WiSIM DIA Method Set-upOrbitrap Fusion MS
MS Instrument Control Software v.1.1 SP1
2
WiSIM DIA, how does it work?
• Unique to the Orbitrap Fusion MS: • Three HR/AM SIM are used to cover precursors ions from 400 to 1000
m/z MS1 ultra high resolution quantification
• In parallel with each HR/AM SIM scan, 17 sequential Ion Trap CID MS/MS with 12 amu isolation windows are acquired to cover the associated 200 amu HR/AM SIM mass range MS2 confirmation
SIM 400-600 amu SIM 600-800 amu SIM 800-1000 amu
406 418 430 442 598 606 618 630 642 898 806 818 830 842 998
tSIM Orbitrap
tMS2 Ion Trap
3
WiSIM DIA, how does it work?
SIM 400 – 600 m/z
MS2 No. 1………………MS2 No.17
1st MS1
SIM 600 – 800 m/z
Ion
Tra
pO
T24
0k
3.6
sec
SIM 800 – 1000 m/z
2nd MS1MS2 No. 1………………MS2 No.17
Ion
Tra
p
3rd MS1MS2 No. 1………………MS2 No.17
Ion
Tra
pO
T24
0kO
T24
0k
17 Windows of 12 amu each
17 Windows of 12 amu each
17 Windows of 12 amu each
4
WiSIM DIA Method Set-up at a glance:
SIM• 400-600, 600-800, 800-1000 m/z• 200 amu ion trap isolation width• Resolution@ 240 000• AGC: 3e4• 50 ms Max injection • Profile
tMS2• 12 amu quadrupole isolation• 150-1850 m/z mass range• Ion Trap detection• AGC: 5e4• 35 ms Max Injection • CID 30 • Rapid Scan rate
X3
X3
6 experiments:
406 514418 526430 538442 550454 562466 574478 586490 598502
606 714618 726630 738642 750654 762666 774678 786690 798702
806 914818 926830 938842 950854 962866 974878 986890 998902
5
WiSIM DIA Method Set-up:
6 experiments:-Three HR/AM tSIM-Three Ion trap tMS2 CID
6
Experiment 1:
SIM 400-600 amu SIM 600-800 amu SIM 800-1000 amu
406 418 430 442 598 606 618 630 642 898 806 818 830 842 998
7
Experiment 2:
SIM 400-600 amu SIM 600-800 amu SIM 800-1000 amu
406 418 430 442 598 606 618 630 642 898 806 818 830 842 998
8
Experiment 2:
SIM 400-600 amu SIM 600-800 amu SIM 800-1000 amu
406 418 430 442 598 606 618 630 642 898 806 818 830 842 998
9
Experiment 3:
SIM 400-600 amu SIM 600-800 amu SIM 800-1000 amu
406 418 430 442 598 606 618 630 642 898 806 818 830 842 998
10
Experiment 4:
SIM 400-600 amu SIM 600-800 amu SIM 800-1000 amu
406 418 430 442 598 606 618 630 642 898 806 818 830 842 998
11
Experiment 4:
SIM 400-600 amu SIM 600-800 amu SIM 800-1000 amu
406 418 430 442 598 606 618 630 642 898 806 818 830 842 998
12
Experiment 5:
SIM 400-600 amu SIM 600-800 amu SIM 800-1000 amu
406 418 430 442 598 606 618 630 642 898 806 818 830 842 998
13
Experiment 6:
SIM 400-600 amu SIM 600-800 amu SIM 800-1000 amu
406 418 430 442 598 606 618 630 642 898 806 818 830 842 998
14
Experiment 6:
SIM 400-600 amu SIM 600-800 amu SIM 800-1000 amu
406 418 430 442 598 606 618 630 642 898 806 818 830 842 998
15
WiSIM-DIA advantages
• Low detection limits, high selectivity is achieved using 240,000 SIM scan.
• High sensitivity Ion Trap CID MS2 is used for confirmation of the low abundant components.
• Unsurpassed speed with 2 detectors working in parallel allows for 3.6 s cycle.
• LOD and LOQ down to 10 attomole.• CVs are better than 15 % for 85 % of compounds, better than
20% for 95% of compounds.• Linear dynamic range up to 4 orders of magnitude.