2390 a0119 a new stiffness parameter in air puff induced...

1
A New Stiffness Parameter in Air Puff Induced Corneal Deformation Analysis Cynthia J. Roberts, PhD 1 ; Ashraf M. Mahmoud, BS 1 ; Jeffrey P. Bons, PhD 2 ; Arif Hossain, MS 2 ; Ahmed Elsheikh, PhD 3-4 ; Riccardo Vinciguerra, MD 5 ; Paolo Vinciguerra, MD 6-7 ; Renato Ambrósio Jr, MD, PhD 8-9 1 Department of Ophthalmology & Visual Science, and Department of Biomedical Engineering, The Ohio State University, Columbus, OH, USA; 2 Department of Mechanical & Aerospace Engineering, The Ohio State University, Columbus, OH USA; 3 School of Engineering, University of Liverpool, Liverpool, United Kingdom; 4 NIHR Biomedical Research Centre for Ophthalmology, Moorfields Eye Hospital NHS Foundation Trust and UCL Institute of Ophthalmology, United Kingdom; 5 Department of Surgical Sciences, Division of Ophthalmology, University of Insubria, Varese, Italy; 6 Eye Center, Humanitas Clinical and Research Center, Via Manzoni 56, Rozzano (MI) – Italy; 7 Vincieye Clinic, Milan, Italy; 8 Rio de Janeiro Corneal Tomography and Biomechanics Study Group – Rio de Janeiro, Brazil; 9 Department of Ophthalmology, Federal University of São Paulo – São Paulo, Brazil Purpose To investigate a new stiffness parameter in corneal deformation analysis and compare responses in normal (NL) and keratoconic (KC) subjects, matched for intraocular pressure (IOP). Discussion Results All DCR’s evaluated showed a significant difference between NL and KC, except peak distance, as shown in Table 1. The KC group had lower SP values, thinner pachymetry, shorter applanation lengths, greater absolute values of applanation velocities, earlier first applanation times and later second applanation times, greater HC deformation and HC deflection amplitudes, and lower HC radius of concave curvature (greater concave curvature). All DCR’s evaluated showed a significant relationship with SP in both groups, as shown in Table 2 and Figure 5. Stiffer eyes were associated with greater pachymetry, longer applanation lengths, lower absolute value of applanation velocities, later first applanation times, earlier second applanation times, lower HC deformation and HC deflection amplitudes, shorter peak distances, greater HC radius of concave curvatures (flatter), and higher values of IOP FEM . Conclusions Dr. Roberts, Dr. P. Vinciguerra, and Dr. Ambrósio are consultants to, and had their Corvis provided by Oculus Optikgeräte GmbH. Dr. Elsheikh has received research funding from Oculus Optikgeräte GmbH . No other author has any financial or proprietary interest in any material or method discussed. Acknowledgements A0119 2390 References 1. Gallagher B, Maurice D. Striations of light scattering in the corneal stroma. Journal of ultrastructure research 1977; 61:100-14. 2. Roberts CJ, Mahmoud AM, Liu J, Sharalaya Z, Mauger TF, Lembach RG, Hendershot AJ, Kuennen R, Klyce SD. Conservation of Arclength in Keratoconic and Normal Corneas with Air Puff Induced Deformation. Invest Ophth Vis Sci. 2012;53: ARVO E-Abstract 6893. 3. Meek KM, Tuft SJ, Huang Y, Gill PS, Hayes S, Newton RH, Bron AJ. Changes in collagen orientation and distribution in keratoconus corneas. Invest Ophthalmol Vis Sci 2005; 46:1948–56. Methods A new stiffness parameter (SP) is defined as the resultant pressure at inward applanation, divided by corneal deflection amplitude at highest concavity (HCDeflectAmp). The spatial and temporal profiles of the Corvis ST air puff (Oculus, Wetzlar, Germany) were characterized using hot wire anemometry from 0 to 16mm from the nozzle. Measured velocity was correlated in time with the pressure profile exported by the Corvis ST, measured within the nozzle. The z position of the cornea at the time of inward applanation was used to calculate an adjusted air pressure value (adjAP1) at the time and position of first applanation. An algorithm to correct IOP estimation based on finite element modeling, termed IOP FEM , was used for the equation: SP = (adjAP1 - IOP FEM )/ HCDeflectAmp. Linear regression analyses between dynamic corneal response parameters (DCR’s) and SP were performed on a retrospective dataset of 180 KC eyes Figure 2: Top: Measured velocity (red) and Corvis-exported pressure signal (green), both time- synchronized by the photo cell signal (blue) Bottom: Centerline Velocity Distribution as a function of distance from the nozzle. Figure 3: Superimposed frames extracted from a single exam, showing A: Cornea in the Predeformation phase (pseudocolored blue), at maximal corneal deflection (pseudocolored red), and at maximal whole eye movement (pseudocolored white); B: Cornea at maximum deflection (Highest Concavity) with illustration of displacement from predeformation anterior surface arc (blue line); and C: Correction for whole eye motion by aligning all corneal positions to that at predeformation. § Keratoconic eyes demonstrated less resistance to deformation than normal eyes with similar IOP. § All of the deformation parameters investigated showed a significant relationship with the new stiffness parameter. § This may be useful in future biomechanical studies comparing populations Figure 1: Above: Experimental set up for hot wire anemometry; and Below: Locations for measurement of air puff velocity relative to the nozzle Figure 5: Regressions of the stiffness parameter against pachymetry and selected DCR’s in both normal and keratoconic corneas. Regressiong statistics are given in Table 2, but all show a significant relationship. A: Pachymetry showing that thicker corneas tend to be stiffer; B: A1 Velocity showing that stiffer corneas have lower velocities due to greater resistance to deformation; C: DA Ratio, showing that stiffer corneas have less difference in deformation between the center and periphery; D: Deflection Amp Max = HC DeflAmp, showing that stiffer corneas show less deflection; E: HC Radius, showing that stiffer corneas are flatter at highest concavity (HC); F: HCdarclength, showing opposite behavior between normal and keratoconic corneas. As the cornea passes into a state of concavity, the collagen fibers crimp, and the arclength shortens, similar to what occurs in the posterior stroma with an edematous cornea. [1] The shortest arclength occurs at maximum deformation. [2] In normal eyes, stiffer corneas have greater resistance to deformation and thus lower deformation/deflection amplitudes and less change in arclength with less collagen crimping. However, keratoconic corneas that are stiffer have greater shortening of their arclength. It is possible that the disruption of the collagen organization that is known to occur in keratoconus [3] leads to this behavior. Table 2: Regression Analysis Statistics between Stiffness Parameter and Dynamic Corneal Response Parameters N=158 in NL and KC Normal Keratoconus P value age 40 ± 16 35 ± 12 p =.0028 Pachymetry (µm) 542 ± 34 471 ± 36 < .0001 A1 Length (mm) 1.82 ± .08 1.72 ± .16 < .0001 A1 Vel (mm/ms) .16 ± .02 .17 ± .03 < .0001 A1 Time (ms) 7.19 ± .28 7.00 ± .28 < .0001 A1 DeflAmp (mm) .09 ± .01 .10 ± .02 < .0001 DA Ratio (unitless) 4.37 ± .42 5.81 ± .1.38 < .0001 DefA Ratio (unitless) 5.06 ± .65 7.16 ± 4.82 < .0001 DA (mm) 1.09 ± .10 1.18 ± .12 < .0001 HCDeflAmp (mm) .91 ± .11 1.01 ± .13 < .0001 HCDeflArea (mm 2 ) 3.39 ± .55 3.54 ± ..53 0.013 HCdarclength (mm) -.14 ± .02 -.12 ± .03 < .0001 Peak Distance (mm) 5.08 ± .26 5.05 ± .24 0.226 HC Radius (mm) 7.08 ± .79 5.62 ± 1.00 < .0001 InvRadMax (1/mm) .168 ± .02 .22 ± .04 < .0001 WEM Max (mm) .29 ± .07 .27 ± .06 0.0065 A2 Length (mm) 1.73 ± .31 1.53 ± .41 < .0001 A2 Vel (mm/ms) -.40 ± .08 -.47 ± .11 < .0001 A2 Time (ms) 21.78 ± .37 21.96 ± .39 < .0001 SP (mmHg/mm) 11.9 ± 3.5 7.7 ± 3.1 <.0001 Normal (N = 482) Keratoconus (N = 180) Slope KC,NL Pachymetry (µm) R 2 = .2107; p < .0001 R 2 = .1926; p < .0001 +, + A1 Length (mm) R 2 = .0234; p = .001 R 2 = .0600; p = .001 +, + A1 Vel (mm/ms) R 2 = .2107; p < .0001 R 2 = .0446; p = .005 -, - A1 Time (ms) R 2 = .5699; p < .0001 R 2 = .4916; p < .0001 +, + A1 DeflAmp (mm) R 2 = .2265; p < .0001 p = .87 +, 0 DA Ratio (unitless) R 2 = .2106; p < .0001 R 2 = .2395; p < .0001 -, - DefA Ratio (unitless) R 2 = ..0676; p < .0001 R 2 = .0460; p < ..0040 -, - DA (mm) R 2 = .4929; p < .0001 R 2 = .5189; p < .0001 -, - HCDeflAmp (mm) R 2 = .5818; p < .0001 R 2 = .5607; p < .0001 -, - HCDeflArea (mm 2 ) R 2 = .5303; p < .0001 R 2 = .4645; p < .0001 -, - HCdarclength (mm) R 2 = .0569; p < .0001 R 2 = .0237; p = .041 +, - Peak Distance (mm) R 2 = .5968; p < .0001 R 2 = .3605; p < .0001 -, - HC Radius (mm) R 2 = .0976; p < .0001 R 2 = .2006; p < .0001 +, + InvRadMax (1/mm) R 2 = .0846; p < .0001 R 2 = .2247; p < .0001 +, + WEM Max (mm) R 2 = .0221; p = .001 p = .47 +, 0 A2 Length (mm) R 2 = .0906; p < .0001 R 2 = .0665; p = .001 +, + A2 Vel (mm/ms) R 2 = .4661; p < .0001 R 2 = .3027; p < .0001 +, + A2 Time (ms) R 2 = .2581; p < .0001 R 2 = .2448; p < .0001 +, + Z Y X Hot Wire probe Photocell Sensor 0 20 40 60 80 100 0 0.2 0.4 0.6 0.8 1 Time (ms) Normalized value Signal detection Photocell response (Corvis LED) Velocity measurement at x = 2mm Corvis internal pressure signal A 0 5 10 15 20 25 30 35 40 0 20 40 60 80 100 120 140 Time [ms] Velocity [m/s] Centerline velocity distribution (Temporal) z = 2mm z = 4mm z = 6mm z = 8mm z = 10mm z = 12mm z = 14mm z = 16mm B and 482 NL eyes. DCR’s from a subset of 158 eyes of 158 subjects in each group were matched for IOP FEM and compared using t-tests. Significance threshold was p < 0.05. A B C D E F Figure 6: Regression analysis of HC Deflection Amp vs HC darclength, showing that in normal corneas, the greater the depth of deformation, the greater the change in arclength (shorter arclength at maximum deflection). However, in keratoconic corneas, the greater depth of deformation was associated the least change in arclength, possibly due to disruption of collagen in keratconus. (NL: R 2 = .2627, p< .0001; KC: R 2 = .0307, p < .0185). Table 1: Mean ± Standard Deviation in IOP FEM -Matched t-test Comparison Note: In DefA Ratio, one KC eye was excluded as an outlier. Note: In DefA Ratio, one KC eye was excluded as an outlier. Figure 4: Phases of Deformation

Upload: trinhkhanh

Post on 18-Aug-2019

213 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 2390 A0119 A New Stiffness Parameter in Air Puff Induced ...livrepository.liverpool.ac.uk/3006904/1/Roberts - Corvis stiffness... · exam, showing A: Cornea in the Predeformation

A New Stiffness Parameter in Air Puff Induced Corneal Deformation AnalysisCynthia J. Roberts, PhD1; Ashraf M. Mahmoud, BS1; Jeffrey P. Bons, PhD2; Arif Hossain, MS2; Ahmed Elsheikh, PhD3-4; Riccardo Vinciguerra, MD5; Paolo Vinciguerra, MD6-7; Renato Ambrósio Jr, MD, PhD8-9

1Department of Ophthalmology & Visual Science, and Department of Biomedical Engineering, The Ohio State University, Columbus, OH, USA; 2Department of Mechanical & Aerospace Engineering, The Ohio State University, Columbus, OH USA;3School of Engineering, University of Liverpool, Liverpool, United Kingdom; 4NIHR Biomedical Research Centre for Ophthalmology, Moorfields Eye Hospital NHS Foundation Trust and UCL Institute of Ophthalmology, United Kingdom;

5Department of Surgical Sciences, Division of Ophthalmology, University of Insubria, Varese, Italy; 6Eye Center, Humanitas Clinical and Research Center, Via Manzoni 56, Rozzano (MI) – Italy; 7Vincieye Clinic, Milan, Italy; 8Rio de Janeiro Corneal Tomography and Biomechanics Study Group – Rio de Janeiro, Brazil; 9Department of Ophthalmology, Federal University of São Paulo – São Paulo, Brazil

Purpose To investigate a new stiffness parameter in corneal deformation analysis andcompare responses in normal (NL) and keratoconic (KC) subjects, matchedfor intraocular pressure (IOP).

Discussion

Results

All DCR’s evaluated showed a significant difference between NLand KC, except peak distance, as shown in Table 1. The KCgroup had lower SP values, thinner pachymetry, shorterapplanation lengths, greater absolute values of applanationvelocities, earlier first applanation times and later secondapplanation times, greater HC deformation and HC deflectionamplitudes, and lower HC radius of concave curvature (greaterconcave curvature). All DCR’s evaluated showed a significantrelationship with SP in both groups, as shown in Table 2 andFigure 5. Stiffer eyes were associated with greater pachymetry,longer applanation lengths, lower absolute value of applanationvelocities, later first applanation times, earlier secondapplanation times, lower HC deformation and HC deflectionamplitudes, shorter peak distances, greater HC radius of concavecurvatures (flatter), and higher values of IOPFEM.

Conclusions

Dr. Roberts, Dr. P. Vinciguerra, and Dr. Ambrósio are consultants to, and had their Corvis provided by Oculus Optikgeräte GmbH. Dr. Elsheikh has received research fundingfrom Oculus Optikgeräte GmbH . No other author has any financial or proprietary interest in any material or method discussed.

Acknowledgements

A01192390

References1. Gallagher B, Maurice D. Striations of light scattering in the corneal stroma. Journal of ultrastructure research 1977; 61:100-14.2. Roberts CJ, Mahmoud AM, Liu J, Sharalaya Z, Mauger TF, Lembach RG, Hendershot AJ, Kuennen R, Klyce SD. Conservation of Arclength in

Keratoconic and Normal Corneas with Air Puff Induced Deformation. Invest Ophth Vis Sci. 2012;53: ARVO E-Abstract 6893.3. Meek KM, Tuft SJ, Huang Y, Gill PS, Hayes S, Newton RH, Bron AJ. Changes in collagen orientation and distribution in keratoconus corneas. Invest

Ophthalmol Vis Sci 2005; 46:1948–56.

MethodsA new stiffness parameter (SP) is defined as the resultant pressure at inwardapplanation, divided by corneal deflection amplitude at highest concavity(HCDeflectAmp). The spatial and temporal profiles of the Corvis ST air puff(Oculus, Wetzlar, Germany) were characterized using hot wire anemometryfrom 0 to 16mm from the nozzle. Measured velocity was correlated in timewith the pressure profile exported by the Corvis ST, measured within thenozzle. The z position of the cornea at the time of inward applanation wasused to calculate an adjusted air pressure value (adjAP1) at the time andposition of first applanation. An algorithm to correct IOP estimation basedon finite element modeling, termed IOPFEM, was used for the equation: SP =(adjAP1 - IOPFEM)/ HCDeflectAmp. Linear regression analyses betweendynamic corneal response parameters (DCR’s) and SP were performed on aretrospective dataset of 180 KC eyes

Figure 2: Top: Measured velocity (red) and Corvis-exported

pressure signal (green), both time-synchronized by the photo cell

signal (blue) Bottom: Centerline Velocity Distribution as a function

of distance from the nozzle.

Figure 3: Superimposed frames extracted from a single exam, showing A: Cornea in the Predeformation phase (pseudocolored blue), at maximal corneal deflection

(pseudocolored red), and at maximal whole eye movement (pseudocolored white); B: Cornea at maximum deflection (Highest Concavity) with illustration of displacement from

predeformation anterior surface arc (blue line); and C: Correction for whole eye motion by aligning all corneal

positions to that at predeformation.

§ Keratoconic eyes demonstrated less resistance to deformation than normal eyes with similar IOP.

§ All of the deformation parameters investigated showed a significant relationship with the new stiffness parameter.

§ This may be useful in future biomechanical studies comparing populations

Figure 1: Above: Experimental set up for hot

wire anemometry; and Below: Locations for measurement of air puff velocity relative to the

nozzle

Figure 5: Regressions of the stiffness parameter against pachymetry and selected DCR’s in both normal and keratoconic corneas. Regressiong statistics are given in Table 2, but all show a significant relationship. A: Pachymetry showing that thicker corneas tend to bestiffer; B: A1 Velocity showing that stiffer corneas have lower velocities due to greater resistance to deformation; C: DA Ratio, showing that stiffer corneas have less difference in deformation between the center and periphery; D: Deflection Amp Max = HC DeflAmp,showing that stiffer corneas show less deflection; E: HC Radius, showing that stiffer corneas are flatter at highest concavity (HC); F: HCdarclength, showing opposite behavior between normal and keratoconic corneas. As the cornea passes into a state of concavity, thecollagen fibers crimp, and the arclength shortens, similar to what occurs in the posterior stroma with an edematous cornea. [1] The shortest arclength occurs at maximum deformation. [2] In normal eyes, stiffer corneas have greater resistance to deformation and thuslower deformation/deflection amplitudes and less change in arclength with less collagen crimping. However, keratoconic corneas that are stiffer have greater shortening of their arclength. It is possible that the disruption of the collagen organization that is known tooccur in keratoconus [3] leads to this behavior.

Table 2: Regression Analysis Statistics between Stiffness Parameter and Dynamic Corneal Response Parameters

N=158 in NL and KC Normal Keratoconus P valueage 40 ± 16 35 ± 12 p =.0028Pachymetry (µm) 542 ± 34 471 ± 36 < .0001A1 Length (mm) 1.82 ± .08 1.72 ± .16 < .0001A1 Vel (mm/ms) .16 ± .02 .17 ± .03 < .0001A1 Time (ms) 7.19 ± .28 7.00 ± .28 < .0001A1 DeflAmp (mm) .09 ± .01 .10 ± .02 < .0001DA Ratio (unitless) 4.37 ± .42 5.81 ± .1.38 < .0001DefA Ratio (unitless) 5.06 ± .65 7.16 ± 4.82 < .0001DA (mm) 1.09 ± .10 1.18 ± .12 < .0001HCDeflAmp (mm) .91 ± .11 1.01 ± .13 < .0001HCDeflArea (mm2) 3.39 ± .55 3.54 ± ..53 0.013HCdarclength (mm) -.14 ± .02 -.12 ± .03 < .0001Peak Distance (mm) 5.08 ± .26 5.05 ± .24 0.226HC Radius (mm) 7.08 ± .79 5.62 ± 1.00 < .0001InvRadMax (1/mm) .168 ± .02 .22 ± .04 < .0001WEM Max (mm) .29 ± .07 .27 ± .06 0.0065A2 Length (mm) 1.73 ± .31 1.53 ± .41 < .0001A2 Vel (mm/ms) -.40 ± .08 -.47 ± .11 < .0001A2 Time (ms) 21.78 ± .37 21.96 ± .39 < .0001SP (mmHg/mm) 11.9 ± 3.5 7.7 ± 3.1 <.0001

Normal (N = 482) Keratoconus (N = 180)Slope

KC,NLPachymetry (µm) R2 = .2107; p < .0001 R2 = .1926; p < .0001 +, +A1 Length (mm) R2 = .0234; p = .001 R2 = .0600; p = .001 +, +A1 Vel (mm/ms) R2 = .2107; p < .0001 R2 = .0446; p = .005 -, -A1 Time (ms) R2 = .5699; p < .0001 R2 = .4916; p < .0001 +, +A1 DeflAmp (mm) R2 = .2265; p < .0001 p = .87 +, 0DA Ratio (unitless) R2 = .2106; p < .0001 R2 = .2395; p < .0001 -, -DefA Ratio (unitless) R2 = ..0676; p < .0001 R2 = .0460; p < ..0040 -, -DA (mm) R2 = .4929; p < .0001 R2 = .5189; p < .0001 -, -HCDeflAmp (mm) R2 = .5818; p < .0001 R2 = .5607; p < .0001 -, -HCDeflArea (mm2) R2 = .5303; p < .0001 R2 = .4645; p < .0001 -, -HCdarclength (mm) R2 = .0569; p < .0001 R2 = .0237; p = .041 +, -Peak Distance (mm) R2 = .5968; p < .0001 R2 = .3605; p < .0001 -, -HC Radius (mm) R2 = .0976; p < .0001 R2 = .2006; p < .0001 +, +InvRadMax (1/mm) R2 = .0846; p < .0001 R2 = .2247; p < .0001 +, +WEM Max (mm) R2 = .0221; p = .001 p = .47 +, 0A2 Length (mm) R2 = .0906; p < .0001 R2 = .0665; p = .001 +, +A2 Vel (mm/ms) R2 = .4661; p < .0001 R2 = .3027; p < .0001 +, +A2 Time (ms) R2 = .2581; p < .0001 R2 = .2448; p < .0001 +, +

Z

Y

X

HotWireprobe

PhotocellSensor

0 20 40 60 80 1000

0.2

0.4

0.6

0.8

1

Time (ms)

Nor

mal

ized

val

ue

Signal detection

Photocell response (Corvis LED)Velocity measurement at x = 2mmCorvis internal pressure signal

A

0 5 10 15 20 25 30 35 400

20

40

60

80

100

120

140

Time [ms]

Vel

ocity

[m/s

]

Centerline velocity distribution (Temporal)

z = 2mmz = 4mmz = 6mmz = 8mmz = 10mmz = 12mmz = 14mmz = 16mm

B

and 482 NL eyes. DCR’s from asubset of 158 eyes of 158 subjects ineach group were matched for IOPFEMand compared using t-tests.Significance threshold was p < 0.05.

A B C D E F

Figure 6: Regression analysis of HCDeflection Amp vs HC darclength, showingthat in normal corneas, the greater the depthof deformation, the greater the change inarclength (shorter arclength at maximumdeflection). However, in keratoconiccorneas, the greater depth of deformationwas associated the least change in arclength,possibly due to disruption of collagen inkeratconus. (NL: R2 = .2627, p< .0001; KC:R2 = .0307, p < .0185).

Table 1: Mean ± Standard Deviation in IOPFEM-Matched t-test Comparison

Note: In DefA Ratio, one KC eye was excluded as an outlier. Note: In DefA Ratio, one KC eye was excluded as an outlier.

Figure 4: Phases of Deformation