1© Dr. A. KucherTechnical Service
Lumax 540 (VR-T, DR-T, HF-T, DX)
Thoracic Impedance
CRT
Global Technical Services
08.12.2011
2© Dr. A. KucherTechnical Service
Lumax 540Thoracic Impedance – Possible Parameter Settings
Option 1
Option 2
Option 3
Please note: As long as HM is OFF, there is no access to TI
Option 4
If HM is ON, TI can be enabled ordisabled individually
TI (HM) = ON and TI (statistics) = OFF means that the TI-values will only be transmitted via HM and not be storedInside the ICD.
TI (HM) = ON and TI (statistics) = ON means that the TI-values will be transmitted via HM and storedinside the ICD.(The interrogation time during next FU is significantly longer if TI (statistics) is ON)
√
3© Dr. A. KucherTechnical Service
Lumax 540Thoracic Impedance – Trend Curve via HM
4© Dr. A. KucherTechnical Service
Lumax 540Thoracic Impedance – Trend Curve stored in the ICD