Page 1: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

18O & D/H Analysis of Solids by TC/EA-IRMS

Page 2: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Collect your samples

Page 3: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Sample Preparation

Material needs to be dry and homogenous

Organisms can be analyzed wholeif small enough

Separate samples are required for hydrogen and oxygen

Page 4: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Sample Weights

Sample should contain:400 µg oxygen

80 µg hydrogen

Page 5: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Encapsulating SamplesWeigh samples into

silver capsules

A crimper plate helps shape capsules

The Good

The Bad

The Ugly

Page 6: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Organizing Trays

Load samples across rows of 96-wells trays

Do not leave empty wells between samples

Label trays uniquely and clearly

Place samples for different analyses in different trays

Page 7: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Sealing trays

For small samples, place an index card or parafilm between tray and lid before sealing

Tape lid securely to tray – use lab tape

Page 8: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

International Shipments

Use a shipping company that brokers their own packages (such as Fedex or DHL)

Customs documentsSpecify plant genus and species

Specify for “chemical analysis”

Permits are required for bird feathers

Email [email protected] the appropriate permit

Page 9: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Paperwork

Complete an Analysis Order Form for each analysis type

Complete a sample list for each Analysis Order From

Email forms to [email protected]

Include printed copies with samples

Page 10: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

IRMS Basics

(1) Ionize gas

(2) Focus ion beam

(3) Deflect ions

(4) Detect masses

Page 11: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Pyrolysis (High Temperature)TC/EA-IRMSHD & 18O in solids

H2

CO

Gas separation in GC column

CombustionReactor (1400°C)

GCColumn

H2OTrap

Flow toIRMS

HeFlow

Auto sampler

Page 12: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Chromatography

Page 13: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Standards - Hydrogen

International standard (IAEA-CH-7) and inter-laboratory keratin standards are included in all runs

In-house standards have been calibrated against same standards

All results are expressed with respect to VSMOW

Page 14: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Standards - Oxygen

International standards (IAEA-601, IAEA-602, USGS-8568) are included in all runs

In-house standards have been calibrated against same standards

All results are expressed with respect to VSMOW

Page 15: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Raw Data

Reviewed in reprocessorPoor chromatography

Evaluate reference gas values

Confirm placement of standards

Page 16: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Hydrogen Corrections

All corrections are performed on raw ratios

Memory, if needed

Drift over time

H3+ / Size correction

Scale expansion

Page 17: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Oxygen Corrections

Memory

Drift

Size

Scale expansion

Page 18: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Exchangeable hydrogen

Carboxyl and hydroxyl groups will exchange hydrogen atoms with water vapor

Results in shift of deuterium value

Page 19: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Exchangeable Hydrogen - Keratin

Accepted keratin standards allow measurement of keratin samples by “like treatment”

Standards and samples equilibrate with lab atmosphere and are analyzed together

Page 20: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Exchangeable Hydrogen - Keratin

Correcting to known values of keratin standards determines value of non-exchangeable hydrogen of keratin samples

Only works for keratin

Page 21: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Exchangeable Hydrogen

For non-keratin samples, percent of exchangeable hydrogen must be calculated

Page 22: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Dual-Equilibration

By equilibrating a material with waters of different deuterium values, the percent of exchangeable hydrogen and deuterium value for the non-exchangeable fraction can be determined

Page 23: 18 O & D/H Analysis of Solids by TC/EA-IRMS. Collect your samples

Oxygen – Memory effect

Oxygen is prone to carryover from one sample to the next

Varies with age of reactor and ash accumulation

Can be corrected for by mass balance

Best precaution – run replicates