3020 lecture 10 - eds wds fib

Upload: ales-nagode

Post on 10-Feb-2018

219 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    1/65

    Energy Dispersive X-ray Spectroscopy

    EDS, EDX, EDAX, etc.

    Detect X-rays emitted by incident electrons Determine compositional information at high spatial

    resolution

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    2/65

    Bremsstrahlung X-rays

    Incident e- undergo

    Coulombic repulsion with

    charged atomic species inthe sample, e- energy is

    lost

    Conservation of energy and

    momentum requires thatphotons are produced

    X-ray photons due to

    braking (Bremsstrahlung)

    gives rise to an energycontinuum

    ( )

    E

    EEZiI opcontinuum

    =

    Gives rise to background in EDS

    measurements, must be removedbefore quantification

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    3/65

    Inelastic Scattering - X-ray emission

    After the incoming e- beam excites inner shell e- in the sample, theexcited e- decay back to a ground state

    In order to obey energy conservation, they emit an X-ray The X-ray energy is dependent upon the elements present within the sample

    Can quantify composition by measuring these energies (EDX)

    May also emit an Auger e-

    Most X-rays generated are re-absorbed by the sample (will discusslater in quantitative analysis)

    X-ray energies are labeled by the type of transition (see graphprevious page):

    decay of L to K shells: K decay of M to K shells: K decay of M to L shells: L etc.

    K1 = K to LII

    K3 = K to LI, but is not an allowed transition (see table in book CD)

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    4/65

    Inelastic Scattering Characteristic X-ray

    Formation

    Bohr model - electronsaround an atom willequilibrate in certain energylevels around an atom

    Energy levels are dependentupon the nuclear mass andthe number of electrons

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    5/65

    More Characteristic X-rays

    Excited e- decay time ~ 1ps

    The probability of a transition is given by:

    Incident electron energy (in practice, we want an accelerating voltage 1.5xthe energy of the highest X-ray peak)

    Critical Ionization Energy (Ec) a certain E is required in order to excite a

    characteristic X-ray (greater than X-ray E) in general, as Z decreases, Ec

    decreases - table of all Ec available in table on book CD Cross section for ionization

    Fluorescence Yield

    Ratio of X-rays produced to the number of shell ionizations

    Increases with Z, resets with each X-ray line (K, L, M)

    )log(1051.62

    20 Uc

    UE

    bnQ s

    c

    ss=

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    6/65

    X-ray Detection

    How detect the energy

    dispersion of the emitted

    X-rays?

    Similar to the p-n BSE

    detector

    X-rays hit a

    semiconductor, produceelectron-hole pairs

    The number of pairs

    produced is proportional to

    the energy of the X-rays Si(Li) detector diffusion

    produced p-n junction that

    has intrinsic region ~mm

    range under reverse bias

    Si(Li) detectors: must be kept cold (77K) to

    stop E field induced diffusion and reduce thermal

    noise

    Other detectors: HPGe, proportional counters,

    microcalorimeters, silicon drift detectors

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    7/65

    X-ray detection

    The number of e-

    /hopairs created

    equals the X-rayenergy divided by the

    energy needed per

    charge (3.8eV/charge

    in Si(Li) detectors)

    ex: for a 5keV X-

    ray, 1316 carriers are

    produced, which

    results in a charge of2x10-16C for most

    detectors

    How do we measure

    such a small charge?

    EDS Windows: block stray light, avoid oil and ice

    contamination, sometimes seal vacuum

    Window Materials: Be (old systems), polymer films (TW,UTW, SUTW, most common in new systems), windowless

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    8/65

    X-ray detection

    The charge is

    converted to voltage

    by the FET

    The voltage pulse is

    then shaped and

    processed by a set of

    amplifiers

    Problem: want to

    average out noise

    (need many pulses),but this must happen

    before another X-ray

    comes into the

    detector and charges

    pile up

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    9/65

    X-ray pulse processing

    Solution: use 2

    amplifiers

    Fast amplifier -

    determines when a

    pulse is coming in or

    not, and if there are 2

    overlapping pulses -

    discriminator andpile-up inspector

    decide whether to

    reject or accept the

    pulse

    Slow amplifier -

    actually does the

    pulse shaping so the

    peak height can be

    converted into X-rayenergy

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    10/65

    Time Constants and Dead Time

    Important stuff from all the pulse

    processing

    Time Constant- time allowed for

    the pulse processor to evaluate themagnitude of the pulse

    short time constant, more counts per

    second, but with decreased energy

    resolution

    long time constant gives better

    resolution, but smaller cout rate

    Dead Time - period of time between

    when a pulse is processed and the

    detector shuts off to when thedetector can measure another pulse

    (closely related to the time constant)

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    11/65

    EDS Spectra

    Bremsstrahlung background

    Characteristic peaks illustrating the presence ofindividual elements

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    12/65

    Analyzing EDS Data

    What we know:

    How X-rays are generated How the X-rays are measured in a detector

    How the pulse processor turns the detector signal into

    X-ray energies

    What we can do with this info:

    Analyze what elements are in the samples Quantify how much of each element is in the sample

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    13/65

    Qualitative EDS

    Qualitative EDS - find out what elements are present pick highest peak intensity and highest energy first

    match each of the peak lines with the energies of the X-rays

    Today - all systems are computerized use software to scan through the element lists and determine what is present by

    which peaks line up with what is in your spectrum careful of SUM PEAKS!

    *ALL measurable peaks* (K, L, M) must match with the elements you think are

    there - just one line ID does not mean it is correct

    Problems - many elements have peaks that overlap with otherelements

    KMnKCr

    KCrKV

    KVKTi

    K

    K

    Al, P, S, Cl

    SrMOs

    LineInterferes withLineElementNote: Dont

    forget what

    you coated your

    samples with!

    Au/Pd can

    overlap with many

    elements

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    14/65

    Si Escape Peaks

    It is probable that an incoming X-ray will

    ionize the Si within the detector material,

    producing electron/hole pairs, Auger

    electrons, and characteristic Si X-rays

    Some of these characteristic Si K X-rays

    can escape from the detector

    This shows itself as an extra peak within

    the EDS spectra with an energy equal tothe characteristic X-ray energy minus the

    Si K X-ray energy (1.74eV)

    The probability of forming escape peaks

    increases as the energy of the incoming X-ray approaches 1.74eV (see graph)

    Si escape peaks will not occur for X-ray

    energies less than the Si K ionization

    energy (1.838ev)

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    15/65

    Detector Resolution

    Energy spread of the detector is

    dependent upon several factors:

    natural line width of the characteristicX-rays

    electronic noise (cool with LN2)

    pulse processing

    abruptness of the p-i-n junction in theSi(Li)

    Full width at half maximum

    (FWHM) is determined at optimal

    conditions and standardized for agiven composition (Mn), time

    constant, and dead time

    Typical ideal resolutions ~137-

    140eV

    If the FWHM is known at oneenergy, it can be determined at

    other energies by:

    FWHM= [2.5(EEref) + FWHMref2]1/ 2

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    16/65

    Quantitative Analysis

    Following qualitative analysis (what elements are present),

    we may be able to quantitatively determine whatcomposition is present

    What controls X-ray intensity?

    Composition (number of X-rays at a certain energy being emittedfrom the sample)

    X-rays produced = (ionizations/e-)*(X-rays/ionization)*(atoms/mole)

    *(moles/g)(g/cm3

    )(cm) Detection process (how many X-rays emitted get measured)

    X-rays detected = (X-rays/ionization)*(fraction emitted towards

    detector)*(detector efficiency)*(dead time)*(signal processing

    efficiency) etc..

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    17/65

    More Quantitative Analysis

    Best way to elminate all of those efficiency factors, etc.is to do Standards Basedanalyses

    take a sample of known composition (preferrably close to onethat you are trying to measure) and collect spectra using goodcollection statistics (30% dead time, long collection times,etc.) and standard beam conditions (the same gun bias, spot

    size, collection area, etc. you are going to use on your sample) calculate the composition of your sample from:

    ZAF = correction for Z (atomic #), A (absorption) and F(fluorescence)

    Ci

    Cst

    ZAF(Ii)

    ZAF(Ist)

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    18/65

    ZAF Correction Factors

    Z - larger atoms, more BSE (with relation to standard)

    A - X-rays must escape the surface without being reabsorbed

    dependent upon the (z) distribution of x-ray production with depth into sample

    F - fluorescence - X-ray energy from one element is enough to fluoresce an X-rayfrom a neighboring atom, typically zero for

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    19/65

    More ZAF

    (z) distribution is dependent

    upon accelerating voltage!

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    20/65

    Special Topics - Quant EDS of Thin

    Films Thin Film Correction

    Model with Monte Carlo

    Need to know substrate

    comp.

    Errors large if thickness of

    film unknown High spatial resolution if no

    substrate (STEM)

    Can also use back to bulkanalysis - use low kV such

    that interaction volume is

    similar to film thickness

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    21/65

    Special Topics - Quant EDS of Particles

    Particle Analyses

    Particle size vs. interaction

    volume Alters take off angles

    Measure many particles tolook for inhomogeneities /

    irregularities in spectra

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    22/65

    Special Topics - Quant EDS

    Surface Roughness Artifacts

    Minimize by tilting sample towards

    detector

    Low Z Analyses

    Use lowest possible U to avoid

    large A factors and deal with

    detection efficiency problems

    Use windowless detectors (can

    make results suspect)

    z corrections important

    Quantitative analysis very difficult

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    23/65

    EDS Summary

    Measure composition qualitatively using software, making sure alllines match up

    Measure standard composition and ZAF correction factors

    Measure sample, quantitatively determine composition usingcorrection factors

    Quantitative musts:

    choose standards and conditions carefully be aware of ZAF corrections - if ZAF moves significantly away from 1, bad data

    Good things about EDS Fast spectra collection

    Easy to operate Small, low cost system

    Bad things about EDS Spectral artifacts

    Poor resolution

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    24/65

    WDS Intro

    Wavelength Dispersive Spectroscopy (WDS)

    Used in electron probe microanalysis (EPMA)

    Use crystal of known crystallography and takeadvantage of Braggs law to increase resolution

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    25/65

    WDS Analysis System

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    26/65

    WDS Crystals

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    27/65

    WDS Analysis

    Advantages

    High resolution if take off angle remains same Low Z analysis

    Lower background

    Disadvantages Slow (need many counts)

    Large detector can limit imaging resolution

    Mechanical movements needed

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    28/65

    EDS / WDS Sampling

    How determine composition of your sample? BSE imaging can illustrate possible compositional changes in the

    specimen Composition from a specific position: spot mode

    Composition from set of specific positions: multipoint spectrumcollection

    Composition from an area: area scan mode Composition from a boundary: line scan mode

    Composition from an array of specific positions: mapping(spectral imaging, dot mapping, X-ray mapping, etc)

    Allows for determination of phase distributions

    Allows for determination of minor phases

    Quantitative mapping possible

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    29/65

    EDS Mapping - Setup

    Qualitative mapping

    Max count rate using high beam currents, analytical WD, tilt sample towards detector

    Small time constant, low dwell times at each pixel

    Ex: 1000cps, 512 x 512 bitmap, 30% deadtime

    1 sec dwell time: 375,000 sec, 6240 min, 104 hours!

    Better solutions: decrease dwell time (lower total counts) or use smaller energy range for

    analysis

    Ex: 0.02sec dwell time: 7,500 sec, 125 min, 2 hours

    Long scan times: need drift correction

    Quantitative mapping Need sufficient counts and energy resolution to do background subtraction, ZAF

    corrections

    Usually sacrifice spatial resolution to prevent long scans

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    30/65

    Data Recall (all images courtesy EDAX)

    Produce line

    profiles from

    spectral images

    Recall EDS

    data from

    each pixel(voxel)

    Produce

    element

    maps

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    31/65

    EDS Mapping

    Digital Dot Mapping (spectra at each pixel)

    Window around energy region of interest

    Displaying data:

    Grayscale - intensity of peaks given grayscale value

    (need for scaling the data)

    Color - intensity again scaled to give bright / dark areas,different colors given to different elements

    Modern EDS systems allow for determination and display of

    composition at each pixel or along line profiles

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    32/65

    Quant Map - ZAF Corrected

    Pb M

    S K

    Pb M ZAFPb L ZAF

    BSE S K ZAF

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    33/65

    Live Spectral Mapping

    Frame averaging gives gradually better EDS data at each pixel,

    displays live maps either elementally or as overlay

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    34/65

    Electron Energy Loss Spectroscopy

    EELS

    Typically used in the TEM / STEM

    Can be done in parallel and serial modes Measures the amount of energy lost by an electron after

    it passes through a sample of some thickness

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    35/65

    EELS

    Bend electron beam around hemispherical

    analyzer to separate e- of different energies in

    space Can also select a certain window of e- energies to

    image with (energy filtered TEM, EFTEM or GIF)

    Make maps of composition without STEM

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    36/65

    EELS

    Information gathered from EELS

    Characteristic energy loss gives elemental composition,

    complementary to EDS Zero loss peak intensity gives sample thickness if mean

    free path length is known

    Plasmon losses give information about majority carriermobility

    Ability to map electronic conductivities, etc.

    At high energy resolution, can get information about

    bandgap Need specialized electron source

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    37/65

    FIB

    Focused Ion Beam

    Controlled milling / machining using ion beams

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    38/65

    Ion - Solid Interactions

    Elastic / Inelastic collisions

    Sputtering: Elastic collisions,

    momentum transferred fromincident ions into the targetatoms, forming secondary ions(SI)

    aka backsputtering

    SIs can be measured by its ownplate detector, gives rise todifferent contrast than SEs, similarto BSE

    Inelastic scattering: producesSE, phonons, plasmons, etc. Nuclear

    Electronic

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    39/65

    Ion - Solid Interaction Ranges

    R = range

    Rp = projected range

    Xs = projected rangealong a vector normal

    to the surface

    Rr = radial range

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    40/65

    Collision Cascades

    Collision cascade: moving sea of particles within a solidunder ion bombardment

    3 regimes: I. Single knock-on (M1M2, Eo is high)

    Majority of atoms within the spike volume move during thecollision cascade

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    41/65

    More Ion - Solid Interactions

    Modeling energy loss in amorphous solids Universal screening function leads to the stopping power of the

    target material Take into account both conservation of momentum and the

    interatomic potential

    Best for MSE scientists to model with SRIM calculations (MonteCarlo, www.srim.org)

    Ion Implantation Flux - time rate of flow of energy (ions / cm2 / s)

    Fluence - number of particles incident on a specific area (ions /cm2) during a certain time interval

    Dose - quantity of ions absorbed by a medium (ions / cm2)

    Beam Current / Current Density - time rate of flow (Amps (C/s) orAmps / cm2)

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    42/65

    More Ion - Solid Interactions

    Sputtering - sputteringyield = # ejected particles

    per incident ion Depends on target,incident energy, angle ofincidence

    In FIB, sputter yield variesbetween 0.1 and 100

    Ejected ion energiestypically between 2-5eV

    Sputter yield tables

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    43/65

    Dual beam:

    FEGSEM

    FIB

    EDS

    Omniprobe

    EBSD

    Mounted directlyunder FIB column

    Nova 200 Nanolab

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    44/65

    The FIB Instrument

    Similar to SEM

    optically Accelerate ions into a

    solid sample - but

    why? Attachments:

    GIS

    Liftout / manipulator Dual Beam

    EDS, EBSD

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    45/65

    FIB Column

    Optically similar to SEM

    Consists of:

    Gun (next slide)

    Condenser Lens

    Beam defining aperture (changes spot size and beam current

    alone, no real change in condenser setup)

    Beam blanker - Faraday cage which brings beam out of optic

    axis and into bulk material

    Objective lens - changes focal length

    Scan coils

    Stigmators

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    46/65

    FIB Sources

    Liquid Metal Ion System (LMIS)

    heat Ga metal above melting temperature

    Ga flows to a W tip with radius ~2-5m

    use field emission to form 2-5nm Ga tip (Taylorcone)

    extract Ga+ ions and accelerate them down the

    column

    Ga flow continuously replenishes source

    i

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    47/65

    FIB Imaging

    SE detection

    same as SEM (E-T detector)

    spatial resolution limited by large interaction volume, aberrations (largeenergy spread for incident beam)

    SI detection

    can image insulators!!! - as long as charging is not too severe

    similar to BSE - only line of sight ions hit the detector Topographical contrast

    surface roughness introduces contrast into the electron image

    Channeling contrast

    ion channeling into individual grains at different orientations givesdifferent backscatter yields, thus contrast

    FIB I C l Ch li C

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    48/65

    FIB Ion Column Channeling Contrast

    FIB I i P bl

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    49/65

    FIB Imaging - Problems

    Ion beam can damage / alter the surface you are

    trying to image

    Sputtering can take the surface of interest away

    Beam induced grain growth

    Secondary Ga phase formation

    FIB Milli

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    50/65

    FIB Milling

    Effects of Z, crystal orientation, incidence angle

    Damage (amorphization, theatre curtain)

    effects of accelerating voltage, beam current, angle of

    incidence

    Redeposition

    Grain Growth

    Ga secondary phase formation

    FIB Milli Th t C t i Eff t

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    51/65

    FIB Milling - Theatre Curtain Effect

    FIB Milli R d iti

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    52/65

    FIB Milling - Redeposition

    FIB Milling Effects of Orientation

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    53/65

    FIB Milling - Effects of Orientation

    FIB Milling Parameters

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    54/65

    FIB Milling - Parameters

    Can mill using patterns Rectangle - mill top to bottom, reverse, or side to side

    Cross section (swimming pool)

    Cleaning cross section (one line at a time progressively towardsendpoint)

    Circle or donut (mill top to bottom, reverse, side to side, or annular)

    Can also mill using bitmap (24-bit RGB)

    R: not used G: 0 - 255 (0 = blanking, 255 = no blanking)

    B: dwell time (0 = 100ns, 255 = determined by user)

    Only works in 100ns steps

    If user defines 500ns dwell time, only 5 milling levels allowed If user defines 500s dwell time, 5000 levels allowed (better

    depth control

    Can ultimately define depth by changing the number of passes

    More FIB Milling Parameters

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    55/65

    More FIB Milling Parameters

    Computer analyzes milling data pixel by pixel

    Max number of points = 1Million

    Pitch (x,y) = size of mill set in UI / #pixels in image

    Pattern resolution = HFW / pixels in length

    Overlap = resolution / spot size (given by selected beam

    current aperture)

    Decreasing overlap (necessary at low beam

    currents) changes milling rate and possibly

    structure

    FIB TEM Specimen Preparation

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    56/65

    FIB TEM Specimen Preparation

    Geometries

    H bar or Swimming Pools

    Wedge polishing / FIB finishing

    Omniprobe Wedge FIB and total liftout technique

    Plan-view

    In-situ and Ex-situ liftout Artifacts

    Ga implantation / contamination - reduce with low kV

    cleaning Theatre curtain / Amorphization - prevent with Pt

    Spotting

    FIB TEM Specimen Prep

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    57/65

    FIB TEM Specimen Prep

    In-situ Liftout

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    58/65

    In-situ Liftout

    Dual Beam FIB

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    59/65

    Dual Beam FIB

    Allows for 3-Dreconstructionsfrom serial sections

    IMOD (U.Coloradosoftware)

    AMIRA(Mercurysoftware,tgs.com)

    EDS - TEM prep

    EBSD - TEM prep

    X-C SEMpreparation

    FIB SIMS

    FIB Auger

    Courtesy: IMOD

    Other Ion Systems in the FIB

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    60/65

    Other Ion Systems in the FIB

    Gas Injection Systems (GIS), i.e., Deposition (e-beam, i-beam)

    Allow for deposition / etching of structures using gases other than the Ga beam

    precursor chemical is heated and injected into the beam path of the electrons / ions via along needle

    precursor is converted into metallic species

    Methylcyclopentadienyl Pt trimethyl - (CH3)3(CH3C5H4)Pt

    Tungsten Hexacarbonyl, W(CO)6

    Trimethylamine Alane, (CH3)3NAlH3

    TEOS, TMOS + H2O - SiO2 Phenanthrene, pyrene, napthalene for C deposition

    On roadmap: Au, Pd, Co, Fe

    Enhanced Etching chemistries

    gaseous chemical is injected into the Ga beam path, enhances the etch rate by reacting

    with the target material Halogens used most often (XeF2, Cl2, Br2, I2) for Si, SiO2, Al, W

    also H2O for PMMA and polyimide

    Delineation etch (2,2,2-trifluroroacetamide) removes insulator, leaving Si intact

    CuRx - reduces Cu channeling / preferential sputtering

    Selective C mill - removes organics and plastics, leaving inorganics intact

    GIS Deposition with e- and ions

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    61/65

    GIS Deposition with e and ions

    e- beam deposition

    1. Lower dep rate (~0.2)

    2. No Ga contamination

    3. No surface damage

    4. C and porosity content

    increased5. Smaller structures

    possible

    Ion Beam Deposition

    1. High dep rate

    2. Ga contamination

    3. Residual surface

    damage

    4. Effects of beam taillimit minimum feature

    size

    3-D Nanofabrication

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    62/65

    3 D Nanofabrication

    Use GIS

    deposition to

    produce nanoscale3-D structures

    Usefulness:

    connect lines in IC

    devices

    lithography mask

    repair

    MEMS

    Milling cap layers

    3-D Nanofabrication

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    63/65

    3 Na o ab ca o

    3-D Etches

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    64/65

    Useful for

    variable etchingrate materials

    and devices

    FIBIX

  • 7/22/2019 3020 Lecture 10 - Eds Wds Fib

    65/65

    X-ray spectroscopy induced by Ga beam

    No Bremsstrahlung X-rays, greatly reduced background

    and higher resolution at low Es

    Very sensitive at low Es

    No high E X-rays detectable

    NeednA to get decent spectra

    Sample will not last long in spot mode at these high

    beam currents