Measurement Measurement SystemSystem

Measurement System 1

Physical Measurement Method

(Metode Pengukuran Fisika)SF 091306

Gatut YudoyonoPhysics department

Page 2: 2a Measurement System

Measurement System 2

Measure what can be measured and make measurable what cannot be measured

(Galileo Galilei, 1600)

A measurement system is comprised of the equipment used

to sense an experiment’s enviroment, to change and condition

what is sensed into a recordable form, and to record its values.

A measurement system’s main purpose is to produce an accurate an accurate

numerical value numerical value of the measurand.

Ideally, the recorded value should be the exact value of the physical

variable sensed by the measurement system.

Page 3: 2a Measurement System

Measurement System 3

Present-day applications of measuring instruments measuring instruments can be classified into three major areas.

The firstfirst of these is their use in regulatingregulating trade, applyinginstruments that measure physical quantities such as length, volume and mass in terms of standard units.

The secondsecond application area of measuring instruments is in

monitoring functionsmonitoring functions.

The gardener uses a thermometer to determine whether he should turn the heat on in his

greenhouse or open the windows if it is too hot.

Use as part of automatic feedback control systemsautomatic feedback control systems

forms the third application area of measurement systems.

Page 4: 2a Measurement System

Measurement System 4

Pengelompokan alat ukur 0:

Some quantities can thus be measured directly, other can be

measured only indirectly.

The great majority of physics measurements are indirect

measurements.

The Young’s modulus of a brass wire cannot be experimentally

determinated by finding how many times the unit of Young’s modulus,

usually determinated by measuring a force and three lengths, and from

them calculating the Young’s modulus.

Page 5: 2a Measurement System

Measurement System 5

Pengelompokan alat ukur 1:

1.Monitoring suatu proses dan operasi

Contoh:

a.Termometer, barometer, anemometer yg digunakan BMG

untuk melihat kondisi cuaca lingkungan.

Hanya membaca/ display bukan kontrol

b. Alat ukur air, gas, listrik di rumah tangga

2.Mengontrol suatu proses dan operasi

Hasil ukur sebagai variabel umpan balik

Page 6: 2a Measurement System

Measurement System 6

3. Analisis Eksperimental

Menyelesaikan problem engineering:

Teoritis (Hasil bersifat umum, Perlu asumsi model matematis,

Tanpa alat ukur)

Eksperimental (Hasil bersifat khusus, Tanpa asumsi, Perlu

peralatan yg akurat)

Analisis eksperimental / integrasi teori dan eksperimental (Uji

validasi prediksi teoritis; Formulasi kondisi empiris; Menentukan

bahan, komponen, parameter sistem: variabel, performens; Studi

fenomena untuk pengembangan teori)

Komputasi

Page 7: 2a Measurement System

Measurement System 7

Pengelompokan alat ukur 2:

1. Alat ukur absolut (mutlak, standar)

Alat ukur yg menunjukkan besaran dari komponen/besaran yg diukur dg

batas pada konstanta dan penyimpangan alat itu sendiri.

Tidak perlu ditera/ kalibrasi atau dibandingkan dg alat ukur lain

Contoh: Galvanometer, mengukur arus yg ukuran nilainya sesuai dg:

Busur dari simpangan

Jari-jari (panjang jarum)

Jumlah lilitan

Komponen horisontal dari permukaan bumi

Page 8: 2a Measurement System

Measurement System 8

2. Alat ukur sekunder

alat ukur yg menunjukkan harga besaran yg diukur dan dapat

ditentukan hanya dari simpangan alat ukur setelah dikalibrasi

"Tanpa kalibrasi alat ini tidak berguna”

Alat ukur sekunder digolongkan:

alat ukur penunjuk

menunjukkan nilai sesaat pada waktu pengukuran

contoh: ammeter, voltmeter

Alat ukur pencatat

Menunjukkan rekaman selama waktu tertentu

Alat ukur terintegrasi

Page 9: 2a Measurement System

Measurement System 9

Pengelompokan alat ukur 3:

Active and passive instruments

Instruments are divided into active or passive

ones according to whether the instrument

output is entirely produced by the quantity

being measured or whether the quantity being

measured simply modulates the magnitude of

some external power source.

Page 10: 2a Measurement System

Measurement System 10

Passive pressure gauge.

The pressure of the fluid is translated into a movement of a pointer against a scale. The energy expended in moving the pointer is derived entirely from the change in pressure measured: there are no other energy inputs to the system.

Page 11: 2a Measurement System

Measurement System 11

Petrol-tank level indicator.

The change in petrol level moves a potentiometer arm, and the output signal consists of a proportion of the external voltage source appliedacross the two ends of the potentiometer.

The energy in the output signal comes from the external power source: the primary transducer float system is merely modulating the value of the voltage from this external power source.

Page 12: 2a Measurement System

Measurement System 12

One very important difference between active and passive

instruments is the level of measurement resolution that can be

obtained.

In terms of cost, passive instruments are normally of a

more simple construction than active ones and are

therefore cheaper to manufacture.

Page 13: 2a Measurement System

Measurement System 13

Pengelompokan alat ukur 4:

Null-type and deflection-type instruments

Passive pressure gauge:a good example of a deflection type of instrument,

A deflection type of instrument, where the value of the quantity being measured is displayed in terms of the amount of movement of a pointer.

Page 14: 2a Measurement System

Measurement System 14

Deadweight pressure gauge:

a null-type instrument.

Weights are put on top of the piston until the downward force balances the fluid pressure.

Weights are added until the piston reaches a datum level, known as the null point.

Pressure measurement is made in terms of the value of the weights needed to reach this null position.

Page 15: 2a Measurement System

Measurement System 15

Pengelompokan alat ukur 5:

Analogue and digital instruments

An analogue instrument gives an output that varies

continuously as the quantity being measured changes.

A digital instrument has an output that varies in discrete steps and so can only have a finite number of values.

Rev counter.

Page 16: 2a Measurement System

General architecture of a measurement system

Three main functions can be distinguished:

data acquisition

data processing

data distribution

Measurement System 16

Page 17: 2a Measurement System

Data acquisitionData acquisition:: to obtain information about the measurement

object and convert it into an electrical signal. All data is transferred

via a single connection to the data processing element.

Data processingData processing: includes processing, selecting or otherwise

manipulating measurement data in a prescribed manner. This task is

performed by a microprocessor, a micro-controller or a computer.

Data distributionData distribution: to supply processed data to the target object. The

multiple output indicates the possible presence of several target

instruments such as a series of control valves in a process installation.

Measurement System 17

Page 18: 2a Measurement System

Measurement System 18

Three main functions of data acquisition

Three main functions of data distribution

Page 19: 2a Measurement System

The element of a measurement system:

SensorSensor

TranduserTranduser

Signal conditionerSignal conditioner

Signal processorSignal processor

Page 20: 2a Measurement System

1.1. The sensorThe sensor is a device that senses a physical stimulus and convert it into a signal. (This signal usually is electrical, mechanical, or optical)

2.2. The tranducerThe tranducer is the device that changes (transduces) the signal into the desired quantity (be it electrical, mechanical, optical or another form). In the most general sense, a tranducer transforms energy from one form to another.

3.3. The signal contionersThe signal contioners, in essence, puts the signal in its final form to be processed and recorded.

4.4. The signal processor The signal processor

Page 21: 2a Measurement System

Sensor can categorized into domain, according to the type of physical variables that they sense.

Chemical: chemical concentration, composition, and

reaction rate

Electric: current, voltage, resistance, capacitance,

inductance, and charge

Magnetic: magnetic field intensity, flux density, and

magnetization

Mechanical: displacement or strain, level, position,

velocity, acceleration, force, torque, pressure, and flow rate

Radiant: electromagnetic wave intencity, wavelength,

polarization, and phase

Thermal: temperature, heat, and heat flux

Page 22: 2a Measurement System

SensorSensor always are based on some physical principle or law.

These include, but are not limited to, the sensor’s:•Operational bandwidth•Magnitude and frequency response over that bandwidth•Accuracy•Voltage or current supply requirement•Physical dimensions, weight, and materials•Enviromental operating conditions (pressure, temperature, relative humidity, air purity, radiation)•Type of output (electrical, mechanics)•Further signal conditioning requirement•Operational complexity•Cost

The choise of either designing or selecting a particular sensor start with identifying the physical variable to be

sensed and the physical principle or law associated with that variable. Then, the sensor”s input/output characteristics

must be identified.

Page 23: 2a Measurement System

Measurement System 23

The element of a measurement system

Page 24: 2a Measurement System

Measurement System 24

1. Primary sensing element

Menerima energi dari medium yg diukur hasil tgt besar yg diukur

Besaran terukur selalu diganggu oleh besaran yg diukur tidak

mungkin pengukuran sempurna

Alat yg baik, didesain untuk meminimalisasi efek gangguan

2. Variable conversion element Mengkonversi besaran dari elemen sensing

Merupakan elemen fungsional, bukan fisis

Page 25: 2a Measurement System

Measurement System 25

3. Variable Manipulation Element

Memanipulasi variabel dari elemen sebelumnya

Contoh : sinyal yg lemah diperkuat

4. Data Transmission Element

5. Data Presentation Element

Page 26: 2a Measurement System

Measurement System 26

Pengukur Tekanan

Page 27: 2a Measurement System

Measurement System 27

Page 28: 2a Measurement System

Calibration and ResponseCalibration and Response

A single numberA single number has more genuine and permanent value than an expensive library full of hypothesesfull of hypotheses. (Robert J. Mayer, c. 1840)

Measurement system and their instrument are used in experiment to

obtain measurand value that usually are either steady or varying in time.

The error arise in the measurand value simply because the instrument

are not perfect; their output do not precisely follow their inputs.

These errors can be quantified through the

process of calibrationcalibration.

Page 29: 2a Measurement System

In a calibration, a known input value (called the standart) is applied to the

system and then is output is measured.

Calibrations can either be static (not a function of time) or dynamic

(both the magnitude and the frequency of the input signal can be a

function of time).

Calibration can be performed in either sequential or random steps.

In sequential, the input increased systematically and then decreased.

In random, the input is changed from one value to another in no

particular order.

Page 30: 2a Measurement System

Measurement System 30

STATIC Response CharacteristicsSTATIC Response Characteristics

Page 31: 2a Measurement System

Measurement System 31

The STATIC SENSITIVITY, the slope of the calibration curve at a particular input value

The curve is linear, K will constant.

The ABSOLUTE ERROR

The RELATIVE ERROR

The ACCURACY

1

)( 1xxdx

dyxKK

valueindicated - valuetrue eabs

valuetrueerel

abse

rele1ecal

Page 32: 2a Measurement System

Measurement System 32

1. Accuracy and inaccuracy (measurement uncertainty)

The accuracy of an instrument is a measure of how close the output reading of the instrument is to the correct value.

Inaccuracy is the extent to which a reading might be wrong, and is often quoted as a percentage of the full-scale (f.s.) reading of an instrument.

If, for example, a pressure gauge of range 0-10 bar has a quoted inaccuracy of ±1.0% f.s. (± 1% of full-scale reading), then the maximum error to be expected in any reading is 0.1 bar. This means that when the instrument is reading 1.0 bar, the possible error is 10% of this value.

The term measurement uncertainty is frequently used in place of inaccuracy.

Page 33: 2a Measurement System

Measurement System 33

2. Precision/repeatability/reproducibilityPrecision is a term that describes an instrument's degree of

freedom from random errors.

If a large number of readings are taken of the same quantity by a

high precision instrument, then the spread of readings will be very

small.

Repeatability describes the closeness of output readings when the

same input is applied repetitively over a short period of time, with the

same measurement conditions, same instrument and observer,

same location and same conditions of use maintained throughout.

Reproducibility describes the closeness of output readings for the

same input when there are changes in the method of measurement,

observer, measuring instrument, location, conditions of use and time

of measurement.

Page 34: 2a Measurement System

Measurement System 34

Page 35: 2a Measurement System

Measurement System 35

3. Tolerance

Tolerance is a term that is closely related to accuracy and defines

the maximum error that is to be expected in some value.

The electric circuit components such as resistors have

tolerances of perhaps 5%. One resistor chosen at random from

a batch having a nominal value 1000 and tolerance 5% might

have an actual value anywhere between 950 and 1050 .

4. Range or span

The range or span of an instrument defines the minimum and maximum values of a quantity that the instrument is designed to measure.

Page 36: 2a Measurement System

Measurement System 36

5. Linearity

It is normally desirable that the output reading of an instrument is linearly proportional to the quantity being measured.

Page 37: 2a Measurement System

Measurement System 37

6. Sensitivity of measurement

The sensitivity of measurement is a measure of the change in instrument output that occurs when the quantity being measured changes by a given amount.

Thus, sensitivity is the ratio:

The sensitivity of measurement is therefore the slope of the straight line drawn on.

For example, a pressure of 2 bar produces a deflection of 10 degrees in a pressure transducer, the sensitivity of the instrument is 5 degrees/bar (assuming that the deflection is zero with zero pressure applied).

Page 38: 2a Measurement System

Measurement System 38

Example

The following resistance values of a platinum resistance thermometer were measured at a range of temperatures. Determine the measurement sensitivity of the instrument in ohms/oC

SolutionIf these values are plotted on a graph, the straight-line relationship between resistance change and temperature change is obvious. For a change in temperature of 30oC the change in resistance is 7 . Hence the measurement sensitivity = 7/30 = 0.233 /oC

Page 39: 2a Measurement System

Measurement System 39

7. Threshold

This minimum level of input is known as the threshold of the instrument.

As an illustration, a car speedometer typically has a threshold of about 15 km/h. This means that, if the vehicle starts from rest and accelerates, no output reading is observed on the speedometer until the speed reaches 15 km/h.

8. Resolution

When an instrument is showing a particular output reading, there is a lower limit on the magnitude of the change in the input measured quantity that produces an observable change in the instrument output.

Using a car speedometer as an example again, this has subdivisions of typically 20 km/h. This means that when the needle is between the scale markings,we cannot estimate speed more accurately than to the nearest 5 km/h. This figure of 5 km/h thus represents the resolution of the instrument.

Page 40: 2a Measurement System

Measurement System 40

9. Sensitivity to disturbance

All calibrations and specifications of an instrument are only valid under controlled conditions of temperature, pressure etc. As variations occur in the ambient temperature etc., certain static instrument characteristics change, and the sensitivity to disturbance is a measure of the magnitude of this change.

Such environmental changes affect instruments in two main ways, known as zero drift (bias) and sensitivity drift.

Zero drift or bias describes the effect where the zero reading of an instrument is modified by a change in ambient conditions

Sensitivity drift (scale factor drift) defines the amount by which aninstrument's sensitivity of measurement varies as ambient conditions change.

Page 41: 2a Measurement System

Measurement System 41

Page 42: 2a Measurement System

Measurement System 42

Example A spring balance is calibrated in an environment at a temperature of 20 oC and has the following deflection/load characteristic.

It is then used in an environment at a temperature of 30 oC and the following deflection/ load characteristic is measured.

Determine the zero drift and sensitivity drift per oC change in ambient temperature.

Page 43: 2a Measurement System

Measurement System 43

10. Hysteresis effects

The non-coincidence between these loading and unloading curves is known as hysteresis.

Two quantities are defined: •maximum input hysteresis •maximum output hysteresis

Page 44: 2a Measurement System

Measurement System 44

Hysteresis is most commonly found in instruments that contain springs, such as the passive pressure gauge

Devices like the mechanical flyball (a device for measuring rotational velocity) sufferhysteresis from both of the above sources because they have friction in moving partsand also contain a spring.

Hysteresis can also occur in instruments that contain electricalwindings formed round an iron core, due to magnetic hysteresis in the iron.

This occurs in devices like the variable inductance displacement transducer, the LVDT and the rotary differential transformer.

Page 45: 2a Measurement System

Measurement System 45

11. Dead space

Dead space is defined as the range of different input values over which there is nochange in output value.

Page 46: 2a Measurement System

Measurement System 46

DYNAMIC CharacteristicsDYNAMIC Characteristics

The dynamic characteristics of a measuring instrument describe its

behaviour between the time a measured quantity changes value and

the time when the instrument output attains a steady value in

response.

As with static characteristics, any values for dynamic characteristics

quoted in instrument data sheets only apply when the instrument is

used under specified environmental conditions. Outside these

calibration conditions, some variation in the dynamic parameters can

be expected.

Page 47: 2a Measurement System

Measurement System 47

In any linear, time-invariant measuring system, the following general relation can be written between input and output for time (t) > O:

where qi is the measured quantity, q0 is the output reading and ao... an, bo... bm are constants.

If we limit consideration to that of step changes in the measured quantity only, then the equation reduces to:

Page 48: 2a Measurement System

Measurement System 48

If all the coefficients a1 . . . an other than a0 in the equation are assumed zero, then:

where K is a constant known as the instrument sensitivity as defined earlier.

A potentiometer, which measures

motion, is a good example of such

an instrument, where the output

voltage changes instantaneously as

the slider is displaced along thepotentiometer track.

Page 49: 2a Measurement System

Measurement System 49

If all the coefficients a2...an except for ao and a1 are assumed zero inthe equation then:

Any instrument that behaves according to the equation is known as a first order instrument.

If d/dt is replaced by the D operator:

Defining K = bo/ao as the static sensitivity and = a1/ao

Page 50: 2a Measurement System

Measurement System 50

The liquid-in-glass thermometer is a good example of a first order instrument. It is well known that, if a thermometer at room temperature is plunged into boiling water, the output e.m.f, does not rise instantaneously to a level indicating100oC but instead approaches a reading indicating 100oC in a manner similar to that shown in the figure.

Page 51: 2a Measurement System

Measurement System 51

If all coefficients a3... an other than ao, a1 and a2 in the equation are assumed zero, then we get:

K (static sensitivity), (undamped natural frequency) and (damping ratio)

This is the standard equation for a second order system and any instrument whose response can be described by it is known as a second order instrument.

Page 52: 2a Measurement System

Measurement System 52

Page 53: 2a Measurement System

Measurement System 53

For case (A) where = 0, there is no damping and the instrument output exhibits

constant amplitude oscillations when disturbed by any change in the physical

quantity measured.

For light damping of = 0.2, represented by case (B), the response to a step

change in input is still oscillatory but the oscillations gradually die down.

Further increase in the value of reduces oscillations and overshoot still more, as

shown by curves (C) and (D), and finally the response becomes very overdamped

as shown by curve (E) where the output reading creeps

up slowly towards the correct reading.

Page 54: 2a Measurement System

Measurement System 54

Clearly, the extreme response curves (A) and (E) are grossly unsuitable

for any measuring instrument. If an instrument were to be only ever

subjected to step inputs, then the design strategy would be to aim

towards a damping ratio of 0.707, which gives the critically damped

response (C).

As the form of the input variable changes, so the best value for varies,

and choice of becomes one of compromise between those values that

are best for each type of input variable behaviour anticipated.

Commercial second order instruments, of which the accelerometer

is a common example, are generally designed to have a damping

ratio () somewhere in the range of 0.6-0.8.

Page 55: 2a Measurement System

Measurement System 55

• Directly / indirectly.

• Monitoring/kontrol/ Analisis

• Absolut/sekunder

• Active and passive

• Null-type and deflection-type

• Analogue and digital instruments