NEW KERARING SI-6FOR KERATOCONUS TREATMENT
ASCRS, BOSTON 2010
FRANCISCO SANCHEZ LEON MDINSTITUTO NOVAVISION CD. MÉ[email protected] +52555 3438882
NO FINANCIAL INTEREST TO MEDIPHACOS CO.
150°X 200m
150° X 300m
![Page 2: 150°X 200 m](https://reader035.vdocuments.us/reader035/viewer/2022062315/568161a6550346895dd164f8/html5/thumbnails/2.jpg)
Pourpose
Evaluate the corneal remodeling using the new Keraring SI 6 for the Keratocous Treatment.
![Page 3: 150°X 200 m](https://reader035.vdocuments.us/reader035/viewer/2022062315/568161a6550346895dd164f8/html5/thumbnails/3.jpg)
©1997 – 2006 IntraLase Corp. All rights reserved. Confidential Mkt Doc 470 Rev. A
Corneoplastic Techniques
Mechanical methods Intracorneal segmentsIncisional procedures
Biochemical methods Cross-linking UV lightConductive Keratoplasty (CK)
Mix methodsExcimer laser ablationInlays (Acufocus)
![Page 4: 150°X 200 m](https://reader035.vdocuments.us/reader035/viewer/2022062315/568161a6550346895dd164f8/html5/thumbnails/4.jpg)
Optimized prismatic design
Scalene triangular shape SI-6 compared to Isosceles triangle in SI-5.
Improves effect and optimizes total internal reflection, reducing glare and halos
![Page 5: 150°X 200 m](https://reader035.vdocuments.us/reader035/viewer/2022062315/568161a6550346895dd164f8/html5/thumbnails/5.jpg)
800µ base width instead of 600µ
Wider base provides 25% greater implant volume and compensates for reduction of effect in the larger optical zone
![Page 6: 150°X 200 m](https://reader035.vdocuments.us/reader035/viewer/2022062315/568161a6550346895dd164f8/html5/thumbnails/6.jpg)
When should Keraring SI-6 implantation be considered?
Mesopic pupil diameter ≥ 5mm BCVA is ≥ 0.6 Moderete Sphere ≤ 5 D Cylinder value is ≤ 5 D Segment thickness should not exceed 60% of the
minimum pachymetry at the 5.5 optical ring zone When PRK is planned as a second procedure. When phakic IOL implantation is planned as a
second procedure
![Page 7: 150°X 200 m](https://reader035.vdocuments.us/reader035/viewer/2022062315/568161a6550346895dd164f8/html5/thumbnails/7.jpg)
Different Arc Lengths for each Refraction and Topography appearance
![Page 8: 150°X 200 m](https://reader035.vdocuments.us/reader035/viewer/2022062315/568161a6550346895dd164f8/html5/thumbnails/8.jpg)
New Keraring SI-6IS-6 IntraLase 60 Hz parameters Depth in Cornea 80% at 5.5 mm Incision Axis at steepest meridian Entry Cut Length 1 mm
Inner Diameter 6.0 mm Outer Diameter 6.8 mm
Ring Energy 1.5 Entry Cut Energy 1.3
![Page 9: 150°X 200 m](https://reader035.vdocuments.us/reader035/viewer/2022062315/568161a6550346895dd164f8/html5/thumbnails/9.jpg)
New Keraring SI-6150°Asymmetric Thickness Implantation
-3.50/-4.00X135°=20/60
-0.50/-1.25 X 30°=20/30
150°X 200m
150° X 300m
![Page 10: 150°X 200 m](https://reader035.vdocuments.us/reader035/viewer/2022062315/568161a6550346895dd164f8/html5/thumbnails/10.jpg)
+1.50 / -7.50 X 20°= 20/60
Plano/-1.25 X 45°= 20/30
New Keraring SI-6120°symmetric implantation
120°X 250m
![Page 11: 150°X 200 m](https://reader035.vdocuments.us/reader035/viewer/2022062315/568161a6550346895dd164f8/html5/thumbnails/11.jpg)
Single Inferior Segment for PMD Corneoplasty Achivement
Preop Postop
![Page 12: 150°X 200 m](https://reader035.vdocuments.us/reader035/viewer/2022062315/568161a6550346895dd164f8/html5/thumbnails/12.jpg)
New SI-6 KeraringEarly Conclusions Keraring SI-6 achieves Corneoplasty purposes. Keraring SI-6 is not intended to replace the IS-5,
but rather intends to offer the surgeon additional tools to customize the corneal remodeling according to each individual patient’s needs