4 review of literature
TRANSCRIPT
Review Of Literature
Page 6
HERBERT D. AYERS et al (1960)12 conducted a detail duplication test to
evaluate elastic impression materials. Conclusions from the study were that a difference
can be demonstrated in the ability of various impression materials to reproduce fine
surface detail. Polysulfide rubbers, hydrocolloid and silicone materials appear to be
excellent, while alginates are definitely inferior both in their inherent behavior and in
combination with the three die materials tested. The ability of the impression material to
register detail from the original die exceeded the reproduction capabilities of the gypsum
cast material tested. A difference was noted in the behavior of the stones tested in regard
to reproduction of small detail.
BERGMAN, JAN OLSSON, BERGMAN (1980)13 conducted a study to
determine the influence of disinfectants on the dimensional stability, surface detail and
accuracy of elastomeric impression materials. The results showed, in some cases, effect
of certain disinfectants during the impression material disinfection procedures. The shape
of specimen was built in Teflon mould. On its surface, spiral grooves were marked to
record the measurement variations. Each specimen was analyzed under the ZEISS
microscope under controlled humidity and temperature. A total of one hundred and
twenty specimens divided into groups of fifteen, were made for each type of material
(Permelastic Light, Regular and President Viscous Light, Regular, Heavy and Viscous
Reprosil Light, Regular and Sticky; Optosil and Xantopren, and Impregum). Specimens
from the control group were dipped in distilled water for one hour, then washed in
distilled water for 1 minute, dried in an air jet and measured. Then the specimens were
left on a plate-glass for 24 hours before measuring again. These measures were also made
for the other seven groups of immersions in seven different disinfectant solutions. 3
Review Of Literature
Page 7
levels of surface change were determined: 1. the visible change, 2. slight change of the
surface and 3. Significant change. Two observers classified the specimens with
agreement of 94%. This work showed that the differences found in linear dimensions of
all the materials studied are within the margin of measurement error for immersion times
of 1 hour and thereafter 24 hours.
MC’CABE (1981),14 conducted a study in which he immersed Silicones,
Polyether, Polysulfide, reversible hydrocolloid and irreversible hydrocolloid in solutions
of 1%Sodium Hypochlorite, 2% Alkaline Glutaraldehyde and 4% Formalin for 16
hours. The evaluation of the materials for dimensional stability showed that
1. Silicones can be disinfected by this method in any of the solutions.
2. Polysulfide showed better stability in a solution of 2% glutaraldehyde.
3. Polyether showed significant dimensional change in aqueous solutions of both
glutaraldehyde and formaldehyde.
4. Alginate did not change significantly only in solution of 1% sodium hypochlorite
RHODES et al. (1985), 15 studied the effects of glutaraldehyde based solutions
on commercially available elastomeric materials. The objective of this investigation was
to evaluate the effects of solutions with different pH on the physical properties of
polysulfide, silicone [addition and condensation] and polyether. The results indicated that
the times and dilutions tested did not cause significant changes in any of the tested
materials.
Review Of Literature
Page 8
SETCOS, J.C et al. (1985)16 studied the effects of 12 disinfectant solutions on
the linear dimension of the polyether Impregum. They made impressions of a metallic
standard model, then it was washed, shaken and placed under water for 15
minutes. Models were cast in stone after this time, except for the control group which
was built immediately after molding. Neither solution was capable of causing change in
material in the time tested, indicating this method of disinfection that can be used without
restrictions.
JOHANSEN, STACKHOUSE JR (1987)17 conducted a study to measure and
compare the linear dimensional change of elastomeric impression materials submitted to
immersion in solution of 2% glutaraldehyde. The authors concluded that polyether stay
stable on the bench, but expand when immersed in solution considerably. Polysulfide and
condensation silicone contract 0.3 to 0.4% during 16 hours of immersion, not showing
any significant difference from the control group.
WATKINSON (1988)18 studied the routine impression disinfection procedures in
40 Departments, 15 undergraduate schools / British graduate schools. Only in 15 of the
40 Departments use was made of disinfecting agents for impression disinfection, while in
18 departments no disinfection procedure was used. In the other seven, the impressions
were washed with water to reduce surface contamination.
TULLN, COMMET; MOON (1988)19 conducted a study to determine which
specific disinfectant solution should be used for immersion disinfection of the impression
materials; that caused least significant dimensional changes in impression. An acrylic
master model was constructed to represent half of the mandible from central incisor to
Review Of Literature
Page 9
second molar. Points of reference for the model were constructed of stainless steel. Trays
were made with individual relief of 2mm to 6 mm for elastomers and hydrocolloids. The
impression materials tested were Polysulfide, Polyether, Addition Silicone and
Alginate. After 15 minutes of immersion in solutions of Iodoform, Sodium Hypochlorite
or Neutral Glutaraldehyde, templates were washed, dried and poured immediately, except
for Polysulphides [waited 1 hour for pouring]. The impressions from control were
washed, dried and immediately poured without going through any kind of treatment.
After waiting an hour for the casts to set completely, the model was separated from the
mold and measured 5 times each, in an optical microscope with an accuracy of 0.0001
inch. Only four measures showed statistical significance of dimensional changes, though
these distances are not clinically relevant. For the author, a change of 0.31%, as found in
addition silicone and in alginate, when projected on a preparation of 5 mm, which means
only 15 microns in marginal adjustment of the crown. This is not significant compared to
25 microns required by the film of cement specified by the ADA.
JAMES A. COTTONE, et.al (1990)9 conducted a study to determine the efficacy
of disinfection / sterilization protocols recommended by manufacturers of impression
material. Minimum distortion of poly sulphide rubber base material occurs with 2% acid
glutaraldehyde, 0.5% or 1% sodium hypochlorite, or 0.1% povodine iodine and
concluded that none of impression manufactures had an appropriate, complete
disinfection protocol.
PEUTZFELDT et al.(1990)20 studied the effect of immersion in disinfectant
solution on the surface texture of two types of polyether, five types of silicones and three
types of irreversible hydrocolloids. The specimens were constructed with Plaster of Paris
Review Of Literature
Page 10
from a metal block with a rough surface. The casts poured from the control group were
stored at room temperature for 24 hours. The other impressions were immersed for one
hour in distilled water, chloramine 2%, 70% ethanol, chlorinated tri-sodium phosphate
4% and 2% glutaraldehyde. Followed over 24 hours, the grooves of the plaster models
were evaluated with a profilo-meter. Three of the five addition silicones showed no
change after treatment with disinfectant. In the seven remaining materials, reproduction
of detail was altered by some of the solutions tested. From this, the authors conclude that
disinfection causes reduced capacity for detail reproduction and in some cases improved
surface texture.
LANGENWALTER; AQUILINO; TURNER (1990)21 investigated the
dimensional stability of polysulfide, polyether and addition silicone from a stainless steel
model built according to no 19 specification of the ADA. The impressions were
immersed in a solution of 0.0075% Iodoform, 0.05% sodium hypochlorite and 2%
glutaraldehyde for 10 minutes. The results indicated that the disinfectants tested did not
cause a statistically significant change in the linear dimension of the impression
materials.
In 1990, MINAGI et al22 studying the effect of the solution glutaraldehyde on the
dimensional stability of hydrophilic silicones of various consistencies showed that
disinfection by immersion for the times of 10, 20, 30, 60 and 120 minutes at 20 ° c
caused slight expansion (0.03%) in tested materials. They also showed that this type of
solution can be used for disinfection of hydrophilic silicones.
Review Of Literature
Page 11
MATHYAS et al (1990).23 compared measurements of models poured from
impressions treated for 10 minutes under immersion or spray of solution of o-
phenylphenol, glutaraldehyde, sodium hypochlorite, phenol and formalin with those of
untreated impressions. All materials were stable during the various test conditions. There
was also no difference in the accuracy of models from the group treated by dipping
or spraying.
LAUB L.W ; SANDRIK (1990).24 determined the effect that the immersion of
impressions of elastomers in disinfectant solutions based on glutaraldehyde had on the
surface hardness characteristics of the plaster casts. Impressions of line patterns in a
marked block were made with Polyether (Permadyne, Impregum F), Addition Silicone
(Express, Permagum) and Condensation Silicone (elasticon). After polymerization in
water bath at 35 ° c the specimens were immersed in four kinds of solution of
glutaraldehyde (Cidex 7, Glutarex, Sterall, Sporicidin) for 10 minutes. The control group
was kept in ambient conditions and immersed in de-ionized water. The plaster models
were poured 1 hour after polymerization and subjected to verification of knoop hardness
24 hours later. The cast surface quality was assessed with the semi-electronic microscope
with increased magnification of 1500x. The difference in hardness was not significant
(p ≤ 0.01) for dry specimens of the control groups and de-ionized water and impressions
obtained from Impregum F, Express and Elasticon. The hardness of the gypsum casts in
the control group was greater. Casts obtained from Impregum F showed no significant
differences compared to the control group bodies immersed in water and the group
immersed in Sterall. There is a significant decrease of hardness after immersion in
glutarex, cidex and sporicidin.
Review Of Literature
Page 12
SAMARANAYAKE, HUNJAN; JENNINGS (1991)25 investigated the
contamination and persistence of oral flora on irreversible hydrocolloid and elastomeric
impressions. The first part of the study was innoculation of Streptococcus mutans,
Escherichia coli, Staphylococcus aureus and Candida albicans on the surface of two
types of elastomers (Addition Silicone - Provil and Polysulfide Permelastic) and two
types of Irreversible Hydrocolloids ( Kromopan Blueprint and Aseptic). Samples of the
impression materials were removed in the form disk 13 mm in diameter. After a period of
5 hours in culture, the microorganisms were counted. In the second part of the study, 21
impressions of edentulous and dentulous patients were taken and the count of bacteria
was estimated and reported. The results showed 2 to 5 times more retention of bacteria in
irreversible hydrocolloids compared with the surface of the elastomers. In all impression
materials the number of viable bacteria present after 5 hours, decreased 65% to 98%,
except in the hydrocolloid containing disinfectant in its constitution. This showed total
destruction of microorganisms in less than three minutes. Among the findings seen, the
microbial load of the impression is higher in dentate than in edentulous, and this charge is
reduced rapidly with time. However, this does not eliminate the need for compulsory
disinfection of impressions.
GERHARDT, WILLIAM (1991)26 three major factors should be considered
when the molds are disinfected: the behavior of impression materials, the stability of the
solution disinfectant and effectiveness of the disinfectant solution. His study found that
sodium hypochlorite used to disinfect dental impressions is stable. The results indicated
the interference of three factors on the stability namely time, the conditions of storage
and use.
Review Of Literature
Page 13
PHILLIPS (1993) 8 classified impression materials as inelastic and elastic. As
per this author elastomers in general, can be disinfected in several solutions provided that
the period of exposure to disinfectant is short. A prolonged immersion in solution can
promote measurable distortions in the impression, and certain agents can reduce the
surface hardness of the poured plaster model. The Polyethers are most likely to change
dimensionally, if the immersion time is longer than 10 minutes.
C.PETER OWEN et.al (1993)27 conducted a study on disinfection of impression
material to prevent viral cross contamination. The polyether impression material expands
when immersed in disinfectant solution for prolonged period so it is not a material choice
when sterilisation is required. He concluded that choice of suitable impression material is
limited by the long immersion time required for sterilization.
The behavior of condensation silicone, polyether, polysulfide and addition
silicone immersed in Sodium Hypochlorite, Glutaraldehyde, Povidone Iodine and Ethyl
Alcohol were studied by ODA; MATSUMOTO, SUMI (1995).28 His work showed that
polysulfide and condensation silicone contract over time, but this contraction decreases
when immersed in disinfectant. Addition silicone shows excellent dimensional
stability. However, hydrophilic addition silicone and polyether expand considerably
when immersed in disinfectant solution, especially in ethanol. The hydrophobic addition
silicone provides excellent dimensional stability in the absence or presence of
disinfectant.
Review Of Literature
Page 14
THOUATI et al (1996)29 showed the influence of three disinfectant solutions on
the dimensional stability of seven elastomers. The impression materials tested are Optosil
P / L Xantoprem Blue Zetaplus / Tixoflex, Provil P / Lcd Elite (High and Low Viscosity)
and Dense Eurogum X1 / X3 Fluid. These materials have been dispensed and mixed
according to the manufacturer's specifications in an environment with controlled
temperature using the technique of double impression. The impressions were subjected to
four different conditions including immersion for 30 minutes in a solution of Quaternary
Ammonium Amphoteric Agent 2% (Dentasept), immersion for 60 minutes in 10%
Formaldehyde (Gigasept), immersion for 30 minutes in Sodium Hypochlorite 5.25%
(Hypochlorite) and poured immediately without immersion (control). Their results
indicated a contraction of 0.07% to 0.51 in the materials that have not been subjected to
disinfectant action. Condensation silicones showed half of the Zetaplus, Optosil
accuracy. However, taking into account the contraction of material when it polymerizes,
these expansions become smaller in absolute values for all materials.
MARIA DEL PILAR RIOS, et.al. (1996)11 conducted a study to evaluate the
accuracy and stability of impression materials when treated with disinfectant solutions.
He concluded that high level disinfectant solutions that belong to different chemical
groups at different concentrations did not affect the accuracy and dimensional stability of
polyether and poly vinyl siloxane impression material for 30 or 60 min.
According to POULOS, ANTONOFF (1997)10 many impression materials are
dimensionally unstable when immersed in a solution. The possibility of distortion is the
cause of much concern, especially in hydrophilic material that can absorb liquid.
Glutaraldehyde solutions cause slight expansion in rubbery silicone hydrophilic materials
Review Of Literature
Page 15
after immersion, so it is the most compatible solution that could be used with this
material. Because of its gel structure, hydrocolloids undergo changes dimensionally
immediately after gelation. The immersion of this material may cause distortion. There
are two methods available for disinfecting irreversible hydrocolloid (alginate). They are:
1 - spraying with a disinfectant solution diluted according to manufacturer's instructions,
placing in a plastic sealing bag and letting it stand for the time recommended by the
manufacturer (usually 10 to 30 minutes), then removing the bag, washing in water current
at room temperature and pouring as usual . 2 - Replacing the water used for mixing
irreversible hydrocolloid (alginate) with a solution disinfectant (provided by
manufacturer) of 0.2% or 0.01 % Chlorhexidine Titratabe Iodine. This results in an
internal disinfection and allows the operator to pour the impression immediately. 3-
Reversible hydrocolloid is disinfected by washing after removal from the mouth and
submerged in a solution disinfectant, preferably for 30 minutes, then washing with water
at room temperature and poured immediately. "
JOHNSON et al (1998).30 showed in another study that Irreversible
Hydrocolloids, Polyethers and Addition Silicones can be disinfected by immersion;
however this led to decreased accuracy of impressions and plaster models that can be
clinically significant. Specific combinations of irreversible hydrocolloid and
disinfectants, as is the case with the Jeltrate-Iodoform and Palgaflex-Glutaraldehyde
yielded the higher precision usually obtained. The surface quality of casts and models
tested in Polyether and Addition Silicone groups if disinfected or not, was higher than
that in the Irreversible Hydrocolloid. However, the quality of surface models from
irreversible hydrocolloid was enhanced by immersing the impressions in Glucose-And
Review Of Literature
Page 16
Phenol-Glutaraldehyde. This work gave indirectly the dimensional change of impression
materials by measuring distances of several clinically relevant type IV plaster models
reproduced from a master model of resin that represented a mandibular dentulous
arch. This model contained master stainless steel pins in each first molar occlusal surface
and lingual surface of the central incisors that served as reference for taking
measurements. Still, the area of a right pre-molar was receiving a modified stainless steel
pin with 12 degrees of convergence cervical-occlusal, mimicking a premolar prepared to
receive a full crown. The impression materials that were tested are alginates
(Palgaflex,Jeltrate), polyether (Impregum F) and addition silicone (President). The
disinfectants used were Iodoform (Biocide), glyco-glutaraldehyde (Impresept of) and
phenol-glutaraldehyde (Sporicidin). Ten minutes after the setting time, the impressions
were removed from master model and rinsed for 10 seconds. They were then immersed in
solution disinfectant for 10 minutes, while the control remained on the bench for this
same time. As a result, this study demonstrated that with polyether impressions, dies
generated are smaller in size than the original preparation, both from disinfected and not
disinfected impressions. Although the addition silicones have shown a significantly
higher degree of accuracy compared to the polyether, this difference is not clinically
relevant.
XAVIER LEPE et al (2002).31 There is an ongoing effort by dental
manufacturers to create impression materials with improved wetting properties. They
compared wettability, imbibitions and mass change of various recently introduced mixed
low-viscosity addition silicone and polyether materials before and after immersion
Review Of Literature
Page 17
disinfection. Within the limitations of this study imbibitions for the 2 polyether materials
were significantly higher (p<.001). Polyether materials lost significantly more (0.6% to
0.8%) and Aquasil LV gained significantly more (0.6%) mass in air.
MOREIRA DA SILVA et al (2004)32 conducted a study on effect of disinfectant
solutions on stability of impression materials. He concluded that there was no statistically
significant difference between the dimensional stability of silicone impressions immersed
in the 1% sodium hypochlorite solution or in 2% glutaraldehyde solution.
WALKER MP et al (2005)33 evaluated and compared the dimensional accuracy
and surface detail reproduction of two hydrophilic vinyl polysiloxane and two polyether
impression materials when used under dry and moist conditions. The surface detail
evaluation using criteria similar to ADA specification 19 indicated that moisture had a
significant effect on detail reproduction of vinyl polysiloxane materials. Under dry
conditions, all materials exhibited satisfactory detail reproduction 100% of the time;
however, under moist conditions, only 29% of PVS impressions produced satisfactory
detail reproduction, while 100% of the polyether impressions still met the criteria for
surface detail. Thus, although moisture may not adversely affect the dimensional
accuracy of either polyether or hydrophilic vinyl polysiloxane material, evidence
suggested that Polyether material is more likely to produce impressions with superior
surface detail reproduction in the presence of moisture.
CHANDUR P.K., WADHWANI et al (2005)34 conducted a study to assess the
accuracy of 2 types of fast setting impression materials when disinfected with acid
glutaraldehyde. The impression materials evaluated demonstrated sufficient accuracy.
Review Of Literature
Page 18
Immersion disinfection did not adversely affect the accuracy of any of the impression
materials. The working cast and working dies from regular and fast setting polyether
demonstrated an increase in all dimensions when compared to the master model and
stainless steel complete crown preparation. The working casts from the fast set VPS were
larger than the master model, whereas working dies showed a reduction in mesio-distal
dimension and height compared to the stainless steel complete crown preparation. The
new fast setting polyether and VPS materials demonstrated dimensional accuracy
equivalent to a traditional polyether.
TATSUO ENDO et al (2006), 35 worked on the dimensional accuracy of a new
polyether impression material. Their study showed that the vinyl polysiloxane material is
reasonably dimensionally stable during 24 hours of storage, irrespective of the
surrounding humidity. Impressions from polyether should preferably be cast with stone
within a few hours after impression making. During 24 hours of storage, the materials
show significantly increased shrinkage. Although polyether impressions show relatively
high shrinkage after 1 hour, they demonstrate acceptable dimensional accuracy after 24
hours only, when the impressions are not exposed to long-term relative humidity above
50%.
MARY P. WALKER et al (2007),36 investigated the effect of disinfection on
surface quality and dimensional stability of more recent, reformulated vinyl polysiloxane
and polyether materials and concluded that polyether significantly expanded following
disinfection with NaOCl or dual phenol, while there was no significant change in vinyl
polysiloxane dimensional accuracy following disinfection. Both non-disinfected and
disinfected vinyl polysiloxane and poly ether impressions exhibited increasing shrinkage
Review Of Literature
Page 19
over the period of time. There was a significant adverse effect on the surface quality of
poly ether with increasing exposure to NaOCl. Although disinfection does not appear to
adversely affect dimensional accuracy/stability of the more recent VPS and PE materials,
not all disinfectant solutions produce optimal impression surface quality with the newly
formulated poly ether.
SOO-HWA KIM et al. (2007), 37 conducted studies to evaluate the wettability of
the surface of the impression materials. The polymerization process and time might affect
the wettability of impression materials. The additional silicone impression material
showed distinct decrease of initial hydrophobicity to very hydrophobic levels, during and
after working time. He concluded that contact angle of dental impression materials is
effected by amount of time elapsed after mixing and disinfection and the hydrophobicity
of impression materials used should be considered when taking impressions and making
gypsum models.
OSAMA AL-JABRAH et al (2007), 38 stated that impression materials may act
as a vehicle for the transfer of microorganisms, and disinfecting them can completely
eliminate microorganisms carried by impressions.
HANDAN YILMAZ et al (2007),39 conducted a study on effect of disinfectants
on the dimensional stability of polyether impression materials and concluded that the
disinfectants tested for 10 min caused no significant dimensional change in polyether
material compared to control group.
DC JAGGER et al (2007). 40 studied the effect of range of disinfectants on
dimensional accuracy of some impression materials and concluded that for all the
Review Of Literature
Page 20
materials the changes in dimensions were small. These changes however are of no
clinical significance for fixed prosthodontic procedures.
DARIO MELILLI et al (2008). 41 conducted a study on effect of immersion
disinfectant procedures on dimensional stability of elastomeric impression material and
concluded that the dimensional change of all materials was within ADA specifications.
E. KOTSIOMATI et al (2008) 42 conducted a study on accuracy and stability of
impression materials subjected to chemical disinfection and concluded that polyethers
can be effectively disinfected by spraying; hydrophobic elastomers can be safely
immersed in disinfectants.
JENS JOHANNES BOCK ET AL (2008)43 conducted a study on influence of
different disinfectants on primary impression materials and concluded that casts from
silicone impressions were more accurate. Different disinfectant methods have only
marginal influence on dimensional stability and surface quality of dental casts.
WALA M. AMIN, et al (2009)44 concluded that sodium hypochlorite,
glutaraldehyde affected addition silicones very marginally. Of all the disinfectants
employed 0.5% sodium hypochlorite effected the least changes in the dimensions of
impression materials. Use of 0.5% sodium hypochlorite is recommended for 10 min
disinfection.
DIDEM ATABEK , et al (2009)45 evaluate the time dependent effects of
disinfection agents on impression materials and concluded that to rinse impression
materials only with water before casting is not an effective method for preventing cross-
contamination. Such a short time of 3 min. application of either 7.5% Povidone iodine or
Review Of Literature
Page 21
1% sodium hypochlorite was found sufficient for disinfecting impressions. In routine
dental practice the disinfection process of impression material in short period of 3
minutes can be practical, easy and effective method for clinicians. Disinfecting
impressions in a short period of 3 minutes as being a very practical method will stop
cross contamination and spreading of the infections in dental profession.
STOBER T, JOHNSON GH, SCHMITTER M. (2010)46 conducted a study to
evaluate accuracy of the newly formulated vinyl siloxanether elastomeric impression
material. Impressions were made from a modified dentoform master model containing a
simulated crown preparation. Dimensional changes between the master model and
working casts were assessed. VSE impressions demonstrated acceptable accuracy for
clinical use with immersion disinfection, since the results for VSE were comparable to
the results for PE and VPS materials, and the differences as compared to the master
model were small.
J. BUSTOS et al (2010)47 conducted a microbiological and SEM study on effect
of immersion disinfection with 0.5% sodium hypochlorite and 2% glutaraldehyde on
silicone and concluded that immersion in o.5% NaOCl and 2% glutaraldehyde for 10 min
completely eliminated bacteria in impressions. Immersion for 5-10 min significantly
inhibited bacterial growth in silicone impression materials. In SEM study immersion in
both disinfectants did not significantly affect surface quality.
N. ALMORTADI et al (2010)48 conducted a survey on disinfection of dental
impression and its compliance to accepted standards and concluded that a wide range of
solutions at different dilutions of same product was used by dentists to disinfect
impressions. 37.2% rinsed the impression with water, 2.6% always brushed debris away
Review Of Literature
Page 22
before disinfection and 24.7% informed laboratory of disinfection. Irrespective of the
disinfection status of the received impression, 50% of technicians disinfected
impressions, 64.7% were confident that impressions received by them are disinfected by
dentists.
K. SAMRA et al (2010)49 conducted a comparative study to evaluate efficacy of
different disinfectant systems on addition silicone materials of Indian and international
origin and concluded that glutaraldehyde and hypochlorite were effective in reducing the
microbial load, persistence. Carriage of microbial load was twice in alginate impressions
as compared to addition silicone group.
CINTIA LARA ODA CARVALHAL et al (2011)50 studied the influence of
immersion period in two disinfectant solutions on dimensional change of elastomeric
impression materials and concluded that with the exception of poly di-methyl siloxane for
5 or 10 min, there were no difference between disinfectant solutions. Immersion time
influenced dimensional change of different impression materials. Up to 20 min there was
no significant dimensional change for impression materials.
EMADE WADIE ESTAFANOUS et al (2012)51 evaluated disinfection of
bacterially contaminated hydrophilic polyvinyl siloxane and polyether impressions and
concluded that no bacterial growth was found on any stone casts resulting from
disinfected impression materials and no bacterial growth on selective media from stone
casts disinfected. Ten minutes either with spray or immersion disinfection is an effective
time to disinfect the impression materials. Disinfectants investigated in this study will
effectively disinfect PVS and polyether impression material.
Review Of Literature
Page 23
RAVI KUMAR R et al (2012) 52 evaluated the effects of chemical disinfection,
autoclaving and microwave sterilization on five elastomers and concluded that sterilizing
elastomers by autoclave and microwave is associated with mild linear dimensional
changes. Although statistically significant dimensional changes were observed, it is
important to note that the dimensional changes were less than 0.3%, which is well within
the ADA recommended dimensional change (< 0.5%). Therefore elastomeric material
can be safely sterilized using autoclave and micro wave energy because chemical
disinfection alone is less lethal and does not eliminate all bacterial forms.