatmentsite

Plantype

Plan performance

QS WE CI HI Number of segments MU Time (s)

HN IMRT 0.92 79.91 1.83 1.26 129 1775 40

VMAT 0.84 71.63 1.82 1.23 108 960 312

Prostate VMAT_P 0.13 N/A 1.54 1.02 364 810 1080

IMRT 0.32 63.44 1.43 1.23 112 1418 36

VMAT 0.31 55.43 1.41 1.22 116 1180 294

VMAT_P 0.13 N/A 1.21 1.03 182 1076 1210

IMRT 0.13 15.9 1.83 1.14 166 1178 32

Lung VMAT 0.13 15.8 1.84 1.12 138 1162 216

VMAT_P 0.11 N/A 1.63 1.03 108 626 760

Fig. 6 The transaxial plan dose distribution of prostate case in (a) IMRT plan and (b) VMAT plan

Yan et al. Radiation Oncology (2018) 13:101 Page 7 of 13

Page 8: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

Plan quality can be further improved by increasing thenumber of beams. However, as more beams included, thecomputation time for plan optimization and dose calcula-tion could be increased substantially. We speculate that onlyin certain cases, e.g., with complex target shape requiringsubstantial leaf motion, it would be beneficial to increasegantry position sampling. It is observed that the number ofsegments for VMAT plan is less than or similar to that ofIMRT plan in three cases. The total MU for VMAT plan isalso less than or similar to that of IMRT plan in three cases.Overall, the delivery time for VMAT plan is significantly re-duced compared to IMRT plan. Plan optimization timesvary case by case but all within 5 min. Compared with theselected commercial treatment planning systems, the run-ning time of plan optimization is reduced by 3–4 times.Also, due to the nature of gradient-based optimization algo-rithm, the optimization solution is repeatable as long as theinitial setting of plan parameters is the same.

The current objective function is quadratic functionwhich can be solved efficiently by gradient decent method.However, for real clinical use it is desired to incorporatemore complex objectives and constraints such asdose-volume histogram (DVH) and general equivalent uni-form dose (gEUD). For DVH constraints, the current algo-rithm is applicable with minor modification. Those voxelswhose doses exceed the given threshold will be identifiedwhen comparing actual DVH to expected DVH. Theweights for those voxels will be adjusted automatically inorder to minimize the DVH difference. For gEUD con-straints, the form of objective function will be changed asdescribed by Wu [34] and can be solved by gradient des-cent method which is similar to FMD algorithm employedin this study. The workflow shown in Fig. 2 is also feasiblefor gEUD-based optimization algorithm. It is also possibleto use gEUD-based objective to constrain mean dose fornormal tissue by setting parameter α to small and positive

Fig. 7 The DVH of prostate case in IMRT plan and VMAT plan

Fig. 8 The transaxial plan dose distribution of lung case in (a) IMRT plan and (b) VMAT plan

Yan et al. Radiation Oncology (2018) 13:101 Page 8 of 13

Page 9: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

value. For more challenging clinical cases with complexgeometry and spatial distribution of anatomical structures,the overlapped region between PTV and OARs should behandled respectively and supplementary structures wouldbe used.For the first time the new VMAT plan optimization ap-

proach was implemented on an in-house developed plan-ning platform. The progressive sampling strategy iscombined with gradient-descent-based FMO algorithmfor a high-performance VMAT planning system. Thiswork demonstrates a new way to transform the existingoptimization algorithms originally designed for IMRT tothe new algorithm for VMAT planning. Currently, the dy-namic arcs are planned with a single 360° with 2° anglespacing, varying dose rate, and constant gantry speed. Sen-sitivity to parameters such collimator angle, couch angle,arc length, and delivery time are not explored methodic-ally. However these parameters may provide plan qualityimprovements for some cases. In addition, there may be aneed for multiple arcs for cases not included in this study.Determining optimal use of these parameters and the po-tential automation of the parameter settings are subjectsto ongoing investigation. It is also necessary to performphantom verification in order to implement this algorithmfor clinical application.

ConclusionA new approach was developed which is based on fastgradient descent algorithm and mixed integer program-ming technique to provide a high-performance VMATplanning. Results from clinical case studies demon-strated that plan quality of VMAT plans is similar to or

better than that of IMRT plans. The optimization timeand the number of segments are reduced considerably.This work demonstrates a way to transform the existingoptimization algorithms originally designed for IMRT tothe new algorithm designed for VMAT planning.

Appendix 1The objective function of classic FMD is defined asbelow.

min f xð Þ ¼Xi

Xj

Xk

Wijk Dijk−Pijk� �2 ð1Þ

where Dijk ¼XNn¼1

An;ijkxn

N is the total number of beamlet and xn is the fluenceof beamlet n. Dijk denotes the dose absorbed by voxelvijk. An,ijk is the dose deposition coefficient defining thedose contribution per unit fluence of beamlet n to voxel(i, j, k). Pijk is the constraint doses for target volume(TV), critical organs (CO), and normal tissue (NT). Wijk

is the weighting factors assigned for TV, CO, and NT.Pijk and Wijk are predefined according to clinical require-ment. The objective is to find minimum of f(x) subjectto xn > 0 for all beamlet n∈N. A non-synchronous updat-ing scheme as shown below is used which allows onlyone beamlet adjusted at a time.

xm l þ 1ð Þ ¼ xm lð Þ þ Δxm;m∈N ð2Þ

Here

Fig. 9 The DVH of lung case in IMRT plan and VMAT plan

Yan et al. Radiation Oncology (2018) 13:101 Page 9 of 13

Page 10: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

Δxm ¼ PNn¼1 BmnxnðlÞ þ Cm ,

Bmn ¼−Xi

Xj

Xk

W ijkAm;ijkAn;ijk ;

Xi

Xj

Xk

W ijkA2m;ijk

Cm ¼

Xi

Xj

Xk

WijkAm;ijkPijk

Xi

Xj

Xk

WijkA2m;ijk

.

l is the iteration number and m is the index of beamletadjusted in l–th iteration. Only one beamlet adjusted at oneiteration. If all beamlets are adjusted one time, there are Niterations and it is called one cycle. Usually it takes few cy-cles (~ 10) to reach a satisfactory solution for one beam set.To accommodate the progressive sampling strategy of

VMAT optimization, FMD algorithm was modified toaccount for the increasing number of beams. Theoptimization process starts from the initial beam setconsisting of few beams (usually 5), and then the fluencemaps are achieved by FMO and processed by leaf se-quencing. The beam set keeps expanding and newbeams are added. The beams from previous beam set arecalled old beams and the beams newly added are callednew beams. In a given beam set, the fluence maps of oldbeams are fixed and the fluence maps of new beams areoptimized by FMO for several cycles. The beam set willcontinuously grow until the maximal number of beamsreaches. For each beam set, there are many beamletscorresponding to the given number of beams. Assumingthere are NT beamlets in T-th beam set and NT =NT

new

+NTold where NT

new and NTold are the numbers of

beamlets for new and old beams. xrnew denotes the flu-

ence of r-th beamlet of new beam while xsold denotes the

fluence of s-th beamlet of old beam. For a given voxelvijk, its dose is calculated based on the fluences of allxr

new and xsold.

Dijk ¼XNT

new

r¼1

Ar;ijkxnewr þ

XNTold

s¼1

As;ijkxolds ð3Þ

Here r and s are the index of beamlets of new and oldbeams. Accordingly, the objective function (1) can be re-written as below.

f xð Þ ¼Xi

Xj

Xk

wijk

Pijk−XNT

new

r¼1

Ar;ijkxnewr þ

XNTold

s¼1

As;ijkxolds

0@

1A

24

352

ð4ÞAlternatively, it can be expressed as.

f xð Þ ¼Xi

Xj

Xk

wijk

Pijk−XNT

new

r¼1

Ar;ijkxnewr

0@

1A−

XNTold

s¼1

As;ijkxolds

24

352

ð5Þ

Then the expression (5) can be extended as.

f xð Þ ¼Xi

Xj

Xk

½wijk Pijk−XNT

new

r¼1

Ar;ijkxnewr

0@

1A

2

−2wijk Pijk−XNT

new

r¼1

Ar;ijkxnewr

0@

1A

XNTold

s¼1

As;ijkxolds þ wijk

XNTold

s¼1

As;ijkxolds

0@

1A

2

ð6Þ

f xð Þ ¼Xi

Xj

Xk

wijk Pijk−XNT

new

r¼1

Ar;ijkxnewr

0@

1A

2

−Xi

Xj

Xk

2wijk Pijk−XNT

new

r¼1

Ar;ijkxnewr

0@

1AXNT

old

s¼1

As;ijkxolds

24

35

þXi

Xj

Xk

wijk

XNTold

s¼1

As;ijkxolds

0@

1A

2

ð7Þ

To minimize the objective function in (2) with respectto the fluence of m-th beamlet, its gradient at l + 1 iter-ation should be zero.

∂ f x l þ 1ð Þ½ �∂xm

¼ 0 ð8Þ

The gradient of objective function at l + 1 iteration isexpressed as.

∂ f x l þ 1ð Þ½ �∂xm

¼Xi

Xj

Xk

2wijk Pijk−XNT

new

r¼1

Ar;ijkxnewr l þ 1ð Þ

0@

1A

−Am;ijk� �

−Xi

Xj

Xk

2wijk

XNTold

s¼1

As;ijkxolds −Am;ijk� �

24

35

ð9Þ

For m∈, the update formula (2) is.

xnewm l þ 1ð Þ ¼ xnewm lð Þ þ Δxnewm ;m∈NnewT ð10Þ

Put (8) and (9) together we have

Yan et al. Radiation Oncology (2018) 13:101 Page 10 of 13

Page 11: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

Xi

Xj

Xk

2wijk Pijk−XNT

new

r¼1

Ar;ijkxnewr l þ 1ð Þ

0@

1A

−Am;ijk� �

−Xi

Xj

Xk

2wijk

XNTold

s¼1

As;ijkxolds −Am;ijk� �

24

35 ¼ 0:

ð11Þ

Replace xnewr ðl þ 1Þ in (11) with [xnewr ðlÞ+Δx] in (10),we have

Xi

Xj

Xk

wijkAm;ijk Pijk−XNT

new

r¼1

Ar;ijkxnewr lð Þ þ Am;ijkΔxm

24

35

8<:

9=;

¼Xi

Xj

Xk

wijkAm;ijk

XNTold

s¼1

As;ijkxolds

ð12Þ

Equation (12) can be expressed as

Xi

Xj

Xk

wijkAm;ijk Pijk� �

−Xi

Xj

Xk

wijkAm;ijk

XNTnew

r¼1

Ar;ijkxnewr lð Þ� �

−Xi

Xj

Xk

wijkAm;ijkAm;ijkΔxm

¼Xi

Xj

Xk

wijkAm;ijk

XNTold

s¼1

As;ijkxolds :

ð13Þ

With proper arrangement of both sides of Eq. (13),it is

Xi

Xj

Xk

wijkPijkAm;ijk−Xi

Xj

Xk

wijk

XNTnew

r¼1

Am;ijkAr;ijkxnewr lð Þ−

Xi

Xj

Xk

wijk

XNTold

s¼1

Am;ijkAs;ijkxolds ¼

Xi

Xj

Xk

wijkA2m;ijkΔxm:

ð14Þ

Then Δxmis calculated as

Δxm ¼

Xi

Xj

Xk

wijkPijkAm;ijk−Xi

Xj

Xk

wijk

XNTnew

r¼1

Am;ijkAr;ijkxnewr lð Þ−

Xi

Xj

Xk

wijk

XNTold

s¼1

Am;ijkAs;ijkxolds

Xi

Xj

Xk

wijkA2m;ijk

:

ð15Þ

Bnewm ¼ −

Xi

Xj

Xk

wijk

XNTnew

n1¼1

Ar;ijkxr lð ÞAm;ijk

Xi

Xj

Xk

wijkA2m;ijk

¼ −XNT

new

r¼1

xnewr lð Þ

Xi

Xj

Xk

wijkAm;ijkAr;ijk

Xi

Xj

Xk

wijkA2m;ijk

:

ð15aÞ

(15a) can also be expressed as Bnewm ¼ PNT

newr¼1 xnewr ðlÞ

Bnewmr ,where Bnew

mr ¼ −

Xi

Xj

Xk

wijkAm;ijkAr;ijk

Xi

Xj

Xk

wijkA2m;ijk

; r∈NTnew.

Boldm ¼

−Xi

Xj

Xk

wijk

XNTold

s¼1

As;ijkxolds Am;ijk

Xi

Xj

Xk

wijkA2m;ijk

¼XNT

old

s¼1

xolds

−Xi

Xj

Xk

wijkAm;ijkAs;ijk

Xi

Xj

Xk

wijkA2m;ijk

ð15bÞ

(15b) can also be expressed as Boldm ¼ PNT

olds¼1 xolds Bold

ms ,

where Boldms ¼

−Xi

Xj

Xk

wijkAm;ijkAs;ijk

Xi

Xj

Xk

wijkA2m;ijk

; s∈NTold .

Cm ¼

Xi

Xj

Xk

wijkAm;ijkPijk

Xi

Xj

Xk

wijkA2m;ijk

ð15cÞ

Finally, the updating formula is below.

xm l þ 1ð Þ ¼ xm lð Þ þXNT

new

r¼1

Bnewmr x

newr lð Þ

þXNT

old

s¼1

Boldms x

olds þ Cm; m∈NT

new ð16Þ

Yan et al. Radiation Oncology (2018) 13:101 Page 11 of 13

Page 12: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

Note that xolds is solved in 1-th to T-1-th beam set andconstant in T-th beam set, while xnewr is variable to besolved in T-th beam set. Cmis calculated based on wij-

k,Am, ijkand Pijk which can be pre-computed prior to theoptimization of 1-th beam set. Bold

m is calculated basedon constantswijk,Am, ijk,Pijk and xolds . Bold

m can bepre-computed prior to the optimization of T-th beamset. Bnew

m is calculated based on constantswijk,Am, ijk,Pijkand variable xnewr . Bold

m can’t be pre-computed and shouldbe calculated online.

Appendix 2The objective function of the MILP model is defined asbelow.

minXNi¼1

XMj¼1

Icij−Acij

� ð1Þ

Here Icij is the original fluence of bixel (beam pixel)

after FMO, and Acij is the unknown fluence of bixel for

the single aperture. N and M are the dimensions of flu-ence map indexed with ij. For achieving the best uni-form fluence mapAcin approximating non-uniformfluence mapIc, the following constraints are required.

lcij þ rcij ¼ 1−bcij ð2Þ

lcijþ1≤ lcij ð3Þ

rcijþ1≥rcij ð4Þ

Acij ¼ Acbcij ð5Þ

Acij≤ I

cij ð6Þ

XMj¼1

lcij−XMj¼1

lpij

≤vmax

sϕΔϕ ð7Þ

XMj¼1

lcij−XMj¼1

l fij

≤vmax

sϕΔϕ ð8Þ

XMj¼1

rcij−XMj¼1

rpij

≤vmax

sϕΔϕ ð9Þ

XMj¼1

rcij−XMj¼1

r fij

≤vmax

sϕΔϕ ð10Þ

The superscripts c, p, and f represent current beam,previous beam and following beam.The constants in the above constraints are defined as:dmax: Maximal dose rate.vmax: Maximal leaf speed.sϕ: Gantry angular speed.

Δϕ: Angular spacing between two beams.Icij : Contains the original fluence of bixel ij resulted by

FMO.Ac: Maximal fluence of current beam and is less than

vmaxsϕ dmax.The variables in the above constraints are defined as:lcij : Binary variable indicates whether bixel ij of current

beam is blocked by left leaf or not (1: close; 0: open).rcij : Binary variable indicates whether bixel ij of current

beam is blocked by right leaf or not (1: close; 0: open).bcij: Binary variable indicates whether bixel ij of current

beam is open or not (1: open; 0: close).Acij: Float variable contains the fluence of bixel ij.

lpij : Binary variable indicates whether bixel ij of pervious

beam is blocked by left leaf or not (1: close; 0: open).

l fij : Binary variable indicates whether bixel ij of follow-

ing beam is blocked by left leaf or not (1: close; 0: open).rpij : Binary variable indicates whether bixel ij of pervious

beam is blocked by right leaf or not (1: close; 0: open).

r fij : Binary variable indicates whether bixel ij of following

beam is blocked by right leaf or not (1: close; 0: open).The meanings of constraints are explained as below.Constraint (2): requires bixel ij either blocked by leaf

or open.Constraint (3): requires mechanical continuity of left leaf.Constraint (4): requires mechanical continuity of right leaf.Constraint (5): requires fluence of beam either zero or

fluence Ac.Constraint (6): requires fluence of beam is less than

fluence Iijc.

Constraint (7): requires the distance difference betweenleft leaf positions in current and previous beams less thanvmaxsϕ Δϕ , which is the maximal distance that a leaf cantravel between two neighboring control points or angles.Constraint (8): requires the distance difference between

left leaf positions in current and following beams less thanvmaxsϕ Δϕ which is the maximal distance that a leaf can travelbetween two neighboring control points or angles.Constraint (9): requires the distance difference between

right leaf positions in current and previous beams lessthan vmax

sϕ Δϕ , which is the maximal distance that a leaf cantravel between two neighboring control points or angles.Constraint (10): requires the distance difference between

right leaf positions in current and following beams lessthan vmax

sϕ Δϕ , which is the maximal distance that a leaf cantravel between two neighboring control points or angles.

AbbreviationsCI: Conformity index; CO: Critical organ; DAO: Direct aperture optimization;DVH: Dose volume histogram; DVH: Dose-volume histogram; FMD: Fastmonotonic descent; FMO: Fluence map optimization; gEUD: generalequivalent uniform dose; HI: Homogeneity index; IMAT: Intensity modulatedarc therapy; IMRT: Intensity modulated radiation therapy; MILP: Mixed integer

Yan et al. Radiation Oncology (2018) 13:101 Page 12 of 13

Page 13: A fast optimization approach for treatment planning of ......fluence-map optimization algorithm and leaf-sequencing algorithm, are explained in brief. These algorithms were developed

linear programming; MLC: Multiple leaf collimator; MU: Monitor unit;NT: Normal tissue; OAR: Organ at risk; PTV: Planning target volume;QS: Quality score; ROI: Region of interesting; TV: Target volume;VMAT: Volumetric modulated arc therapy; WE: Weighted error

FundingThis work was supported by National Key Projects of Research and Developmentof China (2016YFC0904600) and National Science Foundation of China(NSFC11275270).

Authors’ contributionsHY, JD, YL designed the study. All authors performed critical review of themanuscript and finally approved the manuscript.

Ethics approval and consent to participateNot applicable.

Competing interestsThe authors declare that they have no competing interests.

Publisher’s NoteSpringer Nature remains neutral with regard to jurisdictional claims inpublished maps and institutional affiliations.

Received: 18 October 2017 Accepted: 17 May 2018

References1. Teoh M, Clark CH, Wood K, Whitaker S, Nisbet A. Volumetric modulated arc

therapy: a review of current literature and clinical use in practice. Br J Radiol.2011;84(1007):967–96.

2. Bedford JL. Treatment planning for volumetric modulated arc therapy. MedPhys. 2009;36(11):5128–38.

3. Rao M, Yang W, Chen F, Sheng K, Ye J, Mehta V, Shepard D, Cao D.Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT:plan quality, delivery efficiency and accuracy. Med Phys. 2010;37(3):1350–9.

4. Yu CX, Tang G. Intensity-modulated arc therapy: principles, technologiesand clinical implementation. Phys Med Biol. 2011;56(5):R31–54.

5. Yu CX. Intensity-modulated arc therapy with dynamic multileaf collimation:an alternative to tomotherapy. Phys Med Biol. 1995;40(9):1435–49.

6. Shepard DM, Earl MA, Li XA, Naqvi S, Yu C. Direct aperture optimization: aturnkey solution for step-and-shoot IMRT. Med Phys. 2002;29(6):1007–18.

7. Crooks SM, Wu X, Takita C, Watzich M, Xing L. Aperture modulated arctherapy. Phys Med Biol. 2003;48(10):1333–44.

8. Bortfeld T, Webb S. Single-arc IMRT? Phys Med Biol. 2009;54(1):N9–20.9. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med

Phys. 2008;35(1):310–7.10. Vanetti E, Nicolini G, Nord J, Peltola J, Clivio A, Fogliata A, Cozzi L. On the

role of the optimization algorithm of RapidArc(®) volumetric modulated arctherapy on plan quality and efficiency. Med Phys. 2011;38(11):5844–56.

11. Verbakel WF, Cuijpers JP, Hoffmans D, Bieker M, Slotman BJ, Senan S.Volumetric intensity-modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study. Int J RadiatOncol Biol Phys. 2009;74(1):252–9.

12. Unkelbach J, Bortfeld T, Craft D, Alber M, Bangert M, Bokrantz R, Chen D, LiR, Xing L, Men C, Nill S, Papp D, Romeijn E, Salari E. Optimizationapproaches to volumetric modulated arc therapy planning. Med Phys. 2015;42(3):1367–77.

13. Rao M, Cao D, Chen F, Ye J, Mehta V, Wong T, Shepard D. Comparison ofanatomy-based, fluence-based and aperture-based treatment planningapproaches for VMAT. Phys Med Biol. 2010;55(21):6475–90.

14. Wang C, Luan S, Tang G, Chen DZ, Earl MA, Yu CX. Arc-modulated radiationtherapy (AMRT): a single-arc form of intensity-modulated arc therapy. PhysMed Biol. 2008;53(22):6291–303.

15. Ulrich S, Nill S, Oelfke U. Development of an optimization concept for arc-modulated cone beam therapy. Phys Med Biol. 2007;52(14):4099–119.

16. Cameron C. Sweeping-window arc therapy: an implementation of rotationalIMRT with automatic beam-weight calculation. Phys Med Biol. 2005;50(18):4317–36.

17. Bzdusek K, Friberger H, Eriksson K, Hardemark B, Robinson D, Kaus M.Development and evaluation of an efficient approach to volumetric arctherapy planning. Med Phys. 2009;36(6):2328–39.

18. Men C, Romeijn HE, Jia X, Jiang SB. Ultrafast treatment plan optimization forvolumetric modulated arc therapy (VMAT). Med Phys. 2010;37:5787–91.

19. Papp D, Unkelbach J. Direct leaf trajectory optimization for volumetricmodulated arc therapy planning with sliding window delivery. Med Phys.2014;41(1):011701.

20. Craft D, McQuaid D, Wala J, Chen W, Salari E, Bortfeld T. Multicriteria VMAToptimization. Med Phys. 2012;39(2):686–96.

21. Matuszak MM, Steers JM, Long T, McShan DL, Fraass BA, Romeijn HE, TenHaken RK. FusionArc optimization: a hybrid volumetric modulated arctherapy (VMAT) and intensity modulated radiation therapy (IMRT) planningstrategy. Med Phys. 2013;40(7):071713.

22. Nguyen D, Lyu Q, Ruan D, O'Connor D, Low DA, Sheng KA. Comprehensiveformulation for volumetric modulated arc therapy planning. Med Phys.2016;43(7):4263.

23. Yan H, Yin FF, Guan HQ, Kim JH. AI-guided parameter optimization ininverse treatment planning. Phys Med Biol. 2003;48:3565–80.

24. Li RP, Yin FF. Optimization of inverse treatment planning using a fuzzyweight function. Med Phys. 2000;27:691–700.

25. Rowbottom CG1, Webb S, Oldham M. Improvements in prostateradiotherapy from the customization of beam directions. Med Phys 1998;25(7 Pt 1):1171–1179.

26. Xia P, Hwang AB, Verhey LJ. A leaf sequencing algorithm to enlargetreatment field length in IMRT. Med Phys. 2002 Jun;29(6):991–8.

27. Langer M, Thai V, Papiez L. Improved leaf sequencing reduces segments ormonitor units needed to deliver IMRT using multileaf collimators. Med Phys.2001;28(12):2450–8.

28. Chen Y, Hou Q, Galvin JM. A graph-searching method for MLC leafsequencing under constraints. Med Phys. 2004;31(6):1504–11.

29. Dai J, Zhu Y. Minimizing the number of segments in a delivery sequencefor intensity-modulated radiation therapy with a multileaf collimator. MedPhys. 2001;28(10):2113–20.

30. Yang R, Dai J, Yang Y, Hu Y. Beam orientation optimization for intensity-modulated radiation therapy using mixed integer programming. Phys MedBiol. 2006;51(15):3653–66.

31. Zhu X, Thongphiew D, McMahon R, Li T, Chankong V, Yin FF, Arc-modulated WQJ. Radiation therapy based on linear models. Phys Med Biol.2010;55(13):3873–83.

32. Bohsung J, Gillis S, Arrans R, Bakai A, De Wagter C, Knöös T, Mijnheer BJ,Paiusco M, Perrin BA, Welleweerd H, Williams P. IMRT treatment planning:- acomparative inter-system and inter-Centre planning exercise of the ESTROQUASIMODO group. Radiother Oncol. 2005;76(3):354–61.

33. Stein J, Mohan R, Wang XH, Bortfeld T, Wu Q, Preiser K, Ling CC, Schlegel W.Number and orientations of beams in intensity-modulated radiationtreatments. Med Phys. 1997;24(2):149–60.

34. Wu Q, Mohan R, Niemierko A, Schmidt-Ullrich R. Optimization of intensity-modulated radiotherapy plans based on the equivalent uniform dose. Int JRadiat Oncol Biol Phys. 2002;52(1):224–35.

Yan et al. Radiation Oncology (2018) 13:101 Page 13 of 13