490series_completemanual
TRANSCRIPT
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 1/21
490 Series
Biological Microscope
OPERATOR’S MANUAL
READ THIS MANUAL BEFORE USING THE MICROSCOPE
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 2/21
2
Contents 490 SERIES
Before use ..............................................................................................
1. Nomenclature ....................................................................................
2. Operation ............................................................................................
2-1 Assembly ...........................................................................2-2 Setting up ................... ................... .................... ................
2-3 Focusing .............................................................................
2-4 Adjusting the condenser and diaphragm .............
2-5 Using the trinocular port ................................. ............
2-6 Setting the stage’s stop limit ............................ ..........
2-7 Adjusting focusing tension ................. .................... ....
2-8 Changing the bulb .................... .................... .................
2-9 Changing the fuse .................. .................... ................... .
2-10 Maintenance ....................................................................
3. General specications .................. .................... ................... ...........
4. Conguration drawing ................. .................... ................... ...........
5. Optional Accessories ................... ................... .................... .............
6. Parameters/technical terms ................... .................... ..................
6-1 Objectives ..........................................................................
6-2 Eyepieces ...........................................................................
6-3 Electrical system ................. .................... ................... ......
6-4 Parameters ........................................................................
6-5 Technical Terms ................................... ................... .........
7. Troubleshooting ...............................................................................
3
4
8
89
9
10
11
11
12
12
12
13
14
15
16
17
17
17
18
18
18
20
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 3/21
3
Before use 490 SERIES
IntroductionCongratulations on the purchase of your new AmScope Microscope!
This manual is designed for the B490 and T490 series microscopes.
Though both have identical bases, the B490 is equipped with a
binocular head (2 eyetubes) and the T490 is equipped with a trinocular
head (2 eyetubes & 1 photo/video port). Please take a few minutes to
familiarize yourself with the features and functions of your new
microscope. If you want more information on microscope, parts, and
accessories, please visit our website at:
www.AmScope.com
We recommend that you study this manual thoroughly before operat-
ing the microscope, and that you keep it on hand for future reference.
If you have additional questions or need assistance, please send us an
email at:
Please include the microscope model number in your email so that wecan identify your model and provide immediate help.
Precautions(1) As the microscope is a precision instument, always handle it with
care, avoiding impact or abrupt movement during transportation. Do
not shake the package.
(2) Do not place the microscope in direct sunlight or in high heat. Keep
it indoors in a dry and clean place with temperatures between 32°-100°
F (0°-40°C), maximum relative humidity: 85%
(3) Avoid touching the lenses on the objective and the eyepiece so thatoil and dirt from your ngerprints do not obstruct your view.
(4) Before turning the power on, make sure that the power supply
voltage is consistent with the voltage of your microscope
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 4/21
4
1. Nomenclature 490 SERIES
Nosepiece
Objective
Mechanical Stage
Condenser,iris diaphragm
Base lens
Base
Eyepiece
Fig. 1: B490
Diopter
Finefocusing knob
Coarsefocusing knob
Arm
BinocularGemel head
Iris adjustment slider
Filter holder
Slide controls
Power switch
Dimmer
Tension knob
Eyetube base
Head -lock screw
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 5/21
5
1. Nomenclature 490 SERIES
C-mount
Fig. 2: T490
Adjustabletrinocular port
TrinocularGemel head
Condenseradjustment knob
Limit-stopknob
Coarsefocusing knob
Finefocusing knob
Condenserscrew
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 6/21
6
1. Nomenclature 490 SERIES
Fig. 3: Bottom
Fig. 4: Condenser, iris and lter holder
FuseLight screw
Halogen
bulb
IrisFilter holder
Condenserscrew Condenser
Iris adjustmentslider
Denitions:
Base Lens Directs the light source towards the slide
Coarse Focusing Knob Used to initially bring the slide into sight and
focus
Adjustable Trinocular Port Threaded port with adjustable length forattachment of digital or video camera
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 7/21
7
1. Nomenclature 490 SERIES
Condenser & Iris Diaphragm
Controls the amount of light that hits the slide
Condenser Adjustment Knob
Moves the condenser up and down
Condenser Screw Screws the condenser in place, secureing it
to the mechanical stage
Dimmer Controls the amount of light that escapes
from the base lens
Diopter Allows the focus to be perfected for both eyes,
independent of each other
Fine Focusing Knob Used to rene focus to a clear image
Head Lock Screw Screws the microscope head into place
Limit-Stop Knob Limits the upward movement of the mechanical
stage in order to avoid damaging the slide and
objective (adjustable).
Mechanical Stage Mechanically moves the slide along an X and Y
axis for optimal positioning
Nosepiece Houses the objective lenses
Rotating Gemel Head Adjusts eyepieces to t the distance between
the user’s eyes for added comfort
Tension Knob Adjusts the tension of the focusing knobs
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 8/21
8
2. Operation 490 SERIES
2-1 Assembly(1) First take the styrofoam container out of the cardboard carton and lay
it on its side, paying attention to which side is labeled up. Remove tape
and open the container carefully so as to avoid dropping and damaging
the optical items. Check carefully to ensure that all parts and accessories
are intact.
(2) Check the parking list below:
One Microscope Body
One Binocular (B490) or Trinocular (T490) Rotating Gemel Head
One Adjustable Trinocular Photo Port (for T490 only)
Four Objectives: 4X, 10X, 40X, and 100X
Two Pairs of Eyepieces: W10X and P16X
Color Filters
One Bottle of Immersion Oil
One 6V/20W Halogen Spare Bulb
One Spare FuseOne Dust Cover
(3) Remove the microscope body from the box and remove the plastic
protective covering. The body of the microscope is composed of the
base, the stage, the arm, and the nosepiece.
(4) Loosen the metal knob located directly above the nosepiece (the
head-lock screw) completely. Place the head, at circular side down, onto
the circular opening at the top of the arm. Make sure the microscope
head is securely on top of the arm and tighten the head-lock screw to
lock the head in place.
(5) For the T490 model only, screw the photo port onto the top of the
microscope head.
(6) Remove the eyetube caps and drop the desired eyepieces into the
eyetubes without touching the lens.
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 9/21
9
2. Operation 490 SERIES
(7) Screw the objectives into the microscope nosepiece from the lowest
magnication to the highest, being careful not to touch the lens.
(8) Plug in the microscope and turn it on. If no light emerges from the
light source, adjust the dimmer knob on the side of the base.
2-2 Setting up(1) Loosen the head-lock screw and adjust the microscope head so that it
is in a comfortable position for observation. Lock the head-lock screw.
With both eyes open, look into the eyepieces. Adjust the interpupillary
distance by holding the eyetubes and rotating the eyepieces either
towards or away from one another, until one circle of light can be seen.
(2) Place the specimen to be studied on a glass slide (or simply use a
prepared slide), and place it on the stage, holding it sungly in place with
the metal slide-holders of the mechanical stage. Using the slide controls,center the specimen over the stage opening, lining it up with the light
and objective lens.
(3) To adjust the illumination, slowly turn the dimmer until the desired
intensity of light is achieved.
2-3 Focusing(1) Turn the nosepiece to choose an objective. It is easiest to use the
lowest magnication rst (usually 4X) to locate and focus the specimen.
As you move up in magnication, you may have to refocus the image a
little bit each time. When using the 100X objective, a little immersion oil
must be put between the objective and the cover glass on the slide in
order to minimize the distortion cause by air. Always wipe the oil o the
objective with either a glass-cleaning paper or cloth after each use.
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 10/21
10
2. Operation 490 SERIES
(2) Begin focusing by rst looking with one eye through the eyetube
without the diopter. Close your other eye. Loosen the limit-stop knob
(located on the inside of the coarse-focusing knob) in order to give the
mechanicalstage a full range of movement and adjust the coarse-
focusing knob to bring the slide into focus. You may also need tore-center the specimen on the stage.
(3) When the object is in your eld of vision, adjust the ne-focusing
knob until you see a clear, sharp image. BE ALERT: When moving the
mechanical stage vertically, be careful not to hit the objective lens with
the slide. Remember to use the limit-stop knob to avoid impact between
the objective and the slide when using the 40X and 100X objectives.
Lock the limit-stop knob at the position where the slide almost touches
the objective.
(4) Once the image is focused through the one eyepiece, look through
the other eyepiece and turn the diopter until the image is clear throughthat eyepiece as well
2-4 Adjusting the condenser and diaphragm(1) Using the condenser-adjustment knob, you can change the distance
between the light condenser and the stage. This allows you to control
the concentration of the light hitting your slide.
(2) By Changing the aperture of the iris diaphragm, you can adjust the
background brightness. Adjust the aperture of the iris diaphragmm
using the iris adjustment slider located directly under the stage.
(3) If you want to use a color lter, swivel the lter holder out, towards
you. You can now place the desired color lter into the circular opening.
Slide the lter holder back to original position before observation.
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 11/21
11
2. Operation 490 SERIES
2-5 Using the trinocular port The AmScope T490 is uniquely designed so that you can focus the
image through the eyepieces and the trinocular port simultaneously.
This “parfocal” feature allows the images through the microscope’s
eyepieces, and those displayed on your computer screen or TV to be
viewed at the same time! You do not need an adapter to attach your
AmScope camera to the trinocular port. However, you may need an
adapter if you have a non-AmScope camera.
Attaching a camera:
(1) If your camera has a C-mount, simply screw the camera onto the
trinocular port.
(2) If your camera has a 23mm mounting size, remove the C-mount from
the top of the trinocular port by loosening the screw. Then, drop the
camera into the trinocular tube.
(3) To focus through the trinocular port, simply turn the middle portion
of the tube.
2-6 Setting the stage’s stop-limit(1) Set the stage to the maximum height desired.
(2) Pull the stop-limit knob towards the stage. The stage’s stop-limit is
now set, allowing you to move the stage down, but not up past this
point.
(3) To reset the stage’s stop-limit, simply push the stop-limit knob back
towards the back panel of the microscope. The stage’s full range of
motion is now restored.
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 12/21
12
2. Operation 490 SERIES
2-7 Adjusting focusing tension(1) To adjust the tension of the focusing knobs, rst locate the black
ridged tension-knob on the inside of the coarse focusing knob.
(2) To decrease tension, rotate the tension adjustment forward, towards
the stage. To increase tension, rotate the tension adjustment backwards,
towards the back panel of the microscope.
2-8 Changing the bulb(1) Before changing the light bulb, rst pull the plug out of the electrical
socket and wait for the lamp to cool down in order to avoid being
burned. Also, make sure to remove the eyepieces or secure them to the
microscope before turning the microscope over!
(2) Loosen the door-screw on the bottom of the microscope and pullopen the small door. Pull the old bulb out and replace it with a new one.
Close the door and tighten the screw.
2-9 Changing the fuse(1) First, remove the eyepieces or secure them to the microscope before
turning the microscope over.
(2) Loosen the door-screw on the bottom of the microscope and pull
opUsing a at-head screwdriver, unscrew the fuse located on the
bottom of the microscope.
(3)Remove the old fuse and replace with a new one. Screw the fuse back
in place.
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 13/21
13
2. Operation 490 SERIES
2-10 Maintenance
(+) All glass surfaces must always be kept clean. Fine dust on the
optical surface should be blown o using a hand blower or gently
wiped o with a soft lens tissue. Carefully wipe o oil or ngerprints on
lens surfaces using tissue moistened with a small amount of 3:7 alcohol
to ether mixture
(+) Do not use this solution to wipe the surface of other components.
These parts, especially those made of plastic, should be cleaned with a
neutral detergent.
(+) Do not assemble or disassemble the microscope’s electrical compo-
nents yourself.
(+) After use, cover the microscope with the provided dustcover. Make
sure to keep it in a dry and clean place in order to prevent rust.
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 14/21
14
3. Specications 490 SERIES
Specications for the 490 Series Biological Microscope
PARTS SPECIFICATIONS WF10X/18mm
WF10X/18mm with pointerWF10X/18mm with reticle
WF16X/13mm
WFEyepiece
WF20X/10mm
X5P
X01P
PlanEyepiece
X61P
4X/0.10
10X/0.25
40X(spring)/0.65
60X(spring)/0.85
DINAchromaticObjectives
100X(spring, oil)/1.25
X4
X0140X(spring)
PlanObjectives
100X(spring, oil)
Binocular sliding head, 360° swiveling45° ViewingHead Trinocular sliding head, 360° swiveling
Binocular Gemel head, 360° swiveling30° ViewingHead Trinocular Gemel head, 360° swiveling
TrinocularPort
C-mount adjustable photo port.
Simultaneous Viewing
Focusing Coaxial focusing system
Nosepiece Quadruple nosepiece
Stage130mm x 140mm mechanical stage
Moving range: 50mm x 76mm
Condenser Abbe, NA=1.25Diaphragm Iris diaphragm
Halogen light with dimmerIllumination
ф50mm mirror
Lamp 6V/20W
Filter Blue/yellow/green
B490 T490
23mm adjustable photo port.
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 15/21
15
4. Conguration drawing 490 SERIES
C C D
V I D E O
C A M E R A
4 0 0 P H
4 9 0 T
H 3 0 G
4 0 0 T H 4 5
4 9 0 B H 3 0 G
4 0 0 B H 4 5
4 0 0 M H 4 5
4 0 0 M S
4 0 0 B L
4 0 0 M R
4 0 0 N P 4
4 0 0 F B
A
1 0 0 X
A 4 0 X
A 1 0 X
A 4 X
P A 1 0 0 X
P A 6 0 X
P A 4 0 X
P A 1 0 X
P A 4 X
W F 1 0 X 1 8
W F 1 0 X 1 8 P
W F 1 5 X 1 6
W F 2 0 X 1 0
E P 5
E P 1 0
E P 1 6
4 9 0
S e
r i e s
B i o l o g i c a l M i c r o s c o p e
C o n g u r a t i o n d r a w i n g
4 0 0 C D 1 . 2
5
D R Y 1 0 0
O I L 1 0 0
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 16/21
16
5. Optional Accessories 490 SERIES
NAME DESCRIPTION MODEL # PURPOSES5X EP5X23 For obtaining 20X, 50X, 200X
and 500X magnicationpowers
20X EP20X23 For obtaining 80X, 200X, 800Xand 2000X magnicationpowers
25X EP25X23 For obtaining 250X and 2500Xmagnication powers
10X with pointer EP10X23P For easier identifying of objects
Eyepiece
10X with reticle EP10X23R For measuring objects
2X A2X For obtaining 20X and 32Xmagnication powers
5X A5X For obtaining 50X and 80Xmagnication powers
20X A20X For obtaining 200X and 320Xmagnication powers
60X A60X For obtaining 600X and 960Xmagnication powers
Plan 4X PA4X
Plan 10X PA10X
Plan 40X PA40X
Objective
Plan 100X PA100X
For obtaining high quality highresolution images
Dry DarkeldCondenser
DK-DRY100 For obtaining low powerdarkeld imagesDarkeld
Condenser Oil DarkeldCondenser
DK-OIL100 For obtaining high powerdarkeld images
CMOS Digital
MD400 (350K)MD600 (1.3MP)MD700 (2.0MP)MD800 (3.0MP)MD900 (5.0MP)
For obtaining still or live digitalimages
CCD Digital MD1000 For obtaining high quality stillor live digital images
Camera
CCD TV Video CCD-NP
Optional Accessories There is a variety of accessories available that can enhance your
microscope’s capabilities and give it additional features. Some of them
are listed in the table below:
For live images on TV
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 17/21
17
6. Parameters/Technical terms 490 SERIES
Aluminum case AC-400 For carrying microscope aroundCase
Hard wood case WC-400 For storage
Simple PCS Phase contrast imagesPhaseContrast Kit Turret PCT High quality phase contrast
images
Stage Warmer TCS-100 For controlling stagetemperature
FluorescentKit
FK-YX Fluorescent images
Please nd more accessories on our website at: www.AmScope.com
6-1 Objectives
6-2 Eyepieces
Type MagnicationNumericalAperture
(N.A.) Medium
Par-focalDistance
(mm)
MagnicationMarks
(color ring)
4X 0.10 Air 45 Red10X 0.25 Air 45 Yellow
40X 0.65 Air 45 Light blue
60X 0.85 Air 45 Deep blue
DINAchromaticObjective(195mm)
100X 1.25 Cedaroil 45 White
4X 0.10 Air 45 Red
10X 0.25 Air 45 Yellow
40X 0.65 Air 45 Light blue
PlanObjective(195mm)
100X 1.25Cedar
oil45 White
Type Wide eld eyepiece Plan eyepiece
Magnication 10X 15X 20X 5X 10X 16X
Field of view(mm)
ф18 ф13 ф11 ф18 ф18 ф15
Optional Accessories cont’d...
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 18/21
18
6. Parameter/technical terms 490 SERIES
6-3 Electrical system There are two options for electrical systems. Both are equipped with
6V/20W halogen lamps.
(1) 220V∼240V power supply: 220V∼240V ±10%, 50Hz
This electrical system is CE and GS certied
(2) 100V∼120V power supply: 100V∼120V ±10%, 60Hz
This electrical system is UL certied
6-4 Parameters(1) Total magnication: 20X∼1600X
(2) Field of view: ф0.8mm∼ф4.5mm
(2) Mechanical tube length: 160mm
(4) Object to primary image distance: 195mm
(5) Fine focusing sensitivity: 0.002mm
6-5 Technical terms(1) Total magnication= magnication of objective multiplied by magni-
cation of eyepiece
(2) Field of view= line field of view of the eyepiece divided by themagnification of the objective
(3) Object to primary image distance= n·sinα (max), N.A. is an important
parameter that marks the features of the objective and condenser. The “n”
is the refractive index of the medium (air or immersion oil) between the
·
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 19/21
19
6. Parameter/technical terms 490 SERIES
objective lens and the specimen. The “α ” is 1/2 the aperture angle.
The larger the N.A., the high the resolution of the objective
(4) Object to primary image distance = the distance between the
object plane to the primary image plane. The conjugate distance isxed
(5) Mechanical tube length = the distance between the objective
shoulder and the ocular shoulder
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 20/21
20
7. Troubleshooting 490 SERIES
Troubleshooting for the 490 Series Biological Microscope
O P T I C A L I S S U E S
The nosepiece is misaligned
There is no cover glass on the slide
The cover glass is not standard
The specimen is upside down
SYMPTOM CAUSE REMEDY
One side of theeld of view is
darker
Turn the nosepiece until it
clicks into place
Stain or dust has accumulated on the
condenser, object, eyepieces, or base
lens
Obstructions
are observed in
the eld of
view
Clean the lens
Stains has accumulated on the
specimen
Stains have accumulated on the lens
Clean the specimen
Clean the lens
Unclear
Image
The immersion oil has accumulated on
the dry objective
No immersion oil used with the 100Xobjective
Air bubble in the immersion oil
Used wrong oil
The aperature is not open to an
appropriate diameter
Stain or dust has accumulated on the
lens in the inlet of the head
The condenser is not in the r ight position
Add the cover glass
Use glass w/0.17mm thickness
Flip it over
Clean lens thoroughly
Use immersion oil
Pop the air bubbles
Adjust the iris diaphragm
Use standard cedarwood oil
Clean the lens
Adjust the condenser
One side of the
eld of view is
dark or the
image moves
while focusing
The eld of
view is not
bright enough
The specimen slide is not xed
The nosepiece is not in the right position
The iris diaphragm is not big enough
The condenser is not in the right position
Stain or dust has accumulated on the
condenser, objective, eyepieces, or base lens
Secure slide to the stage with
clips
Turn the nosepiece until it
clicks into place
Adjust the diaphragm
Adjust the condenser
Clean the lens
8/2/2019 490Series_CompleteManual
http://slidepdf.com/reader/full/490seriescompletemanual 21/21
21
7. Troubleshooting 490 SERIES
Troubleshooting for the 490 Series Biological Microscope cont’d...
O P T I C A L I S S U E S
The mechanical stage is not properly secured
Tension knob is too tight
SYMPTOM CAUSE REMEDY
The color of the
image is notaccurate
Adjust the brightness-
adjustment knob
No lter is used
The objective
touches the
cover glass
Use desired lter
The slide is not secured correctly Adjust the slide
The coarse-focusing knob
won’t raise the
stage
The tension knob is too loose
Limit-stop knob locked
Limit-stop knob locked
Tighten the mechanical stage
Loosen it a little
Tighten it a little
Loosen the limit-stop knob
and lock it after refocusing
The
microscope
does not light
up
The bulb burns
out frequently
The bulb is not inserted correctly
The bulb burned out
The fuse burned out
The voltage is too high
Used wrong bulb
Insert it correctly
Replace the bulb
Replace the fuse
Use correct power supply
Replace w/correct bulb
The brightness-adjustment knob is not
in the right position
M E C H A N I C A L I S S U E S
The cover glass is upside down
The cover glass is not standard Use glass w/0.17mm thickness
Flip the cover glass over
Unable to
move the slide
smoothly
The microscope is unplugged
Check the power cable ‘s plug
Focus-knob
does not turn
Stage declines
by itself
The ne-
focusing knob
won’t raise the
stage
Loosen the limit-stop knob
and lock it after refocusing
E L E C T R I C A L I S S U E S
The bulb
ickers The bulb is close to burning out Replace the bulb
The fuse burns
out frequently The voltage is too high
Use the correct power supply
or get a voltage adapter