a comparison of pm10 monitors at a kerbside site in the northeast of england
TRANSCRIPT
![Page 1: A comparison of PM10 monitors at a Kerbside site in the northeast of England](https://reader036.vdocuments.us/reader036/viewer/2022080110/575075f01a28abdd2e9c168d/html5/thumbnails/1.jpg)
Atmospheric Environment 37 (2003) 4425–4434
A comparison of PM10 monitors at a Kerbside sitein the northeast of England
Monica Pricea,*, Susan Bulpitta, Michael B. Meyerb
aSchool of Health, Natural & Social Sciences, University of Sunderland, Sunderland, UKbRupprecht & Patashnick Co., Inc., Albany, NY, USA
Received 16 December 2002; received in revised form 26 June 2003; accepted 3 July 2003
Abstract
There is a need for a consistent measurement technique for both PM10 and PM2.5 that is capable of providing real-
time data suitable for determining the effects of particulate pollution on human health. Rupprecht and Patashnik have
developed a TEOMs monitor configuration that increases collection efficiency for the semi-volatile mass fraction, when
present. By operating at a lower setpoint temperature the system offers a real-time monitor that removes particle bound
moisture and promises to improve comparability with the European Union (EU) reference gravimetric method. Trials
with the device, a conventionally operated TEOM and a Partisols gravimetric monitor have shown that in the
northeast of England the loss of organics and nitrates may not be the major cause of the observed differences between
the monitors. Instead the data presented in this study indicate that it is the retention of particle bound water by the EU
reference method that may be causing the observed differences.
The presence and amounts of moisture associated with particles depends on the chemical composition and size range
of the particles as well as the ambient relative humidity. As both of these factors vary spatially and temporally it is
problematic to apply scaling factors to make data collected by the TEOM comparable to data collected by the EU
reference method. In addition, whether particle bound moisture, some of which may be absorbed after sampling should
be included in air quality standards needs further investigation.
r 2003 Elsevier Ltd. All rights reserved.
Keywords: Atmospheric particles; Measurement; Moisture; Semi-volatile matter
1. Introduction
Within the European Union (EU) the specified
reference method for monitoring PM10, EN12341, is a
gravimetric technique which requires that the PM10
mass fraction is collected upon a filter. Filters are
normally exposed for 24 h; although this can be
extended if the concentration of particulate matter is
thought to be low. Preparation of filters for exposure
must be carried out under strict, specified conditions.
Prior to, and after exposure, equilibration must take
place at 2071�C and a relative humidity of 5075% for
24 h (unless further equilibration is indicated). Filters
have to be weighed using a balance accurate to within
10 mg (BSI, 1998). Under the terms of the first daughter
directive on air quality, 1999/30/EC, EU member states
must monitor concentrations of atmospheric PM10 using
methods that demonstrate equivalence to the reference
method. Member states may use alternative techniques
providing they can illustrate a consistent relationship to
the reference method. Any results obtained must be
adjusted by a relevant factor to show equivalence with
the results that would have been obtained, should the
reference method have been used.
The requirement for filter equilibration and the nature
of the operation of gravimetric monitors means that
reference method results are often not obtained for a
ARTICLE IN PRESS
AE International – Europe
*Corresponding author.
E-mail address: [email protected] (M. Price).
1352-2310/$ - see front matter r 2003 Elsevier Ltd. All rights reserved.
doi:10.1016/S1352-2310(03)00582-X
![Page 2: A comparison of PM10 monitors at a Kerbside site in the northeast of England](https://reader036.vdocuments.us/reader036/viewer/2022080110/575075f01a28abdd2e9c168d/html5/thumbnails/2.jpg)
period of up to 2 weeks. However, in terms of the
determination of health effects and the issue of public
information, there is a need for real-time data. In order
to meet this requirement many countries are adopting
the tapered element oscillating microbalance (TEOM
Monitor), which has been in use world-wide since 1988,
to give an automatic, near real time measurement of
particulate matter (Meyer et al., 2000).
The TEOM method is based on the simple physical
law that the frequency of mechanical oscillation of an
element is directly proportional to the mass of an
element. The TEOM method uses a standard regulatory
PM10 inlet, operating at 16:7 l min�1; to select the
particles. These then pass through an isokinetic
flow splitter from which 3 l min�1 passes through the
16 mm diameter filter cartridge (13:5mm active area
sampling diameter) connected to the top of a narrow
oscillating hollow tapered glass tube. The collection
of particles by the filter at the free end of the
tapered glass tube will alter its effective mass, which in
turn will change its resonant frequency. As more
particles deposit on the filter the tube’s natural
frequency of oscillation decreases (QUARG, 1996). A
microprocessor converts the oscillation frequency to
mass and then to mass concentrations, which are
updated every 2 s: The inlet and the sensing system of
the TEOM are operated at a constant temperature of
50�C in order to minimise the adsorption/desorption
effects of atmospheric moisture on the microbalance
filter (Mukerjee et al., 1999).
Concerns have arisen about the TEOM monitor due
to the default operating temperature of 50�C: It is
believed that in addition to the removal of water vapour,
organic compounds and a proportion of the sulphate
and (more likely) nitrate may also be lost. This has
resulted in the TEOM monitor sometimes appearing to
give lower concentrations than the reference method in
Northern Europe (Muir, 2000).
A number of studies have presented data concerning
the loss of semi-volatile species on the filter of the
TEOM relative to the manual gravimetric method (Allen
et al., 1997; Ayers et al., 1999; Meyer et al., 1992).
APEG (1999) concluded that in the UK at concentra-
tions around 50 mg m�3 the TEOM tends to under-read
compared with a gravimetric sampler by between 15%
and 30%. In areas where semi-volatile components
constitute a high fraction of PM10 it would be desirable
to maintain an operating temperature as low as possible,
or as close to the reference method equilibration
temperature, to minimise the loss of these semi-volatile
fractions.
In order to provide a real-time monitor that compares
more readily with the EU reference method, Rupprecht
and Patashnick have developed a system that enables the
sample air stream flowing into a TEOM monitor to be
operated at a lower temperature. The sample equilibra-
tion system (SES) continuously conditions the flows of
the TEOM monitor to remove moisture, allowing the
instrument to be operated year-round at a temperature
closer to that of the gravimetric reference method
(Meyer et al., 2000). A reduction in the operating
temperature of the TEOM monitor should allow the
greater retention of semi-volatile material.
In order to remove moisture the SES incorporates a
Nafions dryer, which continuously conditions both the
sample and bypass flows of the TEOM monitor. The
SES enables filter-based mass measurements by the
TEOMmonitor under consistent, dry conditions (Meyer
et al., 2000) and at a reduced temperature. In a
validation study using a monitor fitted with a Nafion
drier Eatough et al. (2003) have demonstrated that fine
particles are not lost during the passage of air through
the drier system.
Initial trials with a TEOM monitor fitted with
an SES, to monitor PM2.5, were carried out in
Albany, New York, USA (Meyer et al., 2000). In this
study it was found that the TEOM monitor plus SES,
operated at 30�C; consistently recorded a higher
concentration of particulate matter than the TEOM
monitor operating at 50�C: The SES TEOM monitor
also showed a better relationship to a gravimetric
monitor than the conventional TEOM monitor
configuration.
One of the first sites in the EU to monitor particulate
matter utilising an SES modified TEOM was Trimdon
Street, Sunderland, northeast England. At a kerbside
monitoring station, PM10 has been monitored using
three machines: a TEOM Series1400a monitor operating
at 50�C; a Partisol—Plus model 2025 air sampler, and a
TEOM series 1400a PM monitor fitted with an SES,
operating at 30�C: This paper reports the findings of a
trial conducted between November 2000 and August
2001.
2. Methods
2.1. Sites
All three monitors were sited at the City of Sunder-
land kerside monitoring unit (sited 1m from the kerb) in
the northeast of England. The site is situated adjacent to
a busy road junction with traffic flows of
2000 vehicles h�1: Traffic flows are consistently high
throughout the day. Fig. 1 illustrates the location of
Sunderland within the northeast of England and the
location of the sampling site within the City.
2.2. Monitors
Monitoring for PM10 was carried out using a Partisol
Plus 2025 air sampler, a tapered element oscillating
ARTICLE IN PRESSM. Price et al. / Atmospheric Environment 37 (2003) 4425–44344426
![Page 3: A comparison of PM10 monitors at a Kerbside site in the northeast of England](https://reader036.vdocuments.us/reader036/viewer/2022080110/575075f01a28abdd2e9c168d/html5/thumbnails/3.jpg)
microbalance TEOM series 1400 operating at 50�C
and a TEOM monitor plus SES drier operating
at 30�C; all manufactured by Rupprecht &
Pataschnik Co. The addition of a drier to the
TEOM device has been described elsewhere (Meyer
et al., 2000). Monitoring using the co-located
devices commenced at the site in November 2000.
Total nitrogen oxides, nitric oxide and nitrogen
dioxide are also monitored at the site by Sunderland
city council.
Partisol filters were equilibrated pre- and post-
sampling according to the requirements of EN12341
and weighed using a balance accurate to 0:001 mg; withreplicate weighings of each filter being made. Initial
investigations of the method had shown that there was
no further weight loss from filters after the recom-
mended equilibration period. Filters were exposed in the
monitor for a period of 24 h:Flow rates through both TEOM monitors are
16:7 l min�1 with 3 l min�1 passing through the sampling
head to a 13:5 mm filter. The Partisol has a flow rate of
16:7 l min�1 passing through a filter of 47 mm diameter
(active area of 42 mm), thus the filter face velocities of
the air passing through the filters are comparable.
Particulate matter was collected onto PTFE-coated glass
fibre filters (Pall Corporation).
Meteorological data for the sampling period was
obtained from the University meteorological station in
Sunderland.
3. Results and discussion
3.1. Analysis of PM10 data
The PM10 data collected by the three monitors
appears in Figs. 2–4 and the correlation coefficients in
Table 1. Good correlation is displayed between the
TEOM and the TEOM plus SES all year round.
However, the correlation between these monitors and
the Partisol sampler appears to be seasonal, with the
strongest correlation occurring during the summer
months. Figs. 2–4 display the time series for each
sampler (where the TEOM and TEOM plus SES are
averaged over 24 h periods to match the sampling period
of the Partisol). During periods of elevated PM10 in the
winter and spring, the Partisol sampler records higher
mass than the TEOM and the TEOM plus SES.
However, during the summer all three instruments tend
to record elevated PM10 concentrations.
The wintertime recorded differences between the
Partisol and TEOM monitors occurred during episodes
of elevated PM10, with the gravimetric monitor record-
ing a greater increase than the TEOM monitor. The
wintertime discrepancy in PM10 concentrations as
measured by the Partisol and TEOM has been observed
in previous studies such as that by Green et al. (2000).
This seasonal difference in the relationship is shown in
the strength of correlations (Table 1). In November the
correlation coefficient for the relationship between the
ARTICLE IN PRESS
0 5kms
Scale
0 0.5km
Scale
N
N
Morpeth
Biyth
A19
A69Newcastle
Chester-le-Street
Durham
Gateshead
Area ofdetail map
ENGLAND
A1(
M)
A19
Middlesborough
Hartlepool
Washington
A1(
T)
South Shields
SUNDERLAND
Weatherstation
Monitoring site
City centre
River Wear
NorthSea
Fig. 1. Location of Sunderland in the northeast of England and the site within the city of Sunderland.
M. Price et al. / Atmospheric Environment 37 (2003) 4425–4434 4427
![Page 4: A comparison of PM10 monitors at a Kerbside site in the northeast of England](https://reader036.vdocuments.us/reader036/viewer/2022080110/575075f01a28abdd2e9c168d/html5/thumbnails/4.jpg)
Partisol monitor and the TEOMmonitor is 0.59 whereas
in August it is 0.89. Throughout the time period the
results obtained for the TEOM and the TEOM plus SES
show a strong relationship.
Recorded differences between the TEOM and gravi-
metric monitors are well documented. They include the
work of Salter and Parsons (1999), who used a TEOM
1400a series monitor operating at 3 l min�1 airflow and a
Partisol Model 2000 air sampler operating at
16:7 l min�1: In this study the TEOM was operated at
50�C and results were collected for PM10 over a period
of 100 days from April to October 1997. The results
obtained showed that although the TEOM and Partisol
results correlate well at low values of PM10 as particulate
ARTICLE IN PRESS
0
20
40
60
80
100
120
140
1/3 5/3 9/3 13/3 17/3 21/3 25/3 29/3 2/4 6/4 10/4 14/4 18/4 22/4 26/4 30/4 4/5 8/5 12/5 16/5 20/5 24/5 28/5
date
PM
10 µg
m-3
Partisol TEOM TEOM+SES
Fig. 3. PM10 ðmg m�3Þ results for March 2001–May 2001.
Fig. 2. PM10 ðmgm�3Þ results for November 2000–January 2001.
M. Price et al. / Atmospheric Environment 37 (2003) 4425–44344428
![Page 5: A comparison of PM10 monitors at a Kerbside site in the northeast of England](https://reader036.vdocuments.us/reader036/viewer/2022080110/575075f01a28abdd2e9c168d/html5/thumbnails/5.jpg)
concentrations increased the Partisol was found to
record a higher concentration than the TEOM. A study
by Cyrys et al. (2001) evaluated the difference between a
TEOM 1400a series monitor operated at 50�C and air
flow of 3 l min�1 with a Harvard–Marple impactor,
using Teflon membrane filters (PTFE). Data were
collected for the period March–April 1999 and the
TEOM consistently recorded lower values than the
Harvard Impactor.
Both of the above studies collected data over a
relatively short time period even though the composition
of particulate matter is reported to vary on a seasonal
basis (Harrison et al., 1997). The differing composition
of sampled PM10 may lead to both spatial and temporal
differences in the recorded discrepancies between
monitors. The results collected in this study illustrate
the seasonal differences and highlight the necessity for
monitoring over an extended time scale.
In terms of the compositional differences between
PM10, as sampled by gravimetric monitors and the
TEOM monitor operated at 50�C; a number of studies,
including those by Green et al. (2001), Chung et al.
(2001) and Allen et al. (1997) suggest that the elevated
temperature of the TEOM monitor causes the loss of
semi-volatile material including sulphate and nitrate.
The results from the present study would suggest that
this is not the case at the Sunderland site. The TEOM
monitor plus SES drier operated at 30�C; records a
comparable value to the TEOM monitor operated at
50�C; and for some periods of the study this was lower.
These results contrast to those reported byMeyer et al.
(2000) in an initial trial with the SES carried out at
Albany, New York, USA during the summer. During
the Albany study temperatures were high, average
daytime temperature was 30�C dropping to 10–15�C
ARTICLE IN PRESS
0
10
20
30
40
50
60
70
80
90
100
1/6 7/6 13/6 19/6 25/6 1/7 7/7 13/7 19/7 25/7 31/7 6/8 12/8 18/8 24/8 30/8
date
Partisol TEOM TEOM+SES
PM
10 µg
m-3
Fig. 4. PM10 ðmgm�3Þ results for June 2001–August 2001.
Table 1
Correlation coefficients for the different monitors
TEOM SES
Full data set Partisol 0.675a 0.743a
TEOM — 0.867a
November Partisol 0.589a 0.502a
TEOM — 0.879a
December Partisol 0.746a 0.680a
TEOM — 0.952a
January Partisol 0.403a 0.608a
TEOM — 0.867a
March Partisol 0.668a 0.664a
TEOM — 0.931a
April Partisol — 0.650a
TEOM — 0.936a
May Partisol 0.737a 0.845a
TEOM — 0.918a
June Partisol 0.794a 0.841a
TEOM — 0.932a
July Partisol — —
TEOM — —
August Partisol 0.887a 0.854a
TEOM — 0.975a
aSignificant correlation at pp0:05:
M. Price et al. / Atmospheric Environment 37 (2003) 4425–4434 4429
![Page 6: A comparison of PM10 monitors at a Kerbside site in the northeast of England](https://reader036.vdocuments.us/reader036/viewer/2022080110/575075f01a28abdd2e9c168d/html5/thumbnails/6.jpg)
at night. The recorded relative humidity was high at
night reaching 95–100% but lower during the daytime
dropping to 40–50%. The results obtained by Meyer
et al. showed an increased quantity of PM2.5 monitored
by the TEOM plus SES monitor as compared to the
conventional TEOM monitor. The results obtained by
the TEOM plus SES being comparable to those
obtained by a gravimetric monitor.
At the Sunderland site, weather conditions are
different to those in Albany, during the winter
temperatures range from nighttime lows of –5–8�C
and daytime temperatures of 5–12�C. Relative humidity
is high throughout the day and night (80–100%). In the
summer, temperatures are found to range from night-
time lows of 6–20�C, to daytime temperatures of 12–
25�C. Relative humidity in the summer is lower ranging
from 55–85%. At this site both TEOM monitors
correlate well with the gravimetric monitor during the
summer but not during the winter. The contrasting
results indicate that the difference between PM10
monitors will be dependent upon a range of factors,
including meteorology. The results from one study
should therefore not be applied to another area of
contrasting meteorology and particulate composition.
In the winter at the Sunderland site, the TEOM plus
SES and the TEOM operating at 50�C show a good
correlation, suggesting that the loss of semi-volatile
materials may not be occurring at this site.
In order to obtain an accurate measure of atmo-
spheric particulate matter, the retention of semi-volatile
material will be of importance if the composition of the
atmospheric aerosol comprises materials that will be lost
at 50�C, and under certain circumstances this may
comprise materials other than nitrate. Moisture could
also be considered as a semi-volatile material lost by the
50�C TEOM monitor. The default operating tempera-
ture of the TEOM monitor is chosen to minimise the
amounts of particle bound water monitored whereas
gravimetric monitors rely on filter equilibration to
remove moisture.
A study carried out in Canada (Mignacca and Stubbs,
1999) showed that a reduction in the operating
temperature of the TEOM from 50�C to 30�C resulted
in a 22% increase in measured PM10. However, it is
unclear whether this increase is due to the retention of
particle bound moisture or semi-volatile nitrate and
organics. Without a full knowledge of the composition
of the atmospheric aerosol it is not possible to determine
the nature of the differences recorded in terms of the
composition of the material lost. In some areas the
material lost could be organics or nitrate, in other areas
it may be particle bound moisture.
Atmospheric aerosol is a complex mixture of inor-
ganic and organic components where organic species can
represent up to 50% of the aerosol mass depending on
location (Cruz and Pandis, 2000). Whilst the water
uptake of the inorganic fraction has been widely
investigated and is well understood the knowledge
gained concerning the organic fraction has been limited.
In terms of the inorganic fraction at a low relative
humidity aerosol particles comprising inorganic salts are
solid (Seinfeld and Pandis, 1998). However, as the
relative humidity increases a threshold value will be
reached above which water will be rapidly absorbed by
the particle. The relative humidity at which this occurs is
characteristic for the aerosol composition and is termed
the deliquescence relative humidity. Values for typical
components of atmospheric aerosols are: NaCl—75.3%,
NH4NO3—61.8%, (NH4)2SO4—79.9% (Seinfeld and
Pandis, 1998). These are all values of relative humidity
exceeded in Sunderland routinely in the winter but also
during the spring and summer. Seinfeld and Pandis
(1998) note that not all aerosol species exhibit deliques-
cence behaviour, some such as H2SO4 are hygroscopic
and the water content associated with them changes
smoothly as the relative humidity increases or decreases.
Studies with the organic fraction have shown that a
significant fraction of the particulate organic carbon in
the atmosphere is water soluble organic carbon (Yang
et al., 2003) which due to its affinity with water plays a
role in aerosol–cloud interactions, wet scavenging and
the formation of atmospheric haze. The affinity of the
material for water should also perhaps be considered in
terms of particle bound moisture and the measurement
of PM10.
Whilst the effects of moisture on the inorganic and
organic fraction has been considered the effects of
moisture absorption by internal mixtures of salts and
insoluble components, for example soot, is more difficult
to predict (Ebert et al., 2002). The authors report that
organic components can alter the hygroscopic behaviour
of salt particles severely.
The quantity of water associated with PM10 will
therefore be dependent on its composition and the
relative humidity of the atmosphere. This is likely to
vary both seasonally and temporally and with local and
regional pollution sources. The role of particle bound
moisture in mixed aerosols is an area that requires
further research not only in the atmosphere but also by
particles collected on the filters of particulate monitors.
Whilst the TEOM is operated at 50�C in order to
minimise particle bound water, equilibration of Partisol
filters according to EN12341(BSI 1998) is carried out to
minimise the loss of volatile material which may include
particle bound moisture. The results obtained in the
present study suggest that any water associated with the
PM10 collected on the filters may also be retained. The
correlation between the 30�C TEOM plus SES drier and
the TEOM operated at 50�C and the observed
differences to the Partisol, most noticeably in the winter,
would suggest that moisture plays a role in the
differences between monitors at the Sunderland site.
ARTICLE IN PRESSM. Price et al. / Atmospheric Environment 37 (2003) 4425–44344430
![Page 7: A comparison of PM10 monitors at a Kerbside site in the northeast of England](https://reader036.vdocuments.us/reader036/viewer/2022080110/575075f01a28abdd2e9c168d/html5/thumbnails/7.jpg)
As noted previously the deliquescent relative humidity
for particulate matter commonly found in the atmo-
sphere is of the order of 70% RH. However, although
the particles will absorb moisture at a relative humidity
of 70% they do not start to release this moisture until a
much lower RH is achieved, the hysteresis phenomenon.
For (NH4)2SO4 this may not occur until the relative
humidity is reduced to 30–40% (Seinfeld and Pandis,
1998) hence the equilibration relative humidity chosen
for Partisol filters (50%) may not be sufficiently reduced
to ensure moisture is removed from the particulate mass.
If this is the case the differences between gravimetric
monitors and real-time monitors, operated to minimise
moisture, will be dependent on the amount of moisture
associated with the particulate matter collected on the
filter of the gravimetric monitor. This in turn will be
determined by the composition of the aerosol and
atmospheric relative humidity.
3.2. Relationship to meteorological parameters
In order to review the role of particle bound water
further, the data have been compared to the meteor-
ological data collected for the same period. Tables 2 and
3 present data for the meteorological conditions at times
when the greatest discrepancy ð> 20 mg m�3Þ is found
between the monitors and also at times of good
agreement (o5 mg m�3). The meteorological data were
averaged for the 24 h period of interest to allow
comparison with the PM10 data.
Rainfall was recorded on several of the days when
high discrepancies between the Partisol and TEOM and
TEOM plus SES monitors were observed. In addition
high NOx concentrations have been recorded on all of
the dates with the exception of 9/6, suggesting that both
pollution levels, including a high proportion of fresh
aerosol and atmospheric moisture have been high at the
same time. The effect of the timing of a rainfall event on
the particle bound moisture content of aerosols collected
on filters remains unclear. For example, if rainfall occurs
before sampling, the particles may be washed out of the
atmosphere thus reducing their concentration. However,
if rainfall occurs after the filter sample has been collected
by the gravimetric monitor, the movement of a moist air
stream over the collected particles could lead to the
absorption of moisture on both particles and the filter.
Of the remaining dates other than 21/12 the dew point
had been reached indicating condensation of atmo-
spheric moisture will be occurring.
The timing of increased atmospheric moisture content
and the presence of high particle concentrations may
explain the contrasting results obtained in the initial
study carried out with the TEOM plus SES (Meyer et al.,
2000) and the results obtained in Sunderland. In the
initial trials relative humidity was high at night and
reduced during the day whereas in Sunderland relative
ARTICLE IN PRESS
Table
2
Datesofgreatest
discrepancy
(>20mg
m�3)betweenthePartisolandTEOM
monitors
Date
Partisol
TEOM
TEOM+SES
NO
xMin
temp
Maxtemp
Dew
point
RH
Rainfall
Rainfall
Sunshine
Meanwindspeed
Winddirection
(mgm
�3)
(mgm
�3)
(mgm
�3)
(ppb)
(�C)
(�C)
(�C)
(%)
(mm)
(h)
(h)
(ms�
1)
7/12/00
71.5
40.9
39.3
108
6.4
10.0
695
16.0
7.2
0.6
3.8
SW
11/12/00
75.2
55.8
45.0
94
6.8
12.6
986
00
2.4
3.8
SW
18/12/00
82.4
32.2
33.1
147
�0.5
5.4
098
00
00.7
S
21/12/00
66.2
38.1
38.9
93
6.6
7.2
691
00
02.8
SE
18/1/01
81.3
31.1
42.1
131
0.7
3.2
189
4.9
50
0.5
Calm
6/3/01
108
46.5
52.9
68
07.4
082
3.3
2.5
3.4
4.3
SE
7/3/01
87.2
43.7
35.3
107
1.7
14.3
7100
00
5.3
2.2
Calm
9/6/01
93.9
48.9
60.6
23
7.9
15.3
885
14.5
7.6
1.7
2.7
W
M. Price et al. / Atmospheric Environment 37 (2003) 4425–4434 4431
![Page 8: A comparison of PM10 monitors at a Kerbside site in the northeast of England](https://reader036.vdocuments.us/reader036/viewer/2022080110/575075f01a28abdd2e9c168d/html5/thumbnails/8.jpg)
humidity is high throughout both the day and night. At
the Sunderland site the RH is high when atmospheric
particulate pollution is high whilst in the Albany study
the relative humidity was rising as particulate levels may
have been falling.
Table 3 shows that the meteorological conditions
found when there is a good agreement between the
monitors have tended to be dry with some sunshine
recorded. However, there are occasions when rainfall
has been recorded or the dew point reached and there is
no discrepancy between the monitors. The variation in
composition of atmospheric particulate matter will
determine the amounts of moisture adsorbed and this
could account for the variability in results obtained.
A thin film of water molecules coats many substances
under ambient conditions (Romakkaniemi et al., 2001)
and adsorption has been shown to occur at relative
humidities below the deliquescent relative humidity. Due
to its importance, the hygroscopic properties of aerosol
particles has been studied for a long time (Ebert et al.,
2002) including the work of Hanel (1976). A number of
studies have considered the adsorption of moisture by
atmospheric particulate matter including that by Hameri
et al. (2000), who have measured the hygroscopic
growth of ultrafine aerosol particles. However, in the
gravimetric measurement of atmospheric PM10
and PM2.5 it is assumed that once a particle is
deposited on a filter it remains inert even in the presence
of an air stream of high relative humidity containing a
range of gaseous pollutants. The work of Hameri et al.
(2000) and Romakkaniemi et al. (2001) would
suggest that this is unlikely to be the case and that
collected particles will continue to adsorb moisture at
relative humidities at or above the deliquescent relative
humidity for the sampled particles. In Sunderland,
particularly during the winter, the recorded ambient
relative humidity is above the deliquescent relative
humidity for particulate matter commonly sampled by
gravimetric monitors.
Although the gain of moisture by atmospheric
particulate matter associated with filters has been
illustrated by a number of researchers including the
work of Jarrett et al. (2001) its subsequent removal or
retention during filter conditioning has yet to be
determined. When the relative humidity decreases the
deliquescent RH will be reached, however, the
particle does not lose all of its associated moisture at
this or a decreased relative humidity (Seinfeld and
Pandis, 1998). In order to regain a solid state, salt nuclei
need to be formed and salt crystals must grow around
them. In the atmosphere this occurs at a relative
humidity significantly lower than the deliquescent
relative humidity. For salts such as ammonium nitrate
with a deliquescent relative humidity of 61.8% this may
well be below the relative humidity of 50% chosen for
filter conditioning.
ARTICLE IN PRESS
Table
3
Datesofgoodagreem
ent(o
5mg
m�3)betweenthePartisolandTEOM
monitors
Date
Partisol
TEOM
TEOM+SES
NO
xMin
temp
Maxtemp
Dew
point
RH
Rainfall
Rainfall
Sunshine
Meanwindspeed
Winddirection
(mgm
�3)
(mgm
�3)
(mgm
�3)
(ppb)
(�C)
(�C)
(�C)
(%)
(mm)
(h)
(h)
(ms�
1)
6/11/00
23.0
21.4
19.3
28
4.8
9.6
496
23.3
16
06.1
Calm
16/12/00
21.2
18.1
18.1
50
�1.0
4.2
�4
74
00
1.2
1.9
W
29/1/01
22.4
24.9
18.9
0.4
6.6
191
00
7.7
1.5
W
24/4/01
52.4
53.1
51.1
89
3.3
10.3
794
0.6
1.5
7.2
2.3
Calm
9/5/01
45.1
47.9
41.9
35
4.5
12.6
784
00
2.3
2.3
NE
31/5/01
20.8
20.6
18.3
20
9.4
17.4
866
2.3
0.4
12.6
5.3
W
3/6/01
15.2
16.8
14.0
12
6.0
14.7
358
00
6.8
3.8
N
1/8/01
26.7
24.9
22.5
35
12.0
24.5
10
64
00
12.3
1.9
W
14/8/01
21.6
22.5
20.9
29
18.8
24.7
17
70
0.7
0.4
10.1
3.9
SW
M. Price et al. / Atmospheric Environment 37 (2003) 4425–44344432
![Page 9: A comparison of PM10 monitors at a Kerbside site in the northeast of England](https://reader036.vdocuments.us/reader036/viewer/2022080110/575075f01a28abdd2e9c168d/html5/thumbnails/9.jpg)
4. Conclusion
Particulate matter both PM10 and PM2.5 will comprise
a range of chemical species plus variable amounts of
particle bound water, where the amount of moisture will
be dependent on the composition of the atmospheric
aerosol and the relative humidity of the atmosphere.
This will vary both temporally and spatially.
The default operating temperature chosen for the
TEOM of 50�C; minimises the amounts of particle
bound moisture monitored. This has led to criticisms
that the TEOM under-records atmospheric particulate
matter due to the loss of semi-volatile materials such as
organics and nitrates. However, particle bound moisture
could be classified as a semi-volatile constituent of
atmospheric aerosols which is lost by the TEOM but
retained by gravimetric monitors. The results of the
monitoring campaign carried out in Sunderland, using a
TEOM fitted with a drier and operated at a lower
temperature, would suggest that at this site the gravi-
metric monitor is retaining a greater proportion of
atmospheric moisture than the TEOM monitor.
At this site it would therefore not be possible to apply
a factor to adjust the TEOM results to make them
comparable with the EU reference method, in that the
recorded differences are not consistent but governed by
aerosol composition and atmospheric relative humidity.
In addition application of a factor would unnecessarily
penalise local industry in that the increased amounts of
particulate matter recorded by the gravimetric monitor
could be due to atmospheric moisture and not an
increased anthropogenic input.
In any review of monitors and air quality standards
the role of particle bound moisture needs to be
evaluated.
References
Allen, G., Sioutas, C., Koutrakis, P., Reiss, R., Lurmann,
F.W., Roberts, P.T., 1997. Evaluation of the TEOM
method for measurement of ambient particulate mass in
urban areas. Journal of the Air and Waste Management
Association 47, 682–689.
APEG, 1999. Source apportionment of airborne particulate
matter in the United Kingdom. Airborne Particulates
Expert Group.
Ayers, G.P., Keywood, M.D., Gras, J.L., 1999. TEOM vs.
manual gravimetric methods for determination of PM2.5
aerosol mass concentrations. Atmospheric Environment 33,
3717–3721.
BSI, 1998. BS EN 12341:1998. Air quality—determination of
the PM10 fraction of suspended particulate matter—
reference method and field test procedure to demonstrate
reference equivalence of measurement methods. British
Standards Institute.
Chung, A., Chang, D.P.Y., Kleeman, J., Perry, K.D.,
Cahill, T.A., Dutcher, D., McDougall, E.M., Stroud, K.,
2001. Comparison of real-time instruments used to monitor
airborne particulate matter. Journal of the Air and Waste
Management Association 51, 109–120.
Cruz, C.N., Pandis, S.P., 2000. Deliquescence and hygroscopic
growth of mixed inorganic–organic aerosol. Environment
Science and Technology 34, 4313–4319.
Cyrys, J., Dietrich, G., Kreyling, W., Tuch, T., Heinrich, J.,
2001. PM2.5 measurements in ambient aerosol: comparison
between Harvard Impactor (HI) and the tapered element
oscillating microbalance (TEOM) system. The Science of
the Total Environment 278, 191–197.
Eatough, D.J., Long, R.W., Modey, W.K., Eatough, N.L.,
2003. Semi-volatile secondary organic aerosol in urban
atmospheres: meeting a measurement challenge. Atmo-
spheric Environment 37, 1277–1292.
Ebert, M., Inerle-Hof, M., Weinbruch, S., 2002. Environmental
scanning electron microscopy as a new technique to
determine the hygroscopic behaviour of individual aerosol
particles. Atmospheric Environment 36, 5909–5916.
Green, D., Fuller, G., Barratt, B., 2000. ‘Marylebone Road
(‘Supersite’), Annual Report, 199. http://www.seiph.umds.
ac.uk/detr/ss reports/reportar98.htm.
Green, D., Fuller, G., Barratt, B., 2001. Evaluation
of TEOM ‘correction factors’ for assessing the EU
Stage 1 limit values for PM10. Atmospheric Environment
35, 2589–2593.
Hameri, K., Vakeva, M., Hansson, H.C., Laaksonen, A., 2000.
Hygroscopic growth of ultrafine ammonium sulphate
aerosol measured using an ultrafine tandem differential
mobility analyzer. Journal of Geophysical Research-Atmo-
spheres 105, 22231–22242.
Hanel, G., 1976. The properties of atmospheric aerosol
particles as functions of the relative humidity at thermo-
dynamic equilibrium with the surrounding moist air.
Advances in Geophysics 19, 74–183.
Harrison, R.M., Deacon, A.R., Jones, M.R., 1997. Sources and
processes affecting concentration of PM10 and PM2.5
particulate matter in Birmingham (UK). Atmospheric
Environment 31 (24), 4103–4117.
Jarrett, R.P., Clark, N.N., Gilbert, M., Ramamurthy, R., 2001.
Evaluation and correction of moisture adsorption and
desorption from a tapered element oscillating microbalance.
Powder Technology 119 (2–3), 215–228.
Meyer, M., Lijek, J., Ono, D., 1992. Continuous PM10
measurements in a woodsmoke environment, PM10 stan-
dards and non-traditional particulate controls. In: Chow,
J.C., Ono, D.M. (Eds.), Journal of the Air and Waste
Management Association 21 (1), 24–38.
Meyer, M.B., Patashnick, H., Ambs, J.L., Rupprecht, E., 2000.
Development of a sample equilibration system for the
TEOM continuous PM monitor. Journal of the Air and
Waste Management Association 50, 1345–1349.
Mignacca, D., Stubbs, K., 1999. Effects of the equilibration
temperature on PM10 concentrations from the TEOM
method in the Lower Fraser Valley. Journal of the Air and
Waste Management Association 49 (10), 1250–1254.
Muir, D., 2000. The suitability of tapered element oscillating
microbalance (TEOMs) for PM10 monitoring in Europe.
The use of PM10 data as measured by TEOM for
ARTICLE IN PRESSM. Price et al. / Atmospheric Environment 37 (2003) 4425–4434 4433
![Page 10: A comparison of PM10 monitors at a Kerbside site in the northeast of England](https://reader036.vdocuments.us/reader036/viewer/2022080110/575075f01a28abdd2e9c168d/html5/thumbnails/10.jpg)
compliance with the European Air Quality Standard.
Atmospheric Environment 34, 3209–3212.
Mukerjee, S., Shadwick, D.S., Bowser, J.J., Carmichael, L.Y.,
1999. Application of the dual fine particle sequential
sampler, a tapered element oscillating microbalance and
other air monitoring methods to assess transboundary
influences of PM2.5. Field Analytical Chemistry and
Technology 3 (3), 201–217.
QUARG, 1996. Airborne particulate matter in the United
Kingdom. Third report of the Quality of Urban Air Review
Group. May 1996, Department of the Environment. ISBN 0
0952077132.
Romakkaniemi, S., Hameri, K., Vakeva, M., Laaksonen, A.,
2001. Adsorption of water on 8–15nm NaCl and
(NH4)2SO4 aerosols measured using an ultrafine tandem
differential mobility analyzer. Journal of Physical Chemistry
A 105 (35), 8183–8188.
Salter, B., Parsons, L.F., 1999. Field trials of the TEOM and
partisol for PM10 monitoring in the St Austell china
clay area, Cornwall, UK. Atmospheric Environment 33,
2111–2114.
Seinfeld, J.H., Pandis, S.N., 1998. Atmospheric Chemistry and
Physics. From Air Pollution to Climate Change. Wiley
Interscience, New York.
Yang, H., Qianfeng, L., Jian Zhen, Y., 2003. Comparison of
two methods for the determination of water-soluble organic
carbon in atmospheric particles. Atmospheric Environment
37, 865–870.
ARTICLE IN PRESSM. Price et al. / Atmospheric Environment 37 (2003) 4425–44344434