103012 c 230
TRANSCRIPT
-
8/13/2019 103012 c 230
1/28
Chem. 230 10/30 Lecture
-
8/13/2019 103012 c 230
2/28
Announcements
Quiz 3 Today
HW Set 4 will be posted
What we are covering today Quantification in Chromatography
Mass Spectrometry
-
8/13/2019 103012 c 230
3/28
Quantitation in ChromatographyOverview
Performance Measures
Detector Response
Levels of Detection and Quantification Data Smoothing
Integration
Calibration Methods
-
8/13/2019 103012 c 230
4/28
Quantitation in ChromatographyPerformance Measures
Precision How reproducible a measurement is
Accuracy How close measured concentration is to true value
Sensitivity The ability to measure small concentrations or amounts of
analyte
Selectivity Can be an issue in quantification when overlapping/interfering
peaks occur
% Recovery % of analyte added to sample that is measured in sample
-
8/13/2019 103012 c 230
5/28
Quantitation in ChromatographyDetector Response
Concentration Type vs. Mass Flow Type In concentration type, signal depends on analyte in sample cell;
so generally flow independent In mass flow type, signal depends on mass transport to detector
(e.g. in FID without compounds entering flame, no signal willresult)
Note: for some mass flow (HPLC-ABDs and HPLC-MS) transportefficiency depends on liquid flow so signal is not directlyproportional to flow rate
Time
Concentration Detector
flow off
flow on
Time
Mass Flow Detector
flow off flow on
-
8/13/2019 103012 c 230
6/28
Quantitation in ChromatographyDetector Response
Concentration Type - examples:
PID (GC)
UV-Vis (HPLC)
Fluorescence (HPLC) Mass Flow Type - examples:
FID (GC)
NPD (GC)
-
8/13/2019 103012 c 230
7/28
Quantitation in ChromatographyDetector Response
Detector Signal
Depends on concentration of analyte or mass ofanalyte reaching detector
Most (but not all) detectors give linear response overportion of detectable range
Detector Noise
Present in all detectors
High and low frequency types
Ability to Detect Small Quantities Depends onSignal (Peak Height) to Noise Ratios
-
8/13/2019 103012 c 230
8/28
Quantitation in ChromatographyLevels of Detection and Quantification
Noise can have high and low frequency parts
Ways of defining noise peak to peak (roughly 5)
standard deviation (more accurate way)
Signal = peak height
high frequency
component
low frequency component
peak
to
peak
nois
e
-
8/13/2019 103012 c 230
9/28
Quantitation in ChromatographyLevels of Detection and Quantification
Limit of Detection (LOD): minimum detectable signal can be defined as S/Npeak-to-peak= 2
to 3.3 minimum detectable concentration = concentration needed to
get S/Npeak-to-peak= 2 or S/= 3.3
Calculate as 2N/m where m = slope in peak height vs. conc.calibration plot Minimum detectable quantity = (minimum detectable
conc.)(injection volume)
Limit of Quantification (LOQ): Calculated in similar fashion as LOD Lowest concentration to give an reasonable conc. (e.g. can be
auto-integrated using software) Typically 5LOD
-
8/13/2019 103012 c 230
10/28
Quantitation in ChromatographyData Smoothing
Data should be digitized with a frequency ~20/peak width High frequency noise (where fnoise>> fsignal) can be removed by filtering
see example below note: overfiltering results in reduction of signal and loss of resolution overfiltering result also can occur if detector response is too slow (or cell volume
is too large Difficult to remove noise with frequency similar to or lower than peaks
300
350
400
450
500
550
15 17 19 21 23
time
response
Raw DataFiltered Data
Excess Filtering
-
8/13/2019 103012 c 230
11/28
Quantitation in ChromatographyIntegration
Integration of peakshould give: peak height peak area peak width (often just peak
area/peak height) Difficulty comes from
determining if a peak is apeak (or just noise), andwhen to start the peak
and end the peak. Can use auto
integration or manualintegration
we want to pick
up this peak
but not these
noise spikes
-
8/13/2019 103012 c 230
12/28
Quantitation in ChromatographyIntegration
Other issues inintegration (besidesnoise peaks)
start and ends topeaks
how to splitoverlapping peaks
-
8/13/2019 103012 c 230
13/28
Quantitation in ChromatographyIntegration
Peak Height vs. Peak Area Reasons for using peak area
peak area is independent of retention time
(assuming linear response), while the peak heightwill decrease with an increase in retention time
peak area is independent of peak width, while thepeak height will decrease if the column isoverloaded (non-linear response)
Reasons for using peak height Integration errors tend to be smaller if samples are
close to the detection limits
-
8/13/2019 103012 c 230
14/28
Quantitation in ChromatographyLOD/LOQ example
Determine the LODs and LOQ for thefollowing example. Determine it for the4.6 min peak if the concentration is 0.4 ngL-1. Use the 3.3 and 2N LOD defintions.
-
8/13/2019 103012 c 230
15/28
Quantitation in ChromatographyCalibration Methods
External Standard most common method standards run separately and calibration
curve prepared samples run, from peak areas,
concentrations are determined best results if unknown concentration comes
out in calibration standard range
Internal Standard Common for GC with manual injection
(imprecisely known sample volume) Useful if slow drift in detector response Standard added to sample; calibration and
sample determination based on peak arearatio
F = constant where A = area and C = conc.(X = analyte, S = internal standard)
SX
SX
CC
AAF
/
/
Area
Concentration
AX/AS
Conc. X (constant conc. S)
-
8/13/2019 103012 c 230
16/28
Quantitation in ChromatographyCalibration Methods
Standard Addition Used when sample matrix affects
response to analytes Commonly needed for LC-MS with
complicated samples Standard is added to sample (usually
in multiple increments) Needed if slope is affected by matrix Concentration is determined by
extrapolation (= |X-intercept|)
Surrogate Standards Used when actual standard is not
available Should use structurally similar
compounds as standards Will work with some detector types
(FID, RI, ABDs)
Area
Concentration
Added
Analyte
Concentration
0 bmXA
mbX /
standards in water
-
8/13/2019 103012 c 230
17/28
QuantitationAdditional (Recovery Standards + Questions)
Recovery Standards
Principle of use is similar to standard addition
Standard (same as analyte or related compound)
added to sample, then measured (in addition to directmeasurement of sample)
Useful for determining losses during extractions,derivatization, and with matrix effects
expected
unknowntotal
expected
recovered
amount
100amount-amount
amount
100amountrecovered%
-
8/13/2019 103012 c 230
18/28
QuantitationSome Questions/Problems
1. Does increasing the flow rate improve the sensitivity ofa method?
2. Does the use of standard addition make more sensewhen using a selective detector or a universaldetector?
3. Is a matrix effect more likely with a simple sample or acomplex sample?
4. Why is the internal standard calibration more commonwhen using manual injection than injection with anautosampler?
-
8/13/2019 103012 c 230
19/28
QuantitationSome Questions/Problems
5. A scientist is using GC-FID to quantitate hydrocarbons. The FID isexpected to generate equal peak areas for equal numbers ofcarbons (if substances are similar). Determine the concentrationsof compounds X and Y based on the calibration standard (1-octanol). X = hydroxycyclohexane and Y = hydroxypentane.
Compound 1-octanol cC6-OH cC5-OH
Area 3520 299 1839
Conc. (ugmL-1)
10.0 ? ?
-
8/13/2019 103012 c 230
20/28
QuantitationSome More Questions/Problems
6. A chemist is using HPLC with fluorescence detection. He wants tosee if a compound co-eluting with a peak is quenching(decreasing) the fluorescence signal. A set of calibrationstandards gives a slope of 79 mL g-1and an intercept of 3. Theunknown gives a signal of 193 when diluted 4 mL to 5 mL (using
1 mL of water). When 1.0 mL of a 5.0 g mL-1standard is addedto 4.0 mL of the unknown, it gives a signal of 265. What is theconcentration of the unknown compound and is a significantquenching (more than 10% drop in signal) occurring?
-
8/13/2019 103012 c 230
21/28
QuantitationSome More Questions/Problems
6.7. A chemist is testing an extraction process for removing DDT fromfish fat. 8.0 g of fat is first dissolved in 50 mL of 25% methylenechloride in hexane. The 50 mL is divided into two 25 mLportions, one of which is spiked by adding 2.0 mL of 25.0 ng mL-1DDT. Each portion is run through a phenyl type SPE cartridge
and the trapped DDT is eluted with 5.0 mL 100% methylenechloride. The methylene chloride is evaporated off, and thesample is redissolved in 0.5 mL of hexane and injected onto a GC.The un-spiked sample gives a DDT conc. (in 0.5 mL of hexane) of63 ng mL-1, while the spiked sample gives a DDT conc. of 148 ng
mL-1. What is the % recovery? What was the original conc. ofDDT in the fat in ppb?
-
8/13/2019 103012 c 230
22/28
Mass SpectrometeryOverview
Applications of Mass Spectrometry
Mass Spectrometer Components
GC-MS LC-MS
Other Applications
-
8/13/2019 103012 c 230
23/28
Mass SpectrometeryApplications
Direct Analysis of Samples Most common with liquid or solid samples
Reduces sample preparation
Main problem: interfering analytes
Off-line Analysis of Samples Samples can be separated through low or high
efficiency separations
More laborious
Chromatographic Detectors generally most desired type since this allows
resolution of overlapping peaks
-
8/13/2019 103012 c 230
24/28
Mass SpectrometeryApplications
Purposes of Mass Spectrometry Quantitative Analysis (essentially used as any other
chromatographic detector) Advantages:
selective detector (only compounds giving same ion fragmentswill overlap)
overlapping peaks with same ion fragment can be resolved(through deconvolution methods)
semi-universal detector (almost all gases and many solutes inliquid will ionize)
very good sensitivity
Disadvantages cost
requires standards for quantification
-
8/13/2019 103012 c 230
25/28
Mass SpectrometeryApplications
Purposes of Mass Spectrometry - continued Qualitative Analysis/Confirmation of Identity
With ionization method giving fragmentation, few compounds willproduce the same fragmentation pattern
Even for ionization methods that dont cause fragmentation, theparent ion mass to charge data gives information about thecompound identity.
Some degree of elemental determination can be made based onisotopic abundances (e.g. determination of # of Cl atoms in smallmolecules).
Additional information can be obtained from MS-MS (furtherfragmentation of ions) and from high resolution mass spectrometry
(molecular formula) if those options are available. Isotopic Analysis Mass spectrometry allows analysis of the % of specific isotopes
present in compounds (although this is normally done by dedicatedinstruments)
An example of this use is in drug testing to determine iftestosterone is naturally produced or synthetic
-
8/13/2019 103012 c 230
26/28
Mass SpectrometeryInstrumentation
Main Components:
Ion source (more details on subsequent slides)
Analyzer (more details on subsequent slides)
Detector: most common is electron multiplier
Anode
Cathode
Dynodes
M+e-e-
I
Detection Process:
Ion strikes anode
Electrons are ejected
Ejected electrons hitdynodes causing acascade of electronreleases
Current of electronshitting cathode is
measured
-
8/13/2019 103012 c 230
27/28
Mass SpectrometeryInstrumentation
Ion Sources For Gases
Electron Impact (EI): electrons from heated element
strike molecules M + e-=> M+*+ 2e-
M+is the parent ion
Because M+*often has excessenergy, it can fragment further,usually producing a smaller ion
and a radical Fragmentation occurs at bonds,
but electronegative elementstend to keep electrons
e-
+
e-
gas stream M
CH3-Br+*
CH3+ + Br
Main
fragment
CH3+ Br+
Minor or unobserved
fragment
-
8/13/2019 103012 c 230
28/28
Mass SpectrometeryInstrumentation
Ion Sources
For Gases
Chemical Ionization (CI):
Can produce positive or negative ions First, a reagent gas reacts with a corona discharge to
produce a reagent ion: CH4=> => CH5+(more likely
CH4H+)
Then the reagent ion transfers its charge to a molecule:
M + CH5+=> MH+ (one of largest peak has mass tocharge ratio of MW + 1)
Less fragmentation occurs, so more useful for identifyingthe parent ion