Elevator Furnace (+1200°C to +1700°C)
LABEC Elevator Furnaces offer a stable loading system for furnaces up to 1700ºC. Due to the electrically driven bottom, the BL series of furnaces can be easily loaded with samples. With their all round heating, these furnaces achieve outstanding temperature uniformity. The furnace chamber is separated from the heating elements due to the use of a tubular plasma ceramic retort placed vertically in the furnace. This design prevents a certain degree of contamination of both the load and heating elements. It is also recommended for gassing processes.
Features include:• Temperature Max 1200ºC to 1700ºC.• High quality molybdenum disulfide heating
elements or Kanthal A1 element spiral wound.• Furnace chamber lined with first class,
durable fibre materials.• Outstanding temperature uniformity
due to all round furnace chamber heating.• Tubular ceramic retort to prevent a certain
level of contamination and to improve temperature uniformity.
• Furnace chamber with a volume of 2 litres, up to 10 litres.
• Precise, electric spindle drive with push button operation.
• Casing is fan cooled and either painted or stainless steel.
• Adjustable air inlet through the floor.• Exhaust air vent in the roof.• Type B thermocouple
(Type K on 1200ºC models.)• Switchgear with SCR control.
Loading platform
Controller - Fitted with a 50 stage programmable digital microprocessor based PID temperature controller, controls the temperature within the range +300°C to +1200°C or +1700°C.
Data logging - The controller can be supplied with RS232/RS485 communication ports which can be connected directly to a data loggers or computers to collect information. Software is available for programming and monitoring.
www.labec.com.au
Laboratory Equipment Pty Ltdemail: [email protected] Ph: 02 9560 2811 • Fax: 02 9560 6131
Serv
ing
the
Scie
ntifi
c, M
edic
al a
nd R
esea
rch
indu
stri
es s
ince
194
5
2014
SPECIFICATIONS
ModelHeated Length (mm)Int. Diameter (mm)Temp. MaxTemp. Setting AccuracyTemp. FluctuationTemp. SensorHeating SystemInterior MaterialExterior MaterialGas Purge
Options
Safety FeaturesPower (W)Weight (kg)
Light Weight Ceramic Fibre
Inlet and outlet for purging atmosphere.
BLF-200/XX BLF-300/XX
200 300
120 160
300
4800
Epoxy Coated Steel / Stainless Steel
Cable entry port, timer, programmable controller, custom diametre tubes, single or multiple heat zones, high temperature models RS232 connection.
Over Temperature Alarm, temperature protection cut off, earth leakage breaker (ELB).
1200°C to 1700°C
1°C
+/- 5%
Type B or K
Electric Elements
XX Specify which temperature i.e BLF-200/12 (temperature range for both models 1200ºC to 1700ºC)