9081-13-000 bremer halo

12
Product Information BREMER HALO SYSTEM External Bracing

Upload: others

Post on 16-Apr-2022

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 9081-13-000 Bremer Halo

Product Information

BREMER HALO™

SYSTEM

External Bracing

Page 2: 9081-13-000 Bremer Halo
Page 3: 9081-13-000 Bremer Halo

1

BREMER HALO™

SYSTEM

External Bracing

Contents

Product InformationBremer Halo™ System 23-D Crown 3HIFix™ Skull Pin 4AirFlo™ Vest 5Bremer™ Classic II Vest System 6Paediatric System 6Adjusting Ring 6Universal and Trippi-Wells Tongs 7Kyphosis Vest 7

Page 4: 9081-13-000 Bremer Halo

2

ADVANCED TECHNOLOGY

WORKING FOR YOU AND

YOUR PATIENTS

The Bremer Halo™ System is an

advanced bracing system with

over 20 years of clinical usage.

Designed for security, strength,

and comfort, it offers ease of

application and cost-effective

treatment for many cervical

pathologies.

Bremer’s crown and vest systems

are fully MRI compatible up

to 1.5 Tesla and CT compatible

to the level of the crown.

BREMER HALO™ SYSTEM

Page 5: 9081-13-000 Bremer Halo

3

3-D CROWN

A FULLY ADJUSTABLE HALO CROWN

• Optimized anterior pin position – Movable anterior pin sliders

allow optimal medial/lateral pin positioning.

• Crown adjustments are more secure and efficient allowing:

a) Unlimited flexion/extension range.

b) Up to 127mm of A/P adjustment in normal cases with more

available for unusual circumstances.

c) Up to 127mm of distraction in normal cases with more

available for unusual circumstances.

• One size fits all – Posterior pin holders adjust to fit various headsizes allowing one crown to fit most adults.

• Optimized pin/skull angle – Anterior pin sliders available in 0°

and 12° angles to optimize pin/bone angle for straight or sloped

foreheads.

• Capital arch allows a strong, solid crown without additional

weight, as well as reducing posterior expansion.

• Torque Limiting Caps provide accuracy and ease of use,

simplifying and shortening crown application time.*

• Unique “hinged” design enables fast, efficient full chest CPR access.

• This unique, fixation system allows below the equator posterior

pin positioning, enhancing security.

*Note: For use on adult patients with sufficient bone quality only.

Page 6: 9081-13-000 Bremer Halo

4

HIFIX™ SKULL PIN*

IMPROVED HALO SKULL PINS

• The HIFix™ Skull Pin utilises a unique cutting head to

significantly reduce bone destruction.

• As a result, HIFix™ Skull Pins provide significantly increased

initial resistance to transverse loading and reduced skull pin

motion at the pin site as compared to the conventional Bremer™

skull pin, (as indicated in Table 1).

• Provides the opportunity for significantly improved skull

pin fixation1.

• Lightweight titanium alloy design increases strength and

reduces pin threading problems.*

• Knurled distraction now allows rapid, accurate Halo distraction.*

• Standard application procedure for ease of use.

*Note: For use on adult patients with sufficient bone quality only.

SKULL PIN MOTION AT PIN SITE**

HIFix™-type skull pin Standard skull pin HIFix™-type skull pin Standard skull pin

FAILURE STRENGTH OF SKULL PINS**

.43mm

3.15mm 2010N

1128N

Table 1** Halo pin loosening: a biomechanical comparison of experimental and conventional designs.

Journal of Biomechanics 31 (1998) pp397-400. Michael J. Voor, Cyna Khalily.

Page 7: 9081-13-000 Bremer Halo

5

AIRFLO™ VEST

A VERSATILE VACUUM FORMED HALO VEST

• Two sizes fit 98% of all adults, reducing inventory and simplifying

vest selection.

• Short vest length worn at or above the xiphoid, along with soft

medially placed shoulder straps allows enhanced patient range of

motion while reducing skull pin stress and cervical motion.

• The superstructure is highly adjustable, producing a wedge effect

which rests on major muscle groups. This offers exceptional

flexibility to meet patients’ needs while remaining lightweight

and strong.

• Specially engineered toothed joints are exceptionally strong and do

not come loose, improving patient safety and freeing the carer

from periodic retorquing of loose vest joints.

• Large posterior openings maximise ventilation, improving patient

comfort and reducing potential for skin problems.

• Roll-edge construction significantly reduces redundant tissue

capture, thereby improving patient comfort by allowing for a

lighter weight and smaller size, while improving vest strength.

• Lightweight and slim profile, providing improved patient comfort.

Page 8: 9081-13-000 Bremer Halo

6

BREMER™ CLASSIC II VEST SYSTEM

The Bremer™ Classic II vest system, compatible with all Bremer

Rings and Crowns as well as Ace Trippi-Wells* and Universal Tongs*,

is designed to fit most adults and provides a cost-effective alternative

to the AirFlo™ vest.

PAEDIATRIC SYSTEM

The Bremer™ Paediatric System, with two vest sizes and two

ring sizes, is designed to optimize halo traction for very

small children and infants.

ADJUSTING RING

The Bremer™ Adjustable Ring is compatible with the AirFlo™,

Bremer™ Classic II and Paediatric systems and provides a cost

effective alternative to competitive designs.

*Note: Requires Catalogue no CLP010

Page 9: 9081-13-000 Bremer Halo

7

UNIVERSAL AND TRIPPI-WELLS TONGS

The Ace tong systems are designed to provide maximum fixation for

traction while providing ease of use and the ability to be attached to

an Ace or Bremer Halo™* vest for mobile halo fixation.

KYPHOSIS VEST

The Bremer™ Kyphosis vest, allowing a closer approximation of

patient anatomy, is specially designed to provide ease of use and

flexibility of application for highly kyphotic patients.

*Note: Requires Catalogue no CLP010

Page 10: 9081-13-000 Bremer Halo

8

Notes

Page 11: 9081-13-000 Bremer Halo
Page 12: 9081-13-000 Bremer Halo

DePuy International LtdSt Anthony’s RoadLeeds LS11 8DTEnglandTel: +44 (113) 270 0461Fax: +44 (113) 202 5992

0086

DePuy AcroMedTM is a joint venture with Biedermann Motech GmbH.

BremerTM, Bremer HaloTM System, AirfloTM Vest, HIFixTM and the External Bracing icon are trademarks of DePuy AcroMed, Inc.

This publication is not intended for distribution in the USA.

© 2001 DePuy International Ltd. All rights reserved.

Cat No: 9081-13-000

Revised: 12/01

References

1. Voor, M.J., Khalily, C.,: Halo pin loosening: a biomechanical comparison of experimental and conventional designs, J Biomech 31: 4, 397-400 (1998).