27.hg calibrator 81i spec
TRANSCRIPT
-
8/10/2019 27.Hg Calibrator 81i Spec
1/3
1 of 1
SR.NO
1
2
3
4
5
6
7
8
9
10
11
12
13
14
DATA SHEET DATA SHEET NO SHEET
EID/130/DS/02
327, TTC industrial area,MERCURY CALIBRATOR
ITEM CODE
Pawane, Navi Mumbai 81I-BNNNCA
roducedorexhibitedtothirdpartieswithoutanywrittenpermissionofTFS
DESCRIPTION SPECIFICATION
MANUFACTURER THERMO FISHER SCIENTIFIC, USA
MODEL NO 81i-BNNNCA
DISPLAY LCD
MASS FLOW MEASUREMENT ACCURACY +/- 2% SET POINT OR +/- 1% FS, WHICHEVER IS LESS FROM
-20 TO 100% FS
REPEATABILITY OF MASS FLOW CONTROL +/- 2% OF READING OR 1% OF FULL SCALE, WHICHEVER IS LESS
-20 TO 100% FS
LINEARITY OF MASS FLOW MEASUREMENTS +/- 0.5% FS
ZERO AIR FLOW CONTROLLER 20 SLM
SPAN GAS FLOW CONTROLLER 50 SCCM
HG SOURCE CONTROLLER 0- 15C
HG CONCENTRATION STD OUTPUT 3 - 50 g/dscm
LINEARITY +/- 2% FS
OPERATING TEMPERATURE 5- 40C
POWER REQUIREMENT 220 VAC 50HZ 250W
SIGNAL OUTPUT ANALOG OUT (4-20mA)
15
16
17
18
NOTE :
10/20/2011 0 SS MNS SIEMENS
ThisdocumentisthepropertyofThermoFisherScientificIndiaPvtLtd.andnottobecopiedorrepr
MOUNTING TYPE ON 19 RACK
16.75(W) X 8.62(H) X 23(D)PHYSICAL DIMENSIONS
APPROXIMATE WEIGHT 21.8 KG
MODEL DE-CODIFICATION
81I MERCURY CALIBRATOR
B = 220 VAC 50/60 Hz
N = HYDRATOR (STANDARD)
N = NOT AVAILABLE YET (FUTURE OPTION)
N = NOT AVAILABLE YET (FUTURE OPTION)
C = 6 CHANNEL 4-20mA CURRENT OUTPUT, (STANDARD)
A = BENCH MOUNTING (STANDARD)
ISSUED FOR APPROVAL
DATE REV PREPARED CHECKED APPROVEDDESCRIPTION
-
8/10/2019 27.Hg Calibrator 81i Spec
2/3
The Thermo Scientific Mercury Calibrator,
Model 81iMercury Calibrator is one of
the four major components of the Mercury
Freedom System. The Model 81iutilizes
a vapor generator that allows standard
calibration and dynamic spiking into the
extraction probe. A wide calibration range
lets you directly calibrate the analyzer at
post-dilution concentrations. The calibrator
is ideally suited for daily zero/span checks
and routine converter efficiency and
linearity testing.
Designed to meet the provisions of 40CFR
Parts 60 and 75, the Mercury Freedom
System provides a complete mercury
monitoring solution that reports elemental,
ionic, and total mercury in exhaust stacks
from both coal-fired boilers and waste
incinerators.
The Model 81ialso offers the benefit
the iSeries platform features flexible
communications, increased serviceab
and an easier to use interface.
Ethernet connectivity provides efficien
remote access, allowing the user to
download measurement information d
from the instrument without having to
on site. Captive hardware and slide-a
modules allows easy access for servic
periodic maintenance.
You can easily program short-cut keys
allow you to jump directly to frequent
accessed functions, menus or screens
The larger interface screen can displa
to five lines of measurement informat
while the primary reporting screen rem
visible.
Key Features
Performs standard calibration upstream of the
inertial filter
Capability to perform diagnostic calibrations
directly to analyzer and critical orifice
Long lasting Hg source
Uses Peltier cooler vapor pressure control
and mass flow control
Wide concentration range for post-dilution
calibration
Ideal for periodic converter efficiency and
linearity testing
p
Mercury Calibrator, Model 81iPrecision source of mercury for automated calibrations and checks
-
8/10/2019 27.Hg Calibrator 81i Spec
3/3
Gas Dilution:
Accuracy of each mass flow measurement +2% of reading or 1% of full scale, whichever is less - 20 to 100% full scaleLinearity of mass flow measurement 0.5% of full scale
Repeatability of mass flow measurement +2% of reading or 1% of full scale, whichever is less - 20 to 100% full scaleMass Flow Controller:
Zero Air Flow Controller 20 slmSpan Gas Flow Controller 50 sccmHg Source Control 0O- 15OCHg Concentration Standard Output 3 - 50 g/dscmLinearity +/- 2 % full scale
Operating Temperature 5O- 40OC
Power Requirements 250 wattsSize and Weight 16.75"(W) x 8.62"(H) x 23"(D), 48 lbs. 425 mm (W) x 219 mm (H) x 584 mm (D), 21.8 kg
Outputs Selectable Voltage, RS232/RS485, TCP/IP, 10 Status Relays, and Power FailIndication (standard). 0-20 or 4-20 mA Isolated Current Output (optional)
Inputs 16 Digital Inputs (standard), 8 0-10 VAC, Analog Inputs (optional)
CDA(20-40 psig)
Bypasssolenoid
Hg source
Hg scrubberExhaust
MFC 10-50 sccm
MFC 20-20 lpm Zero airto analyzer
Calibration gas
to analyzer
Cal gas to probeZero air to probe
RegulatorSet to 25 psig
NO
NC
C V5
NO
NCCNCC
Tempcontrol
V1
V2
V3
V4
C
C
NC
NCNO
NO
Model 81i Hg Calibrator
Hg Calibrator, Model 81i
To maintain optimal product performance, you need immediate access to experts worldwide, as well as priority status when your air quality equipme
needs repair or replacement. We offer comprehensive, flexible support solutions for all phases of the product life cycle. Through predictable, fixed-co
pricing, our services help protect the return on investment and total cost of ownership of your Thermo Scientific air quality products.
Product Specifications
Environmental
Instruments Division
Air Quality Instruments
This specification sheet is for informational purposes only and is subject to change without notice. Thermo Fisher Scientific makes no warranties, expressed or implied, in this product summary.
2009 Thermo Fisher Scientific, Inc. All rights reserved Thermo Fisher Scientific, Inc.
This product is manufactured in a plant whose quality management system is ISO 9001 certified.
Lit_Model#AQI_01/09