MOTOR ARMATURE TESTER 6920
ENGLISH MANUAL
1st Oct 2008
MICROTEST
CATALOG 1. BEFORE USE.......................................................................................................................1
1.1 Product Introduction..........................................................................................................1 1.2 Product Characteristic .......................................................................................................1
2 SPEC INSTRUCTION.........................................................................................................2
3 SAFETY INSTRUCTION ...................................................................................................4
4 INSTALLATION .................................................................................................................6
4.1 System Setup Procedure....................................................................................................6 4.2 Front Panel Function .........................................................................................................7 4.3 Back Panel Function..........................................................................................................9 4.4 Fixture Panel Function ....................................................................................................11
5 OPERATION INSTRUCTION.........................................................................................12
5.1 Before Use.......................................................................................................................12 5.2 Function Instruction ........................................................................................................12
5.2.1 Function Enter Setting.............................................................................................12 5.2.2 Self-Testing ..............................................................................................................12 5.2.3 Fixture Pnumatic Control........................................................................................13 5.2.4 Temperature Testing Configuration ........................................................................13
5.3 System Instruction...........................................................................................................15 5.3.1 System Entering Setting ...........................................................................................15 5.3.2 System Environment Setting ....................................................................................15 5.3.3 Test Configuration Setting.......................................................................................16 5.3.4 System Time Setting .................................................................................................18 5.3.5 Password Setting .....................................................................................................19
5.4 File Management.............................................................................................................20 5.4.1 File Management Setting.........................................................................................20 5.4.2 Load File .................................................................................................................20 5.4.3 Create New File.......................................................................................................21 5.4.4 File Saving Instruction ............................................................................................21 5.4.5 Delete File ...............................................................................................................22
5.5 Setting Instruction ...........................................................................................................22 5.5.1 Test Setting Menu ....................................................................................................22 5.5.2 Fixture Setting .........................................................................................................23
MICROTEST
5.5.3 Resistance Test Setting ............................................................................................24 5.5.4 Insulation Test Setting .............................................................................................25 5.5.5 Hipot Test Setting ....................................................................................................26 5.5.6 Impulse (Surge) Test Setting....................................................................................27
5.6 Start Testing ....................................................................................................................33 5.6.1 Test Progress Procedure .........................................................................................33 5.6.2 Testing Message Analysis ........................................................................................35
6. CALIBRATION PROCEDURE .......................................................................................37
6.1 Enter Configuration Procedure........................................................................................37 6.2 High Voltage Output Calibration ....................................................................................38 6.3 High Voltage Output Testing ..........................................................................................38
6.3.1 High Voltage Output Setting....................................................................................38 6.3.2 High Voltage Value Setting .....................................................................................39
MICROTEST
1
1. BEFORE USE
1.1 Product Introduction
MT-6920R is a tester which exclusive made for Motor Armature tester and it includes 8 4-wire testing independent testing points. The testing items are including DC Resistor testing and Impulse / Surge Testing which can well examined your product and also speed up the testing.
1.2 Product Characteristic
Speedy, time saving, efficiency Programming control and analyze Simple editing and easy learning Simultaneously testing available Sole testing available Waveform display and storage Non-destructiveness analyzing Code function Error stop function Pass / Fail alarming Temperature compensation
MICROTEST
2
2 SPEC INSTRUCTION DC Resistor
Measuring Range 0.01mΩ~100KΩ
Basic Accuracy Low Resistance: 0.1mΩ~1Ω ± (0.2% ± 1mΩ) High Resistance: 1Ω~100kΩ ± 0.1%
Hipot
Measuring Voltage DC 500V ~ 3kV ±5%
Leakage Voltage
DC 0.1 ~ 10mA
Measuring Time
0.1 ~ 99.9 second continuously adjustable
Stable Accuracy
2%
Insulation Resistor
Measuring Voltage
100V ~ 1kV ± 5%
Measuring Range
1MΩ ~ 9.9GΩ
Measuring Time
0.1 ~ 99.9 second continuously adjustable
Stable Accuracy
2%
MICROTEST
3
Layer Short
Measuring Voltage 200Vp-p ~ 3kVp-p ±1%
Measuring Item
Waveform area, waveform area difference, waveform tremor, waveform comparison, flutter value
Stable Accuracy
2%
Other Spec
Measuring Pin
Depending on Samples
Display / Sound Device
Pass / Fail graphic display / Sound device
Power Supply
AC 220V / 110V ± 10% 50/60Hz
Temperature
10℃ 40℃
Humidity
Less than 60% relative
MICROTEST
4
3 SAFETY INSTRUCTION The instrument is not suitable using in outdoor especially in humid and dusty environment,
impropriate using may cause electronic shock. Before using the instrument, please read the safety
instruction completely in order to prevent any accident while using the instrument.
1. Security Signal
Caution
High voltage
Protect the GND terminal. Make sure the terminal connect the ground conductor before using the instrument
Warning. Improper use of instrument may cause the instrument and DUT
breakdown
Warning. Improper use of instrument may caused injured or death
2. Electronic Shock To avoid electronic shock, we strongly suggest to wearing rubber gloves before testing
3. Ground Conduction Please make sure ground conductor at the back panel is well connected. Do not touch
the exterior case to avoid any accident 4. Power
The instrument built with 110V / 200V AC power. Please switch the AC power at the back panel in order to match with the power which provide in the using area. Ensure that only fuses with require type and current can be replaced, also please take off the power before changing the fuses
Model Mark (Input Voltage Range) Fuse
6905 98~132V/192~264V 5A, 250V
6920 98~132V/192~264V 5A, 250V
6908 98~132V/192~264V 5A, 250V
MICROTEST
5
5. Connect testing cable with HV output port Please make sure the power is off when HV cable plug in the ports of the tester, and ensure that the exterior of cables are not broken or come off
6. Warm Up Warm up the machine at least 15 minutes in order to make the testing value accurate. The instrument can still work normally without any warm up
7. Exterior Control Do not touch HV port and DUT during operation
8. Breakdown Stop the operation and contact your agent or Microtest immediately while there’s abnormal
condition with the testers
9. End Up Turn off the power when the tester is not in use and do not turn the power on and off continuously
10. Placing The instrument must be placed under temperature 5°C~40°C and humidity 80%RH. Also, in order to make the instrument accurate and safe, please do not place the instrument under sunshine, high temperature, high humidity, unstable and dusty places
11. Emergency Turn off and power off if DUT or main instrument is burning
12. Generally Guide
(1) Do not put any inflammables or heavy articles above the instrument
(2) Do not let the instrument under any serious hit
(3) Do not turn on and plug in the power while cleaning the instrument
(4) Do not dismantle or fix the instrument while the instrument is under unusual
circumstances and contact your agent or Microtest Corp. directly
MICROTEST
6
4 INSTALLATION
4.1 System Setup Procedure
Start
6920 Setup
F6920 Fixture Setup
Measuring Items and
Golden Sample Setup
6908 and Accessories
Setup
Finish
Start Testing
MICROTEST
7
4.2 Front Panel Function
(1) Core 1: Testing Output Terminal for 1st of Fixture
(2) Core 2: Testing Output Terminal for 2nd of Fixture
(3) CH1: Testing Output Terminal for 6920 and 6908
(4) CH2: Testing Output Terminal for 6920 and 6908
(5) 1st Fixture Connector Slot: From the left to right order is 1 ~ 4 PIN.
(6) 2nd Fixture Connector Slot: From the left to right order is 5 ~ 8 PIN.
MICROTEST
8
※There is no 2nd Fixture Connector Slot without 48 Channels or Double Fixtures Option.
MICROTEST
9
4.3 Back Panel Function
(1) 6905 and 6920 HV Output Port (Connect by RED cables)
(2) 6905 and 6920 HV Input Port (Connect by BLACK cables)
(3) Temperature Sensor Port
(4) Solenoid Valve Connect Port for 1st Fixture
(5) Solenoid Valve Connect Port for 2nd Fixture
(6) RS-232 Port: Connect 6905 and 6920
(7) Printer Port
MICROTEST
10
(8) Remote Port (For Foot Switch)
(9) Ground Conduction Port
(10) Connect Ports for 6920 and 6908
MICROTEST
11
4.4 Fixture Panel Function
(1) DUT Diameter Adjuster: Use hexagon screw to adjust the elasticity of DUT
(2) DUT Installation: Place DUT @ no. 2. The Pin number is counted as counter
clockwise (Pin 1~Pin 24)
(3) Horizontal Adjuster: Make sure the probes are needled to the metal clips of DUT
(4) Solenoid Valve Port: Connect AIR 1 of 6908 to the fixture solenoid valve, and
connect to AIR 2 if it is the 2nd fixture.
MICROTEST
12
5 OPERATION INSTRUCTION
5.1 Before Use
1. The instrument built in 110V / 220V AC power. Please switch AC power at the back panel to match with the current in local area
2. Make sure the power is off when HV cables plug in the ports of the tester, also please ensure that exterior of cables are not broken
3. Warm up the testers at least 15 minutes in order to reach an accurate value. The instruments can still work normally without any warm up.
5.2 Function Instruction
5.2.1 Function Enter Setting
Press FUNC key under the main menu
Figure 5.2.1-1
5.2.2 Self-Testing
Press S1 under Function Menu and enter Self Test mode and start the self testing. Test items are CPU, RAM, ROM, EEPROM, CLOCK…etc. If test result is NG, power off the tester and contact your agent or Microtest Corp. to repair the instrument
MICROTEST
13
Figure 5.2.1-2
5.2.3 Fixture Pnumatic Control
Press S4 to enter Fixture Pnumatic Control. Select Fixture 1 or 2 to set up the power of the fixture
Figure 5.2.3.1
5.2.4 Temperature Testing Configuration
PessS4 under Function Menu and enter Temperature Testing Configuration mode. The function helps the testing progress under usual temperature and which make the test results more accuracy
MICROTEST
14
Figure 5.2.3.1
Configuration: Press S2 and manager code under Temperature Testing Configuration mode and start the calibration. Press S1 and S2 to adjust the value and make it same as the standard temperature
Figure 5.2.3-2
Fahrenheit / Centigrade: Press S3 under Temperature Test Configuration mode
to switch the temperature module.
MICROTEST
15
Figure 5.2.3.3
5.3 System Instruction
5.3.1 System Entering Setting
Press SYS at the main menu to enter System Setup Menu
Figure 5.3.1.1
5.3.2 System Environment Setting
Press S1 to enter System Setup. If want to modify the contents, use navigation key to moved the cursor and press S1 to modify the contents
MICROTEST
16
Figure 5.3.2.1
Line Frequency: Line Frequency can be switched from 50Hz to 60Hz
Tester ID No. : For the user who has several instruments. Default value is 1, the maximum value is 999, use numeric key to edit and press ENTER to finish setting
Key Lock: Keypad lock function
Warning! Key Lock function must key in secret code
LCD Contrast: Use S1 to adjust the brightness of LCD screen
Remote Output: Press S1 to setup control type, Standard, Extend, Unused
Power On Self-Test: Press S1 to Lock or Unlock the Power On Self-Test
WARNING: If self testing is NG, please turn off power immediately and contact
your agent or Microtest Corp. immediately
Language: Press S1 to change language
5.3.3 Test Configuration Setting
Press S2 to enter Test Configuration, use navigation key to move cursor and press S1 to modify the contents
MICROTEST
17
Figure 5.3.3.1
Temp. Correction: Press S1 to turn on / off the function
Warning: Please turn off Temperature Correction function during
configuration.
Test Alarm: Press S1 to select setup item as follow…
None No alarm whether the result is pass or fail
All Alarm when the results are pass and fail
Pass Alarm only DUT is pass
Fail Alarm only DUT is fail
Test Data Font: Press S1 to select the size of data font
Show Testing Wave: Press S1 to select the display of waveform; use S4 to enter
graphic setup
MICROTEST
18
Figure 5.3.3.2
Break On Failed Step: Press S1 to turn on / off the function. Stop testing when result is failed
Auto Print Test Data: Press S1 to turn on / off the function. Print out the information when the sample is fail
Print Test Data: Print out all information or fail information
Fixture Type: Press S1 to select Auto / Manual
Trigger Delay: Use numeric key to enter delaying time
5.3.4 System Time Setting
Press S3 under System Setting menu to enter System Time Setting
Figure 5.3.4.1
MICROTEST
19
5.3.5 Password Setting
Press S4 under System Setting and enter the password into a text box.
After enter the original code (6920), re-enter the new password (maximum 8 digits number) and re-enter the new numbers again to confirm the new password
Figure 5.3.5.1
Figure 5.3.5.2
MICROTEST
20
Figure 5.3.5.3
5.4 File Management
5.4.1 File Management Setting
Press FILE under Main Menu to enter File Management
Figure 5.4.1.1
5.4.2 Load File
Press S1 under File Management Menu to enter Load File. If file is using, the figure will shows “@” in front of file name. Use navigation key to move the cursor and press S1 to open the file.
MICROTEST
21
Figure 5.4.2.1
5.4.3 Create New File
Press S3 under File Management to enter Create New File. Use navigation key to move the cursor and press S3 to select the alphabet, press S4 to delete and key in the file name press S1 to save the file name.
Figure 5.4.3.1
5.4.4 File Saving Instruction
Press S4 under File Management and enter Save As. Use navigation key to move the cursor and press S3 to select the alphabets and press S4 to delete and key in the file name press S1 to save the file name.
MICROTEST
22
Figure 5.4.4.1
5.4.5 Delete File
Press S5 under File Management and enter Delete File. Use navigation key to move the cursor you want to delete and press S1 to delete.
Figure 5.4.5.1
5.5 Setting Instruction
5.5.1 Test Setting Menu
Press SET under Main Menu to enter Test Setting Menu
MICROTEST
23
Figure 5.5.1.1
Caution: Please be noted that the testing items for 6920 includes: Insulation, Hipot,
Surge, Resistance, O/S and more. Insulation and Hipot are only for the testing which connect with 6905, if the instrument does not connect with 6905, please turn off these 2 items.
5.5.2 Fixture Setting
Press S1 under to enter Fixture Setting mode
Figure 5.5.2.1
Fixture: Set up the numbers of fixture
Commutator Bars: Key in the numbers directly and press enter. The range is between 3~48 which is depends on the spec of fixture
MICROTEST
24
Fix 1 Start Channel: Key in the pin numbers directly before press enter. Values are 1, 7, 13, 19, 25, 31, 37 and 43, the values are correspond to the output ports of 6908 cable. There will be 4 sets of cable connectors if instrument is equipped with 24CH, the first pin of each cable connector corresponds to 1, 7, 13, 19 pin; if instrument is equipped with 48CH, there will be 8 sets of cable connectors, the first pin of each cable connector corresponds to 1, 7, 13, 19, 25, 31, 37, 43.
Fix 1 Start Channel: Key in the pin numbers directly before press enter. Values are 1, 7, 13, 19, 25, 31, 37 and 43, the values are correspond to the output ports of 6908 cable. Please be noticed the instrument 6908 you purchased is equipped with 24CH or 48CH.
Test Function Setting: Press Test under Fixture Setting Mode to enter Test Function Setting. Use Selec to choose the test item.
Figure 5.5.2.2
5.5.3 Resistance Test Setting
Press S2 under Test Function Setting to enter Resistance Test Setting
MICROTEST
25
Figure 5.5.3.1
Speed: Press Selec to choose testing speed FAST / MED / SLOW
Fail Restest: Use Selec to turn on / off the fucntion
R Limit: Use numeric key to enter the resistance limit which is the standard to
discriminate PASS / FAIL
Weld Limit: Use numeric key to enter the resistance limit which is the standard to
discriminate PASS / FAIL
Temp. Comp.: Use Selec to turn on / off the function. The function is transfer the
testing result to the value when tested under 20°c
Caution: Please turn off the function while using resistance calibration or DUT is not
made by copper
5.5.4 Insulation Test Setting
Press S3 under test Setting Menu to enter Insulation mode
MICROTEST
26
Figure 5.5.4.1
Voltage: Move the cursor to Voltage and use numeric key to set up voltage value, the maximum voltage is 1.0kV, the minimum voltage is 0.1kV
Dwell: High voltage output sequential times. Use numeric key to set up, maximum
value is 99.9 second, minimum value is 0.1 second
Max/Min: Use numeric key to set up values, the maximum value is 9999MΩ,
the minimum value is 0MΩ
Arc Sen.: The smaller value, the higher sensitivity. Use numeric key to set up
values, if value is 0 then the function is off.
5.5.5 Hipot Test Setting Press S4 under Testing Function Setting to enter Hipot Test Setting.
MICROTEST
27
Voltage / Frequency: Move the cursor to Voltage and use numeric key to set up
voltage value, the maximum value is AC 5.0kV, DC 6.0kV, the minimum voltage is 0.1kV. After the setting, press navigation key to move the cursor to frequency, use S1 to select frequency
Dwell: High voltage output sequential times. Use numeric key to set up, maximum
value is 99.9 second, minimum value is 0.1 second
Ramp: The maximum value is 99.9 second, the minimum value is 0.1 second
Max/Min: Use numeric key to set up values, the maximum value for AC is 10mA, DC
is 5mA, the minimum value is 0 for both AC and DC
Arc Sen.: The smaller value, the higher sensitivity. Use numeric key to set up
values, if value is 0 then the function is off
Offset: To deduce the leakage voltage which made by testing equipment or fixtures. Use numeric key to enter the leakage voltage value and tester will deduce the error automatically
5.5.6 Impulse (Surge) Test Setting
Press S5 under Testing Setting mode to enter Impulse / Surge Setting
Figure 5.5.6.1
Bar +/-: Move the cursor to Voltage and use numeric key to set up the bar channel.
Bar +/- cannot use the same channel.
Voltage: Move the cursor to Voltage and use numeric key to set up voltage value, the maximum voltage is 3.0kV, the minimum voltage is 0.2kV
Mode: Impulse / Surge Testing is tested by waveform comparison. Move cursor to
MICROTEST
28
press selec to choose the mode
Learn: Impulse / Surge Testing is tested by waveform comparison, according to
this, the instrument has to learn standard waveform to compare waveform
of DUT. Move cursor to Learning, after entering, connect standard sample with
Hipot output terminal and press Learn. If the waveform is too wide or too
narrow, press S2 to widen or S3 to narrow, or move the cursor to width and key
the value directly. Under this mode, it can also adjust voltage and testing mode
after learning. After back to Impulse / Surge testing after learning, press Exit
to exit.
Figure 5.5.6.2
Figure 5.5.6.3
MICROTEST
29
Area Size: This item is to check the difference between testing waveform and standard
waveform. Move the cursor to Area Size and use numeric to set up
tolerance rate and then press TEST to enter Testing Mode (please note that
the instrument need to learn standard waveform before enter this mode) and
use S4 ~ S5 to set up the range. After the setting, press TEST to start
pre-test while there’ll be a difference value displayed at the right corner
and then press EXIT to exit the setting if all values are correct.
Figure 5.5.6.4
Figure 5.5.6.5
Differential Area Size: Move cursor to differential area size and press S1(testing
option), and use numeric key to enter tolerance rate. Press
TEST to enter Testing Mode and use S4~S5 to set up testing
MICROTEST
30
range. After the setting, press TEST to start the pre-test.
In the meantime, there’ll be a difference value displayed at
the right corner, and then press S6 to exit the setting if all
the values are correct.
Figure 5.5.6.6
Figure 5.5.6.7
Flutter Value: Move cursor to Flutter value and press Selec (Test Selecting) key, use
numeric key to enter standard values, if the test result is larger than the
value, the result is fail. Press Test to enter Testing Mode and
use S4~S5 to set up testing range, after the setting, press Test to start
pre-test. In the meantime, there’ll be a flutter value display at the right
corner to determine if the results are in the range, press Exit to exit
MICROTEST
31
setting.
Figure 5.5.6.8
Figure 5.5.6.9
Waveform Comparison: Move the cursor to Waveform Comparison and press Selec
(testing option), and use numeric key to enter tolerance rate.
Press Test to enter Testing Mode and use S4~S5set up testing
range, after the setting, press Test to start pre-test. In the
meantime, the result will display at the right corner, press Exit
to exit.
MICROTEST
32
Figure 5.5.6.10
Figure 5.5.6.11
Corona Value: Move the cursor to Corona Value and press Selec(Testing Option), using
numeric key to enter the standard value. Press Test to enter Pre-Testing
Mode and then press Test to start the pre-test, in the meantime there’ll be
a corona value displayed at the right corner and then press Exit to exit
the setting mode if all values are correct.
MICROTEST
33
Figure 5.5.6.12
Figure 5.5.6.13
5.6 Start Testing
Put DUT on the fixture after confirm there’s no voltage output and the indicator light of voltage is not lightening
5.6.1 Test Progress Procedure 1. Press TEST under Main screen. After see READY TO TEST, press TRIG to start the
combination testing
MICROTEST
34
Figure 5.6.1.1
Figure 5.6.1.2
2. When press TRIG, Hipot will be outputted and the indicator light will be lightened. After testing,
the test results will be shown on the screen also the Test Completed description
MICROTEST
35
Figure 5.6.1.3
3. If Impulse / Surge short is opened, press S6 after the testing in order to check the waveform and related values. Press S6 again to check another testing result
Figure 5.6.1.4
4. If the test result is failed, it will be marked to let the operator easy to identify
Figure 5.6.1.5
5.6.2 Testing Message Analysis
Hipot
Test Result Description
HI Current is over the standard value
LO Current is below the standard value
MICROTEST
36
BREAK DOWN DUT Hi-Pot is fail
ARCING DUT have corona value
LINK ERR De-connected with 6905
Insulation
Test Result Description
HI Current is over the standard value
LO Current is below the standard value
BREAK DOWN DUT Hi-Pot is fail
ARCING DUT have corona value
LINK ERR De-connected with 6905
Resistance Coils and Balance
Test Result Description
HI Current is over the standard value
LO Current is below the standard value
MICROTEST
37
6. CALIBRATION PROCEDURE DC calibration available, please ensure to warm up the instrument at least 30 minutes in order to make the testing value correct.
6.1 Enter Configuration Procedure Press FUNC key under main menu and then press S2 to enter H/V CAL. The password must be keyed in when using H/V CAL
Warning: When using the function, please do not touch output terminal or connect any
load into the output terminal
Figure 7.1.1
Figure 7.1.2
MICROTEST
38
6.2 High Voltage Output Calibration
Use S1 and S2 to adjust the voltage, and please adjust the value of CODE and READING the same if possible. After the adjustment, press ENTER to move next time
Figure 7.2.1
UP: Press S1 to increase CODE value
DOWN: Press S2 to decrease CODE value
SAVE: Press S4 to save CODE
DEFAULT: Press S5 to reset default
QUIT: Press S6 to quit the calibration
6.3 High Voltage Output Testing
Connect the positive terminal of DC meter to CH7 of the instrument and negative terminal to CH8. The result should be minus value due to the voltage is negative while testing
6.3.1 High Voltage Output Setting
Press FUNC under the main menu to enter System mode, and then press S3 to enter H/V Test mode
MICROTEST
39
Figure 7.3.1.1
6.3.2 High Voltage Value Setting
Use numeric key to setup high voltage value. High voltage range is 0.2kV~5KV.
If the setting value smaller than 0.2kV and larger than 5KV, the instrument will
insert 0.2kV and larger than 5KV, the instrument will insert 0.2kV and 5KV
automatically
ON: Press S1 to show setting voltage
OFF: Press S2 to close output voltage
UP: Press S3 to increase voltage value 0.1kV
DOWN: Press S4 to decrease voltage value 0.1kV
EXIT: Press S6 to exit voltage testing mode
MICROTEST
40