62 D I E U R O P E MARCH 2017 By Dr A Sauter CT Pulmonary Angiography (CTPA) as a sub-mSv modality: dose reduction using new iterative reconstruction algorithms Computed tomography pulmonary angiography (CTPA) is the most com- monly used imaging modality to con- firm/rule-out suspected pulmonary embolism (PE) [1-3]. CTPA is available at all times in most hospitals, can be performed quickly and is relatively cost efficient. With recent improvements in CT technology, high diagnostic accu- racy can be achieved, with sensitivities of up to 92% and specificities of up to 95% [4]. However, CTPA involves a high radiation exposure with a currently estimated average dose of 10.7 mSv [5]. Thus it should be noted that the use of CTPA is limited in pregnant women, as well as in patients with chronic kidney failure and/or iodine allergy. In such patients, the use of CTPA must be con- sidered carefully; it may be that eventu- ally CTPA is not possible in these cases. Over the last few decades, the num- ber of CT scans carried out worldwide increased dramatically : in the United States the number of CT scans grew rapidly from approximately 3 million in 1980 to about 62 million in 2006 [6], This has resulted in a corresponding increase in collective radiation expo- sure, which has been associated with a potentially increased lifetime risk of malignancy and a danger of damage to the gametes, especially with younger or pregnant patients [6, 7]. In addition CTPA there is a substantial radiation exposure to the breast; ionizing radia- tion is an established risk factor for breast cancer. [8]. For all these reasons, methods for reducing radiation dose are highly desirable, particularly for younger patients. Various techniques for dose reduction have been developed and partially implemented in daily clini- cal routine, following the principle of “as low as reasonably achievable” (ALARA) [9], In addition to the use of shielding techniques, reduction of anatomical scan coverage and new methods such as z-axis modulation, the dose can be reduced by lowering tube peak kilo voltage (kVp) or tube current (mA). [10]. Several studies have shown that acceptable image quality of CTPA can be maintained when tube voltage is reduced to 100 kVp or even 80 kVp, resulting in significantly lower radia- tion exposures [11, 15]. However, reduced tube voltage leads to higher image noise, especially in obese patients. To compensate for image noise, higher tube currents are needed, leading to only limited reduction of radiation dose [12]. In order to further decrease radiation dose, additional reduction of tube current must be aimed for. Depend- ing on the individual patient’s habitus, different tube currents are needed to obtain suitable image quality. However, highly reduced tube current can lead to impaired image quality, making missed diagnoses more likely [13, 14]. Advances in processing power and the development of new and faster algorithms have enabled wide-spread clinical use of iterative reconstruction methods. These iterative reconstruc- tion algorithms make improvements in image quality and consequent reduc- tion of radiation dose possible, even in obese patients [15, 17, 18]. Iterative reconstruction has been widely intro- duced into clinical routine over the last five years and first generation iterative algorithms have already led to a sig- nificant reduction of image noise in comparison to filtered back projection (FBP) reconstruction [16]. STUDY DESIGN In this study a new iterative recon- struction algorithm was compared to a first-generation iterative reconstruc- tion algorithm (iDose) and standard filtered back projection (FBP) at differ- ent simulated tube current levels [19]. The Author Dr Andreas Sauter, Laboratory for Advanced Computed Tomography Imaging, Department of Diagnostic and Interventional Radiology, Technische Universität München, Munich, Germany. email: [email protected] This article summarizes a recent evaluation of a new iterative reconstruction algorithm for the detection/ rule-out of pulmonary embolism (PE) in ultra-low dose computed tomography pulmonary angiography (CTPA). The results show that by using a next generation iterative reconstruction algorithm “Iterative Model Reconstruction” (IMR) for the detection of PE, a reduc- tion in dose of up to 75% is possible in CTPA, while maintaining full diagnostic confidence. This would result in a mean effective dose of below 1.0 mSv for CTPA. RADIATION DOSE REDUCTION

Upload: others

Post on 05-Jul-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: - RAdiAtion doSe Reduction...64 DI EUROPE MARCH 2017 2nd - 4th October 2017 Langenbeck-Virchow Haus, Berlin, Germany ICIS 2017 will focus on quantitative and functional oncologic imaging

62 D I E U R O P E MARCH 2017

By Dr A Sauter

CT Pulmonary Angiography (CTPA) as a sub-mSv modality: dose reduction using new iterative reconstruction algorithms

Computed tomography pulmonary angiography (CTPA) is the most com-monly used imaging modality to con-firm/rule-out suspected pulmonary embolism (PE) [1-3]. CTPA is available at all times in most hospitals, can be performed quickly and is relatively cost efficient. With recent improvements in CT technology, high diagnostic accu-racy can be achieved, with sensitivities of up to 92% and specificities of up to 95% [4].

However, CTPA involves a high radiation exposure with a currently estimated average dose of 10.7 mSv [5]. Thus it should be noted that the use of CTPA is limited in pregnant women, as well as in patients with chronic kidney failure and/or iodine allergy. In such patients, the use of CTPA must be con-sidered carefully; it may be that eventu-ally CTPA is not possible in these cases.

Over the last few decades, the num-ber of CT scans carried out worldwide increased dramatically : in the United States the number of CT scans grew

rapidly from approximately 3 million in 1980 to about 62 million in 2006 [6], This has resulted in a corresponding increase in collective radiation expo-sure, which has been associated with a potentially increased lifetime risk of malignancy and a danger of damage to the gametes, especially with younger or pregnant patients [6, 7]. In addition CTPA there is a substantial radiation exposure to the breast; ionizing radia-tion is an established risk factor for breast cancer. [8].

For all these reasons, methods for reducing radiation dose are highly desirable, particularly for younger patients. Various techniques for dose reduction have been developed and partially implemented in daily clini-cal routine, following the principle of “as low as reasonably achievable” (ALARA) [9], In addition to the use of shielding techniques, reduction of anatomical scan coverage and new methods such

as z-axis modulation, the dose can be reduced by lowering tube peak kilo voltage (kVp) or tube current (mA). [10]. Several studies have shown that acceptable image quality of CTPA can be maintained when tube voltage is reduced to 100 kVp or even 80 kVp, resulting in significantly lower radia-tion exposures [11, 15]. However, reduced tube voltage leads to higher image noise, especially in obese patients. To compensate for image noise, higher tube currents are needed, leading to only limited reduction of radiation dose [12].

In order to further decrease radiation dose, additional reduction of tube current must be aimed for. Depend-ing on the individual patient’s habitus, different tube currents are needed to obtain suitable image quality. However, highly reduced tube current can lead to impaired image quality, making missed diagnoses more likely [13, 14].

Advances in processing power and the development of new and faster algorithms have enabled wide-spread clinical use of iterative reconstruction methods. These iterative reconstruc-tion algorithms make improvements in image quality and consequent reduc-tion of radiation dose possible, even in obese patients [15, 17, 18]. Iterative reconstruction has been widely intro-duced into clinical routine over the last five years and first generation iterative algorithms have already led to a sig-nificant reduction of image noise in comparison to filtered back projection (FBP) reconstruction [16].

Study deSignIn this study a new iterative recon-struction algorithm was compared to a first-generation iterative reconstruc-tion algorithm (iDose) and standard filtered back projection (FBP) at differ-ent simulated tube current levels [19].

the Authordr Andreas Sauter,

Laboratory for Advanced Computed Tomography Imaging,

Department of Diagnostic and Interventional Radiology, Technische Universität München, Munich, Germany. email: [email protected]

This article summarizes a recent evaluation of a new iterative reconstruction algorithm for the detection/rule-out of pulmonary embolism (PE) in ultra-low dose computed tomography pulmonary angiography (CTPA).

The results show that by using a next generation iterative reconstruction algorithm “Iterative Model Reconstruction” (IMR) for the detection of PE, a reduc-tion in dose of up to 75% is possible in CTPA, while maintaining full diagnostic confidence. This would result in a mean effective dose of below 1.0 mSv for CTPA.

RAdiAtion doSe Reduction

Page 2: - RAdiAtion doSe Reduction...64 DI EUROPE MARCH 2017 2nd - 4th October 2017 Langenbeck-Virchow Haus, Berlin, Germany ICIS 2017 will focus on quantitative and functional oncologic imaging

MARCH 2017 D I E U R O P E 63

The objective of the study was to evaluate the diagnostic performance of a new gen-eration iterative reconstruction algorithm (IMR) in low and ultra-low dose CTPA with a particular focus on image quality and diagnostic confidence in the diagnosis of PE.Sixteen patients with PE were enrolled into this single centre study. PE was sub-classified as central, segmental, or subseg-mental. Multiple localizations were also possible. All patients were examined using a 256-slice multidetector CT (Brilliance iCT; Philips Healthcare) using a standard CTPA protocol.The CTPA raw data were then used to simulate CTPA scans with reduced tube currents, resulting in dose levels (DL) of 50%, 25%, 12.5%, 6.3% and 3.1% of the original dose. All other imaging param-eters remained identical. All CTPA images so obtained (i.e. tube current levels of 100% - 3.1%) were then reconstructed with FBP, iDose, and IMR, resulting in a total of 288 datasets (6 different levels of tube current, 3 different algorithms, 16 patients). Each data set was independently evaluated by two blinded radiologists according to the following criteria: image noise, classic arte-facts, ring artefacts, image quality, diagnos-tic confidence in detection of central and peripheral PE as well as overall diagnostic confidence and artificial image appearance.The radiologists carrying out the rating were asked to evaluate central, segmen-tal and subsegmental pulmonary arteries in every data set as positive or negative for PE. PEs in multiple localizations were possible. As there is no gold standard for the detection of PE, the clinical report, together with a re-evaluation by the author of the clinically available images were used to determine the correctness of ratings. In order to measure image quality objec-tively, two contrast-to-noise ratios (CNR) were determined: one for the main pul-monary artery as a central vessel (cen-tral contrast to noise radio – cCNR) and one for a segmental artery as a peripheral vessel (peripheral contrast to noise ratio - pCNR).

ReSultS: The mean effective dose for all examina-tions was 3.6 ± 2.9 mSv.

For images reconstructed with iDose and IMR, the median subjective image quality was higher than that of images reconstructed with FBP at all dose levels from 100% to 12.5% DL. At a DL of 3.1%, median image quality was 1 (i.e. unac-ceptable for diagnostic purposes) for all algorithms. The median image quality was higher after reconstruction with IMR than after reconstruction with iDose at DL of 6.3% and 25%.Using iDose and IMR reconstruction, sig-nificantly higher diagnostic confidence could be obtained than with FBP recon-struction (p < 0.01). This conclusion was true for all divisions of PE (central, periph-eral, overall) and to all dose levels.At lower dose levels, IMR-images pro-duced a higher median diagnostic confi-dence than iDose images (higher median in IMR-images than in iDose-images at 12.5% DL for all divisions of PE and at 6.3% DL for central and peripheral PE).In IMR-images the median diagnostic confidence in the detection of PE was 4 (completely confident) for central PE at DL of 25%, 50% and 100%.On the basis of the full-dose images, no PE (central, segmental, and subsegmental) was missed by either rating radiologist in IMR-reconstructed images at dose levels

of 12.5% to 100%, resulting in a sensitivity of 100%. The same conclusion was found with iDose at dose levels of 25%-100%. With FBP reconstruction, only at 100% DL was every PE identified. With the use of FBP-reconstruction at 12.5% DL, only 35% of all central, 23% of all segmental and 19% of all subsegmental PEs could be identified. At all dose levels, cCNR and pCNR were significantly better in IMR-images than in iDose and FBP-images (p < 0.01). In addition, cCNR in IMR-images at DLs of 50%, 25% and 12.5% was significantly bet-ter than in iDose-images at 100% DL (p < 0.05).

SummARy And outlook:Over the last few years, it has been shown that the use of iterative reconstruction algorithms has made possible significant dose reduction via reduced tube-current and tube-voltage [13]. However, even further reduction of radiation exposure is desirable particularly as CTPA is fre-quently used in clinical routine. As a reduction of tube voltage below 80 kVp is not reasonable due to the k-edge of iodine, further reduction of radiation dose has to be obtained by reducing tube current. The results of the present study show

Figure 1. Comparison of FBP, iDose and IMR at dose levels of 3.1-100%Comparison of FBP (top), iDose (middle) and IMR (bottom) at different dose levels (from left to right: 100% - 50% - 25% - 12.5% - 6.3% - 3.1%). Emboli are located in the right pulmonary artery, the left upper and lower lobar artery and in several segmental arteries of the left lung. Note the good detectability even at ultra-low dose levels with IMR. With iDose emboli are also detectable but image quality is inferior to IMR. With FBP, emboli are not certainly detectable at lower dose levels.

Page 3: - RAdiAtion doSe Reduction...64 DI EUROPE MARCH 2017 2nd - 4th October 2017 Langenbeck-Virchow Haus, Berlin, Germany ICIS 2017 will focus on quantitative and functional oncologic imaging

64 D I E U R O P E MARCH 2017

2nd - 4th October 2017Langenbeck-Virchow Haus, Berlin, GermanyICIS 2017 will focus on quantitative and functional oncologic imaging and the multidisciplinary management of oncology patients. There will be a spectrum of sessions ranging from innovative and state-of-the-art imaging, practical hands-on-workshops, essential refresher courses and live-case workshops.

For further information please contact:ICIS, 140-148 Borough High Street, London, SE1 1LB E: [email protected]

cancerimagingsociety.org.uk

Delegates will received 17 CPD points from the Royal College of Radiologists (Monday - 6, Tuesday - 6, Wednesday - 5). CME accreditation will be sought from the European Accreditation Council Continuing Medical Education (EACCME).

Programme planning committee: Stefan Diederich ICIS President (DE), Bernhard Gebauer (DE), Dow-Mu Koh (UK), Beth McCarville (US), Anwar Padhani (UK), Andrea Rockall (UK), Heinz-Peter Schlemmer (DE), Harriet Thoeny (CH).

Call for abstracts: Deadline 29th May 2017Submit online at icimagingsociety.org.uk

In association with:

• 2 Keynote lectures ‘Cancer Imaging in the era of precision medicine’ and ‘Immunotherapy: Imaging challenges’

• 8 Computer hands-on workshops using Siemens workstations on evaluation of cancer of prostate, breast, lung, liver, lung, ovary, uterus and whole-body imaging

• 25 sessions addressing structured reporting, response evaluation, complications of medical and radiation therapy, metastatic disease, joint sessions with radiation and medical oncologists, precision medicine, radiogenomics, prostate, female pelvis, liver, pancreas, lung, lymphoma, CEUS and many more

• 4 Scientific Sessions dedicated to proffered papers, and a poster exhibition with prizes for the best overall paper and poster

• 4 live-case workshops Transarterial chemoembolization (TACE) in HCC / liver metastases, Radioembolization (SIRT), Microwave ablation of a pulmonary tumour in metastasis or small lung cancer, Radiofrequency ablation of liver metastasis

International Cancer Imaging Society Meeting & 17th Annual Teaching Course

Cancer Imaging and Intervention: a Multidisciplinary Approach

Call for abstracts: Deadline 29th May 2017Submit online at icimagingsociety.org.uk

that it is possible to dramatically reduce tube current using new iterative recon-struction algorithms. In our study, IMR, a second generation iterative reconstruction algorithm, provided excellent diagnostic confidence in the detection of PE even in ultra-low-dose images acquired with a simulated mean effective dose of only 0.9 mSv. However, according to the ALARA principle, dose should not be lowered to the lowest possible value but only to the value at which a very high diagnostic con-fidence is guaranteed. We suggest a dose reduction of no greater than 75% since, down to this dose level the diagnostic confidence in the detection of peripheral PE using images reconstructed with IMR could be rated as probably con-fident or even higher. Although in our study no PE was missed down to a dose level of 12.5% (0.45 mSv), we neverthe-less consider that the dose should not be reduced as far as this in order to be abso-lutely sure of maintaining image quality and diagnostic confidence.Over the last few years, it has been shown that the radiation dose per CT scan could be dramatically reduced through the appli-cation of more powerful iterative recon-struction algorithms. However, the biggest achievement in dose reduction was through

the replacement of FBP by the first genera-tion of iterative reconstruction algorithms. In the future, it can be foreseen that further radiation dose reduction can be achieved by the combination of new technologies such as spectral photoncounting CT [20] or sparse sampling CT in combination with iterative reconstruction algorithms.

concluSionUsing next generation iterative recon-struction algorithms such as IMR, the radiation dose for CTPA examinations can be lowered dramatically, while provid-ing suitable image quality and diagnostic confidence even in sub-mSv images.

ReFeRenceS:1. Mayo J, Thakur Y. Pulmonary CT angiography as first-

line imaging for PE: image quality and radiation dose considerations. AJR American journal of roentgenol-ogy. 2013; 200(3): 522-8. .

2. Schoepf UJ, Costello P. CT angiography for diagnosis of pulmonary embolism: state of the art. Radiology. 2004; 230(2): 329-37. .

3. Remy-Jardin M, Pistolesi M, Goodman LR, Gefter WB, Gottschalk A, Mayo JR, et al. Management of sus-pected acute pulmonary embolism in the era of CT angiography: a statement from the Fleischner Society. Radiology. 2007; 245(2): 315-29.

4. Stein PD, Fowler SE, Goodman LR, Gottschalk A, Hales CA, Hull RD, et al. Multidetector computed tomography for acute pulmonary embolism. The New England journal of medicine. 2006; 354(22): 2317-27.

5. Takahashi EA, Yoon H-C. Four-Year Cumulative

Radiation Exposure in Patients Undergoing Computed Tomography Angiography for Suspected Pulmonary Embolism. Radiology Research and Practice. 2013; 2013: 5.

6. Brenner DJ, Hall EJ. Computed tomography--an increas-ing source of radiation exposure. The New England journal of medicine. 2007; 357(22): 2277-84. .

7. Parker L. Does radiation associated with diagnostic CT scans in children and young people increase their subsequent risk of developing cancer? Arch Dis Child Educ Pract Ed. 2013; 98(6): 239.

8. Hill DA. et al. Medical radiation, family history of can-cer, and benign breast disease in relation to breast cancer risk in young women, USA. Cancer Causes Control, 2002; 13(8): 711.

9. Laqmani A, Buhk JH, Henes FO, Klink T, Sehner S, von Schultzendorff HC, et al. Impact of a 4th generation iterative reconstruction technique on image quality in low-dose computed tomography of the chest in immu-nocompromised patients. RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2013; 185(8): 749-57.

10. Kalra MK, Rizzo S, Maher MM, Halpern EF, Toth TL, Shepard JA, et al. Chest CT performed with z-axis modulation: scanning protocol and radiation dose. Radiology. 2005; 37(1): 303-8.

11. Szucs-Farkas Z, Megyeri B, Christe A, Vock P, Heverhagen JT, Schindera ST. Prospective ran-domised comparison of diagnostic confidence and image quality with normal-dose and low-dose CT pulmonary angiography at various body weights. European rRadiology. 2014;24(8):1868-77.

12 . Marin D, Nelson RC, Schindera ST, Richard S, Youngblood RS, Yoshizumi TT, et al. Low-tube-voltage, high-tube-current multidetector abdomi-nal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm--initial clinical experience. Radiology. 2010; 254(1): 145-53.

13. Henes FO, Groth M, Begemann PG, Bley T, Adam G, Regier M. Impact of tube current-time and tube volt-age reduction in 64-detector-row computed tomogra-phy pulmonary angiography for pulmonary embolism in a porcine model. Journal of thoracic imaging. 2012; 27(2): 113-20.

14. MacKenzie JD, Nazario-Larrieu J, Cai T, Ledbetter MS, Duran-Mendicuti MA, Judy PF, et al. Reduced-dose CT: effect on reader evaluation in detection of pulmonary embolism. AJR American journal of roentgenology. 2007; 189(6): 1371-9.

15. Noël PB, Köhler T, Fingerle AA, Brown KM, Zabic S, Münzel D, et al. Evaluation of an iterative model–based reconstruction algorithm for low-tube-voltage (80 kVp) computed tomography angiography. Journal of Medical Imaging. 2014; 1(3): 033501.

16. Noel PB, Fingerle AA, Renger B, Munzel D, Rummeny EJ, Dobritz M. Initial performance characterization of a clinical noise-suppressing reconstruction algo-rithm for MDCT. AJR American journal of roentgenol-ogy. 2011; 197(6): 1404-9.

17 Willemink MJ, Takx RA, de Jong PA, Budde RP, Bleys RL, Das M, et al. Computed tomography radiation dose reduction: effect of different iterative recon-struction algorithms on image quality. J Comput Assist Tomogr. 2014; 38(6): 815-23.

18. Noel PB et al. Initial performance characteriza-tion of a clinical noise-suppressing reconstruction algorithm for MDCT. AJR Am J Roentgenol. 2011; 197(6): 1404.

19. Sauter A, Koehler T, Fingerle AA, Brendel B, Richter V, Rasper M, et al. Ultra Low Dose CT Pulmonary Angiography with Iterative Reconstruction. PLOS

ONE. 2016; 11(9):e0162716. 20. Muenzel D. et al. Spectral Photon-counting

CT: Initial Experience with Dual-Contrast Agent K-Edge Colonography. Radiology, 2016 Dec. : p. 160890.

Figure 2. Overall diagnostic confidence Medians of overall diagnostic confidence in detection of PE shown as box-and-whisker plots at different dose levels (3.1 – 100% of the original dose level). IMR performs significantly better, providing high diagnostic confidence also at 25%, 12.5% and 6.3% of the original radiation dose.* = significance compared to iDose at the corresponding dose level (* = p<0.05, ** = p<0.01)# = significance compared to FBP at the corresponding dose level (p<0.01) Values are shown as the mean of the medians of both raters.

Radiation dose Reduction