Page 1: 004-10061-00 -Service Manual: QuickClean Ultrasonic

FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY

Model Numbers:Service and

Parts Manual

QuickCleanUltrasonic

QC1-01QC3(R)-01QC6(R)-01

004-10061-00 Rev. AA8 (1/21/20)Language of origin: English

Page 2: 004-10061-00 -Service Manual: QuickClean Ultrasonic

ii© Midmark Corporation 2018 004-10061-00 (07/11/18 )

Tabl

e of

Con

tent

sGENERAL INFORMATION

Symbols .................................................................iiiOrdering Parts .......................................................iiiSerial Number Location .......................................iiiSpecifications ....................................................... ivMaintenance ......................................................... vi

TROubLEshOOTINGView Cycle History ........................................... A-2View Temperature Sensor Readings ............. A-3Troubleshooting Chart ..................................... A-4

TEsTING & REpAIRTemperature Sensor Test ................................ B-2 Level Sensor Test ........................................ B-3Replace Level Sensor ...................................... B-4Test PSB Fuses (Power Supply & Generator PCB ) ... B-5Test Fuses (Main Power Socket) .............................. B-6Test IGBT’s ........................................................ B-7 Adjust LCD Display Contrast .......................B-8Replace LCD Display ....................................... B-9Replace Power Switch ................................... B-10Replace Power Socket ...................................B-11Ground Continuity Test .................................. B-12Isulation Resistance Test .............................. B-13 Transducer Testing .................................... B-14

Gen

eral

Info

Sect

ion

A

Sect

ion

E

Sect

ion

B

Sect

ion

CSe

ctio

n D

(*) Indicates multiple pages due to model / serial number break(s).

AccEss pROcEduREsRemove Base Plate - Table Top .................... C-2Remove Tank Assembly - Table Top ............. C-3Remove Recessed Unit from Counter Top ... C-4Remove Base Plate - Recessed ................... C-5Remove Tank Assembly - Recessed ............. C-6

WIRING dIAGRAMsTable Top Unit ................................................... D-2Recessed Unit ................................................... D-3Recessed Display Panel Assembly ............... D-4

pARTs LIsTTable Top (QC1-01, QC3-01,QC6-01) ................... E-2Recessed (QC3R-01, QC6R-01) .......................... E-3Basket .............................................................. E-12Beaker Holder Assembly ............................... E-13

Page 3: 004-10061-00 -Service Manual: QuickClean Ultrasonic

iii© Midmark Corporation 2018 004-10061-00 (11/13/18)

Gen

eral

Info

rmat

ion WARNING

Indicates a potentially hazardous situation which could result in serious injury if not avoided.

Equipment Alert Indicates a potentially hazardous situation which could result in equipment damage if not avoided.

caution Indicates a potentially hazardous situation which may result in minor or moderate injury if not avoided. It may also be used to alert against unsafe practices

symbols Ordering partsThe following information is required when ordering parts: • Serialnumber&modelnumber • Partnumberfordesiredpart (Refer to Section E: Exploded Views & Parts Lists) For warranty parts orders, call Midmark’s Technical Service Department with the required information. Hours:8:00amto5:00p.m.EST(MondaythruFriday) Phone:1-800-Midmark(1-800-643-6275)

serial Number Location

NoteUsed for special instructions or additional information.

serial Number

Page 4: 004-10061-00 -Service Manual: QuickClean Ultrasonic

iv© Midmark Corporation 2018 004-10061-00 (06/08/17)

SpecificationsOperating EnvironmentAmbientTemperature 41°Fto104°F(5to40°C)Maximumrelativehumidity 80%R.H.inroomtemperaturesupto88°F(31°C)decreasinglinearly

to50%R.H.at104°F(40°C)Altitudeabovesealevel Upto6,560ft(2000m)OperatingEnvironment Indooruseonly

UL 61010-1:2012CAN/CSA C22.2 No. 61010-1:2012

Safetyrequirementsforelectricalequipmentformeasurement,controlandlaboratoryuse

FCCCFR47Part15PARTB EMCComplianceEN 60601-1-2:2007 EMCCompliance

(ECmedicaldevicesemissionsGroup1ClassA)EN 61326:2013 EMCcompliance

(EC,emissionsGroup1ClassA)

Electrical Requirements

standards

NOTE: Iftheequipmentisusedinamannernotspecifiedbythemanufacturer,theprotectionprovidedbythe equipmentmaybeimpaired.

Connecttoa115VAC60Hzfullygroundedsupplyviaa3pinplug.Itcanbedangeroustooperateanultrasonicbathwithoutagrounded connection.

Thepowerplugisfittedwitha10Afuse(T10AH250V).NEVERFITAFUSEOFAHIGHERRATING.Ensurethatexcesspowercableisstoredneatly.

115ElectricalDetailsPower supply: 115V @ 60 HzUltrasonic frequency: 32 to 38kHzPollution degree: 2Installation category: II

Note: Supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.

Page 5: 004-10061-00 -Service Manual: QuickClean Ultrasonic

v© Midmark Corporation 2018 004-10061-00 (06/08/17)

WorkingCapacity

Tank capacitymax.

TotalPower(W)

Ultrasonic power

(W)QC1 1.2 gal

(4.5L)1.3 gal(5L)

110 100

QC3(QC3R)

3.3 gal(12.5L)

3.4 gal (13 L)

210 200

QC6(QC6R)

6.6 gal(25L)

7.4 gal (28 L)

410 400

TankExternalDimensionsin(mm)

TankInternal Dimensionsin(mm)

BasketInternal Dimensionsin(mm)

A B C D E F G X Y Z

QC1 13.5(345)

7.8(200)

10.2(260)

10.4(265)

11.8(300)

5.9(150)

5.9(150)

10.4(265)

4.5(115)

4.3(110)

QC3 14.7(375)

13.5(345)

16.5(420)

10.4(265)

13.0(330)

11.8(300)

5.9 (150)

11.6(295)

10.4(265)

4.3(110)

QC6 21.6(550)

13.7 (350)

16.5(420)

14.3(365)

19.9(505)

11.8(300)

7.9(200)

18.3(465)

10.4(265)

6.3(160)

dimensions

capacity / power

Page 6: 004-10061-00 -Service Manual: QuickClean Ultrasonic

vi© Midmark Corporation 2018 004-10061-00 (11/13/18)

Maintenance

Maintaining Your Ultrasonic

Itisimportanttokeepyourbathclean.Notonlywillcontaminatedliquidreducetheperformanceofthebath,itmayalsodamageit.Changethecleaningliquidregularly.Yourcleaningprocesswilldeterminehowoftentochangetheliquid–themoresoiledyouritems,themoreoftenyouwillneedtochangetheliquid.

Beforecleaningtheequipmentalwaysswitchoffanddisconnectitfromthepowersupplyandallowittocooldowntolessthan40°C/104°F.Cleanbywipingwithadampsoapyclothandrinse.

Thebaseofthebathgeneratestheultrasonicactivitybyvibratingatveryhighspeeds.Ifanycon-taminantsareincontactwiththebath,theyactasanabrasive,causingwearonthemetalsurface.Inextremecases,thebathwilldevelopholesandstarttoleak.

Perform Periodic “Foil” TestFollow the foil test guidelines in accordance with your local regulations.

Purpose: UsedtoDeterminetheEfficiencyofanUltrasonicCleaner.• Periodictestingoftheultrasoniccleanertoassurethatitisoperatingproperly.• Testingshouldalsobeperformedwheneverdeteriorationinthecleanerperformanceissuspected.• Foiltestarerelativelyeasytoperformandwillprovideapermanentrecordforthefuturecomparative

evaluationoftheultrasoniccleanersperformance.

Page 7: 004-10061-00 -Service Manual: QuickClean Ultrasonic

A-1© Midmark Corporation 2018 004-10061-00 (06/08/17)

Sect

ion

ATroubleshooting

Calling for ServiceContactyourMidmarkAuthorisedDealer.Modelandserialnumberinformationwillberequiredwhencallingforserviceandcanbefoundonthebackoftheunit.

Page 8: 004-10061-00 -Service Manual: QuickClean Ultrasonic

A-2© Midmark Corporation 2018 004-10061-00 (11/13/18)

Temp at Cycle End

Cycle Time Setting

View Cycle History View Cycle History.A) To view information about the last 5 cycles, press and hold the top and bottom arrow keys while turning the power switch on.

cycle Configuration

degas On

Leap On

Level On

cycle passed

1

2

3

4

5

6

789

10

11

12

13

14

15

TAbLE 1

Abort Reason description

1 Abort by user2 Abort by Level Low4 Abort by Low sonics

8 Abort by power Failure16 Abort Water Over Temperature32 Abort heater Over Temperature64 Abort unit Over Temperature

TAbLE 2

Top Arrow Key

Bottom Arrow Key

Temp at Start of Cycle °C Temp Setting for

Start of Cycle (N/A For This Model)

Cycle Configuration (SEE TABLE 1)

Aborted Cycle Reason (SEE TABLE 2)

Last 5 Cycles (A thru E)

Cycle Power Setting

Page 9: 004-10061-00 -Service Manual: QuickClean Ultrasonic

A-3© Midmark Corporation 2018 004-10061-00 (11/13/18)

View Temperature Sensors

View Temperature Sensors

A) To view the current values of the temerature sensors on the display, press and hold the bottom arrow key while turning the power switch on.

Note: Temperature sensor failures for ch1 & ch3 would be indicated by the one of the two readings.

197 Open sensor

0 shorted sensor

Top Arrow Key

Bottom Arrow Key

Channel 1 Water Temp in °C Channel 1 °F Software Mode

N/A

Channel 2 (N/A For This Model). Should Display 50-51

Channel 3 Base Plate Temp in °C

Page 10: 004-10061-00 -Service Manual: QuickClean Ultrasonic

A-4© Midmark Corporation 2018 004-10061-00 (11/13/18)

Troubleshooting chart

Problem Symptom Probable Cause Check CorrectionLiquid Level Low (repeatedly) Liquid Level Low (repeatedly) LiquidisnottouchingtheLevel

Sensor.CheckthattheLevelSensorandsurroundingareaiscleanof contaminant.

Filltanktorequiredlevel.

Detergentnotadded. (watertopure)

Checkcorrectdetergentisbeingused.

Adddetergent.

Disconnection or failure of Level SensorComponent.

CheckconnectiontoControlBoardandresistanceofLevelSensor.(See Level Sensor Test)

ReconnectorreplaceLevelSensoriffaulty.(see testing & repair)

Over Temp (repeatedly) Over Temp (repeatedly) Liquidisabove60°C/140°F. (TankTemperatureSensor)

Checkfluidtemperature,ensuretemperatureisbelow60°C/140°F.

Allowfluidtocoolordrainsomefluid&replacewithcoolerfluid.

Rapid Temperature Fluctuations

Rapid Temperature Fluctuations

Disconnection or failure of TemperatureSensor.

CheckconnectiontocontrolboardandresistanceofTemperatureSensor. (See Temp Sensor Test)

ReconnectorreplaceTemperatureSensoriffaulty.(see testing & repair)

Sensor Fault / Over Temp (in Cycle)

Sensor Fault / Over Temp (in Cycle).Erratic temp readings

TemperatureSensoropencircuitordisconnected.

CheckconnectiontoControlBoardandresistanceofTemperatureSensor.(See Temp Sensor Test)

ReconnectorreplaceTemperatureSensoriffaulty.(see testing & repair)

Sensor Fault / (in Standby) Sensor Fault / (in Standby) TemperatureSensorclosecircuit.

CheckresistanceofTemperatureSensor.(See Temp Sensor Test)

ReplaceTemperatureSensoriffaulty. (see testing & repair)

Unit Over Temp Unit Over Temp Transistorshavereached100°C/212°F. (BasePlateTemperatureSensor)

Checkwithuserthefrequencyofuse(avoidconsistentusewithouttimetocool).

Allowtimetocool.

Checkthatthebathhasnotbeenoverloaded(largemass).

Removeexcessiveitemsfromtankandrestartcycle.

Checkthatthebathhasnotbeenrunwithoutsufficientfluid.

Filltanktorequiredlevel.

Checkthatthelevelsensorhasnotbeenturnedoffintheadvancedmenuandrunwithoutsufficientfluid.

sonics Low Sonics Low Sonicsfeedbackexceedinglimits.

Checkthatthetankhasnotbeenoverloaded(largemass).

Removeexcessiveitemsfromtankandrestartcycle.

UltrasonicGeneratorPCBFuseblown.

Check2xFusesonUltrasonicGeneratorPCB.(See IGBT Test)

Replacefuses.(See PCB Fuses)

IGBTsblown. Check2xIGBTsonUltrasonicGeneratorPCB.(See IGBT Test)

ReplaceBasePlateAsm.

WARNING some procedures require power to be connected with covers removed. Line voltage is present. use extreme caution to prevent electric shock.

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 11: 004-10061-00 -Service Manual: QuickClean Ultrasonic

A-5© Midmark Corporation 2018 004-10061-00(9/24/18)

Troubleshooting chart

Problem Symptom Probable Cause Check CorrectionNo Display No Display Nopower. Checkthatmainpowerisconnect-

ed,CheckthatmainPowerSwitchisinONposition.

Connectmainsandswitchonunitatthe rear.

Nopower. Checkmainfuses. Replacefusesatrearofbath.

RibboncablebetweenControlPCBandPowerSupplyPCBis disconnected.

Checkribboncableconnection. Reconnectribboncable.

PowerSupplyPCBfusesblown.

CheckPowerSupplyPCBfuses. ReplacePowerSupplyPCBfuses.

FaultyLCD. Checkalltheabove. ReplaceLCD.

Backlight without text on LCD Display

Backlight without text on LCD Display

Contrast not set. Checkiftextappearswhile adjustingthepotontheControlPCB.

Adjustcontrast. (LCD Display Contrast)

Pixels missing from LCD Display

Pixels missing from LCD Display

FluidenteredLCDDisplay. Eliminatethesourceoffluid. ReplaceLCDDisplay.

Touch Pad buttons not responding

Touch Pad buttons not responding

FluidenteredTouchPad. Eliminatethesourceoffluid. ReplaceTouchPad.

FaultyTouchPad. Checkalltheabove. ReplaceTouchPad.

Fluid Leaking Fluid Leaking Plumbing. Checkallplumbingjoints. Tighten,sealorreplaceaffectedjoints.

Severe cavitation erosion. Checkforexcessivepittinginthebaseofthetank,visiblefromoutside.

Replaceultrasonic.

No Power No Power Notpluggedintopoweroutlet. Checkmainpowerconnection. Connecttopower.

IECfusesblown. Over current. ReplaceIECFuses.(See Power Socket Fuses)

FaultyPowerSwitch. CheckswitchilluminatesinOnposition.

Ifnotilluminated,replacePowerswitch. (See Power Switch)

GFIistripped CheckifGFIistripped ResetGFIWeak/FaultyGFIoutlet. Checkifsterilizerworkson

differentGFIoutletContactanelectriciantoreplaceGFIoutlet

Voltageleakthrough components

Checkallelectricalcomponentsandwireconnectionsfor deterioration

Secure/Clean/Replace compromisedcomponent.

Page 12: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-1© Midmark Corporation 2018 004-10061-00 (11/13/18)

Sect

ion

BTesting & Repair Temperature Sensor Test ................................ B-2

Replace Tank Temperature Sensor ............. B-3Replace Base Plate Temperature Sensor .... B-4Level Sensor Test ............................................. B-5Replace Level Sensor ...................................... B-6Test PSB Fuses (Power Supply & Generator PCB ) ... B-7Test Fuses (Main Power Socket) .............................. B-8Test IGBT’s ........................................................ B-9 Adjust LCD Display Contrast ..................... B-10Replace LCD Display ......................................B-11Replace Power Switch ................................... B-12Replace Power Socket .................................. B-13

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 13: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-2© Midmark Corporation 2018 004-10061-00 (11/13/18)

Temperature Sensor Test NoteIt is not necessary to remove tank from case to test Temperature Sensor. Only remove tank when tank sensor replacement is necessary.

Control PCB

Degrees°F 0 1 2 3 4 5 6 7 8 950 103.90 104.12 104.34 104.55 104.77 104.98 105.20 105.42 105.63 105.85

60 106.07 106.28 106.50 106.71 106.93 107.15 107.36 107.58 107.79 108.01

70 108.23 108.44 108.66 108.87 109.09 109.30 109.52 109.73 109.95 110.17

80 110.38 110.60 110.81 111.03 111.24 111.46 111.67 111.89 112.10 112.32

90 112.53 112.75 112.96 113.18 113.39 113.61 113.82 114.04 114.25 114.47

100 114.68 114.90 115.11 115.33 115.54 115.76 115.97 116.18 116.40 116.61

110 116.83 117.04 117.26 117.47 117.68 117.90 118.11 118.33 118.54 118.76

120 118.97 119.18 119.40 119.61 119.82 120.04 120.25 120.47 120.68 120.89

130 121.11 121.32 121.53 121.75 121.96 122.18 122.39 122.60 122.82 123.03

140 123.24 123.46 123.67 123.88 124.09 124.31 124.52 124.73 124.95 125.16

Resistance in Ω (Ohms)To Tank SensorTo Base Plate Sensor

Test Temperature Sensor.A) Turn power off, unplug and drain water from tank.B) Turn unit upside down to remove bottom plate. (see access procedures).C) Set meter to Ώ.D) Remove appropriate red & white Temp Sensor wires from the Control PCB by pushing down on the tab and pulling out the wire.E) Measure resistance, refer to the table below. Resistance should correspond with the ambient temperature. i.e 75°F should be approximately109.3Ω. F) If the reading is not representative of the table, replace the sensor. (see testing & repair).

To Tank Sensor

Base Plate Temperature SensorRecessed

Table Top

Tank Temperature Sensor

To Base Plate Sensor

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 14: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-3© Midmark Corporation 2018 004-10061-00 (11/13/18)

Test Fill Level Sensor.A) Turn power off, unplug and drain water from tank.B) Remove the Base Plate following the procedures shown in this manual (see access procedures). C)SetmetertoΏD) Remove yellow level sensor wire from the control PCB.E) Measure the conductivity between the level sensor and the tank (Test A) should read open. If the conductivity is not open, check sensor insulation or replace. (see level sensor replacement)F) Measuring the conductivity between the level sensor inside the tank (TestB)andtheendofthecableshouldreadaround1Ω.Checkforthe correct assembly of the Level sensor. If conductivity is open, replace sensor. (see level sensor replacement)

Fill Level Sensor Test

Test Meter Reading Required ActionTest A OL when tank

is emptyIftheconductivityisnotopen,checksensor insulationorreplace.

Test b < 1 ohm Wire Harness OK Replaceifreadings

are out of range.

Fill Level Sensor

TEST BTEST

A

NoteIt is not necessary to remove tank from case to test Level Sensor. Only remove tank when tank sensor replacement is necessary.

Control PCB

To Fill Level Sensor

To Fill Level Sensor

Recessed

Table Top

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 15: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-4© Midmark Corporation 2018 004-10061-00 (11/13/18)

Replace Fill Level sensor

step 1: Replace Fill Level sensor.A) Turn power off, unplug and drain water from tank.B) Remove the tank following the procedures shown in this manual (see access procedures).C) Remove yellow wire from Control PCB by pressing on tab and pulling wire out of connector.D) Remove nuts (2), plain washers (3), fill level sensor harness & Nylon washers (2) from outside of tank. E) Remove screw and nylon insulator from in side of tank. F) Reassemble in reverse order.

M3 x 12mm screw and Insulator are inside tank

Nylon Washer

Level Sensor Harness

M3 Nyloc Nut

Nylon Insulator

M3x12 Dome Head Screw

M4 Plain Washer

M3 Plain Nut

NoteFill level sensor screw and plain nut must be tight enough to form a good seal. Overtightening will lead to insulator deforming and possible leaking.

M3 Plain Washer

Control PCB

To Fill Level Sensor

To Fill Level Sensor

Recessed

Table Top

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 16: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-5© Midmark Corporation 2018 004-10061-00 (11/13/18)

check pcb Fuses

Power Supply PCB

FUSE 2

FUSE 1

Ultrasonic Generator PCB

FUSE 2

FUSE 1

Test Meter Reading Required ActionFuse conductivity test 0-1Ω. Fuseisgood

Open or OLFuseisbadreplacewith

samevaluefuse..

Test Fuses.A) Turn power off, unplug and drain water from tank.B) Turn unit upside down and remove bottom plate (see access procedures).C)SetmetertoΩD) Remove fuses from top of holder E) Measuringacrossthefuseshouldread0-1Ω.Iffusemeasuresopenor OL, replace the fuse.

NoteTo remove the fuses on the Power Supply, & Ultrasonic Generator PCB’s carefully pull up on the top of the holder.

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 17: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-6© Midmark Corporation 2018 004-10061-00 (11/13/18)

check Main power socket Fuses

step 1: Test Main power socket Fuses.A) Unplug the power cord from the main power socket at the rear of the unit. B) Grip top and bottom tabs of the fuse cover located at the bottom of the main power socket and pull straight out.C) Rotate the end of the fuse cover down to expose the fuses.D) Remove fuses, test, and replace any bad fuses with fuses of the same rating.

Tabs

Fuse Cover

Main Power Socket

Test Meter Reading Required ActionFuse conductivity test 0-1Ω. Fuseisgood

Open or OLFuseisbadreplacewith

samevaluefuse..

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 18: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-7© Midmark Corporation 2018 004-10061-00 (11/13/18)

Ultrasonic Generator PCB

Test IGBT’s. A) Turn power off, unplug and drain water from tank.B) Turn unit upside down and remove bottom plate (see access procedures).C) Set meter to D) Place the black meter lead on the center leg of the IGBT and the red lead on one of the side legs then the other.E) Each measurement should be between 350 - 700.

Test Meter Reading Required ActionIGbT Test between

350 - 700Readingofbetween 350-700forbothtest theIGBTisgood

A reading outsideof 350 - 700

Readingoutsideof 350-700foreithertestIGBTisbadandneeds

replaced

check IGbT’s

IGBT

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 19: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-8© Midmark Corporation 2018 004-10061-00 (11/13/18)

Adjusting Lcd display contrast

Maximum Contrast

Optimum Contrast

Minimum Contrast

step 1: Adjusting Lcd contrast.A) Gain access to the Control PCB using the procedures shown in this manual. (see LCD display removal)B) Locate (VR7) on the Control PCB & turn it to adjust the LCD display to the optimum contrast level shown. C) Rotating Pot clockwise will make dispaly lighter and counterclockwise will make display darker.

WARNING some procedures require power to be connected with covers removed. Line voltage is present. use extreme caution to prevent electric shock.

Page 20: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-9© Midmark Corporation 2018 004-10061-00 (11/13/18)

Replace Lcd display Table Top

step 1: Remove and Replace Lcd display.A) Turn power off, unplug and drain water from tank.B) Remove the tank following the procedures shown in this manual. (see access procedures) C) Remove sensor wires going to the Control PCB by pressing in on tab & pulling on wire. D) Remove ribbon cable by pulling outward on the side tabs and removing connector.E) Remove front cover mounting torx screws(6), remove front cover.F) Remove outer mounting torx screws (4) attaching the Control PCB and LCD Display .G) Disconnect the touch pad ribbon cable connecting the Control PCB to the Touch Pad.H) Remove stand off screws holding the two circuit boards together.I) Separate the boards by pulling them apart at the connection.J) Reassemble unit with new LCD Display by reversing these directions.

Mounting Screws

Front cover Mounting screws

Ribbon CableConnection

Remove Wires

Board Connection

Touch Pad Ribbon Cable

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 21: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-10© Midmark Corporation 2018 004-10061-00 (11/13/18)

Replace power switch

step 1: Replace power switchA) Turn power off, unplug and drain water from tank.B) Remove Bottom Plate Assembly (see access procedures). C) Remove the brown & blue wires from the back of the switch.D) Press in on the side tabs and push the switch out of the case.E) Replace switch, install wires and re-assemble.

To Case

Main Filter Power Socket

Power Switch

Power Switch

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 22: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-11© Midmark Corporation 2018 004-10061-00 (11/13/18)

Replace power socket

Power Socket

To Case

Main Filter Power Socket

Power Switch

step 1: Replace power socketA) Turn power off, unplug and drain water from tank.B) Remove Bottom Plate Assembly (see access procedures). C) Remove the brown, blue & green wires from the back of the socket.D) Remove the screws and pull the socket out of the case.E) Replace socket, install wires and re-assemble.

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 23: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-12© Midmark Corporation 2018 004-10061-00 (11/13/18))

Ground continuity Test

step 1: Ground continuity TestA) Turn power off and unplug main power socket at the rear of the unit. B) Visually inspect cord, plugs, and etc. for damage or cracks.C) Set meter to continuity test mode as shown.D) Connect one test lead to the power plug ground pin and other lead to a conductive surface on the QuickClean™ Ultrasonic Cleaner.

Note: Refer to multimeter instructions for specific operation and testing procedures.

Continuity Test Mode

Ground Pin

Conductive Surface

Test Meter Reading Required ActionGround continuity Test Less than 1Ω.

/ Meter beeps Groundisgood

Greater than 1Ω/ Meter doesn’t

beep

Groundisbad.Verifythatcordsandwiringareundamaged,routed,connectedproperlythentestunitagain.Ifunitcontinuestofail,

contactMidmark’sTechnicalService,

NoteUse a standard multimeter to perform this test.

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 24: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-13© Midmark Corporation 2018 004-10061-00 (11/13/18)

Insulation Resistance Test

step 1: Insulation Resistance TestA) Turn power off and unplug main power socket at the rear of the unit. B) Connect large test lead to both the hot and neutral pin and connect the other test lead to the ground pin as shown.C) Push the power switch to the on position on the QuickClean™ Ulrasonic Cleaner.D) Set Insulation resistance meter to 500V and run insulation test.

Note: Refer to insulation resistance tester instructions for specific operation and testing procedure.

Meter Setting

Large Test Lead

Test Meter Reading Required Action

Insulation Resistance Test

Greater than 1 MΩ. Insulationisgood.

Less than 1 MΩ.

Insulationisbad.Verifythatcordsandwiringareundamaged,routed,&connectedproperlythentestunitagain.Ifunitcontinuestofail,

contactMidmark’sTechnicalService.

WARNING To prevent electrical shock, fire, or personal injury, do not touch the Quickclean™ ultrasonic cleaner during testing and make sure the test has been completed before disconnecting.

Equipment Alert Only perform Insulation Resistance Test after performing and passing the Ground Continuity Test.

Ground Pin

NoteAn Insulation Resistance Tester is required to perform this test.

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 25: 004-10061-00 -Service Manual: QuickClean Ultrasonic

B-14© Midmark Corporation 2018

Generator Board

004-10061-00(4/30/19)

step 1: Inspect Transducer ceramic Rings.A) Turn power off and unplug main power socket at the rear of the unit. B) Remove bottom cover. Visually inspect the Ceramic Rings on each transducer for cracks or missing pieces.C) If all transducers are not in good condition Contact Midmark’s Technical Service.

Test Meter Reading Required Action

Transducer Resistance / “Ω” Ohm

Test

Open circuit Resistanceisgood.

closed circuit Resistanceisbad.ContactMidmark’s TechnicalService.

Transducer capacitance Test

=/- 5%

Qc1 = 7nFQc3 = 14nF Qc6 = 28nF

Capacitanceinrange.Transducersaregood.

Capacitanceoutofrange.TransducersarebadContactMidmark’sTechnicalService.

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Transducer Testing step 2: Test Transducers.A) Disconnect red and black leads from Generator Board, briefly connect them together to discharge the transducers. B) Use meter to measure resistance across the red and black wires. C) If meter is equipped, change the meter setting to measure capacitance, connect leads to the red and black wires

Ceramic Rings

Bottom of Tank

Page 26: 004-10061-00 -Service Manual: QuickClean Ultrasonic

C-1© Midmark Corporation 2018 004-10061-00 (06/08/17)

Sect

ion

CAccess procedures

Remove Base Plate - Table Top .................... C-2Remove Tank Assembly - Table Top ............. C-3Remove Recessed Unit from Counter Top ... C-4Remove Base Plate - Recessed ................... C-5Remove Tank Assembly - Recessed ............. C-6

Page 27: 004-10061-00 -Service Manual: QuickClean Ultrasonic

C-2© Midmark Corporation 2018 004-10061-00 (11/13/18)

step 1: Remove base plate Assembly.A) Turn power off, unplug, and drain water from tank and place the front (display side) of ultrasonic down on a soft cloth to prevent damage to the display.B) Remove the four rubber feet (each corner) and screw(s) (number of screws depends on model) from bottom of the ultrasonic.C) Separate bottom plate from ultrasonic and rotate top of plate down and away from unit. This allows you to place the bottom plate on the table with out disconnecting wires.

Remove base plate Assembly (Table Top)WARNING

To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 28: 004-10061-00 -Service Manual: QuickClean Ultrasonic

C-3© Midmark Corporation 2018 004-10061-00 (11/13/18)

step 1: Remove base plate & disconnect Wiring.A) Remove Bottom Plate Assembly. (see base plate removal)B) Disconnect red and black power wires at the Ultrasonic Generator PCB by loosening the screws on the connector and removing the wires. C) Disconnect yellow level sensor wire & red and white thermal sensor wires (going to side of tank) from Control PCB using a small screwdriver to push on the tabs while pulling wires from connector.

step 2: Remove Tank Assembly.A) Remove nyloc nuts, washers & terminal holding the ground wire from the tank studs in the corners.B) Depress and hold the quick fitting elbow retaining ring (shown in red) toward the elbow to release it’s grip on the metal tube coming from the bottom of the tank. C) Slowly remove the tank from the case. The metal tube extending into the elbow will come out of the elbow fitting as the tank is removed.D) Reassemble in reverse order.E) Replace any zip ties that were cut during disassembly.

Remove Tank Assembly (Table Top)

Ultrasonic Generator PCB

Power Supply PCB

Quick Fitting Elbow

Red and White Thermal Sensor Wires

Quick Fitting Elbow

M4 Nyloc Nut, M5 Washer, & Ground Terminal

Yellow Level Sensor Wire

Equipment Alert To prevent damage to the wires ensure they are routed correctly and secured with zip ties.

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 29: 004-10061-00 -Service Manual: QuickClean Ultrasonic

C-4© Midmark Corporation 2018 004-10061-00 (11/13/18)

step 1: Remove ultrasonic From counter Top.A) Unthread and remove remote panel cable plug from socket at bottom of unit. B) Remove power cord from power connection and hose from drain elbow.C) Locate retaining clips in each corner holding onto under side of counter top, unthread screws almost all the way & pull down on bottom of clips.D) Rotate bottom of retaining clips towards the side of the unit allowing the ultrasonic to clear the inside of the hole in the counter top.E) Push/pull ultrasonic up and out of the counter top.

Remove ultrasonic From counter Top (Recessed)

Clip Bottom

Retaining Clip

Remote Cable Plug

Socket

Power Connection

On/OffSwitch

Drain Valve

Drain Elbow

Hose Connection

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 30: 004-10061-00 -Service Manual: QuickClean Ultrasonic

C-5© Midmark Corporation 2018 004-10061-00 (11/13/18)

Remove base plate Assembly (Recessed)

step 1: Remove base plate Assembly.A) Place front (plug/power switch) side of unit down and remove the screws from bottom plate (number of screws depends on model). B) Separate bottom plate from ultrasonic and rotate top of plate down and away from unit. This allows you to place the bottom plate on the table without disconnecting wires.

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 31: 004-10061-00 -Service Manual: QuickClean Ultrasonic

C-6© Midmark Corporation 2018 004-10061-00 (11/13/18)

step 1: disconnect Wiring.A) Remove Bottom Plate Assembly (See base plate removal).B) Disconnect red and black power wires at the Ultrasonic Generator PCB by loosening the screws on the connector and removing the wires. C) Disconnect yellow level sensor wire & red and white thermal sensor wires (going to side of tank) from Control PCB using a small screwdriver to push on the tabs while pulling wires from connector.

step 2: Remove/Replace Tank Assembly.A) Remove nyloc nut and washer from the tank stud in each corner.B) Remove the terminal holding the ground wire to the tank stud located in the same corner as the power switch.C) Depress and hold the quick fitting retaining ring (shown in red) toward the fitting to release it’s grip on the metal tube coming from the bottom of the tank. D) Slowly remove the tank from the case. The metal tube extending into the elbow will come out of the elbow fitting as the tank is removed.E) Reassemble in reverse order.F) Replace any zip ties that were cut during disassembly.

Remove Tank Assembly (Recessed)

Ultrasonic Generator PCBYellow Level Sensor Wire

Power Supply PCB

Quick Fitting

Red and White Thermal Sensor Wires

Quick Fitting

Main Power Wires

To Front Panel Asm

M4 Nyloc Nut, M5 Washer, & Ground Terminal

Equipment Alert To prevent damage to the wires ensure they are routed correctly and secured with zip ties.

WARNING To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

Page 32: 004-10061-00 -Service Manual: QuickClean Ultrasonic

D-1© Midmark Corporation 2018 004-10061-00 (06/08/17)

Sect

ion

DWiring diagrams Table Top Unit ................................................... D-2

Recessed Unit ................................................... D-3Recessed Display Panel Assembly ............... D-4

Page 33: 004-10061-00 -Service Manual: QuickClean Ultrasonic

D-2© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00 (11/13/18)

Filter

Qc1-01,Qc3-01,Qc6-01all

Power Switch

Power Connection

Power Supply

Ultrasonic Generator

Control & Display PCB

Touchpad

Temperature Sensor Bottom Plate

Level Sensor

Temperature Sensor Tank

Wiring diagram Table Top

Fuses:10 amp slow blow

Fuses:5 amp slow blow

Fuses:10 amp slow blow

Page 34: 004-10061-00 -Service Manual: QuickClean Ultrasonic

D-3© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00(9/24/18)

Wiring diagram Recessed

Qc3R-01,Qc6R-01all

Power Switch

Filter

Power Supply

Ultrasonic Generator

Temperature Sensor Tank

Temperature Sensor Bottom Plate

Level Sensor

IDCBreakoutPCB

Power Connection

Fuses:10 amp slow blow

Fuses:5 amp slow blow

Fuses:10 amp slow blow

Page 35: 004-10061-00 -Service Manual: QuickClean Ultrasonic

D-4© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00 (11/13/18)

Qc3R-01,Qc6R-01all

LCD Display

To Fill Level Sensor

Brow

nG

reyYellowG

reenP

inkB

lueR

edD

ark Orange

White

Purple

Orange

Black

Base Plate

Con

trol B

oard

Touchpad

From Ultrasonic Generator PCB

IDCBreakoutPCB

Wiring diagram Recessed display

OrangeDark Orange White Black BlueRedBrown Purple

To Tank Temp SensorTo Bottom Plate Temp Sensor

GreyGreenPinkYellow

Page 36: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-1© Midmark Corporation 2018 004-10061-00 (06/08/17)

Sect

ion

EExploded Views & parts Lists

XXX-XXXX-XX

XXX-XXXX-XX

XXX-XXXX-XX

Table Top ...........................................E-2Recessed...........................................E-3

Page 37: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-2© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00 (11/13/18)

Table Top unit

Qc1-01,Qc3-01,Qc6-01all

Base Plate Assembly ...................E-8Aluminium plate with Power Supply and Ultrasonic Generator PCBs mounted to it. Attached to the case by screw(s) and screws holding the rubber feet.

Case ................................................E-6S/S case, power switch, main filter, power inlet and ball valve to drain. All other assem-blies are fitted to the case. The size varies depending on model

Tank Assembly ..............................E-4S/S Tank, Temperature sensor, Liquid Level Sensor, drain outlet and Transducers.

Transducers per model:QC1-01 x 2 TransducerQC3-01 x 4 TransducerQC6-01 x 8 Transducer

The tank is attached to the case by internal nuts, wired to the Base Plate Assembly and Front Panel Assembly. A series of plumbing connects to the drain.

Front panel Assembly ...............E-10ABS moulding that consists of Keypad, LCD Display, Control PCB and Label. Attached to case by internal screws.

Page 38: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-3© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00 (11/13/18)

Recessed unit

Qc3R-01,Qc6R-01all

Base Plate Assembly ...................E-9Aluminium plate with Power Supply, Ultrasonic Generator PCB, And XXX PCB’s mounted to it. Attached to the case by screw(s).

Case .................................................E-7S/S case, power switch, main filter, power inlet and ball valve to drain. All other assem-blies are fitted to the case. The size varies depending on model

Tank Assembly ..............................E-5S/S Tank, Temperature sensor, Liquid Level Sensor, drain outlet and Transducers.

Transducers per model:QC3R-01 x 4 TransducerQC6R-01 x 8 Transducer

The tank is attached to the case by internal nuts, wired to the Base Plate Assembly and Front Panel Assembly. A series of plumbing connects to the drain. Fixing Clips secure the unit to the counter top from the bottom.

Front panel Assembly ...............E-11ABS moulding that consists of Keypad, LCD Display, Control PCB and Cable connecting to the base plate. Attached to cabinet by screws with covers.

Page 39: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-4© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00 (07/11/18)

Qc1-01,Qc3-01,Qc6-01all

9

8

1 002-10107-0010

76

5 4

2 a) 053-10032-00 b) 053-10033-00 c) 053-10034-00

Tank Assembly Table TopItem Description Qty

1 FillLevelSensor(includes 4 thru 10) 1

2 TankLida) QC1-01b)QC3-01c) QC6-01

1

3 TemperatureSensor 1

4 LevelSensorScrew 1

5 LevelSensorInsulator 1

6 LevelSensorNylonWasher 2

7 LevelSensorPlainNut 1

8 LevelSensorHarness 1

9 LevelSensorNylocNut 1

10 LevelSensorPlainWasher 3

3

Page 40: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-5© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00(3/29/19)

Qc3R-01,Qc6R-01all

2 a) 029-10151-00 b) 029-10152-00

Tank Assembly RecessedItem Description Qty

1 FillLevelSensor(includes 6 thru 12) 1

2 TankLid(includes item 3)a) QC3R-01b)QC6R-01

1

3 MidmarkLabel 1

4 RetainingClip 4

5 TemperatureSensor 1

6 LevelSensorScrew 1

7 LevelSensorInsulator 1

8 LevelSensorNylonWasher 2

9 LevelSensorPlainNut 1

10 LevelSensorHarness 1

11 LevelLockNylocNut 1

12 LevelSensorPlainWasher 3

11

10

1 002-10107-0012

9 8

76

4 016-10041-00

061-10180-00 3

5

Page 41: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-6© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00(3/29/19)

Qc1-01thru 602439

case Table TopItem Description Qty

1 QuickFittingElbow 1

2 PVC1/2” 1

3 QuickFittinga) QC1-01 (QC x 3/8” thread)b)QC3-01&QC6-01(QC x 1/2” thread)

1

4 Drain Hose (1/2” ID ) by the inch 48

5 BulkheadFittinga) QC1-01 ( 3/8” )b)QC3-01&QC6-01 (1/2” )

1

6 ShutoffValvea) QC1-01 (3/8” Male/Female)b)QC3-01&QC6-01(1/2” Male/Female)

1

7 Elbowa) QC1-01 (3/8” BSPT x 12 mm barb)b)QC3-01&QC6-01 (1/2” BSPT x 12 mm barb)

1

8 MainPowerSwitch(DPST) 1

9 MainPowerFuse (20mm, 10A (T) A/S, ceramic) 2

10 MainPowerSocket(includes 9) 1

11 MainPowerFiltera) QC1-01 (3 Amp)b)QC3-01&QC6-01(5 amp)

1

5

015-10100-00 10

4 014-0609-00

015-10187-00 9

7 a) 014-10009-00 b) 014-10008-00

6 a) 014-10029-00 b) 014-10028-00

8 015-10194-00

5

3

2

1

11

Qc3-01,Qc6-01all

Page 42: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-6.1© Midmark Corporation 2017 TP204 20-42-FO-00015 Rev A1 C2169

Models:serial Numbers:

Always verify model & serial number 004-10061-00(3/29/19)© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number

Qc1-01602440 to Present

case Table TopItem Description Qty

1 QuickFittingElbow 1

2 PVC1/2” 1

3 QuickFitting (QC x 1/2” thread) 1

4 Drain Hose (1/2” ID ) by the inch 48

5 BulkheadFitting (1/2” ) 1

6 ShutoffValve (1/2” Male/Female) 1

7 Elbow (1/2” BSPT x 12 mm barb) 1

8 MainPowerSwitch(DPST) 1

9 MainPowerFuse (20mm, 10A (T) A/S, ceramic) 2

10 MainPowerSocket(includes 9) 1

11 MainPowerFiltera) QC1-01 (3 Amp)b)QC3-01&QC6-01(5 amp)

1

5

015-10100-00 10

4 014-0609-00

015-10187-00 9

8 015-10194-00

5

3

2

1

11

Qc3-01,Qc6-01all

6 014-10028-00

7 014-10008-00

Page 43: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-7© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00(3/29/19)

Qc3R-01,Qc6R-01all

2

014-0609-00 5

015-10187-00 7

015-10100-00 6

case RecessedItem Description Qty

1 QuickFitting (QC x 1/2” thread) 1

2 ShutoffValve(1/2” Male/Female) 1

3 Elbow (1/2” BSPT x 12 mm barb) 1

4 MainPowerFilter (5 amp) 1

5 Drain Hose (1/2” ID ) by the inch 108

6 MainPowerSocket 1

7 MainPowerFuse(10A (T) A/S,Slow Blo) 1

8 MainPowerSwitch 2

3 a) 014-10008-00

2 a) 014-10028-00

015-10194-00 8

1

4

Page 44: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-8© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00 (1/21/20)

Qc1-01,Qc3-01,Qc6-01all

base plate Assembly Table TopItem Description Qty

1 PowerSupplyPCB 1

2 Screw AR

3 RibbonCable 1

4 PowerSupplyPCBFuses (10A (T) A/S,Slow Blo)

2

5 BasePlateAssembly(includes 6 thru 9)a) QC1-01b)QC3-01c) QC6-01

1

6 TemperatureSensor 2

7 UltrasonicGeneratorPCBFuses (5A (T) A/S,Slow Blo)

2

8 UltrasonicGeneratorPCB 1

9 BasePlate 1

10 BaseHardwareKit(includes 11 thru 13) 1

11 Foot 4

12 Washer AR

13 Screw AR

a) 029-10087-00 5b) 029-10089-00 c) 029-10090-00

2

4 015-10187-00

7 015-10187-01

8

2

002-10181-00 10

11

13

12

96

015-10130-00 1

015-10144-00 3To Display PCB

Page 45: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-9© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00 (1/21/20)

Qc3R-01,Qc6R-01all

base plate Assembly RecessedItem Description Qty

1 IDCBreakoutPCB 1

2 PowerSupplyPCB 1

3 RibbonCable 1

4 PowerSupplyPCBFuses (20mm, 10A (T) A/S, ceramic) 2

5 Screw AR

6 BasePlateAssembly(includes 7 thru 11)a) QC3R-01b)QC6R-01

1

7 TemperatureSensor 1

8 12pinconnector 1

9 UltrasonicGeneratorPCBFuses (20mm, 5 A (T) A/S, ceramic) 2

10 UltrasonicGeneratorPCB 1

11 BasePlate 1

12 Washer AR

13 Screw AR

015-10130-00 2

015-10106-00 1

9 015-10187-01

015-10155-00 8

a) 029-10158-00 6b) 029-10161-00

015-10144-00 3

5

5

5

10

11

12

13

7

4 015-10187-00

Page 46: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-10© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00(3/29/19)

1 a) 002-10177-00 b) 002-10179-00

Qc1-01,Qc3-01,Qc6-01all

8 002-10166-00

511

061-10180-00 2

Front panel Assembly Table TopItem Description Qty

1 FrontPanelAssembly(includes 2 thru 4)a)QC1-01&QC3-01b)QC6-01

1

2 MidmarkLabel 1

3 TouchPad NA

4 FrontPanel NA

5 Screw AR

6 AngledPlateLeft 1

7 AngledPlateRight 1

8 FrontPanelPCBKit(includes 9 & 10) 1

9 DisplayPCB 1

10 ControlPCB 1

11 Screw AR

3

4

11

6

5

7

910

Page 47: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-11© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00(3/29/19)

Qc3R-01,Qc6R-01all

16

Front panel Assembly RecessedItem Description Qty

1 FrontPanelAssembly(includes 3 thru 14) 1

2 FrontPanelKit(includes 4 thru 7) 1

3 Screw&ScrewCoverKit(includes 4 & 5) 1

4 ScrewCover 4

5 Screw 4

6 TouchPad 1

7 Frontcover 1

8 AngledPlateLeft 1

9 AngledPlateRight 1

10 DisplayPCB 1

11 Screw AR

12 Bracket 1

13 ControlCableKit 1

14 ControlPCB 1

15 BackCover 1

16 Screw AR

13 015-10112-00

029-10141-00 1

16

15

8

11

12

12

9

7

11

11

54

6 2 002-10167-00

3 002-10182-00

10

14 002-10312-00

Page 48: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-12© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00 (07/11/18)

1 a) 002-10007-00 b) 002-10008-00 c) 002-10009-00

Qc1-01,Qc3(R)-01,Qc6(R)-01all

basketItem Description Qty

1 Basket (Includes Item 2)a) QC1-01b)QC3(R)-01c) QC6(R)-01

1

2 BasketRubberFootKit(contains4feet) 1

2 002-10185-00

Page 49: 004-10061-00 -Service Manual: QuickClean Ultrasonic

E-13© Midmark Corporation 2018

Models:serial Numbers:

Always verify model & serial number 004-10061-00(3/29/19)

1 a) 9A612001 b) 9A613001 c) 9A614001

Qc1-01,Qc3-01,Qc6-01all

beaker holder AssemblyItem Description Qty

1 BeakerHolderAssembly(includes 2 thru 5)a) QC1-01b)QC3(R)-01c) QC6(R)-01

1

2 Beakerholder 1

3 O-ring AR

4 Beaker AR

5 Lid AR

5 053-1778-00

4 053-1779-00

3 053-1782-00

2

Page 50: 004-10061-00 -Service Manual: QuickClean Ultrasonic

SUBJECT TO CHANGE WITHOUT NOTICE