A “Short” Update on the P2IT MEBT Emittance Scanner Vic Scarpine, et al PIP-II Meeting 24 Jan 2017

Upload: others

Post on 03-Oct-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: A “Short” Update on the P2IT MEBT...MEBT Allison-Style Emittance Scanner Parameter LEBT MEBT Kinetic Energy 30 keV 2.1 MeV Nominal Beam Current 5 mA 5 mA Max/Min Beam Current 10

A “Short” Update on the P2IT MEBT

Emittance Scanner

Vic Scarpine, et al

PIP-II Meeting

24 Jan 2017

Page 2: A “Short” Update on the P2IT MEBT...MEBT Allison-Style Emittance Scanner Parameter LEBT MEBT Kinetic Energy 30 keV 2.1 MeV Nominal Beam Current 5 mA 5 mA Max/Min Beam Current 10

Emittance Scanners for P2IT

• Fast, accurate transverse emittance measurements critical for

commissioning and quantifying the operation of the warm front-end of

PIP-II

• Emittance measurements for P2IT LEBT utilizes a Allison style emittance

scanner

– Fairly well proven and studied at different laboratories

• Choice of emittance measurement device in P2IT MEBT was not a clear

cut a choice

– Quad scans with profile measurements

• Ok but rms only, no phase space

– Slit-slit scanner or slit-profile scanner

• Phase space measurement but fairly slow

– Allison scanner??

• Has never been done at MeV energies

– High beam power material damage

– “Stiff” beam difficult to bend

1/24/2017Vic Scarpine - MEBT Emittance Scanner2

Page 3: A “Short” Update on the P2IT MEBT...MEBT Allison-Style Emittance Scanner Parameter LEBT MEBT Kinetic Energy 30 keV 2.1 MeV Nominal Beam Current 5 mA 5 mA Max/Min Beam Current 10

Allison Scanner for Source/LEBT Emittance

Measurements

Water cooled Allison scanner – CW operation

• Developed in collaboration with SNS

• Adjustable entrance slits

• Status: – Operational in LEBT

– Different locations along LEBT

– Final location in ion source chamber

Vic Scarpine - MEBT Emittance Scanner3 1/24/2017

-H

Page 4: A “Short” Update on the P2IT MEBT...MEBT Allison-Style Emittance Scanner Parameter LEBT MEBT Kinetic Energy 30 keV 2.1 MeV Nominal Beam Current 5 mA 5 mA Max/Min Beam Current 10

LEBT Emittance Scanner at Ion Source

1/24/2017Vic Scarpine - MEBT Emittance Scanner4

Other measurements locations during LEBT commissioning:• After solenoid 1• ~At RFQ entrance

Page 5: A “Short” Update on the P2IT MEBT...MEBT Allison-Style Emittance Scanner Parameter LEBT MEBT Kinetic Energy 30 keV 2.1 MeV Nominal Beam Current 5 mA 5 mA Max/Min Beam Current 10

“Typical” LEBT Emittance Control Panel Image

1/24/2017Vic Scarpine - MEBT Emittance Scanner5

Page 6: A “Short” Update on the P2IT MEBT...MEBT Allison-Style Emittance Scanner Parameter LEBT MEBT Kinetic Energy 30 keV 2.1 MeV Nominal Beam Current 5 mA 5 mA Max/Min Beam Current 10

MEBT Allison-Style Emittance Scanner

Parameter LEBT MEBT

Kinetic Energy 30 keV 2.1 MeV

Nominal Beam Current 5 mA 5 mA

Max/Min Beam Current 10 mA/0.5 mA 10 mA/0.5 mA

Nominal Beam radius (rms) 2.5 mm 2 mm

Max CW Beam Power 300 W 21 kW*

Nominal Max Beam Pulse CW 20 msec

Measurement X range (resolution) ± 30 mm (100 mm) ± 15 mm (25 mm)

Measurement X’ range (resolution) ± 80 mrad (0.5 mrad) ± 12 mrad (0.25 mrad)

Signal Dynamic Range 103 103

1/24/2017Vic Scarpine - MEBT Emittance Scanner6

*Limit beam power to < 8 W/mm2

Page 7: A “Short” Update on the P2IT MEBT...MEBT Allison-Style Emittance Scanner Parameter LEBT MEBT Kinetic Energy 30 keV 2.1 MeV Nominal Beam Current 5 mA 5 mA Max/Min Beam Current 10

From MEBT

Emittance

Scanner FRS

1/24/2017Vic Scarpine - MEBT Emittance Scanner7

Parameter Unit Value

Motion and positioning

Vertical range of measurement positions with respect to the beam line axis

mm ± 15

Absolute accuracy of vertical position of the sensor mm ≤ 1

Reproducibility of vertical position mm ≤ 0.1

Resolution of vertical position mm ≤ 0.025

Sensor horizontal alignment with respect to beam line axis mm ≤ 1

Sensor yaw alignment with respect to beam line axis mR 10

MAES module pitch adjustment range mR ± 35

MAES module pitch adjustment resolution mR 1

MAES module pitch angle stability thru measurement region mR 0.2

Scanner dimensions

Scanner slit-to-slit distance mm 320

HV deflector plate length mm 300

Minimum HV deflector plate width mm > 40

Gap between teeth on HV plates mm 5.5 ±0.2*

Angle of the teeth surface on HV plates degree > 2

Teeth depth mm ≤ 0.5

Front slit gap mm 0.2

Rear slit gap mm 0.2

Front slit/Rear slit/HV plates axial alignment mm < 0.2

Electric

HV amplitude on each plate kV -1 to +1

Time Slice Resolution mS 1

Page 8: A “Short” Update on the P2IT MEBT...MEBT Allison-Style Emittance Scanner Parameter LEBT MEBT Kinetic Energy 30 keV 2.1 MeV Nominal Beam Current 5 mA 5 mA Max/Min Beam Current 10

Assembly Underway

1/24/2017Vic Scarpine - MEBT Emittance Scanner8

Dry-Run Assembly

Page 9: A “Short” Update on the P2IT MEBT...MEBT Allison-Style Emittance Scanner Parameter LEBT MEBT Kinetic Energy 30 keV 2.1 MeV Nominal Beam Current 5 mA 5 mA Max/Min Beam Current 10

More Assembly Pics

1/24/2017Vic Scarpine - MEBT Emittance Scanner9

At alignment

In VacuumChamber

Page 10: A “Short” Update on the P2IT MEBT...MEBT Allison-Style Emittance Scanner Parameter LEBT MEBT Kinetic Energy 30 keV 2.1 MeV Nominal Beam Current 5 mA 5 mA Max/Min Beam Current 10

MEBT Beamline

1/24/2017Vic Scarpine - MEBT Emittance Scanner10

Installed in Dec 2016 for horizontal emittance measurements• Rotate entire vacuum chamber for vertical measurements

Page 11: A “Short” Update on the P2IT MEBT...MEBT Allison-Style Emittance Scanner Parameter LEBT MEBT Kinetic Energy 30 keV 2.1 MeV Nominal Beam Current 5 mA 5 mA Max/Min Beam Current 10

First Beam Measurements Jan 6, 2017

1/24/2017Vic Scarpine - MEBT Emittance Scanner11

Page 12: A “Short” Update on the P2IT MEBT...MEBT Allison-Style Emittance Scanner Parameter LEBT MEBT Kinetic Energy 30 keV 2.1 MeV Nominal Beam Current 5 mA 5 mA Max/Min Beam Current 10

Summary

• First(?) high-energy Allsion-style emittance scanner has been

designed, constructed, commissioned and is making beam

measurements

• Calibration of scanner looks good

– Position and angle

• Many measurements taken of 2.1 MeV beam under a variety

of beam conditions – for future talk

– Used extensively for last two week of beam operations

• Special thanks to Matt Alvarez (Mech Eng) who's hard work

made this instrument possible

• Also thanks to all the work by Instrumentation, Mechanical

and Alignment departments and also all of the PIP2IT

department

1/24/2017Vic Scarpine - MEBT Emittance Scanner12