Technical Publications Direction 2159823–100 Revision 5 GE Medical Systems 1.X / 2.X to CT/i Upgrade Manual Copyrighted 1998 by General Electric Company Operating Documentation

Upload: luis-fernando-garcia-s

Post on 09-Dec-2015

16 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 1.x,2.x to Cti Upgrade

TechnicalPublications

Direction 2159823–100Revision 5

GE Medical Systems1.X / 2.X to CT/i Upgrade Manual

Copyrighted 1998 by General Electric Company

Operating Documentation

Page 2: 1.x,2.x to Cti Upgrade
Page 3: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

i

THIS SERVICE MANUAL ISAVAILABLE IN ENGLISH ONLY.WARNING

��IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGEOTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITYTO PROVIDE TRANSLATION SERVICES.

��DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THISSERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.

��FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TOTHE SERVICE PROVIDER, OPERATOR OR PATIENT FROM ELECTRICSHOCK, MECHANICAL OR OTHER HAZARDS.

CE MANUEL DEMAINTENANCE N’ESTDISPONIBLE QU’EN ANGLAIS.

AVERTISSEMENT

��SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANSUNE AUTRE LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’ILINCOMBE DE LE FAIRE TRADUIRE.

��NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTSTANT QUE LE MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ETCOMPRIS.

��LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNERCHEZ LE TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DESBLESSURES DUES À DES DANGERS ÉLECTRIQUES, MÉCANIQUESOU AUTRES.

DIESES KUNDENDIENST–HANDBUCHEXISTIERT NUR IN ENGLISCHERSPRACHE.

WARNUNG

��FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHEBENÖTIGT, IST ES AUFGABE DES KUNDEN FÜR EINEENTSPRECHENDE ÜBERSETZUNG ZU SORGEN.

��VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVORDIESES KUNDENDIENST–HANDBUCH NICHT ZU RATE GEZOGENUND VERSTANDEN WURDE.

��WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZUVERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DESBEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHESCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.

ESTE MANUAL DE SERVICIOSÓLO EXISTE EN INGLÉS.AVISO

��SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMSSOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ESRESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DETRADUCCIÓN.

��NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SINHABER CONSULTADO Y COMPRENDIDO ESTE MANUAL DESERVICIO.

��LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DARLUGAR A QUE EL PROVEEDOR DE SERVICIOS, EL OPERADOR O ELPACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSASELÉCTRICAS, MECÁNICAS O DE OTRA NATURALEZA.

ESTE MANUAL DE ASSISTÊNCIATÉCNICA SÓ SE ENCONTRADISPONÍVEL EM INGLÊS.

ATENÇÃO

��SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA,QUE NÃO A GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA,É DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOSDE TRADUÇÃO.

��NÃO TENTE REPARAR O EQUIPAMENTO SEM TERCONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIATÉCNICA.

��O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO ASEGURANÇA DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘CHOQUES ELÉTRICOS, MECÂNICOS OU OUTROS.

IL PRESENTE MANUALE DIMANUTENZIONE È DISPONIBILESOLTANTO IN INGLESE.

AVVERTENZA

��SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMSRICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE ÈTENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE.

��SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURASOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED AVERNECOMPRESO IL CONTENUTO.

��NON TENERE CONTO DELLA PRESENTE AVVERTENZAPOTREBBE FAR COMPIERE OPERAZIONI DA CUI DERIVINOLESIONI ALL’ADDETTO ALLA MANUTENZIONE,ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONEELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

Page 4: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

ii

IMPORTANT! . . . X-RAY PROTECTIONX-ray equipment if not properly used may causeinjury. Accordingly, the instructions hereincontained should be thoroughly read andunderstood by everyone who will use the equipmentbefore you attempt to place this equipment inoperation. The General Electric Company, MedicalSystems Group, will be glad to assist and cooperatein placing this equipment in use.

Although this apparatus incorporates a high degree ofprotection against x-radiation other than the useful beam,no practical design of equipment can provide completeprotection. Nor can any practical design compel theoperator to take adequate precautions to prevent thepossibility of any persons carelessly exposing themselvesor others to radiation.

It is important that everyone having anything to do withx-radiation be properly trained and fully acquainted with

the recommendations of the National Council onRadiation Protection and Measurements as published inNCRP Reports available from NCRP Publications, 7910Woodmont Avenue, Room 1016, Bethesda, Maryland20814, and of the International Commission on RadiationProtection, and take adequate steps to protect againstinjury.

The equipment is sold with the understanding that theGeneral Electric Company, Medical Systems Group, itsagents, and representatives have no responsibility forinjury or damage which may result from improper use ofthe equipment.

Various protective material and devices are available. It isurged that such materials or devices be used.

CERTIFIED ELECTRICAL CONTRACTORSTATEMENT

All electrical installations that are preliminary to positioning ofthe equipment at the site prepared for the equipment shall beperformed by licensed electrical contractors. In addition,electrical feeds into the Power Distribution Unit shall beperformed by licensed electrical contractors. Otherconnections between pieces of electrical equipment,calibrations, and testing shall be performed by qualified GEMedical personnel. The products involved (and theaccompanying electrical installations) are highlysophisticated, and special engineering competence is

required. In performing all electrical work on these products,GE will use its own specially trained field engineers. All ofGE’s electrical work on these products will comply with therequirements of the applicable electrical codes.

The purchaser of GE equipment shall only utilize qualifiedpersonnel (i.e., GE’s field engineers, personnel of third-partyservice companies with equivalent training, or licensedelectricians) to perform electrical servicing on theequipment.

DAMAGE IN TRANSPORTATION

All packages should be closely examined at time ofdelivery. If damage is apparent, have notation “damage inshipment ” written on all copies of the freight or expressbill before delivery is accepted or “signed for” by aGeneral Electric representative or a hospital receivingagent. Whether noted or concealed, damage MUST bereported to the carrier immediately upon discovery, or inany event, within 14 days after receipt, and the contentsand containers held for inspection by the carrier. Atransportation company will not pay a claim for damage ifan inspection is not requested within this 14 day period.

Call Traffic and Transportation, Milwaukee, WI (414)827–3449 / 8*285–3449 immediately after damage isfound. At this time be ready to supply name of carrier,delivery date, consignee name, freight or express billnumber, item damaged and extent of damage.

Complete instructions regarding claim procedure arefound in Section “S” of the Policy & Procedure Bulletins.

Page 5: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

iii

OMISSIONS & ERRORS

Customers, please contact your GE Sales or Servicerepresenatives.

GE personnel, please use the GEMS CQA Process toreport all omissions, errors, and defects in thisdocumentation.

Page 6: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

iv

NOTES

Page 7: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–1

Chapter 1

CT/i Upgrade Manual

Terminology

Internet Address Upgrade–20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Boot Level Upgrade–72. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Operating System Level (Irix) Upgrade–72. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Applications Level Upgrade–72. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 8: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–2

Table Of Contents

Section Description Page

Section 1Work Flow Chart Upgrade–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Section 2Before you Start the Upgrade Upgrade–13. . . . . . . . . . . . . . . . . . . . . . . . . . .

2.1 Collect RP system data Upgrade–13. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2 Write Down System Internet Addresses Upgrade–13. . . . . . . . . . . . . . 2.3 Record ISO Values Upgrade–14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.4 Record Collimator Data Upgrade–14. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5 Record N Numbers Upgrade–15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.6 Write Down Tube Usage Infromation Upgrade–15. . . . . . . . . . . . . . . . . 2.7 Write Down Window / Level Hardkey Values Upgrade–16. . . . . . . . . . 2.8 Record Reconfig Information Upgrade–17. . . . . . . . . . . . . . . . . . . . . . . 2.9 IP Addresses in CT/i Upgrade–19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10 IP Addresses in RP System Upgrade–20. . . . . . . . . . . . . . . . . . . . . . . . 2.11 Check CT/i Modem is Available Upgrade–20. . . . . . . . . . . . . . . . . . . . .

Section 3Equipment Removal Upgrade–21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3.1 REMOVE SCU Upgrade–21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2 REMOVE GENESIS CONSOLE Upgrade–22. . . . . . . . . . . . . . . . . . . . 3.3 Remove CT/i Console from Skid Upgrade–23. . . . . . . . . . . . . . . . . . . . 3.4 Open Console and Attach Keyboard Shelf Upgrade–24. . . . . . . . . . . .

3.4.1 Open Console Upgrade–24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4.2 Keyboard Table Top Bracket Attachment and Console

Placement Upgrade–25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4.3 Keyboard Table Top Attachment Upgrade–26. . . . . . . . . . . . . . . . . . . . . 3.4.4 Seismic Mounting Upgrade–28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Section 4System Power and Ground Guidelines Upgrade–29. . . . . . . . . . . . . . . . . .

4.1 Electrical Inspection Upgrade–29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2 Identification Upgrade–30. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3 Cable Color Identifiers Upgrade–30. . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 9: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–3

Section Description Page

Section 5Wiring Upgrade–31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5.1 Introduction Upgrade–31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2 How to Use this Section Upgrade–31. . . . . . . . . . . . . . . . . . . . . . . . . . . .

INDIGO COMPUTER CONSOLE

5.3 Power Detail Upgrade–32. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.4 Incoming Power Detail w/Global or Compact PDU Upgrade–33. . . . . 5.5 Incoming Power Detail w/GE PDU Upgrade–34. . . . . . . . . . . . . . . . . . . 5.6 Computer Carrier Detail Upgrade–35. . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.7 Specialix Detail Upgrade–36. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.8 Back Bulkhead Detail Upgrade–37. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.9 Display Monitor Detail Upgrade–38. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.10 Rx Monitor Detail Upgrade–39. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OCTANE COMPUTER CONSOLE

5.11 Octane Rear Access Upgrade–40. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.12 Remove Shipping Strap and Bracket Upgrade–41. . . . . . . . . . . . . . . . . 5.13 Modem Installation 1 Upgrade–42. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.14 Modem Installation 2 Upgrade–43. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.15 DASM Installation 1 Upgrade–44. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.16 DASM Installtion 2 Upgrade–45. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.17 DASM Installation 3 Upgrade–46. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.18 DASM Isntallation 4 Upgrade–47. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.19 Power Detail Upgrade–48. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.20 Incoming Power Detail w/Global or Compact PDU Upgrade–49. . . . . 5.21 Incoming Power Detail w/GE PDU Upgrade–50. . . . . . . . . . . . . . . . . . . 5.22 Computer Carrier Detail Upgrade–51. . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.23 Back Bulkhead Detail 1 Upgrade–52. . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.24 Back Bulkhead Detail 2 Upgrade–53. . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.25 Display Monitor Detail Upgrade–54. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.26 Rx Monitor Detail Upgrade–55. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5.27 Keyboard / Trackball Detail Upgrade–56. . . . . . . . . . . . . . . . . . . . . . . . . 5.28 Gantry/Console Cable Connections Upgrade–57. . . . . . . . . . . . . . . . . .

5.28.1 E–Stop/Intercom/Scan Control Upgrade–57. . . . . . . . . . . . . . . . . . . . . . 5.28.3 Autovoice Upgrade–58. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.28.5 SCOM/FEP Fiber Upgrade–61. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5.29 System Ground Connections Upgrade–63. . . . . . . . . . . . . . . . . . . . . . .

Page 10: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–4

Section Description Page

Section 6COMPUTER INTEGRATION Upgrade–72. . . . . . . . . . . . . . . . . . . . . . . . . . . . .

DEFINITIONS Upgrade–726.1 Introduction Upgrade–72. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2 Tools Needed: Upgrade–72. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3 Requirements/Assumptions: Upgrade–73. . . . . . . . . . . . . . . . . . . . . . . . 6.4 Service Key Install Upgrade–73. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.5 Power Up RCU Upgrade–74. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.6 Keyboard Configuration Upgrade–74. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.7 Default Passwords Upgrade–75. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.9 Touch Calibration Upgrade–76. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.8 DASM Checkout Upgrade–75. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.7 Passwords Upgrade–75. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.11 ReConfig Upgrade–77. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.14 Install Options Upgrade–79. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.16 Verify Options Upgrade–81. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.18 KVTest and Z–Align Baselines Upgrade–83. . . . . . . . . . . . . . . . . . . . . . 6.19 Monitor Set–up Upgrade–84. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Section 7Table / Gantry Characterization Upgrade–88. . . . . . . . . . . . . . . . . . . . . . . . .

7.1 Collimator Characterization Upgrade–88. . . . . . . . . . . . . . . . . . . . . . . . . 7.2 Tilt Characterization Upgrade–91. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3 Elevation Characterization Upgrade–92. . . . . . . . . . . . . . . . . . . . . . . . . . 7.4 Cradle Characterization Upgrade–94. . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.5 Interference Test Upgrade–96. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Section 8MX200 Tube Upgrade Upgrade–98. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Section 9Tube Calibration Upgrade–99. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

9.1 New Tube Seed Shift Upgrade–99. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.2 Auto mA Cal Upgrade–101. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 11: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–5

Section Description Page

Section 10System Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Upgrade–102

10.1 Q–Cal Upgrade–103. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.2 Crosstalk Calibration Upgrade–105. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.3 Alpha Vector Cal Generation Upgrade–106. . . . . . . . . . . . . . . . . . . . . . . . 10.4 Hot ISO Upgrade–107. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.5 Air/Phantom Calibrations Upgrade–108. . . . . . . . . . . . . . . . . . . . . . . . . . .

Section 11Packing Material For Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Upgrade–110

11.1 Console Packing Upgrade–110. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.2 SCU Packing Upgrade–110. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.3 Label Return Material Upgrade–111. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

11.3.1 Return Information Upgrade–112. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.3.2 Return Information Upgrade–114. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 12: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–6

List Of Illustrations

Illustration Page

Illustration 1–1Upgrade Flowchart Upgrade–11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Illustration 1–2CT/i Network Devices Upgrade–19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Illustration 1–3RP Network Devices Upgrade–20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Illustration 1–4Console Stabilizer Pad Installation Upgrade–23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . Illustration 1–5Console Front Cover Latches Upgrade–24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Illustration 1–6Console Top and Keyboard Shelf Mounting Upgrade–25. . . . . . . . . . . . . . . . . . . . . . Illustration 1–8CT/i Console Power Wiring Detail Upgrade–32. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Illustration 1–11CT/i Console Computer Carrier Wiring Detail Upgrade–35. . . . . . . . . . . . . . . . . . . . Illustration 1–12CT/i Console Specialx Wiring Detail Upgrade–36. . . . . . . . . . . . . . . . . . . . . . . . . . . . Illustration 1–13CT/i Console Back Bulkhead Wiring Detail Upgrade–37. . . . . . . . . . . . . . . . . . . . . . Illustration 1–32CT/i Console Keyboard / Trackball Wiring Detail Upgrade–56. . . . . . . . . . . . . . . . . . Illustration 1–14CT/i Console Display Monitor Wiring Detail Upgrade–38. . . . . . . . . . . . . . . . . . . . . . Illustration 1–15CT/i Console Rx Monitor Wiring Detail Upgrade–39. . . . . . . . . . . . . . . . . . . . . . . . . . Illustration 1–33Gantry A1A5 Panel Location Upgrade–60. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Illustration 1–34CT/i Gantry Interconnect Cable Connections Upgrade–60. . . . . . . . . . . . . . . . . . . . . Illustration 1–36System Ground Wiring Upgrade–63. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 13: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–7

Section Description Page

Illustration 1–37Non EMC HiSpeed CT/i Interconnect Diagram Upgrade–66. . . . . . . . . . . . . . . . . . . Illustration 1–38EMC HiSpeed CT/i Interconnect Diagram Upgrade–67. . . . . . . . . . . . . . . . . . . . . . . Illustration A–1Elevation Distance Measurement Upgrade–93. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Illustration A–2Cradle Characterization Pin Locations Upgrade–95. . . . . . . . . . . . . . . . . . . . . . . . . .

Page 14: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–8

List Of Tables

Table Page

Table 1–1 System Interconnect Drawing Matrix Upgrade–31. . . . . . . . . . . . . . . . . . . . . Table 1–2 CT/i Ground Wire Sizes and Connections Upgrade–64. . . . . . . . . . . . . . . . .

Page 15: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–9

REVISION HISTORY

REV DATE REASON FOR CHANGE

A . . . . . . . . . . . . . . . . . . 24 July, 1996 . . . . . . . Initial Draft of document.

0 . . . . . . . . . . . . . . . . . .

1 . . . . . . . . . . . . . . . . . .

2 . . . . . . . . . . . . . . . . . .

18 Sept, 1996 . . . . . . .

18 Oct, 1996 .. . . . . . .

02 June, 1997 .. . . . .

Initial Release.

Pilot Feedback.

3.6 Software Release

3 . . . . . . . . . . . . . . . . . . 27 Oct 1997 . . . . . . . . 4.0 Software Release

4 . . . . . . . . . . . . . . . . . . 18 Mar 1998 . . . . . . . . Octane Clinical Release

5 . . . . . . . . . . . . . . . . . . 31 Dec 1998 . . . . . . . . Changes for 5.3 M4 Octane, 4.1 M4 Indigo

Page 16: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–10

NOTES

Page 17: 1.x,2.x to Cti Upgrade

SECTION 10

Does this upgradeinclude a MX200tube upgrade?

Get RP System Info:ISO, IP Addresses,Tube Slice, etc...

Route and terminatemonitor video &power cables inconsole base.Illustration 1–8 &Illustration 1–11

Disconnect &Remove GENESISConsole

SECTION 3.2

Lower console tofloor and removeshipping dollies

Remove consolefront cover and coveron front of scanchassis

Attach keyboard shelfto console sides

Install ConsoleStabilizer LegLeveling Pads beforelowering console tofloor.

Disconnect SCUcables and removeSCU

Remove CT/iConsole from skid

Unplug GENESISConsole Power Cordat PDU. TapeConsole end orremove cable.

Move SCU cables toCT/i Console location

Disconnect SCUground connection atTable JunctionGround Bus. Tapeboth ends or remove.

Move Consoleto/near final position

Attach keyboard shelfto shelf supports

Place keyboardassembly on thekeyboard shelf

Position bothmonitors on consoletop

Terminate cables atCT/i console

Power On

Computer Integration Table / GantryCharacterization

Calibrate System

Verify Images

SECTION 2.0

SECTION 3.1

SECTION 5

SECTION 6

SECTION 3.3

SECTION 3.3

Option – If performing a MX200tube upgrade do so at this point

SECTION 7

SECTION 8

No

Yes

Calibrate TubeSECTION 9

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–11

SECTION 1WORK FLOW CHART

Illustration 1–1UPGRADE FLOWCHART

Page 18: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–12

NOTES

Page 19: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–13

SECTION 2BEFORE YOU START THE UPGRADE

This section describes things that should be completed beforedisconnecting or removing any of the HSA RP 1x/2.x system pieces.

2.1 Save RP System State InformationThis section describes the procedure to share system specific files from the RP CTScanner to a CT/i CT scanner.

In this document output from the system is shown in this font/style: (ex. login:) ,,

site specific output is shown in this font/style: (ex. <suiteID>) ,

and commands to be typed by the user are shown in this font/style: (ex. cd/usr/g) .

2.2 Write Down System Internet Addresses(1) Login as root on the RP OC.

<suitId> _OC0 Login: root<enter>

Password: <Root Password>

<suite_id> _OC0 1: ypcat hosts|grep <suite_id>

This represents an example output of the primary IP addresses of interest.

192.20.20.2 <suiteID> _SBC0 <suiteID>_sbc0

192.20.20.1 <suiteID> _OC0 suti_oc0

The following IP address is associated with the optional second ethernet adapterboard that provides access to networks outside of the CT suite.

3.7.52.101 <suiteID> _OC0–gate suti_oc0–gate

Record number <suiteID> _SBC0: ________________________

Record number <suiteID> _OC0: ________________________

Record number <suiteID> _OC0–gate :________________________

Page 20: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–14

2.3 Record iso ValuesFrom a Unix Shell

Type: su – <enter> Log in as root.

Type: grep isoSmallSpot /usr/g/g_config/ScanHardware.cfg <enter>

Record number: ________________________

Type: grep isoLargeSpot /usr/g/g_config/ScanHardware.cfg <enter>

Record number: ________________________

2.4 Record Collimator Data

Collimator Data

Closed

1mm

3mm

5mm

7mm

10mm

CRC

Page 21: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–15

2.5 Record N numbers

Small Large

80 kV

100 kV

120 kV

140 kV

2.6 Write Down Tube Usage InformationThe RP system must be up and running for this procedure.

(1) Open a Unix Shell

<suite_id> _OC0 1: tubeUsage

Record current tube slice count:

Patient: ________________________

Non–Patient: ________________________

Transfer the current tube slice count into your site records. The slice count for CT/iwill start with zero after the software is reconfigured.

Page 22: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–16

2.7 Write Down Window / Level Hardkey Information

Window / Level Hard Key Values

Anatomy Window Level

Head

Mediastinum

Spine

Abdomen

Lung

Vertebrae

Page 23: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–17

2.8 Record Recofig Information(1) Shutdown Applications.

(2) Open a Unix Shell.

(3) As genesis on the RP OC.

<suitId> _OC0 Login : sd –i<enter>

(4) Wait for Applications to shutdown.

(5) Login as root on the RP OC.

<suitId> _OC0 Login: root<enter>

Password : <Root Password>

<suite_id> _OC0 2 : reconfig

(6) Record the following information from the reconfig screens.

Suite ID

Hostname

Hosptial Name

Service ID

Starting Patient Exam

Starting Optical Disk

Timezone

Doctor’s Title

Tube Housing PN

Tube Housing SN

Tube Insert PN

Tube Insert SN

Page 24: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–18

Camera Type

DAS Part Number

DAS Type

ACRNEMA Address

Suite ID of PAN Host

Hostname of PAN Host

Subnet Netmask

Gateway Netmask

(7) Go back to the first screen and exit without changes.

Page 25: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–19

2.9 IP Addresses in CT/iIllustration 1–2CT/I NETWORK DEVICES

ec0

du0

SGI IP22

vd0

Modem

ei0

SBC

DLAN To Gantry STC

Bit3GIO64

Bit3VME

vd0

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

VMEBusExternal

Network

gateway

InSite

OBC0

addressaddress

addressSBC Control LANaddress

du0 MODEM DEVICE MNEMONIC

ec0 IP22 ETHERNET TRANSCEIVER DEVICE MNEMONIC

vd0 BIT3 DEVICE MNEMONIC

ei0 SBC ETHERNET TRANSCEIVER DEVICE MNEMONIC

SBC0address

Page 26: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–20

2.10 IP Addresses in RP SystemIllustration 1–3RP NETWORK DEVICES

ei0 MOTOROLA CPU ETHERNET TRANSCEIVER DEVICE MNEMONIC

hk0 HAWK ETHERNET TRANSCEIVER DEVICE MNEMONIC

ei0 SBC ETHERNET TRANSCEIVER DEVICE MNEMONIC

hk0 CLA (HAWK) ETHERNET TRANSCEIVER DEVICE MNEMONIC

hk0

Motorola

Modem

SBC

DLAN To Gantry STC

ExternalNetwork

gateway

SBC0

address

address

SBC Control LANaddress

Hawk

ei0

CLA

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

VMEBus

OC0address

hk0ei0

Ethernet

2.11 Check CT/i Modem is AvailableInsure that a unique Internet Address is available for the CT/i InSite Modem.

HiSpeed CT/i has made a change to the way the OnLine Centers communicate withthe scanner. In the past the protocols used to communicate with the scanner wherebased upon standard/special serial protocols and special security access controls.

HiSpeed CT/i makes use of a Point to Point Protocol (PPP) to communicate with theOnLine Centers. The PPP allows standard TCP/IP connectivity tools to be used as ifthe modem connection where part of a TCP/IP based network. In addition, multiplelevels of access security are used to insure that un–authorized users cannot ‘get intothe system’.

For PPP to work correctly for the InSite connection, a unique IP address must beassigned to either the CT/i modem or to the SGI computer gateway.

Note 1 If the Host Computer (GENESIS) had a second ethernet board to connect as agateway to the facilities network and had unique IP addresses assigned by thefacilities Network Administrator, then that number may be used for the SGI HostComputer IP Gateway Address and may also be used for the Modem IP Address.

Note 2 If the Host Computer (GENESIS) does not have a Unique IP address assigned bythe facilities Network Administrator, then a unique number for the InSite Modem canbe obtained from the OnLine Center during the InSite Connection Check–Out.

Page 27: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–21

SECTION 3EQUIPMENT REMOVAL

This section tells you how to remove the Console and the SCU Cabinet.

3.1 REMOVE SCU(1) Disconnect these cables from the SCU and protect from damage. These cables

will be moved and re–connected to the HiSpeed CT/i Computer/Console:

> Disconnect SCU Power Cable (2118139 or 46–278722G1) at SCU.

Note 3 In case the SCU power cord was cut to length and re–terminated for the RP system,a new Power Cord 2118139 is included in the Upgrade kit of parts.

> Disconnect Scan Abort Cable (2118148 or 46–297565G1) at SCU.

> Disconnect Control LAN Cable (2118150 or 46–296454G2) at SCU

> Disconnect DAS Data Fiber (2117848–2) at FEP Board in SCU.

> Disconnect Network cable connection if present

> Disconnect Camera Video and Pixel CLock Cables

Note 4 The HiSpeed CT/i System requires a Camera Host Control Interface for filmingfunctions.

> Disconnect Camera Control Cable

(2) Disconnect remaining SCU cables and discard.

(3) Disconnect the SCU Ground wire at the Table/Gantry Junction Ground Bar. Tapeboth ends or remove ground wire from conduit or duct.

(4) Move SCU cabinet out of the work area.

Page 28: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–22

3.2 REMOVE GENESIS CONSOLE(1) Disconnect the cables from the GENESIS Operators Console.

DANGER FAILURE TO DISCONNECT THE GENESIS CONSOLE POWER CORD AT THEPDU COULD RESULT IN A SHOCK WHEN THE SYSTEM IS TURNED BACK ON.

(2) Disconnect the GENESIS Console power cable from the PDU.

(3) Tape up the Console end of the old power cord or remove the cord from theconduit or duct.

(4) Remove the trim covers over the Console Stabilizer Leg hardware on one side ofthe console and remove the hardware.

DANGER REMOVAL OF THE HARDWARE THAT ATTACHES THE GENESIS CONSOLE TOTHE FLOOR OR RISERS IN THE NEXT STEP MAKES THE GENESIS CONSOLELESS STABLE AND EASIER TO KNOCK OVER.

WE RECOMMENDED THAT YOU ATTACH THE CONSOLE SHIPPING DOLLIESTO THE CONSOLE BEFORE YOU REMOVE ANY OF THE CONSOLEANCHORING HARDWARE.

(5) Remove GENESIS Console Floor anchors if the console is attached to the flooror remove anchors that attach the console to the risers.

(6) Move the GENESIS Console to a location that will allow access to the hardwarethat attaches the stabilizer legs.

(7) Attach the GENESIS Console Shipping Dolly and Fisch–paper provided in theUpgrade Kit to the side of the console that the stabilizer leg was just removedfrom.

(8) Remove the trim covers over the Console Stabilizer Leg hardware on the otherside of the console and remove the hardware.

(9) Attach the GENESIS Console Shipping Dolly and Fisch–paper provided in theUpgrade Kit to the side of the console that the stabilizer leg was just removedfrom.

(10)Move the GENESIS Console out of the work area.

Page 29: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–23

3.3 Remove CT/i Console from Skid(1) Raise the CT/i Console off of the shipping skid using the adjustable dolly casters.

» Raise the console enough to ‘clear’ the wood blocking attached to the shippingskid.

» Roll the console away from the shipping skid.

(2) Move the console to very near it’s final position. Do not lower the console to thefloor yet!

Note 5 Even though the CT/i Console has internal casters, use the Shipping Dollies to movethe console any significant distance.

The larger diameter casters on the shipping dollies make it easier to move over doorand elevator thresholds and cause less shock to the console.

(3) Install the Console Stabilizer Leg Leveling Pads before lowering the Console tothe floor. See Illustration 1–4.

Illustration 1–4CONSOLE STABILIZER PAD INSTALLATION

» Screw the pads all the way into the Stabilizer Legs.

» After the console has been placed in it’s final position lower the pads so that theconsole does not rock forward.

(4) Lower the Console to the floor and remove the shipping dollies.

Page 30: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–24

3.4 Open and Assemble Console

3.4.1 Open Console(1) Front Cover Removal

> Unlatch the two fasteners at the top corners of the console front cover, rotatecounter–clockwise to the 5 o’clock position. See Illustration 1–5.

Illustration 1–5CONSOLE FRONT COVER LATCHES

Un–latched Latched

RotateCounter–Clockwise to Unlatch

(2) Once the front cover is unlatched slide the top edge forward 1/2 inch then lift thecover up and away from the console.

Page 31: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–25

3.4.2 Adjust Table Top Height and Console Placement

The Console should have arrived with the Monitor top mounted, starting with thefourth bolt from the bottom. The Keyboard Table Top should be attached startingwith the third hole from the bottom. This arrangement will provide for optimumoperator comfort.

Note 6 Your site may require the console to be adjusted from this position.

The important factor in setting the height is operator visibility of the patient beingscanned. If you need to adjust from this starting position keep the relationship ofmounting the Keyboard Table Top one hole lower then the Console Table Top, seeIllustration 1–6 for clarification.

Illustration 1–6CONSOLE TOP AND KEYBOARD SHELF MOUNTING

START WITH THE FOURTH HOLE FROM THE BOTTOM.THIS WILL LEAVE 2 1/2 HOLES SHOWING.SPEC 8 +/– 1 FOOT POUND

START WITH THE THIRD HOLE FROM THE BOTTOM.SPEC 8 +/– 1 FOOT POUND

ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ

2”

Wall

Minimum

Note 7 When placing the console in it’s final position keep a 2” minimum distance from thewall to maintain console airflow.

Page 32: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–26

3.4.3 Attach Keyboard Table Top – Select Illustration that matches your system.

Illustration 1–7ATTACH KEYBOARD TO TABLE TOP

Use spring loaded handles to attach keyboard top to support arms.

You may want to delay installing the top until the Octane computer is wired in Sec-tion 2 .

Page 33: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–27

NOTES

Page 34: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–28

3.4.4 Seismic Mounting

See Pre–Installation Manual 2152916–100.

Page 35: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–29

SECTION 4SYSTEM POWER AND GROUND GUIDELINES

This section provides reference information used to insure a goodpower installation for the system.

At this point, you MUST physically inspect the System Power and GroundConnections per the following:

» The system should have a dedicated ground. The ground wire should originate atthe supplying transformer and shall have no secondary ground connected to ituntil the ground bar in our PDU.

» The system ground wire should be copper and sized equal to or larger than the 3phase supply wires, with a minimum size of AWG 1/0. The ground wire shouldtake the same physical path (conduit) as the supply wires.

» The ground wire impedance from the system disconnect, including the groundrod, shall not have an impedance greater than 2 ohms to earth.

» The neutral from the 480 volt system should never be brought into the PDU.

» 3 phase wires, power cables, neutrals, and grounds should be routed in thesame conduit or raceway duct. These cables must not be routed in the sameconduit or raceway duct as system control cables.

CAUTION Check all the ground connections on the system for tightness and propertermination. Check ground connection at the supply transformer. Checkground from transformer to its earth connection.

Note 8 If you suspect that a system performance problem is related to, or caused by, apower/grounding issue, call:

GEMS – Americas GEMS – Europe GEMS – Asia

Power/Grounding Hotline

1–800–321–7937

Press 1 for OLC

Press 7 for Power and Ground

Tel: (33 1) 39 20 00 07 Taiwan: (886 2) 713 4150

Australia: (61 2) 318 9100

SMS: (82 2) 553 9737

YMS: (81 426) 56 0033

(ask for technical support)

4.1 Electrical InspectionArrange for electrical code inspection, if required.

Page 36: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–30

4.2 Identification» CT2 – Gantry» CT1 – Patient Table» PM – Power Distribution Unit

» OC – Operator’s Console» WL – X–Ray on Warning Light

All system cables are identified using the system component designators (seeabove). Each end of a system cable has a label and may have a color near theconnector (see below). This indicates the component and the jack identifier of thecomponent.

4.3 Cable Color IdentifiersThe ends of the cables may be marked with a color, either red, orange, blue, greenor brown, to help with the cable installation. The chart indicates which color isassociated with which sub component.

Sub–component Color

Gantry Blue

Table Yellow

PDU Red

Console Green

Page 37: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–31

SECTION 5WIRING

5.1 IntroductionThis upgrade manual supports 2 Console and 4 PDU types:

CONSOLE: INDIGO 2136400–2OCTANE 2169455

PDU: 46–296220G146–327106G12113764 EMC/NON–EMC2133533 EMC/NON–EMC

5.2 How to Use this SectionThis section contains detailed drawings of the individual cable connections and systeminterconnect drawings.

Use Sections 5.3 through 5.10 for a Console with an Indigo Computer.

Use Sections 5.19 through 5.26 for a Console with an Octane Computer.

To determine which system interconnect drawing to use consult Table 1–1, SYSTEMINTERCONNECT DRAWING MATRIX

Table 1–1SYSTEM INTERCONNECT DRAWING MATRIX

PDU Type Console Type

Indigo PN2138400–2 Octane PN2169455

46–327106G1 Illustration 1–39 page Upgrade–68 Illustration 1–42 page Upgrade–71

46–296220G1 Illustration 1–39 page Upgrade–68 Illustration 1–42 page Upgrade–71

2113764 EMC Illustration 1–38 page Upgrade–67 Illustration 1–41 page Upgrade–70

2133533 EMC Illustration 1–38 page Upgrade–67 Illustration 1–41 page Upgrade–70

2113764 NON EMC Illustration 1–37 page Upgrade–66 Illustration 1–40 page Upgrade–69

2133533 NON EMC Illustration 1–37 page Upgrade–66 Illustration 1–40 page Upgrade–69

Page 38: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–32

5.3 Power DetailÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

Console PN2138400–2

Illustration 1–8 CT/I CONSOLE POWER WIRING DETAIL

Front View

Notes:� Shown for reference, cable already

installed.� Option: Shown for reference, cable may or may not be installed on this

system.

P1 P3 P5 P7

P10 P12 P14 P16

P2 P4 P6 P8

P9 P11 P13 P15

P# DEVICE P# DEVICE

P1 RX MONITOR P9 VIEB �P2 EMC CHASSIS � P10 DASM �P3 KIEB � P11 IMAGE PRINTER �P4 MODEM � P12 VCR �P5 VIDEO SPLITTER � P13 REMOTE MONITOR �P6 PRINTER � P14 SPAREP7 DISPLAY MONITOR P15 SPAREP8 SPARE P16 COMPUTER CARRIER �

Page 39: 1.x,2.x to Cti Upgrade

Front View

Console Power 2118139From : PDU A4J5

P1 P3 P5 P7

P10 P12 P14 P16

P2 P4 P6 P8

P9 P11 P13 P15

12

Indicates Cable Run Number

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–33

5.4ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

Console PN2138400–2

Incoming Power Detail – For a System with Global or Compact PDUIf you have a PDU PN 46–329106G1 or 46–296220G1 see Section 5.5.

Illustration 1–9 CT/I CONSOLE INCOMING POWER WIRING DETAIL

Page 40: 1.x,2.x to Cti Upgrade

Front View

Console Power 2118139 or 46–278722G1 or 46–327187G1From : RP Console Wall Outlet

P1 P3 P5

P10 P12 P14

P2 P4 P6

P9 P11 P13

12

Indicates Cable Run Number

You will need this Section 5.5 if your RP Console waspowered from a wall outlet and not a power cord comingfrom the PDU. See Illustration 1–39 cable run 12 for moreinformation. In this situation two options exist for cable run12.

(1) You can re–use the power cord ( 46–278722G1 or46–327187G1 ) which connected the SCU to the wall powerfor cable run 12.

or

(2) If your site requires a more pliable power cord. You canre–work the power cord 2118139 that came with the kit. Toperform the re–work follow the steps below.

(a) Cut cable 2118139 to the desired length measuring fromthe female connector end.

(b) Remove the male plug from the power cord which wasused on the RP Console that was just removed.

(c) Install the male plug on cable 2118139.see detail (a) for plug wiring.

Brown – Z

Y

Black

X – blank

Blue

CenterYellow/Green

detail a

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–34

5.5ÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

Console PN2138400–2

Incoming Power Detail – For a System with a PDU PN 46–329106G1 or 46–296220G1If you have a Global or Compact PDU see Section 5.4

Illustration 1–10 CT/I CONSOLE INCOMING POWER WIRING DETAIL

Page 41: 1.x,2.x to Cti Upgrade

BIT3 1VCR Video 2Specialix 1

Rx Video2142221From : Rx Monitor

Notes:1 Shown for reference, cable already

installed.2 Option: Shown for reference, cable may or may not be installed on this

system.

Display Video2142221 – Color or2143804 – B/WFrom : Display Monitor

Front View

Audio In 1

Audio Out 1

SCSI 1 1

Inter Suite 1Line Printer 1

Inter Suite 1

Mouse:Route throughconsole top usinghole provided.

Keyboard 1

Camera SCSI 0 1

Insite Modem 1Security Key 1

110 VAC 1

1

Indicates Cable Run Number

3

5

J33

J35

J37

J39

J40

J41

J42

J43

J44

J45

J46

J47

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–35

5.6 Computer Carrier DetailÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

Console PN2138400–2

Illustration 1–11 CT/I CONSOLE COMPUTER CARRIER WIRING DETAIL

Page 42: 1.x,2.x to Cti Upgrade

Touch 2139581 �From : Rx Monitor

Front View

Notes:� Shown for reference, cable already

installed.� Option: Shown for reference, cable may or may not be installed on this

system.

� �

Indicates Cable Run Number

2

���� �� �� ����

Indicates Cable Run Number

����������������

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–36

5.7 Specialix DetailÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

Console PN2138400–2

Illustration 1–12 CT/I CONSOLE SPECIALIX WIRING DETAIL

Page 43: 1.x,2.x to Cti Upgrade

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

A10J8 – E–Stop

A10J9 – Intercom

A10J4 – Scan Control

Rear View

Keyboard 2139579From : Keyboard

DAS Data 2117848–2From : Gantry A1A5–J5

or SCOM

LAN 2118150 or 46–296454G2From : Gantry A1A5–J3

Scan Abort 2118148 From : Gantry A1A5–J4

or Scan Abort 46–297565G1From : PDU

Indicates Cable Run Number

4

8

9

10

11

J19

J20

J17

J18

Scan Control

Intercom

Existing cables disconnectedfrom RP Console

E–Stop2139578 *

*Alternate method for cable run 8 isto run a completely new cable PN2139577 from J20 to A1A5–J1

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–37

5.8 Back Bulkhead DetailÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

Console PN2138400–2

Illustration 1–13 CT/I CONSOLE BACK BULKHEAD WIRING DETAIL

Page 44: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–38

5.9 Display Monitor DetailÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

Console PN2138400–2

Illustration 1–14 CT/I CONSOLE DISPLAY MONITOR WIRING DETAIL

DISPLAY MONITOR COLOR OPTION

R G B

2142221 VideoFrom: Computer CarrierRight Hand Video Plug

Rear View

Rx MonitorDisplay Monitor

110VACFrom: P7

Indicates Cable Run Number

3

Page 45: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–39

5.10 Rx Monitor DetailÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

Console PN2138400–2

Note 9 The Rx monitor is color and may have a touch option.

Illustration 1–15 CT/I CONSOLE RX MONITOR WIRING DETAIL

Rear View

Rx MonitorDisplay Monitor

110VACFrom: P1

RX MONITOR

R G B

2142221 Video From: Computer CarrierLeft Hand Video Plug

2139581 Touch 1From: Specialix #2 Notes:

1 Option: Shown for reference, cable may or may not be installed on this

system.

Indicates Cable Run Number

2

1

Page 46: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–40

5.11 Octane Rear AccessÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

To install cables at the rear of the Octane in the following sub–sections it will benecessary to gain access to the rear of the Octane computer.

(1) Remove the screws at locations A and B.

(2) Slide the drawer forward.

(3) Remove the velcro strap and metal bracket, see Illustration 1–17, pageUpgrade–41.

Note 10 To prevent overheating of the Octane computer you must remove the velcro strapbefore turning on power.

Note 11 Once cable installation is complete return to step (4)

(4) Reverse the procedure to secure the Octane back into the console. Be aware oftight clearance points while sliding the tray back in.

» Location C, between the Octane and SCSI connector.

Illustration 1–16 CT/I CONSOLE

Front View – Console

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ ÉÉ

OCTANE

A B

C

Page 47: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–41

5.12 Remove Shipping Strap and BracketÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

Illustration 1–17 STRAP DETAIL

Many items removed for clarity

Remove strap before turning on power.The strap is blocking cooling air vents.Store strap for future use.

Remove and store bracket.

WARNING: Strap must be removed before power is turned on. The strap is blocking cooling air vents.

Page 48: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–42

5.13 Modem Installation (Hardware Option)ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

How to Install

(1) Some Consoles will come with the modem already installed.

(1) Refer to Illustration 1–18 and Illustration 1–19.

(2) Feed cable 2206138, modem power and phone line through the shelf provided for the modemfrom the back of the console to the front.

(3) Connect cable 2206138, power and phone line to the modem.

(4) Place the modem in the console on the shelf provided, see Illustration 1–18.

(5) Plug the modem power cord/pack into P12, see Illustration 1–15.

Illustration 1–18 MODEM DETAIL 1

MODEM

Modem Cables Shown; Serial Line In Power

Place Modem Here

Page 49: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–43

5.14 Modem Installation (Hardware Option)ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

(6) Connect the other end of cable 2206138 to serial port 1 of the Octane.

Illustration 1–19 MODEM DETAIL 2

2206183

OCTANE

MODEM

Connect to Serial Port 1

Many items removed for clarity

Page 50: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–44

5.15 Dasm Installation (Hardware Option)ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

How to Install

(1) Refer to Illustration 1–20.

(2) Remove screws (items C) and cover (item A) and set aside.

(3) Verify/Remove third jumper pin from ribbon cable (view B).

Note 12 Following this step will ensure the DASM is set to SCSI id 1.

(4) Replace cover and secure with screws.

Illustration 1–20 DASM DETAIL 1

BOTTOMVIEWDASM

Page 51: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–45

5.16 Dasm Installation (Hardware Option)ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

(5) Refer to Illustration 1–21, Analog DASM only.

(6) Attach cable 46–296980G1 to front of DASM.

Illustration 1–21 DASM DETAIL 2

Analog DASM

Many items removed for clarity

Page 52: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–46

5.17 Dasm Installation (Hardware Option)ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

(7) Refer to Illustration 1–22, disconnect the Central Data Box end of SCSI cable A.

Illustration 1–22 DASM DETAIL 3

Rear View of Octane and Central Data Box

SCSI Cable Configuration without DASM

Many items removed for clarity

OCTANECENTRAL DATA BOX

A

Page 53: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–47

5.18 Dasm Installation (Hardware Option)ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

(8) Refer to Illustration 1–23.

(9) Install SCSI 3 to 1 Adapter to the back of the DASM.

(10)Attach the free end of cable A to adapter installed in step (9).

(11)Attach cable B between the Central Data Box and DASM as shown.

(12)Plug DASM power cord into J10, see Section 5.10.

(13)Re–attach remaining cable(s) from camera to DASM.

Illustration 1–23 DASM DETAIL 4

Many items removed for clarity

Use SCSI 3 to1 Adapter forcable comingfrom Octane

OCTANE

CENTRAL DATA BOX

DASM

A

B

Place the DASM in-side the console nextto the Octane. Nomounting is provided.

Page 54: 1.x,2.x to Cti Upgrade

Notes:� Shown for reference, cable already installed.� Option: Shown for reference, system may or may not have this cable

J1 J3 J5 J7

J10 J12 J14 J16J2 J4 J6 J8

J9 J11 J13 J15

Front View

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–48

5.19 Console PowerÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

Illustration 1–24 CT/I CONSOLE RX WIRING DETAIL

J# Device J# Device

J1 Rx Monitor J9 Vieb �

J2 Powered Transceiver � J10 DASM �

J3 Kieb � J11 Image Printer �

J4 VCR � J12 Modem �

J5 Printer � J13 Remote Monitor �

J6 Video Splitter � J14 Spare

J7 Display Monitor J15 Central Data Box �

J8 Octane � J16 EMC Chassis �

Page 55: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–49

5.20ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

Incoming Power Detail – For a System with Global or Compact PDUIf you have a PDU PN 46–329106G1 or 46–296220G1 see Section 5.21

Illustration 1–25 CT/I CONSOLE INCOMING POWER WIRING DETAIL

Console Power 2118139From: PDU A4J5

J10 J12 J14 J16

J9 J11 J13 J15

12

Front View

Indicates Cable Run Number

Page 56: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–50

5.21ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

Incoming Power Detail – For a System with a PDU PN 46–329106G1 or 46–296220G1If you have a Global or Compact PDU see Section 5.20

Illustration 1–26 CT/I CONSOLE INCOMING POWER WIRING DETAIL

You will need this Section 5.21 if your RP Console waspowered from a wall outlet and not a power cord comingfrom the PDU. See Illustration 1–39 cable run 12 for moreinformation. In this situation two options exist for cable run12.

(1) You can re–use the power cord ( 46–278722G1 or46–327187G1 ) which connected the SCU to the wall powerfor cable run 12.

or

(2) If your site requires a more pliable power cord. You canre–work the power cord 2118139 that came with the kit. Toperform the re–work follow the steps below.

(a) Cut cable 2118139 to the desired length measuring fromthe female connector end.

(b) Remove the male plug from the power cord which wasused on the RP Console that was just removed.

(c) Install the male plug on cable 2118139.see detail (a) for plug wiring.

Brown – Z

Y

Black

X – blank

Blue

CenterYellow/Green

detail a

J10 J12 J14 J16

J9 J11 J13 J15

12

Front View

Console Power: 2118139 or 46–278722G1 or 46–327187G1From : RP Console Wall Outlet

Indicates Cable Run Number

Page 57: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–51

5.22 Computer CarrierÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

Illustration 1–27 COMPUTER CARRIER DETAIL

R L

HeadphoneCable toBulkhead �

Notes:� Shown for reference, cable already installed.

Indicates Cable Run Number

MicrophoneCables toBulkhead �

Mouse: Routethrough top usinghole provided

SCSI Cable toCentral DataBox or DASM �

ÀÀ

3D

Cable to modem

ÀÀ

3D

Rx Video2142221From: RxMonitor

PCI BIT 3 �

110 VAC �

2 1

Cable to NetworkTransceiver �

Keyboard � Keyboard Mouse

Headphone

IN

OUT

IN

OUT

Optical

Coaxial

Display Video2142221 – ColorFrom: DisplayMonitor

5

Security Cable �

1

3

Rear View of Octane

Front ViewNote: Cable runs 1 and 3 may already be

connected at the Octane end. Themonitor end of these cables can befound at the rear cable exit of theconsole. The monitor power cordswill also be plugged into the powerpanel and be attached to the videocables.

Page 58: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–52

5.23 Back Bulkhead 1ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

Illustration 1–28

Rear View

Keyboard 2139579From: Keyboard

DAS Data: 2117848–2From: Gantry A1A5–J5

LAN 2118150From: Gantry A1A5–J3

Scan Abort 2118148 From: Gantry A1A5–J4

4

9 10 11

Indicates Cable Run Number

Front View

Routed from frontof FEP board.

See Illustration 1–29 on page Upgrade–53 for cable run 8.

Page 59: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–53

5.24 Back Bulkhead 2ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

Illustration 1–29

Rear View

8

Indicates Cable Run Number

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

Existing cables disconnected from

RP Console

*Alternate method for cable run 8 isto run a completely new cable PN2139577 from J20 to A1A5–J1

Scan Control

Intercom

E–Stop

2139579 *

A10J4 – Scan Control

A10J9 – Intercom

A10J8 – E–Stop

Page 60: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–54

5.25 Display Monitor Color OptionÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

Illustration 1–30

DISPLAY MONITORCOLOR OPTION

R G B

2142221 VideoFrom: Computer Carrier Top Video Plug

Rear View

Rx MonitorDisplay Monitor

110VACFrom: P7

3

Indicates Cable Run Number

Page 61: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–55

5.26 Rx Display MonitorÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

Illustration 1–31

110VACFrom: P1

RX MONITOR

R G B

2142221 Video From: Computer CarrierBottom Video Plug

1

Indicates Cable Run Number

Rear View

Rx MonitorDisplay Monitor

Page 62: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–56

5.27 Keyboard / Trackball DetailConnect the Trackball and Keyboard cables as shown. Take note that the angledend of the keyboard cable connects to the keyboard not the straight end.

Do not connect the mouse to the keyboard. Route the mouse cable through theconsole top and connect it directly to the computer carrier. The mouse is cable run5.

Illustration 1–32 CT/I CONSOLE KEYBOARD / TRACKBALL WIRING DETAIL

Trackball

Keyboard Cable

Indicates Cable Run Number

4

6

Page 63: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–57

5.28 Gantry/Console Cable Connections

Select the console type that applies: INDIGO, Section 5.28.1 or OCTANE, Section 5.28.2.

5.28.1 E–Stop/Intercom/Scan ControlÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇ

Console PN2138400–2

(1) At the console; cables A10J4 (Scan Control), A10J8 (E–Stop) and A10J9(Intercom) have been disconnected from the RP Console. Use adapter cable2139578 at the console end of the cable to make the connection to CT/i ConsoleRear Bulkhead J20. See Illustration 1–13 and Illustration 1–39.

Note 13 Alternate Method for cable run 8.

If desired, the Console Gantry Signal Cable may be removed from A1A5 J1 andreplaced with Upgrade Kit cable 2139577 between Gantry A1A5 J1 and CT/i ConsoleRear Bulkhead J20.

(1) Connect new cable from upgrade kit 2139577; from Console J20 to Gantry A1A5J1. See Illustration 1–13.

5.28.2 E–Stop/Intercom/Scan ControlÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

ÉÉÉÉÉÉ

Console PN2169455

(1) At the console; cables A10J4 (Scan Control), A10J8 (E–Stop) and A10J9(Intercom) have been disconnected from the RP Console. Use adapter cable2139578 at the console end of the cable to make the connection to CT/i ConsoleRear Bulkhead J20. See Illustration 1–29 and Illustration 1–42.

Note 14 Alternate Method for cable run 8.

If desired, the Console Gantry Signal Cable may be removed from A1A5 J1 andreplaced with Upgrade Kit cable 2139577 between Gantry A1A5 J1 and CT/i ConsoleRear Bulkhead J20.

(1) Connect new cable from upgrade kit 2139577; from Console J20 to Gantry A1A5J1. See Illustration 1–29.

Page 64: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–58

5.28.3 Autovoice ÇÇÇÇÇÇ

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÇÇ

ÇÇÇÇ

Console PN2138400–2

Note 15 This cable is not needed with software revision 4.1 or greater. The LAN Cable nowcarries this function.

Page 65: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–59

5.28.4 Autovoice ÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉ

Console PN2169455

Note 16 This cable is not needed with software revision 5.3 or greater. The LAN Cable nowcarries this function.

Page 66: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–60

Illustration 1–33 GANTRY A1A5 PANEL LOCATION

Gantry Front View

A1 A5 panelDetail A

Illustration 1–34 CT/I GANTRY INTERCONNECT CABLE CONNECTIONS

A1A5 Panel

J4

J3

J6

2139577

2118150

Detail A

Used for VX only (not used on RP or IG)

2118146

J1

J2

2118148

J5

2117848–2

(J6 does not existon an EMC gantry)EMC gantry has 4–position TS–1.(Terminal Strip 1.)

**

J78

15

11

10

9

Indicates Cable Run Number

CAUTION ** Use extreme caution when installing the fiber optic DAS data cable. (PartNumber 2117848–2). It is fragile and should not be stepped on, kinked or bentsharply.

Page 67: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–61

5.28.5 SCOM/FEP FiberLook at the SCU and determine if your system is currently using a FEP Fiber link tothe gantry. If it is this Section 5.28.5 is complete, continue with Section 5.29.

If not continue with the following steps 5.28.5.1 and 5.28.5.2.

5.28.5.1 SCOM Board

The SCOM board is a static sensitive device, be sure to use ESD procedures. Use wrist strapprovided.

Remove the cover to the STC and determine if the SCOM board has a FEP Fiberconnection.

If it does leave the board installed and continue with step 5.28.5.2.

If it does not replace the board using the following steps.

(1) Disconnect the coax cables attached to the Axial board in the STC.

(2) Remove the SCOM Board

(3) Verify the dip switch on the SCOM board (S1) has switch 1 on and switches 2, 3and 4 are off.

(4) Install the SCOM board in the Gantry STC. The SCOM board is installed in theA5 slot (far right).

(5) Install the ground lead from the SCOM board to the STC chassis.

(6) Connect the coaxial cables attached to the Axial board in the STC.

5.28.5.2 FEP Fiber Cable(1) Route the fiber cable from the SCU to the gantry and ty–rap as shown in

Illustration 1–35. Verify that the fiber cable does not interfere with the gantryshrouds when the gantry is tilted.

(2) Connect the fiber cable to the fiber connector on the SCOM board in the STC.Connect the black ground wire (46–327241G1) from the SCOM board to theSTC power supply mounting plate as shown on Illustration 1–35 detail A.

(3) Install the new STC cover provided (2101947). Extra cover screws and washersare included, if needed.

Page 68: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–62

Illustration 1–35 FEP FIBER CABLE ROUTING

2117848–2 Fiber Optic Cable

Ty–Rap cable to existing harness andto the outside of the flexible trough

Detail A

SCOM

Attach SCOM ground wire tolower screw.46–327241G1

SCOM Lead Data(See Detail A).

Page 69: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–63

5.29 System Ground ConnectionsIllustration 1–36 SYSTEM GROUND WIRING

20

������ ����

������������

�����������

�� �������� ��

��

��

��

����# 2*

The 1 inch braid isattached to the tablewith two metalplates,46–297427P1 .Position the braidunder each plateand screw down theplate with thetwo bolts.

CT1 RACEWAYGROUND BUS

DETAIL

���

�����������

# 1/0

VAULTGROUND

* use ground wire supplied with system for these grounds

�� ����� ��

���������

# 2*

# 2*

#1/0*

ETC

1

SEE TABLE BELOWFOR CONNECTIONS

Page 70: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–64

Table 1–2CT/I GROUND WIRE SIZES AND CONNECTIONS

AWG CONNECTION P.M.

# INSTALLED

Connection To Connection To

#1/0** EARTH GROUND PDU

#2 OC CT1 RACEWAY #10

#2 CT2 CT1 RACEWAY #9

#2 CT1 CT1 RACEWAY #8

#1/0 PM CT1 RACEWAY #11

GND Braid (2) CT1 CT2

Page 71: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–65

NOTES

Page 72: 1.x,2.x to Cti Upgrade
Page 73: 1.x,2.x to Cti Upgrade

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

Console PN2138400–2

2139577 Intercom, Scan, E–Stop Adapter q

NEW CABLE FROMUPGRADE KIT

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

A1A5J3

A1A5J5

GANTRY �CT2(BLUE)

A2A7–J1

A2A7–J2

A2A7–J3

A2A7–J7

A1A5J1

A2A7–J8

TABLE

GANTRY

A2A7–J5

A1A5J2

A1A5J4

A1TS15, 6, 8

A1TS19, 10, 11

FIXED

A2 TS1–1TS2–1GND

A4 J4

A1 TS1–1TS1–2TS1–3

A4 J2

A4 J3

A4 J5

A1TS11, 2, 3

q

2117848–2 DAS Data � q

2118150 LAN q

2118148 Scan Abort q

q

DAS Data Fiber Con-nected to FEP

CONSOLE

2118139 Console Power q

NOTES:1 Non–EMC gantry has a 12–position TS1.2 Use extreme caution when installing the fiber optic

das data cable, part #2117848–2. It is fragile andshould not be stepped on, kinked or bent sharply.

3 A variety of Monitors are available as options, seepages 7, 11–12 for instructions.

4 Option.∨

TABLECT1(YELLOW)

PDUPM (RED)

q

MOUSE q

q Signal

q Tilt Motor Power

q LAN

q Tilt Limit Switch

q 2118141 Table Power

FIXED

FIXED

FIXED

FIXEDq Footswitch

A2A7–J6 FIXEDq Footswitch

Raceway Foot-switches

q HVDC Power q 46–297080G1 (HVDC)

q Axial Motor qPower 46–327155G1

q Signal q 2118146

q 120VDC Power q 2119603

��2139579 Keyboard q

PDU

KEYBOARD

Back Bulkhead

Back Bulkhead

Back Bulkhead

Power Panel

TrackBall �

q

Back Bulkhead J20

COMPUTERCARRIER

Indicates Cable Run Number

12

11

10

9

6

4

5

13

14

15

16

17

q

q

RXMONITOR

3

2

1 qSpecialx #2

��2142221 Video q

��2142221 Video q

��2139581 Touch q

q

8

HATCHED CABLES AREMOVED FROM THE SCUCABINET TO THE NEWCT/I CONSOLE

NON EMC ONLY

DISPLAYMONITOR

Use Existing cable with adapter 2139577 or runnew cable 2139577 Intercom, Scan, E–Stop q

A NEW 2118139CABLE ISINCLUDED IN THEUPGRADE KIT.

q

How to use this page: Check box when cable end is installed. q �

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–66

Illustration 1–37NON EMC HISPEED CT/I INTERCONNECT DIAGRAM

USE ON PDU 2133533, 2113764

Page 74: 1.x,2.x to Cti Upgrade

ÇÇÇÇÇÇ

ÇÇÇ

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÇÇÇÇÇÇ

Console PN2138400–2

2139577 Intercom, Scan, E–Stop Adapter q

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

GANTRY

A1A5J2

A1TS11, 2

A1TS13, 4

A2 TS1–1TS2–1

A4J4

A1 TS1–1TS1–2

A4J2

A4 J3

A4 J5

NOTES:� EMC gantry has a 4–position TS1.� Use extreme caution when installing the fiber optic

das data cable, part #2117848–2. It is fragile andshould not be stepped on, kinked or bent sharply.

� A variety of Monitors are available as options, seepages 7,11–12 for instructions.

� Option.

PDUPM (RED)

q HVDC Power q 2118143�(550)

q Axial Motor qPower 2118145

q Signal q 2118146

q 120VDC Power q 2118142

PDU

* Shields grounded withclamp at PDU cabinet andgantry base

Line FilterTop: BlueMiddle: Green/YellowBottom: Brown

A1A5J3

A1A5J5

GANTRY �CT2(BLUE)

A2A7–J1

A2A7–J2

A2A7–J3

A2A7–J7

A1A5J1

A2A7–J8

TABLE

A2A7–J5

A1A5J4

FIXED

q

How to use this page: Check box when cable end is installed. q

2117848–2 DAS Data � q

2118150 LAN q

2118148 Scan Abort q

DAS Data Fiber Con-nected to FEP

CONSOLE

TABLECT1(YELLOW)

q

MOUSE q

q Signal

q Tilt Motor Power

q LAN

q Tilt Limit Switch

q 2118141 Table Power

FIXED

FIXED

FIXED

FIXEDq Footswitch

A2A7–J6 FIXEDq Footswitch

Raceway

Footswitches

��2139579 Keyboard q

KEYBOARD

Back Bulkhead

Back Bulkhead

Back Bulkhead

Power Panel

TrackBall �

q

Back Bulkhead

Computer Carrier

Indicates Cable Run Number

12

11

10

9

6

4

5

q

DISPLAYMONITOR

q

RXMONITOR

3

2

1 qSpecialx #2

��2142221 Video q

��2142221 Video q

��2139581 Touch q

q

13

14

15

16

17

2118139 Console Power q

HATCHED CABLES AREMOVED FROM THE SCUCABINET TO THE NEWCT/I CONSOLE

8NEW CABLE FROMUPGRADE KIT

EMC ONLY

Use Existing cable with adapter 2139577 or runnew cable 2139577 Intercom, Scan, E–Stop q

A NEW 2118139CABLE ISINCLUDED IN THEUPGRADE KIT.

q

q

q

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–67

Illustration 1–38EMC HISPEED CT/I INTERCONNECT DIAGRAM

USE ON PDU 2133533, 2113764

Page 75: 1.x,2.x to Cti Upgrade

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ46–327187G1 or 46–278722G1 or 2139577

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

XCVR

SRU POWER

EXP INTERLOCK

DAS DATA Fiber Optic Cable

A3–K3, TS5

A3A3J19

A3A3J15CONSOLE EMERGENCY STOP

46–296986G1

SCAN CONTROL/AUTOVOICE

J546–297264G1

J8A1A5

A1A5

A1A5J6

46–297565G1

46–211848–2

46–297089G1

(PM)

PDU

GANTRYCONSOLE (CT2)

INTERCOM A10–J9

A10–J4

ETHERNET

A10–J8

2139578

Adapter Cable

Mouse

Keyboard

2142221

Video

2142221Video

2139581

Touch Option

RxMonitor

DisplayMonitor

SCOM inSTC

46–296454G5

LAN

CABLES MOVE FROM SRU TO CT/I CONSOLE

CONSOLEWALLOUTLET

CABLE MOVES FROMSRU TO CT/I CONSOLE

NEW CABLES FORUPGRADE

Back BulkheadJ17

DLAN

Back BulkheadJ20

Intercom Audio / E–Stop /Scan Control

Back BulkheadJ18

Exp I‘lock / Scan Abort

DAS Fiber OpticConnect to front of FEP

Power Panel

2139579Keyboard

J17

SGI Computer CarrierJ41

Back BulkheadJ9

SGI Computer CarrierSlot 5

SGI Computer CarrierSlot 2

SpecialX MTASP2

TrackballÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉ

HATCHEDCABLES AREMOVED FROMTHEIR ORIGINALLOCATIONS ATTHE SRU TO THECT/I CONSOLE

SHADEDCABLES ARENEW FROMUPGRADE KIT.

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

HATCHEDCABLES AREMOVED FROMTHEIR ORIGINALLOCATIONS ATTHE GENESISCONSOLE TOTHE CT/ICONSOLE

10

9

8

64

5

3

2

1

12

11

Indicates Cable Run Number

CABLESMOVE FROMGENESISCONSOLE TOCT/I CONSOLE

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

Console PN2138400–2

Indicates Cable Run Number

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–68

Illustration 1–39HISPEED CT/I INTERCONNECT DIAGRAM

USE ON PDU 46–296220G1. 46–327106G1

Page 76: 1.x,2.x to Cti Upgrade

ÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

Console PN2169455

2139577 Intercom, Scan, E–Stop Adapter q

NEW CABLE FROMUPGRADE KIT

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

A1A5J3

A1A5J5

GANTRY �CT2(BLUE)

A2A7–J1

A2A7–J2

A2A7–J3

A2A7–J7

A1A5J1

A2A7–J8

TABLE

GANTRY

A2A7–J5

A1A5J2

A1A5J4

A1TS15, 6, 8

A1TS19, 10, 11

FIXED

A2 TS1–1TS2–1GND

A4 J4

A1 TS1–1TS1–2TS1–3

A4 J2

A4 J3

A4 J5

A1TS11, 2, 3

q

2117848–2 DAS Data � q

2118150 LAN q

2118148 Scan Abort q

q

DAS Data Fiber Con-nected to FEP

CONSOLE

2118139 Console Power q

NOTES:1 Non–EMC gantry has a 12–position TS1.2 Use extreme caution when installing the fiber optic

das data cable, part #2117848–2. It is fragile andshould not be stepped on, kinked or bent sharply.

3 A variety of Monitors are available as options, seepages 7, 11–12 for instructions.

4 Option.∨

TABLECT1(YELLOW)

PDUPM (RED)

q

MOUSE q

q Signal

q Tilt Motor Power

q LAN

q Tilt Limit Switch

q 2118141 Table Power

FIXED

FIXED

FIXED

FIXEDq Footswitch

A2A7–J6 FIXEDq Footswitch

Raceway Foot-switches

q HVDC Power q 46–297080G1 (HVDC)

q Axial Motor qPower 46–327155G1

q Signal q 2118146

q 120VDC Power q 2119603

��2139579 Keyboard q

PDU

KEYBOARD

Back Bulkhead

Back Bulkhead

Back Bulkhead

Power Panel

TrackBall �

q

Back Bulkhead J20

COMPUTERCARRIER

Indicates Cable Run Number

12

11

10

9

6

4

5

13

14

15

16

17

q

q

RXMONITOR

3

1

��2142221 Video q

��2142221 Video q q

8

HATCHED CABLES AREMOVED FROM THE SCUCABINET TO THE NEWCT/I CONSOLE

NON EMC ONLY

DISPLAYMONITOR

Use Existing cable with adapter 2139577 or runnew cable 2139577 Intercom, Scan, E–Stop q

A NEW 2118139CABLE ISINCLUDED IN THEUPGRADE KIT.

q

How to use this page: Check box when cable end is installed. q �

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–69

Illustration 1–40NON EMC HISPEED CT/I INTERCONNECT DIAGRAM

USE ON PDU 2133533, 2113764

Page 77: 1.x,2.x to Cti Upgrade

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉÉÉÉ

Console PN2169455

2139577 Intercom, Scan, E–Stop Adapter q

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

GANTRY

A1A5J2

A1TS11, 2

A1TS13, 4

A2 TS1–1TS2–1

A4J4

A1 TS1–1TS1–2

A4J2

A4 J3

A4 J5

NOTES:� EMC gantry has a 4–position TS1.� Use extreme caution when installing the fiber optic

das data cable, part #2117848–2. It is fragile andshould not be stepped on, kinked or bent sharply.

� A variety of Monitors are available as options, seepages 7,11–12 for instructions.

� Option.

PDUPM (RED)

q HVDC Power q 2118143�(550)

q Axial Motor qPower 2118145

q Signal q 2118146

q 120VDC Power q 2118142

PDU

* Shields grounded withclamp at PDU cabinet andgantry base

Line FilterTop: BlueMiddle: Green/YellowBottom: Brown

A1A5J3

A1A5J5

GANTRY �CT2(BLUE)

A2A7–J1

A2A7–J2

A2A7–J3

A2A7–J7

A1A5J1

A2A7–J8

TABLE

A2A7–J5

A1A5J4

FIXED

q

How to use this page: Check box when cable end is installed. q

2117848–2 DAS Data � q

2118150 LAN q

2118148 Scan Abort q

DAS Data Fiber Con-nected to FEP

CONSOLE

TABLECT1(YELLOW)

q

MOUSE q

q Signal

q Tilt Motor Power

q LAN

q Tilt Limit Switch

q 2118141 Table Power

FIXED

FIXED

FIXED

FIXEDq Footswitch

A2A7–J6 FIXEDq Footswitch

Raceway

Footswitches

��2139579 Keyboard q

KEYBOARD

Back Bulkhead

Back Bulkhead

Back Bulkhead

Power Panel

TrackBall �

q

Back Bulkhead

Octane Computer

Indicates Cable Run Number

12

11

10

9

6

4

5

q

DISPLAYMONITOR

q

RXMONITOR

3

1

��2142221 Video q

��2142221 Video q q

13

14

15

16

17

2118139 Console Power q

HATCHED CABLES AREMOVED FROM THE SCUCABINET TO THE NEWCT/I CONSOLE

8NEW CABLES FROMUPGRADE KIT

EMC ONLY

Use Existing cable with adapter 2139577 or runnew cable 2139577 Intercom, Scan, E–Stop q

A NEW 2118139CABLE ISINCLUDED IN THEUPGRADE KIT.

q

q

q

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–70

Illustration 1–41EMC HISPEED CT/I INTERCONNECT DIAGRAM

USE ON PDU 2133533, 2113764

Page 78: 1.x,2.x to Cti Upgrade

ÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

Console PN2169455

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ46–327187G1 or 46–278722G1 or 2139577

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉSRU POWER

EXP INTERLOCK

DAS DATA Fiber Optic Cable

A3–K3, TS5

A3A3J19

A3A3J15CONSOLE EMERGENCY STOP

46–296986G1

SCAN CONTROL/AUTOVOICE

J546–297264G1

J8A1A5

A1A5

A1A5J6

46–297565G1

46–211848–2

46–297089G1

(PM)

PDU

GANTRYCONSOLE (CT2)

INTERCOM A10–J9

A10–J4

A10–J8

2139578

Adapter Cable

Mouse

Keyboard

2142221

Video

2142221Video

RxMonitor

DisplayMonitor

SCOM inSTC

46–296454G5

LAN

CABLES MOVE FROM SRU TO CT/I CONSOLE

CONSOLEWALLOUTLET

CABLE MOVES FROMSRU TO CT/I CONSOLE

NEW CABLES FORUPGRADE

Back BulkheadJ17

DLAN

Back BulkheadJ20

Intercom Audio / E–Stop /Scan Control

Back BulkheadJ18

Exp I‘lock / Scan Abort

DAS Fiber OpticConnect to front of FEP

Power Panel

2139579Keyboard

J17

SGI Computer CarrierJ41

Back BulkheadJ9

SGI Octane ComputerTop Video Head

SGI Octane ComputerBottom Video Head

TrackballÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉ

ÉÉÉÉÉÉÉ

HATCHEDCABLES AREMOVED FROMTHEIR ORIGINALLOCATIONS ATTHE SRU TO THECT/I CONSOLE

SHADEDCABLES ARENEW FROMUPGRADE KIT.

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

ÇÇÇÇÇÇ

HATCHEDCABLES AREMOVED FROMTHEIR ORIGINALLOCATIONS ATTHE GENESISCONSOLE TOTHE CT/ICONSOLE

10

9

8

64

5

3

1

12

11

Indicates Cable Run Number

CABLESMOVE FROMGENESISCONSOLE TOCT/I CONSOLE

Indicates Cable Run Number

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–71

Illustration 1–42HISPEED CT/I INTERCONNECT DIAGRAM

USE ON PDU 46–296220G1. 46–327106G1

Page 79: 1.x,2.x to Cti Upgrade
Page 80: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–72

SECTION 6COMPUTER INTEGRATION

DEFINITIONS

Boot Level

This is the software level where no software is running other than what can be run out ofCPU firmware. On GENESIS this was often referred to as ‘Prom Monitor’ or ‘BootPrompt’.

Operating System Level (Irix)

This is the software level in between ‘Boot Level’ and ‘Applications Level’. In GENESIS,this was often referred to as the Unix or Operating System level. The HiSpeed CT/isystem will normally start and login as user ‘ctuser’ leaving the User Interface ready forselection of Irix Commands or start–up of the Scanner Applications Software.

Applications Level

This is the software level where the Scanner specific software has been initialized andthe system could be used to: scan, archive, display, film, network, etc...

6.1 IntroductionThis section describes the reconfig, options installation and touch monitor calibration fora HiSpeed CT/i System which has had the operating system, platform and diagnosticsoftware previously loaded.

6.2 Tools Needed:» GE Service Key (J25)

» Option MOD (optional)

Page 81: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–73

6.3 Requirements/Assumptions:(1) All stations in a suite must have the same SUITE ID.

(2) The name of all hosts in the suite and their respective Internet/Ethernet numbers.

(3) The first time a ‘reconfig’ is executed on the system the following information isneeded: Internet (IP) addresses. If the system is to be connected to a networkcontact your system administrator to obtain the IP address for the computers in thesuite. If the system is stand–alone (not connected to any network) the defaultinternet numbers can be used.

CAUTION This procedure instructs the operator to REGEN DATABASEwhen executing a ‘reconfig’. This is required during the initialinstallation or Upgrade of a System because the Databases arenot compatible due to the load/duplication processes used byManufacturing to create these disks. During an initial Installationor Upgrade it is assumed that disks do not contain anyCalibration or Image Data. The ‘reconfig’ process, as describedin this procedure, will destroy any Image or Calibration Data thatexists on the System. If a ‘reconfig’ must be executed to changeSite Specific parameters it is strongly recommended that all Info,Image and Calibration Data be saved on Tape to avoid possibleloss.

6.4 Install Service Key(1) Install the service key into J25, located on the front of the console.

Illustration 1–43SERVICE KEY LOCATION

Service Key(top connector)

Page 82: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–74

6.5 Power Up RCU(1) Power up the RCU.

(2) When the window appears ‘Starting up the system . . .’

Select Stop for Maintenance

6.6 Keyboard ConfigurationThe CT/i System can be used with 1 of 4 keyboards;

› o English – US (default)

› o German – DE

› o Swedish – SE

› o French – FR

The computer leaves the factory configured for the English – US keyboard. If your systemis using that keyboard then continue with section 6.6.2.

6.6.1 Non–English Keyboard Configuration

A screen will pop up with several icons:

› Start System

› Install System Software

› Run Diagnostics

› Recover System

› Enter Command Monitor

› Select Keyboard Layout

(1) Use the mouse and click on Select Keyboard Layout

(2) Select the appropriate keyboard type. The currently supported keyboards are:

› o English – US (default)

› o German – DE

› o Swedish – SE

› o French – FR

(3) Select ”Apply” to reconfigure.

Page 83: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–75

6.6.2 Start System(1) Select ”Start System”.

(2) Stop the CT/i Auto–Start by selecting Cancel when the pink attention

window appears.

6.7 PasswordsAll forward production HISPEED CT/i systems have the following default passwords:

root: #bigguy

ctuser: 4$apps

insite: 2getin

service: 4rhelp

diag: <enter>

These defaults were changed because of customer and field requests for tighter security,especially those operating on networks. We suggest that the default passwords be changedONLY by customer request. Contact your support center with the new passwords.

6.8 DASM Checkout (Hardware Option)

Note 17 If a DASM was installed in Section 5, perform the following steps.

(1) Open a UNIX shell as ctuser.

(2) Type/enter scsistat

Note 18 Observe the following output. Some of the output has been removed to make it generic.The important parts of the output are; Device 1 1 Disk, this shows that the DASM wassuccessfully installed as scsi device 1. You should see 3 devices after the DASM, theseare your central data, MOD and CD–ROM. If these 3 devices are listed it means that theinstallation of the DASM did not disrupt the SCSI 1 bus.

Device 0 1 DiskDevice 1 1 Disk ............. DASMDevice 1 3 OpticalDevice 1 4 CommDevice 1 6 CD–ROM

Page 84: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–76

6.9 Touch Calibration (Hardware Option)If your system has the touch option it will be necessary to calibrate this feature. Follow theinstructions below for touch calibration.

(1) In a ‘Unix Shell’ window , become root: su – <Enter> (then enter the rootpassword)

(2) Type : cd /usr/g/xtouch <Enter>

(3) Type : setup <Enter>

Note 19 The ‘Serial Device Name’ field should contain: /dev/ttya2

(4) Click on software calibrate

(5) Touch on lower left corner

(6) Touch on upper right corner

(7) Click on save settings

(8) Click on exit

Note 20 These settings will not take effect until the next time ctuser logs in.

(9) In the ‘Unix Shell’ type: exit <Enter> to log out as root.

6.10 Restore iso Spot ValuesRestore the HSA RP iso Large and Small Spot values to the CT/i System that was savedin section 2.3 on page Upgrade–14 of this direction.

(1) Open a UNIX Shell on the OC.

(2) Type: jot /usr/g/config/ScanHardware.cfg

(a) change isoSmallSpot line to the number recorded in Section 2.3.

(b) To find the value move the mouse to the jot top menu bar to Select pulldown a menu and select Search Forward . Enter isoSmallSpot into thesearch filed box and push enter. The isoSmallSpot filed will now behighlighted. Use the backspace key to remove the default value.

(c) change isoLargeSpot line to the number recorded in Section 2.3.

(d) Use the left mouse button select File Save at the top of the window.

(e) Use the left mouse button select File Exit at the top of the window.

(3) Type: cd /usr/g/config

Copy the ScanHardware.cfg file to the SBC.

(4) Type: rcp ScanHardware.cfg<suiteID>_sbc0:/usr/g/g_config/scanhardware.cfg

Page 85: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–77

6.11 Reconfig

6.11.1 Rconfig the OC

Note 21 When you reconfigure the OC, you MUST reconfig the SBC. The document collectorbox that arrived with your system, contains an OC/SBC System Reconfigurationprocedure, PN 2168070–100 for your reference.

Applications must not be up.

(1) Type su – <enter> to change to root

(2) Type cd /usr/g/scripts

(3) Type reconfig .

(4) Select config <enter> to open the reconfig UIF.

(5) Select Regenerate Database? |Yes| on the first configuration screen, to erase theformer calibration information.

(6) Complete the remaining configuration screen data field selections. Refer to theOC/SBC System Reconfiguration procedure, PN 2168070–100 (located in thedocument collector box that arrived with your system), for illustrations of theConfiguration screens, and explanations of the data fields.

(7) When you finish all the configuration screen data parameters, select |Accept| at thetop of the screen.

(8) Select |Yes| when the screen prompts you to re–boot now.

6.11.2 Rconfig the SBC

Note 22 The OC reconfig program cannot change the timezone on the SBC. Before runningreconfig on the SBC perform the timezone command below.

(1) Open a UNIX shell.

(2) Type/enter cu sbc

(3) Login as root

(4) Type/enter reconfig

Note 23 Wait for the SBC reconfig to complete. The software automatically reboots the sbc whenit completes the SBC reconfig.

(5) Exit the cu session type ~ .

i.e.: to exit the cu session press the following keys: <tilde><dot><Enter>

Page 86: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–78

6.12 Set Date and Time(1) Open a UNIX Shell on the OC.

(2) Type/enter su root to change to root.

(3) Type/enter setdate to set the date and time.

(4) Type exit <enter>

6.13 Applications Start–Up(1) In the Console window type: st <Enter>

Page 87: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–79

6.14 Install Option(s)Note 24 This system may have a MOD which contains options that the customer has

purchased. If this system has such an MOD install it at this time.

(1) Select |Service|

Service

(2) Select |SYSTEM INTEGRATION|

SYSTEMINTEGRATION

(3) Select |COMPUTER INTEGRATION|

COMPUTER INTEGRATION

TABLE / GANTRY INTEGRATION

GENERATOR CALIBRATION

DAS INTEGRATION

ALIGNMENTS

SYSTEM TESTS

(4) Select |START|

. . . START. . .

Page 88: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–80

(5) Select |Install Options|

Computer Integration

Dismiss

RP Diags

Restore System State

Install Options

_ _ _

Computer Diags

Display DiagsOperator I/O Diags

Verify Options

New Tube Config

(6) Insert the options MOD into the top MOD drive.

(7) Select the options in the left–hand column to install the corresponding software.

(8) Select |install|

6.15 Restore Window / Level Hardkey ValuesRestore the HSA RP System Information saved in section 2.1 on page Upgrade–13 of thisdirection.

The CT/i system must be up and running the CT applications. Connect and startup the CT/isystem as normal before starting this procedure.

Restore the Window / Level values in Image Works.

(1) Select the Image Works desktop.

(2) Select Viewer.

(3) Select User Prefs on the Viewer.

(4) Fill out the Window /Level table for the CT values from the table you saved in Section 2.7 .

(5) Click on Save Defaults.

Page 89: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–81

6.16 New Tube Config(1) Select |New Tube Config|

Computer Integration

Dismiss

Restore System State

Install Options

Verify Options

– – –

Computer Diags

Display Diags

Operator I/O Diags

RP Diags

New Tube Config

(2) Answer the questions.

Note 25 You must perform a system shutdown and startup for the tube usage count to be correct.Contiune with steps (3) thru (5) to perform this.

(3) Select |Utilities|

UTILITIES

Page 90: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–82

(4) Select |Applications Shutdown|

.

.

.

.

.

.

.

.

.

.

Applications Shutdown

(5) After Applications have been shutdown, in the Console shell type: st <Enter>

6.17 Verify OptionsIf options were installed perform this section.

(1) Select |Verify Options|

Computer Integration

Dismiss

Restore System State

Install Options

Verify Options

– – –

Computer Diags

Display Diags

Operator I/O Diags

RP Diags

New Tube Config

(2) A shelltool will open: Verify that the options loaded in the previous section are

listed.

Page 91: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–83

6.18 KVTest and Z–Align BaselinesKVTest baseline and Z–Align baseline are not saved from the RP System. KVTest baselineand Z–Align baseline will need to be re–established prior to the next tube change.

Page 92: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–84

6.19 Monitor Set–upThis chapter describes the steps required to set up the two Display Monitors. The flowchartin Illustration 1–44 provides an overview of the Monitor Setup procedure. Use the detailedprocedures in the following sections to set up the monitors.

Illustration 1–44SETUP FLOWCHART

Start

Factory

End

Pre–Set

SetColor

Adjust Brightnessand Contrast

Adjust H/V Size,Center and Tilt

Adjust ConvergencePincushion

Select the 6500K color temperature

Page 93: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–85

6.19.1 Monitor Manual Use

The following steps are the proper order to follow for adjustment. You will need to reference themonitor manual that arrived in the monitor box for specific control procedure.

6.19.2 Factory Pre–Set(1) Reference the monitor manual and reset all adjustments to factory preset levels.

6.19.3 Center the Raster(1) Reference the monitor manual and center the raster.

› The monitors normally have a blank area to the left and right of the display,to maintain the correct aspect ratio.

6.19.4 Set Monitor Color Temperature(1) Reference the monitor manual and select the 6500K color temperature.

6.19.5 Monitor Brightness and Contrast Set–up

Color monitors have a lower light output than black and white monitors. Take care when you adjustthe monitor brightness & contrast on CT/i systems.

If the operators tell you they think the image on the monitor looks “softer” than the image on the film,and they like the film, adjust the monitor Brightness and Contrast to compensate for its lower lightoutput. Use the following procedure to adjust the monitor Brightness and Contrast, until the operatorsees the anatomical structure (window width) at the expected window levels.

(1) Adjust the ambient room light to the level used by the operator during imagereview.

(2) Display the Image Works desktop.

(3) Display the SMPTE pattern:

› Use the browser to select Exam 1000, which contains the SMPTE pattern,and enlarge the image to full screen display.

–or–

(a) Select the Service Desktop

(b) Select |Trouble Shooting|

(c) Select |Display Processing|

(d) Select |Install SMPTE|

(4) Increase the Contrast to maximum.

(5) Increase the Brightness to maximum.

(6) Decrease the Brightness, until the raster just fades into, and matches, themonitor screen background.

Page 94: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–86

6.19.6 Adjust the Raster Horizontal and Vertical Size and Tilt.(1) Reference the monitor manual and adjust the horizontal, vertical raster size and

tilt.

6.19.7 Adjust Pincushion(1) Reference the monitor manual and adjust the pincushion.

6.19.8 Adjust Convergence(1) Display the Service Desktop Manager.

(2) Select |Operator I/O|

(3) Select |Host Diags|

(4) Select the |Monitor| icon.

(5) Select |Convergence| from the monitor menu, to display the convergencepattern on the exam Rx monitor.

Note 26 Swap video cables to adjust convergence on the display monitor.

(6) Reference the monitor manual and adjust the convergence.

(7) If you swapped cables to display the convergence pattern on the Display monitor,restore the cables to the original configuration.

Page 95: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–87

NOTES

Page 96: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–88

SECTION 7TABLE GANTRY CHARACTERIZATION

7.1 Collimator Characterization(1) Select SERVICE

Service

(2) Select SYSTEM INTEGRATION

SYSTEMINTEGRATION

(3) Select TABLE / GANTRY INTEGRATION

COMPUTER INTEGRATION

TABLE / GANTRY INTEGRATION

GENERATOR CALIBRATION

DAS INTEGRATION

ALIGNMENTS

SYSTEM TESTS

Page 97: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–89

(4) Select START

. . . START. . .

(5) Select Characterization

Table/Gantry Installation

HV DC Enable Switch

Characterization

– – – –

Power Supply Checks

Collimator

Rotor Verification

Dismiss

Page 98: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–90

(6) Select Collimator Data

Mechanical Characterization Tool

File Help

Cradle

HardwareReset

TableElevation

Tilt

CollimatorData

CharactLimits

ViewValues

ViewLog

Characterize

Data View

(7) Perform

enter the characterization values that were recorded in SECTION 2.4.

Note 27 Use the <tab> key to move to the next data input field for collimator characterizationvalues.

Select . . . . . Continue

Note 28 Note message which says that characterization was completed successfully (if youentered new values).

Page 99: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–91

7.2 Tilt Characterization(1) Select Tilt

CollimatorData

Mechanical Characterization Tool

File Help

Cradle

HardwareReset

TableElevation

Tilt

CharactLimits

ViewValues

ViewLog

Characterize

Data View

(2) Perform Tilt Characterization per the instructions given on the screen.

Page 100: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–92

7.3 Elevation Characterization(1) Remove upper right and lower right table covers.

Small L–shaped measurement block is stored on top of the bigger measurement block. Removesmall L–shaped measurement block from its storage position and install it in its characterizationposition at the right rear of the table. Move measurement plate on right side of table to ‘out’position. Tighten mounting screws for both plates.

(2) Select Table Elevation

CollimatorData

Mechanical Characterization Tool

File Help

Cradle

HardwareReset

TableElevation

Tilt

CharactLimits

ViewValues

ViewLog

Characterize

Data View

(3) Perform Elevation Characterization per the instructions given on the screen.

Note 29 When measuring elevation distance, measure as vertically as possible. Measurebetween the bottom surface of the appropriate measurement block and the center ofthe rear leg upper pivot pin, per Illustration 7–2 below.

This measurement is critical. Elevation characterization will fail if this measurement isincorrect.

Re–install measurement plates to original positions.

Page 101: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–93

Illustration A–1ELEVATION DISTANCE MEASUREMENT

Weldment Plate

MeasureThisDistance

Top of WeldmentPlate

Measurement from bot-tom of rear measure-ment block for lowerpositionmeasurement

Measurement from bot-tom of right side mea-surement blockfor upper position mea-surement

MeasureThisDistance

Rear leg upper pivot pin

Rear leg upper pivot pin

UPPER POSITION DISTANCE = 36.44 INCHES(925.6MM)

LOWER POSITION DISTANCE = 14.7 INCHES(373.4MM)

Weldment Plate

Gantry

ÀÀÀÀÀÀÀÀÀ

ÀÀÀÀÀÀÀÀÀÀÀÀ

etc board

ÀÀÀÀÀÀÀÀÀ

ÀÀÀÀÀÀÀÀÀÀÀÀ

etc board

Gantry

Top of WeldmentPlate

CT38885A

CT38886A

Page 102: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–94

7.4 Cradle Characterization(4) Select Cradle

TableElevation

CollimatorData

Mechanical Characterization Tool

File Help

Cradle

HardwareReset

Tilt

CharactLimits

ViewValues

ViewLog

Characterize

Data View

(5) Perform Cradle CharacterizatIon per the instructions given on the screen.

IMPORTANT CRADLE CHARACTERIZATION INFORMATION

Note 30 The following changes apply to the on–screen directions given for thealignment light and end–of–cradle position characterization steps, for tablesequipped with the METAL–FREE cradle only.

» – In the alignment light procedure (3rd and 4th screens), use the BLACKDOT on the cradle right–side bumper, instead of the black stripe, for bothsteps. Note that the table height must be at least 40mm from isocenter inorder for the black dot to reach the external scan plane alignment lightbefore the cradle reaches its home position.

» – In the end–of–cradle position procedure (5th screen), use the sameBLACK DOT on the cradle right–side bumper, instead of the end of thecradle.

Page 103: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–95

Illustration A–2CRADLE CHARACTERIZATION PIN LOCATIONS

ÀÀÀÀÀÀÀÀÀÀÀÀ

etc board

CT38886A

Cradle Characterization PinStorage Location(Under Front Cover)

Cradle Characterization PinFront Characterization Location

Gantry

Cradle Characterization Pin Rear Characterization Location

Page 104: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–96

7.5 Interference TestAt this point you must do a hardware reset to download the new characterization valuesto the ETC, OBC, and STC before doing the table/gantry interference test on this page.

(1) Select Hardware Reset

Cradle

TableElevation

CollimatorData

Mechanical Characterization Tool

File Help

HardwareReset

Tilt

CharactLimits

ViewValues

ViewLog

Characterize

Data View

Note 31 If the following elevation and/or tilt display readings are incorrect, DO NOT adjustlimit switches. Re–characterize elevation and/or tilt.

Note 32 When discussing gantry tilt on this page, – (minus sign) refers to having the top ofthe gantry tilted away from the table. + (plus sign) refers to having the top of thegantry tilted toward the table.

Note 33 Pressing the ‘fast’ pushbutton only will cause the elevation limits for the presentelevation/tilt combination to be displayed.

(2) Move table to the first position, down and out:

(a) Move the cradle to the home position.

(b) Using the Down pushbutton, lower the table to the minimum height.

(c) Verify that the elevation display reads 577.5 +/– 3 mm.

(3) Move Elevation to maximum:

(a) Using the Up pushbutton, raise the table to the maximum height.

Page 105: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–97

(b) Verify that the elevation display reads 25.5 +/– 3 mm.

(4) Check Elevation minimum clearance with zero tilt:

(a) Advance the cradle from the home position about 36 inches.

(b) Tilt the gantry to –23 degrees.

(c) Use the Down pushbutton to lower the table until it stops automatically.

(d) Verify that the elevation display reads 115.5 +/– 3 mm.

(e) Verify that minimum clearance between table and gantry covers is no lessthan 1/2 inch.

(5) Check Elevation clearance with gantry tilted minus 20:

(a) Tilt the gantry to –20 degrees.

(b) Use the Down pushbutton to lower the table until it stops automatically.

(c) Verify that the elevation display reads 209.5 +/– 3 mm.

(d) Verify that minimum clearance between table and gantry covers is no lessthan 1/2 inch.

(6) Check Tilt Interference with table raised:

(a) Using the Up pushbutton, raise the table until the elevation display readsbetween 119.5mm and 209.5mm.

(b) Tilt the gantry to –10 degrees.

(c) Tilt the gantry back and verify that it stops at –22.0 +/– 0 degrees.

(d) Tilt the gantry forward and verify that it stops at +18.5 +/– 0 degrees.

(7) Verify +/– 30 degree tilt:

(a) Return the cradle to home position.

(b) Verify that the gantry will now tilt +/– 30.0 degrees.

(c) Leave gantry tilted at –30 degrees.

(8) Check Cradle movement below 209mm:

(a) Raise the table enough to allow cradle movement.

(b) Advance the cradle from the home position about 12 inches.

(c) Using the Home Button , return the cradle to the home position, tilt thegantry to 0 degrees and lower the table to minimum height.

(d) Verify that no cradle movement is possible below 209.5 +/– 3mm.

(e) Verify that the cradle Lock Indicator on the gantry display is lit.

(9) Select File, Exit

(10)Select Dismiss

(11)Select Cleanup

Page 106: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–98

SECTION 8MX200 TUBE UPGRADE

If your system will be upgrading to the MX200 tube do so at this point in this upgradeprocedure. The MX200 upgrade procedure will complete any remaining steps for the CT/iUpgrade. You will not be returning to complete any remaining sections of this manual.

Page 107: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–99

SECTION 9TUBE CALIBRATION

9.1 New Tube Seed ShiftUse the following sequence to run a partial Install New Tube, and set the calseed values on a newsystem.

(1) Select |Generator Calibration|

COMPUTER INTEGRATION

TABLE / GANTRY INTEGRATION

GENERATOR CALIBRATION

DAS INTEGRATION

ALIGNMENTS

SYSTEM TESTS

(2) Select |Start|

START

Page 108: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–100

(3) Select |New Tube Seed Shift|

Generator Installation and Verification

Dismiss

mA Meter Check

kV Test

Auto mA Cal

HHS Scans

HHS Report

Ground Leakage Test

Heat Soak & Seasoning

– – – –

New Tube Seed Shift

Bleeder Setup (DDC)

Rise and Fall Time (DDC)

Internal Scan Time (DDC)

Heat Soak & Seasoning (Diag)

(4) The screen prompts you to confirm the displayed tube type.

(5) Verify the installed and configured tube types match.

(6) Press (Start Scan) when it illuminates, to automatically run the filament current seedshift program.

(7) ABORT the program after the filament current seed shift scans, and before ductilitywarm–up

Page 109: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–101

9.2 Auto mA CalIf you have just run a Heat Soak and Seasoning, wait 30 minutes before you run the Auto mACal.

(1) Select |Auto mA Cal|

Generator Installation and Verification

Dismiss

mA Meter Check

kV Test

Auto mA Cal

HHS Scans

HHS Report

Ground Leakage Test

Heat Soak & Seasoning

– – – –

New Tube Seed Shift

Bleeder Setup (DDC)

Rise and Fall Time (DDC)

Internal Scan Time (DDC)

Heat Soak & Seasoning (Diag)

(2) Press (Start Scan) when it flashes.

(3) Complete the scans.

Page 110: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–102

SECTION 10SYSTEM CALIBRATION

All Calibrations plus n–number correction must be performed to insure optimum image quality.

The Q–Cal procedure must be run before calibration and scanning can begin.

Performing a Q–Cal destroys all old phantom CALS. The following warm–ups, etc. shouldbe strictly observed:

» The DAS should have power turned ON for at least 2 hours before calibrating.

» A very cold tube is required. There should be no exposures for at least one andone half hours before Q–Cal is started (this also means no tube warm–up).Tests that do not interfere with tube cooling will be completed during this wait.

» Verify that the mylar window is centered on the patient alignment lights. If themylar window is not properly centered, one of the metal bands of the mylarwindow assembly could be shading the detector, causing Q–Cal to fail.

Page 111: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–103

10.1 Q–Cal(1) Select SERVICE

Service

(2) Select SYSTEM INTEGRATION

SYSTEMINTEGRATION

(3) Select SYSTEM TESTS

COMPUTER INTEGRATION

TABLE / GANTRY INTEGRATION

GENERATOR CALIBRATION

DAS INTEGRATION

ALIGNMENTS

SYSTEM TESTS

(4) Select START

. . . START. . .

Page 112: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–104

(5) Select Qcal

A Shelltool will open.

System Tests

FastCal

Perpendicular Alignment

Tomo Plane Indication

Exposure Time Accuracy

System Scanning Tests

Set / Adjust N#s

Image Analysis

Phantom Self Calibration

Image Series

––––

Qcal

xtalk

alpha vector

Hot ISO

Air cal

P cal

System Reli Scanning

Save System State

Heat Soak & Seasoning

Streak Test

Teflon Pin Test

(6) Select Cal Types in order

1 – Cold Qcal 40 scans 2 – Heat Qcal 30 scans 3 – Hot Qcal 40 scans

Note 34 If the time since the last Q–Cal was done is greater than 96 hours, Q–Cals may beredone to give a higher degree of confidence that the image series will pass. However,this is not a requirement.

Page 113: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–105

10.2 Crosstalk Calibration(1) Select xtalk

System Tests

FastCal

Perpendicular Alignment

Tomo Plane Indication

Exposure Time Accuracy

System Scanning Tests

Set / Adjust N#s

Image Analysis

Phantom Self Calibration

Image Series

––––

Qcal

xtalk

alpha vector

Hot ISO

Air cal

P cal

System Reli Scanning

Save System State

Heat Soak & Seasoning

Streak Test

Teflon Pin Test

A Shelltool will open.

(2) Select Cal Types in order

4 – Xtalk Air 4 scans

Be sure the table is referenced so the table height is reported on the gantry. Position the tableas low as possible. Adjust the phantom holder so it is as high as possible with respect tothe table. Place the Crosstalk phantom on the phantom holder and center it using thealignment lights. Raise the table 180 mm from the centered position. Now execute theCrosstalk phantom scans:

Select Cal Type

5 – Xtalk Phantom 8 scans

Page 114: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–106

10.3 Alpha Vector Cal GenerationNote 35 If Auto–mA calibration has not been run recently ( <4hours), do so now.

(1) Select alpha vector

System Tests

FastCal

Perpendicular Alignment

Tomo Plane Indication

Exposure Time Accuracy

System Scanning Tests

Set / Adjust N#s

Image Analysis

Phantom Self Calibration

Image Series

––––

Qcal

xtalk

alpha vector

Hot ISO

Air cal

P cal

System Reli Scanning

Save System State

Heat Soak & Seasoning

Streak Test

Teflon Pin Test

A Shelltool will open.

(2) Select Cal Types in order

6 – Afterglow 2 scans

Page 115: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–107

10.4 Hot ISO Cal Generation(1) Select Hot ISO

System Tests

FastCal

Perpendicular Alignment

Tomo Plane Indication

Exposure Time Accuracy

System Scanning Tests

Set / Adjust N#s

Image Analysis

Phantom Self Calibration

Image Series

––––

Qcal

xtalk

alpha vector

Hot ISO

Air cal

P cal

System Reli Scanning

Save System State

Heat Soak & Seasoning

Streak Test

Teflon Pin Test

(2) Perform the scans

Page 116: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–108

10.5 Perform Air and Phantom Calibrations

(1) Select

Scanner Utilities

(2) Select Detailed Calibration

(3) Refer to Illustration 1–45. Select the following screen parameters:

(a) Calibration Type: Air, Small SFOV, Large FOV

(b) KV: Select all the kV settings.

(c) Aperture: Scaled.

Illustration 1–45DETAILED CALIBRATION SCREEN

Calibration Small Large Type SFOV SFOV

Detailed Calibration

KV 80KV 100KV 120KV 140KV

Aperture Scaled 1mm 3mm 5mm 7mm 10mm

AirAdjust CTNumber...

(4) Run the selected Calibrations.

(5) Perform N number correction.

Page 117: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–109

NOTES

Page 118: 1.x,2.x to Cti Upgrade
Page 119: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–110

SECTION 11PACKING MATERIAL FOR RETURN

This section describes the packing for shipment of the GENESIS Console and theSCU cabinet.

11.1 Console PackingThe console should already be on the shipping dollies with the fisch–paper supplied in theUpgrade Kit between the dollies and the console legs. Verify this now.

(1) Raise the dollies enough to move the console over the 48” x 40” wooden skid that theCT/i console was delivered on.

(2) Ensure that no part of the console is extending out over the edges of the skid. THisincludes the ground stud on the back of the console.

(3) Using the dollies as a template, drill 4 holes in the skid, and bolt the console to the skidusing the Lag Screws 5/16” x 1 1/2” and washers provided in the Upgrade kit returnmaterials.

(4) Drill 2 holes in the console lumber brace and bolt it between the dollies using theprovided hardware.

(5) Place a piece of “Kim–Pack” across the top of the console to protect the bezel.

(6) Place cardboard sheets (1 set 46–303734P1) across the top and front of the console.Bend and tape in place.

(7) Place a plastic bag over the entire console.

11.2 SCU Packing(1) Install the shipping brace diagonally across the back of the SCU cabinet.

(2) Attach the two smaller yellow brackets to the large yellow bracket loosely with theprovided hardware.

(3) Position the braces diagonally across the back of the scu cabinet between the cabinetvertical rails and install the hardware to hold it in position.

(4) Tighten the brace hardware to keep the cabinet from twisting during shipment.

(5) Attach the cabinet keys to the cabinet using tape.

(6) Place the SCU cabinet on to the skid (LD00506) provided in the Upgrade Kit ReturnMaterials.

(7) Bolt the SCU in place on the skid using the “L” brackets and hardware provided in theUpgrade Kit Return Materials.

(8) Cover the SCU with a plastic bag provided in the Upgrade Kit Return Materials.

Page 120: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–111

(9) Place cardboard sheets (one set 46–166887P1) on the sides of the cabinet and bendtham into place around the front and rear of the cabinet.

(10)Tape the cardboard to the Cabinet.

11.3 Label Return Material(1) Fill in the information on page Upgrade–112 and Upgrade–114 and attach one each to

the console and SCU.

Page 121: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–112

11.3.1 Return Information

Call: GEMS Recycling Center

414–747–6997

Arrange for pick–up of the GENESIS Console, SCU Cabinet, and the CT/i Shipping Dollies.

Fill in the requested information on these sheets and attach to the outside of the GENESISConsole and SCU Cabinet.

FDO:Site Name:

Removal Date:Field Engineer:

SCU Serial Number:Console Serial Number:

HiSpeed CT/i EquipmentRemoval Information

(This page is not a substitute for equipment de–installation cards)

Attention: GEMS Recycling CenterMAXINE or CHRISTINE

Page 122: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–113

NOTES

Page 123: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–114

11.3.2 Return Information

Call: GEMS Recycling Center

414–747–6997

Arrange for pick–up of the GENESIS Console, SCU Cabinet, and the CT/i Shipping Dollies.

Fill in the requested information on these sheets and attach to the outside of the GENESISConsole and SCU Cabinet.

FDO:Site Name:

Removal Date:Field Engineer:

SCU Serial Number:Console Serial Number:

HiSpeed CT/i EquipmentRemoval Information

(This page is not a substitute for equipment de–installation cards)

Attention: GEMS Recycling CenterMAXINE or CHRISTINE

Page 124: 1.x,2.x to Cti Upgrade

DIRECTION 2159823–100REV 5GE MEDICAL SYSTEMS 1.X / 2.X TO CT/I UPGRADE MANUAL

Upgrade–115

NOTES

Page 125: 1.x,2.x to Cti Upgrade
Page 126: 1.x,2.x to Cti Upgrade

CT SERVICE ENGINEERING, NB–918P.O. BOX 414MILWAUKEE WISCONSIN 53201–0414 U.S.A.Fax: (414) 785–5874