2. digital instrumentselectronicsandinstrumentation.weebly.com/uploads/1/0/1/6/... · web...

27
2. DIGITAL INSTRUMENTS INTRODUCTION: The digital instruments display measurement of electrical parameters like voltage and current as discrete numerals instead of pointer deflection on a continuous scale as in analog devices. Numerical read out is advantageous in many applications because it reduces human error, parallax error, increases reading speed and often provides outputs in digital form suitable for further processing or recording. Digital instruments, particularly digital voltmeters or multimeters, are used to measure analog quantity. It is, therefore, necessary to convert the analog signal to an equivalent digital signal. So, analog-to-digital converters (A/D converters) are also used as a main element of a digital instrument. The basic building block of a digital instrument is shown in fig. Fig (a): Building Block of a Digital Instrument The display block may be analog or digital in nature. If an analog readout is desired, it becomes necessary to include a stage involving digital to analog conversion. 2.1 ADVANTAGES OF DIGITAL INSTRUMENTS OVER ORDINARY INSTRUMENTS : 1. It reduces human reading & interpolation errors. 2. It eliminates parallax error. 3. Increases reading speed. 4. Digital instruments have high accuracy up to ± 0.003%. 5. They have high resolution. 6. They possess good reliability and repeatability. 7. Loading effect is eliminated due to their high input impedance. 8. They have high sensitivity. A.A.N.M & V.V.R.S.R POLYTECHNICPage 1 Analog To Digital Converter Signal Processing Display

Upload: others

Post on 26-Jan-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

2. DIGITAL INSTRUMENTS

INTRODUCTION:

The digital instruments display measurement of electrical parameters like voltage and current as discrete numerals instead of pointer deflection on a continuous scale as in analog devices. Numerical read out is advantageous in many applications because it reduces human error, parallax error, increases reading speed and often provides outputs in digital form suitable for further processing or recording.

Digital instruments, particularly digital voltmeters or multimeters, are used to measure analog quantity. It is, therefore, necessary to convert the analog signal to an equivalent digital signal. So, analog-to-digital converters (A/D converters) are also used as a main element of a digital instrument. The basic building block of a digital instrument is shown in fig.

Signal

Processing

Analog

To

Digital

Converter

Display

Fig (a): Building Block of a Digital Instrument

The display block may be analog or digital in nature. If an analog readout is desired, it becomes necessary to include a stage involving digital to analog conversion.

2.1 ADVANTAGES OF DIGITAL INSTRUMENTS OVER ORDINARY INSTRUMENTS:

1. It reduces human reading & interpolation errors.

2. It eliminates parallax error.

3. Increases reading speed.

4. Digital instruments have high accuracy up to ± 0.003%.

5. They have high resolution.

6. They possess good reliability and repeatability.

7. Loading effect is eliminated due to their high input impedance.

8. They have high sensitivity.

9. They have wide range of frequency response.

10. They have wide range of frequency measurement.

11. They are portable due to their small size.

12. The digital instruments also have greater speed.

13. They have very low power consumption.

14. It provides output in digital form suitable for further processing and recording.

15. They have high input range from ±1.000000 to ± 1,000.000 V.

2.2 DIGITAI VOLTMETERS (DVM) :

Digital voltmeters are measuring instruments that convert analog voltage signals into a digital or numeric readout. This digital readout can be displayed on the front panel and also used as an electrical digital output signal.

The digital voltmeter (DVM) displays ac and dc voltages as discrete numbers, rather than as a pointer on a continuous scale as analog voltmeter. A numerical readout is advantageous because it reduces human error, eliminates parallax error, increases reading speed and often output in digital form suitable for further processing and recording.

The DVM is a versatile and accurate instrument that can be used in many laboratory measurement applications. With the development of IC modules, the size, power requirements and cost of DVMs have been reduced, so that DVM compete with analog voltmeters in portability and price.

Digital voltmeters can be classified according to the following broad categories

(a) Ramp-type DVM

(b) Dual slope integrating type DVM (Voltage to time conversion)

(c) Integrating type DVM (Voltage to frequency conversion)

(d) Staircase - Ramp DVM

(e) Successive - Approximation DVM

(f) Continuous-Balance type DVM.

2.2 (a) Ramp-Type DVM:

Operating Principle: The operating principle in the Ramp type DVM is based on the measurement of the time it takes for a linear ramp voltage to, rise from 0V to the level of the input voltage, or to decrease from the level of the input voltage to zero. This time interval is measured with an electronic time interval counter and the count is displayed as a number of digits on electronic indicator.

The block diagram of a Ramp type DVM is as shown in Fig 1 .

FIG 1: Block Diagram of Ramp Type DVM

Description: The Ramp type DVM consists of a ranging & attenuator which is used to select the required range of measurement. A ramp generator is used to generate the ramp voltage. An input comparator continuously compares the dc input voltage with the ramp voltage. The ground comparator compares the ramp voltage with the ground voltage i.e., 0 V. A counter is used to count the number of pulses and the read out gives the digital display.

Working :

At the start of the measurement cycle a ramp voltage is initiated; this voltage can be positive going or negative going. This -ve going ramp is illustrated in Fig. 2.

FIG 2: Voltage to Time Conversion

This negative going ramp is continuously compared with the unknown input voltage. At the instant that the ramp voltage equals to unknown voltage, the comparator generates the pulse which opens the gate. The ramp voltage continues to decrease with time until it finally reaches 0 V and a second comparator generates an output pulse which closes the gate.

An oscillator generates clock pulses which are allowed to pass through the gate to a number of decade counting units which totalize the number of pulses passed through the gate. The decimal number, displayed by the indicator tubes is a measure of the magnitude of the input voltage. The sampled rate multivibrator determines the rate at which the measurement cycles are initiated. The sample rate circuit provides an initiating pulse for the ramp generator to start its next ramp voltage .At the same time, a reset pulse is generated which resets the counter to the zero state.

Advantages and Disadvantages:

The ramp technique circuit is easy to design and its cost is low. Also, the output pulse can be transmitted over long feeder lines. However, the single ramp requires excellent characteristics regarding linearity of the ramp and time measurement. Large errors are possible when noise is superimposed on the input signal. Input filters are usually required with this type of converter.

2.2 (b) Successive Approximation Type Digital Voltmeter:

The successive approximation type DVM works on the principle of balancing the weights as in a simple balance. To understand the concept clearly, let us consider we want to measure the weight of some unknown quantity of sugar. What do we do? First we approximate the weight of sugar to some known weight, If the weight of sugar is more than known weight, then we add some more weights to the known weight. If it is less, then we replace the weight with a lesser value. This process is repeated until the pointer balances the two weights. The successive approximation type DVM uses the same principle.

The basic block diagram of a successive approximation type DVM is as shown in the Fig:3

FIG 3 : Successive Approximation Type DVM

Description : This DVM consists of an input attenuator for selecting the desired range of input voltage and also to attenuate any noise in the given voltage. This selected input is applied to the comparator through a sample and hold circuit. The successive approximation register (SAR) receives its 8 bit input from the ring counter after each clock pulse. This Input is applied to the Digital to Analog converter which converts the digital data into Analog voltage. This voltage is applied as second input to the comparator. The o/p of the AND gate goes high when there is a positive o/p at the comparator. Finally, the digital output is taken out from the successive approximation register with input voltages other than dc; the input level changes during digitization and decision made during conversion are not consistent. To avoid this error, a sample and hold circuit is used and placed in the input directly following the input attenuator. This digital voltmeter is capable of 1000 readings per second.

Working: When the start pulse signal activates the control circuit, the SAR is cleared. Therefore, the output of the SAR is 0000 0000.V OUT of the D/A converter is 0. Now if

V in > V out, the comparator output is positive. During the first clock pulse, the control circuit sets D 7, to 1 and V out jumps to the half of reference voltage. The SAR o/p is 10000000. If V out is greater than V in, the comparator o/p is negative and the control circuit resets D 7. However, if V in is greater than V out, the comparator output is positive and the control circuit keeps D 7 set. Similarly the reset of the bits beginning from D 7 to D0 are set and tested. Therefore the measurement is completed in eight clock pulses.

Example: Suppose the converter can measure a maximum of 5 V. i.e. 5 V corresponds to the maximum count of 11111111. If the test voltage V in = 1V the following steps will take place in the measurement.

FIG 4 : Various Output Levels for Each Bit ( 8-Bit Shows the Voltage Level Very

Nearly Equal to 1V)

2.2 (c) DUAL SLOPE INTEGRATING TYPE DVM : (VOLTAGE TO TIME CONVERSION)

In the ramp techniques, superimposed noise can causes large errors. In the dual ramp technique noise is averaged out by positive and negative ramp using the process of integration

As shown in figure 8(b) the input voltage ‘ei ‘ is integrated with the slope of the integrator output proportional to the test input. After a fixed time equal to t1 the input voltage is disconnected and the input voltage is connected to negative voltage -er the integrator output will have a negative slope which is constant and proportional to the magnitude of the input voltage.

2. DIGITAL INSTRUMENTS

A.A.N.M & V.V.R.S.R POLYTECHNICPage 4

FIG 8 (a) : Block Diagram of Dual Slope Integrator DVM

FIG 8 (b) : Waveforms of Dual Slope Integrator

At the start a pulse reset the counter and the F/F output to logic ‘0’. Si is closed and Sr is open. The capacitor begins to charge as soon as the integrator output exceeds zero the comparator output voltage changes state, which opens the gate so that the oscillator clock pulses are fed to the counter.(when the ramp voltage starts, the comparator goes to state 1, the gate opens and clock pulse drives the counter)when the counter reaches the maximum count, i.e. the counter is made to run for a time ‘t1’ in this case 9999, on the next clock pulse all the digits go to 0000 and the counter activates the F/F to logic level ‘1’. This activates the switch drive, ei is disconnected and -er is connected to the integrator. The integrator output will have negative slope which is constant, i.e. integrator output is now decreases linearly to zero volts. Comparator output state changes again and locks the gate. The discharge time t2 is now proportional to the input voltage. The counter indicates the count during t2. When the slope of the integrator reaches the zero, the comparator switches to state ‘0’ and the gate closes and the capacitor is now discharge with constant slope. As soon as the eo is zero the counter is stopped. The pulses counted thus has the directly relation with the input voltage.

During charging

eo= -1/ RC ∫ot 1 eidt = -ei t 1/ RC(eq-1)

During discharging ( For analysis reference source value is taken as -er)

eo = 1/ RC ∫ot 2 –er dt = -er t 2 / RC(eq-2)

Subtracting Eqs 2 from 1 we have

eo - eo = -er t 2 /RC – (-ei t 1/ RC)

0 = -er t 2 /RC – (-ei t 1/ RC)

er t 2 /RC = ei t 1/ RC

There fore ei = er (t 2/ t 1)(eq-3)

If the oscillator period equals T and the digital counter indicates n 1and n 2 counts respectively.

ei = (n 2 T/ n 1 T) er i.e. ei = (n 2 / n 1) er

Now, n 1 and er are constants. Let K 1= er/ n 1.Then ei =K1 n 2 (eq-4)

From Eq -3 it is evident that the accuracy of the measured voltage is independent of the integrator time constant The times t 1 and t 2 are measured by the count of the clock given by the numbers n 1 and n 2 respectively. The clock oscillator period equals T and if n 1and er are constants, then Eq-4 indicates that the accuracy of the method is also independent of the oscillator frequency.

Advantages & Disadvantages of Dual Slope Technique :

(a) The dual slope technique has excellent noise rejection as the noise is averaged over in

the process of integration.

(b) The speed and accuracy can be easily varied according to demands of the

measurement situation. The speed is high.

(c) The only source of error is the reference voltage. Hence the system is suitable for

accurate measurements. An accuracy of ± 0.05% in 100 ms is available. Accuracy is

independent of oscillator frequency.

d) Filters are not required

e) Highly stable.

Comparison with ramp type voltmeter.

S.NO.

Ramp Type

Dual Slope

1

2

3

4

5

6

Large errors are possible when noise is superimposed on the input signal.

Circuit complexity is low

Input filters are required

Low accuracy and accuracy depends on the stability of the oscillator and the linearity of the ramp slope

Poor Stability

Operating speed slow

Has excellent noise rejection because noise and superimposed ac are averaged out in the process of integration.

Circuit complexity is moderate.

Filters not required

Accuracy of this DVM is high. Accuracy is independent of oscillator frequency.

High Stable

Operating speed high

Comparison of Various types of DVM’s :

System

Accuracy

Input Impedance

Speed

Staircase ramp

Dual –Slope integration

Successive Approximation

High

High

High

Not constant (Low to high)

Very high

Not constant(Low to high)

Medium

Medium

High

2.3 GENERAL SPECIFICATIONS OF DIGITAL VOLTMETERS :

I. Display: 3-1/2 digits, LCD.

2. Input Range: from ± 1.000 V to ± 1000 V, with automatic range selection and overload

indication.

3. Over-Range Indication: Only (1) or (-1) displayed at the MSB position.

4. Absolute Accurcuracy : as high as ±0.005 percent of the reading.

5. Stability : Short-term, 0.002% of the reading for a 24 period; long-term. 0.008% of the

reading for a 6-month period.

6. Resolution : 1 part in 10 6 (1μV can be read on the 1V input range)

7. Input characieristics : Input resistance typically 10 MΩ; input capacitance typically

40 PF.

8.Calibration : Internal calibration standard allows calibration independent of the

measuring circuit.

9. Output signals : Print command allows output to printer; BCD output for digital

processing or recording.

10. Zero adjustment: Automatic

11. Functions : DC volts, AC volts, DC amps AC amps, ohms, diode test.

12. Manual : Switch selection as desired.

13. Temperature: Operating 0°C — 60°C, 70% RH (Relative humidity)

14. Polarity : AUTO negative polarity indication .

15. Low Battery: B mark on LCD readout.

3 ½ - DIGIT

The number of digit positions used in a digital meter determines the resolution. Hence a 3 digit display on a DVM for a 0-1 V range will indicate values from 0-999 mV. Normally , a fourth digit capable of indicating 0 or 1 (hence called a Half Digit) is placed to the left. This permits the digital meter to read values above 999 up to 1999, to give overlap between ranges for convenience, a process called over-ranging. This type of display is called a 3½ digit display, shown in Fig : 9.

FIG 9 : 3 ½ Digit Display

5,9. RESOLUTION AND SENSITIVITY OF OIGITAL METERS

Resolution

If n = number of full digits, then resolution (R) is 1/10 n .

The resolution of a DVM is determined by the number of full or active digits

used,

I f n=3, R=1/10 n =1/10 3 =0.001 or 0.1%

Sensitivity of Digital Meters

Sensitivity is the smallest change in input which a digital meter is able to detect. Hence, it is the full scale value of the lowest voltage range multiplied by the meter’s resolution.

Sensitivity S = (fs) min x R

where (fs) min = lowest full scale of the meter

R = resolution expressed as decimal

.Example 1: What is the resolution of a 3½ digit display on I V and 10 V ranges?

Solution Number of full digits is 3, Therefore, resolution is 1/10 n where

n =3. Resolution R =1/10 n = 1/103 = 0.001

Hence the meter cannot distinguish between values that differ from each other by less than 0.001 of full scale.

For full scale range reading of 1V, the resolution is 1 x 0.001 = 0.001 V.

For full scale reading of 10 V range, the resolution is 10 V x 0.001 = 0.01 V.

Hence on 10 V scale, the meter cannot distinguish between readings that differ by less than 0.01 V.

2.4 DIGITAL MULTIMETER’S:

The three major classes of digital meters are panel meters, bench type meters and system meters.

All digital meters employ same kind of analog to digital (often dual slope integrating type) and have a visible readout display at the converter output.

Panel meters are usually placed at one location (and perhaps even a fixed range). while bench meters and system meters are often multimeters, i.e. they can read ac and dc voltage currents and resistances over several ranges. Bench meters are intended mainly for stand alone operation and visual operation reading, while system meters provide at least an electrical binary coded decimal output (in parallel with the usual display), and perhaps sophisticated interconnection and control capabilities, or even microprocessor based computing power.

The basic circuit shown in Fig10:(a) is always a dc voltmeter. Current is converted to voltage by passing it through a precision low shunt resistance while alternating current is converted into dc by employing rectifiers and filters. For resistance measurement, the meter includes a precision low current source that is applied across the unknown resistance; again this gives a dc voltage which is digitized and readout as ohms.

FIG 10 : (a) Digital Multimeter

The basic digital multimeter (DMM) is made up of several A/D converters, circuitry for counting and an attenuation circuit. A basic block diagram of a DMM is shown in Fig10 (b). The current to voltage converter shown in the block diagram of Fig. 10 (b) can be implemented with the circuit shown in fig 10 (c).

FIG 11 : (b) Block Diagram of a Basic Multimeter

The current to be measured is applied to the summing junction (∑ i) at the input of the opamp. Since the current at the input of the amplifier is close to zero because of the very high input impedance of the amplifier, the current I R is very nearly equal to I i, the current I R causes a voltage drop which is proportional to the current, to be developed across the resistors. This voltage drop is the input to the A/D converter, thereby providing a reading that is proportional to The unknown current.

FIG 13 : (c) Current to Voltage Converter

Resistance is measured by passing a known current, from a constant current source, through an unknown resistance. The voltage drop across the resistor is applied to the A/D converter, thereby producing an indication of the value of the unknown resistance .

2.5 SPECIFICATIONS OF DIGITAL MULTIMETER:

1. Display :3 ½ digit LCD with a maximum reading of 1999.

2. Polarity : Automatic, (-)negative polarity indication.

3. Over range indication : (OL) or (-OL) is displayed.

4. Operating Environment : 0 0 c to 50 0 c at <70% R.H. (Relative Humidity)

5. Input Impedance :10MΩ in all DCV and ACV ranges.

6. Power requirements :9V Battery.

7. Dimensions: 91 mm (w) Χ 170 mm (L) Χ40 mm (H)

8.Weight : 330 gms

9. DC volts :

Ranges : 200 mv,2V,20V,200V, 1000V

Resolution :100 μV

Accuracy : ± 0.5% of reading

10. AC volts :

Ranges : 200mV,2V, 20V,200V,750V

Accuracy : ± 1% of reading (40-60Hz)

Resolution : 100 μV

11. DC current :

Ranges : 200 μA,2mA,20 mA,200mA,2A,10A.

Accuracy : ±0.5% of reading

12 AC current :

Ranges :2 mA,200mA,10 A

Accuracy : ± 1% of reading

13. Resistance :

Ranges : 200Ω,2KΩ,20KΩ,200KΩ,2MΩ,20MΩ

Accuracy : ±0.5% of reading

14. Diode test :

Test Current 0.8 mA ±0.3mA

Open current volts :3.0 V

15.Transistor,h f e :

Ranges :0-1000

Base current : 10 μ A DC approx. (V c e=3.0V DC)

2.6 DIGITAL FREQUENCY METER :

Principle of Operation

The signal waveform is converted to trigger pulses and applied continuously to an AND gate, as shown in Fig 14. A pulse of 1s is applied to the other terminal, and the number of pulses counted during this period indicates the frequency.

FIG 14 : Principle Of Digital Frequency Measurement

The signal whose frequency is to be measured is converted into a train of pulses, one pulse for each cycle of the signal. The number of pulses occurring in a definite interval of time is then counted by an electronic counter. Since each pulse represents the cycle of the unknown signal, the number of counts is a direct indication of the frequency of the signal. Since electronic counters have a high speed of operation, high frequency signals can be measured.

Basic Circuit of a Digital Frequency Meter :

The block diagram of a basic circuit of a digital frequency meter is shown in Fig 15.

FIG 15 : Basic Circuit of a Digital Frequency Meter

The signal may be amplified before being applied to the Schmitt trigger. The Schmitt trigger converts the input signal into a square wave with fast rise and fall times, which is then differentiated and clipped. As a result, the output from the Schmitt trigger is a train of pulses one pulse for each cycle of the signal.

The output pulses from the Schmitt trigger are fed to a START/STOP gate. When this gate is enabled, the input pulses pass through this gate and are fed directly to the electronic counter, which counts the number of pulses.

When this gate is disabled, the counter stops counting the incoming pulses. The counter displays the number of pulses that have passed through it in the time interval between start and stop. If this interval is known, the unknown frequency can be measured.

The block diagram of a digital frequency meter is shown in fig 16.

FIG 16 : Diagram of a Frequency Meter

The input signal is amplified and converted to a square wave by a Schmitt trigger circuit. In this diagram, the square wave is differentiated and clipped to produce a train of pulses, each pulses separated by the period of the input signal. The time base selector output is obtained from an oscillator and is similarly converted into positive pulses.

The first pulse activates the control Flip-Flop, this gate control F/F provides an enable signal to the AND gate. The trigger pulses of the input signals are allowed to pass through the gate for a selected time period and counted. The second pulse from the decade frequency divider changes the state of control F/F and removes the enable signal from the AND gate, thereby closing it. The decimal counter and display unit output corresponds to the number of input. Pulses received during a precise time interval; hence the counter display corresponds to the frequency.

The frequency of the input signal‘s computed as

F=N / t

Where,

F= Frequency of the input signal

N = Number of pulses counted

t = Duration of gate pulses

In some applications it is desirable to measure the timeperiod of the signal,the timeperiod is computed as

T = N / f

Where,

T = Time period of input signal.

N = Number of pulses countyed,

f = Clock frequency.

2.7 SPECIFICATIONs OF DIGITAL FREQUENCY METER :

The front panel of digital frequency counter is shown in fig 17 and also its specifications are shown below :

FIG 17 : Front Panel of Digital Frequency Meter

The specifications of digital frequency meter are listed below.

1. Frequency range: 10 Hz - 300 MHz (max)

2. No. of digits: 5 to 8 digits (LED display)

3.Response time: 0.2 seconds

4.Input sensitivity: kHz range

10Hz- I MHz,50mVrms

MHz range

1MHz-300MHz, 40mVrms

5.Resolution: 0.1 Hz (at 10 sec gate time)

100 Hz (at 0.01 sec gate time)

6.Time Base: 10 MHz (Crystal osc ±10 ppm)

7.Power supply: 230 V 50Hz a.c

8.Impedance: 3M Ω

9.Dimensions: 250 mm x 170mm x 70mm

2.8 BLOCK DIAGRAM OF DIGITAL LCR METER (DIGITAL IMPEDANCE METER ) :

FIG 18 : Block Diagram of Digital LCR Meter

The above block diagram of digital LCR meter, explains,

1. Switch circuit2. Conversion circuits

3. Phase sensitive detector4. A/D converter

5. Decade counter6. Digital read out

1. Switching Circuit:

(a) Manual Switch

(b)Automatic/Inside Semiconducting Switch

For selecting measurement of Inductance (L), capacitance (C), resistance (R) finding unknown values of above components, front panel controls of LCR meter has manual/ Automatic switches, whose can switch for required flow for measuring unknown values switch position 1, 2, 3 meant for measuring unknown inductance (L),capacitance‘C’, resistance (R), as indicated in above block diagram (Fig 18).

2. Conversion Circuit:

(a) For measurement of unknown inductance

(b)For measurement of unknown capacitance.

FIG 19 : Conversion Circuit of Inductive Impedance into voltage

Conversion circuit is consisting of Op-amp voltage given to non-inverting terminal and measure the current in unknown Inductance’ L’, (i.e.) the current is ‘I ‘ through resistor gives voltage developed across ‘R’.

Input voltage = 1.6 V

Resistor = 1 k Ω

I=1.6V/1K Ω = 1.6mA

Then V L = I . X L = I . (2πfL) =1.6 x 2π x 1kHz x assume some value

ie, 100 mH

= 1 V(rms)

If L= Assumed value= 200 mH

Then V L = 1.6 x 2π x 1kHz x 200mH

= 3V

i.e., we conclude that voltage developed across ‘L’ is directly proportional to the

Inductive impedance.

For Measuring Capacitance Value :

FIG 20 : Conversion Circuit for Capacitive Impedance into Voltage

Input voltage is developed across the capacitor and the output voltage is measured across the resistor

I = V / Xc

V R = I R

V = Input voltage = 1.6 V (Vrms)

f = 1KHz

R = 1 kΩ

Assume ‘C’ value is 0.1 μF

Then I =V/Xc = V(2πfc) = 1.6x 2π x1kHz x 0.1μF

= 1mA

V = IR =1mA x1 kΩ = 1V (rms)

If we assume ‘C’ value as 0.3 μF

Then I= 3mA,

V R = 3V

Which indicates voltage developed across ‘R’ is directly proportional to the capacitive impedance.

3. Phase Sensitive Detector:

• In case of ‘L’ measurement. it is a transistorized shifter, or Op-amp shifter circuit, is employed to solve the Inductor voltage into

(a) Quadrature

(b) In-phase voltages

i.e., Rs+ jω ‘Ls’ form, and fed to digital measuring circuit for display series equivalent circuit inductance Ls, which in turn dissipation factor ‘D’ and Q-factor (Q =1/D) values.

• In case of ‘C’ measurement, the phase sensitive detector gives the resistor voltages into quadrature and in-phase components, proportional to the capacitive current. The displayed capacitance measurement is that of parallel equivalent circuit (Cp), and also ‘D’ dissipation factor and Q-factor (Q=1/D).

For Measurement of Resistance using LCR Meter :

Taking Op-amp as current to voltage converter, known current is passing through constant current source and in feed back path unknown resistor ‘R’ is connected and observe the voltage drop across the unknown resistor, the voltage drop across the resistor is applied to A/D converter, which converts. The analog data values into digital values and standard values are maintained to get the final output as indication of the value of the unknown resistance ‘R’.

FIG 21 : Circuit Diagram of Current to Voltage Conversion

The digital RCL meter shown in Fig : 22 can measure inductance, capacitance, resistance, conductance, and dissipation factor. The desired function is selected by pushbutton. The range switch is normally set to the automatic (AUTO) position for convenience.However, when a number of similar measurements are to be made, it is faster to use the appropriate range instead of the automatic range selection. The numerical value of the measurement is indicated on the 3 1/2-digit display, and the multiplier and measured quantity are identified by LED indicating lamps.

2.9 SPECIFICATIONS OF DIGITAL LCR METER:

1. Display: 3-1/2 display(L/C/R-maximum 9999 display)

2. Parameters: D/Q- maximum 999 display

3. Tolerance mode: 1%, 5%, 10%

4. Measurement: 120Hz &41kHz

5. Measurement rate: 1 measurement per second, normal

6. Calibration: calibrates the meter internal parameters as well as external connector residues

7. Resistance: Range 10Ω to 10MΩ(±0.6%+5 dgts)

8. Inductance: Range 1mH to 1000H (±0.3%+5dgts)

9. Capacitance: Range 10nF to 10mf (±2.5%+5dgts)

10. Operation temp: 0 0c to 40 0c

11. Power: 6W max

12. Power requirement: AC 230/50 Hz

13. Dimensions: 211(L)x260(W)x71(H)

14. Weight: 1.64 Kg

Unit-1

1. List the advantages and disadvantages of PMMC meter (apr-10,nov-09)

2. Explain about rectifier type ammeter(apr-10)

3. Explain the FET-input electronic voltmeter(apr-10,nov-09,apr-09,nov-07)

4. Explain the capacitance measurement using the Schering bridge(apr-10)

5. Explain the series type ohm meter(apr-10,apr-09)

6. Explain high voltage and current probes(apr-10,nov-08)

7. Applications of wheat stone bridge(nov-09)

8. Explain loading effect of a voltmeter(nov-09,apr-09,nov-08,nov-07)

9. Explain how to extend the range of DC ammeter(nov-09,apr-09)

10. Explain the inductance measurement using Maxwell’s bridge(nov-09,nov-08)

11. Explain the working of basic differential voltmeter(apr-09,nov-08)

12. Explain the resistance measurement using wheat-stone bridge(apr-09,nov-07)

13. Principle of extending the range of DC voltmeter(nov-08)

14. Explain the working of shunt type ohmmeter (nov-08)

15. Explain the working principal of PMMC(nov-07)

UNIT-2

1. List the specifications of digital multimeter (apr-10,apr-09)

2. Explain the accuracy of a frequency meter(apr-10)

3. Explain the working of successive approximation type DVM(apr-10,nov-09)

4. List the specifications of digital frequency meter(apr-10,nov-09,apr-09,nov-07)

5. Explain the working of digital frequency meter(apr-10apr-09)

6. Mention the advantages of digital instruments over analog instruments(nov-09,nov-07)

7. List the specifications of digital LCR meter(nov-09,apr-09,nov-08)

8. List the specifications of DVM(nov-09,nov-08)

9. Explain the ramp type DVM(apr-09,nov-07)

10. Explain the working of digital multimeter(nov-08,nov-07)

11. Explain the dual slope integrating type DVM(nov-08)